CHAUVIN ARNOUX C.A 742 IP2X Voltage Detector User Manual

June 12, 2024
CHAUVIN ARNOUX

CHAUVIN ARNOUX C.A 742 IP2X Voltage Detector User Manual

Thank you for purchasing a CA 742 Voltage Detector.
For the best results from your instrument:

  • read this operating manual carefully,
  • conform to the precautions for use.

WARNING, danger! The operator must refer to these instructions when this danger symbol appears.
Equipment protected by double insulation.
Equipment suitable for live work.
Battery.
Earth.
The CE marking indicates compliance with the European Low Voltage Directive (2014/35/EU), Electromagnetic Compatibility Directive (2014/30/EU), and Restriction of Hazardous Substances Directive (RoHS, 2011/65/EU and 2015/863/ EU).
The UKCA marking certifies that the product complies with the requirements that apply in Great Britain, especially regarding low-vol.tage safety, electromagnetic compatibility, and the restriction of hazardous substances.
The trash can with a line through it indicates that, in the European Union, the product must be subject to selective disposal in accordance with directives WEEE 2012/19/EU. This equipment must not be treated as household waste.

Definition of measurement categories

  • Measurement category IV corresponds to measurements at the source of low-voltage installations. Example: power supply, counters and protection devices.

  • Measurement category III corresponds to measurements in building installations.
    Example: distribution panel, circuit breakers, machines or fixed industrial devices.

  • Measurement category II corresponds to measurements taken on circuits directly connected to low-voltage installations. Example: Power supply for home electrical appliances and portable tools.

PRECAUTIONS FOR ASAZ

This device is protected against voltages up to 600V with respect to earth in measurement category IV.
The protection of the device can be compromised if it is used differently than specified by the manufacturer and thus put the user at risk.

  • Do not exceed the maximum rated voltage and current in the measurement category. Do not use your instrument on networks whose voltage or category exceeds that specified.
  • Respect the conditions of use, namely the temperature, humidity, altitude, degree of pollution and the place of use.
  • When handling the test samples, keep your fingers behind the physical guard.
  • Use connection accessories whose measurement category and service voltage are at least equal to that of the installation being measured.
  • Do not use the device if it is open, damaged or poorly disassembled, or its accessories if they appear damaged.
  • The device must be kept clean so that the condition of the cable insulators, housings and accessories can be checked. Any component whose insulator is damaged (even partially) must be sent for repair or scrapped.
  • The device is designed to be used by qualified personnel and in accordance with national safety regulations.
  • We recommend wearing personal protective equipment if the environment in which the device is used makes it necessary.
  • All troubleshooting and metrological checks must be performed by competent, accredited personnel.

PRESENTATION

Delivery condition
CA 742 voltage detector
Delivered in blister pack with:

  • a red test probe 2 mm in diameter,
  • a black lead terminated by a test probe 2 mm in diameter,
  • a band,
  • where AAA or LR3 alkaline batteries,
  • a manual in 5 languages,
  • a verification certificate.

Accessories and options

  • A red test probe 2 mm in diameter
  • Black lead with test probe 2 mm diameter
  • IP2X accessories
  • Carrying bag
  • CA 751 2P+T adapter

For accessories and spare parts, visit our website: www.chauvin- arnoux.com

FUNCTIONS
The CA 742 is a voltage detector with lights.
It meets the requirements of standard IEC 61243-3.
The main function of the CA 742 is for the absence of voltage. It detects dangerous voltages, meaning voltages exceeding the ELV (extra low voltage: 50 Vac or 120 VDC).
Its other functions are:

  • Indication of a voltage between 12 and 690 Vac or 750 Vdc, with indication of polarity.
  • Continuity indication.
  • Indication of the phase position.

The voltages indicated on the CA 742 are nominal voltages. Make sure that the device is used on networks on standardized volume

CA 742

ON THE BACK
When the device is not in use, the test probes can be stored on the back.

USED

This device is a detector. The indications it provides must not be used for measurement purposes.

SELF-TEST
Before using the CA 742, perform a self-test. This checks the integrity of the leads, the correct operation of the electronic circuit, and the battery voltage.
Connect the red test probe to the + terminal and the black lead to the COM terminal.
Touch the 2 test samples together and press the AUTO TEST button. Press as long as necessary.

  • If all the indicators of the device except ELV light and the acoustic signal sound, the device is working properly and can be used.

  • If all other lights are on, the batteries must be replaced (see §4.2).

  • If every third indicator is off, the leads and test samples must be checked. Check that they are properly connected and that the test samples are actually touching, then press the AUTO TEST button again.
    If the problem persists, replace the batteries and try again. If the problem persists with new batteries, the lead and/or test samples must be replaced.

  • If no lights are on, replace the batteries (see §4.2). If the problem persists with new batteries, the device is defective and must be sent for repair.

After each check, perform a self-test to confirm correct operation of the device.
In a noisy environment, make sure you can actually hear the device’s audio signal.
Note:  If the AUTO TEST button is pressed for more than 10 seconds, if the test samples do not touch each other, the device will switch to standby.

VOLTAGE DETECTION
Connect the red test probe to the + terminal and the black lead to the COM terminal.
Keep your hands behind the guards of the device and the test probe.

Place the test samples on the item being tested and maintain firm contact.
There is no need to switch on the CA 742, because it starts up automatically.

If the voltage is present:

  • AC the indicators light to indicate its value and the + (green) and – (orange) indicators light.

  • DC, the indicators light to indicate its value and the + indicator (green) or the – indicator (orange) lights to indicate the polarity.

  • dangerous ( >50 Vac or 120 Vdc): the higher the voltage, the faster the ELV indicator (red) flashes; the device also sends audible beeps.

  • ELV:  Extra Low Voltage. This redundant indicator shows that the voltage is greater than the ELV.

The first two indicators of the bar graph are green to indicate that the voltage is not dangerous and the device does not beep. The remaining indicators are red and the device beeps.
If only the ELV light is on, the batteries are low or missing.

If there is a disturbing voltage in the vicinity of the element being tested, the device can indicate the presence of an operating volume on the element being tested.

Continuity Detection
Connect the red test probe to the + terminal and the black lead to the COM terminal.
Keep your hands behind the guards of the device and the test probe.
Place the test samples on the item being tested and maintain firm contact.

Keep the button pressed.

If no voltage is detected, the CA 742 performs a continuity test.

  • If the resistance is below the continuity limit, indicators 12 to 230 will flash one after the other. The device emits a continuous audible signal.

  • If the resistance is above the continuity threshold, the device displays nothing and emits no sound.

PHASE DETECTION
The CA 742 performs unipolar phase detection. This means that connecting a single test probe is sufficient to determine if a phase is present.
Attention:  phase detection cannot replace a lack of voltage test.
To operate properly, the phase detection function must be used on networks that are referred to earth.
It can be used, zBample, to locate the phase at an outlet.
Connect the red test probe to the + terminal
Keep your hands behind the guards of the device
Place the test samples on the element being tested and maintain a firm contact.

If the test probe is actually in phase, the Ph (phase) indicator will flash and the device will beep.

Attention:  if the Ph indicator does not flash, this does not mean that there is no dangerous voltage at the outlet.

USE CA 751 (OPTION)
If you bought a CA 751 2P+T adapter, you can test for the absence of voltage between the phase and neutral on an outlet.
Note:  pairing the CA 742 with the CA 751 derates the combined product to measurement category II, 250V.

Connect the CA 751 to the terminals of the CA 742, then refer to the operating instructions of the CA.751.

characteristics

Reference conditions

Amount of influence Reference values
Temperature 23 ± 5 °C
Relative humidity 30 to 75% RH
Supply voltage 3 ± 0.1V
Frequency of the measured signal DC or 45 to 65 Hz
Type of signal sinusoidal
External electric field <1V/m
DC external magnetic field <40A/m

Electrical characteristics

VOLTAGE

  • Nominal voltage is:  12, 24, 50, 127, 230, 400, 690 Vac/Vdc and 750 Vdc.
  • Operating frequency:  DC and 16.67 at 800Hz.
  • Maximum input current:  3.5 mArm.
  • Input impedance > 300 kΩ.
  • Response time  500 ms.

The indicator corresponds to voltage V lights before the voltage reaches 85% V.
If no light lights, the voltage present is <12V.
The CA 742 must only be used on networks in the standardized voltagas
operating cycle:  30s (maximum duration for which the device can remain connected to a live element) – 240s (minimum rest time during which the detector is not connected to a live element must be).

CONTINUITY
Continuity detection is disabled when a voltage > 1V is present. Trigger threshold between 100Ω and 150Ω.
Test current ≤1mA
Open-circuit voltage ≤3.3V

PHASE IDENTIFICATION
120 Wattage<690 Vac and 45Hz≤Frequency ≤65Hz
400 Wattage<690 Vac and 16.67Hz≤Frequency <45Hz

ENVIRONMENTAL CONDITIONS

The device is of type N. It must be used under the following conditions:
% RH

  1. Reference domain
  2. Operating room
    -15 to +45°C and 20 to 95% RH without condensation. (max. 35°C at 95% RH)
  3. Range in storage (without batteries)
    -40 to +70°C and 20 to 95% RH without condensation.
    If an extended period of non-use is anticipated, or for storage, withdraw the batteries from the housing.
    For indoor and outdoor use.
    Pollution degree: 2.
    Altitude:  <2000 m.

POWER
The CA 742 is powered by two 1.5V alkaline batteries (type AAA or LR3).
Battery mass: about 2 x 26 g
The battery life is 7,000 10 second measurements.
The batteries can be replaced with rechargeable batteries, but the life between charges will be much shorter.

CHARACTERISTICS OF CONSTRUCTION

  • Measures (LxWxH)  163 x 64 x 40mm

  • Weight  about 210g

  • Gold wire  length of 90cm

  • Protection class  IP65 according to IEC 60529
    IK04 according to IEC 62262

  • Drop  2 meters

COMPLIANCE WITH INTERNATIONAL STANDARDS
Two-pole voltage detector according to IEC 61243-3 ed. 3 of 2015.
The device complies with IEC/EN 61010-1 or BS EN 61010-1 600V, CAT IV.

ELECTROMAGNETIC COMPATIBILITY
Emissions and immunity in an industrial setting conforming to IEC/EN 61326-1 or BS EN 61326-1.

MAINTENANCE

Apart from the batteries, the instrument contains no parts that can be replaced by personnel who have not been specially trained and accredited. Any unauthorized repair or replacement of a part by an “equivalent” can seriously damage safety.

Cleaning
The device must be kept perfectly clean.
Disconnect the instrument completely.
Use a soft cloth, dampened with soapy water. Rinse with cloth and dry quickly with a dry cloth or forced air. Do not use alcohol, solvents or hydrocarbons.

REPLACEMENT OF BATTERIES
If during the self-test only half of the indicators light up, you need to replace the batteries.

  • Disconnect everything connected to the device.
  • Using a screwdriver, unscrew the two captive screws from the battery compartment cover on the back of the device.
  • Take back the used batteries and replace them with two new batteries (AAA or LR3 1.5V alkaline batteries).
  • Close the battery compartment cover and make sure it is completely and correctly closed.
  • Screw the two screws back.
    Used batteries must not be treated as normal household waste. Take them to the appropriate recycling collection point.

WARRANTY

Unless otherwise stated, our warranty is valid for 24 months from the date the equipment was sold. The extract from our general conditions of sale is available on our website.
www.chauvin-arnoux.com/en/general-terms-of-sale
The guarantee does not apply in the following cases:

  • Inappropriate use of the equipment or use with incompatible equipment;
  • Changes made to the equipment without the express permission of the technical staff of the manufacturer;
  • Work on the device performed by a person not approved by the manufacturer;
  • Adaptation to a particular application not foreseen in the definition of the equipment or not specified in the user manual;
  • Damage caused by shock, falls or floods.

FRANCE
Chauvin Arnoux
Phone:  + 33 1 44 85 44 85
Fax:  + 33 1 46 27 73 89
info@chauvin-arnoux.com
www.chauvin-arnoux.com

INTERNATIONAL
Chauvin Arnoux
Phone:  + 33 1 44 85 44 38
Fax:  + 33 1 46 27 95 69
Our international contacts www.chauvin-arnoux.com/contacts

Read More About This Manual & Download PDF:

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

CHAUVIN ARNOUX User Manuals

Related Manuals