Precautions for Use
Use batteries
- Please use right original AC/DC adapter to charge; Do not keep the device in charging status when not use; and be sure fully charge the device if re-use after long-term placed.
- Keep the device away from water, fire or high-temperature environment.
- Do not dissemble, open or destroy the inside lithium battery, or set it on fire, which may lead to explosion or release toxic substances; Please keep the device away from strong shock or vibration.
- Please obey local laws and regulations to solve the discarded products.
Avoiding condensation problems
As much as possible, avoid sudden temperature changes. Do not attempt to use
the device immediately after moving it from a cold to a warm location, or
raising the room temperature suddenly, as condensation may form with in the
device.
Description
On/Off and Auto-off
Short press” “to turn on the device with auto-off function· ‘ to cancel the function ‘ long press until “appears After device on short press to turn on/off backlight,long press to turn off the device.
Optical power meter and Offset(self-calibration)Setting
Turn on the device and enter OPM interface, short press key to switch unit and
check REF value; long press to set current value as new REF value. Press key
to select the wavelength; long press to open WAVE ID function.
Long press and key at the same time to ON/OFF Wavehmgth turn on the device and
enter the Offset setting interface; press up/down key to set,the range is
[-5,+ 5]; then press key to confirm the setting.
Data Storge
Long press” ” to save the current OPM value.Max 500 data storage;the initial
data will be covered after more than 500 data are saved.Short press to check
the saved data, switch pages by” “and ” “. VFL ~
Press” ” again to exit data storage.
VFL/Cable Test and LED Lighting
Short press to switch VFL output mode. Long press to turn on/off cable test
mode .
Short press to turn on/off LED lighting.
Cable Test Module
Press the module lock “press” key to pull out the module to test network cable, and push it back after testing;then make sure the module is locked tightly.Long press”\ >~1f I” to open the network test.and the measurement will start auto matically after the network cable is connected.
Power Indicator
Four levels indication of power detection
Remain 80%—100%
Remain 40%—80%
Remain 20%—40%
Remain less than 20%
Charge
The device has a charging function. When the power is less than 20%, you
should promptly shut down it and recharge. Long time undervoltage will shorten
the lifetime of the rechargeable battery.
Insert Micro USB cable to 5V/1 A charging adapter, power bank or computer to
charge the device, and charging time should be within 24 hours.
Meter Maintenance
General maintenance
Optical fiber connect the adapter should avoid contacting with hard objects
and keep clean.Should be stored in a dry and ventilated place to avoid
moisture.
When unuse for long time,charge the battery every 3 months ,and charge 50%~80%
of the power.
Troubleshooting
Failure | Failure Cause | Solution |
---|---|---|
Cannot boot | Battery is dead | Charge the battery |
Immediately shutdown after booting | Low battery | Charge the battery |
The operation is invalid after starting up | The instrument program is | |
disordered |
Reboot
Large measurement error| The end face of the light port is dirty| Clean the light port
Specification
MT-7615 | MT-7616 | |
---|---|---|
Detector Type | lnGaAs | |
Wavelength Response Range | 700-1700nm | |
OPM Connector | 2.5mm universal connector | |
Measurement Wavelengths | 850/980/1270/1300/1310/1490/1550/1577/1625/1650nm | |
Measurement Range | -70~+10dBm | |
Resolution | 0.01dB |
Accuracy*
| ±0.2dB (131011550nm)
±0.3dB(850/980/1300/1490/1625/1650nm)
± 1dB (127011577nm)
Unit| dB,dBm, mW, uW, nW
Modulation Detection| 270/1k/2kHz
Data Storage| 500 groups
Wavelength Memory| Yes
WAVE ID| Yes(Only for Prokits’ light source products)
User Self-calibration| Yes
REF| Yes
Back Light| Yes
Output Power| 15mW| 30mW
Testing Distance*| 10-12km @SM| 14-16km@SM
VFL Connector| 2.5mm universal connector
Wavelength| 650±20nm
Operation Mode| CW/2Hz Glint
Network Application| UTP network cable
Interface| RJ45 (UTP/8P8C)
Testing Type| Network sequence
---|---
Testing Distance| 600m
Illumination| LED
Auto Power-off| Yes(no operation in 10 minutes)
Low Battery Indicator| Battery level
Charging Socket| Micro USB
Working Time| >50 hours(only OPM)
Power Source
| 3.7V/450mAh
Li-ion
Operation Temperature| -10~50’C
<90%RH
Size| 985927mm
Weight| 100g
• at 20″C ±2″C, 40%-60%RH, with standard testing fiber
• Accuracy range, +3dBm–60dBm, Others as following:
±0.BdB, +3dBm-+6dBm, -60dBm–65dBm
±3.0dB, +6dBm-+1 0dBm, -65dBm–70dBm