AKO-58110 Calibration kit Installation Guide
- June 12, 2024
- AKO
Table of Contents
Calibration kit
AKO-58110
5812H002 Ed.03
Installation Guide
Installation
The AKO-58110 calibration kit enables the testing, verification of accuracy, reset to zero and calibration of the transmitter.
|
---|---
Install the kit, attaching it to the transmitter window as shown in the
image.| Press hard until it fits together perfectly.
|
Connect the regulator of the gas pump to the free end of the transparent
tube.| To extract it, pull out from the lower end.
Operation test (Bump Test)
Materials required: AKO-58110 calibration kit + Gas cylinder*
EN-378 and F-GAS international standards require verification of the correct
operation of the transmitter at least once per year. Please check what current
local regulations specify for such cases. Always ensure that you comply with
current local regulations.
- Before starting the verification procedure, check the Pre-Alarm configuration ( ). This should be enabled in order to check it has been activated. It is advisable to disable the Pre Alarm delay ( ) and Alarm delay ( ) to speed up the verification process.
- Install the AKO-58110 calibration kit.
- Open the shut-off valve of the gas cylinder and wait.
- After a few moments, the transmitter concentration reading will begin to rise until it reaches the Pre-Alarm and then Alarm level.
- Check that both the Pre-Alarm and the Alarm signal correctly and that the corresponding relays are activated.
- Close the shut-off valve of the gas cylinder, disconnect the cylinder and remove the calibration kit from the transmitter.
*** The transmitter may take a few moments to go back to showing the
concentration values from before the verification procedure. Once the
verification procedure is complete, remember to readjust the Pre-Alarm ( ) and
delay ( ) parameters to the values prior to the beginning of the verification
procedure.
Use a cylinder with a type of gas that is suitable to the sensitivity of the transmitter to be calibrated. When calibrating the AKO-575400 universal gas transmitter, use a R134a gas cylinder.
Verification of accuracy (AKO-575xxx only)
Materials required: AKO-58110 calibration kit + Calibrated gas cylinder*
AKO recommends verifying the accuracy of the transmitter at least once per
year. Please check what current local regulations specify for such cases.
Always ensure that you comply with current local regulations.
- Install the AKO-58110 calibration kit.
- Open the shut-off valve of the gas cylinder and wait.
- After a few moments, the transmitter concentration reading will begin to rise. Wait for the reading to stabilize.
- Compare the reading displayed with the calibrated value of the gas cylinder. If the accuracy is sufficient for the intended application, it is not necessary to carry out a calibration. If the opposite is true, calibrate the transmitter.
- Close the shut-off valve of the gas cylinder, disconnect the cylinder and remove the calibration kit from the transmitter.
** The transmitter may take a few moments to go back to showing the
concentration values from before the verification procedure.
Reset to zero (AKO-575xxx only)**
OPTION A: CLEAN AIR
-
Before beginning the reset to zero procedure, ensure that the area of the premises is free of Freon gas and any other substances that could affect the transmitter, and that the transmitter has been operating for at least 20 minutes.
-
Enter the programming menu by pressing the SET key for 10 seconds and access parameter . The unit will
request a confirmation code ( ). Use keys and to enter code 63 and press SET. -
Use keys and to select option 1 and press SET. This will start the process.
-
During the process, the display will alternate between showing the gas concentration and the CAL message and the illuminated symbol. This process will last for between 30 seconds and 5 minutes. On completion, if the reset to zero process has been successful, the display will show “ ” and emit a long beep.
OPTION B: WITH NITROGEN GAS
Materials required:
AKO-58110 calibration kit + Nitrogen gas cylinder
- After initiating the reset to zero process, ensure that the transmitter has been operating for at least 20 minutes.
- Install the AKO-58110 calibration kit.
- Enter the programming menu by pressing the SET key for 10 seconds and access parameter . The unit will request a confirmation code ( ). Use keys and to enter code 63 and press SET.
- During the process, the display will alternate between showing the gas concentration and the CAL message and the symbol illuminated. This process will last for between 30 seconds and 5 minutes. On completion, if the reset to zero process has been successful, the display will show “End” and emit a long beep.
- Close the shut-off valve of the nitrogen gas cylinder, disconnect the cylinder and remove the calibration kit from the transmitter.
**** If any problems have been detected during the reset to zero process, the transmitter will emit three short beeps and display one of the following error codes:
CODE | ERROR |
---|---|
EC | A reading of greater than 300 PPM has been detected during the process. |
Et | The temperature of the sensor has experienced a variation of greater than |
5°C.
Eto| The maximum time period (5 min) has been exceeded and the reading has not
stabilized.
Use a cylinder with a type of gas that is suitable to the sensitivity of the transmitter to be calibrated. When calibrating the AKO-575400 universal gas transmitter, use a R134a gas cylinder.
Calibration of transmitter (AKO-575xxx only)
Materials required: AKO-58110 calibration kit + 2000 PPM calibrated gas
cylinder
Check whether current local regulations require a specific calibration
frequency. Always ensure that you comply with current local regulations.
- After initiating the calibration process, ensure that the transmitter has been operating for at least 20 minutes.
- Reset the transmitter to zero..
- Install the AKO-58110 calibration kit.
- Press the SET key for 10 seconds to enter the programming menu.
- When calibrating an AKO-575400 universal transmitter, configure parameter Gc2 to ALL.
- Access parameter . The unit will request a confirmation code ( ). Use keys and to enter code 63 and press SET.
- Use keys and to select option 1 and press SET. This will start the process.
- Open the shut-off valve of the gas cylinder and wait.
- During the process, the display will alternate between showing the gas concentration and the message and the illuminated symbol. This process will last for up to 15 minutes. On completion, if the calibration process has been successful, the display will show “ ” and emit a long beep.
- Close the shut-off valve of the calibrated gas cylinder, disconnect the cylinder and remove the calibration kit from the transmitter.
**** If any problems have been detected during the calibration process, the transmitter will emit three short beeps and show one of the following error codes:
CODE | ERROR |
---|---|
Et | The temperature of the sensor has experienced a variation of greater than |
- Eto| The maximum time period (15 min) has been exceeded and the reading has not stabilized.
**** The transmitter may take a few moments to go back to showing the
concentration values from before the calibration procedure.
AKO-575400 only: Once the verification procedure has been completed, remember
to adjust parameter Gc2 to the value prior to the start of the calibration
procedure.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>