seca RS-232 Adapter Box User Guide

June 4, 2024
seca

Quick Reference Guide seca 480
RS-232 adapter box

for product combinations based on seca 651/650 and seca 655/654
from firmware version R1.4

INTENDED USE

SAFETY PRECAUTIONS

WARNING! Injury from falls
► Route connecting cables in such a way that neither user nor the patient can trip over them.

CAUTION!
Patient hazard, damage to the device
Additional devices which are connected to medical electrical devices must provide evidence of compliance with the relevant IEC or ISO standards (e.g. IEC 60950 for data-processing devices). Furthermore, all configurations must comply with the requirements of standards for medical systems (see IEC 60601-1-1 or section 16 of edition 3.1 of IEC 60601-1 respectively). Anyone connecting additional devices to medical electrical devices is considered a system configured and is therefore responsible for ensuring that the system complies with the requirements of standards for systems. Your attention is drawn to the fact that local laws take precedence over the above-mentioned requirements of standards. In the event of any queries, please contact your local specialist dealer or Technical Service.

SCOPE OF DELIVERY

Pos. Pcs.
a RS-232 adapter box 1
b • Standard: interface cable for PC
• Option: Interface cable for dialysis system 1
c Quick reference

Note:
The “RS-232 Specifications (G3E)“ Art.-No. 30-35-00-019 are also required for proper setup. The document is available from such Service or as a  download from the Seca website (user account and password required)| 1

DEVICE OVERVIEW

No.|
---|---
1| RS-232 adapter box
2| SDB input cable (SDB: seca Device Bus)
3| SDB output to display
(for seca 651/650 only)
4| RS-232 output

MARKINGS

Symbol Product
Name and address of the manufacturer, date of manufacture
Unique Device Identification
Model number
Serial number, consecutive
ProdID Product identification number, consecutive
Mat.No Variant number
TempRg Temperature range for operation
IP21 Type of protection in accordance with IEC 60529:

• Protection against ingress of solid foreign bodies with a diameter of over 12.5 mm
• Protection against access with fingers
• Protection against drips
| The device conforms to EU directives
| Medical Device according to Regulation (EU) 2017/745
| Do not dispose of devices with household waste
| Follow instructions for use
Packaging
| Fragile. Do not throw or drop.
| Arrows indicate top of the product. Transport and store in an upright position.
| Protect from moisture
| Packaging material can be disposed of through recycling programs
| Permitted min. and max. temperature for transport and storage
| Permitted min. and max. moisture for transport and storage
| Permitted min. and max. air pressure for transport and storage

CONNECTING RS-232 CABLE

COMPATIBLE SCALES AND PRODUCT COMBINATIONS

CONNECTING SECA 651/650

CONNECTING SECA 655/654-BASED PRODUCT COMBINATIONS

SETTING DATA FORMAT

NOTE:
The setting shown is an example. The required data format depends on the device/PC connected.

NOTE:
You can also select the device mode „Basic“ or „Advanced“ here. For details refer to the instructions for use of the measuring device.

MEASURING

NOTE:

  • When using the RS-232 adapter box, measuring results are displayed and transmitted in SI units only.
  • Data transmission starts automatically as soon as stable measuring results are displayed.

HYGIENIC TREATMENT

Cleaning

NOTICE:
► Do not use aggressive or abrasive cleaning agents.
► Use a soft cloth dampened with mild soapsuds to wipe down the device with it.

Disinfecting

NOTICE:
► Use alcohol-based disinfectant (e.g. 70 % ethanol) exclusively
► Follow the instructions on the disinfectant.
► Moisten a soft cloth with disinfectant and wipe down the device with it.

TROUBLESHOOTING

Entry RS232 not shown in scale menu:
• Scale still connected to power supply while connecting the adapter box
– Restart scale as described in corresponding instructions for use Measuring results are not transmitted:
• Faulty cable connections
– Check cables for damage
– Ensure that plugs clicked properly into place
• Faulty data connection
– Refer to instructions for use of device/PC connected
– Refer to “RS-232 Specifications (G3E)“ Art. No. 30-35-00-019

TECHNICAL DATA

Dimensions
•Length:
•Width:
•Height:| 110mm 100mm 29 mm
---|---
Weight:| 0.2 kg
Ambient conditions, operation
•Temperature:
•Air pressure:
•Humidity| +10 °C to +40 °C
700 hPa – 1060 hPa
20%-80%
no condensation
Ambient conditions, storage
•Temperature:
•Air pressure:
•Humidity:
•Warm-up time from lowest storage temperature to operational temperature – At ambient temperature 20 °C:
– At ambient temperature 20 °C and with condensation:
•Cooling time from highest storage temperature to operational temperature (at ambient temperature 20 °C)| -10 °C to +65 C
700 hPa – 1060 hPa
0% – 95 %
no condensation
8 h 24 h 8 h
Ambient conditions, transport
•Temperature:
•Air pressure:
•Humidity:| -10 *C to +65 °C
700 hPa – 1060 hPa 0%-95% no condensation
Power supply:| seca Device Bus
Type of protection in accordance with IEC 60529:| IP21
Duty cycle:| Continuous Duty
A medical device in accordance with Regulation (EU) 2017/745:| Class I
Compatibility:
•seca 474
•seca 651/650
•seca 655/654| Firmware version 1.0 from Firmware version 1.4 from Firmware version 1.4
Interlaces
Adapter box:
•SDB (Seca Device Bus) – Input
– Output
•RS-232:
RS-232 cable:| 1 x cable with plug lx socket
3,5 mm jack socket
3,5 mm jack plug
D-SUB plug (depending on variant: female/male)

SPARE PARTS

Interface cable for PC (3,5 mm jack plug/D-SUB plug, female): 08-06-16-085
Interface cable for dialysis system (3,5 mm jack plug/D-SUB plug, male):

08-06-16-167

Medical Measuring Systems and Scales since 1840
seca gmbh & co. kg Hammer Steindamm 3-25 22089 Hamburg Germany
T +49 40 20 00 00 0
F 49 40 20 00 00 50
E info@seca.com 

seca operates worldwide with headquarters in Germany and branches in:
seca france seca united kingdom seca north america essay
schweiz seca zhong guo seca nihon seca mexico
seca austria seca polska seca middle east seca Suomi seca america latina seca asia pacific seca danmark seca benelux seca lietuva
and with exclusive partners in more than 110 countries.
All contact data at seca.com