seca 685 Instruction manual and guarantee
DEVICE DESCRIPTION
1.1 Intended use
Electronic multi-functional scale seca 685 is mainly used in hospitals,
medical practices and in-patient care facilities in accordance with national
regulations.
The scale is for conventional determination of weight and establishment of
general state of nutrition; it assists the physician supervising treatment in
making a diagnosis or deciding on a course of treatment.
To make an accurate diagnosis, however, other specific examinations have to be
ordered by the physician and their results taken into account, in addition to
determining a weight value.
1.2 Description of function
In addition to conventional determination of weight, the seca 685 provides
a function for determining body mass index. To this end, height is entered on
the keypad and the body mass index associated with the weight value is
calculated automatically. Length measuring devices from the seca 360° wireless
system can transmit height to the seca 685 wirelessly.
Via the seca 360° wireless wireless network, measuring results can be
transmitted wirelessly to a seca wireless printer or to a PC equipped with
seca analytics PC software and the seca USB wireless adapter.
The seca 685 can be moved on casters.
Use the scale only for the purpose named in the section entitled “Intended
use” on page 59.
1.3 User qualification
Assembly
Devices that are shipped partially assembled may only be mounted by
sufficiently qualified persons such as specialist dealers, hospital
technicians or seca service technicians.
Operation
The device may only be operated by healthcare professionals.
SAFETY INFORMATION
2.1 Safety information in these instructions for use
DANGER!
Used to identify an extremely hazardous situation. If you fail to take note of
this information, serious irreversible or fatal injuries will occur.
WARNING!
Used to identify an extremely hazardous situation. If you fail to take note of
this information, serious irreversible or fatal injuries may result.
CAUTION!
Used to identify a hazardous situation. If you fail to take note of this
information, minor to moderate injuries may result.
NOTICE!
Used to identify possible incorrect usage of the device. If you fail to take
note of this information, you may damage the device, or the measured results
may be incorrect.
NOTE
Includes additional information about use of the device.
2.2 Basic safety instructions
Handling the instrument
► Please take note of the information in these instructions for use.
► Keep the instructions for use in a safe place. The instructions for use are
a component of the device and must be available at all times.
DANGER!
Risk of explosion
Do not use the device in an environment in which one of the following gases
has accumulated:
–oxygen
– flammable anesthetics
– other flammable substances/air mixtures
CAUTION!
Patient hazard, damage to device
► Additional devices which are connected to electrical medical devices must
provide evidence of compliance with the relevant IEC or ISO standards (e.g.
IEC 60950 for dataprocessing devices). Furthermore, all configurations must
comply with the requirements of standards for medical systems (see IEC
60601-1-1 or Section 16 of the 3rd edition of IEC 60601-1 respectively).
Anyone connecting additional devices to electrical medical devices is
considered a system configurer and is therefore responsible for ensuring that
the system complies with the requirements of standards for systems. This also
applies to additional devices recommended by seca. Your attention is drawn to
the fact that local laws take precedence over the above-mentioned requirements
of standards. In the event of any queries, please contact your local
specialist dealer or Technical Service.
► Have servicing and verification carried out regularly as described in the
relevant section of this document.
► Technical modifications may not be made to the device. The device does not
contain any parts for servicing by the user. Only have servicing and repairs
performed by an authorized seca Service partner. You can find service partners
in your area at www.seca.com or by sending an
e-mail to service@seca.com.
► Only use original seca accessories and spare parts, otherwise seca will not
grant any warranty.
CAUTION!
Patient hazard, malfunction
► Keep other electrical medical devices, e.g. high-frequency surgical devices,
a minimum distance of approx. 1 meter away to prevent incorrect measurements
or wireless transmission interference.
► Keep HF devices such as cell phones a minimum distance of approx. 1 meter
away to prevent incorrect measurements or wireless transmission interference.
► The actual transmission output of HF equipment may require minimum distances
of more than 1 meter. Details can be found at
www.seca.com.
Preventing electric shock
WARNING!
Electric shock
► Set up the device so that the power supply socket is easy to reach and the
device can be disconnected from the power supply quickly.
► Ensure that your local power supply matches the information on the power
supply unit.
► Do not touch the power supply unit with wet hands.
► Do not use extension cables or power strips.
► Make sure that cables are not pinched or damaged by sharp edges.
► Make sure that cables do not come into contact with hot objects.
► Do not operate the device at an altitude of more than 3000 m above sea
level.
Avoiding infections
WARNING!
Risk of infection
► Before and after every measurement, wash your hands to reduce the risk of
cross-contamination and nosocomial infections.
► Hygienically reprocess the scale regularly as described in the respective
section in this document.
► Make sure that the patient has no infectious diseases.
► Make sure that the patient has no open wounds or infectious skin
alterations, which may come into contact with the device.
Avoiding injuries
WARNING!
Injury from falls
► Ensure that the device is positioned firmly and level.
► Route connecting cables (if present) in such a way that neither user nor
patient can trip over them.
► The device is not designed as a standing aid. Assist people with limited
motor skills when they are getting up, e.g. from a wheelchair.
► Make sure that the patient does not step onto and off the weighing platform
right at the edges.
► Make sure that the patient steps onto and off the weighing platform slowly
and safely.
WARNING!
Danger of slipping
► Ensure that the weighing platform is dry before the patient steps onto it.
► Ensure that the patients feet are dry before he or she steps onto the
weighing platform.
► Make sure that the patient steps onto and off the weighing platform slowly
and safely.
Preventing device damage
NOTICE!
Damage to device
► Ensure that no liquids enter the device. They can damage the electronics.
► Switch off the device before disconnecting the power supply unit from the
mains socket.
► For devices with power supply operation: Disconnect the power supply unit
from the mains socket if you intend to not use the device for a longer period
of time. Only this way it can be ensured that the device is de-energized.
► For devices with battery or rechargeable battery operation: If you do not
use the device for an extended period of time, remove batteries or
rechargeable batteries. Only this way it can be ensured that the device is de-
energized.
► Make sure not to drop the device.
► Do not expose the device to any impacts or vibrations.
► Perform function controls regularly as described in the relevant section in
this document. Do not operate the device if it is damaged or not working
properly.
► Ensure that there is no heat source in the immediate vicinity. Do not expose
to direct sunlight. The excessive temperature could damage the electronics.
► Avoid rapid temperature fluctuations. When the device is transported so that
a temperature difference of more than 20°C occurs, it must stay turned off for
at least 2 hours before it can be turned on again. Otherwise, condensation
water will form which can damage the electronics.
► Use the device only in the ambient conditions outlined in “Intended use”.
► Store the device only in the storage conditions outlined in “Intended use”.
► Use only chlorine and alcohol-free disinfectants which are explicitly
suitable for acrylic sheet and other sensitive surfaces (active ingredient:
quaternary ammonium compounds, for example).
► Do not use aggressive or abrasive cleaning agents.
► Do not use organic solvents (e.g. white spirit or petroleum spirit).
Handling measuring results
WARNING!
Patient hazard
This device is no diagnostic device. It simply assists the treating physician
in establishing a diagnosis.
► In order to make a precise diagnosis and initiate therapeutic measures,
besides determination of the weight, further targeted examinations must be set
up by the physician, and their results must be considered.
► The responsibility for diagnosis and treatment lies with the treating
physician.
NOTICE!
Inconsistent measuring results
► Before you electronically save measurement values determined using this
device and use them further (e.g. in seca PC software or in a hospital
information system), make sure that the measurement values are plausible.
► If measurement values are transmitted to seca PC software or a hospital
information system, make sure prior to further use that the measurement values
are plausible and are assigned to the correct patient.
Handling packaging material
WARNING!
Risk of suffocation
Packaging material made of plastic foil (bags) is a choking hazard.
► Keep packaging material out of reach of children.
► In the event that the original packing material may not be available
anymore, only use plastic bags with security holes in order to reduce the risk
of suffocation. Use recyclable materials if possible.
NOTE
Keep the original packing material for future use (e.g. returning for
servicing).
OVERVIEW
3.1 Controls
No. | Control | Function |
---|---|---|
1 | Display housing | Central control and display element, can be swiveled 180° |
to left and right
2| | Switch scale on and off
3| | Arrow key
• During weighing:
– press briefly: activate hold function
– press and hold: activate tare function
• In the menu:
– select submenu, select menu item
– increase value
4| | Arrow key
• During weighing:
– press briefly: activate BMI function
– press and hold: call up menu
• In the menu:
– select submenu, select menu item
– reduce value
5| | Enter key
During weighing (if wireless network is set up):
– press briefly: send measuring result to devices ready to receive it (PC with
USB wireless adapter)
– press and hold: print out measuring result (wireless printer)
• In the menu:
– confirm selected menu item
– save set value
6| Display| Display element for measuring results and to configure the device
7| Folding seat| For patients with restricted mobility who cannot stand for
the whole measuring process
8| Latch| Secures the folding seat in the upright position
9| Transport casters| The scale can be moved on these casters.
10| Spirit level| Shows whether the device is horizontal
11| Power supply connection| Serves to connect the power pack supplied
12| Foot screw| 4 pcs, for precise alignment
3.2 Symbols in the display
| Symbol| Meaning
---|---|---
A| | Operation with power pack
B| | Non-calibratable function active
C| | Memory location currently in use
D| | Weighing range currently in use see “Technical data” on page 104
3.3 Marking on the device and on the rating plate
Text/symbol | Meaning |
---|---|
Name and address of manufacturer, date of manufacture | |
Model number | |
Serial number, consecutive | |
GAL | Value in m/s² (verified models) |
• Gravitational acceleration on earth
• Depends on the intended location
ProdID| Product identification number, consecutive
Approval Type| Type designation of design approval
| Follow instructions for use
| Electrical medical device, type B
e| Value in units of measurement used to classify and verify a scale
d| Value in units of measurement which gives the difference between two
consecutive display values
| Active weighing range
| Class III scale to directive 2014/31/EU
| Device complies with EU standards and directives.
• M: Conformity label according to Directive 2014/31/EU governing non-
automatic weighing instruments (verified
models)
• 18: (Example: 2018) Year in which conformity verification was performed and
the CE label was applied (verified
model)
• 0102: Notified body metrology (verified models)
• 0123: Notified body medical products
| Symbol of the US Federal Communications Commission (FCC)
FCC ID| Device license number from the Federal Communications Commission (FCC)
IC| Device license number from Industry Canada
| Rating plate on the power supply connection socket
• x-y V: required supply voltage
• max xxx mA: maximum current consumption
• : Note polarity of device plug
• : Operate device with direct current
| Do not dispose of device with household waste
3.4 Marking on the packaging
| Protect from moisture
---|---
| Arrows indicate top of product
Transport and store in an upright position
| Fragile
Do not throw or drop
| Permitted min. and max. temperature for transport and storage
| Permitted min. and max. moisture for transport and storage
| Not sterile
| Do not reuse
| Open packaging here
| Packaging material can be disposed of through recycling programs
3.5 Menu structure
Other functions are available to you in the menu for the device. This
enables you to configure the device perfectly to suit your needs (details from
page 88).
BEFORE YOU REALLY GET STARTED …
4.1 Scope of supply
Components
No. | Component | Pcs. |
---|---|---|
A | Weighing platform | 1 |
B | Left-hand side part | 1 |
C | Left-hand cover with casters | 1 |
D | Folding seat | 1 |
E | Display support with display housing | 1 |
F | Railing with top cross-piece | 1 |
G | Bottom cross-piece | 1 |
H | Right-hand side part | 1 |
I | Right-hand cover with opening for power supply connection | 1 |
J | Power pack with adapters (model-dependent: power pack with Euro-connector) |
1
Fastenings
No. | Component | Pcs. |
---|---|---|
a | Countersunk screw M6 x 12 | 16 |
b | Threaded plate | 4 |
c | Cap to cover countersunk screws |
– black, for side covers
– white, for railing| 8
8
d| Hexagon bolt M6 x 16 (fitted to the weighing platform)| 6
e| Toothed ring J 6.4 (fitted to the weighing platform)| 6
f| Hexagon nut M6 (fitted to the weighing platform)| 6
g| Pan-head screw M6 x 20| 4
h| Washer 6.4| 4
i| Pan-head screw M8 x 20| 4
j| Washer 8.4| 4
k| Pan-head tapping screw B 3.5 x 9.5| 4
l| Cap for display support| 2
| Allen wrench, WAF 4 mm, not shown| 1
| Allen wrench, WAF 5 mm, not shown| 1
| Phillips screwdriver, size 1, not shown| 1
| Socket wrench WAF 8/10, not shown| 1
| Open-jawed wrench WAF 10, not shown| 1
| Instructions for use, not shown| 1
4.2 Assembling the device
We recommend that two people perform the assembly, as large individual
parts have to be aligned with one another and screwed together.
Fitting the side parts
You need the following fastenings for this assembly step (fitted to the
weighing platform):
- Remove the nuts and toothed rings on both sides of the weighing platform.
- Leave the screws in the bores.
- Place the weighing platform on its front edge and hold it in this position.
- Put the right-hand side part on the screws on the weighing platform.
- Put one toothed ring and one nut on each of the screws.
- Tighten up the nuts.
- Repeat steps 4. to 6. for the left-hand side part.
Fitting the side covers
You need the following fastenings for this assembly step:
- Put one threaded plate on each end of the righthand side part.
- Ensure that the bores on the threaded plates and the bores in the side part are in line over one another.
- Put the right-hand cover on the right-hand side part.
- Secure the cover in position on the side part using the two screws at each end.
- Repeat steps 1. to 4. for the left-hand cover.
- Set up the weighing platform.
- Put the black cover caps on the screws.
Assembling the railing
You need the following fastenings for this assembly step:
- Put the railing on the uprights of the side parts.
- Screw the railing onto the side parts using two screws per upright.
- Put the white cover caps on the screws.
Fitting the bottom cross-piece
You need the following fastenings for this assembly step:
- Put the bottom cross-piece on the side parts.
- Ensure that the bores in the cross-piece and the bores in the side parts are in line over one another.
- Screw the cross-piece firmly to the side parts.
Assembling the display support
You need the following fastenings for this assembly step:
-
Insert the loose end of the display cable in the top cross-piece so that the connector points towards the right-hand side part.
Caution!
Risk of tripping
If the display cable is pulled too far through the upright, it forms a loop which patients and users may trip over.
► Do not remove the cable ties on the display cable. The cable ties determine the length of the display cable. This allows it to be conveniently connected to the electronics box without forming a loop. -
Push the display cable through the upright until the connector protrudes at the bottom end of the upright.
-
Ensure that the display cable is routed in the top cross-piece in such a way that it cannot become trapped in subsequent assembly steps.
-
Put the caps on the cross-piece.
-
Ensure that the bores of the caps are precisely over the bores of the cross-piece.
-
Put a screw in each of the outer bores of the caps and tighten them up.
-
Put the display support on the cross-piece.
-
Ensure that the bores in the cross-piece are precisely over the inner bores of the caps and of the cross-piece.
-
Put a screw in each of the bores of the display support and tighten it up.
-
Connect the display cable to the electronics box on the weighing platform.
Fitting the folding seat
You need the following fastenings for this assembly step:
NOTE
►The folding seat is intended for patients who cannot stand for the whole
measuring process.
►The folding seat must be assembled so that it can be folded down in the
direction of the weighing platform.
- Screw the folding seat to the left-hand side part.
- Remove the cable ties on the legs of the folding seat.
Final steps
- Ensure that all parts are correctly located.
- Ensure that all screws are fully tightened.
4.3 Establishing power supply
Power is supplied to the scale with a power pack.
Connecting the power pack
WARNING!
Using the wrong power packs may cause bodily injury or damage to the
device
Conventional power packs may deliver a higher voltage than is indicated on
them. The scale may overheat, catch fire, melt or short-circuit.
► Use only original seca plug-in power packs with a 9 V or controlled 12 V
output voltage.
- If required, insert the power supply connector necessary for your power supply into the power pack.
- Insert the connector of the power pack into the connecting socket of the scale.
- Plug the power pack into a power supply socket.
- Perform a function check as described in the section entitled “Function check” on page 101.
OPERATION
CAUTION!
Bodily injury
Perform a function check as described in the section entitled “Function check”
on page 101 before using the device each time.
5.1 Making the scale operational
Moving the scale
- Tilt the device until it can be moved freely on the casters.
- Move the device in this position to the desired setup or storage location.
Aligning the scale
ATTENTION!
Incorrect measurement due to force shunts
If the scale and housing is in contact with some thing, e.g. a towel, weight
will not be measured correctly.
► Set the scale up so that only its foot screws are in contact with the floor.
-
Place the scale on firm, level surface.
-
Undo the knurled wheels.
-
Level the device by turning the foot screws.
The air bubble in the spirit level must be located in the exact center of the circle. -
Tighten the knurled wheels in the direction of the arrow.
The foot screws are secured against being adjusted.
Swiveling the display housing
The display housing of the device can be swiveled. As a result, it can be
perfectly positioned for every application.
► Swivel the display housing so that you can operate and read it conveniently.
Using the folding seat
Patients with restricted mobility can sit on the folding seat to be
weighed.
CAUTION!
Bodily injury, damage to device
The maximum permitted load for the folding seat is less than the maximum
permitted load for the device.
► Only load the folding seat to the maximum permitted load for the folding
seat, see “Technical data” on page 104.
► Do not use the folding seat if you have to assume that the patient’s weight
exceeds the maximum permitted load of the folding seat.
Folding down the folding seat
- Hold the folding seat steady so that it cannot drop down unintentionally.
- Turn the locking bar of the folding seat clockwise until it no longer protrudes over the seat.
- Fold the seat down.
- Ensure that the legs of the seat are fully extended.
Folding up the folding seat
CAUTION!
Bodily injury
If the folding seat is not correctly secured, it may drop down
unintentionally.
► Ensure that the folding seat is correctly latched when not in use.
► Ensure that the folding seat is not accidentally unlatched.
- Fold the seat up.
- Turn the locking bar of the folding seat counter- clockwise until it is securely engaged in the corre-sponding opening in the side part.
- Ensure that the legs of the seat are fully folded away.
5.2 Weighing
CAUTION!
Patients can be injured if they fall
People with restricted mobility may fall if they try to step onto the scale or
sit down on the scale.
► Support people with restricted mobility when they step onto the scale and
when they sit down.
Switching on the scale
► Press the Start key.
All the elements of the display are shown briefly, then SECA appears in the
display.
The scale is operational when 0.00 appears in the display.
Starting the weighing procedure
-
Ensure that the scale has no load.
-
Ask the patient to step onto the scale.
-
Ask the patient to keep still.
NOTE
If the patient is not steady on his or her legs, offer use of the folding seat. Support the patient as he or she sits down if necessary. -
Read off the measurement result.
Determining extra weight (TARE)
Using the TARE function, you can prevent any extra weight (e.g. a towel or
a support on the weighing platform) from affecting the weighing result.
ATTENTION!
Incorrect measurement due to force shunts
If an extra weight, e.g. a large towel, contacts the surface on which the
scale is placed, weight will not be measured correctly.
► Make sure that extra weights are only placed on the scale’s weighing
surface.
-
Switch on the scale.
-
Place the extra weight on the scale.
-
Hold the (hold/tare) arrow key down until the message “NET” appears in the display.
-
Wait until the display stops flashing and 0.00 appears instead.
-
Weigh the patient as described in the section entitled “Starting the weighing procedure”.
-
Read off the measurement result.
The additional weight is automatically deducted. -
To deactivate the TARE function, press the (hold/ tare) arrow key until the message “NET” is no longer displayed or turn off the scale.
NOTE
The maximum weight which can be displayed is reduced by the weight of the
objects already placed on the scale.
Keeping the measurement result in the display (HOLD)
When you activate the HOLD function, the weight value continues to be
displayed after the weight has been removed from the scale. As a result, you
can attend to the patient before recording the weight.
- Ensure that the scale has no load.
- Switch on the scale.
- Weigh the patient as described in the section entitled “Starting the weighing procedure”.
- Press the (hold/tare) arrow key briefly. The display flashes until a stable weight is measured. The weight is then continuously displayed. The symbol (non-calibratable function) and the message “HOLD” are displayed.
- To deactivate the HOLD function, briefly press the (hold/tare) arrow key. The symbol and the “HOLD” message are no longer displayed.
NOTE
If the auto hold function is activated, the weight value is automatically
displayed permanently until the scale switches itself off or is switched off,
see “Activating the auto hold (AHOLd) function” on page 91.
Determining and rating body mass index (BMI)
Body mass index relates height and weight to one another and consequently
allows more accurate information than ideal Broca weight, for example. A
tolerance range is quoted which is considered ideal for health.
The device has three memories for height. You can enter and save the height of
specific patients. Alternatively, you can save different start values and thus
set a patient’s actual height more quickly.
-
Ensure that the scale has no load.
-
Switch on the scale.
-
Press the (bmi/menu) arrow key briefly. The message “BMI” appears.
The last memory used is displayed (in this case, memory 2). -
You can adopt the memory displayed or select a different memory using the arrow keys.
-
Confirm your setting with Enter (send/print).
Arrows flash in the display. The last height saved to the selected memory is displayed. -
You can accept the height displayed or use the arrow keys to set a different height.
-
Confirm your setting with Enter (send/print).
The height entered is saved and will be available again for the next BMI calculation. NOTE
Make a note of the memory number so that you can call up this height for another BMI calculation. -
Weigh the patient as described in the section entitled Starting the weighing procedure.
The patient’s BMI is automatically calculated and displayed. -
Read off the BMI and compare it with the categories quoted further down.
-
To deactivate the BMI function, briefly press the Enter key (send/print).
BMI | Rating |
---|---|
Below 18.5 | Patient is underweight. There could be anorexic tendencies. |
Weight should be increased to improve well-being and performance. If in doubt,
seek the advice of a consultant.
Between 18.5 and 24.9| The patient’s weight is normal.
Between 25 and 30 (incipient obesity)| The patient is slightly to moderately
overweight. He or she should lose weight if already suffering from a disease
such as diabetes, high blood pressure, gout or disorders of fat metabolism.
Over 30| Weight needs to be lost urgently. The metabolism, the circulatory
system and the bones are all under strain. A strict diet, lots of physical
activity and behavioral training are all recommended. If in doubt, seek the
advice of a consultant.
Sending measurement results to wireless receivers
If the scale is integrated in a seca 360° wireless network, you can send
the measurement results to devices ready to receive them (e.g. wireless
printer, PC with USB wireless adapter) at the touch of a button.
► Press the Enter key (send/print).
– Press key briefly: send measurement results to all devices ready to receive
them
– Press and hold key: print out measurement result on wireless printer
Calculating and printing out BMI automatically
If you use this scale together with a wireless printer and a length
measuring rod from the seca 360° wireless system, you can have BMI calculated
and printed out automatically.
NOTE
The prerequisite for this function is that the devices are registered together
in one wireless group (see “The seca 360° wireless network” on page 93).
-
Perform the weighing operation.
-
Briefly press the Enter key (send/print) of the scale.
The measured value is sent to the wireless printer but not printed. -
Perform the height measurement.
-
Press and hold the Enter key (send/print) of the length measuring rod.
The measured value is sent to the wireless printer.
BMI is calculated.
Height, weight and BMI are printed out.
Automatic switching of weighing range
The scale has two weighing ranges. In weighing range 1 ( ), capacity is
reduced, but the increments in the weight display are finer. In weighing range
2 ( ), you After the scale is switched on, weighing range 1 is active. If a
particular weight is exceeded, the scale automatically switches to weighing
range 2.
To switch back to weighing range 1, proceed as follows.
► Completely remove the load from the scale.
Weighing range 1 is active again.
Switching off the scale
► Press the Start key.
5.3 Other functions (menu)
Other functions are available to you in the menu for the scale. This enables
you to configure the scale to suit your conditions of use perfectly.
-
The description of menu item “rF” can be found in the section entitled “Operating the scale in a wireless group (menu)” on page 95.
Navigating in the menu-
Switch on the scale.
-
Keep arrow key (bmi/menu) pressed until the menu is called up.
The last menu item selected (in this case, auto hold “AHOLd”) appears in the display. -
Press one of the arrow keys until the desired menu item appears in the display (in this case, filtering “FIL”).
-
Confirm your selection with Enter (send/print).
The current setting for the menu item or a submenu are displayed (in this case, level “0”). -
To amend the setting or call up a different submenu, press one of the arrow keys until the desired setting (in this case, level “2”) is displayed.
-
Confirm the setting with Enter (send/print).
You exit the menu automatically. -
To make more settings, call up the menu again and proceed as described.
NOTE
If no key is pressed for about 24 seconds, you exit the menu automatically. -
Press the Start key.
The settings will be saved.
The device switches off.
-
Deleting saved values automatically (ACLr)
To prevent current measurement results remaining in the memory of the
device and leading to an incorrect BMI calculation, you can set the scale so
that measurement results are automatically deleted after 5 minutes.
NOTE
This function is activated at the factory on some models. You can deactivate
the function if desired.
-
Select the item “ACLr” from the menu.
-
Confirm your selection.
-
Select the desired setting:
–on
–off -
Confirm your selection.
You exit the menu automatically.
Permanently saving the additional weight (Pt)
Using the Pre-tare function, you can permanently save an additional weight
and automatically deduct it from a measurement result. For example, you can
save an all- inclusive weight for shoes and clothing and then always deduct
this from the measurement result if a patient is weighed fully-clothed.
The device has three memories for weights. You can save various weights and
call them up individually depending on the initial situation, so that they are
automatically deducted from the measurement result.
-
Select the item “Pt” from the menu.
The message “Pt” appears. -
Confirm your selection.
The last memory used is displayed. -
You can adopt the memory displayed or select a different memory.
-
Confirm your selection.
Arrows flash in the display.
The additional weight saved to the selected memory is displayed. -
You can adopt the saved value or amend it using the arrow keys.
NOTE
When you enter the value “0”, the function is switched off. The message “Pt” is no longer displayed. -
Confirm your selection.
-
Ask the patient to position him or herself on the scale.
The patient’s weight is displayed.
The saved additional weight has been deducted automatically. -
To deactivate the function, select the item “Pt” from the menu again.
-
Confirm your selection.
The function is deactivated.
You exit the menu automatically.
NOTE
When you switch the scale off, the function is switched off. The message “Pt” is no longer displayed when you switch the scale on again.
Activating the auto hold (AHOLd) function
When you activate the auto hold function, the measure- ment result for
every weighing operation continues to be displayed after the weight has been
removed from the scale. It is then no longer necessary to activate the Hold
function manually for each individual weighing operation.
NOTE
This function is activated at the factory on some models. You can deactivate
the function if desired.
-
Select the item “AHOLd” from the menu.
-
Confirm your selection.
The current setting is displayed. -
Select the desired setting:
–On
–Off -
Confirm your selection.
You exit the menu automatically.
Activating beeps (bEEP)
You can set whether a beep is heard every time a key is pressed and when a
stable weight value is achieved.
The latter is significant for the Hold/auto hold function.
NOTE
The “Beep when weight is stable” function is activated at the factory. You can
deactivate this function if desired.
-
Select the item “bEEP” from the menu.
-
Confirm your selection.
-
Select a menu item.
– PrESS: beep when key is pressed
– Hold: beep when weight value is stable. -
Confirm your selection.
The current setting is displayed. -
Select the desired setting:
–On
–Off -
Confirm your selection.
You exit the menu automatically. -
If you also want to activate beeps for the second function, repeat the process.
Setting filtering (FIL)
Filtering (FIL) allows you to reduce interference (e.g. as a result of
patient movement) when determining weight.
-
Select the item “FIL” from the menu.
-
Confirm your selection.
The current setting is displayed. -
Select a filtering level.
– 0: no filtering
– 1: moderate filtering
– 2: high filtering -
Confirm your selection.
You exit the menu automatically.
Restoring factory settings (rESEt)
You can restore the factory settings for the functions below.
Function | Factory setting |
---|---|
Auto hold (AHOLd) | Model-dependent |
Beep (PrESS) | Off |
Beep (Hold) | On |
Filtering (FIL) | 0 |
Autoclear (Aclear) | Model-dependent |
Pre-tare (Pt) | 0 kg |
Height for Body Mass Index (BMI) | 170 cm |
Wireless module (SYS) | Off |
Autosend (ASEnd) | Off |
Autoprint (APrt) | Off |
NOTE
The wireless module is switched off when factory settings are restored.
Information about existing wireless groups is retained. Wireless groups do not
have to be set up again.
-
Select the item “rESEt” from the menu.
-
Confirm your selection.
You exit the menu automatically. -
Switch off the scale.
Factory settings are restored and are available when the scale is switched back on.
THE SECA 360° WIRELESS NETWORK
6.1 Introduction
This device is equipped with a wireless module. The wireless module allows
wireless transmission of measurement results for evaluation and documentation.
Data can be transmitted to the following devices:
- seca wireless printer
- PC with USB wireless adapter
seca wireless groups
The seca 360° wireless network operates with wireless groups. A wireless
group is a virtual group of transmitters and receivers. If several
transmitters and receivers of the same type are going to be operated, up to 3
wireless groups (0, 1, 2) can be set up with this device.
Setting up several wireless groups ensures the reliable transmission of
correctly-addressed measured values if you wish to use several examination
rooms each with comparable equipment.
The maximum distance between transmitters and receivers is approx. 10 meters.
Certain local circumstances such as the thickness and type of walls, may
reduce the range.
The following combination of devices is possible for each wireless group:
- 1 baby scale
- 1 personal scale
- 1 length measuring rod
- 1 seca wireless printer
- 1 PC with seca USB wireless adapter
Channels
Within a wireless group, devices communicate with each other on three
channels (C1, C2, C3). This guarantees that data are transmitted reliably and
without interference.
If you set up a wireless group with this scale, the device suggests three
channels which will ensure optimal data transmission. We recommend adopting
the channel numbers suggested.
You can also manually select the channel numbers (0 to 99), for example if you
wish to set up several wireless groups.
To ensure interference-free data transmission, the channels must be spaced
sufficiently far apart. We recommend a spacing of min. 30 between channel
numbers. Each channel number may only be used for one channel.
Example configuration for channel numbers when setting up 3 wireless groups
within a medical practice:
- wireless group 0: C1=_0, C2=30, C3=60
- wireless group 1: C1=10, C2=40, C3=70
- wireless group 2: C1=20, C2=50, C3=80
Device detection
If you set up a wireless group with the scale, it searches for other active
devices from the seca 360° wireless system. The devices detected are shown in
the display of the scale in the form of modules (e.g. MO 3). The numbers have
the following meaning:
- 1: personal scale
- 2: length measuring rod
- 3: wireless printer
- 4: PC with seca USB wireless adapter
- 7: baby scale
- 5, 6 and 8-12: reserved for system expansion
6.2 Operating the scale in a wireless group (menu)
All the functions you need to operate the device in a seca wireless group can
be found in the “rF” submenu. Information for navigating in the menu can be
found on page 88.
Activating the wireless module (SYS)
The device is supplied with the wireless module deactivated. You have to
activate it before you can set up a wireless group.
-
Switch on the device.
-
Select the “SYS” menu item from the “rF” submenu.
-
Confirm your selection.
-
Select the “On” setting.
-
Confirm your selection.
You exit the menu automatically. -
If you do not wish to make any more settings, press the Start key.
The settings will be saved.
The device switches off.
Setting up a wireless group (Lrn)
To set up a wireless group, follow the instructions below.
-
Switch on the device.
-
Call up the menu.
-
Select the item “rF” from the menu.
-
Confirm your selection.
-
Select the “Lrn” (learn) menu item from the “rF” submenu.
-
Confirm your selection.
The wireless group currently set (in this case: wireless group 0 “Id 0”) is displayed. If wireless group “0” already exists, use the arrow keys to select a different ID (in this case: wireless group 1 “Id 1”). -
Confirm your selection of wireless group.
The scale suggests a channel number for Channel 1 (in this case, “C1 0”). You can adopt the proposed channel number or use the arrow keys to set a different channel number. -
Confirm your selection for Channel 1.
The device suggests a channel number for Channel 2 (in this case, “C230”). You can adopt the proposed channel number or use the arrow keys to set a different channel number.
NOTE
Double-digit channel numbers are displayed without a space. The display “C230” therefore means: Channel “2”, channel number “30”. -
Confirm your selection for Channel 2.
The device suggests a channel number for Channel 3 (in this case, “C360”). You can adopt the proposed channel number or use the arrow keys to set a different channel number. -
Confirm your selection for Channel 3.
The message StOP appears in the display.
The device waits for signals from other wireless-capable devices in range. NOTE
On some devices, a special switch-on procedure has to be followed if they are going to be integrated in a wireless group. Follow the instructions for use for the device in question. -
Switch on the device, e.g. a wireless printer, you want to integrate in the wireless group. When the wireless printer is detected, you will hear a beep.
NOTE
As soon as you have integrated a wireless printer in the wireless group, you then have to select a print option (Menu\rF\APrt) and set the time (Menu\rF\tIME). -
Repeat step 11. for all the devices you want to integrate in this wireless group.
-
Press the Enter key to stop the search procedure.
-
Press one of the arrow keys to have the devices which have been detected displayed (in this case, “MO 3” for a wireless printer). If you have integrated several devices in one wireless group, press the arrow keys several times to ensure that all devices have been detected by the scale.
-
Exit the menu using the Enter key or wait until you exit the menu automatically.
-
If you do not wish to make any more settings, press the Start key.
The settings will be saved.
The device switches off.
Activating automatic send (ASEnd)
You can configure the device so that the measurement results are
automatically sent to all receivers which are ready to receive and are
registered to the same wireless group (e. g.: wireless printer, PC with USB
wireless adapter).
NOTE
If you are using a wireless printer, ensure that “Off” is not set as the print
option (see “Selecting print option (APrt)” on page 98).
-
Switch on the device.
-
Select the “ASEnd” menu item from the “rF” sub-menu and confirm your selection.
-
Select the “On” setting and confirm your selection.
You exit the menu automatically. -
If you do not wish to make any more settings, press the Start key.
The settings will be saved.
The device switches off.
Selecting print option (APrt)
You can configure the device so that the measurement results are
automatically printed out on a wireless printer registered to the wireless
group.
NOTE
This function is only accessible if a seca wireless printer has been
integrated in the wireless group via the “Learn” function.
-
Switch on the device.
-
Select the “APrt’ menu item from the “rF” submenu and confirm your selection.
-
Select the setting relevant for your device combination:
– HI: measurement results from length measuring devices
– MA: measurement results from scales
– HI_MA: measurement results from length measuring devices and scales
– Off: no automatic printing, printing only possible by pressing and holding the Enter key during the weighing procedure. -
Confirm your selection.
You exit the menu automatically. -
If you do not wish to make any more settings, press the Start key.
The settings will be saved.
The device switches off.
Set the time (tIME)
You can configure the system so that the wireless printer automatically
adds a date and time to your measurement results. To do this, you have to set
date and time once on this device and transmit it to the internal clock of the
wireless printer.
NOTE
This function is only accessible if a seca wireless printer has been
integrated in the wireless group via the “Learn” function.
-
Switch on the device.
-
Select the “tIME” menu item from the “rF” submenu.
-
Confirm your selection.
The current setting for “Year” is displayed. -
Set the correct year.
-
Confirm your selection.
-
Repeat steps 3. and 4. accordingly for “month” (Mon), “day” (dAy), hour (hour) and minute (Min).
-
Confirm each of your selections.
After confirming the setting for minute, you automatically exit the menu.
The settings are automatically transmitted to the wireless printer.
The wireless printer automatically adds a date and time to every printout.
NOTE
Follow the instructions for use for the wireless printer for further information about operating it. -
If you do not wish to make any more settings, press the Start key.
The settings will be saved.
The device switches off.
HYGIENE TREATMENT
WARNING!
Electric shock
The device is not de-energized when the on/off key is pressed and the display
goes out. Use of fluids on the device may cause electric shock.
► Ensure that the device is switched off before performing any hygiene
treatment.
► Disconnect the power supply connector before performing any hygiene
treatment.
► Before each hygiene treatment, take the rechargeable battery out of the
device (if present and removable).
► Ensure that no fluids penetrate the device.
CAUTION!
Damage to device
Inappropriate detergents and disinfectants may damage the sensitive surfaces
of the device.
► Use only disinfectants free of chlorine and alcohol which are explicitly
suitable for acrylic sheet and other sensitive surfaces (active ingredient:
quaternary ammonium compounds, for example).
► Do not use caustic or abrasive detergents.
► Do not use organic solvents (e.g. white spirit or petroleum spirit).
7.1 Cleaning
► Use a soft cloth dampened with mild soapsuds to clean the surfaces of the
device.
7.2 Disinfecting
- Check that your disinfectant is suitable for sensitive surfaces and acrylic sheet.
- Follow the instructions on the disinfectant.
- Disinfect the device:
► Moisten a soft cloth with disinfectant and wipe down the device with it.
► Comply with the intervals, see table.
Interval | Component |
---|---|
Before every measurement | Weighing platform |
After every measurement | Weighing platform |
If required | •Display |
• Membrane keyboard
7.3 Sterilizing
This device may not be sterilised.
FUNCTION CHECK
► Perform a function check before each use.
A complete function check includes:
- a visual inspection for mechanical damage
- a visual inspection and function check of the display
- function check of all the controls shown in the section entitled “Overview”.
- a function check of the optional accessories
If you find faults or deviations during the function check, as a first step
try to remedy the fault with the aid of the section entitled “Troubleshooting”
from page 101.
CAUTION!
Bodily injury
If you find faults or deviations during the function check which you are
unable to remedy with the aid of the section entitled “Troubleshooting” from
page 101, you may not use the device.
► Have the device repaired by seca service or an authorized specialist dealer.
TROUBLESHOOTING
Problem | Cause / Solution |
---|---|
No weight displayed with a load on the scale. | The scale has no power supply. |
– Check whether the scale is switched on
0.00 does not appear before weighing.| A load was already on the scale before
it was switched on.
– Remove the load from the scale
– Switch the scale off and back on again
One segment is illuminated either continuously or not at all.| The
corresponding point has a fault.
– Inform the service department
The display “StOP” appears.| Maximum load has been exceeded.
– Remove the load from the scale
The display “tEMP” appears.| The ambient temperature of the scale is too high
or too low.
– Set up the scale in an ambient temperature between +10 °C and +40 °C
– Wait for around 15 minutes until the device has adapted to the ambient
temperature
After switching on for the first time, measurement results are transmitted and
two beeps are heard.| • The device was unable to send measurement results to
the wireless receiver (seca wireless
printer or PC with seca USB wireless adapter).
– Ensure that the scale is integrated in the wireless network
– Ensure that the receiver is switched on
• Reception is being interfered with by HF equipment (e.g. cell phones) in the
vicinity.
– Keep HF equipment at least 1 meter away from transmitters and receivers in
the seca
wireless network. The actual transmission output of HF equipment may require
minimum distances of more than 1 meter. For details, go to
www.seca.com.
NOTE
If this fault is not eliminated, there will be no fresh acoustic warning when
further attempts are made to send data.
Only the “SYS” item is visible in the rF menu.| The wireless module is
deactivated.
– Activate wireless module (see “Activating the wireless module (SYS)” on page
95)
Only the “SYS” and “Lrn” items are visible in the rF menu.| The wireless
module is activated and no wireless group is set up.
– Set up wireless group (see “Setting up a wireless group (Lrn)” on page 95)
The items “APrt” and “tIME” are not visible in the rF menu.| No wireless
printer registered to the wireless group.
– Register the wireless printer to the wireless group via the “Lrn” menu item
(see “Setting
up a wireless group (Lrn)” on page 95)
After calling up the menu, the “rF” item is not displayed.| The scale’s
wireless module is defective.
– Inform seca service
The display “Er:[No.]:11” appears.| The scale has too high a load or too high
a load in one corner.
– Take the load off the scale or distribute the weight more evenly
– Restart the scale
The display “Er:[No.]:12” appears.| The scale has been switched on with too
high a load.
– Remove the load from the scale
– Restart the scale
The display “Er:[No.]:16” appears.| The scale was caused to oscillate, the
zero point could not be determined.
– Restart the scale
The Enter key (send/ print) is pressed and the display “Er:[No.]:71” appears.|
No data transmission possible, wireless module deactivated.
– Activate wireless module (see “Activating the wireless module (SYS)” on page
95)
The Enter key (send/ print) is pressed and the display “Er:[No.]:72” appears.|
No data transmission possible, no wireless group set up.
– Set up wireless group (see “Setting up a wireless group (Lrn)” on page 95)
SERVICING/VERIFICATION
10.1 Information on servicing and verification
We recommend having your device serviced prior to verification.
NOTICE!
Incorrect measurements as a result of poor servicing
► Have servicing and repairs carried out exclusively by seca Service or by an
authorized service partner.
► You can find service partners in your area at
www.seca.com or by sending an e-mail to
service@seca.com.
Have an authorized technician perform verification according to national legal
regulations.
Verification is necessary whenever one or more verification seals are damaged
or the contents of the verification counter no longer match the number on the
applicable verification counter sticker. If verification seals are damaged,
contact seca Service directly.
10.2 Checking the content of the verification counter
This seca scale is verified. Verifications may only be performed by authorized
agencies. To guarantee this, the scale is equipped with a verification counter
to record any change in verification-related data.
If you want to check whether the scale has been properly verified, proceed as
follows:
-
If necessary, turn off the scale.
-
Press and hold any key and start the scale.
The current content of the verification counter flashes in the display for a few seconds. -
Compare the content of the verification counter output with the number given on the verification counter sticker.
Both numbers have to match for the verification to be valid. If the sticker and the verification counter do not match, the scale must be verified. Please contact your service partner or seca Service. Once the scale has been verified, a new, updated verification counter sticker is used to identify the status of the verification counter. The person authorized to perform the verification secures this sticker with an additional seal. The verification counter sticker can be ordered from seca Service.
TECHNICAL DATA
11.1 General technical data
General technical data seca 685
Dimensions of scale
• Depth
• Width
• Height| 965 mm
910 mm
1100 mm
Dimensions of weighing platform
• Depth
• Width
• Height| 850 mm
800 mm
57 mm
Net weight| 59 kg
Temperature range
• Operation
• Storage
• Transport| +10 °C to +40 °C
-10 °C to +65 °C
-10 °C to +65 °C
Air pressure
• Operation
• Storage
• Transport| 700-1060 hPa
700-1060 hPa
700-1060 hPa
Humidity
• Operation
• Storage
• Transport| 30% – 80% no condensation
0% – 95% no condensation
0% – 95% no condensation
Height of digits| 25 mm
Power supply: power pack
• Supply voltage
• Power consumption with wireless module deactivated
• Power consumption with wireless module activated
• Insulated device, protection class II (EN 60601-1)| 12 V
Approx. 20 mA
Approx. 37 mA
Medical device in accordance with Directive 93/42/EEC| Class I with measuring
function
EN 60601-1: electrical medical device, type B|
Wireless transmission
• Frequency band
• Transmission power
• Standards applied| 2.433 GHz – 2.480 GHz
< 10 mW
EN 300 328
EN 301489-1:
EN 301489-17:
11.2 Technical data, weighing
Technical data, weighing
Calibration in accordance with Directive
2014/31/EU| Class III
Maximum load
• Folding seat
• Partial weighing range 1
• Partial weighing range 2| 150 kg
200 kg
300 kg
Minimum load
• Partial weighing range 1
• Partial weighing range 2| 2.0 kg
4.0 kg
Increments
• Partial weighing range 1
• Partial weighing range 2| 100 g
200 g
Tare range| 300 kg
Accuracy on initial calibration
• Partial weighing range 1, to 50 kg
• Partial weighing range 1, 50 kg to 200 kg
• Partial weighing range 2, to 100 kg
• Partial weighing range 2, 100 kg to 300 kg| ±50 g
±100 g
±100 g
±200 g
OPTIONAL ACCESSORIES
seca 360° wireless devices | Article number |
---|
Length measuring rods
• seca 274
• seca 264| Country-specific versions
Country-specific versions
Access ramp| 470-00-00-009
Wireless printers
• seca 360° Wireless Printer 465
• seca 360° Wireless Printer Advanced 466| Country-specific versions
Country-specific versions
PC software
• seca analytics 115| Application-specific license packages
seca 360° Wireless USB Adapter 456| 456-00-00-009
SPARE PARTS
Spare parts | Article number |
---|---|
Power pack with Euro-connector: 230 V~ / 50 Hz/12 V= / 130 mA | 68-32-10-252 |
Switchmode power pack with adapters: 100 – 240 V~ / 50 -60 Hz / 12 V= / 0.5 A |
68-32-10-265
DISPOSAL
Do not dispose of the device with household waste. The device must be disposed of properly as electronic waste. Comply with the national provisions applicable in your country. For further information contact our service department at: service@seca.com
WARRANTY
We offer a two-year warranty from the date of delivery for defects
attributable to faulty material or poor workmanship. This excludes all
moveable parts such as (rechargeable) batteries, cables, power supply units,
etc. Defects which are covered by the warranty shall be rectified free of
charge for customers on production of the sales receipt. No further claims can
be accepted. The costs of shipment in both directions shall be borne by the
customer where the device is not located at the customer’s premises. In the
event of any damage during shipment warranty claims can only be asserted where
the complete original packaging was used for shipment and the scales were
secured inside in the same manner as in the original packaging. You should
therefore keep all packaging.
The warranty shall become null and void where the device is opened by persons
not expressly authorised to do so by seca.
In the event of a warranty issue, please contact your local seca office or the
dealer from whom you ordered the product.
DECLARATION OF CONFORMITY
seca gmbh & co. kg hereby declares that the product meets the terms of the
applicable European directives. The unabridged declaration of conformity can
be found at: www.seca.com.
Medical Measuring systems and Scales since 1840
seca gmbh & co. kg
Hammer Steindamm 3-25
22089 Hamburg : Germany
Telephone +49 40 20 00 00 O
Fax +49 40 20 00 00 50
info@seca.com
seca operates worldwide with headquarters in Germany and branches in:
seca france
seca united kingdom
seca north america
seca schweiz
seca zhong guo
seca nihon
seca mexico
seca austria
seca polska
seca middle east
seca brasil
seca suomi
seca america latina
seca asia pacific
and with exclusive partners in more than 110 countries.
All contact data under www.seca.com