ANYLOAD OCSM Series Light Duty Crane Scale User Guide
- June 12, 2024
- ANYLOAD
Table of Contents
OCSM Series
Light Duty Crane Scale
User’s Guide
Safety Guides
For good performance and precise measurement on daily operation, observe the following safety guides and maintenance recommendations:
- Do NOT overload the scale. This may damage the load cell and will void the warranty.
- Do NOT leave the load hanged on the scale for too long. This will decrease scale’s accuracy and shorten the load cell’s life span.
- Inspect the shackle and hook before using. Check clips, pins and screws properly fitted and installed.
- Check battery frequently. When scale drained its battery, charge the battery with its dedicated charger or replace it with a new one
- Avoid rotating the scale, this may damage the load cell
- Do NOT use scale under thunder or rain.
- Hang the scale on shelf or in dry and well‐ventilated room. Do NOT put the scale on the ground directly.
- Do NOT attempt to repair the scale by yourself. Contact your local dealer or to the Technical Support.
Features
This scale is a combination of sound and proven mechanical design, with nowadays’ most advanced electronics to provide a superb feature sets. It is versatile, reliable, accurate and easy to operate.
- Superb Quality. Strictly in accordance with OIML R76, Chinese GB/T11883‐2002 National Standards, and European CE directives.
- Great Safety. Aluminum‐casting case, high firm hook and ring, dedicated weighing load cell for safety installation.
- Strong Reliability. Cutting‐edge technology, quality integrated circuit for high performance and long time stability.
- Broad Applicability. Popular and applicable in storage, textile, metallurgy industry, and so forth.
- Easy to Use. Infra‐red remote controlling design. Easy to operate on the scale or in distance.
- Complete Function. Division switch, unit conversion, automatic power saver, battery inspection, idle mode, tare set, etc.
Specifications
Accuracy Class | Chinese GB/T 11883‐2002 Class III Equivalent to OIML R76 |
---|---|
Tare Range | 100% F.S. |
Zero Range | 4% F.S. |
Stable Time | ≤10sec |
Overload | 100% F.S. + 9e |
Safety Load | 125% F.S. |
Ultimate Load | 400% F.S. |
Battery | 6V/5Ah lead acid battery. |
Charger | AC220V or 110V input, DC9V/1500mA output |
Op. Temp. | ‐10℃ ~ +40℃ |
Op. Humidity | 20 ℃ ≤90% |
Display | 30mm (1.2inch) LED or 35mm(1.38inch) LCD |
Capacities
model | max. cap. | min. cap. | resolution | division |
---|---|---|---|---|
OCSM‐200lb | 200lb | 2lb | 0.1lb | 2,000 |
OCSM‐400lb | 400lb | 4lb | 0.2lb | 2,000 |
OCSM‐600lb | 600lb | 4lb | 0.2lb | 3,000 |
OCSM‐1000lb | 1000lb | 10lb | 0.5lb | 2,500 |
OCSM‐2000lb | 2000lb | 20lb | 1lb | 2,000 |
Display & Keys
Scale & Remote Controller keys
key | name | function |
---|---|---|
On/Off | press for 1sec, power-on scale. |
press for 1sec, power-off scale.
exit without saving
| Zero| zero scale
with , unit switch
increase digit
| Tare| tare in/out
with , tare set
right scroll digit
| Hold| lock/unlock
with , enter User Setup
confirm
| 2nd| 2nd function
with , enter Password mode
exit and save
| Acc| accumulate weight
with , switch resolution
decrease digit
| Del| delete last weight
with , clear all weight
left scroll digital
| View| view accumulated weight
with , view battery level
| F1| NC
Indicators
indicator | name | note |
---|---|---|
STB | Stable | lit when weight is stable |
ZERO | Zero | lit when weight is at zero |
TARE | Tare | lit when weight is tared |
HOLD | Hold | lit when weight is locked |
lb | lb | lit when weight is in lb |
kg | kg | lit when weight is in kg (some models have no kg label or annunciator) |
Messages
Message | Stand for | note |
---|---|---|
Weight over range | ||
Detect weight | ||
Weight below range | ||
SETUP | User setup | |
END | Save and exit | |
OFF | Power off |
Operations
On/Off
- Press for 1sec, will power‐on scale.
- Scale performs initialization and boot‐up testing then display flashes 3 times, and displays max. cap., battery power, then detects weight and Auto‐Zero.
For information about Auto‐Zero, refer to Scale Setup in Technical Manual.
- Press or for 1sec, power‐off scale.
Scale displays battery power and off message, then cut off power
Zero
- Press or , will zero the scale. Indicator ZERO lights on.
- Scale must not be locked, otherwise displays.
- Scale must be stable, otherwise displays.
- Scale must not be tared, otherwise displays.
- Weight must be in Manual‐Zero Range, otherwise displays.
For information about Manual‐Zero Range, refer to Scale’s Configuration in Technical Manual.
Tare In / Tare Out
-
In gross mode, press or , will store the load as tare. Indicator TARE lights on
-
Scale must not be locked, otherwise displays.
-
Scale must be stable, otherwise displays.
-
Weight must more than 0, otherwise displays.
-
Weight must be lighter than 100% F.S., otherwise displays.
Tare will reduce the apparent overloading range of scale. For example, if a 5000*2kg scale has a 1000kg container as tare, the scale will overload at a new weight of 4018kg (5000 – 1000 + additional 9 divisions). -
In net mode, press or , will remove the tare stored. Indicator TARE lights off.
-
Scale must not be locked, otherwise displays
Lock / Unlock
- Press , will hold the reading in the scale. Indicator HOLD lights on.
- Scale must be stable, otherwise displays.
- Press , unlock scale. Indicator HOLD lights off.
Accumulate
- Press , will accumulate the current weight. will display indicating weight is accumulated. Scale uses displayed weight, so gross or net weight is added into the same accumulator.
- Scale must not be locked, otherwise displays.
- Scale must be stable, otherwise displays.
- Weight must exceed 0, otherwise displays.
- Scale must return zero before new weight can be accumulated otherwise displays
View
- Press , enter the View mode. Display flashes accumulated weight.
- Scale must not be locked, otherwise displays.
- If accumulated weight is zero, displays.
- Press , to view high 5‐digital and low 5‐digital.
- Press or to exit View mode.
Delete
- Press , will delete last accumulated weight. will display indicating last accumulated weight is deleted. Delete function only deletes the last weight.
- Scale must not be locked, otherwise displays.
- If last accumulated weight has been deleted, displays
Clear
- Press , then press , will clear all weights registered. will display indicating all weight are cleared.
Unit Switch
- Press , then press , switching unit between kg, lb, USR. When unit switches to kg, displays, indicator lights on (some models do not have kg label or annunciator). When unit switches to lb, displays, indicator lights on. When unit switches to USR, displays, indicator kg and lb lights off.
- Unit Switch changes unit temporarily. Scale does not save unit unless System Unit is changed.
For more information about USR, refer to Scale’s Configuration in Technical Manual
Tare Set
-
In gross mode, press , then press or , will enter Tare Set mode.
Scale displays 00000, waiting for user input. -
Press or , and or , input weight. Press , enter weight.
Indicator TARE lights on. -
Scale must be in gross mode, otherwise displays.
-
Weight must exceed 0, otherwise displays.
-
Weight must be lighter than 100% F.S., otherwise displays
Resolution Switch
- Press , and then press , switch display resolution
- Scale displays new resolution.
- High resolution offers better accuracy at the cost of longer measuring time and stricter requirement of load’s stability.
- Designed to meet the OIML R76’s directive, the scale has the best (default) performance at 2000 to 3000 division.
- Resolution Switch changes the apparent overloading range of scale. For example, if a 30001kg scale is switched to 30000.5kg, it will overload at 3004.5kg (3000 + 90.5), while by default, it overloads at 3009kg (3000 + 91).
- Default resolution will be restored next time when scale is powered on or enter User Setup. To save changes in resolution for later, enter User Setup and change Resolution.
For information about Resolution, refer to User Setup.
Battery Power View
- Press , and then press to view battery power.
- Scale displays battery voltage, for example , indicating 6.38V.
For information about battery, refer to Battery Maintenance
User Setup
- Press , and then press , enter User Setup mode.
- Message displays.
- Press will enter Resolution.
Resolution
Scale displays resolution to be set. For example , indicating resolution is set to 0.5.
-
Press or , and , change resolution
-
Press or , exit without saving. Press , exit and save.
Designed to meet the OIML R76’s directive, the scale has thebest (default) performance at 2000 to 3000 division -
Press will enter Auto‐Off.
Auto‐Off Time
-
Scale displays auto‐off time, e.g., indicating 15min.
-
Press or , and , change Auto‐Off time.
-
Press or , exit without saving. Press , exit and save.
Auto‐Off function maximizes scale’s battery life against people’s carelessness not to power off scale when it’s not working.
Auto‐Off starts countdown timer when there’s no action or load is stable. Any key pressing or motion in load restarts countdown timer.
Auto‐Off time can be set to: 0 (never auto‐off ), 5min, 15min, 30min, 60min. -
Press will enter Idle Mode setting.
Idle Time
-
Scale displays idle time, e.g., indicating 30sec.
-
Press or , and , change idle time.
-
Press or , exit without saving. Press , exit and save.
To maximize battery life, scale automatically enters Idle Mode, when there’s no action or the load is stable. In Idle Mode, scale works in low‐power consumption status. Any key pressing or motion in load wakes up scale from Idle Mode.
Idle time can be set to: 0 (never ilde), 5sec, 15sec, 30sec, 60sec. -
Press will enter to Brightness / Backlight setting.
Brightness / Backlight
Scale displays LED brightness / LCD backlight status.
-
Press or , and , change LED brightness / LCD backlight status.
-
Press or , exit without saving. Press , exit and save.
Dim LED brightness or turn off LCD backlight saves battery power dramatically.
LED brightness can be set to: 1(dim), 2(normal), 3(bright) 。 -
Press will enter Anti‐Motion
Anti‐Motion
Scale display Anti‐Motion level to be set, e.g., indicating level 1.
-
Press or , and , change Anti‐Motion level.
-
Press or , exit without saving. Press , exit and save.
At the cost of measuring time, Anti‐Motion function intelligently settles weight reading when scale is in motion. The weaker Anti‐Motion is, the faster weight reading displays, but the longer it takes to be stable.
Anti‐Motion can be set to: 0 (off), 1 (weakest), 2 (weak), 3 (normal), 4 (strong), 5 (strongest). -
Press will enter Auto‐Off again.
Battery Maintenance
To maximize battery life, observe the following battery maintenance guides:
- This scale is powered by a 6V rechargeable lead‐acid battery.
- Battery is permanently attached to battery door. To remove battery pack, remove both screws on the access door, pull battery pack straight out, and unplug battery cable from scale.
- Depending on LED brightness or LCD backlight setting, battery works from 40 hours to 100 hours.
- In order to conserve battery life, enable Auto‐Off and Idle Mode, dim LED brightness or turn off LCD backlight.
- Charging time for a completely discharged battery is approximately 8 hours.
- To obtain maximum service life, battery should be stored between ‐20℃ (‐4°F) and +50 ℃ (122°F). Stored batteries should be recharged every three months.
- When charging battery, charging indicator being green indicates lack of power, being red indicates full.
Troubleshooting
Symptom | Possible Cause | Suggested Solution |
---|---|---|
not power-on after is depressed | discharged / defective battery | check battery |
and charge
defective key| press harder and keep pressing 2sec
defective power cable| open front panel, check power cable
defective mainboard| contact representative
display flashes| discharged battery| charge battery
no action taken after key pressed| scale is disturbed| re-plug power cable
defective key| contact representative
weight reading not stable| load in motion| keep load stable
weak Anti-Motion| change Anti-Motion level
damped loadcell or mainboard| dry loadcell or mainboard
defective mainboard| contact representative
weight reading not zero when no load| discharged battery| charge battery
load-cell stressed too long| hang scale in storage
drifting loadcell| contact representative
large error in weight reading| scale not zeroed before applying load| manual
Zero scale before loading
wrong unit| switch to correct unit
scale requires calibration| calibrate scale
defective loadcell or mainboard| contact representative
battery cannot be recharged| defective charge board| contact representative
defective battery
short remote controlling distance| discharged / defective remote battery|
replace remote controller batteries
ANYLOAD Weigh & Measure Inc.
Website: www.anyload.com
Email: info@anyload.com
Fax: +1 866 612 9088
North America Toll Free: 1-855-ANYLOAD (269 5623)
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>