thermoscientific 3950 Reach In Incubator Instruction Manual
- June 12, 2024
- Thermoscientific
Table of Contents
- thermoscientific 3950 Reach In Incubator
- Product Information
- Covered Models
- Intended Use
- 2 CO2 Inject 15 p.s.i. Regulated (103.4 kPa)
- 1 CO2 Inject 15 p.s.i. Regulated (103.4 kPa)
- 2 CO2 Inject 15 p.s.i. Regulated (103.4 kPa)
- 1 CO2 Inject 15 p.s.i. Regulated (103.4 kPa)
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
thermoscientific 3950 Reach In Incubator
Product Information
The Thermo Scientific Reach-In Incubator is a large capacity CO2 incubator that is designed to maintain the optimal environment required for incubating cells and tissues for scientific research and clinical applications. The incubator is available in two models – 3950 T/C and 3951 T/C – and comes with a CO2 sensor. The unit operates at a voltage range of 100-120V or 200-230V at 50/60 Hz.
Covered Models
- 3950 T/C
- 3951 T/C
Intended Use
The Thermo Scientific Reach-In Incubator is intended for maintaining the optimal environment for the incubation of cells and tissues for scientific research and clinical applications. The unit maintains temperature and carbon dioxide (CO2) levels as set by the operator with increased relative humidity.
Non-Intended Use
The Thermo Scientific Reach-In Incubator is not intended for use where electrical or physical contact with the patient is established. It is not intended for use by the general public, within the patient environment, outdoors, or in potentially explosive environments or with flammable materials.
Product Usage Instructions
It is important to read the instruction manual carefully before using the Thermo Scientific Reach-In Incubator. Failure to read, understand, and follow the instructions in the manual may result in damage to the unit, injury to operating personnel, and poor equipment performance.
Caution
All internal adjustments and maintenance must be performed by qualified service personnel.
Warning
If the incubator is not used in the manner specified in the operating manual, the protection provided by the equipment design may be impaired.
Electrostatic Discharge (ESD)
The incubator contains parts and assemblies susceptible to damage by electrostatic discharge (ESD). Therefore, it is important to exercise caution while handling the unit.
Registering Your Warranty
To register your warranty, visit the Thermo Scientific website at www.thermoscientific.com/labwarranty.
User Manual
Forma 29 cu ft Reach-In Incubator
Model 3950 and 3951
Operating and Maintenance Manual 7003950 Rev. 30
Visit us online to register your warranty www.thermoscientific.com/labwarranty
Covered Models
CO2 Sensor*
3950 T/C
3951 T/C
Voltage** 100-120 200-230
- T/C is a thermal conductivity sensor. ** All units are 50/60 Hz.
Preface
Manual Number 7003950
30 ECNM-001355
12/06/21 Updated Exploded Spare Parts Drawings in pages 8-9 and 8-10.
TFI
29 ECNM-001110 11/06/2020 Updated Intended and Non-Intended use statements and
Assembly Drawings in pages 8-5 and 8-6.
TFI
28
41882
2/14/18 Added Intended use and Non-Intended use statement
bpg
27
41283
1/29/18 Updated dwg 3950-8
bpg
26 40087/IN-4743
7/27/16 Changed the label in the remote alarm artwork to a note – pg 6-1
ccs
25
40639
5/3/16 Updated for UL (G Smith), added risk assessment info
ccs
24 31580/IN-4675
10/5/15 Motor chg – part number, expl dwg, schematic
ccs
23
31289
7/18/14 Removed CE reference on pg 7-3
ccs
22 30374/IN-4419
4/22/14 Updated 1900008-06 drawing – pg 8-11 (left hand door swing)
ccs
21 29318//IN-4272
12/14/12 Clarified usage on exploded drawing 3950-08-4 list
ccs
20 28773/IN-4231
10/8/12 Updated Figure 1-6 to current
ccs
19
28267
2/20/12 Added cautions for running high RH or temps
ccs
18 24921/IN-3939
2/16/12 Removed accessory list – pg 7-2
ccs
Thermo Scientific
Reach-In Incubator i
Preface
CAUTION
Contains Parts and Assemblies Susceptible to Damage by
Electrostatic Discharge (ESD)
Important Read this instruction manual. Failure to read, understand and follow
the instructions in this manual may result in damage to the unit, injury to
operating personnel, and poor equipment performance.
Caution All internal adjustments and maintenance must be performed by
qualified service personnel.
Warning If the incubator is not used in the manner specified in this operating
manual, the protection provided by the equipment design may be impaired.
Caution When operating this unit under high temperatures and/or relative
humidity (RH), caution must be observed when the unit is shut down,
unexpectedly or by choice. Condensate can form on the interior of the chamber
as it cools. This includes the heating element surfaces. Heavy condensate can
bridge the ceramic insulators and cause a temporary electrical short to
ground. This electrical circuit is protected by a circuit breaker. However, to
avoid this contingency if these conditions occur, open the door of the unit
and allow it to stand open until the water vapor disperses.
Intended Use: The Thermo Scientific Large Capacity Reach in CO2 Incubators are
intended to maintain an optimal environment for the incubation of cells and
tissues for scientific research and clinical applications. These models
maintain temperature and carbon dioxiode (CO2) levels as set by the operator
with increased relative humidity.
Non-Intended Use: The Thermo Scientific Large Capacity Reach-In CO2 Incubators
are not intended for use where electrical or physical contact with the patient
is established. Not intended for use by the general public. Not intended for
use within the patient environment. Not to be used outdoors. Not intended to
be operated in potentially explosive environments or for use with flammable
materials.
Material in this manual is for information purposes only. The contents and the
product it describes are subject to change without notice. Thermo Fisher
Scientific makes no representations or warranties with respect to this manual.
In no event shall Thermo be held liable for any damages, direct or incidental,
arising out of or related to the use of this manual.
© 2021 Thermo Fisher Scientific. All rights reserved.
ii Reach-In Incubator
Thermo Scientific
Preface
Important operating and/or maintenance instructions. Read the accompanying
text carefully.
Potential electrical hazards. Only qualified persons should perform procedures
associated with this symbol. Equipment being maintained or serviced must be
turned off and locked off to prevent possible injury.
Lifting Hazard Warning. The incubator weighs more than 200lbs (91 kgs). Take
adequate safety measures when moving this device. WEEE Compliance: Thermo
Fisher Scientific has contracted with companies for recycling/disposal in each
EU Member State. For further information, send an email to
weee.recycle@thermofisher.com.
Always use the proper protective equipment (clothing, gloves, goggles, etc.)
Always dissipate extreme cold or heat and wear protective clothing. Always
follow good hygiene practices. Each individual is responsible for his or her
own safety.
Thermo Scientific
Reach-In Incubator iii
Preface
Do You Need Information or Assistance on Thermo Scientific Products?
If you do, please contact us 8:00 a.m. to 6:00 p.m. (Eastern Time) at:
1-740-373-4763 1-800-438-4851 1-877-213-8051 http://www.thermofisher.com service.led.marietta@thermofisher.com www.unitylabservices.com
Direct Toll Free, U.S. and Canada FAX Internet Worldwide Web Home Page Tech Support Email Address Certified Service Web Page
Our Sales Support staff can provide information on pricing and give you quotations. We can take your order and provide delivery information on major equipment items or make arrangements to have your local sales representative contact you. Our products are listed on the Internet and we can be contacted through our Internet home page.
Our Service Support staff can supply technical information about proper setup, operation or troubleshooting of your equipment. We can fill your needs for spare or replacement parts or provide you with on-site service. We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products.
Whatever Thermo Scientific products you need or use, we will be happy to discuss your applications. If you are experiencing technical problems, working together, we will help you locate the problem and, chances are, correct it yourself…over the telephone without a service call.
When more extensive service is necessary, we will assist you with direct factory trained technicians or a qualified service organization for on-the- spot repair. If your service need is covered by the warranty, we will arrange for the unit to be repaired at our expense and to your satisfaction.
Regardless of your needs, our professional telephone technicians are available
to assist you Monday through Friday from 8:00 a.m. to 6:00 p.m. Eastern Time.
Please contact us by telephone or fax. If you wish to write, our mailing
address is:
Thermo Fisher Scientific (Asheville) LLC 401 Millcreek Road, Box 649 Marietta,
OH 45750
International customers, please contact your local Thermo Scientific
distributor.
iv Reach-In Incubator
Thermo Scientific
Preface
Warranty Notes
Information You Should Know Before Requesting Warranty Service · Locate the
model and serial numbers. A serial tag is located on the unit itself. · For
equipment service or maintenance, or with technical or special application
inquiries, contact Technical
Services at 1-800-438-4851 or
1-740-373-4763 (USA and Canada).
Outside the USA, contact your local distributor.
Repairs NOT Covered Under Warranty · Calibration of control parameters.
Nominal calibrations are performed at the factory; typically ±1°C for
temperature, ±1% for gases, and ±5% for humidity. Our service personnel can
provide precise calibrations as a billable service at your location.
Calibration after a warranty repair is covered under the warranty. · Damage
resulting from use of improper quality water, chemicals or cleaning agents
detrimental to equipment materials. · Service calls for improper installation
or operating instructions. Corrections to any of the following are billable
services: 1) electrical service connection 2) tubing connections 3) gas
regulators 4) gas tanks 5) unit leveling 6) room ventilation 7) adverse
ambient temperature fluctuations 8) any repair external to the unit · Damage
resulting from accident, alteration, misuse, abuse, fire, flood, acts of God,
or improper installation. · Repairs to parts or systems resulting from
unauthorized unit modifications. · Any labor costs other than that specified
during the parts and labor warranty period, which may include additional
warranty on CO2 sensors, blower motors, water jackets, etc.
Thermo Scientific
Reach-In Incubator v
Preface
vi Reach-In Incubator
Thermo Scientific
Table of Contents
Section 1
Installation and Start-Up . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . .1-1
Control Panel Keys, Displays, & Indicators . . . . . . . . . . . . . . . . . .
. .1-2 Keypad Operation . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . .1-3 Unit Installation . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . .1-4
Preliminary Cleaning and Disinfecting . . . . . . . . . . . . . . . . . . . .
. .1-4 Installing the Shelves . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . .1-5 Leveling the Unit . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . .1-5 Attaching the Drain Lines . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5 The Water Supply .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Filling the Humidity Reservoir . . . . . . . . . . . . . . . . . . . . . . . .
. . . .1-6 Connecting the Unit to Electrical Power . . . . . . . . . . . . . .
. . . . . .1-7 Connecting the CO2 Gas Supply . . . . . . . . . . . . . . . . .
. . . . . . . . .1-8 Incubator Start-Up . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . .1-9 Setting the Operating Temperature .
. . . . . . . . . . . . . . . . . . . . . . .1-9 Setting the Overtemp Setpoint
. . . . . . . . . . . . . . . . . . . . . . . . . . .1-10 Setting the CO2
Setpoint . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-10
Section 2
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . .2-1
Calibrating the Temperature . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . .2-1 Calibrating the CO2 System . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . .2-2
Section 3
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . .3-1
Disable the ADD WATER Alarm . . . . . . . . . . . . . . . . . . . . . . . . .
. .3-1 Turn All Audible Alarms ON/OFF . . . . . . . . . . . . . . . . . . . .
. . . . . .3-1 Set an Access Code . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . .3-2 Set a Low Temp Alarm Limit . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . .3-2 Enable Temp Alarms to Trip
Relay Contacts . . . . . . . . . . . . . . . . . . .3-3 Set Low CO2 Alarm
Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3 Set
High CO2 Alarm Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . .3-4 Enable CO2 Alarms to Trip Relay Contacts . . . . . . . . . . . . . .
. . . . . .3-4 Set New Zero No. for New CO2 Sensors . . . . . . . . . . . . .
. . . . . . . . .3-5 Set New Span No. for CO2 Sensors . . . . . . . . . . . .
. . . . . . . . . . . . . .3-5 Enable the High Humidity Feature . . . . . . .
. . . . . . . . . . . . . . . . . . .3-6 Select a Primary Tank w/ Gas Guard
Option . . . . . . . . . . . . . . . . . .3-6 Disable the Gas Guard System . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7 Set RS485 Address .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Thermo Scientific
Reach-In Incubator vii
Table of Contents
Section 4 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . .4-1 Sensor Fault Alarms . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . .4-3
Section 5
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . .5-1
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . .5-1 Cleaning the Glass Doors . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . .5-1 Cleaning the Humidity Reservoir .
. . . . . . . . . . . . . . . . . . . . . . . . . . .5-2 Replacing the Power
Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
Discarding/Removing Incubator from Service . . . . . . . . . . . . . . . . .
.5-3
Section 6
Factory Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . .6-1
Connect the Remote Alarm Contacts . . . . . . . . . . . . . . . . . . . . . .
. . .6-1 Connect the RS485 Interface (P/N 190523) . . . . . . . . . . . . . .
. . . . .6-2 Connect Analog Output Boards (P/N 190512, 190543, & 190544) .6-3
CO2 Gas Guard (P/N 1900000) . . . . . . . . . . . . . . . . . . . . . . . . .
. . .6-4
Connect CO2 Gas Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . .6-5 De-activate Gas Guard . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . .6-5 Operation of CO2 Gas Guard . . . . . . . . . . . . .
. . . . . . . . . . . . . . . .6-5 Inner Doors . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . .6-6 Shaker Support
Shelves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.6-6
Section 7 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1
Section 8
Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . .8-1
Assembly Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . .8-2 Exploded Parts Drawings . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . .8-3
Section 9 Electrical Schematics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-1
Section 10 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-1
viii Reach-In Incubator
Thermo Scientific
Section 1 Installation and Start-Up
Figure 1-1. Front View
Figure 1-2. Side View
Figure 1-4. Top View
Thermo Scientific
Figure 1-3. Back View
Reach-In Incubator 1-1
Section 1 Installation and Start-Up
Control Panel Keys, Displays, & Indicators
Tank Low
Figure 1-4. Control Panel Components
Tank Low Indicator – Used with Gas Guard, lights when gas tank is low, if
applicable.
Silence – Mutes the audible alarm. See Section 4 for ringback times. Alarm
Indicator – Light pulses on/off during an alarm condition in the
cabinet. Mode Select Switch – Used to select Run, Setpoints, Calibration and
System Configuration Modes. Message Center – Displays system status. Mode
Select Indicators –
· Run: Run Menu · Set: Set Points Menu · Cal: Calibrate Menu · Config:
Configuration Menu Up and Down Arrows – Increases or decreases the number
values, toggles between parameter values of the selected mode. Enter – Stores
the changed value into computer memory. Heat Indicator – Lights when power is
applied to the heaters. Temp Display Displays temperature continuously
Scroll for Parameters Arrows – Moves the operator through the parameters of
the selected mode. Right arrow goes to next, left arrow goes to previous
parameter. CO2 Inject Indicator – Lights when CO2 is being injected into the
incubator. % CO2 Display – Displays CO2 percentage continuously.
1-2 Reach-In Incubator
Thermo Scientific
Section 1 Installation and Start-Up
Keypad Operation
The Model 3950 Series Reach-In Incubator has four basic modes that allow incubator setup: Run, Setpoints, Calibration and System Configuration.
· Run is the default mode during normal operation. · Set is used to enter system setpoints. · Calibration is used to calibrate various system parameters. · Configuration allows for custom setup of various options.
The chart below shows the selections under each of the modes.
RUN Default Mode
SETPOINT Temperature Overtemp CO2
Base Unit Displays Option Displays
CALIBRATION Temp Offset CO2 Cal
CONFIGURATION Audible Access Code Temp Lo Alarm Temp Relay CO2 Lo Alarm CO2 Hi Alarm CO2 Relay CO2 Z & S #’s RH High On/Off H20 Alarm On/Off Tank Select Gas Guard RS485 Address
Thermo Scientific
Reach-In Incubator 1-3
Section 1 Installation and Start-Up
Unit Installation
Caution Units must be installed against a wall or similar structure in an area of minimum ambient temperature fluctuation. A minimum of six (6) inches clearance is required at the top and back of the incubator plus a minimum three (3) inch clearance on each side for electrical, water and gas hook-ups. At least eight (8) inches clearance above the cabinet is required for service access.
Locate the unit on a firm level surface capable of supporting the unit weight of approximately 500 lbs.
Position the incubator away from doors and windows and heating and air
conditioning ducts.
Warning This incubator weighs approximately 500 lbs. Have sufficient personnel
available when moving.
Preliminary Cleaning and Disinfecting
Disinfect all interior surfaces with a general-use laboratory disinfectant, such as quaternary ammonium, to remove any residue from the production of the incubator. Rinse thoroughly with sterile distilled water, then 70% alcohol. Dry with a sterile cloth as needed. Disinfect the shelf channels and shelves, then rinse with distilled water before installing.
Caution Before using any cleaning or decontamination method except those recommended by the manufacturer, users should check with the manufacturer that the proposed method will not damage the equipment. Accidental spills of hazardous materials on or inside this unit are the responsibility of the user.
Installing the Shelves
The shelves may be installed at any level in the incubator. Install a shelf channel on each side. With the tabs pointing up, attach the channel by locating the rivet into a slotted hole, far end first. Pull the channel toward the front and slide the front rivet on the channel into the slotted hole and press down. Make sure that the channels are opposite each other so that the installed shelf will be level.
Leveling the Unit
Figure 1-5. Shelf Channel
Place a bubble-type level on a shelf inside the incubator. Adjust the feet as
needed; counterclockwise to lengthen or clockwise to shorten. Level the unit
front-to-back and left-to-right.
1-4 Reach-In Incubator
Thermo Scientific
Section 1 Installation and Start-Up
Attaching the Drain Lines Note Drain connections on the back of the unit are
shipped capped.
If humidity is not required, the caps should be left on the drains.
If manual or gravity feed method is used for humidity in the cabinet,
connecting the drains is optional.
If an in-house pure water supply is used to fill the reservoir, it is
recommended that the drains be connected to a floor drain.
Condensate Drain 3/8″ fitting
to 3/8″ ID hose
Back of Unit
1/4″ Water Inlet fitting to 1/4″ ID tubing to in-house water supply shut-off valve (30 PSIG max.)
Chamber Drain (Overflow) 3/8″ elbow with 3/8″ hose barb fitting. Install with
fitting sloping slightly downward.
Back View
Floor
Figure 1-6. Top View of Fitting Connections
The Water Supply
For best operation of the incubator, sterilized distilled, demineralized or de-ionized water should be used in the humidity reservoir. Water purity should be in the resistance range of 50 K to 1 M Ohm/cm, or a conductivity range of 20.0 to 1.0 uS/cm. Refer to ASTM Standard D5391-93 or D4195-88 for measuring water purity.
Distillation systems, as well as some types of reverse osmosis water purity systems, can produce water in the quality range specified. Tap water is not recommended as it may contain chlorine, which can deteriorate the stainless steel. Tap water may also have a high mineral content, which would produce a build-up of scale in the reservoir. High purity or ultra pure water is not recommended as it is an extremely aggressive solvent and will deteriorate the stainless steel. High purity water has a resistance of above 1 M to 18 M Ohm. Even high purity water can contain bacteria and organic contaminants. Water should always be sterilized or treated with a decontaminant, safe for use with stainless steel as well as safe for the product, prior to being introduced into the humidity reservoir.
Thermo Scientific
Reach-In Incubator 1-5
Section 1 Installation and Start-Up
The Water Supply (continued)
Caution Distilled or de-ionized water used in the humidity reservoir must be within a water quality resistance range of 50 K to 1 M Ohm/cm to protect and prolong the life of the stainless steel. Use of water outside the specified range will decrease the operating life of the unit and may void warranty.
The incubator is equipped with a liquid level float switch that detects low water levels in the humidity reservoir and produces an alarm condition. As all applications do not require humidity, this alarm can be disabled. To disable the ADD WATER alarm, see Section 3. If humidity is required in the current application, press the Silence key on the control panel to mute the alarm while the humidity reservoir is being filled.
Filling the Humidity There are three methods for filling the humidity reservoir; an in-house Reservoir pure water supply, the manual and the gravity feed methods.
In-house Pure Water Supply
All incubators provide a connection for using an in-house water supply. Connect a low pressure (not more than 30 PSI) water line, equipped with a manual shut-off valve, to the inlet fitting on the float valve reservoir. The 1/4″ hose connection is located on the rear of the incubator cabinet. With the water turned on, a float valve inside the incubator regulates the water level in the humidity reservoir. When the reservoir is full, a 10-second audible tone is heard and the ADD WATER message no longer displays.
Manual Method
This fill system consists of a funnel attached to the side duct sheet and tubing located inside the incubator. Disinfect the funnel and rinse with sterile water before using. This funnel may be moved to a preferred location by unscrewing the well nuts and disconnecting it from the sidewall. Tighten the nuts after relocation. The holes allotted for funnel use are slightly larger and in a centralized area. See Figure 1-7.
To fill the reservoir, pour sterile distilled water (approximately 2.5 gallons) into the funnel. When the float switch trips, a 10-second audible tone sounds and the ADD WATER message no longer displays. Add 1 gallon after the 10-second tone is heard.
1-6 Reach-In Incubator
Thermo Scientific
Manual Method (continued)
Pour in sterile, distilled water
Section 1 Installation and Start-Up
Center Front Unit Sidewall
Figure 1-7. Manual Feed Method with Funnel
Gravity Feed Method An optional carboy water fill system is also available.
The carboy option
kit includes an instruction sheet that details the installation of the system.
Connecting the Unit See the serial tag on the side of the unit or the Specifications section for
to Electrical Power
electrical specifications. Refer to the electrical schematics at the end of this manual.
Warning Connect the incubator to a grounded, dedicated circuit. The power cord connector is the mains disconnect device for the incubator. Position the incubator so the unit can be easily disconnected.
Plug the provided power cord into the power inlet connector on the back of the
cabinet, then into a grounded, dedicated electrical circuit.
The Model 3950 Series also has an internal outlet located on the right side of
the interior back wall. The outlet is to provide power (230 W maximum) to
accessory equipment. This outlet is not to be used when the temperature is
above 40°C and the RH more than 50%.
Thermo Scientific
Reach-In Incubator 1-7
Section 1 Installation and Start-Up
Connecting the CO2 Gas Supply
Warning High concentrations of CO2 gas can cause asphyxiation. OSHA Standards specify that employee exposure to carbon dioxide in any 8-hour shift of a 40-hour work week shall not exceed the 8-hour time weighted average of 5000 PPM (0.5% CO2). The short-term exposure limit for 15 minutes or less is 30,000 ppm (3% CO2). Carbon dioxide monitors are recommended for confined areas where concentrations of carbon dioxide gas can accumulate.
This incubator is designed for CO2 gas only. Connecting a flammable or toxic gas could result in a hazardous condition.
Gases other than CO2 should not be connected to this equipment. CO2 gas cylinders have a UN1013 label on the cylinder and are equipped with a CGA 320 outlet valve. Check the gas cylinder for the proper identification labels. The CO2 gas supply being connected to the incubator should be industrial grade, 99.5% pure. Do not use CO2 gas cylinders equipped with siphon tubes. A siphon tube is used to extract liquid CO2 from the cylinder, which can damage the pressure regulator. Consult with your gas supplier to ensure that the CO2 cylinder does not contain a siphon tube. Gas cylinders should also be secured to a wall or other stationary object to prevent them from tipping.
A two-stage CO2 pressure regulator is required to be installed on the outlet valve of the gas cylinder. Input pressure to the incubator must be maintained at 15 psig (103.4 kPa) for proper performance of the CO2 control system. A single stage CO2 pressure regulator will not maintain 15 psig (103.4 kPa) to the incubator as the pressure in the CO2 cylinder decreases; therefore, a two- stage regulator is recommended.
If higher purity CO2 is desired inside the incubator (greater than 99.5% pure), the pressure regulator should be constructed with a stainless steel diaphragm, along with specifying the purity of the CO2 from the gas supplier. Follow the manufacturer’s instructions to ensure proper and safe installation of the pressure regulator on the gas cylinder.
Consult your facility safety officer to ensure that the equipment is installed in accordance with the codes and regulations applicable in your area.
The CO2 gas supply being connected should be industrial grade 99.5% pure and should not contain siphon tubes. Install a two-stage pressure regulator at the cylinder outlet. The high pressure gauge at the tank should have 0-2000 psig range. The low pressure gauge, at the incubator inlet, should have a 0-30 psig range. Input pressure to the incubator must be maintained at 15 psig (103.4 kPa), ±5 psig.
1-8 Reach-In Incubator
Thermo Scientific
Section 1 Installation and Start-Up
Connecting CO2 Gas Supply (continued)
Caution Each tank should be connected to one incubator only.
The incubator has serrated fittings on the back of the cabinet to connect the
gas supply (Figure 1-2). The fitting is labeled CO2 Inlet #1 Tank. Secure
connections with clamps. Check all fittings for leaks.
For units having the CO2 Gas Guard option, refer to Section 6.
Incubator Start-Up
With the incubator properly installed, connected to power, the humidity reservoir filled and the unit connected to a gas supply, system setpoints can be entered. The following setpoints can be entered in Set Mode: Temperature, Overtemperature and CO2. To enter Set Mode, press the Mode key until the Set indicator lights. Press the right and/or left arrow keys until the proper parameter appears in the message display center. See Chart 1-1 for more detail.
Setting the Operating Temperature
This incubator has an operating temperature setpoint range of 5.0°C above ambient to 60.0°C. It is shipped from the factory with a temperature setpoint of 10.0°C. At this setting, all heaters are turned off. To change the operating temperature setpoint:
1. Press the Mode key until the Set indicator lights.
2. Press the right arrow until “Temp XX.X” is displayed in the message center.
3. Press up/down until the desired temperature setpoint is displayed.
4. Press Enter to save the setpoint.
5. Press the Mode key until the Run indicator lights for Run mode or press the right/left arrow keys to go to next/previous parameter.
Caution Any equipment placed inside chamber must be rated for unit operating temperature and humidity.
Thermo Scientific
Reach-In Incubator 1-9
Section 1 Installation and Start-Up
Setting the Overtemp Setpoint
Warning The independent overtemp system is designed as a safety to protect the incubator only. It is not intended to protect or limit the maximum temperature of the cell cultures or customer’s equipment inside the incubator if an overtemp condition occurs.
All Model 3950 Series incubators are equipped with a secondary temperature
monitoring system to monitor the air temperature inside the cabinet. This
system is designed as a safety device to turn off all heaters in the event of
a temperature control failure. The temperature control point in the incubator
will be approximately ±1°C of the overtemp setpoint.
The overtemp setpoint is set by the factory (default) at 40°C. However, the
overtemp can be reset over a range from 0.5°C above the operating temperature
setpoint to 65°C.
If the incubator’s operating temperature setpoint is set above the overtemp
setpoint, the overtemp setpoint will automatically update to 1°C above the
temperature setpoint. It is recommended that the overtemp setpoint be
maintained at 1°C over the operating temperature setpoint.
To set the Overtemp setpoint:
1. Press the Mode key until the Set indicator lights.
2. Press the right arrow until Otemp XX.X is displayed in the message center.
3. Press the up or down arrow until the desired Otemp setpoint is displayed.
4. Press Enter to save the setting.
5. Press the Mode key until the Run indicator lights, or press the right or
left arrow to go to the next or previous parameter.
Setting the CO2 Setpoint
All CO2 cells are calibrated at the factory at 37°C, medium humidity and 10% CO2. Therefore, if a temperature setpoint of 37°C has been entered, the humidity reservoir has been filled and the CO2 control is set to run between 0-10%, the CO2 setpoint may be entered immediately. Otherwise, it is important to allow the unit 12 hours to stabilize at the temperature setpoint before entering the CO2 setpoint.
1-10 Reach-In Incubator
Thermo Scientific
Section 1 Installation and Start-Up
Setting the CO2 Setpoint (continued)
All models of the incubator have a CO2 setpoint range of 0.0% to 20.0%. The incubator is shipped from the factory with a CO2 setpoint of 0.0%. At this setting, all CO2 control and alarms are turned off.
To change the CO2 setpoint:
1. Press the Mode key until the Set indicator lights.
2. Press the right arrow until “CO2 XX.X” is displayed in the message center.
3. Press the up/down arrow until the desired CO2 setpoint is displayed.
4. Press Enter to save the setpoint.
5. Press the Mode key until the Run indicator lights to go to Run mode, or press the right/left arrow keys to go to next/previous parameter.
Thermo Scientific
Reach-In Incubator 1-11
Section 1 Installation and Start-Up
Chart 1-1
Set Mode
Press MODE to light SET
To Set:
Operating Temperature
Silence
Mode
Run Set
Cal Config
SET POINTS
Scroll for Parameters
Enter
36.9 Heat
Temp
Inject
4.9 % CO2
Press MODE to move
to CALIBRATE
mode
Mode TEMP
XX.X C
Scroll for Parameters
Press to return
to previous parameter
Numbers increase
Press Enter
Enter
to save setting
Numbers decrease
Over Temperature
Numbers increase
Mode OTEMP XX.X C Enter
Press Enter to save setting
Scroll for Parameters
Press to return to previous parameter
Numbers decrease
Percent CO2
Mode
CO2 XX.X%
Scroll for Parameters
Press to return to previous parameter
Numbers increase Press Enter
Enter
to save setting Numbers decrease
1-12 Reach-In Incubator
Thermo Scientific
Section 2 Calibration
After the unit has stabilized, several different systems can be calibrated. In
the calibration mode, the air temperature, CO2 and RH levels can be calibrated
to reference instruments. To access the calibration mode, press the Mode key
until the Cal indicator lights. Press the right and/or left arrow until the
proper parameter appears in the message center. See Chart 2-1 at the end of
this section for more detail.
Calibration frequency is dependent on use, ambient conditions and accuracy
required. A good laboratory practice would require at least an annual
calibration check. On new installations, all parameters should be checked
after the stabilization period.
Prior to calibration, the user should be aware of the following system
functions. While the unit is in the calibration mode, all system control
functions are stopped so the unit remains stable. Readout of the system being
calibrated will appear on the message center. If no keys are pressed for
approximately five minutes while in Calibration mode, the system will reset to
Run mode so control functions are reactivated.
Caution Before making any calibration or adjustments to the unit, it is
imperative that all reference instruments be properly calibrated.
Calibrating the Before calibration, allow the cabinet temperature to stabilize. Place the
Temperature
calibrated instrument in the center of the chamber. The instrument should be in the airflow, not against the shelf.
Temperature Stabilization Periods
Start-Up – Allow 12 hours for the temperature in the cabinet to stabilize before proceeding.
Presently Operating – Allow at least 2 hours after the display reaches setpoint for temperature to stabilize before proceeding.
Thermo Scientific
Reach-In Incubator 2-1
Section 2 Calibration
Calibrating the Temperature (cont.)
1. Press the Mode key until Cal indicator lights.
2. Press the right arrow until “TEMPCAL XX.X” appears in the message center.
3. Press the up/down arrow to match the display to the calibrated instrument.
4. Press Enter to store calibration.
5. Press the Mode key to return to Run or the right/left arrow to go to next/previous parameter.
Calibrating the CO2 System
Model 3950 Series incubators have a CO2 sensor. The incubator atmosphere is not only effected by the quantity of CO2 present but also by the air temperature and water vapor present in the incubator atmosphere. In monitoring the effects of CO2, air temperature and absolute humidity must be held constant so any change is caused by a change in CO2 concentration.
Changing temperature or changing from elevated humidity to room ambient humidity levels will require re-calibration of the CO2 control.
CO2 Sensor Stabilization Periods
Start-Up – The CO2 sensor has been calibrated at the factory for 37°C and
elevated humidity. Allow the temperature, humidity and CO2 levels in the
chamber to stabilize at least 12 hours before checking the CO2 concentration
with an independent instrument.
Presently Operating – Make sure the chamber doors are closed. Allow at least 2
hours after the temperature and CO2 displays reach their setpoints for chamber
atmosphere stabilization.
2-2 Reach-In Incubator
Thermo Scientific
Section 2 Calibration
Calibrating the CO2 System (continued)
1. Make sure the stabilization periods outlined above are followed.
2. Sample the chamber atmosphere through the sample port with an independent
instrument. Sample the atmosphere at least 3 times to ensure accuracy of the
instrument.
3. Press the Mode key until the Cal indicator lights.
4. Press the right arrow until “CO2 CAL XX.X” is displayed in the message
center.
5. Press the up /down arrows to change the display to match the independent
instrument.
6. Press Enter to store the calibration.
7. Press the Mode key to return to Run or the right or left arrows to go to the next/ previous parameter.
Thermo Scientific
Reach-In Incubator 2-3
Section 2 Calibration
Calibrate Mode
Press MODE to light CAL
Silence
Chart 2-1
Mode
Run Set
Cal Config
CALIBRATE
Scroll for Parameters
Enter
36.9 Heat
Temp
Inject
4.9 % CO2
To Calibrate:
Operating Temperature
Press MODE to move
to SYS CONFIG Mode mode
TEMPCAL XX.X
Scroll for Parameters
Press to return
to previous parameter
Numbers increase
Press Enter
Enter
to save setting
Numbers decrease
CO2
Numbers increase
Mode CO2ZERO XX.X Enter
Press Enter to save setting
Scroll for Parameters
Press to return to previous parameter
Numbers decrease
2-4 Reach-In Incubator
Thermo Scientific
Section 3 Configuration
Several features available in Configuration Mode allow custom setup of the
incubator. These features are listed and described below. All features may not
be necessary in all applications, but are available if needed. To enter
Configuration mode, press the Mode key until the Config indicator lights.
Press the right and/or left arrow until the appropriate parameter appears in
the message center. See Chart 3-1 at the end of this section for more detail.
Disable the ADD All applications do not require humidity. For this reason, the ADD
WATER Alarm
WATER alarm can be disabled. The factory default for the ADD WATER alarm is ON.
1. Press Mode key until Config indicator lights.
2. Press right arrow until H2O ALM XXX is displayed in message center.
3. Press up/down arrow to toggle H2O ALM ON or OFF.
4. Press Enter to save the setting.
5. Press Mode key to return to Run mode or right/left to go to next/previous parameter.
Note Verify that the RH HIGH feature is set to OFF after disabling the ADD WATER alarm.
Turn All Audible The audible alarms can be turned on or off. The factory
setting is ON. Alarms ON/OFF 1. Press Mode key until Config indicator lights.
2. Press right arrow until AUDIBLE XXX is displayed in message center.
3. Press up/down arrow to toggle AUDIBLE ON/OFF.
4. Press Enter to save the setting.
5. Press Mode key to return to Run mode or right/left to go to next/previous
parameter.
Thermo Scientific
Reach-In Incubator 3-1
Section 3 Configuration
Set an Access Code
A 3-digit Access Code can be entered to avoid unauthorized personnel from changing the setpoints, calibration, or configuration. A setting of 000 will bypass the access code. The factory setting is 000.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until ACC CODE XXX is displayed in the message center.
3. Press up or down arrow to change the access code.
4. Press Enter to save the access code.
5. Press the Mode key to return to the Run mode or right/left to go to next/previous parameter.
Set a Low Temp Alarm Limit
The low temperature alarm limit (tracking alarm) is the deviation from the temperature setpoint, which will cause a low temp alarm. The low temp alarm is variable from 0.5° below setpoint to 5.0° below setpoint. The factory setting is 1.0° below setpoint. A minus sign in the display indicates that the alarm setting is below the setpoint.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until TEMP LO -X.X is displayed in the message center.
3. Press up/down arrow to change the low temp alarm limit.
4. Press Enter to save the low temp alarm limit.
5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter.
3-2 Reach-In Incubator
Thermo Scientific
Section 3 Configuration
Enable Temp Alarms to Trip Relay Contacts
The temperature alarms can be programmed to trip the remote alarm contacts. A setting of ON will cause this, a setting of OFF will not allow temperature alarms to trip the contacts. The factory setting is ON.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until TMP RLY XXX is displayed.
3. Press the up/down arrow to toggle the setting ON/OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run or the right/left arrow key to go to next/previous parameter.
Set Low CO2 Alarm Limit
The low CO2 alarm limit (trackiing alarm) is the deviation from the CO2 setpoint, which will cause a low CO2 alarm. The setpoint is variable from 0.5% CO2 below setpoint to 5.0% CO2 below setpoint. The factory setting is 1.0 % CO2 below setpoint. A minus in the display indicates that the alarm setting is below the setpoint.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until CO2 LO -X.X is displayed in the message center.
3. Press up/down arrow to change the low CO2 alarm limit.
4. Press Enter to save the low CO2 alarm limit.
5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter.
Thermo Scientific
Reach-In Incubator 3-3
Section 3 Configuration
Set High CO2 Alarm Limit
The high CO2 alarm limit (tracking alarm) is the deviation from the CO2 setpoint, which will cause a high CO2 alarm. The setpoint is variable from 0.5% CO2 above setpoint to 5.0% CO2 above setpoint. The factory setting is 1.0 % CO2 above setpoint.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until CO2 HI X.X is displayed in the message center.
3. Press up/down arrow to change the high CO2 alarm limit.
4. Press Enter to save the high CO2 alarm limit.
5. Press the Mode key to return to run mode or right/left to go to next/previous parameter.
Enable CO2 Alarms to High and Low CO2 alarms can be programmed to trip the
remote alarm Trip Relay Contacts contacts. A setting of ON will cause this, a
setting of OFF will not allow
CO2 alarms to trip the contacts. The factory setting is ON.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until CO2 RLY XXX is displayed in the message
center.
3. Press up/down arrow to toggle the setting ON/OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run Mode or right/left to go to
next/previous parameter.
3-4 Reach-In Incubator
Thermo Scientific
Section 3 Configuration
Set New Zero # and If a new T/C CO2 sensor is being installed, the two numbers on the
Span # for CO2 Sensors
factory installed sticker on the cell must be entered to calibrate the CO2 in the unit.
Note For the technician’s convenience, a second label containing the two numbers is supplied with the new T/C cell and should be affixed over the original label, inside the electronics drawer.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until T/C ZR# XXXX is displayed in the message
center.
3. Press up/down arrow to change the zero number to match the sticker.
4. Press Enter to save the setting.
5. Press the right arrow until T/C SP# XXXX is displayed in the message
center.
6. Press up/down arrow to change the span number to match the sticker.
7. Press Enter to save the setting. Pressing Enter will erase any CO2 Cal
offset value previously entered into the Calibration menu.
8. Press the Mode key to return to Run mode or right/left to go to
next/previous parameter.
Thermo Scientific
Reach-In Incubator 3-5
Section 3 Configuration
Enable the High Humidity Feature
In normal operation, the incubator will maintain the humidity level above 80%. If a higher RH level is required, the RH HIGH feature may be turned ON. This allows the heater attached to the humidity reservoir to be enabled, resulting in RH levels above 90%. The factory default for the RH HIGH feature is OFF.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until RH HIGH XXX is displayed in the message center.
3. Press up/down arrow to toggle RH HIGH ON or OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter.
Note Verify that the ADD WATER alarm is set to ON after enabling the RH HIGH feature.
Select a Primary Tank On units equipped with the Gas Guard option, a primary tank can be
w/Gas Guard Option
selected. The primary tank will be either Tank 1 or 2. The factory setting is Tank 1.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until TANK SEL X is displayed in the message center.
3. Press the up/down arrow to toggle the setting between 1 and 2.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter.
3-6 Reach-In Incubator
Thermo Scientific
Section 3 Configuration
Disable the Gas Guard System
On units equipped with the Gas Guard option, the Gas Guard system may be
turned ON, or OFF if it is not in use. The factory setting is ON.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until GAS GRD XX is displayed in the message center.
3. Press up/down arrow to toggle the setting ON/OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter.
Set RS485 Address
On units that have the RS485 option, direct communication with the Model 1535 alarm system can be established. Each piece of equipment connected to the 1535 must have a unique address. An address of 0-24 can be entered for the incubator. A setting of 0 is an invalid address that the 1535 will ignore. The factory setting for the RS485 address is 0.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until 485 ADDR XX is displayed in the message center.
3. Press up/down arrow to move the RS485 address.
4. Press Enter to save the RS485 address.
5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter.
Thermo Scientific
Reach-In Incubator 3-7
Section 3 Configuration
Configure Mode
Press MODE to light CONFIG
Page 1 of 2
Chart 3-1
Silence
Mode
Run Set
Cal Config
SYS CONFIG
Scroll for Parameters
Enter
36.9 Heat
Temp
Inject
4.9 % CO2
To Configure:
Water Alarm ON/OFF
Audible ON/OFF
Press MODE to return
to RUN mode
Mode
H2O ALM X.X
Press to return to previous parameter
Scroll for Parameters
Enable water alarm (ON)
Press Enter to
Enter
save the setting
Disable water alarm (OFF)
Mode
AUDIBLE XX
Press to return to previous parameter
Scroll for Parameters
Enter
Turn ON audible alarm
Press Enter to save the setting Turn OFF audible alarm
Access Code Low Temp Alarm Limit Temp Relay ON/OFF CO2% Low Limit
Mode ACC CODE XXX
Press to return to previous parameter
Scroll for Parameters
Enter
Numbers increase Press Enter to save the setting
Numbers decrease
Numbers increase
Mode TEMP LO XX.XC Enter
Press to return to previous parameter
Scroll for Parameters
Press Enter to save the setting
Numbers decrease
Mode TMP RLY XXX
Scroll for Parameters
Press to return to previous parameter
Enable relay contacts (ON) during temp alarms
Enter
Press Enter to
save the setting
Disable relay contacts (OFF)
during temp alarms
Mode CO2 LO X.X%
Press to return to previous parameter
Scroll for Parameters
Numbers increase
Enter
Press Enter to
save the setting
Numbers decrease
CO2% High Limit
Mode CO2 HI X.X%
Press to return to previous parameter
Scroll for Parameters
refer to next page
Numbers increase
Enter
Press Enter to
save the setting
Numbers decrease
3-8 Reach-In Incubator
Thermo Scientific
Section 3 Configuration
Configure Mode (continued) Page 2 of 2
Chart 3-1
refer to previous page
To Configure:
CO2 Relay ON/OFF
Mode CO2 RLY XX
Press to return to previous parameter
Scroll for Parameters
Enter
T/C Zero Number
Mode T/C ZR# -XXXX Enter
Scroll for Parameters
Press to return to previous parameter
T/C Span Number
Mode T/C SP# +XXXX Enter
Press to return to previous parameter
Scroll for Parameters
Enable relay contacts (ON) during CO2 alarms
Press Enter to save the setting Disable relay contacts (OFF) during CO2 alarms
Numbers increase Press Enter to save the setting
Numbers decrease
Numbers increase Press Enter to save the setting
Numbers decrease
CO2 Display ON/OFF RH High ON/OFF
Mode DISP CO2 XX
Press to return to previous parameter
Scroll for Parameters
Mode RH HI XX
Press to return to previous parameter
Scroll for Parameters
Enable CO2 display (ON)
Press Enter to
Enter
save the setting
Disable CO2 display (OFF)
Enable RH High (ON)
Press Enter to
Enter
save the setting
Disable RH High (OFF)
Primary Tank Select
Mode TANK SEL X
Press to return to previous parameter
Scroll for Parameters
Tank 1 prime Press Enter to
Enter
save the setting Tank 2 prime
Gas Guard ON/OFF RS485 Address
Mode GAS GRD XXX
Press to return to previous parameter
Scroll for Parameters
Enter
Mode 485 ADDR XX
Press to return to previous parameter
Scroll for Parameters
Enter
Gas Guard (ON) Press Enter to save the setting
Gas Guard (OFF)
Numbers increase Press Enter to save the setting
Numbers decrease
Thermo Scientific
Reach-In Incubator 3-9
Section 3 Configuration
3-10 Reach-In Incubator
Thermo Scientific
Section 4 Alarms
The Model 3950 Series incubator is equipped with a system that notifies the
user of an alarm condition inside the incubator. All alarms display in the
control panel message center. The following table contains information on all
possible system alarms.
Table 4-1. Alarms
Description No alarm condition exists Temp > Otemp Set point Air Temp Sensor Fault (See Section 4.2) Temperature Controller Failure (See Section 4.2) CO2 Sensor Fault (See Section 4.2) Water Level in Reservoir is Low Inner Door is Open CO2 > CO2 High Tracking Alarm
Message Code SYSTEM OK SYS IN OTEMP AIR SNSR ERR TMP CTRL ERR CO2 SNSR ERR ADD WATER DOOR IS OPEN CO2 IS HIGH
CO2 < CO2 Low Tracking Alarm TEMP < TEMP Low Tracking Alarm Tank 1 is low, switch to Tank 2 (Gas Guard only)
CO2 IS LOW TEMP IS LOW TANK1 LOW
Tank 2 is low, switch to Tank 1 (Gas Guard only) TANK2 LOW
Both tanks are low (Gas Guard only)
TANK 1 and 2 LOW
All alarm delays and ringback times are ±30 seconds.
Delay —-0 min. 0 min. 0 min. 0 min. 0 min. 15 min. 15 min. 15 min. 15 min. 0 min. 0 min. 0 min.
Ringback Relay
—–
—–
15 min. Yes
15 min. No
15 min. YES
15 min. No
15 min. No
15 min. No
15 min. Programmable
15 min. Programmable
15 min. Programmable
—-
No
—-
No
15 min. No
To avoid alarms going off in day-to-day use, some alarms are equipped with a time delay feature. For this to function correctly, the alarm condition must exist for the specified length of time before the message center will display the alarm. This allows for interruptions, such as door openings, to occur without incubator being in a continuous state of alarm.
When an alarm condition exists, press the Silence key to temporarily mute the audible alarm. The message center continues to show the alarm condition. If the alarm condition is not corrected within a specified length of time, the alarm sounds again or “ringback” to remind the user.
When multiple alarm conditions occur, active messages are displayed in the display center one at a time, updating at 5-second intervals. Pressing Silence during multiple alarms causes all active alarms to be muted and to ringback in 15 minutes.
Thermo Scientific
Reach-In Incubator 4-1
Section 4 Alarms
Sensor Fault Alarms
The microprocessor in Model 3950 Series incubators continually scans all available sensors to ensure that they are operating properly. Should an error be detected, the incubator will sound an alarm and display the appropriate message. If such an alarm occurs, contact your local distributor or the Technical Services Department.
4-2 Reach-In Incubator
Thermo Scientific
Section 4 Preventive Maintenance
PREVENTIVE MAINTENANCE Incubators
Your equipment has been thoroughly tested and calibrated before shipment. Regular preventive maintenance is important to keep your unit functioning properly. The operator should perform routine cleaning and maintenance on a regular basis. For maximum performance and efficiency, it is recommended the unit be checked and calibrated periodically by a qualified service technician.
The following is a condensed list of preventive maintenance requirements. See the specified section of the operating manual for further details.
We have qualified service technicians, using NIST traceable instruments, available in many areas. For more information on Preventive Maintenance or Extended Warranties, please contact 5FDIOJDBM4FSWJDFT.
Cleaning and calibration adjustment intervals are dependent upon use, environmental conditions and accuracy required.
Tips for all incubators: · Do NOT use bleach or any disinfectant that has high chlorJOFDPOUFOU · Use sterile, distilled or demineralized water.
· Do not use powdered gloves for tissue cultures.
Preventive Maintenance for.PEFM 3950 Series Incubators
Refer to Action Manual Section
1
Check CO2 tank levels.
Daily Weekly Yearly
9
1
Check CO2 and temperature display versus setpoints, w/ independent
instrument
—
Inspect door latch, hinges and door gasket seal.
9 9
2
Verify and document CO2 and temp calibration, as applicable. See
9
Calibration section.
1 & 5
Perform a complete decontamination procedure. Wipe down interior, shelves and side panels with disinfectant. Drain water reservoir and wipe down with disinfectant. Rinse everything well with sterile distilled water.
Between experiments
More frequent decontamination may be required, depending on use and
environmental conditions
Figure 6-5 *Change CO2 filter (under normal conditions)
9
1
Check operation of the float assembly and water level switch
9
- Qualified service personnel
Thermo Scientific
Reach-In Incubator 4-3
Section 4 Alarms
4-4 Reach-In Incubator
Thermo Scientific
Section 5 Maintenance
Warning If the unit has been in service, turn it off and disconnect the power cord connector before proceeding with any maintenance.
Caution Before using any cleaning or decontamination method not recommended by
manufacturer, users should check with manufacturer that the proposed method
would not damage the equipment.
Cleaning The chamber interior may be cleaned with a general-use laboratory
disinfectant or alcohol.
The cabinet exterior may be cleaned with soap and water or any nonabrasive
commercial glass cleaner. The Thermopane glass door may be cleaned with
commercial glass cleaner or with a solution of ammonia and water.
Warning Alcohol, even a 70% solution, is volatile and flammable. Use it only
in a well ventilated area that is free from open flame. If any component is
cleaned with alcohol, do not expose the component to open flame or other
possible hazard.
Caution Do not use strong alkaline or caustic agents. Stainless steel is corrosion resistant, not corrosion proof. Do not use solutions of sodium hypochlorite (bleach) as they may cause pitting and rust.
Cleaning the Glass Doors
The chamber glass door and the optional independent inner doors may be cleaned using the same disinfectant as used on the incubator interior. It is imperative that they be rinsed with sterile distilled water to remove the disinfectant residue. The doors should then be dried with a sterile soft cloth.
Thermo Scientific
Reach-In Incubator 5-1
Section 5 Maintenance
Cleaning the Glass Doors (continued)
Some precautions in the cleaning and care of the incubator glass doors: Moisture leaches alkaline materials (sodium, Na) from the surface of the glass. Evaporation of the moisture concentrates the alkaline and may produce a white staining or clouding of the glass surface. Cleaning chemicals with a PH above 9 accelerate the corrosion process. Therefore, it is very important to rinse and dry the glass doors after cleaning.
There is no simple method for repairing corroded glass. In most cases, the glass must be replaced.
Cleaning the Humidity Periodic cleaning of the humidity reservoir beneath the incubator floor is
Reservoir
recommended in order to remove bacteria, fungi or any accumulated mineral deposits.
1. Turn off and/or disconnect the water supply to the humidity reservoir.
2. Attach a drain fitting and hose to the drain port on the back of the unit. Run the hose to an appropriate drain.
3. With an adjustable wrench, remove the brass overflow fitting located at the rear of the humidity reservoir. Allow reservoir to drain. Reservoir holds approximately 4 gallons (15.4 L).
4. Disassemble the hose bracket from the incubator floor.
5. Remove the 8 screws securing the floor. Using the circular black plastic grommet, slide the floor outward to expose the humidity reservoir.
6. Remove any remaining water and sediment from the reservoir with a clean sponge.
7. Clean the reservoir with a general-use laboratory disinfectant. Rinse well with sterile distilled water and wipe dry with a clean cloth.
Note On reassembly, brass overflow fitting must be checked for leaks.
Caution Use of chlorinated water, or decontamination products containing chlorine, will deteriorate the stainless steel and cause rust, voiding the warranty.
Caution Do not use strong alkaline or caustic agents. Stainless steel is corrosion resistant, not corrosion proof.
5-2 Reach-In Incubator
Thermo Scientific
Section 5 Maintenance
Replacing the Power Warning De-energize all potential sources of energy to
this unit and Fuses lockout/tagout their controls. (O.S.H.A. Regulation,
Section 1910-147.)
Warning High voltage is present behind control panel. The remote overtemp
alarm system should be installed only by qualified electrical service
personnel.
There are only two replaceable fuses in the incubator.
1. Turn off the incubator’s power switch and unplug the power cord.
2. Remove the top of the unit to access the fuses.
3. Refer to Figure 6-5 for the location of the two fuses.
4. Install the top cover and return the unit to service. If the fuse(s) blow after restoring power to the incubator, contact the Technical Services Department.
Table 5-1. Fuse Replacement Chart
Fuse Voltage andApplication Manufacturers Part # Amp Rating
115 VAC Accessory Outlet
GMC-1 A
1.0 Amp
230 VAC Accessory Outlet
GMC-500 mA
0.5 Amp
115 VAC Interior Outlet
GMC-2.5 A
2.5 Amp
230 VAC Interior Outlet
GMC-1.5 A
1.5 Amp
Rupture Speed Time-Lag Time-Lag Time-Lag Time-Lag
IEC Letter Code T T T T
Discarding/Removing Caution Federal regulations require that doors be removed
from incubators Incubator from Service before units are removed from service
or discarded.
Caution When operating this unit under high temperatures and/or relative
humidity (RH), caution must be observed when the unit is shut down,
unexpectedly or by choice. Condensate can form on the interior of the chamber
as it cools. This includes the heating element surfaces. Heavy condensate can
bridge the ceramic insulators and cause a temporary electrical short to
ground. This electrical circuit is protected by a circuit breaker. However, to
avoid this contingency if these conditions occur, open the door of the unit
and allow it to stand open until the water vapor disperses.
Thermo Scientific
Reach-In Incubator 5-3
Section 5 Maintenance
5-4 Reach-In Incubator
Thermo Scientific
Section 6 Factory Options
Instructions for connecting the incubator to external equipment follow.
Connect the Remote Alarm Contacts
A set of relay contacts is provided to monitor alarms through a RJ-11 telephone style connector on the back of the unit. Refer to Figure 6-3 for the location of the alarm connector. The 12-foot telephone cord (P/N 190388) and RJ11-to-screw terminal conversion box (190392) are available through the Technical Services Department.
The remote alarm provides a NO (normally open) output, an NC (normally closed) output and COM (common). Refer to Figure 6-1.
The contacts will trip on a power outage or an overtemperature condition. The contacts may also be programmed to trip or not trip on temperature alarms and CO2 alarms. See Section 3, Configuration Mode.
(NO) green wire
(COM) yellow wire
(NC) red wire
(not connected) black wire
Alarm Contacts Junction Box
phone cable to incubator alarm contacts RJ-11 connector
All outputs shown in alarm condition
Figure 6-1. Remote Alarm Outputs
Stored product should be protected by an activated alarm system capable of initiating a timely response 24 hours/day. These alarms provide an interconnect for centralized monitoring.
CAUTION
Contains Parts and Assemblies Susceptible to Damage by
Electrostatic Discharge (ESD)
Thermo Scientific
Reach-In Incubator 6-1
Section 6 Factory Options
Connect the RS485 Interface (P/N 190523)
All incubator models can be purchased with the RS485 communications option. This option allows the incubator to be directly connected to a Model 1535 alarm system without the use of a communications module. A junction box is provided with each RS485 option. Refer to Figure 6-2 for wiring details. Figure 6-3 shows the location of the RS485 connector on the back of the incubator.
To allow the incubator and the 1535 to communicate, an address must be allocated on the 1535. Refer to Section 5.8 of the 1535 operating manual. The same address number must be assigned to the incubator. Refer to Section 3, Configuration Mode, of this manual.
Figure 6-2. RS485 Interface Wiring
2 CO2 Inject 15 p.s.i. Regulated (103.4 kPa)
1 CO2 Inject 15 p.s.i. Regulated (103.4 kPa)
RS 485 AUXILIARY HOLE FOR CABLES
ALARM CONTACTS 30V MAX / 1A MAX
Figure 6-3. RS485 Connector Location on Unit Back
6-2 Reach-In Incubator
Thermo Scientific
Section 6 Factory Options
Connect Analog Output Boards (P/N 190512, 190543, & 190544)
Warning The electronics section contains hazardous voltages. Only qualified
electrical service personnel should access this area.
The analog output board is an option that allows the incubator to output
analog signals representing the air temperature and CO2 content of the
incubator interior. There are three different analog output board options
available: 0-1 V, 0-5 V or 4-20 mA signals. Negative display readings will
output 0 V. The outputs do not have isolated grounds. See Table 6-1 for output
specifications of the three boards.
Table 6-1. Analog Output Board Specifications
190512 4-20 mA Output Scaling 4-20 mA Equals
190544 0-1 V Output Scaling 0-1 V Equals
190543 0-5 V Output Scaling 0-5 V Equals
Temperature 0.0-100.0°C
0.0-100.0°C
0.0-100.0°C
CO2
0.0-100.0% CO2
0-100.0% CO2
0-100.0% CO2
For the 0-1 V and 0-5 V boards, the recording device must supply a load
/=1000 ohm. For the 4-20 mA board, the recorder must supply a load of </=100 ohm.
To wire in the analog output board, the shielded 22 gauge, 3-conductor wire, Part # 73041, is recommended, maximum length 50 ft (15.2 m). This wire is also readily available from other vendors including Alpha, Part #2403, and Deerborn, Part # 972203.
Caution Accuracy of the output at the board terminal strip to the incubator display is ±1 unit. There is no calibration from the incubator. Calibration to the incubator display must be at the instrument connected to the output board.
To access the analog board: 1. Turn off the incubator and disconnect it from
the power source. 2. Remove the top of the incubator. 3. Locate the Analog
Output board.
(continued)
Thermo Scientific
Reach-In Incubator 6-3
Section 6 Factory Options
4. Strip the ends of the conductor and wire it to the appropriate terminals of connectors J2 on the analog board. Refer to Figure 6-4.
Figure 6-4. Terminals on Analog Board
Blower Motor
CO2 Sensor
CO2 Valve
Alarm Contacts RS 485 Connection Fuse 1 amp ( .5 amp 3951)
CO2 Filter
Forma
Fuse 2.5 amp (1.5 amp 3951)
Analog Board Microprocessor Board Power Supply
Figure 6-5. Route Wires
5. Route the wires through the auxiliary hole located on the back of the unit. See Figure 6-3 and Figure 6-5.
6. When wiring is completed, replace the unit top and return the unit to service.
CO2 Gas Guard Warning CO2 can be lethal in high concentrations. Refer to the
warnings (P/N 1900000) in Section 1 of this manual.
The 3950 Series incubators can be equipped with a built-in Gas Guard system
that will operate with a CO2 gas supply. The Gas Guard uses two pressure
switches to continuously monitor the pressures of two independent CO2 supplies
and automatically switches from one supply to the other when the supply
pressure drops below 10 psig (0. 690 bar). Do not use the Gas Guard with
multiple incubators.
6-4 Reach-In Incubator
Thermo Scientific
Section 6 Factory Options
CO2 Gas Guard (continued)
The CO2 gas supplies must be equipped with two-stage pressure regulators with gauges. The high pressure gauge at the tank should have a 0-2000 psig range and the low pressure gauge should have a 0-30 psig range. The gas supply to the incubator must be maintained at 15 psig (1.034 bar). Gas pressures below 15 psig will cause nuisance alarms to occur on incubators equipped with the built-in Gas Guard.
Connect CO2 Gas Supplies
The CO2 inlets for the Gas Guard are located on the back of the unit. Using 1/4″ ID tubing, connect one of the CO2 supply tanks to the fitting labeled CO2 Inlet #1 Tank. Connect the second CO2 supply tank to the fitting labeled CO2 Inlet #2 Tank. Install 3/8″ hose clamps to secure the 1/4″ ID tubing to the fittings. Refer to Figure 6-6.
2 CO2 Inject 15 p.s.i. Regulated (103.4 kPa)
1 CO2 Inject 15 p.s.i. Regulated (103.4 kPa)
RS 485 AUXILIARY HOLE FOR CABLES
ALARM CONTACTS 30V MAX / 1A MAX
Figure 6-6. CO2 Connections on Unit Back
De-activate Gas Guard
The built-in Gas Guard is turned ON when shipped from the factory. The tank selected (Tank Sel) has been specified as Tank 1. Refer to Section 3 Configuration, to de-activate the Gas Guard or change Tank Sel from #1 to #2. If the Gas Guard system is not used, the incubator functions normally by supplying CO2 from the supply tank connected to Inlet #1.
Operation of CO2 Gas Guard
With the Gas Guard in operation, the incubator uses the gas supplied by the tank connected to Inlet #1 until pressure drops below 10 psig (68.95 kPa). At this time, the Gas Guard automatically switches to gas supplied through CO2 Inlet #2.
In addition, the incubator automatically changes Tank Sel in Configuration Mode from 1 to 2 to indicate that the incubator is now using gas supplied through Inlet # 2. If the gas supply to Inlet #1 is replenished, the incubator continues to operate using the gas supplied through Inlet #2 unless the operator changes Tank Sel from #2 to #1 through Configuration Mode. Refer to Section 3, Configuration.
Thermo Scientific
Reach-In Incubator 6-5
Section 6 Factory Options
Operation of CO2 Gas Guard (continued)
Audible and visual alarms occur on the control panel when the gas guard
switches from one supply to the other. The audible alarm sounds until the
operator presses the Silence key on the control panel. The visual alarm in the
Message Center reads Tank 1 Low’ while the audible alarm is sounding, but the message is removed when the operator presses the Silence key. However, the
Tank Low’ indicator on the control panel stays lit until the condition is
resolved. The unit operates normally.
Caution Both the audible and visual alarms described above do not ring back once the Silence key is pressed.
If the Gas Guard system does not detect an adequate gas supply at CO2 Inlet #1 or Inlet #2, a visual and audible alarm again occurs on the control panel. The visual alarm in the Message Center reads `Tank 1&2 Low’. The audible alarm continues to ring until the Silence key is pressed. The audible alarm rings back every 15 minutes after the alarm is silenced, if the Gas Guard continues to detect that both gas supply pressures are below 10 psig (68.95 kPa).
Inner Doors
Align the shelves and shelf channels with each of the inner doors to facilitate introduction and removal of trays. Clean these doors with the same care as the single door. See Section 5.
Shaker Support Shelves
Shaker support shelves are reinforced and secured to the walls of the incubator. They have a load limit of (1) shaker or 200 pounds per shelf, one shelf being the floor of the unit. The shaker platform limit is 50 pounds. Shakers must not exceed 250 rpm when used inside this incubator. For shaker power connection, an internal outlet in the upper right corner of the back wall is installed. Casters are installed at the factory for moving the cabinet to the desired location. After the unit is in place and prior to operation, the casters must be removed. The large rubber vibration feet, factory installed, are positioned correctly for operation. Do not adjust.
Caution Any equipment placed inside the chamber must be rated for unit operating temperature and humidity.
6-6 Reach-In Incubator
Thermo Scientific
Section 7 Specifications
- Specifications are based on nominal voltages of 115 V or 230 V in ambient temperature of 22°C to 25°C.
Temperature Control . . . . . . . . . . . .±0.1°C Microprocessor PID Setpoint . . . . . . . . . . . .Digital Touch Pad, 0.1°C Range . . . . . . . . . . . .+5°C above ambient to 60°C Uniformity . . . . . . . . . . . . . . . . . .±0.3°C @ +37°C Tracking Alarm. . .User programmable (low) indicator Overtemp . . .Tracking, user programmable, action and indicator Display . . . . . . . . .Digital, LED, 0.1°C increments
Temperature Safety Type . . .Extreme temperature safety, action and indicator Sensor . . .Thermostat, independent of temperature control system Indicator . .Message center, audible and visual alarms
Relative Humidity Selectable Ranges .Medium (>80%), High (>90%)
CO2 Control: . . . . . . . . . . . .±0.1% microprocessor PID Sensor: . . . . . . . . . . . . . . . . .Thermal conductivity Readability: . . . . . . . . . . . . . . . . . . . . . . . . . .0.1% Range: . . . . . . . . . . . . . . . . . . . . . . . . . . .0 to 20% Inlet Pressure: . . . . . . . . . . . . .15 PSIG (±5 PSIG) Display: . . . . . . . . .Digital LED, 0.1% increments
Thermo Scientific
Reach-In Incubator 7-1
Section 7 Specifications
7-2 Reach-In Incubator
Shelves Dimensions . . .30.62″ W x 25.81″ F-B (77.78 cm x 65.56 cm)
Construction: . . . . . .Solid stainless steel, 2B finish Surface Area . . . .
. .5.4 sq. ft (0.51 sq. m) per shelf Max per Chamber: . . . . .145.8 sq. ft
(13.55 sq. m) Standard: . . . . . . . . . . . . . . . .5 (shipped with unit)
Maximum: . . . . . . . . . . . . . . . . . . .27 (maximum) Clearance: . . . .
. . .Adjustable on 2″ (5 cm) centers
Construction Interior volume . . . . . . . . . . . .29 cu ft (823 liters)
Interior . . . . . . . . . . . .304 stainless steel, 2B finish Exterior . . .
. . . . . . . . . . .18-gauge cold rolled steel Exterior Door . .Heated,
triple pane tempered glass Outer Door Gasket . . . . . . . . . . . . . .Molded
vinyl Insulation . . . . . . . . . . . . . . . . . . . . . . .2″ fiberglass
Fittings Access Port . . .2.4″ (6.1 cm) ID, one port per side CO2 Inlet . . .
. . . . . . . . . . . . . . . . . . .1/4″ barbed Sample Port . . . . . . . . .
.Front mount, 1/4″ barbed Water Inlet . . . . . . . .1/4″ MPT for automatic
fill Overflow Drain . . . . . . . . . . . . . . . . . . .3/8″ MPT Condensate
Drain . . . . . . . . . . . . . . . . . .3/8″ FPT
Electrical Model 3950:100-120 VAC, 50/60 Hz, 1 PH, 10.0 FLA Operating Range,
including fluctuations – 90-125 V, 50-60 Hz, 15 A Breaker power switch Model
3951:200-230 VAC, 50/60 Hz, 1 PH, 6.0 FLA Operating Range, including
fluctuations – 180-250 V, 50-60 Hz, 8 A Breaker power switch Power Switch . .
. . . . . . . . . . .2-pole circuit breaker Accessory Outlet . . . Voltage
equal to the cabinet input, 75 W max., 0.5 mA leakage current Interior Outlet
. . . . Voltage equal to cabinet input, 230 W max, 0.5 mA leakage current
Remote Alarm Contacts . . . Deviation of temperature, CO2 & power. N.O. & N.C.
Thermo Scientific
Section 7 Specifications
Unit BTU Output 115V / 230V: . . . . . . . . . . . . . . . . . . . . .510 BTUH
(150 W)
Dimensions Exterior . . . . . . . . . . . . . . . . .38.0″ W x 80.0″ H x 33.0″
F-B . . . . . . . . . . . . . . . . . . . . . .(96.5 cm x 203.2 cm x 83.8 cm)
Interior . . . . . . . . . . . . . . . . . .31.0″ W x 60.0″ H x 27.0″ F-B . .
. . . . . . . . . . . . . . . . . . . .(78.7 cm x 152.4 cm x 68.6 cm)
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .500 lbs.
(226.8 kg)
Safety Specifications Altitude: . . . . . . . . . . . . . . . . . . . . . . .
. . . .Up to 2,000 meters Temperature: . . . . . . . . . . . . . . . . . . . .
. . . . . . . . .5°C to 40°C Humidity . . .Maximum 80% RH for temperatures up
to 31°C, decreasing linearly to 50% RH at 40°C non-condensing. Mains Supply
Fluctuations . . .not to exceed ±10% of nominal voltage Installation Category
II Pollution Degree 2 Class of Equipment I
1 Installation category (overvoltage category) defines the level of transient
overvoltage which the instrument is designed to withstand safely. It depends
on the nature of the electricity supply and its overvoltage protection means.
For example, in CAT II which is the category used for instruments in
installations supplied from a supply comparable to public mains such as
hospital and research laboratories and most industrial laboratories, the
expected transient overvoltage is 2500 V for a 230 V supply and 1500 V for a
120 V supply.
2 Pollution degree describes the amount of conductive pollution present in the
operating environment. Pollution degree 2 assumes that normally only non-
conductive pollution such as dust occurs with the exception of occasional
conductivity caused by condensation.
Thermo Scientific
Reach-In Incubator 7-3
Section 7 Specifications
7-4 Reach-In Incubator
Thermo Scientific
Thermo Scientific
Section 8 Spare Parts
Part Number 230135 230158 230120 230106 1900621 170164 600210 250118 360157 1900471 290138 191563 290090 400119 190641 980055 515080 227083 770001 600034 950013 505071 190012 180006 197075 190467 380177
Description 1 amp fuse (3950 accessory outlet) 2.5 amp fuse (3950 interior outlet) 0.5 amp fuse (3951 accessory outlet) 1.5 amp fuse (3951 interior outlet) Blower motor service kit Motor capacitor 3 MFD, 370 VAC 300 watt wire- wound heater CO2 solenoid 12 V DC Door switch Add Water float switch with pins Temperature control sensor Microprocessor board CO2 sensor Power supply Gas guard manifold with pressure switches & valve Float valve assembly, 100 PSI Leveler, 2″ diameter Door latch assembly Inline CO2 filter 0.375″ hose clamp 1/4″ vinyl tubing for CO2 connection Stainless steel shelf Stainless steel shelf channel 0 – 60°C chart paper, 1 pen recorder 0 – 100°C chart paper, 2 pen recorder 3/8″ MPT x 3/8″ tube P-trap 3/8″ FPT brass elbow
Reach-In Incubator 8-1
Section 8 Spare Parts
8-2 Reach-In Incubator
Thermo Scientific
Section 8 Spare Parts
Thermo Scientific
Reach-In Incubator 8-3
Section 8 Spare Parts
8-4 Reach-In Incubator
Thermo Scientific
Section 8 Spare Parts
Thermo Scientific
Reach-In Incubator 8-5
Section 8 Spare Parts
8-6 Reach-In Incubator
Thermo Scientific
Section 8 Spare Parts
Thermo Scientific
Reach-In Incubator 8-7
Section 8 Spare Parts
8-8 Reach-In Incubator
Thermo Scientific
Section 8 Spare Parts
Thermo Scientific
Reach-In Incubator 8-9
Section 8 Spare Parts
8-10 Reach-In Incubator
Thermo Scientific
Section 8 Spare Parts
Thermo Scientific
Reach-In Incubator 8-11
Section 8 Spare Parts
8-12 Reach-In Incubator
Thermo Scientific
Section 9 Electrical Schematics
Thermo Scientific
Reach-In Incubator 9-1
Section 9 Electrical Schematics
9-2 Reach-In Incubator
Thermo Scientific
Section 9 Electrical Schematics
Thermo Scientific
Reach-In Incubator 9-3
Section 9 Electrical Schematics
9-4 Reach-In Incubator
Thermo Scientific
Thermo Scientific
THERMO FISHER SCIENTIFIC STANDARD PRODUCT WARRANTY
The Warranty Period starts two weeks from the date your equipment is shipped
from our facility. This allows for shipping time so the warranty will go into
effect at approximately the same time your equipment is delivered. The
warranty protection extends to any subsequent owner during the first year
warranty period.
During the first year, component parts proven to be non-conforming in materials or workmanship will be repaired or replaced at Thermo’s expense, labor included. The Watlow EZ-ZONE PM controller is covered for one additional year for repair or replacement (parts only), provided the unit has not been misapplied. Installation and calibration are not covered by this warranty agreement. The Technical Services Department must be contacted for warranty determination and direction prior to performance of any repairs. Expendable items, glass, filters and gaskets are excluded from this warranty.
Replacement or repair of components parts or equipment under this warranty shall not extend the warranty to either the equipment or to the component part beyond the original warranty period. The Technical Services Department must give prior approval for return of any components or equipment. At Thermo’s option, all non-conforming parts must be returned to Thermo Fisher Scientific postage paid and replacement parts are shipped FOB destination.
THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, WHETHER WRITTEN, ORAL OR IMPLIED. NO WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE SHALL APPLY. Thermo shall not be liable for any indirect or consequential damages including, without limitation, damages relating to lost profits or loss of products.
Your local Thermo Sales Office is ready to help with comprehensive site preparation information before your equipment arrives. Printed instruction manuals carefully detail equipment installation, operation and preventive maintenance.
Section 10 Warranty Information
Reach-In Incubator 10-1
If equipment service is required, please call your Technical Services Department at 1-800-438-4851 (USA and Canada) or 1-740-373-4763. We’re ready to answer your questions on equipment warranty, operation, maintenance, service and special application. Outside the USA, contact your local distributor for warranty information.
ISO 9001
Rev. 5 8/13
REGISTERED
Section 10 Warranty Information
10-2 Reach-In Incubator
THERMO FISHER SCIENTIFIC INTERNATIONAL DEALER WARRANTY
The Warranty Period starts two months from the date your equipment is shipped
from our facility. This allows for shipping time so the warranty will go into
effect at approximately the same time your equipment is delivered. The
warranty protection extends to any subsequent owner during the first year
warranty period. Dealers who stock our equipment are allowed an additional six
months for delivery and installation, provided the warranty card is completed
and returned to the Technical Services Department.
During the first year, component parts proven to be non-conforming in materials or workmanship will be repaired or replaced at Thermo’s expense, labor excluded. The Watlow EZ-ZONE PM controller is covered for one additional year for repair or replacement (parts only), provided the unit has not been misapplied. Installation and calibration are not covered by this warranty agreement. The Technical Services Department must be contacted for warranty determination and direction prior to performance of any repairs. Expendable items, glass, filters, reagents, tubing, and gaskets are excluded from this warranty.
Replacement or repair of components parts or equipment under this warranty
shall not extend the warranty to either the equipment or to the component part
beyond the original warranty period. The Technical Services Department must
give prior approval for return of any components or equipment. At Thermo’s
option, all non-conforming parts must be returned to Thermo postage paid and
replacement parts are shipped FOB destination.
THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, WHETHER
WRITTEN, ORAL OR IMPLIED. NO WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE SHALL APPLY. Thermo shall not be liable for any indirect or
consequential damages including, without limitation, damages relating to lost
profits or loss of products.
Your local Thermo Sales Office is ready to help with comprehensive site preparation information before your equipment arrives. Printed instruction manuals carefully detail equipment installation, operation and preventive maintenance.
Contact your local distributor for warranty information. We’re ready to answer your questions on equipment warranty, operation, mainte-
nance, service and special application.
ISO
9001
Rev. 5 8/13
REGISTERED
Thermo Scientific
thermoscientific.com
© 2021 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are
the property of Thermo Fisher Scientific and its subsidiaries. Specifications,
terms and pricing are subject to change. Not all products are available in all
countries. Please consult your local sales representative for details.
Thermo Fisher Scientific (Asheville) LLC 401 Millcreek Road Marietta, Ohio
45750 United States
A Thermo Fisher Scientific Brand
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>