YODN Hyper S330 Metal Halide Lamp
Product Information
The YODN illuminator model Hyper S330 is a precision optical equipment designed for use with a microscope. The device emits UV light and comes with a liquid light guide and adapter that need to be attached to the microscope before powering it on. The device is not waterproof and should be handled with care to avoid sudden impact. The device has ventilation openings on its sides to release the heat generated during operation. The bend radius of the liquid light guide must be larger than 60 mm to prevent damage. The device should be installed on a sturdy, level table or bench and must be placed on a flat surface to avoid blocking the ventilation openings. The device comes with a power cord provided by the manufacturer.
Product Usage Instructions
Before operating the YODN illuminator model Hyper S330, ensure that you have studied the user manual thoroughly. Keep the user manual in an easily accessible place near the work desk for future reference. Follow the safety precautions listed below:
- Do not power on the device without attaching the liquid light guide and adapter to the microscope.
- Avoid looking directly into the light without protection as it may damage your eyes.
- Avoid fixing the light output in one spot on flammable objects for a long time as it may cause fire.
- Ensure that the device is powered off and unplugged from the outlet before replacing the lamp. Wait until the lamp and its replacement cover have cooled down completely (more than 30 minutes) to avoid accidental electrical shock or burn.
- Install or place the illuminator unit on a sturdy, level table or bench.
- Install this device on a flat surface to avoid blocking the ventilation opening on the sides of the device. Do not place it on a soft surface, i.e. paper or cloth, etc.; this could cause fire.
- Ensure leaving sufficient space for ventilation around this device. Never allow metallic objects to enter from the ventilation opening; this could cause fire or electrical shock.
- Secure sufficient space (larger than 10 cm) around this device to release the heat generated during operation.
- The bend radius of the liquid light guide must be larger than 60 mm to prevent damage. Always use the power cord provided by the manufacturer to ensure that the device operates safely.
- Operate the device in an environment free from direct sunlight, dust, steam, or vibration.
- Keep highly flammable organic solvent away from the illuminator when it is in use.
Read the instruction manual of the designated microscope to ensure that the
microscope and illuminator work together perfectly. Handle the device
carefully and avoid sudden and severe impact.
This user’s manual is made for the YODN illuminator model Hyper S330. To
ensure operation safety, maximize device performance, and to familiarize
yourself fully with the operation of this illuminator, the manufacture
recommends that you study this user’s manual thoroughly before operating this
device. Please retain this user’s manual in an easily accessible place near
the work desk for future reference.
SAFETY PRECAUTIONS
- Never power on the device (turn on the lamp) without the liquid light guide and adapter to the microscope attached. UV light is designed to emit from this illuminator during lamp operation; do not directly look into the light without protection, it may damage your eyes. On the other hand, please avoid fix the light output in a fix spot (especially on flammable objects) for a long time as it may cause fire.
- To replacing the lamp, please make sure the device power is off and unplug the power cord from the outlet, and wait until the lamp and its replacement cover cooled down completely (more than 30 minutes) to avoid accidental electrical shock or burn.
- Install or place the illuminator unit on a sturdy, level table or bench.
- Install this device on a flat surface to avoiding blocking the ventilation opening on the sides of device. Do not place it on a soft surface, i.e. paper or cloth, etc.; this could cause fire.
- Please ensure leaving sufficient space for ventilation around this device. Never allow metallic objects to enter from the ventilation opening; this could cause fire or electrical shock.
- It’s recommended to secure sufficient space (larger than 10 cm) around this device to release the heat generated during operation.
- To prevent the liquid light guide (optional accessory) from damage, the bend radius of the liquid light guide must be larger than 60 mm.
- Always use the power cord provided by the manufacture to ensure the device operated under its original safety design and performance; fail to do so may void the manufacture product warranty. Keep sufficient distance from the power cord to all possible heat sources and avoid direct contact.
- Do not power on the device when the lamp replacement cover is open; this could result in electrical shock, burn, illuminator failure, or light guide damages.
- Please ensure to connect the device power cord to a grounded outlet. Fail to do so may void the intended electrical safety design and device performance.
- Do not touch the power plug during a thunderstorm to avoid accidental electrical shock.
Getting Ready
- This user’s manual only illustrate to the illuminator. Read also the Instruction Manual of the designated microscope to ensure the microscope and illuminator working together perfectly.
- This device is a precision optical equipment; please handle it carefully and avoid sudden and severe impact. This device is not waterproof.
- Do not operate this device where it is subjected to direct sunlight, dust, steam, or vibration.
- Highly flammable organic solvent may cause fire. Keep it away from the illuminator when it is in operation and soon after the illuminator is turned off.
- Do not twist the liquid light guide (optional accessory) excessively to prevent damaged.
- Make sure the device is powered off, unplug the power cord from the electrical outlet, and remove the liquid light guide in the event of transporting the device.
- Please avoid touch the ventilation opening area during the device operation and immediately after power off as its temperature might be high and may cause burns.
- Please avoid turn the device on and off in short period (this device has protective control) as the lamp is not designed to do so and it will shorten the lamp life.
Maintenance and Storage
- Clean the device periodically (every 200 hours as a guide). Before cleaning, unplug the power plug from the outlet and wipe it with a fluff-free soft dry cloth.
- Wipe clean with the soft cloth slightly containing the diluted detergent and use the clean air to blow the dust off if the illuminator and the liquid light guide are very dirty.
- Never disassemble/remodel anything other than those described in this user’s manual. This may cause the optical performance deterioration or incidents; unauthorized disassemble the device will result in product warranty void.
- When the device is idle, please unplug the power cord from electrical outlet. It is recommend to use the dust prevention cover to cover the device if it’s not used on a regular basis or storage; however, please make sure the device is completed cooled off before applying any cover on it.
- Please replace a new lamp when the lamp energy level drops using the replacement lamp authorized by the manufacture. The mercury lamp design is different from the fluorescence, which contains a high pressure gas within the tube. If the lamp is used beyond its designed lifetime, the glass tube may burst due to the accumulated strain. The used mercury lamp must be disposed of as industrial waste and only to the authorized recycle professionals. Before disposing of this product, be sure to follow your local government regulations and rules.
In the event of lamp burst, please follow the procedures below.
- Should the lamp burst, unplug the power cord from the outlet and leave that place while keeping the place ventilate for at least 30 minutes.
- Wait until the lamp and the lamp house have cooled down completely, then use a protective glove to collect the remaining glass with packing tape and store in a sealable package.
- The collected shattered glass and tools used must be sealed completely in a nonmetallic container, and ask the waste disposer to dispose them as it will contain mercury.
- Should you have inhaled mercury steam, consult the doctor immediately and follow his/her instructions.
Device Introduction
Front
Back
Optional Components List
Special
Collimator (Adapter to Microscope):
Device Specifications
Input Voltage | AC 100~240V,Full voltage input |
---|---|
Frequency | 50 / 60 Hz |
Power Consumption | Standby: 5W |
Power On (Lamp Ignition): 165W Power Off(with Cooling): 12.5W
Light Source| YODN AC132W for Hyper S330
Color Temperature (Average): 8000K
Drive mode| AC Type
Light Output Adjustment| 9-level Brightness Adjustment- 0% ~ 100%
Protective Function| Lamp Re-ignition Protection Current Overload Protection
Lamp Ignition Fail Lock
Lamp Hour Recorder| Omron Time Counter
Cooling Function| Forced Electrical Fan to exhaust
Internal Filter| UV / IR Cut Filter
Fuse Type| Glass Tube Fuse -AC250VM 3A (Φ5.2 x 20mm)
Operating Temp.| 10°C ~ 50°C / 20% ~ 80%RH (No condensation)
Dimensions (H x W x D)| 220 x 130 x 365 mm (Except for raised parts)
Weight| 5.5 Kgs
Operation Instruction
Troubleshooting Guide
Under certain conditions, performance of the unit may be adversely affected by factors other than defects. If the problems occur, please review the list and take remedial action as needed. If you cannot solve the problem after checking the list, please contact your sales representative for assistance.
Trouble | Cause | Remedy |
---|---|---|
The power of the equipment can not be turned on | Power cord and plug are not | |
connected properly | Connect them correctly | |
The built-in fuse is blown | Replace the fuse with a new one | |
The lamp cannot be turned on, the cooling fan isn’t start | The cooling fan or | |
the power supply has failed | Contact your sales representative |
The lamp cannot be turned on, the cooling fan is operating ,the lamp switch
flashing red light| The lamp has failed or life has expired| Replace with a
new lamp
The lamp is not mounted properly| Mount the lamp properly
The lamp can be turned on but is unstable| The lamp has failed or life has
expired| Replace with a new lamp
The lamp can be turned on but is dark| The lamp is not mounted properly| Mount
the lamp properly
The light guide is not mounted properly| Mount the light guide properly
The dial value of light volume adjustment is close| Rotate the light volume
adjustment dial toward 100% by each level
The lamp turns off during use| The lamp has failed or life has expired|
Replace with a new lamp
The built-in fuse is blown| Replace the fuse with a new one