YODN Hyper E340 Broadband Excitation Illuminator for Fluorescence Microscope User Manual
- June 9, 2024
- YODN
Table of Contents
Broadband Excitation Illuminator for Fluorescence
Microscope
Hyper E340 User Manual
HYPER E340 User Manual (Edit 2023/03)
Product Introduction
The new generation Hyper E340 can effectively increase the LED light source
intensity by 22% to 143% or more.
The Hyper E340 lighting system provides wide-range LED illumination with a
spectral coverage range of 350 nm to 730 nm, which can satisfy the
fluorescence commonly used in most experimental research, such as DAPI,
GFP/FITC, cherry/Texas Red, Cy3, Cy5 (or Cy7). It can help you work
efficiently and energy-starvingly.
The Hyper E340 lighting system comes with a comprehensive microscope adapter
that can be applied to most existing microscopes or connected to a 3.0mm
liquid light guide. The on/off and light source intensity adjustment of the
light source output are controlled by the electronic control box and adjusted
by the dimming knob.
In addition, the LED has a long lifespan, so you don’t have to worry about
replacing bulbs. YODN Hyper E340 is the best choice for fluorescence
microscopy research.
Cautions and Warnings:
Simple operation to ensure the normal operation of the light box.
Safety instructions:
Before operating the Hyper E340 lighting system, please read and follow all
safety instructions provided. Failure to comply with safety instructions may
result in fire, electric shock, or personal injury and may cause equipment
damage or malfunction. Please keep all safety instructions.
Safety definitions:
Danger: Statements indicating a risk of death, serious injury, equipment
damage, and procedures.
Warning: Statements indicating a risk of personal injury or procedures.
Notice: Statements indicating a risk of equipment damage or procedures.
Safety items:
Warning: Do not use unapproved power sources. Please use the YODN-provided
power source for the Hyper E340 lighting system. The range of the external
voltage input is 100~240 VAC, the voltage output is 12 VDC, and the maximum
current output is 15.0 A.
Danger: Do not stare at UV bright light. The light source output of the
lighting system includes UV light, which may cause harm to the eyes or skin.
Do not stare at the light source output. To avoid exposure of eyes and skin to
the beam, it is essential to wear goggles and protective clothing. The
brightness of the lighting system is higher than that of most commercial
lighting devices and is mainly used for coupling with microscopes or other
biological analysis instruments.
Notice: Do not open the casing of the light box or change external
devices in any way. Opening the lighting system casing will void the product
warranty because there are no repairable or replaceable parts outside the
lighting system.
Notice: Do not place liquid containers on top of the light box. The
spilled liquid may damage the light box.
Notice: Do not drop or suddenly apply external force to the light box.
The vibration caused by falling onto a hard surface, external impact, or
collision may damage or displace the internal components of the lighting
system, causing malfunction.
Note: The liquid light guide should not be subjected to impact or bending
with a minimum bending radius greater than the standard value.
RISK GROUP 3
WARRING IR emitted from this product. Do not look at operating lamp
CAUTION UV emitted from this product . Eye or skin irritation may result from
exposure. Use appropriate shielding.
Illuminator System Accessories
- DC Power Supply: GST160A12-R7B(YODN P/N:H006-0114311)。
- AC Power Cord (optional according to region):
(1)H006-0067312 – Taiwan
(2)H006-0067311 – North America
(3)H006-0068311 – Europe
(4)H006-0069311 – U.K.
(5)H006-0070311 – West Germany
(6)H006-0070311 – South Korea
(7)H006-0071312 – China/Australia
(8)H006-0372311 – Japan
3.2 Adaptor Optional according to the microscope model):
Microscope Model | P/N |
---|---|
Adaptor for Nikon | H600-0024310 |
Adaptor for Nikon-T | H600-0025310 |
Adaptor for Zeiss | H600-0026310 |
Adaptor for Leica | H600-0027310 |
Adaptor for Motic | H600-0028310 |
Adaptor for Evident (Olympus) | H600-0029310 |
3.3-Liquid light guide:Diameter (effective inner diameter)φ3 mm
Description | P/N |
---|---|
Φ3mm*1500mm LLG,Series380,STD | H600-0009311 |
Φ3mm*1800mm LLG,Series380,STD | H600-0010311 |
Φ3mm*3000mm LLG,Series380,STD | H600-0011311 |
LLG Adapter | H600-0030310 |
3.4 Collimator:
Description | P/N |
---|---|
Collimator for Evident (Olympus)(Φ3LLG) | H600-0015311 |
Collimator for Leica(Φ3LLG) | H600-0016311 |
Collimator for Zeiss(Φ3LLG) | H600-0017311 |
Collimator for Nikon(Φ3LLG) | H600-0018311 |
Collimator for Motic(Φ3LLG) | H600-0020310 |
Collimator for Nikon-T(Φ3LLG) | H600-0022310 |
Hyper E340 Function Introduction:
Illuminator:
Before operating the device, please ensure that the optical fiber is correctly
installed into the optical fiber port. Installation method: remove the
protective covers on both sides of the optical fiber, loosen the hand-twist
screw, and slide the optical fiber completely into the optical fiber socket
without interference. After inserting the optical fiber to the bottom
position, gently tighten the hand-twist screw to secure the optical fiber
without loosening, to prevent over-tightening and damaging the optical fiber.
The bending of the optical fiber must not exceed the specifications of the
minimum bending radius (80mm or 3.2 inches). Excessive bending of the optical
fiber may cause permanent deformation and reduce the amount of light output.
Note: Before turning on the light, please ensure that the protective covers on
both sides of the liquid optical fiber have been removed, and the output end
of the optical fiber is placed in a safe position, away from flammable and
unprotected objects. The output light of the optical fiber (including UV light
output) should not be directly viewed by human eyes to avoid high-intensity
light and UV light radiation damage to the operator and others.
Illuminator installation and operation:
After turning on the power on the back of the Hyper E340 light source box, the
light source box enters a standby state, and the startup switch displays
“red.” Before starting the light source, make sure that the display panel
shows an intensity of 000%, and then manually press the front panel startup
switch to turn on or turn off the light source. Pressing the startup switch
can turn on the light source. When the light source is turned on, the startup
switch displays “blue,” and the output intensity is set to the minimum
intensity of 000%. Use the front panel knob to adjust the intensity by
rotating it (for example, 000% → 010% →… → 100%) or press the knob and rotate
it to fine-tune the intensity (for example, 010% → 011% →… → 100%). The light
source intensity is displayed on the panel.
Note: When the illuminator is turned on, take necessary protective
measures to protect yourself and others from exposure to strong light
(including UV light).
The illuminator does not have a safety interlock device. When the optical
fiber is disconnected from the light source box during operation, the light
source box continues to output light. Please take necessary protective
measures to protect personnel and items from exposure to strong light (UV
light).
Before turning off the illuminator, please make sure that the light source has
been turned off. When the light source is turned off, the startup switch
displays “red,” and the cooling fan runs for at least 3 minutes before turning
off the power of the light source box.
Installation and operation:
-
Carefully remove the components from the cardboard box.
-
You can choose to use the Hyper E340 lighting unit through either manual control or remote Operation.
-
Remote operation – RS232: Connect one end of the USB cable to the host and the other end to the computer.
-
Remote operation – TTL: Connect via BNC in order to trigger the LED with an external switch.
-
Connect the power cord to the Power socket on the back of the host, and the other end to a power outlet. Once powered on, the blue light on the power supply will remain on.
※ Please ensure that the DC power supply provided with the product is used, otherwise it may damage the light source and void the warranty. -
Liquid light guide installation method:
a. Secure the light guide adapter to the light source lens port of the Hyper E340 unit.
b. Attach the condenser lens to your microscope’s light receptacle.
c. Remove the protective covers from both ends of the light guide, insert one end into the light guide adapter, and the other end into the condenser lens.
d. After inserting the light guide into the port, gently tighten the screw by hand until the light guide is securely fixed in place. Over-tightening may cause damage to the light guide.
Operating instructions:
- Manual control box – screen display:
1. Startup page
| LED On/Off status memory value
---|---
2 Main menu
3 Submenu -Local All
| LED On/Off status
4 LEDS energy sync adjustment:0~100%
---|---
4 Submenu -Local Channel
| CH1 UV LED intensity setting
CH2 Blue LED intensity setting
CH3 Green LED intensity setting
---|---
| CH4 Red LED intensity setting
5 Submenu -TTL All (TTL Functional status and ALL LEDs intensity setting)
| TTL Functional status (ON/OFF)
4 LEDS energy sync adjustment:0~100%
---|---
6 Submenu -TTL Channel ( LEDS Self-adjustment of energy)
| TTL Functional status (ON/OFF)
CH1 UV Trigger energy setting
CH2 Blue Trigger energy setting
CH3 Green Trigger energy setting
CH4 Red Trigger energy setting
---|---
7 Submenu -Information1
| CH1 UV LED temperature
CH2 Blue LED temperature
CH3 Green LED temperatureCH4 Red LED temperature
---|---
8 Submenu -Information2
| CH1 UV LED hours used
CH2 Blue LED hours used
CH3 Green hours used
CH4 Red LED hours used
---|---
9 Submenu -Version
| Firmware version
---|---
Remote Operation – USB:
Before starting the USB mode for the first time, you can go to YODN official
website (https://www.yodnlighting.com/download/index/13) to download the
driver and operating software.
YODN_EX40 Operating instructions:
Remote operation-TTL:
-
TLL Input specifications:
a. Connector type: BNC (female port)
b. Minimum low level : +0V
c. Maximum low level : +1.5V
d. Minimum high level : +4.0V
e. Maximum high level: +5.5V
f. Typical input current: 450 μA -
BNC(TTL) Port:
Spectral output:
Hyper E340 light box spectrum is as follows:
Product specifications:
Relevant operating and storage environmental conditions and detailed specifications:
| Product Specification
---|---
1| Wavelength range| 350-730 nm
2| LED peaks/FWH| 365/20 nm, 460/40 nm, 560/80 nm, 630/20 nm; (Optional:
405/20 nm, 730/20 nm)
3| External power supply| Universal input 100-240VAC 50/60 H
4| Power consumption| 85W
5| LED on/off response| 1 ms
6| LED expected life| 25000 hours
7| Control Pad| LON/OFF
2.LED intensity control 10-100% 3.Color separation output control
8| External control| TTL Trigger & PC control
(RS232 command /USB B-type connector)
9| Dimension (mm)| Illuminator : 12/365211 Power supply :6716735
10| Weight| Approx. 5.5 Kg (including power supply)
11| Warranty| Illuminator : 24 Months
12| Option| 1.5M/y3 mm liquid light guide/ Adaptor, collimators
13| Certifications| CE(TBA)
Dimensions:
Daily Maintenance and Troubleshooting:
Daily Maintenance:
Please use a suction device to remove any accumulated dust or lint from the
air intake/outlet of the device to ensure stable air supply for cooling. In
dusty or smoky environments, it is recommended to clean the air intake/outlet
with a suction device at least every 6 months.
The Hyper E340 lighting fixture does not have any parts or components that
customers can replace themselves. Opening the outer casing of the lighting
fixture will void the warranty.
Troubleshooting:
Only basic troubleshooting information is provided depending on the condition
of the Hyper E340 lighting fixture. If the Hyper E340 lighting fixture needs
repair, it can only be performed by authorized technicians.
Unable to Start:
a. Make sure the power cord is properly connected to the power outlet.
b. Make sure the power cord is properly connected to the power supply socket.
c. Make sure the power supply connector is properly connected to the Hyper
E340 lighting fixture.
d. If the main power of the Hyper E340 lighting fixture is turned on, the blue
light of the power supply should be constantly lit.
Low Light Intensity:
a. Remove the screw from the collimator lens by hand, insert the light guide
into the collimator lens, and secure the light guide with the screw.
b. Make sure the collimator lens is compatible with the microscope and
securely fix the collimator lens to the microscope.
Control Box Operation Failure: Please make sure the control box connector is securely installed and connected to the Hyper E340 lighting fixture.
Customer Service:
For technical support related to the product, please call +886-3-5637218, send
an email to Sales@yodnlighting.com, or visit
https://www.yodnlighting.com/contact/getInTouch to contact us.
You can also find relevant information on the YODN website at
http://www.yodnlighting.com.
|
---|---
http://weixin.qq.com/r/Ph2hubXEmujlrfHa90gs|
https://www.yodnlighting.com/en
Warranty:
YODN offers a 2-year warranty for the Hyper E340 lighting fixture to end- users, starting from the date of shipment. The warranty for peripheral accessories of the lighting fixture does not cover liquid light guides.
HYPER E340 User Manual (Edit 2023/03)
Documents / Resources
| YODN
Hyper E340 Broadband Excitation Illuminator for Fluorescence
Microscope
[pdf] User Manual