tpi SP341 Sub-mini K-Type Thermometer Smart Probe
Product Information
SP341 K-Type Temperature Smart Probe
The SP341 is a K-Type thermocouple thermometer Smart Probe that connects to
your mobile device and uses the FREE TPI View App to display temperature
readings. It can perform differential temperature measurements with one
instrument and comes with a GK41M air probe. Other probes can be bought
separately.
Product Usage Instructions
-
Step 1: Download the App
Download either the SmartProbe or the View app from either the Apple App Store or the Google Play Store. -
Step 2: Connect the Probe
Connect the K-Type probe with the sub-mini connector to the input connection on the SP341. Ensure that the connector is recessed and polarized (marked + and -). -
Step 3: Turn on the Instrument
Press and hold the power key to turn on the instrument. The SP341 status LED indicator will blink orange if Bluetooth is not connected, and green if Bluetooth is connected. If the battery is low, the LED indicator will blink red. -
Step 4: Use the App
Open the TPI View app on your mobile device to display temperature readings. Data can be displayed, saved, logged, allocated to a specific job, exported & e-mailed in either pdf or csv.
Safety Considerations:
- Never attach the SP341 to a hot surface.
- Always make sure the temperature probe used has the proper range for the application.
- Always use K-Type thermocouples only.
- Do not expose the unit to damp environments for extended periods of time.
Specifications:
The SP341 has been tested with the following standards and is in compliance with the council RED directive 2014/53/EU:
- EN 62479:2010
- EN 61326-1:2013
- EN 61010-1:2010
- ETSI EN 301 489-17 V3.1.1(2017-02)
- ETSI EN 301 489-1 V2.1.1 (2017-02)
- ETSI EN 300 328 V2.2.0 (2017-11)
Warranty:
The SP341 comes with a 3 Year Limited Warranty.
Maintenance:
Do not expose the unit to damp environments for extended periods of time. If
necessary, clean the unit with a dry cloth.
Troubleshooting:
If the SP341 status LED indicator blinks red, replace the batteries using the
instructions provided in the user manual.
Accessories:
Other probes can be bought separately from Test Products International, Inc.
Replacing Batteries:
Replace the batteries using the instructions provided in the user manual.
Test Products International, Inc.
Uses any K-Type probe with a sub-mini connector
Connects to smartphone and uses the FREE TPI View app.
Perform differential temperature measurements with one instrument
Auto field calibrate to +/-1°C system accuracy in the range of -1°C to 49°C using field calibrate mode
3 Year Limited Warranty
- Use standard K-type thermocouple probes with a sub-mini input plug.
- GK41M air probe included.
- Other probes are sold separately.
The SP341 has been tested with the following standards and is in compliance with the council RED directive 2014/53/EU:
INTROdUCTION
- Congratulations: Thank you for purchasing TPI products. The SP341 is easy to use and built to last. It is backed by a 3 year limited warranty.
- Product Description: The SP341 is a K-Type thermocouple thermometer “Smart Probe”. It connects to your mobile device and uses the FREE TPI View App to display temperature readings. Data can be displayed, saved, logged, allocated to a specific job, exported & e-mailed in either pdf or csv.
- Download either the SmartProbe or the View app from either the Apple App Store or the Google Play Store
SAFETY CONSIDERATIONS
- Never attach the SP341 to a hot surface.
- Always make sure the temperature probe used has the proper range for the application.
- Always use K-Type thermocouples only.
- Do not expose the unit to damp environments for extended periods of time.
SPECIFICATIONS
The SP341 has been tested with the following standards and is in compliance with the council RED directive 2014/53/EU: –
- EN 62479:2010
- EN 61326-1:2013
- EN 61010-1:2010
- ETSI EN 301 489-17 V3.1.1(2017-02)
- ETSI EN 301 489-1 V2.1.1 (2017-02)
- ETSI EN 300 328 V2.2.0 (2017-11)
The relevant certificate can be viewed, downloaded, and printed from the TPI
website at the following address:
https://www.tpieurope.com/wp-content/uploads/2019/11/SP341-Certificate-of-
Conformity.pdf.
INSTRUMENT OVERVIEW
- Input connection for K-Type probes. The connector is recessed and polarized (marked + and -). Ensure the K- Type probe is connected properly (see below).
- SP341 status LED indicator.
- Orange and blinking = Bluetooth not connected Green and blinking = Bluetooth connected
- Blinking Red with either of the above = Low battery
- Power Key. Press and hold to turn the instrument on. Press and hold to turn the instrument off.
- Built-in magnet. Attach to metal surfaces for hands-free operation. Ensure surface is not hot.
- Serial number location. The last three digits are the product identification number and are marked in bold. Use this number to distinguish between models when several SP341’s are connected to a phone or tablet.
- Screw to open the battery compartment. Turn the screw counterclockwise until it is just above the compartment door. Lift on the screw to open the compartment.
MAINTENANCE
- Clean the surface of the instrument with a damp cloth.
- Replace the three AAA batteries when the low battery indicator appears.
TROUBLESHOOTING
Temperature seems inaccurate:
- Check the thermocouple probe for incorrect wiring or open connection.
- Low battery.
- The non-K-Type probe is being used.
- Ensure the connection with the phone is working.
SP341 does not turn on:
- Dead or low batteries.
- Not holding down the power switch until the unit turns on.
- Defective POWER ON switch.
SP341 displays “OPEN”:
- No probe connected.
- The temperature probe is bad.
SP341 measures temperature backward:
- Thermocouple probe wired backward.
- Probe connected with reverse polarity.
ACCESSORIES
Included Accessories
- A341SP
- Soft pouch
- GK41M
- General purpose temperature probe
Optional Accessories
- A925
- Carrying case for 4 smart probes
- A927
- Carrying case for SP341 & temperature probes
Test Products International offers a variety of K-Type thermocouple probes for many applications. Visit www.tpieurope.com for additional information.
WARRANTY
This product is warranted to the purchaser against defects in material and workmanship for three years from the date of purchase.
Covered by Warranty:
Repair parts and labor; or replacement of the product at the company’s option.
Normal transportation charges to the purchaser are also covered.
Not Covered by Warranty:
Damages to the product which are the result of abuse, improper use or
maintenance are not covered. Any other expense, consequential damages,
incidental damages, or incidental expenses including damages to property are
not covered. Transportation expenses to the company are not covered.
To Obtain Warranty Performance:
Include with the product: your name, address, phone number, written
description of the problem, and proof of purchase date. Carefully package and
return to:
TPI, Inc.
9615 SW Allen Blvd. Beaverton, OR 97005 USA
503-520-9197
www.testproductsintl.com
TPI Canada
342 Bronte Rd. S., Unit 6 Milton, Ontario L9T 5B7 Canada
905-693-8558
www.tpicanada.com
TPI, Europe Ltd..
Longley House, International Drive Southgate Crawley
West Sussex RH10 6AQ England 44(0) 1293 530196
www.tpieurope.com.
Implied Warranties:
Any implied warranties including implied warranties of merchantability and
fitness for a particular purpose, are limited in duration to three years from
date of purchase. To the extent any provision of this warranty is prohibited
by federal or state law and cannot be preempted, it shall not be applicable.
This warranty gives you specific legal rights, and you may also have other
rights which vary from country to country.
REPLACING BATTERIES
- Turn the battery compartment screw counterclockwise until it is just above the compartment door. Lift on the screw to open the compartment.
- Replace the three AAA batteries.
- Re-install the battery cover and tighten the screw by turning it clockwise.
Thank You!
& Welcome to the TPI Family
We really appreciate you coming onboard, if you aren’t already, and look
forward to serving you over the next few years, and beyond. For the best
experience head over to www.tpieurope.com, create
your account and register your product.
Benefits include:
- Everything in one place
- Create service, calibration and warranty/repair requests
- Download certificates
- Contact sales, support and customer service
- Read correspondence from TPI including Service updates
- Manage your order history
- Shop Online for accompanying products and spares delivered direct to your door
Please do not hesitate to contact us by any of the following methods:
- Tel: 01293 530196
- E-mail: sales@tpieurope.com or service@tpieurope.com
- Website: www.tpieurope.com inc. LIVE CHAT
- Facebook: @tpieurope
- Twitter: @TPIEuropeLTD
Test Products International Europe Ltd.
Longley House, International Drive
Southgate, Crawley, West Sussex RH10 6AQ
Tel:: +44 (0) 1293 530196
www.tpieurope.com.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>