Aseptico AMC-15 Mobile Dental System Instruction Manual

October 30, 2023
Aseptico

Aseptico logo AMC-15
Mobile Dental System
SERVICE MANUALAseptico AMC-15 Mobile Dental System
-

GENERAL SERVICE INFORMATION

  1. This service manual provides information on how access the components most likely to require replacement: the compressor or vacuum pump, the peristaltic pump and the pressure switch. Refer to page 6 for O-ring and gasket replacement instructions. Refer to the schematic drawings included at the end of this manual for electrical and plumbing diagrams of 110V units and 220V units, as indicated on the drawings. Call Aseptico at 866-244-2954 to order parts.

DISASSEMBLY

SIDE PANEL REMOVAL

  1. Drain the air tank by opening the valve (Fig.  2 A ) on the rear panel. Place a container under the drain hose (Fig. 2 B ) to capture any moisture that may be in the tank.Aseptico AMC-15 Mobile Dental System - container under

  2. Remove the water bottles (Fig.  3 A ) from the side panel by unscrewing them from their caps.Aseptico AMC-15 Mobile Dental System - unscrewing

  3. Remove the screw (Fig. 4 A ) from the bottom of the side panel using a 5/32” hex wrench.Aseptico AMC-15 Mobile Dental System - side panel

  4. Remove the 4 bolts (Fig. 5 A ) at the corners of the blue handle using a 3/16” hex wrench.Aseptico AMC-15 Mobile Dental System - blue handle

  5. Remove the right side panel (Fig. 6 A ) from the unit. Do not lift from the blue handle (Fig. 6 B ) as it is not secured to the panel. Also note how the 2 water supply lines (Fig. 6 C ) are threaded through holes in the side panel and will need to be re-inserted through the holes when reattaching the side panel.Aseptico AMC-15 Mobile Dental System - reattaching the side panel

  6. Remove the left side panel (Fig. 7 A ) in the same way as the right side.Aseptico AMC-15 Mobile Dental System - way as the right REAR PANEL REMOVAL

  7. Remove the 4 screws (Fig. 8 A ) from the corners of the rear panel using a Phillips screwdriver.Aseptico AMC-15 Mobile Dental System - Phillips screwdrive

  8. Tip the rear panel (Fig.  9 A ) back to expose the interior of the unit.Aseptico AMC-15 Mobile Dental System - expose the interior FRONT PANEL REMOVAL

  9. Remove the 2 top screws (Fig. 10 A ) from the front panel using a 5/64” hex wrench. Aseptico AMC-15 Mobile Dental System - reattaching the side panel1

  10. Remove the 2 front screws (Fig.  11 A ) from the front panel using a 3/32” hex wrench.Aseptico AMC-15 Mobile Dental System - front panel

  11. Remove the 2 bottom screws (Fig. 12 A ) from the front panel using a 3/32” hex wrench.Aseptico AMC-15 Mobile Dental System -bottom screws

  12. Remove the front panel (Fig. 13 A ) from the unit just far enough to access the internal components as needed.Aseptico AMC-15 Mobile Dental System - components as needed COMPRESSOR (OR VACUUM PUMP) REMOVAL

  13. Disconnect air lines (Fig. 14 A ) using a 5/8” wrench.Aseptico AMC-15 Mobile Dental System - COMPRESSOR

  14. Remove all cable ties (Fig. 15 A ) from wiring using a wire snips.Aseptico AMC-15 Mobile Dental System - Remove all cable

  15. Remove the pressure switch cover (Fig. 16 A ) (located on the air tank) using a Phillips screwdriver.Aseptico AMC-15 Mobile Dental System - Phillips screwdriver

  16. Disconnect the compressor/pump wires (Fig.  17 A ) from the pressure switch using a Phillips screwdriver.Aseptico AMC-15 Mobile Dental System - Phillips screwdriver1

  17. Disconnect the wires (Fig. 18 A ) from the capacitor.Aseptico AMC-15 Mobile Dental System - Disconnect the wires

  18. Remove the ground wire (Fig. 19 A ) using a 5/16” wrench.Aseptico AMC-15 Mobile Dental System - wrench

  19. Locate the 4 bolts (Fig. 20-1 A ) holding the base of the compressor/pump, then locate the holes (Fig. 20-2 A ) in the base of the unit that allow access to those 4 bolts. Remove the 4 bolts using a 5/32” hex wrench.Aseptico AMC-15 Mobile Dental System - bolts using

  20. Remove compressor/pump (Fig. 21 A ) from unit.Aseptico AMC-15 Mobile Dental System - compressor2 PERISTALTIC PUMP REMOVAL

  21. Remove the cable tie (Fig. 22 A ) from the back of the peristaltic pump using a wire snips.Aseptico AMC-15 Mobile Dental System - using a wire snips

  22. Disconnect the red and black peristaltic pump wires (Fig. 23 A ) from where they join the circuit board.Aseptico AMC-15 Mobile Dental System - black peristaltic

  23. Flip up the lever (Fig. 24 A ) on top of the peristaltic pump.Aseptico AMC-15 Mobile Dental System - peristaltic pump

  24. Remove the tubing (Fig. 25 A ) from the peristaltic pump.Aseptico AMC-15 Mobile Dental System - Fig. 25

  25. Remove the screw (Fig. 26 A ) from the peristaltic pump using a #1 Phillips screwdriver.Aseptico AMC-15 Mobile Dental System - Fig. 26

  26. Rotate the front of the peristaltic pump (Fig. 27 A ) 45° counterclockwise, then remove from its base.Aseptico AMC-15 Mobile Dental System - Fig. 27

  27. Remove the 4 corner screws (Fig.  28 A ) from the peristaltic pump base using a 1/16” hex wrench.Aseptico AMC-15 Mobile Dental System - Fig. 28

  28. Remove peristaltic pump base (Fig. 29 A ) from unit.Aseptico AMC-15 Mobile Dental System - Fig. 29

  29. When reinstalling peristaltic pump, make sure that the waste hose direction through the pump (Fig.  30 A ) is left to right, as shown.Aseptico AMC-15 Mobile Dental System - Fig. 30 PRESSURE SWITCH REMOVAL

  30. Note the locations of the 2 black and 2 white wires, then disconnect (Fig. 31 A ) using a Phillips screwdriver.Aseptico AMC-15 Mobile Dental System - Fig. 31

  31. Remove the air line (Fig. 32 A ) from the pressure pump using a 1/2” wrench.Aseptico AMC-15 Mobile Dental System - Fig. 32

  32. Remove the pressure regulator (Fig.  33 A ) from the air tank using a 9/16” wrench.Aseptico AMC-15 Mobile Dental System - Fig. 33

MAINTENANCE

  1. Because of its simple design, the unit requires very little maintenance.
    Call Aseptico at 866-244-2954 to order parts.
    3-Way Syringe

  2. If the syringe starts to leak around the buttons (Fig.  35 A ) or the tip quick-change socket (Fig. 35 B ) due to wear, it is possible to order a syringe rebuild kit from Aseptico. Complete rebuild instructions are included with the kit.Aseptico AMC-15 Mobile Dental System - Fig. 35 Water Bottle Gaskets

  3. Replace water bottle gaskets (Fig.  36 A ) when worn or damaged. Gently peel old gaskets out of fittings and replace with new gaskets.

Aseptico AMC-15 Mobile Dental System - Fig. 36

WARRANTY

Aseptico Inc. warrants its new products against defects in material and workmanship under normal and proper use, care, and maintenance for a period of two (2) years from date of original invoice. This two (2) year warranty does NOT apply nor is it extended to products that are not manufactured by Aseptico. These products may be covered by a separate limited warranty provided by the particular manufacturer, and all claims and questions regarding the same are to be directed to the particular manufacturer.
Expendable components, such as batteries, fuses, light bulbs, and tubing sets installed on Aseptico products are specifically excluded and have no warranty. Consumable goods are warranted for the stated expiration date of such goods.
Repair or replacement of any product(s) or part(s) under this warranty does not extend the term of this warranty, and such product(s) or part(s) shall remain covered by the unexpired portion of the warranty period, or for ninety (90) days from the date of return to Aseptico, whichever is later. This limited warranty applies only to the initial or first installation of the product or part.
During the specific warranty periods set forth above, Aseptico will, at its option, repair or replace the product(s) or particular part(s) that are found to be defective in either material or workmanship in part or whole. Aseptico shall be the sole arbiter of such action.
In the event of alleged defect under warranty, the purchaser is to notify Aseptico’s Customer Service department promptly.
Customer Service will provide Return Material Authorization (RMA) instructions, usually directing that the product be returned for service, shipping prepaid by the buyer or end user, to Aseptico or its designated and authorized warranty service center.
This warranty shall not apply to products (1) that have been subjected to neglect, abuse, misuse, improper installation, inadequate maintenance, or damage due to improper use of cleaning materials or chemicals, or non- compliance with Aseptico’s storage, installation, operation, maintenance or environmental requirements; (2) that have undergone any modification or repair not previously authorized by Aseptico in writing, or service, repair or modification by or from any facility other than an authorized Aseptico service center or technician, or that use non-authorized software or spare or replacement parts; or (3) that fail due to reasonable and normal use or wear and tear, or materials made, furnished or specified by the buyer or end user.
Aseptico does not assume under this warranty any risks or liabilities arising from the clinical use of its products, whether or not such use involves coincidental utilization of products manufactured by others. Under no circumstances will Aseptico be liable or responsible for special, compensatory, incidental, consequential or punitive damages, lost profits, lost sales, or loss of use or loss of business opportunity by or through the use of the product. Aseptico’s sole and maximum liability with respect to the product, other than its obligations set forth above, shall be the total purchase price paid for the product.

NOTES _____

Aseptico logo P.O. Box 1548 • Woodinville, WA 98072-1548
8333 216th St SE • Woodinville, WA 98072-8060
(800) 426-5913 • (425) 487-3157 • Fax: 425-487-2608
www.aseptico.cominfo@aseptico.com
PRINTED IN THE USA