SAEVO S200 Dental Practice Unit Installation Guide

October 30, 2023
SAEVO

SAEVO S200 Dental Practice Unit

Product Information

The Dental Practice Unit is a piece of equipment used in dental practices that requires proper installation and operation. The unit requires compressed air, electricity, clean water, and proper sewer connections to function. The technical specifications for sizing the installations are provided in the user manual. The unit must be installed with a circuit breaker properly sized to protect the installed equipment. The electrical circuits of the equipment must be separated according to their power. The dental practice unit, compressor, vacuum pump, x-ray, and autoclave must be installed on separate circuits. The electrical power points of the equipment must be grounded. Clean, potable water must be used in the dental practice unit. PVC pipes must be used for installations. A water tap must be installed for each dental practice unit. A water register must be installed for the vacuum pump. Proper sewer connections must also be made for the unit to function.

Product Usage Instructions

1. Install the Dental Practice Unit with a circuit breaker properly sized to protect the installed equipment.

2. Separate the electrical circuits of the equipment according to their power. Install the dental practice unit, compressor, vacuum pump, x-ray, and autoclave on separate circuits. Ground all the electrical power points of the equipment.

3. Use only clean, potable water in the dental practice unit. Use PVC pipes for installations. Install a water tap for each dental practice unit and a water register for the vacuum pump.

4. Make proper sewer connections for the unit to function.

5. Refer to the technical specifications provided in the user manual for sizing the installations.

6. Follow the voltage and power requirements for each piece of equipment, as provided in the user manual.

7. Use the Dental Chair, Venturi Vac Plus, Micro Motor, High Rotation, Triple Syringe, X-rays, Vacuum Pump, Autoclave, and Compressor according to their intended usage and as instructed in the user manual.

Property facilities specifications

  • The knowledge of the specifications for the property facilities of the dental practice units contributes to their performance and durability. Therefore, before installing your equipment, read this manual carefully. All the information, illustrations and specifications of this manual are based on the information available at the time of publication. We reserve the right to make changes at any time, both to the product and to this Manual, without prior notification.
  • An authorized Alliage technician, under penalty of losing the guarantee, must carry out the installation of the dental practice units.
  • The design, sizing, and execution of the property facilities must be carried out by specific professionals in accordance with current regulations and legislation.
  • The specifications contained in this manual are based on the normative guidelines of ABNT.

Dental practice unit positioning
To achieve the best distribution of spaces for the dental practice unit, a place must be chosen such that allows the ergonomic installation of the equipment, so that the professional can obtain ideal conditions of ambiance and integration to work, improving his quality and productivity. Therefore, the installation of the equipment must be carried out in accordance with the standards recommended by I.S.O. (International Organization for Standardization) and F.D.I. (International Dental Federation) For analysis of the equipment according to its location in the dental practice unit, I.S.O./F.D.I. determined to divide the room into a few areas. To demarcate these areas, idealize a clock counter, where the center is the axis of the pointers and corresponds to the mouth of the patient lying in the dental chair positioned the horizontally.

  • Around this center, draw three concentric circles, A, B, and C of radius 0.5; 1.0; 1.5 meters respectively.
  • The patient’s head always indicates the 12-hour position. In this way, the axis 6-12 hours, divides the room into two areas: on the dental chair (area of the dentist) and below (area of the assistant).
  • The area limited by circle A, 0.5 meters radius, corresponds to the Transfer Zone, where everything that is transferred to the patient’s mouth must be located, as are the instruments and the manual parts of the Equipment. In this area should also be located the two bench seats, the dentist and his assistant.
  • Circle B of a 1.0-meter radius limits the Working Area (maximum range), which can be reached with the movement of the stretched arm. In this area should be the auxiliary tables and the body of the Equipment.
  • Circle C of 1.5 meters radius limits the total area of the dental practice unit. Fixed cabinets and vats are located in this area, so when opening the drawers, they must be inside circle B.

Compressed air

  • The compressed air installations must be free of solid particles, liquids, and oil. Size the compressed air network between 80 to 120 PSI and install the air compressor close to the supply point to avoid losses.
  • Avoid installing the compressor in sanitary facilities such as bathrooms and lavatories, to minimize air pollution used in dental practice units.
  • An air filter with a pressure regulator must be installed in each dental practice unit using a ¼ x 10mm elbow at the inlet and a ¼ x 6mm elbow at the outlet.
  • Compressed air pipes must have a slope of 5% to 10% in the direction of the airflow and an automatic drain valve at the lowest point of the pipe.
  • In installations, preferably use 10mm Polyurethane tubes.

Electricity

  • Use a circuit breaker properly sized to protect the installed equipment.
  • Separate the electrical circuits of the equipment according to their power. Install the dental practice unit, compressor, vacuum pump, x-ray, and autoclave on separate circuits.
  • Ground all the electrical power points of the equipment.

Water

  • Use only clean, potable water in the dental practice unit.
  • In the installations use PVC pipes.
  • Install the water tap for each dental practice unit.
  • Install water register for the Vacuum Pump.

Sewer

  • The drainage pipes must have a slope of 2% to 5% in the direction of the drain flow.
  • Use separate sections for the drainage of the vats and lavatories, the dental chair, and the vacuum pump. Connect the drain sections only in a siphoned box.
  • In siphoned boxes use a preferentially blind lid or lattice with closure.
  • In joints use only pieces at 45° to avoid blockages and the correct direction to the drain.
  • Install in the drainage network, ventilation branch to avoid negative pressure inside the pipe.
  • Use rigid PVC pipes.

High Power Suction

  • Sewer pipes of high power suction should be made with rigid PVC pipes, with 25mm diameter, joints at 45º, and inclination in the flow direction.
  • The electrical command installations of high-power suction must be made in parallel when used in more than one dental practice unit.

Technical specifications for sizing the installations

Equipment Air consum. l/min Water consum. l/min Voltage (V) Power (VA)
Dental chair 110 / 127 o 220 350
Venturi 25
Vac Plus 35
Micromotor 50 0,09
High rotation 35 0,09
Triple syringe 19 0,13
X-rays 127 of 220 1200
1/2 HP Vacuum Pump 0,4 Bi-volt (key) 1220
1 HP Vacuum Pump 0,4 Bi-volt (key) 1690
12 L Autoclave (see manual) Bi-volt (key) 1600
21 L Autoclave (see manual) Bi-volt (key) 1700
40L Compressor 127 o 220 830 / 1100
65L Compressor 127 o 220 1660 / 2200
150L Compressor     220 2948

DENTAL SET POSITIONING

Consider 1560mm from point R up to point P

  • R – Intersection between point D and the central axis of the dental chair
  • P – Head support Work Point
  1. Corrugated flexible conduit 020mm
  2. 10mm Polyurethane Tube
  3. 3 cables 2,5mm2
  4. Corrugated flexible conduit 020mm
  5. Water inlet connection 01/2″
  6. Weldable nipple with brass cap 020mm × ½”
  7. Weldable 90° elbow 020mm
  8. Weldable rigid PVC tube 020mm
  9. Buffer
  10. Weldable nipple with brass cap025mm×3/4″
  11. Weldable 90° elbow 025mm
  12. Long weldable reduction socket
  13. Weldable rigid PVC tube 040mm
  14. Siphoned box 150×150×50mm
  15. Hose adapter 03/4″
  16. Weldable nipple with brass cap 025mm×3/4″
  17. Weldable 90° elbow 025mm
  18. Weldable rigid PVC tube 025mm
  19. 3 cables 0,75mm (different colors)
  20. Corrugated flexible conduit 020mm

SPARE CONNECTION BOX

ATTENTION
The dental practice units use an integrated connection box at the base of the dental chair. For installations with a standard connection box separated from the base, it is necessary to acquire the kit for a loose connection box (sold separately), in this case, use this installation pattern.

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>