SENSECAP S-pH-01A pH Sensor User Guide
- October 30, 2023
- SENSECAP
Table of Contents
**SENSECAP S-pH-01A pH Sensor
**
Introduction
S-PH-01 transmitter measures the PH of solution or semi-solid substrate. The output signal can be RS485 and Analog Voltage. The sensor is applicable for industrial, water processing, sewerage system, irrigation, smart agriculture etc.
- Can be used for PH measurement
- Output Interface with RS485, Voltage
- High impedance and isolated electrode input
- High accuracy with excellent stability
- Reverse power protection and Built-in TVS/ESD protection
Specificatio ns
Output Interface| Analog Voltage 0-2V
(Output resistance ~0ohm)
| RS485 Modbus-RTU
Power Supply| 3.9-30V/DC| 3.9-30V/DC
Power Consumption| 35mA@24V DC| 35mA@24V DC
Start-up time| < 2 seconds
PH Measurement| High impedance and isolated input; Range: 0-14PH,
Resolution: 0.01PH, Accuracy: +/-0.1PH; Can be used for solution or semi-solid
substrate
Temperature Measurement
(Option)
| Range: -40~80°C, Resolution: 0.1°C, Accuracy: +/-0.5°C
IP Ratings| Electrode: IP68; Transmitter: IP65
Operating Temperature| -40~85°C
Installation| Electrode: 3/4″NPT screw threads; Transmitter: Mounting
hole
Cable Length| Power and Signal Cable:2 meters or Customize; Electrode
Cable: 5 meters
Dimension| Electrode: WidthDiameter 16030mm; 3/4″NPT screw threads
Transmitter: 140mm65mm50mm
Wiring diagrams
Type | Wiring diagram |
---|---|
Analog Voltage Output | Red (V+): Power Supply + Black (G): Power Supply |
– Blue (O1): Analog Output
RS485
Modbus
| Red (V+): Power
Supply + Black (G):
| Power Supply – Yellow
| (T+): RS485+/A/T+
| White (T-): RS485-
| /B/T-
Dimension
Electrode Dimension
Installation Maintenance and Calibration
Installation
Installation locations of Electrodes will vary depending on the system design. The key is to monitor a good representative sample of the whole solution directly after introduction of chlorine. The installation location must allow for complete contact of the scrubber liquid with the probes. Some example installation locations for Electrodes include the following:
- Outlet of packed tower
- Outlet of recycle pump
- Pump bypass line
- Heat Exchanger bypass line
Maintenance
Under normal conditions, electrodes can last anywhere from several months to several years depending on the type of operation, rate of production, strength of product, and quality of the raw materials used in the process. Because each application is different, there is no average life expectancy.
Because the pH responsive glass bulb or flat surface is relatively thin, care should be taken so that the bulb does not become scratched or broken. It is also important that ORP measuring surfaces are not scratched or gouged. The suggestions in this sheet are intended to help avoid these problems. Coating of an electrode’s measuring surface can lead to erroneous readings including shortened span and slow response times. The type of coating determines the type of cleaning technique. Soft Coatings can be removed by vigorous stirring, by use of a squirt bottle or, very carefully, by gently wiping with a soft, clean non-abrasive paper or cloth. Hard Coatings should be chemically removed. The chemical used to remove the coating should be the least harsh chemical that dissolves the coating in 1 or 2 minutes and does not attack the electrode’s materials of construction. For example, a calcium carbonate coating might be removed with 5% HCl (muriatic acid). Oily or Organic Coatings are best removed with detergents or an appropriate solvent that does not attack the electrode’s materials of construction. For example, isopropyl alcohol might be used but acetone should be avoided if the electrode’s body is made of CPVC.
Output Signal Conversion
PH output conversion
Output Interface | Parameters Range | Conversion Formula |
---|---|---|
Analog Voltage Output 0-2V | PH range: 0-14PH | PH=7.00*VOLTAGE. When |
VOLTAGE=1.0V,then PH
=7.00*1.00=7.
RS485| PH range: 0-14PH| PH= (REGISTER VALUE)/100.When REGISTER
Modbus-RTU| | VALUE=7000,then PH= 7000/100=7.00.
| Temperature range: -40-80°C| TEMPERATURE= (REGISTER
| | VALUE)/100.When
| | REGISTER VALUE=2013,then TEMPERATURE
| | = 2013/100=20.13°C.
NOTE: The unit of VOLTAGE is (V)
RS485 Modbus Protocol
Modbus Protocol is widely used to establish master-slave communication between intelligent devices or sensors. A MODBUS message sent from a master to a slave contains the address of the slave, the function code (e.g. ‘read register’ or ‘write register’), the data, and a check sum (LRC or CRC). The sensor is RS485 interface with Modbus protocol. The default serial communication settings is slave address 1, modbus rtu, 9600bps, 8 data bits and 1 stop bit. All communication settings can be changed with modbus command, and take effective after re-power up the sensor.
Following modbus function code are supported by sensor.
Modbus Function Code 0x03 : used for reading holding register.
Modbus Function Code 0x04 : used for reading input register. Modbus
Function Code 0x06 : used for writing single holding register.
Modbus Function Code 0x10: used for writing multiple holding register.
Modbus Register
Parameters| Register Addr. (HEX/DEC)| Dat a Type| Modbus Function
Code(DEC)
| Range and Comments| Default Value
---|---|---|---|---|---
TEMPRATURE| 0x0000 /0| INT16
RO
| 3/4| -4000-8000 for –
40.00~80.00°C.
| N/A
PH
PHVALUE
| 0x0001 /1| UINT16
RO
| 3/4| 0-1400 for 0.00-14.00| N/A
| | | | |
TEMPCOMPENSA TEE N| 0x0020 /32| UINT16
R/W
| 3/6/16| 0: External
Temperature sensor 1: Disabled
2:On board temperature sensor
| 1
| | | | |
PHCALIBRAWAD0 PH calibration point for PH=4.01| 0x0030 /48| UINT16
R/W
| 3/6/16| -2000~2000 for – 2000~2000; Write 0xFFFF to calibrate.| N/A
PHCALIBRAWAD1 PH calibration point for PH=7.00| 0x0031 /49| UINT16
R/W
| 3/6/16| -2000~2000 for – 2000~2000; Write 0xFFFF to calibrate.| N/A
PHCALIBRAWAD2 PH calibration point for PH=10.01| 0x0032 /50| UINT16
R/W
| 3/6/16| -2000~2000 for – 2000~2000; Write 0xFFFF to calibrate.| N/A
| | | | |
SLAVEADDRESS| 0x0200 /512| UINT16
R/W
| 3/6/16| 0-255| 1 or 12
BAUDRATE| 0x0201 /513| UINT16
R/W
| 3/6/16| 0-6
0:1200bps
1:2400bps
2:4800bps
3:9600bps
4:19200bps
5:38400bps
| 3:9600bps
PROTOCOL| 0x0202 /514| UINT16
R/W
| 3/6/16| 0-1
0:Modbus RTU
| 0:Modbus RTU
PARITY| 0x0203 /515| UINT16
R/W
| 3/6/16| 0-2
0:None
1:Even
2:Odd
| 0:None
Parity
DATABITS| 0x0204 /516| UINT16| 3/6/16| 1| 1:8 databits
R/W| | 1:8 databits|
STOPBITS| 0x0205 /517| UINT16
R/W
| 3/6/16| 0-1
0:1 stopbit
1:2 stopbits
| 0:1 stopbit
RESPONSEDELAY| 0x0206 /518| UINT16
R/W
| 3/6/16| 0-255 for 0-2550 milliseconds| 0
ACTIVEOUTPUTINTER
VAL
| 0x0207 /519| UINT16
R/W
| 3/6/16| 0-255 for 0-255 seconds.| 0
NOTE: UINT16:16 bit unsigned integer, INT16:16bit signed integer NOTE: RO: Register is Read Only, R/W: Register is Read/Write NOTE: HEX is Hexadecimal (data with 0x/0X prefix), DEC is Decimal
Modbus Register Detail Description
TEMPERATURE
Data Range| -4000-8000 For -40.00~80.00°C| Default: N/A
Power Down Save| N/A|
Note: Temperature value (Binary complement). Example: When REGISTER =
0x0702 (HEX format), then
VALUE=(0x07256+0x02)/100=17.94°C.When REGISTER=FF05H (HEX format),then
VALUE=((0xFF256+0x05)-0xFFFF-0x01)/100 =(0xFF05-0xFFFF- 0x01)/100=-2.51°C
PH VALUE
Data Range| 0-1400 for 0-14.00PH| Default: N/A
Power Down Save| N/A|
Note: PH value
Example: When REGISTER = 0x02BC (HEX format), then
VALUE=(0x02*256+0xBC)/100=7.00PH
TEMPCOMPENSATEEN
Data Range| 0 : External Temperature
sensor 1: Disabled
2: On board temperature sensor| Default: 1
Power Down Save| YES|
Note: Temperature compensation
PHCALIBRAWAD0
Data Range| -2000~2000| Default: N/A
Power Down Save| YES|
Note: PH calibration AD value for PH=4.01; Immerse the electrode into PH=4.01solutionand wait until the reading value being stable, then write 0xFFFF to this register to calibrate.
PHCALIBRAWAD1
Data Range| -2000~2000| Default: N/A
Power Down Save| YES|
Note: PH calibration AD value for PH=7.00; Immerse the electrode into PH=7.00solutionand wait until the reading value being stable, then write 0xFFFF to this register to calibrate.
Data Range | -2000~2000 | Default: N/A |
---|---|---|
Power Down Save | YES |
Note: PH calibration AD value for PH=10.01; Immerse the electrode into PH=10.01solutionand wait until the reading value being stable, then write 0xFFFF to this register to calibrate.
SLAVEADDRESS — Modbus Slave Address
Data Range| 0-255| Default: 1 or 12
Power Down Save| YES|
Note: Please re-power on the sensor to take effective after set.
BAUDRATE — Serial Comm Baudrate
Data Range| 0-5
0:1200bps
1:2400bps
2:4800bps
3:9600bps
4:19200bp
s
5:38400bps| Default: 3
Power Down Save| YES|
Note: Please re-power on the sensor to take effective after se
PROTOCOL — Serial Comm Protocol
Data Range| 0
0:Modbus RTU
| Default: 0
Power Down Save| YES|
Note: Please re-power on the sensor to take effective after set.
PARITY — Serial Comm Parity
Data Range| 0-2
0:NONE
1:EVEN
2:ODD
| Default: 0
Power Down Save| YES|
Note: Please re-power on the sensor to take effective after set.
DATABITS — Serial Comm Data bits
Data Range| 1
1:8 data bits
| Default: 1
Power Down Save| YES|
Note: Please re-power on the sensor to take effective after set.
STOPBITS — Serial Comm Stop bits
Data Range| 0-1
0:1 stop bit
1:2 stop bits
| Default: 0
Power Down Save| YES|
Note: Please re-power on the sensor to take effective after set.
RESPONSEDELAY — Serial Comm Response Delay
Data Range| 0-255 for 0-2550 milliseconds, 0 for disabled| Default: 0
Power Down Save| YES|
Note: Please re-power on the sensor to take effective after set.
Note: Sensor will delay a period before response to master request command.
Example: When set to 5 and receive a request from master device, then
sensor will delay 5*10ms=50ms, then response to master.
ACTIVEOUTPUTINTERVAL — Serial Comm Active Output Interval time
Data Range| 0-255 for 0-255 seconds, 0 for disabled| Default: 0
Power Down Save| YES|
Note: Please re-power on the sensor to take effective after set.
Note: Sensor will output the data actively without any master request
command.
Note: Only ONE sensor should be on RS485 network, or there will be data
collision and corrupt the data on line.
Note: Use key button to restore the serial comm parameters factory value
to exit the active output mode. Example: When set to 5 then sensor will output
the data every 5 seconds without any master request command.
Modbus Function Code
For description below, data started with 0X/0x means that it’s in HEX format
1 Function Code
Function Code 3 Protocol Example
Master Request: AA 03 RRRR NNNN CCCC
AA | 1 byte | Slave Address,0-255 |
---|---|---|
0x03 | 1 byte | Function Code 3 |
RRRR | 2 byte | Starting Register Addr |
NNNN | 2 byte | Quantity of Register to read |
CCCC | 2 byte | CRC CHECKSUM |
Slave Response: AA 03 MM VV0 VV1 VV2 VV3… CCCC
AA | 1 byte | Slave Address,0-255 |
---|---|---|
0x03 | 1 byte | Function Code 3 |
MM | 1 byte | Register Data Byte Count |
VV0,VV1 | 2 byte | Register Value (High8bits first) |
VV2,VV3 | 2 byte | Register Value (High8bits first) |
… | … | Register Value (High8bits first) |
CCCC | 2 byte | CRC CHECKSUM |
**** Example: Read register 0x0200-0x0201,that is slave address and
baudrate. Master Request:01 03 0200 0002 C5B3
Slave Addr. | 1 byte | 0x01 |
---|---|---|
Function Code | 1 byte | 0x03 |
Starting Register
Addr.
| 2 byte| 0x0200
Quantity of Register
to read
| 2 byte| 0x0002
Checksum| 2 byte| 0xC5B3
Slave Response:01 03 04 00 01 00 03 EB F2
Slave Addr. | 1 byte | 0x01 |
---|---|---|
Function Code | 1 byte | 0x03 |
Register Data Byte
Count
| 1 byte| 0x04
Register Value:
Address
| 2 byte| 0x00(HIGH 8 Bits)
| | 0x01(LOW8 Bits)
Register Value:
Baudrate
| 2 byte| 0x00(HIGH 8 Bits)
| | 0x03(LOW8 Bits)
Checksum| 2 byte| 0xEBF2
6.4.2 Function Code 4 Protocol Example
Master Request: AA 04 RRRR NNNN CCCC
AA | 1 byte | Slave Address,0-255 |
---|---|---|
0x04 | 1 byte | Function Code 4 |
RRRR | 2 byte | Starting Register Addr |
NNNN | 2 byte | Quantity of Register to read |
CCCC | 2 byte | CRC CHECKSUM |
Slave Response: AA 04 MM VV0 VV1 VV2 VV3… CCCC
AA | 1 byte | Slave Address,0-255 |
---|---|---|
0x04 | 1 byte | Function Code 4 |
MM | 1 byte | Register Data Byte Count |
VV0,VV1 | 2 byte | Register Value (High8bits first) |
VV2,VV3 | 2 byte | Register Value (High8bits first) |
… | … | Register Value (High8bits first) |
CCCC | 2 byte | CRC CHECKSUM |
Example: Read register 0x0000-0x0002,that is temperature, PH
value. Master Request: 01 04 0000 0002 71CB
Slave Addr. | 1 byte | 0x01 |
---|---|---|
Function Code | 1 byte | 0x04 |
Starting Register
Addr.
| 2 byte| 0x0000
Quantity of Register
to read
| 2 byte| 0x0002
Checksum| 2 byte| 0x71CB
Slave Response: 01 04 04 08C3 029E 8910
Slave Addr. | 1 byte | 0x01 |
---|---|---|
Function Code | 1 byte | 0x04 |
Register Data Byte
Count
| 1 byte| 0x04
Register Value:
Temperature
| 2 byte| 0x08(HIGH 8 Bits)
0xC3(LOW8 Bits)
Register Value:
PH
| 2 byte| 0x02(HIGH 8 Bits)
0x9E(LOW8 Bits)
Checksum| 2 byte| 0x8910
Temperature =(0x08256+0xC3)/100=2243/100=22.43 °C
PH=(0x02256+0x9E)/100=670/100=6.70PH
6.4.3 Function Code 6 Protocol Example
Master Request: AA 06 RRRR VVVV CCCC
AA | 1 byte | Slave Address,0-255 |
---|---|---|
0x06 | 1 byte | Function Code 6 |
RRRR | 2 byte | Register Addr (High8bits first) |
VVVV | 2 byte | Register Value (High8bits first) |
CCCC | 2 byte | CRC CHECKSUM |
Slave Response: AA 06 RRRR VVVV CCCC
AA | 1 byte | Slave Address,0-255 |
---|---|---|
0x06 | 1 byte | Function Code 6 |
RRRR | 2 byte | Register Addr (High8bits first) |
VVVV | 2 byte | Register Value (High8bits first) |
CCCC | 2 byte | CRC CHECKSUM |
Example: Write Register 0x0020,that is set temperature compensation
Request: 01 06 0020 0000 8800
Slave Addr. | 1 byte | 0x01 |
---|---|---|
Function Code | 1 byte | 0x06 |
Register Addr. | 2 byte | 0x0020 (High8bits first) |
Register Value | 2 byte | 0x0000 (High8bits first) |
Checksum | 2 byte | 0x8800 |
**** Response:01 06 0021 0001 1800
Slave Addr. | 1 byte | 0x01 |
---|---|---|
Function Code | 1 byte | 0x06 |
Register Addr. | 2 byte | 0x0020 (High8bits first) |
Register Value | 2 byte | 0x0000 (High8bits first) |
Checksum | 2 byte | 0x8800 |
6.4.3 Function Code 16 Protocol Example
Master Request: AA 10 RRRR NNNN MM VVVV1 VVVV2 …CCCC
AA | 1 byte | Slave Address,0-255 |
---|---|---|
0x10 | 1 byte | Function Code 0x10 |
RRRR | 2 byte | Starting Register Addr |
NNNN | 2 byte | Quantity of Register to write |
MM | 1 byte | Register Data Byte Count |
VVVV1 | 2 byte | Register Value(High8bits first) |
VVVV2 | 2 byte | Register Value(High8bits first) |
… | … | Register Value(High8bits first) |
CCCC | 2 byte | CRC CHECKSUM |
Slave Response: AA 10 RRRR NNNN CCCC
AA | 1 byte | Slave Address,0-255 |
---|---|---|
0x10 | 1 byte | Function Code 0x10 |
RRRR | 2 byte | Starting Register Addr |
NNNN | 2 byte | Quantity of Register to write |
CCCC | 2 byte | CRC CHECKSUM |
Example: Write Register 0x0200-0x0201,that is set slave address to
1,andbaudrateto 19200bp. Master Request:01 10 0200 0002 04 0001 0004 BACC
0x01 | 1 byte | Slave Addr. |
---|---|---|
0x10(HEX) | 1 byte | Function Code 0x10 |
0x0200 | 2 byte | Starting Register Addr |
0x0002 | 2 byte | Quantity of Register to write |
0x04 | 1 byte | Register Data Byte Count |
0x0001 | 2 byte | Register Value: Slave Address 1 |
0x0004 | 2 byte | Register Value: Baudrate 19200bps |
0xBACC | 2 byte | CRC CHECKSUM |
Salve Response: 01 10 0200 0002 4070
0x01 | 1 byte | Slave Addr. |
---|---|---|
0x10(HEX) | 1 byte | Function Code 0x10 |
0x0200 | 2 byte | Starting Register Addr(High8bits first) |
0x0002 | 2 byte | Quantity of Register to write(High8bits first) |
0x4070 | 2 byte | CRC CHECKSUM |
Software Configuration Utility
You can use software listed below to try reading/writing the register of
sensor:
https://github.com/ed-chemnitz/qmodbus/releases
http://qmodbus.sourceforge.net/