Jion Pharmacy YM-3R9 Mesh Nebulizer User Manual
- June 10, 2024
- Jion Pharmacy
Table of Contents
- Jion Pharmacy YM-3R9 Mesh Nebulizer
- Getting to know your instrument
- Signs and symbols
- Safety information and warnings
- Useful information on the device
- Description of device and accessories
- lnitial use
- Operation
- Cleaning and disinfection
- Storage
- Troubleshooting
- Technical specifications
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Jion Pharmacy YM-3R9 Mesh Nebulizer
Thank you for purchasing this product. Please read the instruction manual carefully and keep it for later use. Be sure to make the manual accessible to other users and observe the information they contain.
Getting to know your instrument
This nebulizer is a mesh type nebulizer for treating the upper and lower respiratory tract. The nebulization and inhalation of medication prescribed or recommended by a physician can prevent against diseases of the respiratory tract, treat the side effects of these disorders and accelerate healing. For more information on possible applications, consult yourphysician or pharmacist. The device is suitable for inhalation at home. Medicines should only be inhaled when instructed by a physician . Inhalation should be performed in a calm and relaxed atmosphere. Inhale slowly and deeply to enable the medicine lo reach the small bronchi deep into the lungs. Exhale normally.
Signs and symbols
The following symbols appear in these instructions.
- Warning Warning instruction indicating a riskofinjuryordamage to health.
- Important Safety note indication possible damage to the uniVaccessory.
- Note Note on important information
Safety information and warnings
Warning
- Before use, ensure that there is no visible damage to the unit or accessories. When in doubt, do not use the unit and contact your dealer or the customer service address provided.
- Using the unit is not a substitute for consultation with or treatmentform a physician. Whenever you have any pain or illness, always contact your physician first.
- If you have health concerns of and kind ,contact your physician.
- 0nly use medicines that have been prescribed orrecommended by your physician or pharmacist.
- lf the unit dose is not functioning correctly or if you begin to feel unwell or feel pain, stop using the unit immediately.
- During use, hold the unit away from the eyes, as some nebu!izer medicines may cause damage to the eyes.
- Do not operate the unit in the presence of flammable gases.
- Do not use the device near electromagnetic transmitters.
- This device is not intended to be used by children or persons with restricted physical, sensory (e.g. insensitivity to pain) or mental abilities or persons lacking in the required experience or knowledge for safe operation of the device, unless supervised or instructed by a person responsible for their safety
- Do not dip the unit into water and fo not use ii in wet rooms. Do not allow any liquid to penetrate the unit.
- Remove the batteries from the device if you will not be using it for an extended period of time. Leaking batteries may damage the device.
Important
- Store the device in a location protected against environmental influences. The device must be stored in the specified ambient conditions.
- Do not use any liquids that have a viscosity (thickness) greater than 3, as these may damage the mesh irreparably.
- Do not use any medicines in powder form (even if dissolved.)
- Do not shake the atomiser, as this may cause liquid to leak, and the device may be damaged.
- Unplug from the socket after using the adapter.
- Switch the unit off immediately if it is defective or malfunctioning.
- The manufacturer shall not be held liable for damage or injuries caused by improper or incorrect use.
Useful information on the device
Accessories
The safe functioning of the device can only be guaranteed if the accessories
recommended by the manufacturer are used.
Protecting against leaks
When filling the medicine container with medicine, ensure that you do not
exceed the maximum mark (8ml). The recommended filling amount is between 2ml
and 8ml. During use, you can tilt the device upto a maximum angle of 45° in
any direction without impairing its nebulizing function or the success of the
treatment. However, please try to hold the device as upright as much as
possible.
Automatic switch-off
The device has an automatic switch-off function . To prevent damage to the
mesh, the device switches off automatically when the medicine/liquid has been
almost entirely used up. Do not use the device if the medicine container or
water tank is empty. The device switches off automatically if the substance to
be sprayed is no longer in contact with the mesh.
Description of device and accessories
For the convenience, the following “1.Medicine container” and “2.Mesh atomiser” will be called by a joint name: medicine cup module (12)
- Medicine container
- Mesh atomizer
- 0n / Off button / Operation LEDs Blue device ready
- Apparatus body
- Battery compartment
- Adapter connector
lnitial use
Before using the unit for the first time
Notes
The atomiser and accessories must be cleaned and disinfected before they are
used
Assembly
- Remove the unit from the packaging.
- Fit the Medicine cup module(12) onto the apparatus body(4). (image 1) When remove the medicine cup module(12), please press the PUSH button, and pull up the medicine cup module. (image 2)
Inserting the batteries
- To open the battery compartment(5) on the underside of the device, apply light pressure and pull the battery compartment cover in the direction of the arrows, then fold the lid up.
- InsertALKALINE or Rechargeable batteries (type AA).
- Ensure that the poles of the batteries are touching their respective poles in the device.
- Fold the battery compartment lid down again and push it back in place, in the opposite direction to the arrows.
Notes
- When changing batteries, ensure that the medicine container is completely empty, otherwise there is a risk of leaks.
- When the red LEDs (3) red light up, replace the batteries, spraying becomes much less effective when the batteries are weak.
- You can operate the device for approx. 60-70 minutes with new alkaline batteries.
- If the unit will not be use for a long time, please take out the battery.
Operating the device with A/C Adapter
Before using the adapter, remove the batteries.
- Insert the USB cable into the connection socket.
- Plug adapter the A/C outlet.
Notes
Remove the batteries when operating the device with the Adapter.
Operation
Preparing the nebulizer
For reasons of hygiene, it is imperative that the nebulizer and accompanying
accessories are cleaned and disinfected after every use. If the therapy
requires that various different medicines be inhaled in succession, ensure
that the medicine cup module is rinsed after every use.
Filling the atomiser
- Open the medicine container (1) by opening the clip, and fill the medicine container with an isotonic saline solution or medicine. Do not overfill! The maximum recommended fill amountis 8ml. (image 4, 5, 6 )
- Use medication only on the advice of a over physician, and check the appropriate inhalation duration and quantity for your needs
- If the recommended dose of the medicine is less than 2ml, add this quantity to at least 4ml using only isotonic saline solution. Viscous medicines may also require dilution. Always follow the instructions of your physician.
- Close the lid of the medicine container and lock the clip in place. (image 7)
Fitting accessories
- Attach the accessories (mouthpiece, adult it mask or child mask) firmly to the assembled medicine container(1).
- Place the device to your mouth and close your lips firmly around the mouthpiece. When using either of the masks, place it over the nose and mouth.
- Start the device using the on/off button (3)
- The spray mist coming from the device and the illuminated blue LEDs(3) show that the device is working perfectly.
Breathing technique
- The right breathing technique is important to ensure that the particles are distributed as widely as possible into the air ways. So that they can enter the air ways and lungs, you must inhale slowly and deeply, hold your breath briefly (5 to 10 seconds) and then exhale quickly.
- You should only use nebulizers to treat respiratory disorders if advised by your physician.
- Certain medicines are only available on prescription.
Notes
You should hold the device as upright as possible. Tilting it slightly will
not, however, affect its function, since the container is leak proof, To
ensure that the device functions correctly, make sure that you do not tilt it
more than 45° in any direction and that the medicine is in contact with the
mesh while in use.
Important
Essential oils, cough medicines, solutions designed for gargling, and drops
for application to the skin or for use in steam baths are wholly unsuitable
for inhalation using the nebulizer.
End of inhalation
- After treatment, switch off the device using the on / off button (3)
- The LEDs (3) go out.
- Once the inhalant has been fully sprayed out, the device will switches off automatically. For technical reasons, a small amount may still remain in the medicine container.
- If applicable, disconnect the adapter (10) from the A/C outlet.
Cleaning and disinfection
Warning
Observe the following hygiene instructions in order to avoid any health risks.
The nebulizer and accessories are intended for multiple use. Please note that
cleaning and hygienic reconditioning requirements differ according to the
different areas of application.
Notes
- Do not put pressure on the mesh with brushes or similar, as this may easily puncture the mesh causing irreparable damage, and will void the warranty. The metal is very thin and can be punctured easily.
- Prior to cleaning the body or the medicine chamber, always remove first the battery or adapter before cleaning as the liquid may leak into the electronic components. Any liquid that penetrates into the unit can damage the electrical parts and impair the function of the device.
Cleaning
- Remove the mouthpiece and any accessories from the atomiser.
- Add 6ml of clean water to the medicine container (1) and switch on the device to first remove any medicine residue from the mesh. (Please use up the entire liquid).
- Remove the batteries.
- If any traces of medical solutions or impurities remain on the gold-coloured contacts on the device and atomiser, use cotton buds to remove them.
Disinfection
Follow the steps listed below carefully in order to disinfect your atomizer
and the accessories. It is advisable to disinfect the individual components
after the last time they are used each day at the latest. Fist, clean the
atomiser and the accessories as described under ‘Cleaning’. You can then
continue with disinfection.
Disinfection with hot water
- Place 5ml of hot water inside the medicine cup module.
- Spray until the water inside the medicine module runs out.
Drying
- Carefully dry all components using a soft cloth.
- Shake the mesh atomiser lightly backwards and forwards(5 to 10 times), so that the water inside is removed.
- Place the individual components on a dry, clean and absorbent surface and allow to dry completely(for at least 4 hours).
Notes
Ensure that all parts are thoroughly dried after cleaning, otherwise there is
an increased risk of bacterial growth. Once all components are fully dry,
reassemble the nebulizer and place the components in a dry, sealed container.
Ensure that the medicine container has dried completely by shaking it.
Otherwise, nebulization may not function properly after reassembling the
device. If this is the case, disassemble the medicine container again and
shake it so that the water can escape. Once it has been reassembled, the
nebulizer should function properly again as usual.
Storage
- Do not store the nebulizer in a damp atmosphere(e.g. in the bathroom) or transport it together with damp objects.
- Store and transport the nebulizer away from direct sunlight.
Troubleshooting
Technical specifications
- Mains connection : 100-240V, 50-60HZ; 0. 15A
- Input : 3V/1A;
- Oscillation frequency : 1OOkHz±10%
- Condensation rate : c:Q.25ml/min.
- Atomized particles :cC 5µm
- Medication flow rate : approx. 0.4ml/min.
- Volume : max. 10ml
- Product weight: 115g±10%
- Size (LxWxH), 45x45x 100mm
- Housing material :ABS
- Operating conditions : Temperature:+5°C ~·+40°C;
- Relative humidity :<80% non-con-densing
- Storage and transport conditions
- Operating conditions : Temperature:-10 ·c -·+45 ·c;
- Relative humidity :<80% non-con-den sing