KIMO KT 120 Data Logger User Manual
- June 10, 2024
- KIMO
Table of Contents
KT 120 Data Logger
Precautions for use
Please always use the device in accordance with its intended use and within
parameters described in the technical features in order not to compromise the
protection ensured by the device.
Symbols used
For your safety and in order to avoid any damage of the device, please follow
the procedure described in this user manual and read carefully the notes
preceded by the following symbol: The following symbol will also be used in
this user manual: Please read carefully the information notes indicated after
this symbol.
Presentation of the device
Use
The KT 120 and KH 120 dataloggers of the HVAC range allow the internal
measurement of temperature only (KT 120) or of temperature and humidity (KH
120). This class of devices is dedicated to the food transport.
The devices have a male USB plug and an integrated software in PDF format
which enables to download and configure the datalogger without specific
software.
Applications
The KISTOCK datalogger is ideal for a temperature and humidity control for the
sensitive foodstuff storage, for example in the food industry or
pharmaceutical domain. It allows to control the temperature and humidity in
refrigerators, cold rooms, food trucks etc.
Therefore, the device guarantees a traceability all along the cold chain. And
at any moment the KISTOCK datalogger allows to edit easily and quickly a data
report in PDF format.
Description of the device
Description of the keys
“OK” key: allows to validate, start or stop the records, display the value
“Selection” key: allows the functions scroll
PC connection
Fixation
The KT 120 and KH 120 KISTOCK dataloggers have a magnetic mounting, so you can
fix it easily.
Technical features
Devices
KT 120 | KH 120 | |
---|---|---|
Units displayed | °C, °F | °C, °F, %RH |
Resolution | 0.1 °C, 0.1 °F | 0.1 °C, 0.1 °F, 0.1 %RH |
External input | USB connector | |
Internal sensor | Temperature | Temperature, humidity |
Type of sensor | NTC | Temperature: NTC Humidity: capacitive |
Measuring range | From -40 to +70 °C | Temperature: From -20 to +70 °C |
Humidity: From 0 to 100 %RH
**Accuracies***
|
±0.4 °C
(-20 °C<T<+70 °C)
±0.8 °C
(beyond)
| Temperature:
From -20 to 0 °C: ±2 % of the reading value ± 0.6 °C From 0 to 30 °C: ± 0.5 °C
From 30 to 70 °C: ± 1.5 % of the reading value Humidity: Accuracy (Repeatability, linearity, hysteresis):
±2 %RH (from 15 °C to 25 °C, from 5 to 95 %RH) Factory calibration uncertainty: ±0.88 %RH Temperature dependence:
±0.04 x (T-20) %RH (if T≤15 °C or T≥25 °C)
Setpoint alarms| 2 setpoint alarms on each channel
Number of points| 50 000
Frequency of measurement| From 1 minute to 24 hours
Working temperature| From -40 to +70 °C| From -20 to +70 °C
Storage temperature| From -40 to +85 °C
Battery life| 3 years| 500 days
European directives| 2011/65/EU RoHS II ; 2012/19/EU WEEE ; 2004/108/EC
EMC ; 2006/95/EC
Housing
Dimensions | 100 x 42.5 x 15.9 mm |
---|---|
Weight | 53 g |
Display | 1-line LCD screen |
Dimensions of screen: 32 x 25.5 mm
Control| 1 OK key
1 Selection key
Material| Compatible with food industry environment ABS housing
Protection| IP65: KT 120
IP40: KH 120
PC communication| 1 USB A male input
Battery power supply| 1 x CR2450 (button battery)
__
Environmental conditions of use
| Air and neutral gases
Humidity: in non condensing conditions Maximum altitude: 2000 m
Dimensions
Guarantee period
KISTOCK dataloggers have 1-year guarantee for any manufacturing defect (return to our After-sales service required).
Display
DATASET is finishedIndicates that one value is being recorded. It flashes: the
DATASET did not start already.
Flashing slowly: DATASET is between 80 and 90 % of the storage capacity.
Flashing quickly: DATASET is between 90 and 100 % of the storage capacity.
Constant: storage capacity full.
Constant: indicates that the batteries have to be replaced.
Indicates the channel number which is measuring.
The displayed values are the recorded
maximum/minimum values for the displayed channels.
Indicates the alarm action type: rising or falling action.
The temperature in °C.
Temperature in °F.
Relative humidity (KH 120).
The values to display selected during configuration via the software will
scroll on the screen every 3 seconds (only with the KH 120).
The display can be activated or deactivated via the KILOG software.
At extreme temperatures, the display can become hardly readable and its
display speed can slow down at temperatures below 0 °C. This has no incidence
on the measurement accuracy.
Functions of keys
OK key: enables to start, stop the dataset (press during >3 seconds) or to
change of scrolling group as described in the tables below.
Selection key: enables the scroll values in the scrolling group as described
in the tables below. * Please see the summary table of the groups organization
on page 10. Only with the KH 120. Groups organisation*
The table below summarises the groups organisation and measured values
available during a measurement dataset. Please see the summary table of the
groups organization on the following page. ** Only with the KH 120. Press key
to change of group.
Press key to scroll the values in the group.
Datalogger configuration with the integrated PDF file
The class 120 KISTOCK dataloggers have an integrated PDF file which allows to
configure quickly and easily the datalogger. Therefore, you can directly
configure your datalogger without opening the KILOG software.
Required configuration : to open this document, you need to use ONLY the “Adobe Acrobat Reader 9 ®” program (or higher), freely downloadable, which allows to read PDF format documents. Ensure you have installed it before starting.
-
Plug the class 120 KISTOCK datalogger on an USB port of your computer**. The following window opens:
-
Click on “Open folder to view files”.
Wait a few seconds (according to the dataset number of points), and a volume appears. -
Double-click on the “Configuration…” PDF file
The integrated configuration file opens: Choose the language: Select the required language:
Choose the date format: Select the required date format:
General information
Dataset name : this field allows to name the dataset.
Comments : this field allows to write comments on dataset.
Configuration
Recorder
Active screen : tick “ Yes ” to activate the screen display or “
No ” to deactivate it.
Management DST : for an automatic management of DST, tick “ Yes ” or
tick “ No ” to deactivate it. If you choose “ Yes ”, the “ Next time
change ” fields become accessible. Dates and times of the next time change
are proposed by default. You can modify them: click on the “ Date ” field
then on to display the calendar. Click on the required date. Click on the
“Hour” field to modify the time on which the next time change will be applied:
the time format is 00:00. On the last field, click on and choose “ +1h ”
to
add an hour or “ -1h ” to subtract an hour. The date and time change will
be
applied on the required date and time and will add or subtract an hour.
Recording
Interval : in the “Interval” field, inform the required interval duration
between two measurements, then select the unit (minutes or hours).
Start type
- Tick “Button” for a start type by button.
- Tick “Date” for a start type by date: inform the required date and time of the start. Click on the “Start date” field then click on to display the calendar and select the required date, or write it manually, with respect to the date format selected previously.
Stop type
Select the required stop type:
- The stop by date is available only if the start type by date has been chosen previously. If you choose “Date”, inform the required stop date and time in the “Stop date” field: click on to display the calendar then select the required date, or write it manually with respect to the date format selected previously.
- The stop by duration allows to determine a recording duration: inform the “Days” and “Hours” fields.
- The stop by number of items allows to determine a required number of measurement before the dataset stops. Inform the “Number of items” field (between 1 and 50 000 points).
- “Total memory” allows to record continuously up to 50 000 points before the dataset stops.
- “Loop” allows to record the values continuously and once the memory capacity reached, the last recorded values overwrite the first.
- “Stop by button”: tick “Yes” to allow a stop by button. Therefore, push the datalogger OK key during 3 seconds to stop the measurement dataset. To not allow it, tick “No”. The stop by button can not be deactivated if the chosen stop condition is “Loop”, “Total memory” or “Number of points”.
Channel parameters
Temperature (KT 120 and KH 120) and humidity (only the KH 120)
- Choose the temperature measurement unit: tick the “°C” or “°F” box.
- Active channel: tick “Yes” to activate the channel or “No” to deactivate it.
- Active alarm: tick “Yes” to activate the alarm or “No” to deactivate it.
If the alarm is activated, inform the “High” and “Low” fields to configure the alarm threshold. Inform the “Delay” field in number of points. According to the measurement interval previously configured, the duration is automatically actualised. For example: if an interval of 1 minute has been configured and that the delay in number of points for the high threshold is 5, the delay duration will be 5 minutes.
-
When the configuration is finished, click on the button to validate.
-
Chose the place to save the configuration: to use this configuration for the next dataset, save the configuration directly on the datalogger on “ Removable disk ”.
Check that the dataset is finished to ensure the new configuration is taken into account.
A message asks you to overwrite the existing file. -
Click on “Yes”.
-
To use this configuration later for another dataset, or to configure another device, you can save it on the required location. To recover it later, click on the button on top of the page, then select the required file on “.xdp” format, click on the button to use it for the next dataset
Datalogger download with PDF report edition
-
Plug the class 120 KISTOCK datalogger on an USB plug of the computer*.
Wait a few seconds, then the following window opens: -
Click on “Open folder to view files”.
The windows explorer opens. -
Double-click on the “Report” PDF file to visualise the dataset report.
-
The computer must be in compliance with the IEC60950 standard.
You can print it or export it in PDF format to integrate it easily to your documents.
Configuration, datalogger download and data processing with the KILOG
software
Please see the KILOG software user manual: “ KILOG-
classes-50-120-220-320 ”.
Device maintenance
Replace the battery
With 500 days to 3 years* battery life, KISTOCK guarantees long-term measurement.
To replace the battery:
- Unlock the battery hatch with a screwdriver or a coin.
- Turn towards the left until the marker aligns in front of the opened padlock symbol.
- Continue to turn until the hatch rises.
- Replace the battery (button battery CR 2450**) in such a way the + pole will be visible.
Only use trademark or high quality batteries in order to guarantee the
announced autonomy.
After the battery replacement, the device must be reconfigured.
Device cleaning
Please avoid any aggressive solvent.
Please protect the device from any cleaning produce containing formalin, that
may be used for cleaning rooms and ducts.
- On the basis of 1 measurement each 15 minutes at 25 °C
** The battery must be in compliance with the 60086-4 standard.
Calibration
All the KISTOCK devices have an integrated adjustment certificate in the
memory in PDF format which can be visualized and printed easily.
A calibration certificate is available as option in paper format.
We recommend to carry out a yearly checking.
Accessories
Only the accessories supplied with the device must be used.
Troubleshooting
Problem | Probable cause and possible solution |
---|
“hi” or “lo” is displayed.
| The measurement range is exceeded. There is a problem with the sensing element.
No value is displayed, only the icons are present.
| The display is configured on “OFF”. Configure it on “ON” with the KILOG
software (see page 14).
The display is completely off and there is no communication with the
computer.|
The battery has to be replaced. (see page 15).
BE CAREFUL! Material damages can happen, so please apply the
precautionary measures indicated.
Once returned to KIMO, required waste collection will be assured in the
respect of the environment in accordance to guidelines relating to WEEE.
www.kimo.fr