GAZELLE G9103 True RMS Digital Multimeters Instruction Manual

June 10, 2024
GAZELLE

GAZELLE G9103 True RMS Digital Multimeters

GAZELLE G9103 True RMS Digital Multimeters product

Overview

G9103 DMM is a small auto range, hand-held 3 5/6 bit true RMS multimeter featuring complete function, novel structure, high reliability and safety as well as large screen for display . It can be used for measuring AC/DC voltage and current, variable frequency voltage (V.F.C), resistance, diode, circuit on-off, capacitance, frequency ratio, CV non-contact AC voltage sensing and so on, is an ideal portable maintenance instrument for users.

 Accessory

Unpack and take out the instrument, please check following attachments carefully for completeness or intactness.In case of any shortage or damage, please contact with your supplier.

  1.  An operating instruction manual
  2. A test pen (CAT III 600V)10 A
  3. Point-type K( nickel chromium nickel silicon) thermocouple
  4.  An optional current clamp

Safety Information

G9103 is designed in compliance with standards such as IC/EN61010-1, pollution grade II, overvoltage (CATIII600) and double insulation standards. Please comply with operation instruction specified in the Manual; otherwise the protection provided by the instrument may be affected.
CE: This Meter complies with the standards EN 61010-1,EN 61010-2-030: in pollution degree 2, overvoltage category (CAT III 600V) and double insulation.
ETL/CETL: CONFORMS TO UL STD 61010-1, 61010-2-030
CERTIFIED TO CA STD C22.2 No. 61010-1, 61010-2-030

  1. It is forbidden to use the product without having rear cap in place, or otherwise there will be electric shocking.
  2. Prior to use, inspect the insulation layer of test pen for intactness, confirming no breakage and broken line.
  3.  When LCD display shows the icon ” *, it is required to replace the battery in time to ensure the measurement accuracy
  4. Range switch shall be set at the correct measurement position.
  5.  In case of electric shock and damage to the instrument, signals being measured shall not exceed rated limit value.
  6. To prevent any damage to the instrument, it is forbidden to change the gear of range switch in measurement.
  7. After each measurement, disconnect table pen and the circuit being measured; after the current measurement, especially the measurement of lame current it is necessary to power  off before disconnecting table per especially the measurement of large current its necessary to power on before disconnecting  table per and the circuit being measured.
  8. In case of electric shock. it is required to be cautious when voltage being measured higher than DC 60V or AC 30Vrms.
  9.  Do not use the product in high-temperature or high-humidity environment, particularly in the damp environment in where me instrument perormance may be severely degraded
  10.  Retrain from changing the internal wiring in the clamp ammeter to guard against damage to the meter and danger.
  11.  Clean the meter case with damo cloth and mild detergent rather than the abrasive material and solvent

Electric Symbol

GAZELLE G9103 True RMS Digital Multimeters 57

 Comprehensive Specification

  1.  Maximum voltage between input termiinal and grounding: see instruction about each input tenninal protection voltage.

  2.  10A terminal (CE) is equipped with: F 10A H 600V fast-acting fuse (<1>6×25) mm

  3.  mA/µA terminal (CE) is equipped with: FF 0.6A H 600V fast-acting fuse (<1>6×32) mm

  4. Maximum display: 6000Refresh 2~3 times per second, display “OL” in case of overrange .
    Capacitance and frequency : 9999 count.
    Duty ratio: 1~99.9%
    Diode: about 3.2V , displaying “OL” in case of overrange.
    Range: auto/manual
    Polarity: auto
    Working temperature: 0’C~40″C
    Relative humidity: ,s;75% when 0’C~30’C, and ,s;50% when 30’C~40’C
    Storage temperature: -10’C~50’C

  5. Electromagnetic compatibility:
    In 1V/m radio frequency (RF) field: Overall accuracy=specified accuracy+ 5% of range, and no specified index for RF over 1V/m.

  6.  Operating ASL: 0~2000m

  7.  Internal battery: AA R6P 1.5Vx2

  8.  Low electricity: LCD displays””.

  9.  Contour dimension: about (175x80x48.5) mm

  10. Weight: about 350g (inclusive of battery)

  11.  Safety Standard: IEC/EN 61010-1: CATIII600V; Pollution grade II

  12. Accreditation: CE

Appearance Structure

  •  Case
  • 2. LCD Display
  • 3/4/7.: Selection buttons
  • 5. Range Switch
  • 6. Measuring input terminal

VII. LCD Display

| Symbol| Instruction
---|---|---
1

**2**

3

4

5

6

| | Data hold prompt
| Negative reading
AC/DC| AC/DC measurement prompt
MAX-MIN| Maximum/Minimum/Maximum-Minimum value measurement prompt
| Under-voltage internal battery
Auto Range| Automatic range prompt

Knobs and Keys for Range selection

Range Location Function
**** AC or DC voltage measurement
Resistance measurement
Measurement of diode PN junction voltage
Measurement of circuit on-off
**** Capacitance measurement
Frequency measurement
Duty ratio masurement
**** Temperature measurement


| AC/DC current measurement
| AC/DC current clamp measurement
GAZELLE G9103 True RMS Digital Multimeters 33| Non-contact AC voltage sensing
| Switch off internal electric power

Button:
RANGE button: it can be used for selecting auto/manual range. After pressing, it will switch one gear of switch, when reading the highest gear, jump to the lowest gear range and in turn. When the time of pressing button is >2s or switching a range, you’ll exit the manual range mode. (Only applicable for VA//I)

MAX/MIN button:
it can be used to automatically enter the manual range mode. In such case, auto shutdown function is disabled and maximum value is displayed, after another pressing on the button, the minimum value will be displayed and values are displayed in turn (maximum value-minimum value). When the time of pressing button is >2s or switching a range, you’ll exit data recording mode (only applicable for V§, h, la and °C/°F)

REL button:
it can be used to automatically enter the manual range mode. The current displayed value will be taken as the reference value and then the difference between the measured value and reference value will be displayed, after another press, you’ll exit the relative measurement mode. (Only applicable for ****)

Hz/% button:
It can be used to select the mode Hz/%, only applicable for the selection of frequency, AC voltage/ current measurement modes.

SELECT button:
it can be used to select range (only applicable for multi-range). Under AC mode, press the button ≥2s, display “UFC”, enter V.F.C measurement mode and measure the variable frequency voltage. After another >2s pressing on the button, display “End” and exit the V.F.C measurement mode.

HOLD button: (Applicable for full range)
It can be used to lock and hold the displayed value, in such case, LCD displays the prompt “ “, after another press, it is unlocked and enter the normal measurement mode.
The backlight will be illuminated when the time of pressing button is ≥2s, after about 15s, the backlight will be automatically turned off; and the backlight will be turned off if pressing the key 2s when the backlight is illuminated.

Measurement Instruction

Check the built-in A 1.5V×2 battery, display will show the symbol ” when lack of power, and then replace battery in time. It is required to pay attention to the symbol ” beside the test pen socket, which reminds one of the fact that in case of measurement safety, testing voltage or current shall not exceed the specified value

AC and DC Voltage Measurement

  1.  Connect the instrument with the load in parallel for measurement.
  2.  When the input impedance of the instrument is about 10M , the load may cause measurement error in the circuit with high impedance. In most cases, the error can be neglected (0.1% or lower) if the circuit impedance is under 10k

Notes:

  • It is forbidden to input voltage higher than 600Vrms, despite of the possibility of measuring higher voltage, as it may damage the instrument.
  •  It is required to avoid the electric shock in measuring high voltage.

Resistance MeasurementGAZELLE G9103 True RMS Digital Multimeters
36 Connect the instrument with the load in parallel for measurement.

Notes:

  • The display will show “OL” when the measured resistance open-circuit or resistance value exceeds the maximum range of the instrument.
  •  Prior to measuring online resistance, it is necessary to switch off all powers in the circuits to be measured, and release all residual charges to ensure the measurement accuracy.
  • In measuring low resistance, a measurement error in resistance of about 0.1 8 – 0.2 Q will be resulted by the test pen. In order to acquire accurate reading, it is required to short circuit the test pen, take REL relative measurement mode to ensure the measurement accuracy.
  •  Check the test pen for any loosening or other reasons in case there is a resistance value no less than 0.50 when test pen is short circuited.
  •  Several seconds may be required for the reading stability when measuring high resistance, which is normal for high resistance measurement.
  •  By using the resistance measurement function, it is allowed to make self-checking of the built-in fuse,GAZELLE G9103 True RMS Digital Multimeters 37
  • No input higher than DC 60V or AC 30V is allowed.

3. Circuit on-off measurementGAZELLE G9103 True RMS Digital Multimeters
38 if the resistance of two terminals to be measured is bigger than 1509, there will be a circuit break and buzzer will make no sound; if the resistance is $100, the circuit is deemed with good conductivity and buzzer wall continuously sound.

Notes

  • Prior to measuring online circuit on-oft, it is necessary to switch off all power supplies in the circuits to be measured and release all resicua charges to ensure the measurement accuracy.
  • To prevent personal injury, it is not allowed to input the voltage higher than DO SOV @ACISIVA

Diode measurementGAZELLE G9103 True RMS Digital Multimeters
39 OL” will be displayed when the diode to be measured is an open circuit or polarity is reversely connected. For Silicon PN junction, the normal value is normally 500

Notes

  •  Prior to measring online diode. it is necessary to o switch off all power supples in the circults to be measured and release all residual charges to ensure the measurement accuracy.
  • Test voltage for diode is about
  •  To prevent personal injury, it is not allowed to input the voltage higher than DC 60V or AC 30V

Capacitance measurement

GAZELLE G9103 True RMS Digital Multimeters 40 The instrument, when without any input, will display a fixed reading which is the internal fixed capacitance value. When measuring small range gear capacitance, the above value shall be subtracted from the value to be measured to ensure the accuracy. The relative measurement REL function can be used to automatically subtract the value to facilitate the measurement.

Notes

  •  The display will show “OL” when the capacitor to be becomes short-circuited or the capacitance value exceeds the maximum range of the instrument.
  •  Generally, several seconds will be taken to measure high-capacity capacitor.
  •  To prevent damage to the instrument and personal injury, it is required before testing to release all residual charges. which is particularly important for capacitor with high voltage.

Frequency/Duty Ratio Measurement

 At the frequency measurement gear, press the button Hz/% to select frequency/duty ratio measurement mode.

Notes

  •  To prevent personal injury, it is not allowed to input the voltage higher than DC 60V or AC 30V

Temperature measurement

Temperature sensor: it is only applicable for K (Ni-Cr and Ni- Si) thermocouple. After startup, “OL’ is displayed complete Celsius or Fahrenheit temperature measurement D connecting the product with K-type temperature sensor

Notes
The point type K (Ni-Cr and N- Si) thermocouple is only applicable for me measurement of temperature under 230 /446 .

AC and DC current measurement Connect the instrument with the load in serial for measurement. AC measurement value will be true RMS.

Notes

  • Before connecting instrument in serial with the loop to be measured, switch off the power supply.
  •  In measurement, it is required to use proper input terminal and function gear; if unable to estimate the current, the measurement should start with the high gear range.
  • Fuses are provided inside the 10A, mA/ input jacks. It is forbidden to connect the table pen test pin in parallel with any circuit, especially the power supply terminal, which may cause damage to the instrument and personal injury. For security purposes, when measuring current higher than 5A, the time of each measurement should be controlled less than 10s and an interval of at least 15min should be maintained.
  •  When measuring AC current online, it is allowed to press the button Hz/% to display online AC frequency duty ratio.
  •  60A AC and DC current clamp measurement .GAZELLE G9103 True RMS Digital Multimeters 44

Connect as shown in the figure with the attached current clamp.

NCV Non-contact AC voltage sensing If need to detect whether there is a AC voltage or electromagnetic field, allow the front end of the instrument be close to the object to be sensed. The analog quantity of sensed AC voltage is about: “EF” is displayed when ,;;;critical voltage. “-” is displayed when G9103 >critical voltage, “- – – -” is designed according to the size of voltage Vd and accompanied by different buzzing sounds to mark the difference of sensed voltage.

Notes
Test pen is not required for measurement when the range knob switch is set at “NCV”.

Other functions:

  • After 2s of full display following startup, enter the normal measurement mode. “ErrE” will be displayed in case of any fault to the internal EEPROM.
  • Auto power-off: the instrument will be “automatically powered off’ to save energy in case of no operation to the knob switch within 15min. Under auto power-off statuts, any press on the any button will “automatically arouse” the instrument, or restart the instrument by turning the knob switch to OFF. Under power-off status, press on SELECT and power on, the buzzer will make five sounds to remind that the auto power-off function has been cancelled. Restarting the instrument after power-off will recover the auto power-off function.
  • Buzzer: A “Beep” sound (about 0.25s) from buzzer means the function button is valid when pressing any button or turning the function button. When measuring voltage or current: AC and DC voltage > about 600V mA gear AC/DC current 590mA When A gear AC/DC current is > 1 OA, buzzer will make continuous sounds to remain the over-range. In addition, the buzzer make make 5 continuous sounds about 1 min prior to auto power-off and then make one long sound prior to power-off. When the auto power-off function is cancelled, 5 sounds will be made by the buzzer every 15min.
  •  Low-voltage detection: it is used to detect the internal VDD. When the voltage is lower than 2.4V, the battery under-voltage symbol “” will be displayed, and the normal operation is available; while when lower than 2.2V, no operation is allowed and only the symbol is displayed after startup.
  • When the battery supply voltage is reduced to 2.6V, the LCD backlight is weak or can not start; measurement functions can still be used normally.

Technical Index

Accuracy: ± (a% reading +b word number), one year of guarantee period. Ambient temperature: 23·c ±5 C C73.4 F ±9°F) Relative humidity: <75%.

DC voltage measurement

Range Resolution Accuracy
60.00mV* 10µV ± (0.5%+2)
600.0mV** 0.1mV
6.000V 1mV

**** ± (0.7%+3)

60.00V| 10mV
600.0V| 0.1V
600V| 1V

1OM . (There will be unstable digital display in case of */** range open- circuit; after connecting with the load, it can be controlled s ±1)

  •  Range: 60mV range for G9103.
  • Range: 600mV range for G9103.

Maximum input voltage: ±600V

AC voltage measurement

Range Accuracy Resolution
60.00mV ± (1.0%+3) 10µV
600.0mV 0.1mV
6.000V ± (0.8%+3) 1mV
60.00V 10mV
600.0V 0.1V
600V + (1.0%+3) 1V
V.F.C 200.0V-600V 0.1/1V ± (4.0%+3)
  • Input impedance: about 10MO.

Display the true RMS. Frequency response: UT139A 45~400Hz, UT139B/C 45~1 kHz.(VFC: 45~400Hz) Assurance of accuracy:5~ 100% range, an allowance of <10 words of residual reading for short-circuit. It will be up to 3.0 when AC crest factor reaches full value (with except for 600V range, which is up to 1.5 when the range reaches the full value). Maximum input voltage: 600Vrms

Resistance measurement

Range Accuracy Resolution
600.00* ± (1.0%+2)
6.000kO ± C0.8%+2)
60.00kO 100
600.0kO 1000
6.000MO ± (1.2%+3)
60.00MO ± (1.5%+5)
  •  Range: measured value = displayed value – short-circuit value of test pen.
  •  Open-circuit: about *1 V
  • Overload protection: 600V-PTC

Circuit on-off, -M- diode measurement

Range Resolution Remark
**** 0.10 Circuit breakage resistance value is set as: > 150 Q, buzzer is

soundless.
Good conductivity is set as: ,s;10 Q buzzer sounds.
| 1mV| Open circuit voltage is about 3.2V
Normal voltage value of silicon PN junction is about 0.5~0.8V.

  • Overload protection: 600V-PTC

Capacitance Measurement

Range Resolution Accuracy
9.999nF 1pF Under REL status: ±(4%+10)
99.99nF~999.9µF 10pF~0.1 µF ±(4%+5)
9.999mF~99.99mF 1µF~10µF ± 10%(,s;2mF)
  • Overload protection: 600V-PTC

For capacitor  1 µF, it is recommended to adopt REL measurement mode to ensure measurement accuracy.

Frequency/duty ratio measurement

Range Resolution Accuracy
9.999Hz~9.999MHz 0.001Hz~0.001MHz ±(0.1%+4)
1%~99.9% 0.1% Not defined
  • Over-load protection: 600V-PTC

Input range a: (DC level is zero )
1OOkHz:100mVrms a 20Vrms

100kHz~1MHz: 200mVrms a 20Vrms
1MHz: 500mVrms a 20Vrms
5MHz~10MHz: 900mVrms a 20Vrms
Duty ratio %: only applicable for measurement 100kHz
Remark:

  • During measurement of AC voltage or AC current, if need to read online frequency value or duty ratio, following input should be met: frequency response: 1 kHz;
  •  AC voltage: mV range input ?,:100mV; V range input ?,: range x6%
  • AC current: input range a 4000/6000µA, 400/600mA, 1 0A range: a range x6%

400/600µA, 40/60mA, 4/6A range: range x60%

Temperature measurement

Range Resolution Accuracy
°C -40~ 1000″C -40~0°C
>0~100°C ±(1.0%+3)
> 100~ 1000°C ±(2.0%+3)

°F

| -40~1832″F| -40~32°F| | ±5

32~212°F| ± (1.5%+5)
212~1832°F| ± (2.5%+5)

  • Over-load protection: 600V-PTC

Remark: The point type K (Ni-Cr and Ni- Si) thermocouple is only applicable for the measurement of temperature under 230″C/446″F

DC current measurement

Range Accuracy Resolution
600.0µA **** ± (0.7%+2) 0.1µA
6000µA 1µA

mA| 60.00mA| 10µA
600.0mA| 0.1mA
A| 6.000A| ± (1.0%+3)| 1mA
10.00A| 10mA

  • Over-load protection:

µA mA range:
F1 fuse: (cp6x32)mm FF0.6A H 600V (CE)
10 A range: F2 fuse: (qi6x25)mm
F 10A H 600V (CE)

AC current measurement

Range Resolution Accuracy
600.0µA 0.1µA ±(1.0+3)
6000µA 1µA
mA 60.00mA 10µA
600.0mA 0.1mA
A 6.000A 1mA
10.00A 10mA

Frequency response: 45~ 1 kHz
Display: true RMS.
Assurance of accuracy: 5~100% range, an allowance of <2 words of residual reading for short-circuit. It will be up to 3.0 when AC crest factor reaches full value.

  • Over-load protection: (the same as the DC current over-load protection)

(60A) current clamp measurement

Range Resolution Accuracy
60A de 0.01A ±(1.0+3)
60A ac ±(1.2+3)
  • Over-load protection: 600V-PTC

Upkeep and Maintenance

warning: Power shall be switched off before opening the rear cover of the instrument; and the test pen shall be away from the input terminal and circuit to be measured.

  1. Conventional upkeep and maintenance
    •  For upkeep and maintenance, wet cloth and mild cleanser rather than abrasive or solution shall be used to clean the meter housing.
    • Please stop using and send for maintenance in case of any abnormal condition about the instrument.
    • The inspection or maintenance for instrument, if necessary, shall be performed by the qualified professional maintenance personnel or designated maintenance department.
  2. Battery or fuse tube replacement Built-in battery shall be replaced in time when LCD displays the under-voltage prompt” “, otherwise measurement accuracy may be affected.
    Battery specification: AA 1.5Vx2

Operating steps:

  1. Set the power switch at “Off”, take the test pen away from the input jack and remove the protective sleeve.

  2. Screw off one screw securing batter rear cover with screw driver, remove the cover and replace:
    The under-voltage used battery and burnout fuse:
    F1 fuse: FF0.6A H 600V (CE)

  3. Screw off the second screw securing the rear cover with screw driver and remove the cover to replace the burnout F2 fuse  mm F 10A H 600V (CE).