Alliedstar AS 200E Digital Optical Scanning Device User Manual
- June 10, 2024
- Alliedstar
Table of Contents
- Alliedstar AS 200E Digital Optical Scanning Device
- Product Informatio
- Technical Specifications
- Product Usage Instructions
- Safety Information
- Regulatory Information
- Regulatory Information
- Technical Specification
- Contact Information
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Alliedstar AS 200E Digital Optical Scanning Device
Product Informatio
The AS 200E is a digital optical scanning device designed for professional use only. It is intended to obtain digital impressions of hard and soft tissues such as teeth, gums, and mucous membrane by means of oral scanning, for oral restoration and orthodontic treatment of malocclusion. The device can be used for both adults and children in clinical practice.
Alliedstar’s intraoral scanners allow dental practitioners to acquire digital impressions with the quality and accuracy required for digital CAD/CAM dental applications. The device’s actual performance is dependent on the user’s training and operating execution, and the user is solely responsible for the accuracy, completeness, and adequacy of the acquired data.
Technical Specifications
Model: AS 200E
Technical Specifications:
- Digital optical scanning device
- Intended for professional use only
- Designed to obtain digital impressions of hard and soft tissues such as teeth, gums, and mucous membrane by means of oral scanning
- Can be used for both adults and children in clinical practice
Environmental Requirements:
- Operating temperature: 10°C to 40°C (50°F to 104°F)
- Storage temperature: -20°C to 70°C (-4°F to 158°F)
- Operating humidity: 10% to 80% non-condensing
- Storage humidity: 10% to 90% non-condensing
Product Usage Instructions
Before using the AS 200E, please read the safety information and regulatory information provided in the user manual carefully. The device is intended for professional use only and is restricted by U.S. Federal law to sale by or on the order of a dentist.
To use the AS 200E:
- Ensure that the device is not exposed to water spray as this can cause an electric shock or malfunction of the unit.
- Place the device in the operating temperature range of 10°C to 40°C (50°F to 104°F) and operating humidity range of 10% to 80% non-condensing.
- Follow the instructions in the user manual to scan the hard and soft tissues such as teeth, gums, and mucous membrane by means of oral scanning.
- After scanning, ensure that the data acquired is accurate, complete, and adequate for digital CAD/CAM dental applications.
If any serious incident occurs in relation to the device, the user must report it to Alliedstar and to the competent authority of its Member State in the European Union.
Notice
The Safety, Regulatory and Technical Specifications User Manual includes
information on the safety instructions, regulatory information, and the
technical specifications of the devices. We recommend that you thoroughly
familiarize yourself with this manual to make the most effective use of your
system.
The information in this document is subject to change. Neither Alliedstar nor
any of its subsidiaries shall be liable for errors contained herein or for
incidental damages in conjunction with the furnishing, performance, or use of
this material. No part of this publication may be reproduced without the
permission of Alliedstar.
This document is originally written in English.
All trademarks and registered trademarks are the property of their respective
holders.
The AS 200E is intended for professional use only.
U.S. Federal law restricts this device to sale by or on the order of a
dentist.
If any serious incident occurs in relation to the device, the user must report
it to Alliedstar and to the competent authority of its Member State in the
European Union.
Manual Name: AS 200E User Manual: Safety, Regulatory and Technical
Specification
Part Number: DD0068
Revision: 1.1
Date: 2022-11
Safety Information
Intended use
The AS 200E is a digital optical scanning device used to obtain digital
impressions of hard and soft tissues such as teeth, gums and mucous membrane
by means of oral scanning, for oral restoration and orthodontic treatment of
malocclusion.
The AS 200E could be used for both adult and children in clinical practice.
Clinical Benefits and Performance Characteristics
Alliedstar’s intraoral scanners benefit a dental practice by enabling
practitioners to acquire digital impressions with the quality and accuracy
required for digital CAD/CAM dental applications. The actual performance of
the device is dependent on the user’s training and operating execution. The
user is solely responsible for the accuracy, completeness, and adequacy of the
acquired data.
Conventions in This Manual
The following special messages emphasize information or indicate potential
risks to personnel or equipment.
WARNING: Warns you to avoid injury to yourself or others by following the safety instructions precisely.
Note: Alerts you to a condition that might cause serious damage.
Important: Alerts you to condition that might cause problems.
Tip: Provides extra information and hints.
Warnings and Safety Instructions
DANGER OF ELECTRIC SHOCK
This is an electrical unit. Do NOT expose it to water spray. Such action can
cause an electric shock or a malfunction of the unit.
IMPORTANT: All known residual risks, contraindications, or undesirable side effects are listed in this manual. If any serious incident occurs in relation to the device, you must report it to Alliedstar and to the competent authority of your Member State in the European Union.
WARNINGS
AS 200E:
- You MUST read and understand this safety information before using the scanner.
- This scanner shall only be used inside hospitals and other professional healthcare facilities and MUST NOT be used near high frequency surgical equipment and the RF shielded room of an ME System for magnetic resonance imaging, where the intensity of electromagnetic disturbance is high.
- Before using the scanner, check the outer surfaces of the unit and any accessories to ensure there are no rough surfaces, sharp edges, or protrusions which may cause a safety hazard.
- You are responsible for the operation and maintenance of the scanner. You MUST have training to use the scanner.
- DO NOT place objects within the field of operation of the unit.
- When the unit is not in use, ensure that the scanner is turned OFF.
- DO NOT use the scanner in conjunction with oxygen-rich environments. This unit is not intended for use with flammable anesthetics or flammable agents.
- DO NOT pull or twist the cable.
- DO NOT drop the scanner or the accessories.
- DO NOT sterilize the scanner.
- DO NOT expose the scanner to water spray or submerge it in water or disinfectant.
- DO NOT expose the scanner to high vibrations.
- DO NOT expose the scanner to ultraviolet radiation directly. The scanner is not designed for ultraviolet disinfection.
- DO NOT stare at the LED emission window.
- When the tip is removed, install the front protective cover to protect the scanner lens window.
- DO NOT remove the cover of any scanner components. The scanner contains no user-serviceable parts. For any repairs, contact a qualified Alliedstar service technician.
- DO NOT replace the cables provided with the scanner with other cables. Doing so may damage the scanner and adversely affect the safety protection and EMC performance of the scanner.
- Any other equipment not complying with IEC-60601 shall be kept at least 1.5 meters away from the patient.
- If the equipment is faulty, turn it OFF, display an “Out of Service” notice, and contact a qualified Alliedstar service technician.
- Using components, accessories, cables, and spare parts other than those specified or provided by the manufacturer of this equipment may impair the safety protection of the scanner and may result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
- No modification of this equipment is allowed.
- Additional multiple outlet strips or extension cords should not be connected to the system.
- The maximum temperature of the applied part may reach to 43 °C; to avoid overheating, do not use it for extended periods.
- To power off the scanner, push the power button for 3 seconds. To isolate the holder from power supply, unplug the USB connector from the USB port.
- DO NOT maintain or service this equipment while it is in use with the patient.
- Connection of the PEMS (Programmable Electrical Medical System) to an IT NETWORK that includes other equipment could result in risks to patients, operators, or third parties. The responsible organization should identify, analyze, evaluate, and control these risks.
- Patients with oral mucosal disease, mental illness, severe respiratory disease, asthma, Parkinson’s disease, hyperactivity disease are forbidden.
- Patients with moderate or severe opening limitation should use it with caution.
Computer:
- Do NOT place any equipment which does not comply with IEC 60601-1 in the immediate vicinity of the patient. Leave at least 1.5 meters distance between the patient and the equipment.
- The scanner is only intended to be connected to a computer that is at least IEC 60950 / IEC 62368, or equivalent standards certified. Connecting the scanner to other equipment may be hazardous.
- See the installation guide for your computer for information about the data processing system, computer, and screen. Leave a sufficient amount of clear space around the computer to ensure that it is properly ventilated.
- Position the screen to avoid light reflections from internal or external lighting for maximum image quality and visual comfort.
Battery:
- Do not dismantle, open, or shred secondary cells or batteries.
- Do not expose cells or batteries to heat or fire. Avoid storage in direct sunlight.
- Do not short-circuit a cell or a battery. Do not store cells or batteries haphazardly in a box or drawer where they may short-circuit each other or be short-circuited by other metal objects.
- Do not remove a cell or battery from its original packaging until required for use.
- Do not subject cells or batteries to mechanical shock.
- In the event of a cell leaking, do not allow the liquid to come in contact with the skin or eyes. If contact has been made, wash the affected area with copious amounts of water and seek medical advice.
- Do not use any charger other than that specifically provided for use with the equipment. Refer to the manufacturer’s instructions or equipment manual for proper charging instructions.
- Do not use any cell or battery which is not designed for use with the equipment.
- Always purchase the battery recommended by the device manufacturer for the equipment.
- Keep cells and batteries clean and dry.
- Wipe the cell or battery terminals with a clean, dry cloth if they become dirty.
- Do not leave a battery on prolonged charge when not in use.
- After extended periods of storage, it may be necessary to charge and discharge the cells or batteries several times to obtain maximum performance.
- Retain the original product literature for future reference.
- Use only the cell or battery in the application for which it was intended.
- When possible, remove the battery from the equipment when not in use.
- Please remove battery using specified tools. Removing the battery by hand is not allowed
- Dispose of the battery properly.
Disposal:
This equipment contains certain materials and chemical compounds incidental to
the manufacture of electrical and electronic equipment, and improper “end-of-
life” disposal of such equipment can result in environmental contamination.
Therefore, this equipment should not be disposed of as ordinary household
waste but should instead be delivered to a designated electrical and
electronic waste disposal or recycling center. For further information on
disposing of electrical and electronic waste, contact the cognizant authority
within the local jurisdiction.
Dispose of the scanner tips according to standard operating procedures or
local regulations for the disposal of contaminated medical waste. For
additional scanner tips, contact your dealer.
Cleaning, Disinfecting, Sterilizing
Perform the following maintenance activities on your scanner and accessories
regularly.
To ensure maximum hygienic safety for the patient, carefully follow the
instructions to prepare the scanner for use.
To ensure maximum hygienic safety for the patient and to minimize the risk of
cross-contamination, carefully perform the following maintenance activities on
your scanner and accessories.
After each patient:
- Clean and disinfect the scanner. See “Clean and Disinfect the Scanner”).
- Clean the tip, and then perform Autoclave Sterilization (See “Clean the Tip” and “Sterilize the Tip”). There are 2 models of tip as following:
Model
|
Size
|
UDI- DI
| Manually Cleaning| 134 ° C
Sterilization
---|---|---|---|---
TP202| Large| (01)06973993441034| √| √
TP203| Small| (01)06973993441041| √| √
Clean and Disinfect the Scanner
General Warnings
WARNINGS
- Read and follow the warnings and personal protection instructions provided in the Safety Data Sheet (SDS) for the disinfectant used to process the scanner.
- You must wear gloves while cleaning and disinfecting the scanner.
- The scanner must be disinfected with a recommended intermediate-level disinfectant solution with tuberculocidal activity between patients.
- DO NOT use a disinfectant containing phenolics or iodophors; doing so will damage the surface coating of the scanner.
- Never put the scanner in an autoclave device or immerse it in water or the disinfectant solution.
- Excessive fluids can damage the scanner.
- Do not use cotton, cloth, or tissues soaked with disinfectant to disinfect the scanner.
Clean the Scanner
If the scanner is visibly contaminated with blood and/or body fluids, you must
clean it before disinfecting it.
To clean the scanner, follow these steps:
- Dampen (do not soak) a lint-free cloth with lukewarm water.
- Remove the blood and/or body fluids with the dampened lint-free cloth.
Disinfect the Scanner
After each patient, the scanner must be thoroughly disinfected.
To adequately disinfect the scanner, follow the disinfectant manufacturer’s
instructions for the appropriate contact time.
Important: If the scanner is visibly soiled, it must be thoroughly cleaned prior to disinfecting. See “Clean the Scanner”.
To disinfect the scanner, follow these steps:
-
Remove the replaceable tip.
-
Remove all visible soil (see “Clean the Scanner”).
-
Use a commercially prepared intermediate level disinfectant wipe. Follow the manufacturer’s instructions for contact time.
Recommended disinfectant wipes: CaviWipes
WARNING: Using a disinfectant that has not been approved may cause damage to the scanner. -
Thoroughly wipe all surfaces of the scanner, DO NOT allow liquid to enter through the gap, air outlet,or pin holes.
WARNING: Do not rinse. -
Allow to air dry.
-
After the scanner has dried, use a clean, lint-free cloth dampened with water to remove residual disinfectant from the surface of the scanner.
Clean and Sterilize the Tip
WARNINGS
- Wear gloves when handling a contaminated scanner tip.
- Read and follow the warnings and personal protection instructions provided in the manufacturer’s SDS for the detergent used to clean the scanner tip.
- Do not soak the scanner tips in disinfectant for a long period.
- Dry the scanner tips thoroughly before mounting onto the scanner.
- Do not use an ultrasonic cleaning machine to clean the scanner tips.
- Do not soak the scanner tip with alcohol-based disinfectants.
Clean the Tip
To manually cleaning the tips, follow these steps:
- Rinse excess soil from the Tips (2 minutes).
- Using a soft brush, apply enzymatic detergent solution (e.g., Metrex EmPower) to all surfaces.
- Rinse under clean, running water (2 minutes).
- Inspect Tips. If the tip is not clean, repeat the steps 1-3.
- Use a lens tissue or lint-free cloth to remove any dust from the mirror in the tip.
Sterilize the Tip
Scanner tips received from the manufacturer are NOT sterilized. You must sterilize the tips before the first use.
Important: If you limit the exposure time at 134°C to not more than 6 minutes, you can autoclave the tip up to 60 cycles.
To sterilize the scanner tips, follow these steps:
- Put the tip into a sealed steam sterilization pouch.
- Place the tips in a steam autoclave for sterilization.
- For TP202 and TP203, exposure temperature should be set as 134°C.
- For TP202 and TP203, exposure time should exceed 3 minutes.
- For TP202 and TP203, exposure time should not exceed 6 minutes.
Important:
- For TP202 and TP203, exposure temperature should be set as 134°C.
- For TP202 and TP203, exposure time should exceed 3 minutes.
- For TP202 and TP203, exposure time should not exceed 6 minutes.
Precautions Before Use
Perform the following activities on your product and accessories before use.
Visually Inspecting the Scanner for Damage
Visually inspect the scanner for damage or signs of deterioration by doing the
following:
- Inspect the scanner’s lens window.
- Inspect around the scanner buttons and contact points.
If damage is noted, do not use the scanner and contact your representative.
Visually Inspecting the Scanner Tips
Visually inspect the scanner tips for signs of deterioration by doing the
following:
- Verify that the tip is not damaged and its components are not detached.
- Verify that the tip mirror does not have any smudges or scratches on it.
If deterioration is noted, replace the tip.
WARNINGS
- The lens window on the scanner is a delicate optical component. Mount the front protective cover to protect the lens window from damage and dirt when the scanner is not in use.
- The mirror in the tip is a delicate optical component. Its clean and undamaged surface is critical to scan quality.
In the event that you see poor scan quality or an unclear video preview in the software, clean the tip mirror and the scanner’s lens window using a microfiber cleaning swab, applying ethanol that is free of impurities.
Regulatory Information
Marking and Labeling Symbols
| Type BF applied part symbol classification in accordance with IEC 60601
standards.
---|---
| In the European Union, this symbol indicates: DO NOT discard this product in
a trash receptacle; use an appropriate recovery and recycling facility.
Contact your local sales representative for additional information on the
collection and recovery programs available for this product.
| Manufacturer
| Date of Manufacture
| Refer to instruction manual/booklet.
| Direct current.
| Sterilizable in a steam sterilizer (autoclave) at 134°C
| Model Number
| Serial number
| Rechargeable battery
| CE marks is a statement by medical equipment manufacturers that meet all the
General Safety and Performance Requirements (GSPR) of all EU related medical
instruments, and the CE mark is also a legal requirement for medical devices
to list in the EU market.
| The Federal Communications Commission (FCC) coordinates domestic and
international Communications by controlling radio, television,
telecommunications, satellites, and cables. Involving more than 50 states,
Columbia, and the territory of the United States to ensure the safety of radio
and wire communications products related to life and property. Many radio
applications, communications products and digital products require FCC
approval to enter the U.S. market.
---|---
| Name of the European authorized representative and address of the registered
place of business.
Regulatory Information
The AS 200E complies with the following regulations:
MDR: (EU) 2017/745 Medical Device Regulation, Cass I following the Rule 5.
FDA Center for Devices & Radiological Health CDRH – Title 21 CFR 872.3661
(USA).
Medical Devices Regulations (Canada).
RoHS: Restriction of the Use of Certain Hazardous Substances in Electrical and
Electronic Equipment, 2011/65/EU Annex II and its amendment Directive (EU)
2015/863
RED: Directive 2014/53/EU The Radio Equipment Directive
FCC: Part 15 of the Federal Communications Commission Rules
ISED: Innovation, Science and Economic Development Canada
Compliance with European and International Standards
- EN / IEC 60601-1: Medical Electrical Equipment, Part 1: General Requirements for Basic Safety and Essential Performance
- ANSI/AAMI ES 60601-1: Medical Electrical Equipment – Part 1: General requirements for basic safety and essential performance
- CAN/CSA-C22.2 No. 60601-1: Medical Electrical Equipment – Part 1: General requirements for basic safety and essential performance
- EN / IEC 60601-1-2: Medical Electrical Equipment, Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests
- EN / IEC 80601-2-60: Medical electrical equipment — Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment
- EN / IEC 62471: Photobiological safety of lamps and lamp systems
- EN / ISO 17664: Processing of health care products – Information to be provided by the medical device manufacturer for the processing of medical devices
- EN / ISO 17665-1: Sterilization of health care products — Moist heat — Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices
- EN / IEC 60601-1-6: Medical Electrical Equipment, Part 1-6: General requirements for basic safety and essential performance – Collateral Standard: Usability
- EN / IEC 62366-1: Medical devices – Part 1: Application of usability engineering to medical devices
- EN / IEC 62304: Medical device software – Software life cycle Processes
- EN ISO 10993: Biological evaluation of medical devices
- ISO 14971: Medical devices – Application of risk management to medical devices
- EN / ISO 15223-1: Medical devices – Symbols to be used with medical device labels, labeling and information to be supplied – Part 1: General requirements
- EN ISO 20417: Medical devices — Information to be supplied by the manufacturer
- ISO 9687: Dentistry – Graphical symbols for dental equipment
- AAMI TIR 12:Designing, testing and labeling reusable medical devices for reprocessing in health care facilities: A guide for medical device manufacturers
- AAMI TIR 30:A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices
- EN / IEC 62133-2: Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications – Part 2: Lithium systems
- EN 50566: Product standard to demonstrate the compliance of wireless communication devices with the basic restrictions and exposure limit values related to human exposure to electromagnetic fields in the frequency range from 30 MHz to 6 GHz: hand-held and body mounted devices in close proximity to the human body.
- EN 301489-1: Electromagnetic compatibility (EMC) standard for radio equipment and services; Part 1: Common technical requirements; Harmonized Standard covering the essential requirements of article 3.1(b) of Directive 2014/53/EU and the essential requirements of article 6 of Directive 2014/30/EU
- EN 301489-17: Electromagnetic Compatibility (EMC) standard for radio equipment and services; Part 17: Specific conditions for Broadband Data Transmission Systems; Harmonized Standard covering the essential requirements of article 3.1(b) of Directive 2014/53/EU
- EN 301893: 5GHz RLAN; Harmonized Standard covering the essential requirements of article 3.2 of Directive 2014/53/EU
Classification in Accordance with EN/IEC 60601-1
Type of protection against electric shock: Internally powered
Degree of protection against electric shock: Type BF Applied Part
Mode of operation: Continuous operation
- Flammable anesthetics : Not suitable for use in the presence of flammable anesthetics or a mixture of flammable anesthetics with air or oxygen or nitrous oxide.
Conformity with EN/IEC 60601-1-2
IEC 60601-1-2 EMC requirements and tests, Medical Electrical Equipment
including CISPR 11 Group 1, Class B.
Electromagnetic Compatibility Precautions
Medical electrical equipment requires special precautions regarding
electromagnetic compatibility (EMC). Medical equipment must be installed and
put into service according to the EMC information provided in this
documentation.
Other equipment can interfere with communications with the AS 200E, even if
the equipment complies with CISPR emissions requirements.
Warning: Portable and mobile RF communications equipment (including
peripherals such as antenna cables and external antennas) should be used no
closer than 30 cm (12 inches) to any part of the AS 200E, including cables
specified by the manufacturer. Otherwise, degradation of the performance of
this equipment could result.
WiFi
The AS 200E intraoral scanner operates with the 802.11a/n/ac protocol.
Following channels may be used for the handpiece:
Channel| Protocol| Central Frequency
(MHz)
| Output Power (Nominal)
(dBm)
---|---|---|---
48| 802.11a| 5240| 14.57
157| 802.11a| 5785| 14.23
Above channels are available for CE, FCC, SRRC and ISED.
This device complies with part 15 of the FCC Rules and contains license exempt
transmitter(s)/receiver(s) that comply with ISED’s license-exempt RSS(s).
Operation is subject to the following two conditions:
- This device may not cause interference.
- This device must accept any interference, including interference that may cause undesired operation of the device.
Note: Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
0.798 W/Kg,10g for CE
0.011 W/Kg,1g for FCC
Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
Limited by local law regulations, the version for North America does not have
a region selection option. The device is for indoor use only and operates in
the 5150-5250 MHz and 5725-5850MHz band to reduce the potential for harmful
interference to co-channel mobile satellite systems.
Note
This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio
frequency energy and, if not in-stalled and used in accordance with the
instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more
of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/ TV technician for help.
Guidance and Manufacturer’s Declaration
Guidance and Manufacturer’s Declaration – Electromagnetic Emission
The AS 200E is intended for use in the electromagnetic environment specified
below. The customer or user of the AS 200E should assure that it is used in
such an environment.
Emissions Test| Compliance| Electromagnetic Environment –
Guidance
---|---|---
RF Emissions CISPR 11| Group 1 Class B| The AS 200E uses RF energy only for
its internal function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity for
Equipment
and Systems (from Non-RF wireless communication equipment)
The AS 200E is intended for use in the electromagnetic environment specified
below (IEC 60601-1-2 Table 4). The customer or the user of the AS 200E should
assure that it is used in such an environment.
Immunity Test| IEC 60601 Test Level| Compliance Level| Electromagnetic Environment –
Guidance
---|---|---|---
Electrostatic discharge (ESD) IEC 61000-4-2| ±8 kV contact
±15 kV air
| ±8 kV contact
±15 kV air
| Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative
humidity should be at least 30%.
---|---|---|---
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
| 30 A/m| 30 A/m| Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital
environment.
Radiated RF IEC 61000-4-3| 3 V/m
80MHz – 2.7GHz
| 3 V/m
80MHz – 2.7GHz
| Environment of a professional healthcare facility.
WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the AS 200E including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment
could result.
NOTE: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the AS 200E is used exceeds the applicable RF compliance level above, the AS 200E should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the AS 200E.
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity for
Equipment and Systems (from RF wireless communication equipment)
For the immunity to proximity fields from RF wireless communications
equipment, the AS 200E is compliant with the test levels specified below,
according to IEC60601-1-2 Table 9. The customer or user of the AS 200E should
assure that it is used in such an environment.
Test Frequency (MHz) | Band (MHz) | Immunity Test Levels |
---|---|---|
385 | 380 – 390 | Pulse modulation 18Hz, 27V/m |
450 | 430 – 470 | FM, ±5 kHz deviation, 1 kHz sine, 28V/m |
710 | 704 – 787 | Pulse modulation 217Hz, 9V/m |
745
780
810| 800 – 960| Pulse modulation 18Hz, 28V/m
870| |
---|---|---
930
1720|
1700 – 1990
| Pulse modulation 217Hz, 28V/m
1845
1970
2450| 2400 – 2570| Pulse modulation 217Hz, 28V/m
5240|
5100 – 5800
|
Pulse modulation 217Hz, 9V/m
5500
5785
Accessories
The use of accessories other than those specified, with the exception of those
sold by the manufacturer of the equipment, as replacement parts for internal
components may result in increased emissions or decreased immunity of the
medical equipment.
Other Equipment
WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify normal operation.
Technical Specification
Model
AS 200E
AS 200E Technical Specifications
Components | Technical Specifications |
---|---|
Weight | Scanner (including battery, without tip): 245 g Tip Large – |
Sterilization (TP202): 13g
Tip Small – Sterilization (TP203): 12g
Color| 3D full color
USB wireless Card Port| USB 3.0
USB charge Port| Type C Port
Field of View| Tip Large – Sterilization (TP202): 16mm x 14mm
Tip Small – Sterilization (TP203): 12mm x 12mm
Depth of Filed| 15mm
Configuration Requirement of Workstation| Processor: Intel® CoreTM i7 9th
Generation, base frequency 2.6 GHz (or better)
Memory: 32 GB (or more) DDR4, frequency 2666 MHz (or better)
Disk: 512G (or more) SSD
Graphics card: NVIDIA® GeForce® GTX 1650 (or better)
Display: 15.6″ FHD (1920 x 1080) (or more) Others: USB 3.0 port
Operating system: Windows 10 (build 18362 or newer) / Win 11, 64 bit Optional: Touch screen
Important: It is MANDATORY to check that your system configuration is compatible with the computer system requirements for the AS 200E software.
AS 200E Environmental Requirements
Components | Environmental Requirements |
---|---|
Operating Temperature | 15°C ~ 30°C |
Transport and Storage Temperature | -10°C ~ 50°C |
Operating Relative Humidity | 10% ~ 65% RH |
Transportation and Storage Relative Humidity | 10% ~ 95% RH |
Operating Atmospheric Pressure | 70 ~ 106 KPa |
Transportation and Storage Atmospheric Pressure | 60 ~ 106 KPa |
Contact Information
Manufacturer
Alliedstar Medical Equipment Co., Ltd.
No.1-4, Floor 3, Unit 2, Building 3, No.222, West third section, Waihuan Road,
Yanjiang
District, Ziyang, 641300, Sichuan, P.R. China
E-mail:
Authorized Representatives
Authorized Representative in the European Community
SUNGO Europe B.V.
Olympisch Station 24 ,1076DE Amsterdam, Netherlands
Alliedstar Medical Equipment Co., Ltd.
For more information, visit: www.allied-star.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>