TESTEC TT-SI 7005 60MHz Active Differential Probe Instruction Manual
- June 10, 2024
- TESTEC
Table of Contents
TESTEC TT-SI 7005 60MHz Active Differential Probe
Safety Summary
To avoid personal injury and/or product damage, review and comply with the following safety precautions. These precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this probe.
** A WARNING statement calls attention to an operating procedure,
practice, or condition, which, if not followed correctly, could result in
injury or death to personnel.
** A CAUTION statement calls attention to an operating procedure,
practice, or condition, which, if not followed correctly, could result in
damage to or destruction of parts or the entire product.
Do Not Work Alone
Do not work alone when working with high voltages.
Inspect the Probe
Inspect the probe and accessories for cracks and frayed or broken leads before
each use. If defects or damages are noted, DO NOT USE the probe.
Dry Conditions
Hands, shoes, floor, and work bench must be dry. Avoid making measurements
under humidity, dampness, or other environmental conditions that might affect
safety.
Do Not Remove the Probe’s Casing
Removal of the probe’s casing may expose you to electric shock. If necessary,
disconnect the inputs and outputs of the probe before opening the case.
Hazardous Contact
To avoid injury, remove jewelry such as rings, watches, and other metallic
objects. Do not touch exposed connections and components when power is
present.
Unexpected Charges
Hazardous voltages may be present in unexpected locations in circuitry being
tested when a fault condition in the circuit exists.
Capacitors inside the instrument may retain a charge even if the instrument is
disconnected from its source of supply.
Use Only in Office-Type Indoor Setting
The probe is designed to be used in office-type indoor environments. Do not
operate the probe:
- In the presence of noxious, corrosive, flammable fumes, gases, vapors, chemicals, or finely-divided particulates.
- In environments where there is a danger of any liquid spilled on the probe.
- In air temperatures exceeding the specified operating temperatures.
- In atmospheric pressures outside the specified altitude limits or where the surrounding gas is not air.
Not for Critical Applications
This probe is not authorized for use in contact with the human body or for use
as a component in a life-support device or system.
Do Not Substitute Parts
Do not install substitute parts or perform any unauthorized modification to
the instrument.
Only Qualified Personnel
Only qualified personnel should use this probe. This differential voltage
probe is designed to be used by personnel who are trained, experienced, or
otherwise qualified to recognize hazardous situations and who are trained in
the safety precautions necessary to avoid possible injury when using such a
device.
Observe Maximum Working Voltage
Do not use any probe above its maximum working voltage ranges. See
specifications on page 7.
Use Proper Power Source
Do not operate this probe from a power source that applies more than the
voltage specified.
Must be Grounded
This probe is grounded by the shell of the BNC connector through the grounding
conductor of the power cord of the measurement instrument. Before making
connections to the input leads of this probe, ensure that the output BNC
connector is attached to the BNC connector of the measurement instrument, and
that the measurement instrument is properly grounded. Whenever it is likely
that the ground protection is impaired, you must make the instrument
inoperative and secure it against any unintended operation.
Terms and Symbols
The following symbols appear on the product or in its documentation:
Direct voltage
Both direct and alternating voltage
Caution, possibility of electric shock
**** Caution, see documentation for details
Earth Ground
Definitions
Measurement Category II (CAT II)
refers to local-level electrical distribution, such as that provided by a
standard wall outlet or plug-connected equipment. Examples of CAT II
measurements would be household appliances, portable tools, and similar
modules.
Measurement Category III (CAT III)
refers to measurements on hard-wired equipment in fixed installations,
distribution boards, and circuit breakers that form part of a building wiring
installation. Other examples are wiring, including cables, bus bars, junction
boxes, switches, socket outlets in the fixed installation, and stationary
motors with permanent connections to fixed installations
Pollution Degree 2
refers to an operation environment where normally only dry, non-conductive
pollution occurs. Temporary conductivity caused by condensation can be
expected.
Working CAT rating is equal to that of the lowest rated element within the
test set-up.
Compliance Statements
EC Declaration of Conformity
The product conforms to the applicable European Union requirements per IEC/EN
61010-031:2015 Safety requirements for electrical equipment for measurement,
control and laboratory use. Part 31: Safety requirements for hand-held probe
assemblies for electrical measurement and test.
EU RoHS Compliance
The probe and accessories conform to the 2011/65/EU RoHS2 Directive.
Disposal of Old Electrical & Electronic Equipment
(Applicable in the European Union and other European countries with separate
collection systems). This product is subject to Directive 2012/19/EU of the
European Parliament and the Council of the European Union on waste electrical
and electronic equipment (WEEE), and in jurisdictions adopting that Directive,
is marked as being put on the market after August 13, 2005, and should not be
disposed of as unsorted municipal waste. Please utilize your local WEEE
collection facilities in the disposition of this product and otherwise observe
all applicable requirements.
This probe is in compliance with IEC/EN 61010-031:2015 CAT III, Pollution
Degree 2.
Introduction
Overview
The TT-SI 7005 differential probe allows safe, accurate measurement between two voltage points where neither point is referenced to ground. The probe is designed for high sensitivity measurements up to 60 MHz bandwidth. Compatible with oscilloscopes from all major manufacturers, the probe is exclusively powered by the included 9 V power adapter.
Features
- Meets IEC/EN 61010-31:2015 safety standard
- Selectable attenuation settings of x1, x5 and x10
- 60 MHz bandwidth @ x10 attenuation setting
- 50 MHz bandwidth @ x5 attenuation setting
- 15 MHz bandwidth @ x1 attenuation setting
- Up to ±35 V (DC + AC peak) common mode
- High accuracy (±2%)
- Power indicator LED
- Over range indicator LED
Initial Inspection
This unit is tested prior to shipment. It is therefore ready for immediate use upon receipt. An initial physical inspection should be made to ensure that no damage has been sustained during shipment. After the inspection, verify the contents of the shipment.
Delivery Content
- Differential probe – TT-SI 7005
- 2 x Grabber, black & red – TT-SI GR2
- 2 x test leads with insulated 4 mm banana plugs – TT-SI TL7 silicone jacket, black & red
- insulated BNC cable, 100 cm – TT-SI BN7
- 9 V power adapter – EU version – TT-SI NT7
- User manual
Using the Probe
WARNING
At the time of powering on the probe, the input leads must not be connected to
an item to be tested. Never operate the probe with the case open.
CAUTION
This probe is used to carry out differential measurements between two points
on the circuit under test. This probe is not designed for electrically
insulating the circuit under test or the measuring instrument.
Inspection Procedure
- Connect the BNC output connector to the vertical input of the oscilloscope.
- Power on the probe.
- Set the attenuation setting on the oscilloscope to match the probe.
- Connect the input of probe to a function generator. Then select a square-wave output of 10 V amplitude and 100 kHz frequency.
- The square-wave will be displayed on the screen of the oscilloscope. This indicates the probe is working properly.
Note that the oscilloscope should show the same voltage and frequency as the function generator.
Getting Started
- Connect the BNC output connector to the vertical input of a general purposed oscilloscope. The oscilloscope must have a ground referenced.
- Connect power adapter.
- Select the proper attenuation ratio. When measuring signals below 3,5 V, switch the attenuation ratio to x1 in order to get higher resolution and less noise ratio. For voltages up to 18 V, set the attenuation ratio to x5 and for voltages up to 35 V set the attenuation ratio to x10.
- The power indicator LED should turn on.
- Plug the supplied grips onto the test leads and connect the circuit under test.
Vertical Scale on Oscilloscope
The actual vertical scale of the oscilloscope is equal to the attenuation
factor multiplied by the range of vertical scale selected on the oscilloscope.
For example, with the probe on factor x00 and the oscilloscope on 0.5 V/div,
the real vertical scale is 10 x 0.5 = 5 V/div.
With the probe on x5, the real vertical scale is 5 x 0.5 = 2,5 V/ div. These
values apply when the oscilloscope is set to the typical 1 MΩ impedance input.
When the scope is set to 50 Ω input, the actual vertical scale will be
doubled: 10 V/div for the x10 setting and 5 V/div for the x´5 setting.
Scope Input Impedance| Probe Attenuation
Setting| Actual Attenuation Setting| Vertical Scale Reading
on Oscilloscope| Actual Vertical Scale of the Oscilloscope
---|---|---|---|---
1 MΩ| x1| x1| 0,5 V/div| 0,5 V/div
1 MΩ| x5| x5| 0,5 V/div| 2,5 V/div
1 MΩ| x10| x10| 0,5 V/div| 5 V/div
50 Ω| x1| x2| 0,5 V/div| 1 V/div
50 Ω| x5| x10| 0,5 V/div| 5 V/div
50 Ω| x10| x20| 0,5 V/div| 10 V/div
Product Overview
Included Accessories
Specifications
All specifications apply to the unit after a temperature stabilization time of 20 minutes over an ambient range of 25°C ± 5°C.
TT-SI 7005 – order number 15176
Bandwidth| 60 MHz @ x10
50 MHz @ x5
15 MHz @ x1
Rise Time| 5,8 ns @ x10 7 ns @ x5 23 ns @ x1
Attenuation| x1, x5, x10
DC-Accuracy| ±2%
CMRR| 80 dB @ 60 Hz
60 dB @ 100 Hz
50 dB @ 1 MHz
Maximum Input Voltage @ x1 (DC + AC peak)| ±3,5 V
Maximum Input Voltage @ x5 (DC + AC peak)| ±18 V
Maximum Input Voltage @ x10
(DC + AC peak)
| ±35 V
Absolute Maximum Input Voltage (each side to ground)| 35 Vrms
Input Impedance (differential)| 1 MOhm // 2,5 pF
Input Impedance (each side to ground)| 0,5 MOhm // 6 pF
Output Voltage Swing| ±4 V using 1 MΩ oscilloscope input
Offset (typical)| ±5 mV
Noise (typical)| 2 mVrms
Source Impedance| 50 Ohm
Power Supply| 9 V power adapter (included)
Weight| 250 g
Dimensions| 195 x 55 x 30 mm
BNC cable length| 100 cm
Input lead length| 55 cm each
Operation Tempearure/Humidity| 0°C to 50°C / 10% to 85% RH
Storage Tempearure/Humidity| -30°C to 70°C / 10% to 90% RH
Safety Standard| IEC/EN 61010-031:2015 CAT III
Pollution Degree| 2
Specifications are subject to change without notice. To ensure the most current version of this manual, please download the current version from our website: www.testec.de
Voltage Derating Curve
The derating curve of the absolute maximum input voltage in common mode is shown as follows:
Cleaning
This probe does not require any particular cleaning. If necessary, clean the
case with a soft cloth.
WARNING
Dry the probe thoroughly before attempting to make voltage measurements.
CAUTION
Avoid immersing or using abrasive cleaners or solvents containing Benzene (or
similar solvents) on the probe as these can cause deterioration of the probe
body and cables.
Service & Warranty Information
Limited One-Year Warranty
Testec Elektronik GmbH warrants these products to be free from defective
material or workmanship for a period of 1 year from the date of original
purchase. Under this warranty, Testec Elektronik GmbH is limited to repairing
the defective device when returned to the factory, shipping charges prepaid,
within the warranty period.
Units returned to Testec Elektronik GmbH that have been subject to abuse,
misuse, damage, or accident, or have been connected, installed, or adjusted
contrary to the instructions furnished by Testec Elektronik GmbH, or that have
been repaired by unauthorized persons, will not be covered by this warranty.
Testec Elektronik GmbH reserves the right to discontinue models, change
specifications, price, or design of this device at any time without notice and
without incurring any obligation whatsoever.
The purchaser agrees to assume all liabilities for any damages and/or bodily
injury which may result from the use or misuse of this device by the
purchaser, his employees, or agents.
This warranty is in lieu of all other representations or warranties expressed
or implied and no agent or representative of Testec Elektronik GmbH is
authorized to assume any other obligation in connection with the sale and
purchase of this device.
Service
If you have a need for calibration or repair services, technical, or sales
support, please contact us:
Testec Elektronik GmbH
Voltastr. 8-10
63303 Dreieich / Germany
Tel.: +49 (0) 69 – 9433350
E-Mail: service@testec.de
www.testec.de