Hanna Instruments HI510 Universal Process Controllers Instruction Manual
- June 11, 2024
- HANNA instruments
Table of Contents
- Single and DualChannel Universal Process Controllers
- Features Displayed on Screen (HI520 display shown)
- Inputs and Outputs
- Data and Event Logging
- Enclosure Features
- Panel-Mount
- Wall-Mount
- Pipe-Mount
- Specifications
- Ordering Information
- ORP and Temperature Industrial Smart Probes
- pH and Temperature Industrial Smart Probes
- Documents / Resources
Hanna Instruments HI510 Universal Process Controllers
Single and DualChannel Universal Process Controllers
- Waterproof IP65 (NEMA 4X) enclosure
- Large backlit LCD
- Multi-color LED status indicators
- Audible alarm
- Tactile rubberized keypad
- Universal mounting
- Universal Hanna digital probe
HI510 and HI520 are advanced process controllers that can be configured for applications requiring monitoring and/ or control of four main water-analysis parameters: pH, ORP, Conductivity, and Dissolved Oxygen.
These controllers feature a digital probe input(s) that automatically detects and updates the controller with the parameter that it measures.
HI520 is Hanna’s first dual-input process controller that accepts virtually any combination of compatible probes.
Designed to adapt to unique process control requirements, users have the option to enable or disable each channel independently.
Additionally, HI520 introduces enhanced industrial control by operating a control-loop system whereby users have the option to run channel control either independently or configure to be triggered sequentially upon reaching the other channel’s set point(s) (1, 2, or both). The device operates a logical channel with built-in mathematical functions.
This function is intended for when the controller works as an analyzer for monitoring high/low parameter levels between two identical inputs with identical measuring configurations.
These Universal Process Controllers offer wall, pipe, and panel mounting options and feature a large backlit dot matrix display for easy viewing and provide an intuitive interface for setup options.
The controllers utilize multi-color LEDs for easy viewing of the instrument’s status including relay activation, in alarm mode, or in hold status. All programming operations are done through the low profile vulcanized rubber keypad or with an RS485 connection to a Modbus-compliant supervisory computer
Features Displayed on Screen (HI520 display shown)
Control Modes
The control mode can be configured to be On/Off, Proportional, or PID.
The mode can be set high or low. High control mode is required if the process
value is too high and needs to be decreased. Low control mode is required if
the process value is too low and needs to be increased.
For On/Off control, the hysteresis band is adjustable, while in Proportional
and PID modes, deviation, control period, and other tuning parameters can be
set to optimize control around a set point.
For HI520, each channel can run control independently or sequentially.
Hold Function
During calibration, cleaning, and configuration the controller automatically
goes into Hold mode. During Hold mode all control loops related are disabled.
The analog outputs may be configured to go to a fixed value or remain at the
last value.
The Hold function can also be triggered manually, using an external digital
input or by entering in Manual mode. This is useful for disabling control when
performing maintenance.
Auto-Cleaning Cycle
Difficult applications often require an almost continuous maintenance of the
probe. Processes with high-suspended solids, fats, oils, pigments, and
microorganisms will coat the pH sensing glass, ORP sensors, and the reference
junction. The cleaning function allows programming of one or more wash cycles
and uses the relays to activate valves, pumps or compressed air based on the
type of washing that is required to maintain probes for reliable results.
Configurable Alarm System
The alarm system is configurable for measured parameters. The alarm can also
be activated by event triggers or abnormal operation. For example, if a dosing
relay remains closed for an excessive period of time or if temperature exceeds
an upper limit during an exothermic neutralization reaction. A blinking red
LED signals an alarm state.
All relays configured for control are inactivated until the alarm state is
resolved.
On HI520 use channel configuration to setup and trigger an alarm.
LCD Information
Local visual indicators of measurement details as well as errors are
displayed. The ? DIAG key provides details of the issue.
? DIAG – Help and Diagnostic Key
The help and diagnostic key (? DIAG) provides information related to errors;
or in setup mode, information about settings.
Inputs and Outputs
Digital Inputs
- Two digital inputs for remotely triggering, cleaning and hold functions
Analog Outputs
- Up to 4 analog outputs and 5 relays used for control and for sending a signal to data loggers, PLC, SCADA and other remote monitoring systems
- 0-20 mA or 4-20 mA
- Scalable in selecting values for the range
- Can be used for control of pumps and valves
- On alarm state can output a 22 mA signal to the monitoring system
Digital Communication
-
The Modbus-compliant unit can be integrated within a Modbus-based network and connected to other industrial electronic devices. The following tasks may be accomplished remotely:
· Monitoring, using the virtual LCD (limited to a single remote control in the entire network)
· Setup
· Loading the Setup configuration file to a controller -
RS485 Digital output for PC and other device connectivity
Relays
- Up to 5 control and 1 alarm electromechanical relays
- Replaceable 5A fuses to protect all relays
- Extractable terminal blocks for easy wiring
- Relay options include single pole double throw (SPDT) and single pole single throw (SPST)
- Control relays can be programmed for On/Off, Proportional, or PID control as well as Cleaning and Hold functions
- Configurable alarm relay
- Relays terminal blocks and their wires paths are separated from the low voltage section for additional safety
Data and Event Logging
Automatic Data-logging
HI510 and HI520 have built-in data logging that stores data at selectable
intervals along with relay control settings and calibration data.
- Data stored in up to 100 lots with each holding 8600 records
- Selectable log interval: from 10 seconds to 3 hours
- Logged Data includes:
- Measured values and alarms for all supported parameters
- Calibration information including solutions used. For pH, the offset and slope is recorded.
Event Log
The Event Log file can hold a maximum of 100 events. It includes errors,
alarms, warnings, calibration events, configuration changes, and cleaning
events.
Password protection
The controllers feature password protected calibration and setup.
Enclosure Features
NEMA 4X Enclosure
The controllers are suitable for indoor or outdoor environments. The NEMA 4X
enclosure ensures the electronics are protected against splashing and hose-
directed water or windblown dirt, dust, rain or sleet. It also provides
corrosion protection for use near salt water.
Cable Glands
To maintain the NEMA 4X enclosure rating during use, the conduit openings and
connection cables are sealed against the environment using the provided cable
glands, seals, and plugs.
Spring Loaded Screws
The front panel is hinged at the front of the enclosure for easy access to
wiring locations.
It features spring loaded screws that won’t fall out when accessed.
USB Type-C Port
Logged data can be transferred to a flash drive as a .csv file using the USB
Type-C port.
A rubberized plug helps protect the port against the ingress of water.
Panel-Mount
HI510-01 Panel-Mount Kit
The controllers can be securely panel mounted through a ½ DIN opening using the HI510-01 Panel-Mount Kit. The kit includes a sealing gasket, dual zinc coated brackets, and associated hardware.
Wall-Mount
The controllers can be surface mounted using the HI510-02 WallMount Kit. The
kit includes a zinc coated mounting plate and associated hardware. The plate
may be oriented vertically or horizontally.
HI510-02 Wall-Mount Kit
Pipe-Mount
The controller can be mounted to a vertical or horizontal pipe using the
HI510-03 Pipe-Mount Kit. The kit includes a zinc coated mounting plate,
associated hardware, and U-bolts to accommodate a 1”, 1 ½”, or 2” pipe.
Specifications
Specifications | HI510 | H1520 |
---|---|---|
Digital Probes | See the following pages | |
Channels | 1 | 2 |
Display | Graphic LCD, 128 x 64 pixel B/W with backlight | |
Digital Inputs | 2 independent galvanically isolated inputs (conf igurable for |
Hold ET Cleaning functions) On state: 5 to 24 Vdc, low or high level
active
Analog Outputs| 2 or 4 independent, galvanically isolated outputs 0- 22 mA
configurable as:
0 – 20 mA; 4 – 20 mA
22 mA as alarm signal, configurable option
Analog Output Accuracy| ±0.2%f.s.
Digital Communication| R5485 serial port -Remote monitoring and control USB-C
port- Retrieve log files and firmware upgrading
Relays| Up to 5 relays (independently configurable for process variables, Hold
& Cleaning functions)
Electromechanical relay SPOT and PST contact outputs
SA – 250 Vac; SA – 30 Vdc (resistive load) Fuse protected:5A, 250V slow blow fuse
Alarm Relay for All Measurement Alarms| Electromechanical relay SPOT contact
output SA – 250 Vac; SA – 30 Vdc (resistive load) Fuse protected:5A, 250V slow
blow fuse
Data Logging| Interval log, up to 100 files, maximum 8600 records on each
stored file. When the maximum limit of 100 stored files is reached, the most
recent file will automatically erase the oldest one.
Event log, maximum 100 records. When the maximum limit is reached, the last record overwrites the oldest one.
Power Supply| 100 – 240 Vac ±10%; 50/60 Hz;15VA; fuse protected (2A, 250V slow
blow fuse)
Power Consumption| 15VA
Installation Category| II
Environment| -20 to 50 °C (-4 to 122 °F); maximum 100% RH non-condensing
Enclosure”| Single case V2 DIN, type 4X, IP65 ingress protection
Weight| Approximately 1.6 kg (3.5 lb.)
Dimensions| Width:144.0 mm (5.7″) Height:144.0 mm (5.7″) Depth:151.3mm
(6.0 “)
Ordering Information| HI510-0540 universal process controller with 5
relays and 4 analog outputs is supplied with 3m power cable, cable gland set,
instrument certificate, and quick reference guide with instructions for manual
download.
H1510-0320 universal process controller with 3 relays and 2 analog outputs is supplied with 3m power cable, cable gland set, instrument certificate, and quick reference guide with instructions for manual download.
HI520-0540 universal process controller with 5 relays and 4 analog outputs is supplied with 3m power cable, cable gland set, instrument certificate, and quick reference guide with instructions for manual download.
H1520-0320 universal process controller with 3 relays and 2 analog outputs is supplied with 3m power cable, cable gland set, instrument certificate, and quick reference guide with instructions for manual download.
Accessories| HI510-01 panel-mount kit| HI76510-10 patch cable, 10 m
(32’91
HI510-02 wall-mount kit| HI76510-15 patch cable,15 m (49’2
“)
HI510-03 pipe-mount kit| H176510-25 patch cable, 25 m (82′)
HI76510-05 patch cable, 5 m (16’51| HI76510-50 patch cable, 50 m
(164′)
- For a water tight seal: Gland seals need to be used over cables and the four screws on the front casing need to be tightened to 13.3 lbf·in (1.5 N·m, maximum 2.0 N·m), of torque.
Specifications | example HI2004-18zz |
---|---|
ORP | Range |
Resolution | 1 mV |
Accuracy | ±2 mV |
Calibration | Single point, value can be adjusted |
±60 mV around measured mV
Temperature| Range| –5.0 to 100.0 °C (23.0 to 212.0 °F)
Resolution| 0.1 °C; 0.1 °F
Accuracy| ±0.5 °C; ±1.0 °F
Calibration| Single point offset (controller setting)
Temperature Source| Automatic (from Probe) Manual
Body| PVDF
Junction| PTFE
Sensor| Material| platinum
Body| glass
Tip shape| flat
Diameter| 22 mm (0.9”)
Insertion length| 21 mm (0.8”)
Maximum Pressure| 6 bar
Threaded Connection| 3/4” NPT external thread for mounting
Wetted Parts| Probe body| PVDF
Sensor body
O-ring
| glass
NBR
Junction| PTFE
Matching pin| titanium
Protection Rating| IP68
Cable Length| zz, see Ordering Information
PTFE junction: Minimizing the potential for clogging and chemically
resistant, PTFE is ideal for samples with high content of suspended solids or
for high-pressure installations.
Ceramic junction: Porous chemically resistant plug that connects the
reference electrode to the process electrically.
Ordering Information
Each probe is supplied with quick reference guide with probe quality certificate
Choose your configuration:
xx| 04
14
| PTFE junction
Ceramic junction
---|---|---
y| 1| Platinum sensor; –5.0 to 100.0 °C (23.0 to 212.0 °F); ±2000 mV
2| Gold sensor; –5.0 to 100.0 °C (23.0 to 212.0 °F);± 2000 mV
8| Smart probe, with RS485 connection
zz| 00 supplied with DIN connector (without cable). See page 15.56 (HI510 and
HI520 accessories section) for patch cable ordering codes.
05, 10, 15, 25, 50 fixed cable length (in meters)
See compatible probe extension cables on page page 15.56, see additional mounting accessories on page 15.25, see probe holders on page 15.134 and 15.130.
ORP and Temperature Industrial Smart Probes
HI2004-18 • HI2014-18
HI2004-28 • HI2014-28
HI510 and HI520 Universal Process Controller compatibleHI20X4-18 and HI20X4-28 are ORP and temperature probes designed for use with the Hanna Instruments® HI510 or HI520. The system is designed to monitor and control disinfection chemicals or follow and control a critical oxidation (or reduction) reaction. mV measurements are auto-compensated for barometric pressure and temperature.
- HI2004-18 and HI2014-18 platinum-sensor series are designed to provide the best response over a wide range of applications. Used in reducing processes such as chlorine dosing in pools and spas or chromate reduction.
- HI2004-28 and HI2014-28 gold-sensor series are designed for oxidative processes such as cyanide oxidation in the mining industry.
The probe features a flat tip (ideal for solutions containing aggressive chemicals) which virtually eliminates dirt deposits and significantly reduces maintenance.
Suitable for continuous measurement of ORP required for monitoring and/or controlling oxidizers and reducing agents in water treatment monitoring, industrial effluent treatment, and swimming pools, the probe can be submersed/immersed using the 3/4” NPT threads, or installed directly in-line or in a flow cell installation, using the lower sensor threads.
The probe is also available with an integral connector that permits cable connections after installation.
Patch cables may be purchased separately to connect between the probe and controller up to 50 meters (164 ft).
- Rugged, chemically-resistant PVDF body
- 3/4” NPT external thread
- 6 bar (87 psi) maximum pressure at 25 °C (77 °F)
- Built-in temperature sensor for measurement
- Digital probe stores model, firmware, serial number, and calibration information
- Minimum maintenance
Dimensions
pH and Temperature Industrial Smart Probes
HI1006-18 • HI1006-38 • HI1006-48
HI1016-18 • HI1016-38 • HI1016-48
HI510 and HI520 Universal Process Controller compatible
These industrial pH probes are intended for industrial process control when
paired with the HI510 or HI520 Universal Process Controller.
- HI1006-18 and HI1016-18 series, designed for low conductivity or low-temperature process environments
- HI1006-38 and HI1016-38 series, designed for extended pH range or high-temperature process environments
- HI1006-48 and HI1016-48 series, designed for process environments where hydrofluoric acid is present
An integral temperature sensor measures water temperature and adjusts the
probe signal over the specified temperature range.
The flat tip eliminates deposits that can foul the sensor and significantly
reduces necessary maintenance.
The PVDF body material is easy to clean and disinfect, and resistant to most
chemicals (e.g. solvents, sodium hypochlorite), ultraviolet light, and fungal
growth.
Suitable for continuous measurement of pH required in applications such as
urban wastewater treatment, industrial effluent treatment, and surface water
monitoring, the probe (and accessories) can be installed directly in-line,
immersed in a tank, or in a flow cell.
The probe is available with an integral connector that permits connections
after installation.
Patch cables may be purchased separately to connect between the probe and
controller up to 50 meters (164 ft).
- Rugged, chemically-resistant PVDF body
- Specialized glass sensor for fast stabilization and accurate results
- 3/4” NPT external thread for mounting
- Built-in temperature sensor for measurement and compensation
- Matching pin helps avoid typical problems caused by grounding loop current
- Digital probe stores model, firmware, serial number, and calibration information
Specifications | HI1006-18zz – example |
---|
pH
| Range| 0.00 to 12.00 pH
Resolution| 0.1 pH or 0.01 pH
Accuracy| ±0.02 pH
Calibration| Up to three-points (option to select from five standard buffers)
Temperature| Range| –5.0 to 80.0 °C (23.0 to 176.0 °F)
Resolution| 0.1 °C; 0.1 °F
Accuracy| ±0.5 °C; ±1.0 °F
Calibration| Single point offset (controller setting)
Temperature Compensation| Automatic| 0.0 to 80.0 °C (32.0 to 176.0 °F)
Temperature Source| Automatic (from Probe) Manual
Body| PVDF
Junction| PTFE
Sensor| Body| LT glass
Tip shape| flat, self-cleaning
Diameter| 22 mm (0.9”)
Insertion length| 17 mm (0.7”)
Maximum Pressure| 6 bar (87 psi) at 25 °C (77 °F)
Threaded Connection| 3/4” NPT external thread for mounting
Wetted Parts| Probe body| PVDF
Sensor body| Glass
O-ring| NBR
Junction| PTFE
Matching pin| Titanium
Protection Rating| IP68
Cable Length| zz, see Ordering Information
PTFE junction: Minimizing the potential for clogging and chemically
resistant, PTFE is ideal for samples with high content of suspended solids or
for high-pressure installations.
Ceramic junction: Porous chemically resistant plug that connects the
reference electrode to the process electrically.
LT glass: Fast stabilization and accurate results at lower temperatures
or lower ion content.
HT glass: Fast stabilization and accurate results at higher temperatures
and over a larger pH range.
HF glass: Fast stabilization and accurate results for aggressive
applications that have fluoride ions (F¯<2 g/L, temperature<60 °C, pH>2).
Ordering Information
Each probe is supplied with quick reference guide with probe quality
certificate.
Choose your configuration:
xx | 06 | PTFE junction |
---|---|---|
16 | Ceramic junction |
y
| 1| Low Temperature (LT) glass sensor, titanium matching pin
–5.0 to 80.0 °C (23.0 to 176.0 °F)
0.00 to 12.00 pH
3| High Temperature (HT) glass sensor, titanium matching pin 0.0 to 100.0 °C (32.0 to 212.0 °F)
0.00 to 14.00 pH
4| Fluoride-resistant (HF) glass sensor, titanium matching pin
–5.0 to 60.0 °C (23.0 to 140.0 °F)
0.00 to 10.00 pH
8| Smart probe, with RS485 connection
zz
| 00 supplied with DIN connector (without cable). See Accessories section for patch cable ordering codes.
05, 10, 15, 25, 50 fixed cable length (in meters)
Accessories| HI76510-05| Patch cable, 5 m (16’5”)
HI76510-10| Patch cable, 10 m (32’9”)
HI76510-15| Patch cable, 15 m (49’2”)
HI76510-25| Patch cable, 25 m (82’)
HI76510-50| Patch cable, 50 m (164’)
Dimensions
Documents / Resources
|
Hanna Instruments HI510 Universal Process
Controllers
[pdf] Instruction Manual
HI510 Universal Process Controllers, HI510, Universal Process Controllers,
Process Controllers, Controllers
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>