Product Information: HI6321 Advanced Conductivity Benchtop Meter

The HI6321 Advanced Conductivity Benchtop Meter is a laboratory and industrial use device that measures conductivity. The package contents include:

  • HI6321 Conductivity USB-C meter
  • Electrode arm holder (3)
  • Probe (4)
  • Power cable (5)
  • HI70031, HI70039, and HI70030 conductivity solutions (20mL each)
  • HI70000 electrode rinse solution (~20mL)

The meter has a capacitive touchscreen interface with back, home, and menu keys. It also has channel settings, users, system, measurement log recall, help, and settings options.

The meter conforms to CE European Directives and UK Standards and is ISO 9001 certified. It comes with a two-year warranty, and the probe has a six-month warranty against defects in workmanship and materials when used for their intended purpose and maintained according to instructions.

Product Usage Instructions

  1. Connect the electrode arm holder to the meter and plug in the probe.
  2. Connect the power cable to the meter and plug it into the mains.
  3. Press the ON/OFF button to start the meter.
  4. Remove the transparent film that protects the capacitive touchscreen prior to operating the meter.
  5. Use the electrode rinse solution to clean the probe before use.
  6. Select the appropriate conductivity solution based on your measurement needs.
  7. Dip the probe into the selected conductivity solution and wait for the reading to stabilize.
  8. Use the capacitive touchscreen interface to navigate through the different options and settings.
  9. For account creation and logins, power on the meter and tap “Users” followed by “Account Management” to enter the account creation settings.
  10. For technical support, email tech@hannainst.com or visit www.hannainst.com.

Dear Customer, thank you for choosing Hanna Instruments.

Package contents

  • HI6321
  • HI7631233 four-ring conductivity probeHANNA instruments HI6321 Advanced Conductivity Benchtop Meter-fig-1
  • 1413 µS/cm conductivity standard solution (4 sachets)
  • 5000 µS/cm conductivity standard solution (2 sachets)
  • 12880 µS/cm conductivity standard solution (2 sachets)
  • Electrode rinse solution (2 sachets)HANNA instruments HI6321 Advanced Conductivity Benchtop Meter-fig-2
  • Electrode arm with base plate and screws (mounting)HANNA instruments HI6321 Advanced Conductivity Benchtop Meter-fig-3
  • 24 VDC power adapter and USB-C to USB-A cableHANNA instruments HI6321 Advanced Conductivity Benchtop Meter-fig-4
  • Probe quality certificate, instrument quality certificate, quick reference guide

Hanna® is committed to developing and deploying digital solutions with a positive impact on the environment and climate.
Please scan the QR code or follow the link to download the user manual. https://manuals.hannainst.com/HI6321

HANNA instruments HI6321 Advanced Conductivity Benchtop Meter-
fig-6

Getting started

Identify the base plate (1) – to be attached to the L or the R side – and mounting screw (2) and proceed as illustrated here.HANNA instruments HI6321
Advanced Conductivity Benchtop Meter-fig-7

With the arm holder (3) in position and the probe (4) plugged in, connect the power cable (5) and plug into the mains. Press the ON/OFF button (6) to start.
Remove the transparent film that protects the capacitive touchscreen prior to operating the meter.HANNA instruments HI6321 Advanced Conductivity Benchtop
Meter-fig-8

Operating language & Access to support

  • At first power on default operating language is English. Browse the language window and select operating language from list of supported options.
  • To access support, tap and ? select.
  • On first access the benchtop’s name is used as default user name and no password is required.

Device handling & Safety precautions

  • Transport the unit with all cables removed.
  • Do not puncture the capacitive touchscreen.
  • Do not insert objects into the ports, spaces around keys other than the intended cable, USB drive.

Capacitive touchscreen

HANNA instruments HI6321 Advanced Conductivity Benchtop Meter-
fig-10

  • return to a previous hierarchical menu level
  • access to the measurement screen & configured profile
  • access to:
    • Users – login & rights configuration
    • System Settings – system configuration, connectivity & printing
    • Measurement Settings – measurement configuration
    • Log Recall – logged measurement data
    • Help – support
  • from the measurement screen, tap to access the measurement menu

Probe usage & Calibration tips

Accessories: rinse container, calibration container, calibration standard, DI water
Prior to placing the probe into a standard:

  • Remove the transparent protective ring (1).
  • Rinse in DI water and gently shake off excess water.
  • Immerse the probe in rinse container ensuring that probe’s holes (3) are completely submerged. Swirl and remove.
  • Immerse and swirl the probe several times to prime the probe. HANNA instruments HI6321 Advanced Conductivity Benchtop Meter-fig-15

Calibration tips

  • Ensure the probe is correctly plugged in.
  • Use fresh standards with a value close to that of the sample.
  • Pour fresh standard in calibration container, immerse the probe, gently swirl, and remove.
  • Immerse and swirl the probe several times.
  • Gently tap the probe to remove trapped air bubbles.
  • Wait for the temperature to stabilize and the probe to acclimate to the solution.
  • Calibrate frequently and discard standard after use.

Conductivity measurement tips

  • Work with recently calibrated probes.
  • Substitute calibration standard with measured sample and follow calibration recommendations.
  • Use the arm holder for easy probe transfer in and out of container and correct probe positioning, away from container walls or bottom.
  • TDS measurement: use correct TDS conversion factor for measured sample.
  • Record measured data once the reading is stable.
  • Clean and rinse off any residual sample on the sensors (2) after measurement to keep the probe in optimum condition.HANNA instruments HI6321 Advanced Conductivity Benchtop Meter-fig-16

Logging in & Creating a new account

  1. Power the meter and tap
  2. Tap followed by .
  3. Tap Edit Users to enter Account Management.HANNA instruments HI6321 Advanced Conductivity Benchtop Meter-fig-18
  4. Tap to enable Account Creation and Logins. Tap to return.
  5. Tap Logout to enter Users screen. Tap the plus symbol avatar.
  6. Input user name and tap .
  7. Enter password and tap  Re-enter password to confirm.

All Hanna instruments conform to the CE European Directives and UK Standards, and our production facilities are ISO 9001 certified. HI6321 is warranted for a period of two years and the probe for six months against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. For technical information e-mail us at tech@hannainst.com. Visit www.hannainst.com for more information about Hanna Instruments and our products.