T3DAQ1-16
Data Acquisition System
User Manual
T3DAQ1-16 User Manual
General Safety Summary
Read the following precautions carefully to avoid any personal injuries, or
damage to the instru- ment or products connected to it. Use the instrument
only as specified.
Use only the power cord supplied for the instrument.
Ground the instrument.
The instrument is grounded through the ground conductor of the power cord. To
avoid electric shock, always connect to grounded outlets. Make sure the
instrument is grounded correctly before connecting its input or output
terminals.
Connect the signal wire correctly.
To avoid damage, observe input polarity and maximum voltage/current ratings at
all times.
Observe all terminal ratings and signs on the instrument to avoid fire or
electric shock.
Before connecting to the instrument, read the manual to understand the
input/output ratings.
Do not operate with suspected failures.
If you suspect that the instrument is damaged, contact the Teledyne LeCroy
service department immediately.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep the surface of the instrument clean and dry.
Avoid touching exposed circuits or wires.
Do not touch exposed contacts or components when the power is on.
Do not operate without covers.
Do not operate the instrument with covers or panels removed.
Use only the fuse specified for the instrument.
Use proper overvoltage protection.
Use anti-static protection.
Operate in an anti-static protected area. Ground measurement cable conductors
before connecting to the instrument to discharge any static electricity before
connecting the cables to the instrument.
Observe ventilation requirements.
Ensure good ventilation. Check the vent and fan regularly to prevent
overheating.
Safety Terms and Symbols
The following terms may appear on the instrument:
DANGER: Direct injury or hazard may occur.
WARNING: Potential injury or hazard may occur.
CAUTION: Potential damage to instrument/property may occur.
CAT I1: IEC Measurement Category I, applicable for making measurements on
‘other’ circuits that are not directly connected to mains. See p. v.
CAT II: IEC Measurement Category II, applicable for making measurements on
circuits connected directly to utilization points (socket outlets and similar
points) of the low voltage mains installation. See p. v.
- CAT I as defined in IEC/EN 61010-031:2008.
Note that Measurement Category I was removed in IEC/EN 61010-031:2015 and replaced by ‘O’, indicating ”other cir- cuits that are not directly connected to mains.”
Saftey Symbols
The following symbols may appear on the instrument:
Operating Environment
Temperature: 0 °C to 40 °C
Humidity: 5% to 90% relative humidity (non-condensing) up to +30°C. Upper
limit derates to 50% relative humidity (non-condensing) at +40°C.
Altitude: ≤ 2000 m
Use indoors only.
Pollution Degree 2. Use in an operating environment where normally only dry,
non-conductive pollution occurs. Temporary conductivity caused by condensation
should be expected.
AC Power
Input Voltage & Frequency:
100-120 V at 50/60 Hz
200-240 V at 50/60 Hz
Manual AC selection with a slide switch.
Power Consumption: 20 W maximum
Mains Supply Connector: CAT II per IEC/EN 61010-1:2010, instrument intended to
be supplied from the building wiring at utilization points (socket outlets and
similar).
Fuse Type
Current Input Terminal: 250 VAC F 10 A, 3 AG
AC Mains: 250 VAC F 300 mA, 5×20 mm
Input Terminal Protection Limitation
Protection limitation is defined for the input terminal.
-
Main Input (HI and LO) Terminals
HI and LO terminals are used for Voltage, Resistance, Capacitance, Continuity, Frequency, Diode and Temperature measurement. Two protection limitations are defined:
• HI-LO protection limitation: 1000 VDC or 750 VAC. This is the maximum measurable voltage. The limitation can be expressed as 1000 Vpk.
• LO ground protection limitation: LO terminal can “float” 500 Vpk relative to the ground safely. The maximum protection limitation of HI terminal relative to the ground is 1000 Vpk. Therefore, the sum of the “float” voltage and the measured voltage can’t exceed 1000 Vpk. -
Sampling (HISense and LOSense) Terminals
HISense and LOSense terminals are used for 4-wire Resistance measurement. Two protec- tion limitations are defined:
• HISense-LOSense protection limitation: 2000 Vpk.
• LOSense-LOSense protection limitation: 2 Vpk. -
Current Input (I) Terminal
The I terminal is used for current measurement. The maximum current which can go through the I terminal is limited to 10 A by the fuse on the back panel.
NOTE: Voltage on the current input terminal corresponds to voltage on the LO terminal. To keep good protection, only use a fuse of the specified type and value to replace this fuse
IEC Measurement Category II Overvoltage Protection
To avoid the danger of electric shock, the Data Acquisition Unit provides
overvoltage protection for line-voltage mains connections that meet both of
the following conditions:
- The HI and LO input terminals are connected to the mains under Measurement Category II conditions described in the warning below.
- The maximum line voltage of the mains does not exceed: 600 VAC for the instrument.
WARNING: IEC Measurement Category II includes electrical devices
connected to mains at an outlet on a branch circuit, such as most small
appliances, test equipment, and other devices that plug into a branch outlet
or socket.
The DAQ is capable of making measurements with the HI and LO inputs connected
to mains in such devices (≤ 600 VAC) or to the branch outlet itself.
However, the HI and LO terminals of the DAQ can’t be connected to mains in
permanently installed electrical devices such as the main circuit-breaker
panels, sub-panel disconnected boxes and per- manently wired motors. Such
devices and circuits are prone to exceed the protection limits of the DAQ.
Limits for Measurements on Other Circuits Not Directly Connected to Mains Max.
rated input voltage: 1000Vrms
Transient overvoltage: 4000Vpk
WARNING: Voltages above 600 VAC can only be measured in circuits that are
isolated from mains. However, there may be transient over voltage in circuits
that are isolated from mains. The DAQ is able to withstand occasional
transient overvoltage up to 4000 Vpk. Please don’t use this instrument to
measure circuits where transient overvoltage may exceed this level.
Daily Maintenance and Cleaning
Maintenance
Protect the liquid crystal display from direct sunlight when storing or using
the instrument.
NOTE: To avoid damage to the instrument or test leads, please don’t place
them in mist, liquid or solvent.
Cleaning
Regularly clean the instrument and test leads.
- Wipe the external dust off the instrument and test leads using a soft rag. Be careful not to scratch the display screen when cleaning. Do not allow any liquid to enter the instrument.
- Use a damp soft rag to clean the instrument after removing the power plug. Use 75% iso- propyl alcohol or water solvent to get a more thorough cleaning.
NOTE: To prevent the surface of the instrument or test leads from damage, please don’t use any corrosive or chemical cleaning reagents. Please make sure the instrument is fully dry before re- connecting the power to avoid short circuits or personal injury.
Introdution to T3DAQ1-16
Teledyne Test Tools T3DAQ1-16 is a 16 channel Data Acquisition System
incorporating the latest 4.3 inch (10.92 cm) dual-display technology which can
be configured to show Data Histograms, Data fluctuation Trends, Bar Graph,
Statistics or the traditional Number mode, all in an easy to use interface.
The T3DAQ features 12 multi-purpose + 4 current channels and supports various
mea- surement functions. It provides a convenient and versatile solution for
test applications that require multiple measurement points or signals and is
an ideal tool for R& D burn-in and production testing. A great feature of the
Teledyne Test Tools T3DAQ is its ability to make highly accurate True RMS AC
Voltage and Current measurements, meaning no loss of accuracy even when
measuring com- plex voltage and current waveforms.
The T3DAQ provides various measurement modes to satisfy a wide range of
application require- ments. 12 Multi-purpose channels can be configured
individually for various scan requirements. The Channel configuration function
allows user to configure different parameters such as mea- surement type,
range, scan mode, number of scan cycles, and duration of scan. Similarly, the
scan control menu provides flexibility to schedule, start, and stop the scan.
Scanning data can be viewed live during a scan and can also be saved for
future use. An internal 1 GB flash memory is available for data logging and to
save configuration files. The T3DAQ also features a USB Host interface to use
a USB flash drive to collect data without being connected to a PC. The T3DAQ
is equipped with LAN and USBTMC interfaces for remote control of the device.
The device can be controlled remotely using EasyDMM software. The graphical
interface of the soft- ware makes it easy to configure measurements, setup and
execute scans or troubleshoot the design. Live measurement data can be viewed
in numeric digits or graphically by trend chart, bar graph and histogram. The
data can be manually or automatically exported as CSV files for further
analysis.
Main features:
- 4.3” TFT-LCD Display
- Dual display, Chinese and English Menu
- Built-in front panel accessible help system
- File management (support for U-disc and local storage)
- Real 6-5 digit (2,200,000 count) readings resolution
- True-RMS AC Voltage and AC Current measuring
- 1 GB flash memory for mass storage configuration files and data files
- Built-in cold terminal compensation for thermocouple
- Standard interface: USB Device, USB Host, LAN
- USB & LAN remote interfaces support common SCPI command set
© 2021 Teledyne LeCroy, Inc. All rights reserved.
Teledyne Test Tools is a brand and trademark of Teledyne LeCroy, Inc. Other
product or brand names are trademarks or requested trademarks of their
respective holders. Specifications, prices, availability and delivery subject
to change without notice.
Chapter 1
Quick Start
General Inspection
Please check the instrument according to the following steps.
-
Inspect the shipping container.
Keep the shipping container and packaging material until the contents of the shipment have been completely checked and the instrument has passed both electrical and mechanical tests. It is always good practice to save the shipping container and packaging for use when returning the power supply to Teledyne LeCroy for service or calibration.
The consigner or carrier will be responsible for damage to the instrument resulting from ship- ping. Teledyne LeCroy will not provide free maintenance or replacement in this instance. -
Inspect the instrument.
If the instrument is found to be damaged, defective or fails in electrical or mechanical tests, please contact the Teledyne LeCroy service department immediately. -
Check the accessories.
Please check that you have received the accessories on the packing list:
1 DAQ
1 Test Lead Set
1 USB Cable
If the accessories are incomplete or damaged, please contact Teledyne LeCroy immediately.
Dimensions
Handle Adjustment
To adjust the handle position of the DAQ, grip the handle by the two sides and
pull outward. Then, rotate the handle to the appropriate position.
Front Panel
A USB Host
Users can store the current state or measurement data into a USB storage
device. Users can also read the state files or updated firmware from a USB
storage device.
B Power Key
Turn the instrument on or off.
C LCD Display
The instrument provides a 4.3 inch high resolution color TFT-LCD display
screen with 480*272 pixels that displays the function menus, measurement
parameter settings, system status, and prompt messages.
D Menu Operation Keys
Press any softkey to activate the corresponding menu.
E Measurement and Assistant Function Keys
| DC Voltage/Current Measurement
---|---
| AC Voltage/Current Measurement
| 2-Wire/4-Wire Resistance Measurement
| Frequency/Capacitance Measurement
| Continuity/Diode Test
| Temperature Measurement/Enable Multiple Scan Card Function
| Enable Dual-display Function/Set Up the Utility
| Acquire Function/Help System
| Math Function/Display Function
| Auto Trigger/Stop
| Single Trigger/Hold Measurement Function
,| Return to local control of the instrument (when in Remote mode). Some of
the front panel keys have text above them. This indicates that the key has a
function that you can access by pressing and releasing [Shift] before pressing
the key.
F Range and Direction Keys
| Increase the measurement range
---|---
| Decrease the measurement range
| Select auto or manual range
| Set up measurement parameter Move the cursor Page up or down
| Set up measurement parameter Move the cursor
| Apply the current setting
G Signal Input Terminals
The measured signal (device) will be connected into the DAQ through these
terminals. Dif- ferent measurement objects have different connection methods.
For details, please refer to “Measurement Connections”.
Rear Panel
A Power Socket
The DAQ accepts two types of AC supplies. Please use the power cord provided
in the accessories to connect the DAQ to the AC power through this socket.
Note: The correct voltage scale must be first selected (through the Voltage
Se- lector) before power connection.
B Power Fuse
The DAQ is already installed with a power fuse before leaving factory. To
change the fuse please:
- Turn off the DAQ and remove the power cord.
- Press down the block tongue using a straight screwdriver (in the direction of the dotted arrow in the figure below) and pull out of the fuse seat.
- Select a proper voltage scale.
- Replace a specified fuse.
- Reinstall the fuse seat into the slot.
C AC Voltage Selector
Select the correct voltage scale (110 V or 220 V) for the AC supply used.
D 16-channel Data Acquisition Module
E USB Device
Connect the PC through this interface. SCPI commands or PC software can be
used to control the DAQ remotely.
F LAN
Through this interface, the DAQ can be connected to the network for remote
control.
G VMC Output
The DAQ outputs a low-true pulse from the [VM Comp] connector after every
measurement.
H Ext Trigger
Trigger the DAQ by connecting a trigger pulse through the [Ext Trig]
connector. Note the external trigger source must be selected.
I Current Input Fuse
The DAQ is already installed with a current Input fuse to provide 10 A maximum
input pro- tection before leaving factory. To replace a new one, please:
(a) Turn off the DAQ and remove the power cord.
(b) Turn the fuse seat counterclockwise as shown in the figure using a
straight screw driver and then pull out the fuse seat
(c) Place a new 10 A specified fuse.
(d) Reinstall the fuse seat into the slot.
J Instrument Kensington Lock Point
A Kensington lock (not supplied) can be used to lock the DAQ to a fixed place
if necessary.
Starting the DAQ
Before connecting the instrument to a power source please adjust the AC
voltage selector on the rear panel of your DAQ according to your local power
supply voltage.
Then connect the power cord as shown in the following figure.
Press the Power key on the front panel to turn on the DAQ. If the DAQ does not start normally, then try the following:
- Make sure the power cord is in good connection and connected to the DAQ and the wall socket.
- Ensure that the wall socket has power and is turned on.
- Try to restart the DAQ, if it fails, check the power fuse and replace with a new one if necessary.
- If the problem still remains, please contact the Teledyne LeCroy service department for help.
Operating Instructions
NOTE: The Scanner Card is not designed to be ”Hot Swappable”.
Please make sure the instrument is turned off before installation or removal
of the Scanner Card.
Hot swapping the card could cause damage to the instrument and is not covered
under warranty.
- Power Off Turn the Power Off and take out The power cord
- Open the DAQ rear panel slot Take off the two screws on the slot corners to remove the optional slot cover. Keep the screws for later reuse
- Connectlon Turn the clamp and insert the wire.
- Insert the Scanner Card Insert the Scanner Card bottom side up. Close the cover by tightening the SUMS.
- Tighten cable Route wiring through strain relief and Cable tie rap Wrap
- Power On Connect the power cord and turn On the power
Scanner Card Connections
An example of 2-wired connection application is shown below.First 12 channels can be used to measure DCV/DCI
ACV/ACI/CAP/FREQ/DIODE/CONT/TEMP
and the last 4 channels are dedicated to current measurement.
4-Wire Resistance measurement example is shown below. CH1 to CH6 can be used for measurements and CH7 to
CH12 can be used as sense channels. To minimize voltage errors, the remote
sense connections (CH7,CH8 etc..) should be made as close to the device-under-
test (DUT) as possible.
Front Panel Operations
Scanner mode functions can be accessed by pressing ”Shift” and ”Temp”
buttons.
The following table gives a short description about different functions available.
Function Menu | Settings | Description |
---|---|---|
Mode | Scan/Step | Set the operation mode |
Time | 0ms-999.999s | Sets the duration between each scan loop (Scan mode) or |
between each scanned channel (Step mode)
Cycles| Auto/Man| Sets the number of scan op- erations
Channel Setup| | Sets the scanned channel range, measurement func-
tion, and measurement pa- rameters
Start| On/Off| Start or stop scan operation
Exit| | Exit the scanner function
-
Operation mode setup
•Scan: Measures all specified channel ranges (Channel MIN-MAX) for each trigger event. Time settings are applied between each scan for the whole channel range.
•Step: Measures a single channel in the specified range (Channel MIN-MAX) at each trigger event. Time settings are applied for each channel. -
Time setup
Use the direction keys to set the duration between each scan loop (Scan mode) or between each scanned channel (Step mode) -
Cycles
•Auto: The instrument will scan specified channel circularly after the scan operation start and you should stop the operation manually.
.•Manual: Sets the number of scan operations by direction keys. The range of the setting is from 1 to 999. After starting the scan operation the instrument will not stop scanning until reaches cycle number. -
Channel setup
Press the [Channel Setup] to enter the setup interface.
The table below gives a brief description of Function menu.
Function Menu | Description |
---|---|
Channel Con- fig | Open/close the channel and set the mea- surement function, |
measurement parame- ters of specified channel.
Low| Set the low value of scanned channel range.
High| Set the high value of scanned channel range.
Press the [Channel Setup] to enter the channel configuration interface and set the channel switch, function, range and speed.
The range setting is applicable for the following functions: DC/AC Voltage
(DCV/ACV), 2/4 Wire Resistance (2W/4W), Capacitance (CAP), Frequency (FRQ).
Ranges available for various functions is given below.
Function Menu | Description |
---|---|
DCV/ACV/ FRQ | Auto, 200mV, 2V, 20V, 200V |
DCI/ACI | 2A (fixed) |
2W/4W | Auto, 200Ω, 2kΩ, 20kΩ, 200kΩ, 2MΩ, 10MΩ, 100MΩ |
The scanner function provides two measurement speeds: Fast (50 reading/s) and
Slow (5 reading/s). (Fast: 1PLC, Slow: 10 PLC)
The speed setting is applicable for the following functions: DC/AC Voltage
(DCV/ACV), 2/4 Wire Resistance (2W/4W)
Operating instructions:
- Move the cursor to choose the wanted parameter by direction keys and the background color of cursor’s position turns to gray.
- Select the current item by pressing “OK” key and the background color of the selected item turns to green.
- Set the parameter by up and down direction keys.
- Press “OK” key again to store the setting of the selected item of which the background turns back to gray. Move the cursor and repeat the prior steps to set the next parameter.
- Press [Done] to save the current settings and return to the higher level menu.
5. Channel range setting
Select [High] or [Low] and then input numerical value by direction keys. Note: The upper limit value should be always bigger than the lower limit value.
6. Start scan operation
Set the [Start] to on to start the scan operation
The upper part of the
interface displays the function, range, and result of the current channel. The
table below records the measurement result of each channel.
7. Enter trend chart and statistics mode (optional operation)
Press ”Shift” and ”Math” to open trend chart and statistics
function.Set the [Inquire Channel] by
direction keys and the interface displays the minimum, average, max- imum,
span, standard deviation, samples and trend chart of measurement results of
the setting channel during the scan operation. Press [Done] to return to
higher level menu.
8. Stop scan operation
Set the [Start] to off to stop the scan operation. If the [Cycles] is set to
manual, then the instrument will stop scan when it reaches the cycle number
9. Store measurement data (optional operation) Press ”Shift” and ”Dual”, then Select [Store/Recall] →
[Store Settings] to enter the storage function interface.
User Interface
Measurement Connections
The DAQ is designed with many measurement functions. After selecting the
desired measure- ment function, please connect the signal (device) under test
to the DAQ according to the method below. Do not switch the measurement
function when measuring as it may cause damage to the DAQ. For example, when
the test leads are connected to the related current terminals, AC voltage
measurement should not be used.
Resistance Measurement (2-wire)
- Resistance Measurement (4-wire)
Connect the test leads and tested circuit as in the diagram below. The HI terminal and HI Sense should be connected to one end of the DUT. The LO terminal and LO Sense should be connected to the other end of the DUT.
Temperature Measurement (For RTD and thermcouple sensor)
Diode Measurement
Connecting to USB and LAN Ports
The DAQ has LAN and USB I/O ports. Connect to the ports as in the diagrams
below:
Using the Built-in Help System
To access the built-in help system, press [Shift] + [Acquire], then use the
direction keys to choose the help item you want. Finally, press [ OK ] to
obtain help. The help listings are as follows:
- Basic Measurements
- Measuring Temperature
- Measuring Capacitance
- Math Function
- Dual-display Function
- Saving and Recalling Information
- Optional Multiple Scan Card
- The convention and Tips for Softkey
Chapter 2 Function and Operation
Measurement Configuration
Most measurement parameters are user-defined. Changing a measurement parameter
will change the measurement precision and speed, as well as the input
impedance. An appropriate measure- ment parameter based on the actual
application will ensure faster measurement or higher mea- surement precision.
The default measurement configurations of the DAQ can ensure the accuracy of
the measurement results in most cases. Users can directly use these defaults
for any measurement or modify the parameters of the measurement function as
required.
The parameters for different measurement functions differ, see the table below
Functions | Parameters |
---|---|
DCV | Range, Integration Time, DC impedance, Auto zero |
DCV | Range, AC filter |
ACV | Range, Integration Time, Auto zero |
DCI | Range, Integration Time, DC impedance, Auto zero |
ACI | Range, AC filter |
OHM (2WR, 4WR) | Range, Integration Time, Auto zero |
CAP | Range |
CONT | Short-circuit resistance |
DIODE | Breakover voltage |
FREQ/PERIOD | Gate time |
TEMP | N/A |
Range
T3DAQ provides auto and manual range selecting modes. In auto mode, the DAQ
selects a proper range automatically according to the input signal. In manual
mode, you can use the front panel key or menu key to set the range. The auto
mode can bring a lot of convenience for users while the manual mode provides
higher reading precision.
Method 1: use the front panel key to set the range.
Method 2: Use the menu key to select the range.
Enter the specific measurement function and select [Range] in the menu to show
the range setting options, as shown in the figure below, then press the menu
operation key to activate the corre- sponding configuration.
Note:
- “overload” will be displayed when the input signal exceeds the currently set range.
- By default, the range is set to Auto at power-on or after a reset.
- Auto mode is recommended if you are not sure about the measurement range in order to protect the instrument and obtain accurate data.
- The range of CONT measurement is fixed at 2kΩ.
Integration Time and Resolution
Integration time is the period during which the DAQ’s analog-to-digital (A/D)
converter samples the input signal for a measurement. The longer the
integration time, the slower the measurement speed will be and the higher the
resolution will be. The shorter the integration time, the faster the
measurement will be and the lower the resolution will be. The integration time
applies to DCV, DCI, 2WR and 4WR measurements.
T3DAQ1-16 expresses the integration time by the number of power line cycles,
the unit is PLC. The DAQ automatically detects the input power line frequency
at power-on. If the frequency is 50Hz, the integration time can be set to
0.005PLC, 0.05PLC, 0.5PLC, 1PLC, 10PLC, 100PLC and the default is 10PLC. If
the frequency is 60Hz, the Integration time can be set to 0.006PLC, 0.06PLC,
0.6PLC, 1PLC, 10PLC, 100PLC and the default is 10PLC.
T3DAQ1-16 can make measurements with reading resolutions of 4½, 5½ and 6½
digits. It auto- matically selects a reading resolution according to the
measurement settings.
- In DCV, DCI and OHM measurements, press [Aperture] to set the integration time, as shown in the diagram below (take DCV measurement for instance). The integration time affects the resolu- tion. The table below shows the relationship between reading resolution and integration time.
Resolution| Integration time
---|---
4½| 0.005PLC/0.006PLC
0.05PLC/0.06PLC
5½| 0.5PLC/0.6PLC
6½| 1PLC
10PLC
100PLC - In ACV, ACI ,FREQ/PERIOD measurements, the resolution is fixed at 6½ digits.
- In CAP measurements, the resolution is fixed at 4½.]
- The instrument always displays 2 digits after the decimal point in CONT measurement.
- In DIODE measurements, the resolution is fixed at 5½.
- In TEMP measurements, the resolution is fixed at 5½.
DC Impedance
DC impedance applies to DCV measurements. The default is “10MΩ”. In the range
of 200mV, 2V or 20V, you can choose “>10GΩ” to reduce the loading error on the
measured object, caused by the multimeter load.
In the range of 200 mV, 2 V or 20 V under DCV measurement, press [Input Z] in
the menu to perform the setting, as shown in the diagram below.
- 10MΩ: set the input impedance in all ranges to 10MΩ.
- 10GΩ: set the input impedances in ranges of 200mV, 2V and 20V to 10GΩ, while in ranges 200V and 1000V, the impedance is still 10MΩ.
Auto Zero
Auto zero (Auto Zero) applies to DCV, DCI, 2WR and 4WR measurements. Enter the
specific measurement function and press [Auto Zero] in the menu to perform the
setting, as shown in diagram below (take DCV measurement for instance).
- ON: the DAQ internally disconnects the input signal and measured circuit after each mea-surement and takes a zero reading. It then subtracts the zero reading from the preceding reading (displaying the difference between the measurement value and zero value during the measurement), in order to reduce the impact of offset voltage from input circuit on measure- ment result.
- OFF: disable the auto zero function.
AC Filter
AC filter applies to ACV and ACI measurements. It can optimize the low
frequency accuracy and minimize the AC settling time. T3DAQ1-16 provides three
types of AC filters (>3Hz, >20Hz, >200Hz).
The AC filter to be used is determined by the input signal frequency. You
should generally select the highest frequency filter whose frequency is less
than that of the signal you are measuring, be- cause the higher frequency
filters result in faster measurements. For example, when measuring a signal
between 20Hz and 200Hz, use the 20Hz filter. If measurement speed is not an
issue, choosing a lower frequency filter may result in quieter measurements,
depending on the signal that you are measuring.
Press [Filter] in the menu of ACV or ACI measurement to show the setting
options, as shown in diagram below (take ACV measurement for instance).
Then press the menu operation key to activate the corresponding configuration.
Short-circuit Resistance
This function only applies to a continuity test. When the measured circuit has
a resistance lower than the short-circuit resistance, the circuit is
considered as connected and the beeper sounds (if sound is on). The default
short-circuit resistance is 50Ω and the setting is stored in non-volatile
memory.
When continuity test is enabled, set the [Threshold] (equal to short-circuit
resistance) using the direction keys. The range is from 1Ω to 2000Ω .
Gate Time
Gate time (also called Aperture Time) applies to the FREQ / PERIOD function.
It decides the res- olution of a low-frequency measurement. The longer the
gate time, the higher the resolution of the low-frequency measurement and the
slower the measurement, and vice versa.
In FREQ / PERIOD measurement, press [Gate Time] to show the setting options,
as shown in diagram below (take FREQ measurement for instance).
The gate time can be set to 1ms, 10ms, 100ms or 1s and the default is 100 ms.
You can select a desired gate time by pressing the corresponding softkey.
Basic Measurement Functions
To Measure DC Voltage
Range: 200mV, 2V, 20V, 200V, 1000V
Max Resolution: 100nV (in the range of 200mV)
Input Protection: 1000 V protection is available on all ranges and a 10% over-
range for all ranges except 1000 V range. If the reading exceeds the range,
“overload” will be displayed.
Operating Steps:
-
Enable the DCV measurement
Press [DCV] on the front panel to enter the DC Voltage measurement interface, as shown in Diagram below. -
Make the connection
Connect the test leads with the measured signal by referring to “Measurement Connec- tions”. -
Set the range
Press [Range] to select a range for the measurement. You can also use the [+], [-], and [Range] keys on the front panel to select the range. Auto (autorange) automatically selects the range for the measurement based on the input. Autoranging is convenient, but it results in slower measurements than using a manual range. Autoranging goes up a range of 110% of the present range, and changes down a range when the measurement is below 10% of the present range. -
Set the Integration
Press [Aperture] and choose the number of power-line cycles (PLCs) to use for the measure- ment. Selecting 100PLC provides the best noise rejection and resolution, but the slowest measurements. -
Autozero setting
Press [Auto Zero] to enable or disable this function. Autozero provides the most accurate measurements, but requires additional time to perform the zero measurement. With au- tozero enabled (On), the DAQ internally measures the offset following each measurement. It then subtracts that measurement from the preceding reading. This prevents offset voltages present on the DAQ’s input circuitry from affecting the measurement accuracy. -
Specify the DC input impedance (Only for Manual 200mV, 2V and 20V ranges) Press [Input Z] to set the DC resistance as “10M” (default value) or 10G”. Users can execute the DC voltage measurement directly without modifying this parameter if they wish.
-
Set the relative value (Optional operation)
Press [Rel] to start or stop the Relative math function. When it is started, the reading dis- played is a value which comes from the result of actual measurement value minus the relative value that has been set. The default relative value is the measurement value when the func- tion is started. (For details, please refer to “Math Functions” in Chapter 2). -
Read the measurement value
The DAQ measures the input signal according to the current measurement settings and dis- plays the measurement result on the screen. -
Perform math operations (advanced)
You can perform math operations (Statistics, Limit, dBm, dB and REL) on every DCV mea- surement reading. For details, please refer to “Math Operations”. -
Display the graph (advanced)
You can analyze the measurement data by using the “Bar Meter”, “Trend Chart” or “Histogram display”. For details, please refer to “Display Mode”.
To Measure DC Current
Range: 200µA, 2mA, 20mA, 200mA, 2A, 10A
Max Resolution: 0.1nA (in the range of 200µA)
Input Protection: a 10A protection fuse is available in all ranges and a 10%
over-range for all ranges except 10A range. If the reading exceeds the range,
“overload” will be displayed.
Operating Steps:
-
Enable the DCI measurement
Press [Shift] and [DCV] on the front panel to enter the DC Current measurement interface, as shown in the diagram below. -
Make the connection
Connect the test leads with the measured signal by referring to “Measurement Connec- tions”. -
Set the range
Press [Range] to select a range for the measurement. You can also use the [+], [-], and [Range] keys on the front panel to select the range. Auto (autorange) automatically selects the range for the measurement based on the input. Autoranging is convenient, but it results in slower measurements than using a manual range. Autoranging goes up to 110% of a range and changes up a range at 110% of the present range, and changes down a range when below 10% of the present range. -
Set the Integration Time
Press [Aperture] and choose the number of power-line cycles (PLCs) to use for the measure- ment. Selecting 100PLC provides the best noise rejection and resolution, but the slowest measurements. -
Autozero setting
Press [Auto Zero] to enable or disable this function. Autozero provides the most accurate measurements but requires additional time to perform the zero measurement. With au- tozero enabled (On), the DAQ internally measures the offset following each measurement. It then subtracts that measurement from the preceding reading. This prevents offset voltages present on the DAQ’s input circuitry from affecting the measurement accuracy. -
Set the relative value (Optional operation)
Press [Rel] to start or stop the Relative math function. When it is started, the reading dis- played is a value which comes from the result of actual measurement value minus the relative value that has been set. The default relative value is the measurement value when the func- tion is started. (For details, please refer to “Math Functions” in Chapter 2.) -
Read the measurement value
The DAQ measures the input signal according to the current measurement settings and dis- plays the measurement result on the screen. -
Perform math operations (advanced)
You can perform math operations (Statistics, Limit and REL) on every DCI measurement reading. For details, please refer to ”Math Functions”. -
Display the graph (advanced)
You can analyze the measurement data by using the “Bar Meter”, “Trend Chart” or “Histogram display”. For details, please refer to “Display Mode”.
To Measure AC Voltage
Range: 200mV, 2V, 20V, 200V, 750V
Max Resolution: 100nV (in the range of 200mV)
Input Protection: 750V protection is available in all ranges and a 10%
overrange for all ranges except 750V range. If the reading exceeds the range,
“overload” will be displayed.
Operating Steps:
-
Enable the ACV measurement
Press [ACV] on the front panel to enter the AC Voltage measurement interface, as shown in the diagram below. -
Make the connection
Connect the test leads with the measured signal by referring to “Measurement Connec- tions”. -
Set the range
Press [Range] to select a range for the measurement. You can also use the [+], [-], and [Range] keys on the front panel to select the range. Auto (autorange) automatically selects the range for the measurement based on the input. Autoranging is convenient, but it results in slower measurements than using a manual range. Autoranging goes up a range at 110% of the present range, and down a range when below 10% of the present range. -
Set the filter
Press [Filter] and choose the filter for the measurement. The instrument provides three dif- ferent AC filters, “>3Hz”, “>20Hz” and “>200Hz”. You should generally select the highest frequency filter whose frequency is less than that of the signal you are measuring. -
Set the relative value (Optional operation)
Press [Rel] to start or stop the Relative math function. When it is started, the reading dis- played is a value which comes from the result of actual measurement value minus the relative value that has been set. The default relative value is the measurement value when the func- tion is started. (For details, please refer to “Math Functions” in Chapter 2.) -
Read the measurement value
The DAQ measures the input signal according to the current measurement settings and dis- plays the measurement result on the screen. -
Perform the math operation (advanced)
You can perform math operations (Statistics, Limit, dBm, dB and REL) on every ACV mea- surement reading. For details, please refer to “Math Operations”. -
Display the graph (advanced)
You can analyze the measurement data by using the “Bar Meter”, “Trend Chart” or “Histogram display”. For details, please refer to “Display Mode”.
To Measure AC Current
Range: 200µA, 2 mA, 20 mA, 200 mA, 2A, 10A
Max Resolution: 0.1nA (in the range of 200µA)
Input Protection: 10A protection is available in all ranges and a 10% over-
range for all ranges except the 10A range. If the reading exceeds the range,
“overload” will be dis-
played.
Operating Steps:
-
Enable the ACI measurement
Press [Shift] and [ACV] on the front panel to enter the AC current measurement interface, as shown in the diagram below. -
. Make the connection
Connect the test leads with the measured signal by referring to “Measurement Connec- tions”. -
Set the range
Press [Range] to select a range for the measurement. You can also use the [+], [-], and [Range] keys on the front panel to select the range. Auto (autorange) automatically selects the range for the measurement based on the input. Autoranging is convenient, but it results in slower measurements than using a manual range. Autoranging goes up a range at 110% of the present range, and down a range when below 10% of the present range. -
Set the filter
Press [Filter] and choose the filter for the measurement. The instrument provides three dif- ferent AC filters, “>3Hz”, “>20Hz” and “>200Hz”. You should generally select the highest frequency filter whose frequency is less than that of the signal you are measuring. -
Set the relative value (Optional operation)
Press [Rel] to start or stop the Relative math function. When it is started, the reading dis- played is a value which comes from the result of actual measurement value minus the relative value that has been set. The default relative value is the measurement value when the func- tion is started. (For details, please refer to “Math Functions” in Chapter 2.) -
Read the measurement value
The DAQ measures the input signal according to the current measurement settings and dis- plays the measurement result on the screen. -
Perform the math operation (advanced)
You can perform the math operations (Statistics, Limit and REL) on every ACI measurement reading. For details, please refer to “Math Functions” in Chapter -
Display the graph (advanced)
You can analyze the measurement data by using the “Bar Meter”, “Trend Chart” or “Histogram display”. For details, please refer to “Display Mode”.
To Measure Resistance
Range: 200Ω, 2kΩ, 20kΩ, 200kΩ, 1MΩ, 10MΩ, 100MΩ
Max Resolution: 100µΩ (in the 200Ω range)
Input Protection: 1000 V protection is available in all ranges and a 10% over-
range for all ranges. If the reading exceeds the range, “overload” will be
displayed.
T3DAQ1-16 provides 2-wire and 4-wire resistance measurements. When the
measured resistance is lower than 100kΩ, the 4-wire resistance measurement is
recommended to reduce the measure- ment error caused by the test lead
resistance and contact resistance between the probe and the testing point.
These two resistances should not be ignored when compared to the measured re-
sistance.
Operating Steps:
-
Enable 2-wire/4-wire resistance measurement
Press [Ω2W] on the front panel to enter the 2-wire resistance measurement interface, as shown in the below diagram. Press [Shift] and [Ω2W] on the front panel to enter the 4-wire resistance measurement inter- face, as shown in the diagram below. -
Make the connection
Connect the test leads with the measured signal by referring to “Measurement Connec- tions”. -
Set the range
Press [Range] to select a range for the measurement. You can also use the [+], [-], [Range] and keys on the front panel to select the range. Auto (autorange) automatically selects the range for the measurement based on the input. Autoranging is convenient, but it results in slower measurements than using a manual range. Autoranging goes up a range at 110% of the present range, and down a range when below 10% of the present range. -
Set the Integration
Press [Aperture] and choose the number of power-line cycles (PLCs) to use for the measure- ment. Selecting 100PLC provides the best noise rejection and resolution, but the slowest measurements. -
Autozero setting
Press [Auto Zero] to enable or disable this function. Autozero provides the most accurate measurements, but requires additional time to perform the zero measurement. With au- tozero enabled (On), the DAQ internally measures the offset following each measurement. It then subtracts that measurement from the preceding reading. This prevents offset voltages present on the DAQ’s input circuitry from affecting the measurement accuracy. -
Set the relative value (Optional operation)
Press [Rel] to start or stop the Relative math function. When it is started, the reading dis- played is a value which comes from the result of actual measurement value minus the relative value that has been set. The default relative value is the measurement value when the func- tion is started. (For details, please refer to “Math Functions” in Chapter 2.) -
Read the measurement value
The DAQ measures the input signal according to the current measurement settings and dis- plays the measurement result on the screen. -
Perform the math operation (advanced)
You can perform the math operations (Statistics, Limit and REL) on every resistance mea- surement reading. For details, please refer to “Math Operations” in Chapter 2. -
Display the graph (advanced)
You can analyze the measurement data by using the “Bar Meter”, “Trend Chart” or “Histogram display”. For details, please refer to “Display Mode”.
To Measure Capacitance
Range: 2nF, 20nF, 200nF, 2µF, 20µF, 200µF, 2mF, 20mF, 100mF
Max Resolution: 1pF (in the range of 2nF)
Input Protection: 1000V protection is available in all ranges. If the reading
exceeds the range, “overload” will be displayed.
Operating Steps:
-
Enable the Capacitance measurement Press [ ] on the front panel to enter the Capacitance measurement interface, as shown in the diagram below.
-
Make the connection
Connect the test leads with the measured signal by referring to “Measurement Connec- tions”. -
Set the range
Press [Range] to select a range for the measurement. You can also use the [+], [-], [Range] and keys on the front panel to select the range. Auto (autorange) automatically selects the range for the measurement based on the input. Autoranging is convenient, but it results in slower measurements than using a manual range. Autoranging goes up a range at 110% of the present range, and down a range when below 10% of the present range. -
Set the relative value (Optional operation)
Press [Rel] to start or stop the Relative math function. When it is started, the reading dis- played is a value which comes from the result of actual measurement value minus the relative value that has been set. The default relative value is the measurement value when the func- tion is started. (For details, please refer to “Math Functions” in Chapter 2.) -
Read the measurement value
The DAQ measures the input signal according to the current measurement settings and dis- plays the measurement result on the screen. -
Perform the math operation (advanced)
You can perform the math operations (Statistics, Limit and REL) on every capacitance mea- surement reading. For details, please refer to “Math Functions”. -
Display the graph (advanced)
You can analyze the measurement data by using the “Bar Meter”, “Trend Chart” or “Histogram display”. For details, please refer to “Display Mode”.
Remember to disconnect power to the DUT and short the legs or connections of the elec- trolytic capacitors before measuring the electrolytic capacitors.
To Measure Frequency or Period
Frequency (Period) Range: From 3Hz to 1MHz (from 0.33s to 1µs).
Input Signal Range: 200mV, 2V, 20V, 200V, 750V.
Input Protection: 750V protection is available in all ranges.
Operating Steps:
-
Enable the Freq/Period measurement
Press [Shift] and [ ] on the front panel, then select [Freq] to enter the frequency measure- ment interface, as shown in the diagram below. Select [Period] to enter the period measurement interface, as shown in the diagram below. -
2. Make the connection
Connect the test leads with the measured signal by referring to “Measurement Connec- tions”. -
Set the range
Press [Range] to select a range for the measurement. You can also use the [+], [-], [Range] and keys on the front panel to select the range. Auto (autorange) automatically selects the range for the measurement based on the input. Autoranging is convenient, but it results in slower measurements than using a manual range. Autoranging goes up a range at 110% of the present range, and down a range when below 10% of the present range. -
Set the gate time
Press [Gate Time] and choose the measurement aperture of 1ms, 10ms, 100ms (default), or 1s. -
Set the relative value (Optional operation)
Press [Rel] to start or stop the Relative math function. When it is started, the reading dis- played is a value which comes from the result of actual measurement value minus the relative value that has been set. The default relative value is the measurement value when the func- tion is started. (For details, please refer to “Math Functions” in Chapter 2.) -
Read the measurement value
The DAQ measures the input signal according to the current measurement settings and dis- plays the measurement result on the screen. -
Perform the math operation (advanced)
You can perform math operations (Statistics, Limit and REL) on every measurement reading .
For details, please refer to “Math Functions”. -
Display the graph (advanced)
You can analyze the measurement data by using the “Bar Meter”, “Trend Chart” or “Histogram display”. For details, please refer to “Display Mode”.
To Test Continuity
Test Current Source: 1mA
Max Resolution: 0.01Ω
Input Protection: 1000V Input Protection
Open-circuit Voltage: <8V
Beep Threshold (short-circuit resistance): from 0Ω to 2000Ω
This function measures the resistance of the circuit with about 1mA current
source. When the measured resistance is lower than the short-circuit
resistance (Threshold), the beeper sounds (if the Beeper is on). Otherwise,
“open” is displayed on the screen.
Operating Steps:
-
Enable the Cont measurement
Press [Cont] on the front panel to enter the Continuity test interface, as shown in the diagram below. -
Make the connection
Connect the test leads with the measured signal by referring to “Measurement Connec- tions”. -
Set the short-circuit resistance (Threshold)
Enter a desired value using the direction keys. The range is from 0Ω to 2000Ω and the default is 50Ω. -
Read the measurement value
The DAQ measures the input signal according to the current measurement settings and dis- plays the measurement result on the screen. -
Perform the math operation (advanced)
You can perform math the operation (Statistics, Limit) on every measurement reading. For details, please refer to “Math Functions”. -
Display the graph (advanced)
You can analyze the measurement data by using the “Bar Meter”, “Trend Chart” or “Histogram display”. For details, please refer to “Display Mode”.
Before testing continuity remember to disconnect the DUT power and discharge all of the DUT capacitors, to avoid damage to the DAQ.
To Test Diode
Test Current Source: 1mA
Voltage Measurement Range: 0V 4V
Max Resolution: 10µV
Input Protection: 1000V Input Protection
Open-circuit Voltage: <8V
This function measures the forward voltage drop on the diode. When the voltage
is lower than the Threshold, the beeper sounds (if the beeper is on).
Operating Steps:
-
Enable the Diode measurement
Press [Shift] and [Cont] on the front panel to enter the Diode test interface, as shown in the diagram below. -
Make the connection
Connect the test leads with the measured signal by referring to “Measurement Connec- tions”. -
Set the Threshold
Enter a desired value using the direction keys. The range is from 0V to 4V and the default is 2V. -
Read the measurement value
The DAQ measures the input signal according to the current measurement settings and dis- plays the measurement result on the screen. If the reading exceeds the threshold, “open” will be displayed. -
Evaluate the results of a measurement
Reverse the probes and measure the forward voltage drop on the diode again. Evaluate the diode according to the following rules:
• If the DAQ displays “open” when in the reverse bias model, it indicates that the diode is normal.
• If the DAQ shows voltage about 0V and the instrument beeps persistently when in for- ward and reverse bias model, it indicates that the diode is short
• If the DAQ shows “open” when in the forward and reverse model, it indicates that the diode is open. -
Perform the math operation (advanced)
You can perform the math operation (Statistics, Limit) on every measurement reading. For details, please refer to “Math Functions”. -
Display the graph (advanced)
You can analyze the measurement data by using the “Bar Meter”, “Trend Chart” or “Histogram display”. For details, please refer to “Display Mode”.
Before testing continuity remember to disconnect the DUT power and discharge all of the DUT capacitors, to avoid damage to the DAQ.
To Measure Temperature
T3DAQ1-16 can directly measure the temperature using TC (Thermocouple) and
THERM (Ther- mistor) sensors.
Operating Steps:
-
Enable the Diode measurement
Press on the front panel to enter the Temperature measurement interface, as shown in the diagram below. -
Make the connection
Connect the test leads with the measured signal by referring to “Measurement Connections”. -
Set the type of sensor
Press [Load] and use the direction keys to choose a desired temperature sensor. Press [De- fine] to view the configurations. Then press [Read] to apply the current temperature sensor configurations. -
Set the display mode
Press [Display] to choose the display mode. The DAQ supports three display modes: Tem- perature Value, Measured Value and All (Temperature Value and Measured Value will be shown on the display together). -
Set the unit of temperature
Press [Units] to choose the unit of temperature. The DAQ supports three units: °C ,°F , °K. -
Set the relative value (Optional operation)
Press [Rel] to start or stop the Relative math function. When it is started, the reading dis- played is a value which comes from the result of actual measurement value minus the relative value that has been set. The default relative value is the measurement value when the func- tion is started. (For details, please refer to “Math Functions” in Chapter 2.) -
Read the measurement value
The DAQ measures the input signal according to the current measurement settings and dis- plays the measurement result on the screen. -
Perform the math operation (advanced)
You can perform the math operation (Statistics, Limit and REL) on every measurement read- ing. For details, please refer to “Math Functions”. -
Display the graph (advanced)
You can analyze the measurement data by using the “Bar Meter”, “Trend Chart” or “Histogram display”. For details, please refer to “Display Mode”.
Dual-display Function
Dual-display function is used to improve the test and measurement functions.
Press [Dual] to open the Dual-display function and the upper right corner will
show “Dual”. Now press a function key if this function can be used as the
second display, it will be displayed in the second Display area. The Main
Display will display the function that is selected before the Dual-display
function is turned on. All the available combinations are listed in the table
below (shade is available).
| Main Display Function
---|---
DCV| DCI| ACV| ACI| FREQ| PERIOD| 2- W ire
R| 4- W ire
R| Cap
Vice
Display
Function| DCV| | | | | | | | |
DCI| | | | | | | | |
ACV| | | | | | | | |
ACI| | | | | | | | |
FREQ| | | | | | | | |
PERIOD| | | | | | | | |
2-Wire R| | | | | | | | |
4-Wire R| | | | | | | | |
Cap| | | | | | | | |
For example, press [DCV]→ [Dual]→ [ACV] to enter the following interface.
Instruction:
-
If the same measurement function is used in both Main and second Display.
• The readings in both displays will update at the same time.
• If math function (dBm, dB) is used in Main Display, when opening the second Display, the math operation will be stopped automatically. The second Display will show the same measurement result as Main Display.
• If math function (Statistics, Limits, Relative) is used in Main Display, when starting the second Display, the result will still be shown in the Main Display, and the second Display will show the same measurement result as the Main Display. -
If different measurement functions are used in both Main and the second Display.
• The readings in both displays will update alternately.
• If math function (dBm, dB) is used in Main Display, when opening the second Display, the math operation will be stopped automatically. The second Display will show the second selected function normally.
• If math function (Statistics, Limits, Relative) is used in Main Display, when opening the second Display, the result will still be shown in the Main Display and the second Display will show the second selected function normally. -
If Temperature function is used in Main Display, set the display mode ( [Temp]→[Display]→[All]). Then the result will be shown in the Main Display and the current measurement value is shown in the second Display.
-
Auto Range is adopted by the second Display. If the same measurement function is used in both displays, so does the range.
-
Measured data in the second Display cannot be saved into “History”.
Utility Function
The Utility function enables users to set up system parameters and interface
parameters of the DAQ.
Press [Shift] and [Dual] to enter the operating menu of Utility function, as
the following diagram shows.
Function Menu | Description |
---|---|
Store/Recall | Store or recall state files. |
Manage File | Create a new file, copy, rename or delete a file. |
I/O Config | Configure LAN. |
Test/Admin | Provide board test function. |
System Setup | Configure instrument’s user settings. |
Store and Recall
The Store/Recall function enables users to store and recall the instrument
state and data files in the local storage as well as in the USB storage. After
entering the function menu of Utility, press [Store/Recall] to enter the
interface as shown in the diagram below.
Function Menu | Description |
---|---|
Store Settings | Store state or data files. |
Recall settings | Recall state files. |
Power On | Select the state that is loaded at power-up. |
Security Erase | Delete all the files stored in local storage and restore the |
instrument to factory default state.
Set to Defaults| Restore the instrument to factory default state.
Done| Return to the higher level menu.
Store Settings
Store settings allows you to save the system configuration (as .xml) or
measurement data (as .csv) into the internal memory or an external USB storage
device. After entering into the function menu of Store/Recall, press [Store
Settings] to enter the following interface.
Table below shows the different functions available.
Function Menu | Settings | Description |
---|---|---|
Browse | Choose the location that file will be saved. | |
File Name | Input the file name. | |
Type | .xml/.csv | Choose the type of file that will be saved. |
Store Data | Store the specified file. | |
Done | Return to the higher level menu. |
Operating Steps:
-
Set the storage directory
Press [Browse] to enter the following interface, then use the direction keys or menu operation keys to choose the storage directory. Press [Select] to set the current directory as storage location and Return to the higher level menu. -
Set the file name
Press [File Name] to enter the following interface and input the name of the stored file.
The method of inputting a file name:
- Press direction keys to select a desired character in the input area.
- Press “OK” key on the front panel to input selected char in the input area.
- Press [Clear All] to clear all input chars.
- Press [Delete Char] to delete the letter on which the cursor is currently placed.
- Press [Previous Char] to move the cursor in the file name area to the previous char.
- Press [Next Char] to move the cursor in the file name area to the next char.
- Press [Done] to save the current file and return to the higher level menu.
- Press [Cancel] to cancel the current operation and return to the higher level menu.
3. Set the type of stored file
Press [Type] to set the type of stored file.
- .xml: save the current system configuration as an “.xml” file.
- csv: save the current measurement result as a “.csv” file.
4. Save the file
Press [Store Data] to store the specified file.
5. Exit
Press [Done] to return to the higher level menu.
Recall Settings
Recall settings allows you to read the system configuration from the internal
memory or an external USB storage device. After entering the function menu of
Store/Recall, press [Recall Settings] to enter the following
interface.Use direction keys or menu
operation keys to choose the state file with the suffix “.xml”.
- Press [Select] to read the file and restore the instrument to a specified state.
- Press [Cancel] to return to higher level menu.
Power On
Select a system configuration to be used at power-on from “Default” and “Last” (configuration at last power-off). The setting will be available at the next power-on.
Security Erase
Press [Security Erase], the instrument will display a prompt message “Are you
sure you want to delete all stored files?”. Press [Yes] to confirm. The
instrument will delete all the files stored in local storage and restore the
instrument to the factory default state.
Set to Defaults
Press [Set to Defaults] and the instrument will be restored to factory default
state.
Manage File
The Manage Files function allows you to create, copy, delete, and rename files
and folders in the instrument’s internal flash memory or on a USB drive
attached to the front panel. It also allows you to capture the current screen
to a bitmap (*.bmp) file. After entering the function menu of Utility, press
[Manage File] to enter the interface as shown in the diagram below.
Press [Action] and select [Folder], [Capture Display], [Copy], [Rename] or [Delete] to do the corre- sponding operation.
- Folder – To create a folder, Browse to the internal or external location for the folder, press File Name, enter a folder name and press Done. Press Create Folder > Done.
- Capture Display – To save a screen capture, Browse to the internal or external location for the screen capture. Press File Name, enter a name and press Done. Press Save Screen > Done.
- Copy – To copy a file or folder, press Copy. Browse to the folder or file to be copied and press Select. Press Copy Path and select an internal or external path for copying. Press Perform Copy > Done.
- Rename – To rename a file or folder, press Rename. Browse to the folder or file to be renamed and press Select. Press New Name, enter a new name and press Done. Press Perform Rename > Done.
- Delete – To delete a file or folder, press Delete and Browse to the folder or file to delete.
Press Select > Perform Delete > Done.
I/O Configuration
Press [I/O Config] to enter the following interface and set up the
parameters.
LAN Settings
The DAQ enables users to operate the instrument remotely by LAN interface and
store or recall internet settings. You can look over current LAN settings and
set up an IP address and subnet mask. After entering into the function menu of
Utility, press [I/O Config]. Select [On ]→[LAN Set- tings]→[Modify Settings]
to enter the following interface.
Function Menu | Description |
---|---|
IP Address | Set up IP address. |
Subnet Mask | Set up subnet mask. |
Gateway | Set up gateway. |
Cancel | Cancel current operation and return to the higher level menu. |
Done | Save all changes and return to the higher level menu. |
Board Test
T3DAQ1-16 provides self-test functions, including Key Test, LCDTest, Beeper
Test and Chip Test.
Operating Steps:
- Press [Shift] and [Dual], then choose [Test/Admin]→[Board Test] to enter the following inter- face.
Function Menu | Description |
---|---|
Test the instrument’s keys. | Test the instrument’s keys. |
LCD | Test the instrument’s LCD screen. |
Beeper | Test the instrument’s beeper. |
Chip | Test the instrument’s chips. |
Done | Return to the higher level menu. |
2. Test the keys (select Keyboard).
Select [Key] to enter the key test interface, as displayed in the diagram
below. The on- screen rectangle shapes represent the keys on the front panel.
Test all keys and knobs and you should also verify that all the back lit
buttons illuminate correctly.
Before testing the keys the key shapes on the screen will be displayed in a
blue color. The on screen keys and knobs will change to a green color when
tested. Press [Done] to exit the test.
3. Test the LCD screen.
Select [LCD] to enter the screen test interface, the screen shows the message:
“Press ‘Change’ to change. Press ‘Done’ to exit”. Press [Change] to start the
test and observe
if the screen has changed color or other display errors as shown in the
diagram below
NOTE:
- Press [Change] to change the color of the screen. There are three colors: red, blue and green.
- Press [Done] to exit the test.
4. Test the beeper.
Press [Beeper] to test the beeper. Under regular circumstances, press [Beeper] one time and the instrument will beep one time.
5. Test the chips.
Press [Chip]→[Start] to enter chip test interface, as shown in the diagram below.NOTE:
• If the test is passed, the corresponding result shows “pass”.
• If the test is failed, the corresponding result shows “fail”.
6. Press [Done] to exit the board test.
Firmware Update
The software of the DAQ can be updated directly via a USB flash drive,
updating the current soft- ware version to the desired software version.
Operating Steps:
- Copy the update file to the USB flash drive.
- Insert the USB flash drive to the USB host interface on the front panel of the DAQ.
- Press [Shift]→ [Dual]→ [System Setup]→ [Firmware Update], then press [Browse] and select the update file. Next, press [Update]→ [Yes] to start updating the system software.
- On completion of the update, the screen will show the message: “Firmware Update Done!” Then you can remove the USB flash drive away.
- Restart the DAQand check the version information. Press [Shift]→ [Dual]→ [System Setup]→ [System Info] to check if the software and hardware version has updated. If not, and the up- date has failed, you will need to update once more following the above operating steps.
- After checking, press [Done] to exit the system information interface.
NOTE:
Do not disconnect the power or turn off the instrument while the instrument is
updating.
System Setup
Press [Shift] and [Dual], then select [System Setup] to enter the following
interface.
Function Menu | Description |
---|---|
Language | Select the display interface language. |
Firmware Update | Update software version. |
Screen | Setup the screen protection function. |
System Info | View system information. |
Done | Return to the higher level menu. |
-
Select language.
The DAQ supports two languages, English and Chinese. Press [Language] to select the Language of the menu. -
Set up the screen protection timer.
Press [Screen] to set screen protection as 1 Min, 5 Mins, 15 Mins, 30 Mins,1 Hour, 2 Hours or 5 Hours according to different demands. Activate the screen saver program and screen saver will be on if no action is taken within the time that you have selected. Press any button to resume. -
View system information.
Press [System Info] to view system information, including start-up times, software version, hardware version, production ID and serial number, as shown in the following diagram
Acquire
Sampling is a process of acquiring and digitizing a signal. The optional
Trigger methods of the DAQ include Auto Trigger, Single Trigger and External
Trigger. Press [Acquire] to enter the interface shown as the following
diagram:
Function Menu | Description |
---|---|
Trg Src | Set the source of trigger. |
Slope | Set the slope polarity of external trigger. |
Delay | Set the delay. |
Samples/Trigger | Set the number of samples per trigger. |
VMC Out | Set the polarity output pulse signal when the sampling signal |
is finished.
Auto Trigger
Auto Trigger parameters that need to be set up include delay, samples/trigger
and VMC out.
Operating Steps:
-
Press [Acquire], then select [Trg Src]→[Auto] or press [Run/Stop] on the front panel directly to enable Auto Trigger.
-
Set the delay.
Delay is the waiting time after the trigger signal is sent out and before the acquiring starts.
Press [Delay] to select Auto or Manual mode. When choosing Manual mode, the Left and Right keys are used to select the numerical digit to adjust. The Up and Down keys are used to change the selected digit value. -
Set the number of samples per trigger.
Press [Samples/Trigger] to set the sample count. Left and Right keys are used to switch the number of a numerical value, Up and Down keys are used to change the selected value.Sample Count
• Sample Count indicates the number of samples taken when the DAQ receives a Single Trigger event (samples per trigger).
• The range of sampling points should be between 1 and 599999999.
• The default value of the Sample Count is 1. -
Set the VMC Out.
The DAQ outputs a pulse signal through the VM COMP interface on the rear panel after the sampling signal is finished. Press [VMC Out] to choose Positive or Negative polarity.
Single Trigger
Single Trigger parameters that need to be set up include delay,
samples/trigger and VMC out.
Operating Steps:
-
Press [Acquire], then select [Trg Src]→ [Single] or press [Single] on the front panel directly to enable Single Trigger.
-
Set the delay.
Press [Delay] to select Auto or Manual mode. -
Set the number of samples or trigger.
Press [Samples/Trigger] to set sample count. -
Set the VMC Out.
The DAQ outputs a pulse signal through the VM COMP interface on the rear panel after the sampling signal is finished.
Press [VMC Out] to choose Positive or Negative polarity.
External Trigger
The external trigger signal will be input via EXT TRIG interface on the rear
panel. External trigger
parameters that need to be set up include delay, samples/trigger, slope and
VMC out.
Operating Steps:
-
Press [Acquire], then select [Trg Src]→ [Ext] to enable the External Trigger.
-
Set the polarity of slope.
Press [Slope] to choose Positive or Negative polarity. -
Set the delay.
Press [Delay] to choose Auto or Manual mode. -
Set the number of samples or trigger.
Press [Samples/Trigger] to set sample count. -
Set the VMC Out.
In External Trigger mode, the DAQ could output a pulse signal through the VM COMP inter- face on the rear panel after the sampling signal is finished.
Help System
T3DAQ1-16 provides a powerful built-in help system. You can recall help
information at any time while using the instrument. You also can get a
functionality help for every button on the front panel or menu soft key by
using the built-in help system. You may also get help about familiar
operations with the help list.
Press [Shift] and [Acquire] to enter the help list, as the following diagram
shows.
Function Menu | Description |
---|---|
Up | Move up the cursor and select the help menu. |
Down | Move down the cursor and select the help menu. |
Select | Select the help information you want to read. |
Done | Return to the higher menu. |
-
Basic Measure.
Get basic measurement types and methods to connect the leads for different measurements. -
Measuring Temperature.
Get the method to measure temperature. -
Measuring Capacitance.
Get the method to measure capacitance. -
Math Function.
Introduction of how to use the math function while making measurements. -
Dual-display Function.
Learn the method to use the dual-display function while you are measuring. -
Saving and Recalling Information.
Introduction of how to store and recall the data/parameter/sensor files. -
The convention and Tips of Soft Key usage.
Get help and tips about using the soft keys. -
Technical Support.
How to obtain technical support.
Explanation:
- In the help menu interface, you can move the cursor and select the corresponding menu by the up and down direction keys and press “OK” to read the help information.
- While reading the help information, you can also look up and down the information by selecting the up and down direction keys.
Math Functions
The DAQ provides five math functions: Statistics, Limits, dBm, dB and
Relative. Choose differ- ent math functions to meet different measurement
needs. Math functions can only be used in DC Voltage, AC Voltage, DC Current,
AC Current, Resistance, Frequency, Period and Temperature measurement. Among
these functions, dBm and dB are only used in DC Voltage and AC Voltage
measurement.
Press [Math] to show the operating menu of math functions on the screen, as
shown in the follow- ing diagram.
Function Menu | Settings | Description |
---|---|---|
Statistics | Reading statistic functions, including max, min, aver- |
age, span, std dev and samples.
Limits| | The Limits function performs Pass/Fail testing accord-
ing to the specified upper and lower limits.
dBm| | The dBm is based on a calculation of power delivered
to a reference resistance, OdBm = lmW
db| | The dB measurement is the difference between the
input signal and a stored relative value.
Rel Value| Val ue/Off| Turn on the relative value function and set up the
value. Or turn off the function.
Explanation:
- Math functions are only applicable to the main display.
- If a measurement function is changed, all math functions will be stopped except Statis- tics.
Statistics
There are many kinds of reading statistic functions, including: Max, Min,
Average and Standard deviation.
Press [Math]→ [Statistics]→ [Show] to enter the interface shown in the
following diagram.
Function Menu | Settings | Description |
---|---|---|
Statistics | Reading statistic functions, including max, min, aver-age, span, |
std dev and samples.
Limits| | The Limits function performs Pass/Fail testing accord-ing to the
specified upper and lower limits.
dBm| | The dBm is based on a calculation of power delivered to a reference
resistance, OdBm = 1 mW
db| | The dB measurement is the difference between the input signal and a
stored relative value.
Rel Value| Value/Off| Turn on the relative value function and set up the
value. Or turn off the function.
Explanation:
- Math functions are only applicable to the main display.
- If a measurement function is changed, all math functions will be stopped except Statis- tics.
Statistics
There are many kinds of reading statistic functions, including: Max, Min,
Average and Standard deviation.
Press [Math]→ [Statistics]→ [Show] to enter the interface shown in the
following diagram.![TELEDYNE T3DAQ1 16 16 Channel Single Slot Data Acquisition
- Fig 72](https://manuals.plus/wp-content/uploads/2023/04/TELEDYNE- T3DAQ1-16-16-Channel-Single-Slot-Data-Acquisition-Fig-72.jpg)
Function Menu | Settings | Description |
---|---|---|
Min | Show the minimum statistics value of current measurement. | |
Average | Show the average statistics value of current measurement. | |
Max | Show the maximum statistics value of current measure- ment. | |
Span | Show the span of current measurement. | |
Std dev | Show the std dev statistics value of current measurement. | |
Samples | Show the number of samples of current measurement. | |
Statistics | Show/Hide | Show or hide the statistics function interface. |
Clear Readings | Clear all current readings and restart statistics. | |
Done | Return to the higher level menu. |
Statistics Function:
-
In statistic function, the first reading is usually set to the maximum or minimum value. When acquiring more readings the current displayed value is always the maxi-
mum/minimum reading among all the measured values. -
The maximum, minimum, average and reading quantities are stored in volatile mem- ory.
Limits
Limits function is available to indicate signals beyond ranges according to
the upper and lower pa- rameters. The following are some measurement functions
which are able to do a limit operation: DC Voltage, AC Voltage, DC Current, AC
Current, Resistance, Frequency, Period, Capacitance and Temperature.
Press [Math]→ [Limits]→ [On] to enter the interface shown in the following
diagram.
The table shows the description of Limits Measurement function menu.
Function Menu | Settings | Description |
---|---|---|
Limits | On/Off | Turn on or turn off the Limits function. |
Low | Set the desired lower limit. | |
Center | Set the desired center value. | |
High | Set the desired upper limit. | |
Status | Show the status of limit test. | |
Low Failures | Show the times that reading is lower than the limit. | |
High Failures | Show the times that reading is higher than the limit. | |
Span | Set the desired span. | |
Beeper | On/Off | When the beeper is on, if the reading is lower or higher than |
limits, the instrument will beep once.
Clear Condition| | Clear all current readings and restart to test.
Done| | Save all changes and return to the higher level menu.
-
How to Set Limits
Select [High], [Low], [Center] or [Span] and then switch to the needed digit using the Left or Right Direction keys and input the numerical value by selecting the Up and Down Direction keys. -
Unit
The unit of Limits is decided by the current measurement function. -
Notes
• When the reading is lower than the set lower limit, the color of the main display will switch from blue to red.
• When the reading is higher than the set higher limit, the color of the main display will switch from blue to red.
• When the reading is lower or higher than the set limits, the Beeper will beep once. (The beeper is turned on.)
The range of Limits function:
- The Limits range is -120% to +120% of the current measurement range.
- The upper limit value should be always bigger than the lower limit value.
- The upper and lower values are stored in volatile memory. They will be set to default values when the power is initially turned on.
dBm
The dBm function is logarithmic and based on a calculation of power delivered
to a reference resistance, relative to 1 milliwatts. This function only
applies to an AC voltage and DC voltage measurements.
Press [Math]→ [dB/dBm]→ [On] and select [Function dBm] to enter the interface
shown in the fol- lowing diagram.
Function Menu | Settings | Description |
---|---|---|
dB/dBm | On/Off | Turn on or turn off dB or dBm function. |
Function | Open the dBm function and the lower right corner of |
the main display will show ”dBm”.
Ref R| | Set the parameter via direction keys: 50Ω – 8000Ω
Done| | Save all changes and return to the higher level menu.
The computation method for dBm:
When the dBm function is turned on, the measured value of voltage is
transformed into dBm ac- cording to the below formula. dBm = 10 ∗
log10[(Reading2/RREF)/0.001W] dB
Each dB measurement is different between the input signal and a stored
relative value, with both values converted to dBm. The dB function applies to
AC voltage and DC voltage measurements only.
Press [Math]→ [dB/dBm On] and select [Function dB] to enter the interface
shown in the following diagram.
Function Menu | Settings | Description |
---|---|---|
dB/dBm | On/Off | Turn on or turn off dB or dBm function. |
Function dB | Open dB function and the lower right corner of the |
main display show “dB”.
Ref R| | Set the parameter via direction keys: 50Ω – 8000Ω
dB Ref Value| | Set the referred value of dB.
Measure Ref
Value| | Set the referred value of measurement.
Done| | Save all changes and return to the higher level menu.
The computation method of the dB:
dB = 10 ∗ log10[(Reading2/RREF)/0.001W] − (dB setting value)
RREF expressed measuring the resistance value in the actual electric circuit.
Range of the dB setting value: -200 dBm to +200 dBm. The default is 0 dBm.
dB value:
- Input a value in dB setting interface using the direction buttons, and then store it as dB setting value.
- Settings of dB value are stored in volatile memory.
Relative Value
Relative value is used for relative measurements. Actual measurement reading
is the difference between measurement value and preset value.
The DAQ Relative Value is available for the following measurement parameters:
DC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Period,
Capacitance and Temperature. Press [Math]→ [Rel Value] to enter the interface
shown in the following diagram.
Function Menu | Description |
---|---|
Value | Select the current measurement value as the preset value. |
Off | Turn off the relative operation function. |
When Rel Value function is turned on, the result of relative measurement will
display on the screen.
Main display = Measurement value – Preset value
Display Mode
The DAQ supports four types of views for measured data: “Number”, “Bar Meter”,
”Trend Chart” and “Histogram”.
Number
Press [Shift] and [Math] to open the menu of display mode and press [Display]
to enter the following interface. “Number” is always the selected mode when
the DAQ is turned on.
Bar Meter
Operating Steps:
- Press [Bar Meter] to enter Bar Meter display mode.
- Press [Horizontal Scale] to set the vertical scale as Default Manual mode.
Table below shows the functions menu and the description.
Function Menu | Description |
---|---|
Low | Set the low value of horizontal scale. |
High | Set the high value of horizontal scale. |
Center | Set the center value of horizontal scale. |
Span | Set the span of horizontal scale. |
Done | Save all changes and return to the higher level menu. |
Trend Chart
Operating Steps:
- Press [Trend Chart] to enter the Trend Chart display mode.Table below describes the Trend Chart Display Mode.
Function Menu | Description |
---|---|
Display Trend | The current select display mode is Trend Chart. |
Recent All | Show recent or all readings. |
Horizontal Scale | Choose the mode of Horizontal Scale. |
Autoscale Once | Automatically set the horizontal scale once. |
Clear Readings | Clear all current readings and restart statistics. |
2. Press [Horizontal Scale] to choose the way to set the horizontal scale as
Default, Auto or Manual mode.
Press [Auto] and the DAQ will set the vertical scale automatically.
Press [Manual] and you can set the vertical scale manually, as displayed in the diagram below.![TELEDYNE T3DAQ1 16 16 Channel Single Slot Data Acquisition
- Fig 79](https://manuals.plus/wp-content/uploads/2023/04/TELEDYNE- T3DAQ1-16-16-Channel-Single-Slot-Data-Acquisition-Fig-79.jpg)
Histogram
Operating Steps:
Press [Histogram] to enter the Histogram display mode.
- Table below describes the Histogram Display Mode.
Function Menu| Settings| Description
---|---|---
Display Histogram| | Currently selected display mode is Histogram.
Binning| | Set Binning as Auto or Manual mode.
Bin Set| | Set the Bin parameters.
Cumulative| On/Off| Turn on or off the Cumulative function.
Clear Readings| | Clear all current readings and restart statistics. - Press [Binning] to set Binning as the Auto or Manual mode. When in Manual mode, press [Bin Settings] to enter the following interface.
Function Menu| Settings| Description
---|---|---
Num.Bins| | Set the number of Bins, 10, 20, 40, 100, 200 or 400 selected.
Low| | Set the low value of the horizontal scale.
High| | Set the high value of the horizontal scale.
Center| | Set the center value of the horizontal scale.
Span| | Set the span of the horizontal scale.
Outer Bins| On/Off| Set the bins beyond the set range or not.
Done| | Save all changes and return to the higher level menu.
Trigger
The DAQ supports the Trigger function. Press [Run/Stop] or [Single] on the
front panel to trigger the DAQ by Auto or Single mode. Auto trigger is
considered as a default when the power is initially turned on.
Auto Trigger
Press [Run/Stop] on the front panel one time and the Auto Trigger will be
started to capture contin- uous readings automatically. The screen will show
“Auto Trigger”. Press [Run/Stop] again and the trigger is stopped.
Single Trigger
Press [Single] on the front panel, the Single Trigger will be started one time
and generate a single reading. The screen will show “Single Trig”.
Explanation
In the Remote Mode, the screen will show “Imme Trig”. Press [Shift] to switch
back to the local mode and the DAQ will choose Auto Trigger automatically.
Hold Measurement Function
Hold Measurement function provides users with a stable reading on the screen
of the front panel.
When the test leads are disconnected, the reading is still held on the screen,
which enables users to view the measured history data.
Press [Shift] and [Single] to open Hold measurement function interface, the
black field just above the screen will show “Probe Hold”, as shown in the
following diagram![TELEDYNE T3DAQ1 16 16 Channel Single Slot Data Acquisition
- Fig 82](https://manuals.plus/wp-content/uploads/2023/04/TELEDYNE- T3DAQ1-16-16-Channel-Single-Slot-Data-Acquisition-Fig-82.jpg)
Function Menu | Settings | Description |
---|---|---|
Probe Hold | On/Off | Turn on or off the Probe Hold function. |
Beeper | On/Off | Turn on or off the Beeper. |
Clear List | Clear all current readings and restart the statistics. | |
Chapter 3 Measurement Tutorial
True RMS AC Measurement
The AC measurement of the DAQ has a true RMS response. The power dissipated in
a resis- tor within a time is proportional to the square of the measured true
RMS voltage, independent of waveform shape. The instrument can accurately
measure true RMS voltage or current, as long as the waveform shape contains
negligible energy above the effective bandwidth.
The AC voltage and AC current functions measure the “AC coupled” true RMS
value, which is to measure the RMS value of the AC component (DC component is
rejected) of the input signal. For sine waves, triangle waves, and square
waves, the AC and AC+DC values are equal since these waveforms do not contain
a DC offset. See the table below.
An AC coupled true RMS measurement is desirable in situations where you are measuring small AC signals in the presence of DC offsets. For instance, measuring an AC ripple present on DC power supplies. There are situations, however, where you might want to know the AC+DC true RMS value. You can determine this value by combining results from DC and AC measurements as the following shows. You should perform the DC measurement using 6.5 digit mode for best AC rejection. ****
Crest Factor Errors (non-sinusoidal inputs)
A common misconception is that ”since an AC Multimeter is a true RMS, its sine
wave accuracy specifications apply to all waveforms.” The shape of the input
signal can dramatically affect mea- surement accuracy. A common way to
describe signal wave shapes is “crest factor”. Crest factor is the ratio of
the peak value to RMS value of a waveform.
The greater the crest factor, the greater the energy contained in high
frequency harmonics. All Multimeters have errors that are crest factor
dependent. (The crest factor errors do not apply for input signals below
100Hz.)
You can estimate the measurement error due to signal crest factor as shown
below: Total Error = Error (Sine wave) + Error (Crest factor) + Error
(Bandwidth)
Error (Sine wave): error for sine wave Error (Crest factor): crest factor
additional error. Error (Bandwidth): estimated bandwidth error as shown below:
C.F.: signal crest factor
F: fundamental frequency of pulse
BW: effective bandwidth of the Multimeter
Example: Calculate the approximate measurement error for a pulse train
input with a crest factor of 2 and a fundamental frequency of 20 kHz. For this
example, assume 1-year accuracy specifications of the Multimeter: ±(0.05%×
reading + 0.03%× range).
Total Error = (0.05%× reading+0.03%× range) + (0.05%× range) + (0.8%× reading)
= 0.85%× reading + 0.08%× range
Loading Errors (AC Voltage)
In the AC Voltage function, the input of T3DAQ1-16 appears as a 1MΩ resistance
in parallel with 100pF of capacitance. The test lead that you use to connect
signals to the Multimeter will also add additional capacitance and loading.
The approximate input resistances of the Multimeter at different frequencies
are listed in the following table.
Table below shows the approximate input resistances at different frequencies.
Input Frequency | Input Resistance |
---|---|
100Hz | 1MΩ |
1kHz | 850kΩ |
10kHz | 160kΩ |
100kHz | 16kΩ |
For low frequencies:
For high frequencies:
F: input frequency
Rs: source resistance
Cm: input capacitance (100pF) plus test lead capacitance
Chapter 4 General Troubleshooting
The commonly encountered failures and their solutions are listed below. When
you encounter those problems, please resolve them using the following steps.
If the problem remains, please contact the Teledyne LeCroy service center and
provide your device Information including serial number.
The screen has no display after pressing the power key.
- Check whether the power cord is fully connected.
- Check whether the power fuse has blown or has failed. If the fuse needs to be changed, please use the specified fuse.
- Restart the instrument after finishing the above checks.
- If the instrument still doesn’t start up properly, please contact the Teledyne LeCroy service center.
The reading doesn’t change when a current signal is input.
- Check whether the test lead is correctly inserted into the HI and LO terminals of current measurement.
- Check whether the current fuse at the back panel has blown.
- Check whether the DCI or ACI measurement function is enabled.
- Check whether the DCI measurement function is used to measure AC current.
The reading is abnormal when a voltage signal is input.
- Check whether the test lead is correctly inserted into the HI and LO terminals for voltage measurements.
- Check whether the DCV or ACV measurement function is enabled.
- Check whether the DCV measurement function is used to measure AC voltage.
The USB storage device cannot be identified.
-
Check whether the USB storage device is in good condition.
-
Make sure the USB storage device you used is a flash storage device. This instrument does not support hardware storage types.
-
Check the capacity of your USB storage device. It is recommended that the capacity of the
USB storage device is no larger than 8G bytes and FAT formatted. -
Restart the instrument, then insert the USB storage device.
-
If the problem persists, please contact the Teledyne LeCroy service center.
Chapter 5 Appendix
Appendix A: Accessories
Standard Accessories:
-
Power cords.
-
Test lead set.
-
USB Cable.
NOTE: -
We suggest that the length of the USB data wire and LAN cable connected to the instrument should be less than 3m to avoid affecting the product performance.
-
All the accessories are available by contacting your local Teledyne LeCroy office.
Appendix B: Warranty summary
Teledyne Test Tools warrants that the products will be free from defects in
materials and work- manship for a period of three years from the date of
shipment from an authorized Teledyne Test Tools distributor. If a product
proves defective within the respective period, Teledyne Test Tools will
provide repair or replacement as described in the complete warranty statement.
To arrange for service or obtain a copy of the complete warranty statement,
please contact your nearest Teledyne Test Tools sales and service office.
Except as provided in this summary or the applicable warranty statement.
Teledyne Test Tools makes no warranty of any kind, expressed or implied,
including without limitation the implied warranties of merchantability and
fitness for a particular purpose. In no event shall Teledyne Test Tools be
liable for indirect, special or conse-
quential damages.
Certifications
Teledyne LeCroy certifies compliance to the following standards as of the time
of publication. Please see the EC Declaration of Con- formity document shipped
with your product for current certifications.
EMC Compliance
EC DECLARATION OF CONFORMITY – EMC
The instrument meets intent of EC Directive 2014/30/EU for Electromagnetic
Compatibility. Compliance was demonstrated to the fol- lowing specifications
listed in the Official Journal of the European Communities:
EN 61326-1:2013, EN 61326-2-1:2013 EMC requirements for electrical equipment
for measurement, control, and laboratory use. 1
Electromagnetic Emissions:
EN 55011:2016+A1:2017, Radiated and Conducted Emissions Group 1, Class A2,3
EN 61000-3-2:2014 Harmonic Current Emissions, Class A
EN 61000-3-3:2013 Voltage Fluctuations and Flickers, Pst = 1
Electromagnetic Immunity:
EN 61000-4-2:2009 Electrostatic Discharge, 4 kV contact, 8 kV air, 4 kV
vertical/horizontal coupling planes 4
EN 61000-4-3:2006+ A2:2010 RF Radiated Electromagnetic Field,
3 V/m, 80-1000 MHz; 3 V/m, 1400 MHz – 2 GHz; 1 V/m, 2 GHz – 2.7 GHz
EN 61000-4-4:2012 Electrical Fast Transient/Burst, 1 kV on power supply lines,
0.5 kV on I/O signal data and control lines 4
EN 61000-4-5:2014+A1:2017 Power Line Surge, 1 kV AC Mains, L-N, L-PE, N-PE 4
EN 61000-4-6:2014 RF Conducted Electromagnetic Field, 3 Vrms, 0.15 MHz – 80
MHz
EN 61000-4-11:2004+A1:2017 Mains Dips and Interruptions, 0%/1 cycle, 70%/25
cycles, 0%/250 cycles 4,5
1 To ensure compliance with all applicable EMC standards, use high-quality
shielded interface cables.
2 Emissions which exceed the levels required by this standard may occur when
the instrument is connected to a test object.
3 This product is intended for use in nonresidential areas only. Use in
residential areas may cause electromagnetic interference.
4 Meets Performance Criteria “B” limits of the respective standard: during the
disturbance, product undergoes a temporary degradation or loss of function or
performance which is self-recoverable.
5 Performance Criteria “C” applied for 70%/25 cycle voltage dips and for
0%/250 cycle voltage interruption test levels per EN61000-4-11.
European Contact:∗
Teledyne GmbH, European Division
Im Breitspiel 11c
D-69126 Heidelberg
Germany
Tel: + 49 6221 82700
AUSTRALIA & NEW ZEALAND DECLARATION OF CONFORMITY – EMC
The instrument complies with the EMC provision of the Radio Communications Act
per the following standards, in accordance with requirements imposed by
Australian Communication and Media Authority (ACMA): AS/NZS CISPR 11:2015
Radiated and Conducted Emissions, Group 1, Class A.
Australia / New Zealand Contacts:∗
RS Components Pty Ltd.
Suite 326 The Parade West
Kent Town, South Australia 5067
RS Components Ltd.
Units 30 & 31 Warehouse World
761 Great South Road
Penrose, Auckland, New Zealand
∗ Visit
teledynelecroy.com/support/contact
for the latest contact information.
Safety Compliance
EC DECLARATION OF CONFORMITY – LOW VOLTAGE
The instrument meets intent of EC Directive 2014/35/EU for Product Safety.
Compliance was demonstrated to the following specifica- tions as listed in the
Official Journal of the European Communities:
EN 61010-1:2010 Safety requirements for electrical equipment for measurement,
control, and laboratory use – Part 1: General requirements
EN 61010-2:030:2010 Safety requirements for electrical equipment for
measurement, control, and laboratory use – Part 2-030: Particular requirements
for testing and measuring circuits
The design of the instrument has been verified to conform to the following
limits put forth by these standards:
- Mains Supply Connector: Overvoltage Category II, instrument intended to be supplied from the building wiring at utilization points (socket outlets and similar).
- Measuring Circuit Terminals: No rated measurement category. Terminals not intended to be connected directly to the mains supply.
- Unit: Pollution Degree 2, operating environment where normally only dry, non-conductive pollution occurs. Temporary conductivity caused by condensation should be expected. Environmental Compliance
END-OF-LIFE HANDLING
**** The instrument is marked with this symbol to indicate that it complies
with the applicable European Union require- ments of Directives 2012/19/EU and
2006/66/EC on Waste Electrical and Electronic Equipment (WEEE) and Batter-
ies.
The instrument is subject to disposal and recycling regulations that vary by
country and region. Many countries pro- hibit the disposal of waste electronic
equipment in standard waste receptacles. For more information about proper
disposal and recycling of your Teledyne LeCroy product, please visit
teledynelecroy.com/recycle.
RESTRICTION OF HAZARDOUS SUBSTANCES (RoHS)
EC DECLARATION OF CONFORMITY – RoHS
Unless otherwise specified, all the materials and processes are compliant with
RoHS Directive 2011/65/EU in its entirety, inclusive of any further amendments
or modifications of said Directive.
CHINA RoHS 2
Unless otherwise specified, all the materials and processes are compliant with
the latest requirements of China RoHS 2. The hazardous substances contained in
the instrument are disclosed in accordance with the standards SJ/T 11364-2014
(Marking for the restricted use of hazardous substances in electronic and
electrical products) and GB/T 26572-2011 (Requirements on concentration limits
for certain restricted substances in electrical and electronic products). The
instrument is marked with an appropriate Environmental Friendly Use Period
(EFUP) symbol. The packaging materials include the appropriate recycling
labels. The below substance disclosure tables (in
Chinese and English languages) provide the required compliance information.
Part Name | Toxic or Hazardous Substances and Elements |
---|
Lead
(Pb)| Mercury
(Hg)| Cadmium
(Cd)| Hemavalent
Chromium
(Cr6+)| Polybrominate
d Biphenyls
(PBB)| Polybrominated
Diphenyl Ethers
(PBDE)
PCBAs| X| 0| 0| 0| 0| 0
Mechanical
FIRIAW At P| 0| 0| 0| 0| 0| 0
Sheet Metal| 0| 0| 0| 0| 0| 0
Plastic Parts| 0| 0| 0| 0| 0| 0
Cable Assemblies| X| 0| 0| 0| 0| 0
Display| 0| 0| 0| 0| 0| 0
Power Supply| 0| 0| 0| 0| 0| 0
Fans| 0| 0| 0| 0| 0| 0
Batteries| 0| 0| 0| 0| 0| 0
Power Cord| 0| 0| 0| 0| 0| 0
Ext Power Supply fif nrecentl| X| 0| 0| 0| 0| 0
Probes (if present)| X| 0| 0| 0| 0| 0
Fuse (if present)| 0| 0| 0| 0| 0| 0
Product Case fif nrPccntl| 0| 0| 0| 0| 0| 0
AdaptersHodules fif nre5entl| 0| 0| 0| 0| 0| 0
Mouse (if present)| 0| 0| 0| 0| 0| 0
0: Indicates that this toxic or hazardous substance contained in all of the
homogeneous materials for this part is below the limit requirement specified
in SJ?T11364-2014.
X: Indicates that this toxic or hazardous substance contained in at least one
of the homogenous materials used for this part is above the limit requirement
specified in SJ/T11364-2014.
EFUP (Environmental Friendly Use Period): 30 years. Use Conditions: Refer to
the environmental conditions stated in the User Manual. EFUP for Probes: 10
years.
CompanyProfile
Teledyne LeCroyis a leading provider of oscilloscopes, protocol analyzers and
related test and measurement solutions that enablecompanies across a wide
range of industries to design and test electronic devices of all types.
Sinceour founding in 1964,we have focused on creating products that improve
productivity by helping engineers resolve design issues faster and more
effectively. Oscilloscopes are tools used by designers and engineers to
measure and analyzecomplex electronic signals in order to develophigh-
performance systems and to validateelectronic designs in order to improve time
to market.
The Teledyne Test Tools brand extends the Teledyne LeCroy product portfolio
with a comprehensiverange of test equipment solutions. This new range of
products delivers
a broad range of quality test solutions that enable engineers to rapidly
validate product and design and reduce time-to- market. Designers, engineers
and educators rely on Teledyne Test Tools solutions to meet their most
challenging needs for testing, education and electronics validation.
Location andFacilities
Headquartered in Chestnut Ridge,New York, Teledyne Test Tools and Teledyne
LeCroyhas sales, service and development subsidiaries in the USand throughout
Europeand Asia. Teledyne Test Tools and Teledyne LeCroy products are employed
across a wide variety of industries, including semiconductor, computer,
consumer electronics,
education,military/aerospace, automotive/industrial, and telecommunications.
Distributed by:
Test Equipment Depot – 800.517.8431 – 5 Commonwealth Ave, MA 01801
TestEquipmentDepot.com
© 2021 Teledyne Test Tools is a brand and trademark of Teledyne LeCroyInc.
All rights reserved.Specifications,prices,availability and deliverysubject to
change without
notice. Product brand or brand names aretrademarks or requested trademarks of
their respective holders.
T3 stands for TeledyneTest Tools.
934368 Rev1
5 Commonwealth Ave
Woburn, MA 01801
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
References
- Teledyne LeCroy - RoHS and WEEE
- Welcome to the Test Equipment Depot, your test equipment supplier of all the best brands.
- Welcome to the Test Equipment Depot, your test equipment supplier of all the best brands.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>