TELEDYNE LECROY T3PS Series T3LCR1002 LCR Meter User Guide
- June 9, 2024
- TELEDYNE LECROY
Table of Contents
TELEDYNE LECROY T3PS Series T3LCR1002 LCR Meter
INTRODUCTION
This quick start guide contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Teledyne LeCroy company.
The information in this quick start guide was correct at the time of printing. However, Teledyne LeCroy continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
SAFETY INSTRUCTIONS
Safety Symbols
These safety symbols may appear in the quick start guide or on the instrument.
WARNING
Warning: Identifies conditions or practices that could result in injury or
loss of life.
CAUTION
Caution: Identifies conditions or practices that could result in damage to the
instrument or to other properties.
When you notice any of the unusual conditions listed below, immediately
terminate operation and disconnect the power cable.
Please contact the Teledyne service center representative for repair of the
instrument. If you continue to operate without repairing the instrument, there
is a potential fire or shock hazard for operators.
The instrument is operating abnormally if:
- The instrument emits abnormal noise, smell, smoke, or a sparklike light during the operation.
- The instrument generates high temperature or electrical shock during operation.
- The power cable, plug, or receptacle on the instrument is damaged.
- Foreign substances or liquid has fallen into the instrument.
The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument.
Failure to comply with these precautions or with specific
WARNINGS elsewhere in this manual may impair the protection provided by
the equipment. In addition it violates safety standards of design,
manufacture, and intended use of the instrument.
-
Disclaimer
Teledyne assumes no liability for the customer’s failure to comply with these requirements. -
Ground The Instrument
To avoid electric shock hazards, the instrument chassis and cabinet must be connected to a safety earth ground by the supplied power cable with earth blade. -
DO NOT Operate In An Explosive Atmosphere
Do not operate the instrument in the presence of inflammable gasses or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard. -
Keep Away From Live Circuits
Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made by qualified maintenance personnel. Do not replace components with the power cable connected.
Under certain conditions, dangerous voltages may exist even with the power cable is removed. To avoid injuries, always disconnect the power and discharge the circuits before touching them. -
DO NOT Service Or Adjust Alone
Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present. -
DO NOT Substitute Parts Or Modify Instrument
Because of the danger of introducing additional hazards, do not install substitute parts or perform unauthorized modifications to the instrument.
Return the instrument to a Teledyne authorized service center for service and repair to ensure that safety features are maintained.
GETTING STARTED
This chapter describes names and functions of the front panel, rear panel, and screen display and provides the basic procedures for operating the T3LCR series.
Overview
The Teledyne T3LCR series is a general-purpose LCR meter for incoming inspection of components, quality control, and laboratory use.
The T3LCR series is used for evaluating LCR components, materials, and semiconductor devices over a wide range of frequencies (10 Hz to 300 kHz) and test signal levels (10.00mV to 2.00V & 100.0uA to 20.00mA).
With its built-in comparator, the T3LCR series can output comparison/decision results for sorting components into a maximum of ten bins. Furthermore, by using the handler interface, the T3LCR series can be easily combined with a component handler, and a system controller to fully automate component testing, sorting, and quality- control data processing.
The list Meas function permits entry of up to 10 frequencies or test signal levels points to be automatically measured.
Main Features
Correction Function| OPEN/SHORT correction Eliminates measurement errors due
to stray parasitic impedance in the test fixtures.
---|---
Comparator Function (Sorting)| • Bin sort The primary parameter can be sorted
into BIN1- BIN9, AUX, OUT and HI/IN/LO for each of the primary measurement
parameters. The sequential mode or tolerance mode can be selected as the
sorting mode.
• Limit Setup Absolute value, deviation value, and % deviation value can be
used for setup.
• BIN count Countable from 0 to 999999
Files| • Up to 10 setup conditions can be written to/read from the built-in
non-volatile memory.
Key Lock| • The front panel keys can be locked.
RS-232| • Complies with SCPI.
Front Panel
Description
1. LCD Display| 2. Soft-key
3. Number Key| 4. Cursor Key
5. Power Switch| 6. Measure key
7. Setup Key| 8. BNC Terminal
9. Trigger Key| 10. ESC Key
11. USB Disk Port (USB-Host)| 12. System Soft-key
Rear Panel
Description
1. Power Cable Receptacle(Outlet)
2. Frame Terminal
3. USB Interface (Type B)
4. RS-232 Interface
5. Handler Interface
Power On
Press the power key for at least 2 seconds. Release power key when the POWER
LED is lit.
Power Off
Press the power key at least 2 seconds. LCR down when you release the power
key.
Warm-up Time
T3LCR series is ready to be used as soon as the power-up sequence has completed. However, to achieve the specification accuracy, first warm up the instrument for 30 minutes.
Connect to the Device Under Test (DUT)
The T3LCR series uses a four terminal measurement configuration that provides easy, accurate and stable measurements and avoids mutual inductance, interference from measurement signals, noise and other factors inherent with other types of connections.
Connect to DUT
WARNING
Identifies conditions or practices that could result in injury or loss of
life. Do not apply DC voltage or current to the UNKNOWN terminals. Applying a
DC voltage or current may lead to device failure.
Connect the measurement sample (DUT) to the test port (or the test fixture,
cables, etc. after the DUT has been completely discharged.
SPECIFICATIONS
The following are the basic specifications for the T3LCR series. For detailed specifications, please see the user manual
General Specification
Display | RGB color TFT-LCD, Size: 3.5” (320×240) |
---|---|
Test Function | Cs-Rs, Cs-D, Cp-Rp, Cp-D, Lp-Rp, Lp-Q, Ls-Rs, Ls-Q, Rs-Q, Rp-Q, |
R-X, DCR, Z-θr, Z-θd, Z-D, Z-Q
Monitor Parameter| Z, D, Q, Vac, Iac, Δ, Δ%, θr, θd, R, X, G, B, Y (2
Parameter)
40 times/s, 10 times/s, 3 times/s
Test Frequency| T3LCR1300: 10Hz-300kHz T3LCR1100: 10Hz-100kHz T3LCR1002: 10Hz-
2kHz
Frequency range and resolution
Frequency range(F) | Resolution |
---|---|
10.00Hz ≤ F ≤ 99.99Hz | 0.01Hz |
100.0Hz ≤ F ≤ 999.9Hz | 0.1Hz |
1.000Hz ≤ F ≤ 9.999Hz | 1Hz |
10.00kHz ≤ F ≤ 99.99kHz | 10Hz |
100.0kHz ≤ F ≤ 300.0kHz | 100Hz |
Frequency Accuracy
Frequency range(F) ±0.01% 4 digit resolution
Display Range
Parameter | Resolution |
---|---|
L | 0.00001uH-9999.99H |
C | 0.00001pF-9999.99mF |
R, X, | Z |
G, B, | Y |
D | 0.00001-9.99999 |
Q | 0.00001-99999.9 |
θd | -179.999°-179.999° |
θr | -3.14159-3.14159 |
DCR | 0.00001Ω-99.9999MΩ |
Δ% | -999999%-999999% |
AC Test signal level | 10.00mV-2.00V (±10%) |
CV: 10.00mV-2.00V (±6%)
100.0uA-20.00mA (±10%)
CC: 100.0uA-20.00mA(±6%) (@2VMax)
DCR Test signal level| +2V, 0.066A(Max.), Output impedance fixed 30Ω
DC Bias: Internal| ±2.5V (0.5%+0.005V)
List Test| 10Step (Frequency/Voltage/Current)
Output impedance| 30Ω, 50Ω and 100Ω
Ranging| Auto, Hold and Nominal range. Total 9 Ranges.
Equivalent Circuit| Serial and Parallel
OPEN/SHORT Test| OPEN/SHORT Zeroing (ALL,SPOT)
Files| built-in 10 files and USB Disk 10 files, 9999 Log File, 999 Picture
File, 10000 Data (.csv)
Beep Feature| OFF/PASS/FAIL
Trigger Mode| Internal, Manual, External and Bus Trigger.
Interface| Handler interface and RS232/USB interface.
Environment
Specification Conditions | Temperature: 23˚C±5˚C, Relative Humidity: < 70%RH |
---|---|
Operating Environment | Temperature: 0~50˚C, |
Relative Humidity: <70%RH
(Indoor use only, Altitude: 2000 meters)
Storage Conditions| Temperature: -10~70˚C, Relative Humidity: <80%RH
Power Source| AC 100V-240V, 50Hz-60Hz
Fuse| 2A Slow-Blow
Maximum rated power| 30W
Weight| 3kg, net
DIMENSIONS
CERTIFICATIONS
Teledyne LeCroy certifies compliance to the following standards as of the time of publication. Please see the EC Declaration of Conformity document shipped with your product for current certifications.
EMC Compliance
EC DECLARATION OF CONFORMITY – EMC
The instrument meets intent of EC Directive 2014/30/EU for Electromagnetic
Compatibility. Compliance was demonstrated to the following specifications
listed in the Official Journal of the
European Communities:
EN 61326-1:2013, EN 61326-2-1:2013 EMC requirements for electrical equipment
for measurement, control, and laboratory use.1
Electromagnetic Emissions:
EN 55011:2016+A1:2017, Radiated and Conducted Emissions Group 1, Class A 2 3
EN 61000-3-2:2014 Harmonic Current Emissions, Class A
EN 61000-3-3:2013 Voltage Fluctuations and Flickers, Pst = 1
Electromagnetic Immunity:
EN 61000-4-2:2009 Electrostatic Discharge, 4 kV contact, 8 kV air, 4 kV
vertical/horizontal coupling planes 4
EN 61000-4-3:2006+ A2:2010 RF Radiated Electromagnetic Field, 3 V/m, 80-1000
MHz; 3 V/m, 1400 MHz – 2 GHz; 1 V/m, 2 GHz – 2.7 GHz
EN 61000-4-4:2012 Electrical Fast Transient/Burst, 1 kV on power supply lines,
0.5 kV on I/O signal data and control lines 4
EN 61000-4-5:2014+A1:2017 Power Line Surge, 1 kV AC Mains, L-N, L-PE, N-PE4
EN 61000-4-6:2014 RF Conducted Electromagnetic Field, 3 Vrms, 0.15 MHz – 80
MHz
EN 61000-4-11:2004+A1:2017 Mains Dips and Interruptions, 0%/1 cycle, 70%/25
cycles, 0%/250 cycles 4 5
- To ensure compliance with all applicable EMC standards, use highquality shielded interface cables.
- Emissions which exceed the levels required by this standard may occur when the instrument is connected to a test object.
- This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic interference.
- Meets Performance Criteria “B” limits of the respective standard: during the disturbance, product undergoes a temporary degradation or loss of function or performance which is self-recoverable.
- Performance Criteria “C” applied for 70%/25 cycle voltage dips and for 0%/250 cycle voltage interruption test levels per EN61000-4-11.
European Contact:*
Teledyne GmbH, European Division
Im Breitspiel 11c
D-69126 Heidelberg
Germany
Tel: + 49 6221 82700
AUSTRALIA & NEW ZEALAND DECLARATION OF CONFORMITY – EMC
The instrument complies with the EMC provision of the Radio Communications Act
per the following standards, in accordance with requirements imposed by
Australian Communication and Media Authority (ACMA):
AS/NZS CISPR 11:2015 Radiated and Conducted Emissions, Group 1, Class A.
Australia / New Zealand Contacts:*
RS Components Pty Ltd. RS Components Ltd.
Suite 326 The Parade West Units 30 & 31 Warehouse World
Kent Town, South Australia 5067 761 Great South Road Penrose, Auckland, New
Zealand
- Visit teledynelecroy.com/support/contact for the latest contact information.
Safety Compliance
EC DECLARATION OF CONFORMITY – LOW VOLTAGE
The instrument meets intent of EC Directive 2014/35/EU for Product Safety.
Compliance was demonstrated to the following specifications as listed in the
Official Journal of the European Communities:
EN 61010-1:2010 Safety requirements for electrical equipment for measurement,
control, and laboratory use – Part 1: General requirements
EN 61010-2:030:2010 Safety requirements for electrical equipment for
measurement, control, and laboratory use –
Part 2-030: Particular requirements for testing and measuring circuits
The design of the instrument has been verified to conform to the following
limits put forth by these standards:
-
Mains Supply Connector: Overvoltage Category II, instrument intended to be supplied from the building wiring at utilization points (socket outlets and similar).
-
Measuring Circuit Terminals: No rated measurement category.
Terminals not intended to be connected directly to the mains supply. -
Unit: Pollution Degree 2, operating environment where normally only dry, non-conductive pollution occurs. Temporary conductivity caused by condensation should be expected.
Environmental Compliance
END-OF-LIFE HANDLING
The instrument is marked with this symbol to indicate that it complies with
the applicable European Union requirements of Directives 2012/19/EU and
2006/66/EC on Waste Electrical and Electronic Equipment (WEEE) and Batteries.
The instrument is subject to disposal and recycling regulations that vary by country and region. Many countries prohibit the disposal of waste electronic equipment in standard waste receptacles. For more information about proper disposal and recycling of your Teledyne LeCroy product, please visit teledynelecroy.com/recycle.
RESTRICTION OF HAZARDOUS SUBSTANCES (RoHS)
EC DECLARATION OF CONFORMITY – RoHS
Unless otherwise specified, all the materials and processes are compliant with
RoHS Directive 2011/65/EU in its entirety, inclusive of any further amendments
or modifications of said Directive.
CHINA RoHS 2
Unless otherwise specified, all the materials and processes are compliant with
the latest requirements of China RoHS 2. The hazardous substances contained in
the instrument are disclosed in accordance with the standards SJ/T 11364-2014
(Marking for the restricted use of hazardous substances in electronic and
electrical products) and GB/T 26572-2011 (Requirements on concentration limits
for certain restricted substances in electrical and electronic products). The
instrument is marked with an appropriate Environmental Friendly Use Period
(EFUP) symbol. The packaging materials include the appropriate recycling
labels. The below substance disclosure tables (in Chinese and English
languages) provide the required compliance information.
Part Name | Toxic or Hazardous Substances and Elements |
---|---|
Lead (Pb) | Mercury (Hg) |
Polybrominated Biphenyls (PBB) | Polybrominated Diphenyl Ethers (PBDE) |
PCBAs | X |
Mechanical Hardware | O |
Sheet Metal | O |
Plastic Parts | O |
Cable Assemblies | X |
Display | O |
Power Supply | O |
Fans | O |
Batteries | O |
Power Cord | O |
Ext Power Supply (if present) | X |
Probes (if present) | X |
Fuse (if present) | O |
Product Case (if present) | O |
Adapters/Modules (if present) | O |
Mouse (if present) | O |
O: Indicates that this toxic or hazardous substance contained in all of the
homogeneous materials for this part is below the limit requirement specified
in SJ/T11364-2014.
X: Indicates that this toxic or hazardous substance contained in at least one
of the homogenous materials used for this part is above the limit requirement
specified in SJ/T11364-2014.
EFUP (Environmental Friendly Use Period): 30 years.
Use Conditions: Refer to the environmental conditions stated in the User
Manual.
EFUP for Probes: 10 years.
ABOUT TELEDYNE TEST TOOLS
Company Profile
Teledyne LeCroy is a leading provider of oscilloscopes, protocol analyzers and related test and measurement solutions that enable companies across a wide range of industries to design and test electronic devices of all types. Since our founding in 1964, we have focused on creating products that improve productivity by helping engineers resolve design issues faster and more effectively. Oscilloscopes are tools used by designers and engineers to measure and analyze complex electronic signals in order to develop high- performance systems and to validate electronic designs in order to improve time to market.
The Teledyne Test Tools brand extends the Teledyne LeCroy product portfolio with a comprehensive range of test equipment solutions. This new range of products delivers a broad range of quality test solutions that enable engineers to rapidly validate product and design and reduce time-tomarket. Designers, engineers and educators rely on Teledyne Test Tools solutions to meet their most challenging needs for testing, education and electronics validation.
Location and Facilities
Headquartered in Chestnut Ridge, New York, Teledyne Test Tools and Teledyne
LeCroy has sales, service and development subsidiaries in the US and
throughout Europe and Asia. Teledyne Test Tools and Teledyne LeCroy products
are employed across a wide variety of industries, including semiconductor,
computer, consumer electronics, education, military/aerospace,
automotive/industrial, and telecommunications.
© 2020 Teledyne Test Tools is a brand and trademark of Teledyne LeCroy Inc.
All rights reserved. Specifications, prices, availability and delivery subject
to change without notice.
Product brand or brand names are trademarks or requested trademarks of their
respective holders.
References
- Teledyne LeCroy - RoHS and WEEE
- Welcome to the Test Equipment Depot, your test equipment supplier of all the best brands.
- Welcome to the Test Equipment Depot, your test equipment supplier of all the best brands.