tyromotion G6 Therapy Plate Instruction Manual
- June 1, 2024
- tyromotion
Table of Contents
G6 Therapy Plate
“`html
Specifications
-
Type description: Medical device
-
Classification: Type BF
-
Protection against electric shock: Yes
-
Electromagnetic compatibility: Yes
-
Country of origin: Not specified
-
Power supply voltage: 5V DC
-
Supply frequency: 50/60Hz
-
Electricity/Power consumption: 2A / 10W
-
Battery: Polymer-lithium-ion, 3.7V, 470mAh
(non-replaceable) -
Radio transmission frequency: ISM Band
(Bluetooth BLE) -
Radiated transmission power: Not
specified -
Operating type: Measurement range
-
Measurement range: < 4 kg
Product Usage Instructions
1. Introduction
We appreciate your choice of a new product from Tyromotion GmbH.
To fully benefit from the options offered by this product read
these instructions for use and save it for future use.
2. Technology
2.1 Overview
The product is classified as a medical device with Type BF
classification, providing protection against electric shock and
electromagnetic compatibility. It operates on a power supply
voltage of 5V DC and has a battery capacity of 470mAh which is
non-replaceable. The device uses Bluetooth BLE for radio
transmission.
2.2 Area of Application
The product is suitable for interference emission measurements
and complies with HF emissions standards according to CISPR Group 1
11.
3. Clinical Application
Detailed information on clinical application, indications,
contraindications, safety concepts, and residual risks are provided
in the user manual.
4. Hardware Configuration
The hardware configuration involves testing the hardware
components and calibrating the sensors as per the provided
instructions.
5. Service Information
The user manual includes guidelines for monthly functionality
checks, periodic checks, cleaning instructions, repair procedures,
and disposal guidelines.
FAQ (Frequently Asked Questions)
Q: Can the battery be replaced?
A: No, the battery is a polymer-lithium-ion
type and must not be replaced.
Q: What is the operating voltage of the
device?
A: The device operates on a power supply
voltage of 5V DC.
Q: How can I deactivate the integrated Bluetooth
adapter?
A: Detailed instructions for deactivating the
integrated Bluetooth adapter are provided in the user manual.
“`
TYMO® G6
INSTRUCTIONS FOR USE
Issue date of instructions for use 2024-03-13, Rev. 3
Contact information:
Manufacturer:
Tyromotion GmbH Bahnhofgürtel 59 8020 Graz AUSTRIA
PHONE +43 316 908 909
FAX
+43 316 231123 9144
EMAIL office@tyromotion.com
WEB www.tyromotion.com
SUPPORT support@tyromotion.com
Distributor (Germany, Switzerland, USA):
Tyromotion GmbH Bahnhofgürtel 59 8020 Graz AUSTRIA
PHONE +43 316 908 909
FAX
+43 316 231123 9144
EMAIL office@tyromotion.com
WEB www.tyromotion.com
SUPPORT support@tyromotion.com
Swiss Authorized Representative
CMC Medical Devices GmbH Rigistrasse 3 CH-6300 Zug SWITZERLAND
UK Responsible Person
CMC Medical Devices Ltd. Office 32 19-21 Crawford Street London, W1H 1PJ
UNITED KINGDOM
Declaration of Conformity This product conforms to the requirements of the
Medical Devices Regulation (2017/745). The CE mark must be removed when
rebuilding the product or when using other than original TYMO® accessories.
EN-1
Table of contents
1 Introduction
7
1.1 Training concept
7
1.2 Symbols
7
1.2.1 Symbols in the instructions for use
7
1.2.2 Symbols on TYMO®
7
1.3 Content
9
1.4 Intended use
9
1.5 Warranty and legal disclaimer
9
2 Technology
11
2.1 Overview
11
2.2 Area of application
12
3 Clinical application
17
3.1 Indications/Contraindications
17
3.2 Safety
19
3.2.1 Safety concept
19
3.2.2 Residual risk
20
4 TYMO®
20
4.1 First steps
20
4.1.1 Charging TYMO®
20
4.1.2 Installation of the Software
21
4.1.3 Installation of the Bluetooth adapter
21
4.1.4 Connecting TYMO® to the PC
21
4.1.5 Deactivate the integrated Bluetooth adapter
22
4.2 Use of TYMO®
22
4.2.1 TYMO® Application
22
4.2.2 TYMO® Rolling Element 1D and 2D
24
4.2.2.1 TYMO® Rolling Element 1D
24
4.2.2.2 TYMO® Rolling Element 2D
25
4.2.3 LED Status TYMO®
26
4.2.4 Battery
26
5 Hardware Configuration
27
5.1 Hardware test
27
5.2 Calibrating the sensors
27
6 Service information
28
6.1 Monthly functionality check
28
EN-2
6.2 Periodic check
29
6.3 Lifetime
29
6.4 Cleaning instructions
29
6.4.1 Disinfection
30
6.4.2 Cleaning process
30
6.5 Repair
30
6.6 Disposal
30
EN-3
Figure 1
< 180 kg < 396.8 lbs
Figure 2
< 180 kg < 396.8 lbs
Figure 3
Figure 4
Figure 5
Figure 6
EN-4
Figure 7 Figure 8 Figure 9
EN-5
Figure 10 EN-6
1 Introduction
We appreciate your choice of a new product from Tyromotion GmbH. To fully
benefit from the options offered by this product read these instructions for
use and save it for future use.
1.1 Training concept
TYMO® is a complex technical device. Users of TYMO® are required to complete a
training course and read the user manual in order to ensure the safety of
patients, users and the device itself. Merely reading the present manual does
not convey sufficient competence for operating TYMO®. Prospective users are
also required to have basic medical training (e.g. physiotherapy/ergotherapy).
Tyromotion GmbH rejects all liability for damages resulting from therapy that
was performed by an untrained user. Prospective users are trained after
delivery of the TYMO®. Users can perform initial and repeated therapies
training with TYMO®. Users are not permitted to instruct other persons in the
usage of TYMO®. Users are trained by a member of Tyromotion GmbH or by another
trainer delegated by Tyromotion GmbH.
1.2 Symbols
1.2.1 Symbols in the instructions for use
Warnings: This symbol indicates a possible risk of injury to your own health
or to the health of others. Be especially mindful of these notices!
1.2.2 Symbols on TYMO®
Follow the instructions for use
Applied part, type BF
EN-7
Do not discard with household waste
CE mark Information about the manufacturer of TYMO®, including the
manufacturer’s full mailing address is displayed next to the factory symbol.
Date of manufacture
Serial number of the device
Medical device The UKCA mark is a product label used for medical devices
marketed in the United Kingdom (England, Wales, Scotland) and shows the
conformity. Shows CH Authorized Representative
On-switch for TYMO® Therapy plate
EN-8
1.3 Content
TYMO® consists of the following components, included in the delivery: · 1x
TYMO® Therapy plate · 1x TYMO® Rolling Element 1D · 1x TYMO® Rolling Element
2D · 1x TYMO® Multipad X · 1x TYMO® power supply · 1x USB charging cable · 1x
USB extension · 1x Bluetooth adapter · 1x USB-Stick (TyroS Software)
1.4 Intended purpose
TYMO® is a monitoring and therapy device for the rehabilitation of patients
suffering from motoric dysfunctions; it is primarily used for neurological
rehabilitation purposes.
The target group not only comprises neurological, but also orthopedic and
pediatric patients with dysfunctional motion- and force control, accuracy,
coordination, body control and balance. Depending on national variances, TYMO®
is typically used in ergotherapy and/or physiotherapy as therapeutic support,
enhancement and intensification in addition to conventional therapy forms.
1.5 Warranty and legal disclaimer
Tyromotion GmbH issues a warranty to the original medical device purchaser
that it shall be free of material and qualitative processing defects for a
period of 12 months under normal usage conditions from the date of
installation on the owner’s premises and that the product complies with the
mechanical and electrical specifications published by Tyromotion (unless the
warranty term is extended by an optional service contract). This warranty is
granted under the provision that the medical device is installed, operated,
and maintained in accordance with the user manual. The customer must submit
all warranty claims to Tyromotion in written form within 60 days of the
occurrence of the problem and before the expiry of the warranty. Tyromotion is
exclusively obligated to repair, exchange or correct faulty or non-compliant
parts at its own discretion in accordance with the warranty. Tyromotion has no
further obligations to the owner in regard to these parts after the repair or
exchange of faulty or non-compliant parts. All repairs or maintenance work
must be performed by an authorized Tyromotion service representative in
accordance with this warranty. The above-mentioned warranty
EN-9
becomes null and void if repairs, maintenance, or other work is performed by
third parties. Moreover, problems resulting from accidents, improper use,
incorrect application, storage damage, negligence as well as device or
component modifications are excluded from the warranty. The above-mentioned
warranty is granted in place of all other warranties, rights or conditions,
and the product is delivered “without deficiency warranty” apart from the
limited warranty. Tyromotion and its third-party suppliers specifically and
unreservedly reject all other explicit or implicit warranties held by the
owner, his personnel and patients, customers, users and any third parties,
unreservedly including all warranties for marketability, applicability for a
specific purpose, non-infringement and any warranties resulting from
performance development, business trends or commercial customs. Tyromotion and
its third-party suppliers do not provide declarations or warranties for
product compliance with the owner’s requirements or for functionality without
interruption, errors, or deficiencies. Tyromotion is in no way liable for
indirect, incidental, specific or consequential damage or for punitive damage
compensation including, among other things, the loss or absence of profits,
yield, goodwill or usage, which the owner or third parties may incur or for
damage to connected equipment, costs for replacement products, installations,
servicing, exchange elements or idle time or for claims from patients,
customers, visitors, the owner’s employees or other persons, regardless
whether submitted within the context of a contractual claim, due to
unauthorized behavior, strict liability or imposed by law or otherwise even
when Tyromotion has been informed about the possibility of such damages.
Tyromotion ‘s liability for damages resulting from or in connection with this
contract may not in any event exceed the purchasing price of the product. Some
jurisdictions limit or exclude the extent of restrictions, the exclusion of
legal means, compensation, or liability, such as liability for gross
negligence or willful misconduct according to or in the abovementioned extent
or do not permit the exclusion of implicit warranties. In such jurisdictions,
the restriction or exclusion of warranties, legal means, compensations, or
liabilities described above may not be valid for the owner. Such restrictions
or exclusions apply according to the highest legally permitted extent even if
they are not valid according to the legally prohibited extent. The owner may
also have other rights that vary depending on the specific country or other
jurisdictions.
EN-10
2 Technology
2.1 Overview
Type description:
TYMO®
Classification:
TYMO® is an active, therapeutic class I medical device according to rule 13 of Medical Device Regulation (EU) 2017/745.
Type of applied part:
Type BF
Protection against electric shock:
Electromagnetic compatibility:
Internally powered medical device
Class B device (CISPR 11) TYMO® is suitable for usage in all establishments
including residential areas and areas that are directly connected to the
PUBLIC SUPPLY GRID, which also supplies residential buildings.
Country of origin: Power supply voltage:
AUSTRIA 100 240V alternating current
Supply frequency:
50/60Hz
Electricity/Power consumption:
Battery:
5V DC / 2A / 10W
Polymer-lithium-ion battery, 3.7V, 470mAh. The battery must not be replaced.
Radio transmission frequency: Radiated transmission power:
Operating type:
Measurement range:
ISM Band (Bluetooth BLE)
Max. 1.79mW Operational area: 10 meters given uninterrupted view between TYMO®
plate and PC Continuous operation
force: 0 1765,8N (0-180 kg) angle : ±90°
Measurement deviation: Weight: Device Requirements:
force: ± 2% of actual value ± 200g angle: ± 3°
< 4 kg
· Windows 10; Home or Professional · Intel or AMD Processor with >= 3 GHz · at
least 8 GB RAM · Display resolution 1280×768 · Microsoft .NET framework 4.5.2
(installed by default) · Unity3D Webplayer Plugin (installed by default) · One
free USB port
EN-11
2.2 Area of application
The product is for indoor use. TYMO® is classified as a medical electronic
device and therefore subject to specific precautionary measures relating to
electromagnetic compatibility (EMC). It is absolutely imperative to observe
the stated indications for EMC. Portable and mobile HF communication devices
may affect the TYMO®.
EN-12
Guidelines and MANUFACTURER`s declaration ELECTROMAGNETIC EMISSIONS
TYMO® is designed for operation in an ELECTROMAGNETIC ENVIRONMENT as indicated
below. The customer or user of TYMO® must ensure that it is used in such an
environment.
Interference emission measurements
Agreement
HF emissions according to CISPR Group 1 11
ELECTROMAGNETIC ENVIRONMENT
Guidelines
TYMO® exclusively uses HF energy for its internal FUNCTIONS. HF emissions are
very low and unlikely to disrupt electronic devices within range.
HF emissions according to CISPR 11
Class B
Harmonics emissions according Class A to IEC 61000-3-2
Emissions of voltage fluctuations/flicker according to IEC 61000-3-3
Not applicable
TYMO® is suitable for usage in all establishments including residential areas and areas that are directly connected to the PUBLIC SUPPLY GRID, which also supplies residential buildings.
Table 1: Guidelines and manufacturer`s declaration Electromagnetic emissions
Guidelines and MANUFACTURER`s declaration ELECTROMAGNETIC IMMUNITY
TYMO® is designed for operation in an ELECTROMAGNETIC ENVIRONMENT as indicated
below. The customer or user of TYMO® must ensure that it is used in such an
environment.
Immunity Test
Electrostatic Discharge (ESD) IEC 61000-4-2
IEC 60601-Test Level
Compliance Level
± 8 kV contact ± 15 kV air
± 8 kV contact ± 15 kV air
ELECTROMAGNETIC ENVIRONMENT – Guidance
Floors should be wood, concrete, or ceramic tile. If floors are covered with
synthetic material, relative humidity should be at least 30%.
EN-13
Electrical Fast Transient/Burst IEC 61000-4-4
Surge IEC 61000-4-5
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
± 2 kV for power supply lines
Not applicable
Mains power quality should be that of a typical commercial or hospital environment.
± 1 kV differential mode
Not applicable
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions, and voltage variations on power supply input lines IEC 61000-4-11
< 5 % UT (> 95 % dip in UT) für 0.5 cycles
70 % UT (30 % dip in UT) for 25 cycles
< 5 % UT (> 95 dip in UT) for 0.5 cycles
70 % UT (30 % dip in UT) for 25 cycles
Mains power quality should be that of a typical commercial or hospital environment.
< 5 % UT (> 95 % dip in UT) for 5 s
Power Frequency
30 A/m
(50/60 Hz) Magnetic Field
IEC 61000-4-8
< 5 % UT (> 95 % dip in UT) for 5 s
30 A/m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
NOTE: UT is the AC mains voltage before application of the test level. Table 2: Guidelines and Manufacturer`s declaration Electromagnetic Immunity
Guidance and Manufacturer’s Declaration– ELECTROMAGNETIC IMMUNITY
TYMO® is designed for operation in an ELECTROMAGNETIC ENVIRONMENT as indicated
below. The customer or user of TYMO® must ensure that it is used in such an
environment.
Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Immunity Test
IEC 60601-Test Level
Compliance Level
EN-14
ELECTROMAGNETIC ENVIRONMENT – Guidance
Recommended Separation Distance:
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
6 Vrms
6 V
150 kHz to 80 MHz
10 V/m
80 MHz to 2.7 GHz
10 V/m
= 0.58
= 0.35 80 MHz to 800 MHz
= 0.7 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site surveya, should be less than the compliance level in each
frequency range.b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTES: – At 80 MHz and 800 MHz, the higher frequency range applies. – These
guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
a Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which TYMO® is used exceeds the
applicable RF compliance level above, the device should be observed to verify
normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating TYMO®. b Over the frequency
range 150 kHz to 80 MHz, field strengths should be less than [4] V/m.
Table 3: Guidelines and Manufacturer`s declaration Electromagnetic Immunity
EN-15
Recommended separation distances between portable and mobile RF communications equipment and TYMO®
TYMO® is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. Users of this device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communication equipment (transmitters) and the device as recommended below, according to maximum output power of the communications equipment
Rated Maximum Output Power of Transmitter W
Separation Distance According to Frequency of Transmitter
150 kHz to 80 MHz = 0.58
80 MHz to 800 MHz
= 0.35
800 MHz to 2.5 GHz = 0.7
0.01
0.058
0.035
0.07
0.1
0.18
0.11
0.22
1
0.58
0.35
0.70
10
1.83
1.11
2.21
100
5.80
3.50
7.00
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTES: – At 80 MHz and 800 MHz, the higher frequency range applies. – These
guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
Table 4: Recommended separation distances between portable and mobile RF
communications
equipment and TYMO®
EN-16
FCC Statement Please take attention that changes or modification not expressly
approved by the party responsible for compliance could void the user’s
authority to operate the equipment.
This device complies with Part 15 of the FCC Rules. Operation is subject to
the following two conditions: (1) This device may not cause harmful
interference, and (2) This device must accept any interference received,
including interference that may cause undesired operation.
3 Clinical application
3.1 Indications/Contraindications
TYMO® is mainly used in neurologic rehabilitation. The target population
includes not only neurologic but also orthopedic and pediatric patients with
deficits in balance and postural control. Typical indications are frequently
linked with problems in weight distribution, weight bearing ability, weight
shifting, and symmetry. Application is possible supporting with the upper
extremity on the TYMO® Therapy plate as well as sitting or standing on the
TYMO®. As in the case of every other therapy, the doctor in charge is
responsible to make medical diagnosis and decide for the type of intervention.
In principle, the same indications and contraindications apply for TYMO®
therapy as those for manually applied therapeutic treatment. Knowledge of the
contraindications is essential in order not to put the patient at risk. Before
applying TYMO® therapy to a patient, check carefully if one or more
contraindications exist. Also, be aware that your patient may have additional
indications and/or contraindications that have not been listed here but may be
relevant. The following listings have no claim to completeness.
Common Indications:
· Stroke (cerebral hemorrhages, ischemic damages) · Traumatic brain injury
(TBI) · Spinal cord injury (SCI) · Brain tumor · Parkinson’s disease · Chronic
diseases, e.g., multiple sclerosis (MS) · Cerebral palsy (CP)
EN-17
· Motor neuron diseases, e.g., amyotrophic lateral sclerosis (ALS) · Muscular
dystrophies · Paralysis due to a herniated vertebral disc · Orthopedic events
such as e.g., amputation, joint surgery, or joint replacement
Absolute Contraindications: The device must not be used!
· Acute pain despite conventional pain therapy · Adjustment and patient
position: Do not carry out training with TYMO® if the
adjustment to the patient’s individually physiologic position is not possible,
especially in case of contractures or severe spasticity (joint is fixed/rigid)
of the trained body region · Insufficient compliance, e.g., children, patients
suffering from severe psychotic diseases or severe neurotic disorders · High
grade ataxia · Severe osteoporosis: risk of fractures · Fractures: Do not
carry out training with unstable or still inadequately consolidated fractures
Relative Contraindications:
Each patient has to be conscientiously assessed by the doctor/therapist in
charge individually to determine if TYMO® therapy is suitable for the patient
in case of:
· Apraxia · Arthritis · Reduced compliance: e.g., children, patients with
cognitive impairments · Consolidated fractures · Epilepsy · Heart pacemakers
and similar devices/implants: Pacemakers can react
differently to external influences. Therefore, knowledge about possible
dangerous influences relevant for each specific device is essential. Inform
patients that magnets are built into the charging connection. The TYMO®
therapy device does not influence heart pacemakers if the distance between
pacemaker and device (or pacemaker and magnets) is not less than 15 cm. ·
Infections · Joint problems: Repetitive training may cause pain and irritation
in case of weak joints. · Neglect · Osteoporosis · Orthostatic circulatory
problems: increased risk of falling. · Pain, e.g., complex regional pain
syndrome (CRPS)
EN-18
· Sensory disorders: Patients with sensory impairment cannot report
potentially occurring pain. Therefore, the therapist must be especially
attentive in such cases.
· Skin problems: Before and after every training carefully check for any skin
problems, existing wounds, pressure marks, and/or skin ulceration, in
particular of body regions in contact with the device.
Please be aware that your patients may have other contraindications that are
not listed here. In the case of questions or feedback, please contact
Tyromotion GmbH (for contact data see start of the document).
3.2 Safety
3.2.1 Safety concept
TYMO® is an active, therapeutic device that is above all to be used with
patients with restricted physical and/or mental abilities. To prevent problems
with the use of the device the following instructions for proper use are also
to be adhered to.
Users are obligated to familiarize themselves with these safety instructions and avoid conditions that may lead to injuries or damage. Users are also required to have basic medical training (e.g. physiotherapy/occupational therapy).
Users ·
·
· · ·
·
Every user must have read the instructions for use prior to using TYMO®.
Tyromotion GmbH rejects any liability for damages to persons or material if
safety regulations and instructions relevant to the usage of TYMO® are not
observed!
The user must assess whether and for how long the patient should perform
therapeutic work independently before starting the therapy. Cognitive
abilities and the general condition of the patient’s health must be taken into
consideration. Only use original accessories from the manufacturer that are
supplied in the product contents. The user must take appropriate measures to
stabilize the user during therapy. If there is any doubt as to the continued
safe use of TYMO® or if any parts should fail or be defective, stop using the
product immediately and contact Tyromotion GmbH or your local dealer. The
cleaning instructions stated in these instructions for use must be strictly
observed.
EN-19
Environment · Only use TYMO® on a firm surface.
Product · No modifications of TYMO® are allowed. · Always contact the
manufacturer for repairs!
Please also observe the instructions for use of the TyroS software.
3.2.2 Residual risk
An unpredictable residual risk remains for therapy despite all safety
precautions. In rare cases, the patient may experience minor pinching or
crushing injuries even during proper operation. However, the probability of
such injuries is very low, and the injuries should not be severe as long as
all safety instructions in the present instructions for use are observed.
Tyromotion GmbH can provide a detailed risk analysis upon request.
If you think that you or someone who used TYMO® has experienced an injury due
to the use of TYMO® please report this to the manufacturer and to your
national health authority.
4 TYMO®
4.1 First steps
The time required for TYMO® to warm/cool from the min./max. storage
temperature between uses until the device is ready for its intended use at
ambient temperature is one hour. Please wait for this period before operating
the unit.
4.1.1 Charging TYMO®
The TYMO® Therapy plate is equipped with a rechargeable battery. Before using
TYMO® for the first time, fully charge the device using the supplied USB cable
and power supply unit (Figure 6). The charging time is 6 hours when the
battery is completely empty.
EN-20
The use of the supplied medical power supply unit for charging the TYMO®
Therapy plate is mandatory. No other power supply units may be used.
Power supply: Adapter Technology, model: ATM012T-W050VU
4.1.2 Installation of the Software
You will find a detailed installation guide in the instructions for use of the
TyroS Software. For this version of TYMO, TyroS version 6.3 or higher can be
used and is compatible.
4.1.3 Installation of the Bluetooth adapter
TYMO® is connected to your PC with Bluetooth technology. To produce this
connection, it is first necessary to connect the Bluetooth adapter included in
the scope of delivery to your PC and to install it. If you have purchased a PC
from Tyromotion this installation will already have been carried out.
Otherwise, connect the Bluetooth adapter to the USB extension and then to a
free USB port of your PC and follow the installation instructions of your PC.
Figure 11: Bluetooth adapter
Figure 12: Bluetooth adapter incl. USB extension
4.1.4 Connecting TYMO® to the PC
Important points before the installation: · The TYMO® Therapy plate must be
fully charged before being used for the first time. · If you previously used
another Bluetooth adapter with your computer, be sure to remove any drivers or
software for this adapter before installing the pluggable adapter. Built-in
Bluetooth adapters must be deactivated (see chapter 4.1.5 and the Quickstart
Guide). · The TYMO® Therapy plate must not be more than 10 meters away from
the PC.
EN-21
· The TYMO® Therapy plate must be turned on. · Installation is completed and
the TYMO® is now ready for use.
4.1.5 Deactivate the integrated Bluetooth adapter
To ensure that TYMO® works correctly with the provided Bluetooth adapter it
might be useful to deactivate the integrated Bluetooth adapter (if available)
in your PC.
If you have purchased a PC from Tyromotion this deactivation has already been
carried out. To deactivate the integrated Bluetooth adapter, go to the control
panel and choose Device Manager. Look for the Bluetooth- (or Bluetooth Radios)
entry. Right-Click on the integrated Bluetooth adapter and choose “Disable”.
4.2 Use of TYMO®
TYMO® evaluates and trains the musculoskeletal system and is specifically used
in therapy of impaired balance and asymmetry of patients. By definition it is
used to display trends. Hence the software consists of integrated measuring
modules and interactive therapy modules with audiovisual feedback.
4.2.1 TYMO® Application
To use TYMO®, carry out the following steps:
· TYMO® can be activated after it has been properly installed (chapter 4.1)
and connected.
· Starting therapy: 1) Switch on the TYMO® Therapy plate by pushing the LED
button for approx. 3 seconds (Figure 5). The LED button should now start
blinking green. 2) Start the TyroS Software 3) The TYMO® Therapy plate
connects itself automatically to your PC via Bluetooth technology as soon as
you start a therapy session with TYMO®. Please note the LED display at the
TYMO® Therapy plate (chapter 4.2.3). This gives information on which status
the TYMO® plate are currently in.
· Performing therapy: 1) Apply now the TYMO® and follow the instructions given
by the TyroS Software. 2) If errors occur during the therapy, they are handled
in the TyroS Software.
EN-22
· Completing therapy: 1) Terminate the application by exiting the TyroS
Software and shutting down the PC. 2) Switch of the TYMO® Therapy plate by
pushing the LED button for approx. 5 seconds (Figure 5). The LED color will
then change to solid violet. Press the button again for two seconds to switch
off the TYMO®. If the TYMO® Therapy plate is disconnected from the software
for more than an hour, it will switch off automatically. 3) Make sure all
TYMO® components are securely stowed after use so they cannot fall to the
floor and thus cause harm to patients.
Please also observe the instructions for use of the TyroS Software.
Despite its robust and high-quality construction, the TYMO® Therapy plate only
has a limited ability to withstand shocks. If the device falls from table onto
the floor, defects can occur in the sensors or the battery. These defects may
possibly not become recognizable until sometime after a fall. Therefore, do
not use the TYMO® Therapy plate after a fall; instead, immediately contact
Tyromotion GmbH or the dealer from whom you bought the TYMO® Therapy plate.
Some application possibilities of TYMO® are shown below.
Figure 13: Static application while sitting Figure 14: Static application for training
(optional with Multipad X on TYMO®)
the movement transition sit to stand.
EN-23
Figure 15: Static or dynamic application with 1D or 2D Rolling Element while
standing.
4.2.2 TYMO® Rolling Element 1D and 2D
The Rolling Elements 1D and 2D are used in the ,,dynamic” mode. The Rolling
Elements are attached to the underside of TYMO® where they are fixed by
integrated magnets (Figure 7, Figure 8 and Figure 9)
Note: As long as there is a Rolling Element attached to the underside of
TYMO®, a static measurement (force measurement) cannot be performed.
Please note that the teeth of the Rolling Element are positioned exactly as
shown in Figure 7, Figure 8 and Figure 9. Otherwise, the Rolling Element could
detach itself from the TYMO® Therapy plate.
Please check the Rolling Elements for mechanical damage before each use. Do
not use the Rolling Elements if there is any damage and contact Tyromotion
GmbH or the dealer from whom you bought the TYMO® Therapy plate.
4.2.2.1 TYMO® Rolling Element 1D
The Rolling Element 1D enables TYMO® to tilt in one direction and prevents a
tilting movement in another direction. The following tilting movements are
possible:
· Forwards/Backwards · Left/Right
EN-24
Depending on the direction in which the TYMO® should be tilted, the Rolling Element 1D must be properly mounted on the underside of the TYMO® Therapy plate.
Figure 16: 1D Mounting direction for the tilt movement left/right
Figure 17: 1D Mounting direction for the tilt movement forwards/backwards
Do not try to tilt TYMO® Therapy plate in the direction which is prevented by
the Rolling Element 1D. This could cause the Rolling Element 1D to detach
itself from the TYMO® Therapy plate and an increased risk of falling is given.
4.2.2.2 TYMO® Rolling Element 2D
The Rolling Element 2D enables a simultaneous tilting movement of TYMO® in
both tilting directions. This makes it possible to perform additional
therapeutic applications.
Figure 18: Mounting 2D Rolling element EN-25
4.2.3 LED Status TYMO®
Observe the LED display on the TYMO® Therapy plate. This provides information
on the status of the TYMO® Therapy plate.
LED-Display Shining ORANGE Flashing ORANGE Shining GREEN Flashing GREEN
Meaning
Necessary Action
Connection with software okay, battery Charging of the battery
level under 15%
(section 4.2.4)
No connection with software, battery Charging of the battery
level under 15%
(section 4.2.4)
Connection with software okay, battery level okay
None
No connection with software, battery
level okay
None
Shining BLUE
Battery is fully charged
None
Flashing BLUE Off
Battery is being charged
Battery of the device is completely empty, or the device is switched off
Table 5: LEDs
None
Charging of the battery or switch on the device (section 4.2.4)
4.2.4 Battery Charging battery:
The integrated battery in the TYMO® Therapy plate can be charged via the TYMO®
power supply included in the procuct content (Figure 6). The TYMO® Therapy
plate switches into the charging mode automatically after connecting the TYMO®
to the power supply.
NOTE: The TYMO® Therapy plate may be used in charging mode.
Duration of Charging:
The duration of charging depends, among other things, upon whether the battery
was fully uncharged at the start of the charging process or not. However, the
charging process should not last longer than 6 hours.
EN-26
Charging Cycles and Lifespan:
Due to the use of high-quality lithium-polymer batteries, it is ensured that
the battery running times still meet the requirements even after many charging
cycles. Nevertheless, no guarantee can be given for the lifespan of batteries
going beyond 12 months after the date of purchase.
If you adhere to the following advice, you can certainly save the battery of
your device and thereby extend its lifespan:
· Do not expose cells or batteries to heat or fire. Avoid direct exposure on
TYMO® to sunlight in order to prevent the battery inside the device from
becoming too hot.
· Use exclusively the power supply provided by Tyromotion GmbH for the
recharging of the battery.
· Do not dismantle, open or shred secondary cells or batteries. · Do not
subject cells or batteries to mechanical shock. · In the event of a cell
leaking, do not allow the liquid to come in contact with
the skin or eyes. If contact has been made, wash the affected area with
copious amounts of water and seek medical advice. · Do not use any cell or
battery which is not designed for use with the equipment.
Battery life:
In use, the battery operating time is minimum 10 hours when fully charged.
5 Hardware Configuration
5.1 Hardware test
The TYMO® hardware test is an integrated diagnostic program that automatically
checks the plausibility of sensor values. It is not necessary to perform this
test in normal operation; it is merely a tool for the manufacturer to easily
prepare a diagnosis.
5.2 Calibrating the sensors
Due to the sensors chosen and the high-quality construction, the TYMO® Therapy
Plate is very robust against long-term effects. nevertheless, it may be
necessary to carry out a calibration.
Proceed with the calibration of the TYMO® Therapy Plate as follows:
EN-27
1. Make sure that the battery is sufficiently charged (for more information
refer to the TYMO® instruction manual).
2. Remove any TYMO® rolling elements attached to the TYMO® therapy plate. 3.
Place the TYMO® therapy plate on an even and stable surface. 4. Ensure that
nothing lays upon the therapy plate during the calibration process and
that nobody stands or presses upon it. 5. Press the button ,,Calibrate
Sensors” in the hardware configuration area in order to
start the calibration. 6. Wait until the progress display is at 100% and the
green instruction window appears
with the text, ,,Calibration OK”.
If the calibration has not run correctly, you will receive the warning message
,,Calibration failed”. In this case, make sure that the TYMO® therapy plate is
ready for use and carry out the above-mentioned points again.
6 Service information
Maintenance of a medical device is wholly the responsibility of the owner of
that device. Failure to maintain a device in accordance with the instructions
for use may invalidate the device’s warranty. Furthermore, failure to maintain
a device may compromise the clinical condition or safety of users.
6.1 Monthly functionality check
The functionality checks described here must be performed monthly. Perform the
checks even if the TYMO® indicates a malfunction (e.g. in case of unusual
sounds, elementary damage, etc.). The person responsible for checking the
device must be trained in handling and operating the TYMO®.
Inspection: Protective covers Rolling Elements 1D and 2D Externally visible deformations
Malfunction:
· Covers shake · Covers missing · Covers damaged
· Rolling elements damaged
or broken
· Parts bent out of shape · Parts asymmetrical · Parts defective
Resulting measure:
· Further use is prohibited · Contact Tyromotion GmbH
· Further use is prohibited · Contact Tyromotion GmbH
· Further use is prohibited · Contact Tyromotion GmbH
EN-28
Cleaning Battery
·
· TYMO® contaminated
·
· Battery life drops below 2h · housing deformations due ·
to an inflated battery occur · ·
Table 6: Inspection points
Further use is prohibited. Clean the contaminated parts as described in
chapter 6.4
Further use is prohibited Contact Tyromotion GmbH
6.2 Periodic check
Periodic checks differ from the checks described in chapter 6.1 as the
legislator may demand the check described here while the checks in chapter 6.1
are intended, among other things, to detect acute damage or wear parts that
require replacement. The operating company of the device itself is responsible
for carrying out both tests.
An interval of one year is defined by Tyromotion GmbH for carrying out the
periodic tests. Periodic tests may only be carried out by service
representatives. The operating company of the device must ensure that the test
intervals imposed for the periodic test are complied with. TYMO® must not be
used if the test intervals have not been complied with.
The periodic test must be carried out according to EN 62353:2014.
6.3 Lifetime
The lifetime of this product in normal use is 7 years if all maintenance and
servicing is carried out in accordance with the manufacturer’s instructions
and demonstrably recorded.
6.4 Cleaning instructions
The following instructions apply to the manual cleaning of medical devices by
Tyromotion GmbH.
Thorough cleaning and wiping are essential for the first time and reuse of
reusable medical devices. Effective cleaning must be performed to achieve
adequate decontamination. The goal of cleaning is to remove any visibly sticky
soil and reduce the number of particles and microorganisms. Cleaning must be
carried out in a manner that minimizes the risks posed by pathogens.
EN-29
The devices of Tyromotion GmbH must be cleaned and disinfected after delivery
before the first and any further use on the patient.
6.4.1 Disinfection
The product can be disinfected with a 70% disinfectant IPA solution. It is
recommended to wipe off any residue and dirt from the product, using a cloth
with warm water and a mild detergent/soap without chlorine and let it dry
before disinfection.
6.4.2 Cleaning process
To minimize the risk of germ transmission, all surfaces that are touched by
both the patient and the user should be periodically cleaned and disinfected
and at least before any further use on the patient.
1. If the patient perspires heavily during use, dry the TYMO® components
after use before disinfecting them.
2. Moisten the disposable cloth, according to the product information
leaflet, only slightly with disinfectant. Wipe the TYMO® components with the
clean, soft and lint-free cloth. Observe the contact time of the disinfectant
used according to the label and product information.
3. Depending on the disinfectant, it may be necessary to wipe the disinfected
area with water after the exposure time.
4. Dry the area wiped with water with a clean, non-abrasive, soft, lint-free
cloth. 5. Always store the TYMO® components in clean and dry rooms or
facilities after
use.
6.5 Repair
Always contact the manufacturer for repairs!
6.6 Disposal
TYMO® must not be disposed of as household waste according to the Directive
2012/19/EU on waste electrical and electronic equipment (WEEE-RL) and the
respective national legislation. The product must be disposed of at the
intended collection point or at a collection point approved for the recycling
of waste electrical and electronic equipment. It can also be returned to
Tyromotion GmbH.
EN-30