VOLTCRAFT 1274960 IR-1600 CAM Infrared Thermometer Instruction Manual
- June 9, 2024
- VOLTCRAFT
Table of Contents
- VOLTCRAFT 1274960 IR-1600 CAM Infrared Thermometer
- Product Information
- Intended use
- Product Usage Instructions
- Delivery content
- Safety instructions
- Operating elements
- Product Description
- Commissioning
- Measuring
- Care and cleaning
- Troubleshooting
- 12. Disposal
- Technical data
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
VOLTCRAFT 1274960 IR-1600 CAM Infrared Thermometer
Product Information
The Infrared Thermometer IR-1600 CAM (Item No. 1274960) is a portable device that measures temperature through infrared radiation. The thermometer has an optical sighting device, a color display, and a camera lens. It also has a laser target pointer and a mini-USB charging port. The device is equipped with a rechargeable Li-ion battery type 18500 and can be used to measure surface temperatures of various objects.
Intended use
The infrared thermometer enables contract-free measurement of surface temperatures. It determines temperature according to the radiated infrared energy that is emitted by every object. Contact-free measurement makes it ideal for dangerous, difficult-to-access, moving or live objects. It cannot measure through transparent media such as glass, plastics, water, etc. Instead, the surface temperature is measured at all times. The temperature measuring range is from -50 to +1600 °C. A double laser target device facilitates the determination of the measuring range.
Additionally, a contact measurement with a thermal element sensor type K is
possible in the measuring range from -50 to +1370 °C, depending on the sensor
used. This contact sensor must only be used on voltage-free media.
The installed sensors can be used to measure ambient temperature and relative
humidity. These measured values can be used to calculate and display the dew
point and the evaporation temperature (wet bulb). There is a mould warning
display.
An integrated camera permits recording images and videos with the measured
data displayed there. The data can be recorded in a microSD memory card
(supplied). A data logger with curve memory is present.
The IR-thermometer as such must not come directly into contact with the
measured contact and surface temperature. A sufficient safety distance and
admissible ambiance conditions must be complied with under all circumstances.
Diagnostic application in the medical area is not admissible.
The degree of emissions can be adjusted to the present surface property of the
measured object. A table of common materials is integrated.
The installed USB interface serves only as a charging interface. Data
transmission is not possible.
A 3.7 V Li-ion battery of type 18500 with a connection plug is used for
voltage supply. No energy supply other than the one indicated can be used.
Charging is performed via the integrated USB charging interface.
The enclosed USB plug-in mains adapter is designed in protection class 2 and
must only be used in dry inner rooms.
For safety and approval purposes, you must not rebuild and/or modify this
product. If you use the product for purposes other than those described above,
the product may be damaged. In addition, improper use can result in short
circuits, fires, electric shocks or other hazards. Read the instructions
carefully and store them in a safe place. Make this prod-uct available to
third parties only together with its operating instructions.
All company names and product names are trademarks of their respective owners.
All rights reserved.
Product Usage Instructions
- Insert the Li-ion battery into the device. A suitable replacement battery can be ordered using the Best.-Nr. 1233684.
- To take measurements, point the device towards the object being measured and press the measurement button. The temperature reading will be displayed on the color screen.
- The laser target pointer can be used to assist in aiming the thermometer. However, never look directly into the laser beam or point it at people or animals.
- The device can be set to display temperature readings in either Celsius or Fahrenheit.
- The thermometer can also be used to measure dew point and wet bulb temperature.
- The device has a mini-USB charging port for recharging the battery. To charge the device, connect it to a power source using a mini-USB cable.
- The thermometer has a camera lens that can be used to capture images of the object being measured. The captured images can be saved to a MicroSD card (not included).
- The device has various settings that can be adjusted, including language, date/time format, and unit of measurement.
- When not in use, turn off the device to conserve battery life.
- Refer to the user manual for troubleshooting tips and proper disposal instructions.
Delivery content
- IR-Thermometer IR-1600 CAM
- Li-ion battery (type 18500)
- Temperature sensor thermocouple type K (measurement range -20 to +250 °C)
- Table stand
- USB cable (Mini-USB to USB)
- USB power adapter
- MicroSD card 8GB
- Transport case
- Operating instructions
Up-to-date operating instructions
Download the latest operating instructions at
www.conrad.com/downloads or scan the QR
code shown. Follow the instructions on the website.
Safety instructions
Read the operating instructions carefully and especially observe the safety information. If you do not follow the safety instructions and information on proper handling in this manual, we assume no liability for any resulting personal injury or damage to property. Such cases will invalidate the warranty/guarantee.
General information
- The device is not a toy. Keep it out of the reach of children and pets.
- Do not leave packaging material lying around carelessly. This may become dangerous playing material for children.
- Meters and accessories have no place in the hands of children. Therefore, be particularly careful when children are present.
- Protect the appliance from extreme temperatures, direct sunlight, strong jolts, high humidity, moisture, flammable gases, steam and solvents.
- Do not operate the meter in rooms or under unfavorable conditions where combustible gases, vapors or dust are or may be present.
- Avoid operation in direct proximity of strong magnetic or electromagnetic fields or transmission aerials. This could affect the measurement.
- Steam, dust, smoke and/or vapours may impair looks and lead to an incorrect measuring result!
- Do not place the product under any mechanical stress.
- If it is no longer possible to operate the product safely, take it out of operation and protect it from any accidental use. Safe operation can no longer be guaranteed if the product:
- is visibly damaged,
- is no longer working properly,
- has been stored for extended periods in poor ambient conditions or
- has been subjected to any serious transport-related stresses.
- Please handle the product carefully. Jolts, impacts or a fall even from a low height can damage the product.
- Consult an expert when in doubt about the operation, safety or connection of the appliance.
- Maintenance, modifications and repairs must only be completed by a technician or an authorized repair centre.
- If you have questions that remain unanswered by these operating instructions, contact our technical support service or other technical personnel.
Laser
- When operating the laser equipment, always make sure that the laser beam is directed so that no one is in the projection area and that unintentionally reflected beams (e.g., from reflective objects) cannot be directed into areas where people are present.
- Laser radiation can be dangerous if the laser beam or its reflection enters unprotected eyes. Therefore, before using the laser equipment, familiarise yourself with the statutory regulations and instructions for operating such a laser device.
- Never look into the laser beam and never point it at people or animals. Laser radiation can seriously damage your eyes.
- If laser radiation enters your eyes, close your eyes immediately and move your head away from the beam.
- If your eyes have been irritated by laser radiation, do not continue to carry out tasks with safety implications, such as working with machines, working from great heights or close to high voltage. Also, do not operate any vehicles until the irritation has completely subsided.
- Do not point the laser beam at mirrors or other reflective surfaces. The uncontrolled, reflected beam may strike people or animals.
- Never open the device. Setting or maintenance tasks must only be executed by a trained specialist familiar with potential hazards. Improperly executed adjustments might result in dangerous laser radiation.
- The product is equipped with a class 2 laser. Laser signs in different languages are included in the package. If the sign on the laser is not written in the language of your country, please affix the appropriate sign onto the laser.
- Caution: if operation settings or procedures other than those described in these instructions are used, it could lead to exposure to dangerous radiation.
Li-ion battery
- The rechargeable battery is permanently built into the product and cannot be replaced.
- Never damage the rechargeable battery. Damaging the casing of the rechargeable battery might cause an explosion or a fire!
- Never short-circuit the contacts of the rechargeable battery. Do not throw the battery or the product into fire. There is a danger of fire and explosion!
- Charge the rechargeable battery regularly, even if you do are not using the product. Due to the rechargeable battery technology being used, you do not need to discharge the rechargeable battery first.
- When charging, place the product on a surface that is not heat-sensitive. It is normal that a certain amount of heat is generated during charging.
Operating elements
-
Visual targeting device (rear/fore sight)
-
Colour display
-
Button “Down” and “Video recording”
-
Button “On/off” and “ESC” (Cancel/Exit)
-
Button “Up” and “Image recording”
-
Button “Enter” for input confirmation
-
Stand connection socket
-
Lens protective cover
-
Measuring button
-
Folding lid for battery compartment and microSD
-
Camera lens
-
Exit openings for double target laser
-
IR measuring lens
-
Integrated sensors for ambient temperature and rel.
humidity -
Connection socket for type K thermocouple sensor 16 RESET button for system reset
-
Mini USB charging socket
-
Chute for microSD memory card
-
Battery compartment
Product Description
Function
Infrared thermometers (IR-thermometers) measure the surface temperature of an
object. The product sensor records the emitted, reflected and passed-through
heat radiation of the object and converts this information into a temperature
value.
The degree of emission is a value used to describe the energy emission
characteristic of a material. The higher this value, the higher the material’s
ability to emit radiation. Many organic materials and surfaces have an
emission de-gree of approx. 0.95. Metal surfaces or glossy materials have a
lower emission degree. This will cause an inaccurate measurement. For this
reason, a matte-black coat of paint or matte adhesive tape should be applied
or the emission degree pre-set accordingly.
IR-measuring optics – D:S ratio
(D:S = measuring distance: measuring area)
To achieve precise measuring results, the measured object must be larger than
the IR measuring spot of the infrared thermometer. The determined temperature
is the average temperature of the measured surface. The smaller the measured
object, the shorter the distance to the infrared thermometer has to be. The
precise measuring spot size can be taken from the following chart. It is also
printed on the device. For precise measurements, the measuring object should
be at least twice as large as the measuring spot.
The meter has a lens of 50:1. The smallest measuring diameter is reached
before the measuring opening at a distance of up to 90 cm. It is 18 mm in
diameter. Keep the sufficient distance to avoid incorrect measurements due to
heating of the IR thermometer. Then the measuring cone expands as shown in the
sketch.
Example: The measuring diameter at a distance of 1.8 m is 36 mm.
Target laser
The target laser is active during measurement depending on the pre-setting. A
warning icon appears in the display with the laser active. Never look into the
laser aperture (12) during measurement.
The target laser is designed as a twin device and marks the inner, approximate
edge area of the measured area (approx. 90%).
When both laser points meet, the smallest measuring area has been reached. It
is 18 mm in diameter.
For large distances, the two laser points move apart according to the
measuring area.
Commissioning
- Before working with the meter, you have to insert the enclosed Li-ion rechargeable battery.
- Insert the rechargeable battery as described in this chapter and charge it.
- The IR-thermometer permits individual system settings for measuring that you should perform after initial commissioning.
Inserting/replacing the rechargeable battery
- The enclosed Li-ion battery must be inserted first at initial commissioning. To open the battery compartment, pull the battery compartment lid (10) away forwards. The battery compartment lid is unlocked. Fold open the battery compartment lid (10) forwards. Connect the plug of the rechargeable battery to the pug-in socket in the battery compartment in the correct polarity. The pug is designed so that the rechargeable battery can only be connected in the proper polarity. Do not apply any force when plugging in the rechargeable battery.
- Close the battery compartment again by folding shut the battery compartment lid (10). Observe that the cables are not caught and the lock latches.
- Do not leave the rechargeable battery in the meter during longer measuring breaks (approx. >2 weeks), since even leak-protected rechargeable batteries may corrode and release chemicals that may harm your health or destroy the device.
A matching Li-ion spare battery of type 18500 is available under the order
number: Item no. 1233684 (please order 1x).
Other rechargeable battery types or non-rechargeable batteries must NOT be
used.
Inserting microSD card
- A microSD card (supplied) can be inserted in the meter for storage expansion and for data transmission to a computer.
- Open the battery compartment (10) to insert the memory card.
- Insert the microSD card into the lateral microSD card chute (18) with the contacts down and outward.
- Carefully push the card into the chute until it latches.
- Close the battery compartment.
After activating of the meter, the inserted microSD card is displayed as a small icon at the upper left screen edge.
To remove the microSD card, briefly push the card until it is unlocked. The card is pushed up by the latching mechanism and can be taken out.
Charging the rechargeable battery
The enclosed Li-ion battery must be fully charged first at initial
commissioning.
It can be charged at any USB interface or the enclosed USB mains unit.
Charge the rechargeable battery no later than when the battery symbol shows no
fill bars in the display anymore.
- Open the rubber cover at the bottom of the meter and connect the enclosed USB charging cable to the mini USB charge socket (17) at the meter. The plug will only fit into the charge socket in the correct orientation.
- Plug the USB plug of the charging cable into an active USB socket at a computer or connect the plug to the en-closed USB mains unit. Charging commences as soon as the mains unit is plugged into a mains socket.
- The display shows charging by the filling (pulsating) battery symbol. If the fill bar is continually visible, charging is completed. The device can be disconnected from the charge cable.
- Pull the USB mains unit from the socket when it is not needed for charging.
Switching the device on/off
- Switch on the meter. For this, hold down the button “ESC” (4) for approx. 3 seconds until a beep sounds. Release the button and the meter will switch on after a brief time. Open the lens protection cover (8).
- After switching on, you are in the main menu “MENU”, from where you can select all functions.
- To switch off, press the button “ESC” again.
System settings
To enter the setting mode, push the button “Up” (5) or “Down” (3) in the main
menu “MENU” until the field “SETTINGS” has a blue background. Then push the
button “Enter” (6). Use the buttons “Up” (5) or “Down” (3) to select the
respective function or change the value. Press the button “ESC” (4) to return
to the previous menu. The function with a blue background is saved.
The setting menu can be terminated at any time by pushing the button “ESC”
(4). The function with a blue background is saved.
Measuring
To receive precise measured values, the infrared thermometer must be adjusted
to the ambient temperature. Let the device adjust to the new ambiance
temperature after a site change.
Longer IR measurements of high temperatures at low measuring distances lead to
heating of the meters and thus to wrong measurements. For precise measured
values: The higher the temperature, the larger should the measuring distance
and the shorter the measuring duration be.
Glossy surfaces will falsify the measuring result in the IR measurement. To
compensate, the surface of glossy parts can be covered with adhesive tape or
matte-black paint. The devices cannot measure through transparent surfaces,
such as glass. Instead, it will measure the surface temperature of the glass.
Display symbols and values
In measuring operation, various parameters and symbols are displayed on the
screen. The following list explains their meaning:
Measurement settings
Further settings can be made in the individual measuring functions and the
values to be displayed can be enabled or disabled there.
To get to the setting menu, push the button “Enter” in a measuring function.
The menu “MEASURE SET” is opened.
Select by pushing the button “Enter” again. You can use the arrows to change
the function/set value. To terminate, push the button “ESC”.
The following functions can be selected and set:
Emission
The emission degree can be set individually in this menu from 0.10 to 1.00 or
selected with pre-set material values. Possible set values are:
ε = x.xx set value 0.10 – 1.00
Cement (0.94)| | Ice/Water (0.96)
Glass (0.92)| | Plastic (0.90)
Human Skin (0.98)| | Wood (0.87)
Alarm High
Here, the upper alarm level (High) of the temperature can be set. You can
enable or disable the function via the selection field.
Possible set values are:
Time value = xx.x set value -50 to +1650 °C
Disable = Function is disabled| | Enable = Function is enabled
Alarm Low
Here, the lower alarm level (Low) of the temperature can be set. You can
enable or disable the function via the selection
field.
Possible set values are:
Time value = xx.x set value -50 to +1650 °C
Disable = Function is disabled| | Enable = Function is enabled
Laser
Here, the target laser during IR measuring can be switched on or off.
Pushing the button “Enter” actuates the slider in the display.
I = enabled, o = disabled
Auto Mode
Here, the permanent measuring can be switched on or off. Pushing the button
“Enter” actuates the slider in the display. I = enabled, o = disabled
The automatic deactivation functions are disabled in permanent measuring mode.
Ensure that you switch off the meter after the measurement is completed.
IR Measurement with the camera image
The menu item “IR CAM” is used for measurements with the activated camera.
This makes it easier to assign the measuring point.
Point the measuring opening (13) at the measured object vertically. Observe
that the measured object is not smaller than the IR measuring area of the
device.
Push and hold the measuring button (9). In the display, the main measured
value and all other enabled parameters are displayed. The displayed main
measured value corresponds to the average surface temperature of the IR
measuring area. During the measurement, the symbol “Start” in the display is
concealed and the symbol “Pause” is shown.
After releasing the measuring button (9), the last measured value is displayed
until automatic deactivation (can be pre-set) for better reading. The symbol
“Pause” goes out and the display “Start” appears to indicate readiness for
measurement.
The device will switch off automatically after the pre-set time once the
measuring button (9) is released. If this function has been disabled, always
switch off the device manually to avoid discharging the rechargeable battery.
To determine the warmest/coldest points of the measuring object, move the
measuring aperture with the measuring button (9) pushed comprehensively over
the surface of the object to be measured. The measured value is updated at a
rate of 150 ms. Additional functions permit automatic display of maximum value
“Max” and minimum value “Min” or average value “Avg” and difference value
“Dif”. These values are displayed above and below the main measuring display.
The additional functions “MIN, MAX, DIF, AVG” always refer to the IR measuring
function.
Picture and video recording
The IR measurement with camera image also has the advantage that the displayed
pictures including all measuring parameters can be recorded as a picture or
video file.
To record the picture, push the button “Picture recording” (5). The current
picture is recorded.
The two arrows can be used to choose saving the picture (arrow “Up” (5)) or
discard the picture (arrow “Down” (3)).
After input, the current live picture is displayed again.
To record the video, push the button “Video recording” (3). The symbol “Video
recording” is displayed and the recording is running. To terminate the
recording, push the button “Video recording” (3) again. A small floppy disc
symbol in the middle of the image signals that the video file is being saved.
The picture or video file can be viewed in the main menu item “Gallery”.
IR Measure
The menu item “IR MEASURE” is used for measurement without the camera. This
makes it easier to read the measurement because of the homogeneous image
background.
Point the measuring opening (13) at the measured object vertically. Observe
that the measured object is not smaller than the IR measuring area of the
device.
Push and hold the measuring button (9). In the display, the main measured
value and all other enabled parameters are displayed. The displayed main
measured value corresponds to the average surface temperature of the IR
measuring area. During the measurement, the symbol “Start” in the display is
concealed and the symbol “Pause” is shown.
The bar display at the lower edge of the screen shows the trend of the current
measured value for a quick overview (rising/falling).
After releasing the measuring button (9), the last measured value is displayed
until automatic deactivation (can be pre-set) for better reading. The symbol
“Pause” goes out and the display “Start” appears to indicate readiness for
measurement.
Contact measurement
Observe that contact temperature measurement is only permitted with objects
that do not turn and are not live. Also observe that contact temperatures are
only possible up to the approved sensor temperature. The approved sensor
temperature of the enclosed wire sensor is -20 to +250 °C. To use the entire
measuring range of the meter, you can purchase optional thermocouple sensors
of type K.
The contact measurement is possible in the measuring functions “IR CAM”, “IR
MEASURE”, “DEW POINT” and “DATA LOGGER”.
In addition to contact-free IR measurement, a thermocouple contact sensor type
K can be connected as well. The contract measurement permits measurement of
the temperature independently of the material and degree of emission of the
object. In addition to the enclosed wire temperature sensor, all conventional
type K thermocouple sensors with miniature plug can also be connected to the
devices
Sensor connection
- Open the lower cover at the meter.
- Push the thermocouple plug of the sensor into the type K connection socket in the proper polarity (15). Observe the mark “+” at the plug and the socket. The plug only fits the socket in the correct polarity.
- Once a thermo sensor is connected to the meter with the type K function active, the display “TK” with the temperature of the sensor appears below the IR measured value in measuring operation.
- Remove the sensor after the end of the measurement and close the cover to prevent the ingress of dirt.
The additional functions “MIN”, “MAX”, “AVG” and “DIF” are not possible for contact measurement. The information always refers to the IR measurement.
Dew point
The menu item “DEW POINT” is used for measurement without the camera. This
makes it easier to read the measurements because of the homogeneous image
background.
Point the measuring opening (13) at the measured object vertically. Observe
that the measured object is not smaller than the IR measuring area of the
device.
Push and hold the measuring button (9). In the display, the main measured
value and all other enabled parameters are displayed. The displayed main
measured value corresponds to the average surface temperature of the IR
measuring area. During the measurement, the symbol “Start” in the display is
concealed and the symbol “Pause” is shown.
The bar display at the lower edge of the screen shows the correspondence of IR
temperature and dew point (DP) in percent for quick overview. It is used as a
warning display for possible mold danger.
Blue range: no danger
Yellow range: possible danger
Red range: acute danger
After releasing the measuring button (9), the last measured value is displayed
until automatic deactivation (can be pre-set) for better reading. The symbol
“Pause” goes out and the display “Start” appears to indicate readiness for
measurement.
The dew point and the evaporation temperature are calculated values that
result from the IR surface temperature, the ambient temperature and the
relative humidity. The sensors for ambient temperature and relative humidity
are installed in the meter (14). Observe the delayed measurement by the
constructional placement of the sensors. Give the meter enough time to adjust
to the ambient conditions (approx. 30 minutes for a stable measured value).
Data logger
The menu item “DATA LOGGER” is used to record the measuring progress.
In data logger mode, when using the supplied microSD card, all measured values
are saved in an editable text file (.txt).
Point the measuring opening (13) at the measured object vertically. Observe
that the measured object is not smaller than the IR measuring area of the
device.
To start recording, push the measuring button (9) briefly. The recording is
started with the pre-set measuring interval. The lock symbol marks the locked
permanent measuring mode. The measuring button is locked and the data are
stored.
In the display, the main measured value is displayed as a figure. The pre-set
high and low alarm levels are always displayed as coloured threshold lines.
To terminate the data logger measurement, push the button “ESC”. The measuring
function is terminated; the data remain in the device memory. The saved data
are saved with time and date designation.
Data view “Gallery”
In the menu item “GALLERY”, the recorded data can be viewed on the meter.
In the submenu, the following data can be selected with the arrow buttons and
the button “Enter”: Video For the saved video files
Picture For the saved picture files
Logs For the saved data logging files
The files can be deleted via the menu item “Memory” in the main menu
“Settings”. Picture files can be deleted in the picture viewing mode.
Individual video and logging files cannot be deleted. It is only possible to
delete (format) the entire memory.
Menu item “Video”
Use the arrows to select the automatically created folder and confirm this
with the button “Enter”. Repeat this for the corresponding video file.
After confirming with the button “Enter”, the video will start.
Use the button “Enter” during playback to display an info line with a progress
bar at the lower edge of the screen. Pushing the button “Enter” again will
stop playback. The symbol “Pause” is displayed. Pushing “Enter” resumes
playback.
All video files are played back in sequence. The arrows can be used to select
the saved video files forward and backward.
The button “ESC” ends playback and returns to the video file menu.
Menu item “Picture”
Use the arrows to select the automatically created folder and confirm this
with the button “Enter”. Repeat this for the corresponding picture file.
After confirming with the button “Enter”, the selected picture is displayed.
The arrows can be used to select the saved pictures forward and backward.
The button “Enter” can be used to open a submenu while viewing a picture.
Remove photograph
Use the button “Enter” to delete the picture. The safety query for deleting
must be confirmed (Yes = delete, No = do not delete)
Information on the photograph
The button “Enter” displays file information on the picture. The field “Exit”
terminates the display.
Exit
The button “Enter” terminates the submenu and returns to picture viewing.
The button “ESC” ends playback and returns to the picture file menu.
Menu item “Logs”
Use the arrows to select the desired file and confirm this with the button
“Enter”. The chart is displayed with the first measured temperature.
Use the arrows to control a cursor across the chart. The right arrow moves the
cursor to the right, the left arrow moves it to the left. The temperature
value at the cursor is displayed at the bottom edge of the screen.
When the cursor reaches the screen edge, the chart is moved on accordingly.
The button “ESC” ends display and returns to the log file menu.
Care and cleaning
General information
The IR thermometer is maintenance-free apart from occasional cleaning and
charging of the rechargeable battery. Always close the safety cover (8) when
the device is not in use.
Always observe the following safety information before cleaning the device.
Cleaning the lens
Remove loose particles with clean compressed air and wipe off any other
deposits with a fine lens brush. Clean the surface with a lens cleaning cloth
or a clean, soft and lint-free cloth. To clean finger prints and other grease
strains, the cloth can be moistened with water or a lens cleaning fluid. Do
not use any acid/alcohol-containing or other solvents and do not use any
rough, linty cloth to clean the lens. Avoid excessive pressure when cleaning.
Cleaning the housing
Do not use any abrasive, chemical or aggressive cleaning media such as
petrols, alcohols or similar for cleaning. They will damage the surface of the
device. Furthermore, the fumes are hazardous to your health and explosive.
Also do not use any sharp-edged tools, screwdrivers, metal brushes, etc. for
cleaning.
Use a clean, lint-free, antistatic and slightly moist cleaning cloth to clean
the device or the measuring lines.
Disconnect the product from the mains before each cleaning.
If you have any questions about handling of the device, contact our technical
support.
Troubleshooting
With this IR thermometer, you have purchased a product built to the latest
state of the art and operationally safe. Nevertheless, problems or errors may
occur.
For this reason, the following is a description of how you can easily remove
possible malfunctions yourself:
Problem | Possible cause |
---|---|
The measuring device does not work. | Is the rechargeable battery flat? |
Check the status.
Wrong measured value indication.| Is the wrong emission degree set? Is the lens contaminated?
Is the measuring surface covered by a glass pane?
Is the operating temperature of the environment exceeded or undercut?
The meter no longer reacts.| The system software has a fault. Push the button “RESET” (16) with a pointy object (e.g. a paper clip) and switch on the device again.
Other repairs than described above must only be performed by an authorised specialist. If you have any questions about handling of the device, contact our technical support.
12. Disposal
Product
Electronic devices are recyclable waste and must not be disposed of in
household waste. At the end of its service life, dispose of the product in
accordance with applicable regulatory guidelines.
Remove any inserted (rechargeable) batteries and dispose of them separately
from the product.
(Rechargeable) batteries
You as the end user are required by law (Battery Ordinance) to return all used
(rechargeable) batteries. Disposing of them in the household waste is
prohibited.
Contaminated (rechargeable) batteries are labeled with this symbol to indicate
that disposal in domestic waste is forbidden. The designations for the heavy
metals involved are: Cd = Cadmium, Hg = Mercury, Pb = Lead (name on
(rechargeable) batteries, e.g. below the trash icon on the left). Used
(rechargeable) batteries can be returned to collection points in your
municipality, our stores or wherever (re-chargeable) batteries are sold.
You thus fulfill your statutory obligations and contribute to the protection
of the environment.
Technical data
- Input voltage………………………………………………Li-ion rechargeable battery 3.7 V 1400 mAh, type 18500
- Charge time……………………………………………….approx. 2 hrs
- Operating time per battery charge…………………approx. 4 hrs continuous operation
- Trigger time……………………………………………….150 ms
- IR spectral range………………………………………..8 – 14 μm
- Emission degree…………………………………………adjustable from 0.10 to 1.00
- Measuring range (IR measurement)………………-50 to +1600 °C (-58 to +2912 °F)
- Measuring range (contact measurement)……….-50 to +1370 °C (-58 to +2498 °F)
- Display resolution………………………………………. 0.1 °C/°F < 1000 °C/°F 1 °C/°F >/= 1000 °C/°F
- IR-lens (D:S)………………………………………………50:1
- Dual laser…………………………………………………. Output < 1 mW, laser class 2,
- Wavelength 630-670 nm
- Display………………………………………………………5.6 cm (2.2”), colour LCD
- Display resolution……………………………………….320 x 240 pixels
- Camera resolution………………………………………640 x 480 pixels
- Image format……………………………………………..jpg
- Video format………………………………………………AVI
- Logging format……………………………………………bin / .txt
- Integrated device memory……………………………300 MB, non-volatile
- External memory chute………………………………..up to 32 GB microSDHC
- Operating temperature………………………………..0 to +50 °C (+32 to 122 °F)
- Operating humidity……………………………………..0 – 100% relative humidity
- Storage temperature……………………………………-10 to +60 °C (+14 to +140 °F)
- Storage humidity…………………………………………10 – 90% relative humidity, non-condensing
- Weight………………………………………………………approx. 450 g
- Dimensions………………………………………………..220 x 62 x 155 mm
- Input voltage (USB mains unit)……………………..100 – 240 V/AC 50 – 60 Hz
- Protection class………………………………………….2 (double or reinforced insulation)
- Output voltage……………………………………………5 V/DC
- Output current…………………………………………….1000 mA
Measuring tolerances
The accuracy is valid for one year at a temperature of +23 to +25 °C (+73 to
+77 °F), and at a relative humidity of less than 75%, non-condensing.
Infrared temperature measuring area
Measuring area °C | Accuracy | Reproducibility |
---|---|---|
-50 to +20 °C | ±5 °C | ±2 °C |
+20 to +500 °C | ±(1.5 % +1.5 °C) | ±1 % or ±1 °C |
+500 to +1000 °C | ±2 % | ±1 % or ±1 °C |
+1000 to +1600 °C | ±2.8 % | ±1.5 % |
Contact temperature measurement type K
Measuring area °C| Accuracy|
---|---|---
-50 to 0 °C| ±3.5 °C|
0 to +1370 °C| ±(1 % + 2 °C)|
Measuring range °F| Accuracy|
---|---|---
-58 to +32 °F| ±7 °F|
+32 to +2498 °F| ±(1 % + 4 °F)|
Air temperature measurement
Measuring area °C | Accuracy |
---|---|
0 to +50 °C | ±1 °C |
±1.5 °C | Residual range |
Measuring range °F | Accuracy |
--- | --- |
+32 to +122 °F | ±1.8 °F |
±3 °F | Residual range |
Dew point temperature measurement
Measuring area °C | Accuracy |
---|---|
0 to +50 °C | ±1 °C |
±1.5 °C | Residual range |
Measuring range °F | Accuracy |
--- | --- |
+32 to +122 °F | ±1.8 °F |
±3 °F | Residual range |
Relative humidity measurement
Measuring range %RH | Accuracy |
---|---|
0 – 40 %RH | ±5 %RH |
40 – 60 %RH | ±4 %RH |
60 – 80 %RH | ±5 %RH |
80 – 100 %RH | ±6 %RH |
Degree of emission of different surfaces
The emission degrees listed in the table are approximated. Various parameters,
such as geometry and surface quality, may influence the emission degree of an
object.
The meter is delivered with a pre-set emission degree of 0.95. It matches for
most non-metal materials. The IR measuring method is only conditionally
suitable for blank metals and requires a special surface treatment (e.g. matte
insulation tape, etc.) or pre-setting of the corresponding degree of emission.
Surface | Emission Degree | Surface | Emission Degree | |
---|---|---|---|---|
Aluminium, blank | 0.04 | Copper, oxidised | 0.60 | |
Aluminium, oxidised | 0.30 | Brass, oxidised | 0.50 | |
Asphalt | 0.90 – 0.98 | Paints, matte | 0.93 – 0.97 | |
Concrete | 0.94 | Foods | 0.93 – 0.98 | |
Ice | 0.96 – 0.98 | Human skin | 0.98 | |
Iron oxide | 0.78 – 0.82 | Plastic | 0.94 | |
Colour | 0.93 | Paper | 0.97 | |
Plaster | 0.80 – 0.90 | Sand | 0.90 | |
Glass/porcelain | 0.92 – 0.94 | Textiles | 0.90 | |
Rubber, black | 0.94 | Water | 0.92 – 0.96 | |
Wood | 0.87 – 0.94 | Brick, plaster | 0.93 – 0.96 | |
Plastic | 0.90 | Cement | 0.94 |
References
- France.fr : Explore France and its wonders - Explore France
- Conrad Electronic » Alle Teile des Erfolgs
- Conrad Electronic » All parts of success
- Conrad Electronic – Alle Teile des Erfolgs
- Conrad Electronic » All parts of success
- Contact
- Download center
- Conrad Electronic Online Shop » Alle Teile des Erfolgs
- Conrad Electronic » All parts of success
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>