VOLTCRAFT VC371 AC DC Mini Clamp Meter Instruction Manual
- June 1, 2024
- VOLTCRAFT
Table of Contents
- VOLTCRAFT VC371 AC DC Mini Clamp Meter
- Product Information
- Product Usage Instructions
- FAQs
- Introduction
- Intended use
- Delivery contents
- Operating Instructions for download
- Safety instructions
- Overview
- Replacing batteries
- Operation
- Measurement
- Testing
- Troubleshooting
- Cleaning
- Technical data
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
VOLTCRAFT VC371 AC DC Mini Clamp Meter
Specifications:
- Model: VC371 40A AC/DC Mini Clamp Meter
- Item Number: 2893199
Product Information
The VC371 40A AC/DC Mini Clamp Meter is a versatile tool for measuring both AC and DC currents. With features like automatic shutdown, display hold function, and the ability to measure maximum and minimum values, this clamp meter is designed to provide accurate measurements in various electrical applications.
Product Usage Instructions
1. Battery Replacement:
To replace the batteries in the clamp meter, follow these steps:
- Ensure the product is switched off.
- Use a Phillips screwdriver to remove the screw of the battery compartment cover.
- Replace the batteries with new ones, observing correct polarity.
- Securely reattach the battery compartment cover.
2. Operation:
Before use, familiarize yourself with the safety instructions provided in the manual. The operation involves:
- Dial Setting: Set the dial to the appropriate range/function before each use.
- Power On/Off: The product is off when the function switch is in the OFF position. Turn off after use.
- Automatic Shutdown: The meter will automatically turn off if inactive. Reactivate by turning off and on again.
3. Display Functions:
- Display Hold: Press the HOLD button to freeze the display.
- Max/Min Value Display: View maximum and minimum values measured.
- Work Light: The built-in light aids visibility in dimly lit environments.
4. Measurements:
AC Current Measurement: Follow these tips for accurate measurement:
- Place the conductor as shown in the diagram within the clamp.
- Note down the measured current.
- Divide the measured current by the number of windings (coils).
FAQs
- Q: How do I download user manuals for this product?
- A: Visit the respective Conrad website for your region (e.g., www.conrad.de for Germany) to find and download the user manual.
- Q: What should I do if the display does not turn on?
- A: Check the battery compartment for correct battery installation and replace batteries if necessary. Ensure the device is switched on as well.
- Q: Can this clamp meter measure both AC and DC currents?
- A: Yes, the VC371 40A AC/DC Mini Clamp Meter can measure both AC and DC currents.
“`
Introduction
Dear customer, Thank you for purchasing this product. If there are any technical questions, please contact: www.conrad.com/contact
Intended use
The product is a clamp meter and can be used to measure and display various
electrical parameters. The product is in accordance with the safety
requirements for Electronic Measuring Equipment, EN 61010-1 and EN
61010-2-032. The product conforms to CAT III 600 V: MEASUREMENT CATEGORY III
is applicable to test and measuring circuits
connected to the distribution part of the building’s low-voltage MAINS
installation. The product is designed for private and commercial use. In
commercial institutions, the accident prevention regulations of the Employer’s
Liability Insurance Association for Electrical Systems and Operating Materials
are to be observed. The product can be used in schools and training centres.
The use must be supervised by trained personnel. The product is intended for
indoor use only. Do not use it outdoors. Contact with moisture must be avoided
under all circumstances. If you use the product for purposes other than those
described, the product may be damaged. Improper use can result in short
circuits, fires, electric shocks or other hazards. The product complies with
the statutory national and European requirements. For safety and approval
purposes, you must not rebuild and/or modify the product. Read the operating
instructions carefully and store them in a safe place. Make this product
available to third parties only together with the operating instructions.
39
All company names and product names are trademarks of their respective owners. All rights reserved.
Delivery contents
Product 2 x 1.5 V AAA batteries Test leads with removable probe tip covers Storage bag Operating instructions
Operating Instructions for download
Use the link www.conrad.com/downloads (alternatively scan the QR code) to download the complete operating instructions (or new/current versions if available). Follow the instructions on the web page.
Description of symbols
This product conforms to the required CE standards and is in compliance with
applicable European (EU) directives. This product is UK conformity assessed
and meets applicable Great Britain directives.
The symbol warns of hazards that can lead to personal injury.
The symbol warns of dangerous voltage that can lead to personal injury by
electric shock.
40
Protection class 2 (double or reinforced insulation, protective insulation).
Application around and removal from HAZARDOUS LIVE conductors is permitted.
Personal protective equipment must be used.
Can test and measure circuits connected to the distribution part of the
building’s low-voltage MAINS installation.
Alternating current (AC)
Direct current (DC)
Earth ground
Polarity markings for direct current (DC) measurement. The symbols indicate
the direction of current for taking measurements.
Safety instructions
Read the operating instructions carefully and especially observe the safety
information. If you do not follow the safety instructions and information on
proper handling, we assume no liability for any resulting personal injury or
damage to property. Such cases will invalidate the warranty/guarantee.
6.1 General
The product is not a toy. Keep it out of the reach of children and pets. Do
not leave packaging material lying around carelessly. This may become
dangerous playing material for children. If you have questions which remain
unanswered by this information product,
contact our technical support service or other technical personnel.
41
Maintenance, modifications and repairs must only be completed by a technician
or an authorised repair centre.
6.2 Handling
Handle the product carefully. Jolts, impacts or a fall even from a low height
can damage the product.
6.3 Operating environment
Do not place the product under any mechanical stress. Protect the appliance
from extreme temperatures, strong jolts, flammable
gases, steam and solvents. Protect the product from high humidity and
moisture. Protect the product from direct sunlight. Do not switch the product
on after it has been taken from a cold to a warm en-
vironment. The condensation that forms might destroy the product. Allow the
product to reach room temperature before you use it. Never operate the product
in direct proximity of strong magnetic or electromagnetic fields or
transmitter aerials or HF generators. Doing so can prevent the product from
functioning properly.
6.4 Operation
Consult an expert when in doubt about the operation, safety or connection of
the product.
If it is no longer possible to operate the product safely, take it out of
operation and protect it from any accidental use. DO NOT attempt to repair the
product yourself. Safe operation can no longer be guaranteed if the product:
is visibly damaged, is no longer working properly, has been stored for
extended periods in poor ambient conditions or has been subjected to any
serious transport-related stresses.
6.5 Batteries
Correct polarity must be observed while inserting batteries.
42
The batteries should be removed from the device if it is not used for a long
period of time to avoid damage through leaking. Leaking or damaged batteries
might cause acid burns when in contact with skin, therefore use suitable
protective gloves to handle corrupted batteries.
Batteries must be kept out of reach of children. Do not leave batteries lying
around, as there is risk, that children or pets swallow them.
All batteries should be replaced at the same time. Mixing old and new
batteries in the device can lead to battery leakage and device damage.
Batteries must not be dismantled, short-circuited or thrown into fire. Never
recharge non-rechargeable batteries. There is a risk of explosion!
6.6 Connected devices
Also observe the safety and operating instructions of any other devices which
are connected to the product.
6.7 Product
Before use, always verify measurements against a known voltage source to
ensure safe operation. If abnormal or erratic operation is detected:
Stop using immediately
have the product inspected by a qualified technician When taking
measurements make sure no objects are trapped between the
clamp jaws (e.g., cables). Do not exceed the maximum permissible measurement
values. Risk of fatal electric shock! The product must never be used if the
housing or
battery compartment cover is open. Risk of electric shock! Use caution when
working with voltages above 30 V/AC
rms (42.4 V peak), 60 V/DC. The rotary switch should be set to the correct
range / function before each use. Inspect the product for damage before each
measurement. Never take meas-
urements if the insulation or product is damaged. Use extreme caution when
working around bare conductors or bus bars as
contact can result in electric shock. When taking measurements always keep the
fingers behind the finger guards.
43
6.8 Test leads and probes
The voltage between the connection points of the meter and the earth potential
must not exceed 600 V AC/DC in CAT III.
Probe assemblies to be used for MAINS measurements should meet EN 61010-031
standard, rated CAT III 600 V, 40 A or better.
Risk of electric shock! Use caution when working with voltages above 30 V/AC
rms (42.4 V peak), 60 V/DC.
Test leads must be disconnected before changing the range / function. The
cables have a wear indicator. If damaged, a second insulation layer in a
different colour becomes visible. Do not use if this occurs, replace
immediately! When taking measurements do not grip beyond the finger barrier or
grip range
markings on the probes. When taking CAT III measurements, probes with cover
caps (max. 4 mm free contact length) must be used to avoid accidental short
circuits.
When using the probes without cover caps, measurements between the meter and
the earth potential must not be performed above the measuring category CAT II.
Prevent short circuits by ensuring test points/connections do not touch when
taking measurements.
Always check the probes and leads for any sign of damage before each use. Do
not use if damaged, replace immediately!
44
Overview
7.1 Product
1
2
3
4
5
13
14 7
6
8 9 10
11
12
1 Current sensing clamp 3 Rotary switch 5 Display
7 Work light 9 Non-contact voltage (AC)
tri-colour LED 11 REL|ZERO, MAX|MIN button 13 COM terminal
2 HOLD button 4 SELECT, V.F.C button 6 Non-contact voltage (AC)
detector tip 8 Finger guard 10 Clamp lever
12 Battery compartment
14
terminal
7.2 Display symbols
Symbol A AC DC
NCV
Description Current (amps) Alternating current Direct current Non-contact voltage
45
Symbol MAX MIN REL
ZERO
k, M
V mV A mA, µA nF mF, µF VFC
Description Maximum value Minimum value Relative mode Overload: range exceeded Automatic shut-off active True root mean square measurement Low battery indicator Display hold active Zero position Minus sign Automatic range selection is active Continuity check Diode test Ohm (unit of electric resistance) Kiloohm (103), Megaohm (106) Volt (unit of electric voltage) Millivolt (10-3) Ampere (unit of current measurement) Milliampere (10-3), Microampere (10-6) Nanofarad (10-9), unit of electrical capacitance Millifarad (10-3), Microfarad (10-6) Low-pass filter activated
46
Replacing batteries
Important: Low battery voltage can affect the accuracy of readings resulting
in electric shock and/or injury: Replace the batteries when the low battery
indicator shows. Rechargeable batteries are not recommended as they
typically have a lower
voltage per cell.
Risk of electric shock! Disconnect the product from any input signals before
replacing the batteries.
Preconditions:
a The power is switched OFF.
1. Use a cross-head screwdriver to remove the compartment cover screw. 2.
Replace the batteries and match the polarity markings shown inside the com-
partment. 3. Replace the battery compartment cover taking care not to over-
tighten the
screw.
Operation
IMPORTANT! Always observe the information contained in the section: Safety
instructions [} 41].
9.1 Rotary switch
The rotary switch should be set to the correct range / function before each
use.
When a function is selected, a beep will sound and the display will update.
47
9.2 Power ON / OFF
The product is switched off when the function switch is in the OFF position.
Switch the power OFF after use.
9.3 Auto shut-off
Auto shut-off is active by default, it is indicated by the symbol. This energy
saving feature will shut power off after approximately 15 minutes of
no activity. To disable auto shut-off: 1. Set the rotary switch to OFF. 2.
Press and hold the SELECT button and then set the rotary switch to any posi-
tion other than OFF. à The symbol will disappear and an alert will sound when
disabled. 3. Auto shut-off will reactivate after the power is switched OFF.
9.4 Display hold
Important: The display hold function freezes the display. The display hold
should be switched OFF before taking measurements.
Press the HOLD button to switch display hold ON/OFF. The hold icon will appear
when display hold is ON.
9.5 Maximum / minimum value display
In this mode the display shows the “MIN” (minimum) or “MAX” (maximum) value
measured. 1. Press and hold the REL/ZERO [MAX/MIN] button to enter maximum /
minimum
value display mode. 2. Press the MAX/MIN button repeatedly to toggle between
modes.
à The display will show “MAX” or “MIN” to indicate which mode is active. 3.
Press and hold the MAX/MIN button to exit this mode.
48
9.6 Work light
Press and hold the HOLD button to switch the work light ON / OFF.
9.7 Relative mode
Relative mode can be used to take differential readings between two test
points or for tracking changes in measurements from a defined reference point
(e.g., line losses). 1. Select a function using the rotary switch: V , V , 4A
, 40A . 2. Take a measurement and note the displayed value. 3. Press the REL
button to activate relative mode.
à “REL” will appear on the display to indicate relative mode is active. 4.
Take another measurement.
à The display will show the difference between the new reading and the initial
reading.
5. Press the REL button to exit relative mode. 6. Switch the power OFF after
use.
Measurement
DANGER
Risk of electric shock! Never exceed the maximum permitted input values for
this product. Use caution when circuits may contain voltages above 30 V/AC rms
(42.4 V peak), 60 V/DC.
IMPORTANT! Always observe the information contained in the section: Safety
instructions [} 41].
49
10.1 Current measurement “A”
Notes: For best results, conductors should be centered in the clamp. The
current sensing clamp is magnetized and a low reading may appear
even when no conductor has been covered. Only one conductor should be
covered by the current sensing clamp.
Tips: For low currents, you can increase the field strength to a measurable
level: 1. Wind the conductor around one side of the current clamp as shown. 2.
Take note of the measured current. 3. Divide the measured current by the
number of winds (coils). à The real current has been derived.
50
10.1.1 Alternating current (AC) measurement
Risk of electric shock! Do not use the clamp on uninsulated conductors.
Unclamp the conductor if (overload) appears on the display. This product is
rated for 50-60Hz. Do not exceed this frequency
range as higher frequencies can dangerously overheat the magnetic circuit.
1. Select a range using the rotary switch: 4A~, 40A~.
2. (Optional step) press and hold the SELECT [V.F.C] button to switch the
low-pass filter ON/OFF. The display will show “VFC” when it is active.
3. Clamp the jaws around the conductor to be measured. à The reading will
show on the display.
4. Carefully remove the clamp from the conductor after taking measurements.
5. Switch the power OFF after use.
Notes: Low-pass filter “VFC”: This software filter is only applicable to AC
ranges. The low-pass filter suppresses noise above 400 Hz.
51
10.1.2 Direct current (DC) measurement
Risk of electric shock! Do not use the clamp on uninsulated conductors.
Unclamp the conductor if (overload) appears on the display.
Notes: For direct current measurements, the polarity of the clamp must match
the
flow of current along the conductor. A minus sign “-” will appear in front of
the reading if the polarities are reversed. Polarity symbols / are indicated
on the front and back of the jaw.
1. Select a range using the rotary switch: 4A , 40A .
2. Press the SELECT button to enter DC mode. à The display will show “DC”.
3. Press the ZERO button to perform a zero adjustment. à The display will
show “ZERO”.
4. Clamp the jaws around the conductor to be measured. à The reading will
show on the display.
5. Carefully remove the clamp from the conductor after taking measurements.
6. Switch the power OFF after use.
52
10.2 Voltage measurement “V”
WARNING: Observe all safety precautions when working on live voltages. 10.2.1
AC voltage measurement
1. Set the rotary switch to: V~. à The display will show “AC” and “V”.
2. (Optional step) press and hold the SELECT [V.F.C] button to switch the
lowpass filter ON/OFF. The display will show “VFC” when it is active.
3. Connect the test leads to the input terminals: – Black test lead to the
negative input terminal: COM. – Red test lead to the positive input terminal:
V.
4. Touch the test probe tips across the circuit or component under test. It
may take a few moments for the reading to stabilize. à The measurement will
show on the display. à The display will show icon (overload) if the measuring
range is exceeded or the circuit is interrupted.
5. Disconnect the test leads and switch the power OFF after use.
53
Notes: Low-pass filter “VFC”: This software filter is only applicable to AC
ranges. The low-pass filter suppresses noise above 400 Hz. 10.2.2 DC voltage
measurement
1. Set the rotary switch to: V . 2. Press the SELECT button to enter DC mode.
à The display will show “DC” and “mV”. 3. Connect the test leads to the input
terminals:
– Black test lead to the negative input terminal: COM. – Red test lead to the
positive input terminal: V. 4. Touch the test probe tips across the object
under test (e.g., circuit, battery). It may take a few moments for the reading
to stabilize. à The measurement will show on the display. à The display will
show icon (overload) if the measuring range is ex-
ceeded or the circuit is interrupted. à A minus sign “-” will appear in front
of the reading if the polarities are re-
versed. 5. Disconnect the test leads and switch the power OFF after use. 54
10.3 Resistance measurement
WARNING: Never test on a live circuit. Remove all power from the circuit and
discharge all capacitors before testing.
1. Set the rotary switch to: . à The display will show “k” or “M” + “.
2. Connect the test leads to the input terminals: – Black test lead to the
negative input terminal: COM – Red test lead to the positive input terminal: .
3. Touch the test probe tips across the circuit or component under test. It
may take a few moments for the reading to stabilize. à The measurement will
show on the display.
4. Disconnect the test leads and switch the power OFF after use. Tips:
Check the test leads for continuity by touching them together. The resistance
value should be approx. 0.5 (inherent resistance of the test leads). For low
resistance measurements (<400 ohms), you can deduct the inherent resistance of
the test leads from the measurements. See section: Relative mode [} 49].
55
10.4 Capacitance measurement
WARNING: Never test on a live circuit. Remove all power from the circuit and
discharge all capacitors before testing.
1. Set the rotary switch to: . 2. Press the SELECT button until the display
shows “n” and “F”. 3. Connect the test leads to the input terminals:
– Black test lead to the negative input terminal: COM. – Red test lead to the
positive input terminal: . 4. Touch the test probe tips across the capacitor
under test. It may take a few moments for the reading to stabilize. à The
measurement will show on the display. à The display will show icon (overload)
if the measuring range is ex-
ceeded or the circuit is interrupted. 5. Disconnect the test leads and switch
the power OFF after use.
56
Tips: If the capacitance being measured is 1µF, perform a zero adjustment to remove stray capacitance in the test leads and internal circuitry. This will improve measurement accuracy. 1. Enter capacitance measurement mode. 2. Press the ZERO button, the display will show “ZERO”. à A zero adjust has been performed.
Testing
DANGER
Risk of electric shock! Never exceed the maximum permitted input values for
this product. Use caution when circuits may contain voltages above 30 V/AC rms
(42.4 V peak), 60 V/DC. IMPORTANT! Always observe the information contained in
the section: Safety instructions [} 41].
57
11.1 Continuity test
WARNING: Never test on a live circuit. Remove all power from the circuit and
discharge all capacitors before testing.
1. Set the rotary switch to: . 2. Press the SELECT button until the display
shows ” “. 3. Connect the test leads to the input terminals:
– Black test lead to the negative input terminal: COM. – Red test lead to the
positive input terminal: . 4. Touch the test probe tips across the circuit or
component under test. à A constant tone will sound if the resistance is <10 .
à The display will show icon (overload) if the measuring range is ex-
ceeded or the circuit is interrupted. 5. Disconnect the test leads and switch
the power OFF after use.
58
11.2 Diode test
WARNING: Never test on a live circuit. Remove all power from the circuit and
discharge all capacitors before testing.
Forward test: Reverse test:
1. Set the rotary switch to: .
2. Press the SELECT button until the display shows ” “.
3. Connect the test leads to the input terminals: – Black test lead to the
negative input terminal: COM. – Red test lead to the positive input terminal:
.
4. Touch the test probe tips across the diode under test.
à Forward test voltages: will show on the display as a voltage reading.
à Reverse test voltages: will show on the display as ” “.
à A defective diode (interrupted): will show on the display as ” ” .
5. Disconnect the test leads and switch the power OFF after use.
59
11.3 Non-contact voltage testing
The non-contact voltage (NCV) detection function can detect AC voltage on
conductors without touching them. Due to the high sensitivity sensor, static
electricity or other sources of energy may trigger the sensor. This is normal
operation.
Risk of electric shock! Before use, always test the voltage detector on a
known live circuit
to ensure safe operation. Insulation type, thickness, and distance from the
voltage source
may affect detection. Always verify measurements using test leads before
touching po-
tentially energized circuits.
1. Set the rotary switch to: NCV. à “NCV” and “EF” will appear on the
display.
2. Place the sensor tip near the conductor. à If AC voltage is present, the
tri-colour LED will light and a beep will sound.
3. If the detected voltage increases: à the beeping will get faster. à the
tri-colour LED changes from green yellow red.
4. Switch the power OFF after use.
60
Troubleshooting
Problem
Possible cause
3x beeps then power OFF
Batteries are low
Display shows: “ErrE” System error
Large idle value after per- Remanence (residual
forming DC zero adjust- magnetism of the meas-
ment
uring coil)
Suggested solution See section: Replacing batteries [} 47] Restart the product
Briefly switch to the AC measuring function (A)
Cleaning
Risk of electric shock! Disconnect the product from any input signals and switch the power OFF before cleaning.
Important:
Do not use aggressive cleaning agents, rubbing alcohol or other chemical
solutions. They damage the housing and can cause the product to malfunction.
Do not immerse the product in water.
Clean the product with a clean, lint-free, antistatic cloth. Lightly moisten
if needed.
Check to make sure the clamp mating surface is clean. Trapped dirt or debris
can result in measurement errors.
61
Disposal
14.1 Product
This symbol must appear on any electrical and electronic equipment placed on
the EU market. This symbol indicates that this device should not be disposed
of as unsorted municipal waste at the end of its service life. Owners of WEEE
(Waste from Electrical and Electronic Equipment) shall dispose of it
separately from unsorted municipal waste. Spent batteries and accumulators,
which are not enclosed by the WEEE, as well as lamps that can be removed from
the WEEE in a non-destructive manner, must be removed by end users from the
WEEE in a non-destructive manner before it is handed over to a collection
point.
Distributors of electrical and electronic equipment are legally obliged to
provide free take-back of waste. Conrad provides the following return options
free of charge (more details on our website): in our Conrad offices at the
Conrad collection points at the collection points of public waste management
authorities or the collection
points set up by manufacturers or distributors within the meaning of the
ElektroG End users are responsible for deleting personal data from the WEEE to
be disposed of. It should be noted that different obligations about the return
or recycling of WEEE may apply in countries outside of Germany.
14.2 (Rechargeable) batteries
Remove batteries/rechargeable batteries, if any, and dispose of them
separately from the product. According to the Battery Directive, end users are
legally obliged to return all spent batteries/rechargeable batteries; they
must not be disposed of in the normal household waste.
62
Batteries/rechargeable batteries containing hazardous substances are labelled
with this symbol to indicate that disposal in household waste is forbidden.
The abbreviations for heavy metals in batteries are: Cd = Cadmium, Hg =
Mercury, Pb = Lead (name on (rechargeable) batteries, e.g. below the trash
icon on the left). Used (rechargeable) batteries can be returned to collection
points in your municipality, our stores or wherever (rechargeable) batteries
are sold. You thus fulfil your statutory obligations and contribute to
environmental protection. Batteries/rechargeable batteries that are disposed
of should be protected against short circuit and their exposed terminals
should be covered completely with insulating tape before disposal. Even empty
batteries/rechargeable batteries can contain residual energy that may cause
them to swell, burst, catch fire or explode in the event of a short circuit.
63
Technical data
15.1 General
Power supply………………………. 2x AAA 1.5 V batteries Measuring categories …………… CAT
III 600 V Current measurement…………… max. 40 A (AC/DC) Display………………………………..
4000 counts (2 3 renewals/sec.) Clamp opening range …………… width: max. 5.5
mm, height: max. 13 mm Automatic shut-off………………… approx. 15 mins Display
type………………………… EBTN Operating altitude ………………… max. 2000 m (above sea level)
Operating conditions…………….. 0 to +28 °C, 95 % RH (non-condensing)
+28 to +40 °C, 75 % RH (non-condensing) >+40 to +50 °C, 45 % RH (non- condensing) Storage conditions……………….. -10 to +50 °C, 80 % RH (non-condensing) Dimensions (W x H x D) ……….. 188 x 51 x 36 mm Weight ……………………………….. approx. 173 g
15.2 Test leads and probes
Rated voltage ……………………… CAT III 1000 V Rated current………………………. 10 A Protection class …………………… II
64
15.3 Specifications
15.3.1 Accuracy Specified accuracy ± (% of the reading + display error in
counts). Accuracy is maintained for 1 year at +23 °C (± 5 °C), 75 % RH (non-
con-
densing). Temperature coefficient: +0.1 x (specified accuracy) / 1 °C. To
avoid impaired measurements, do not operate the product in areas with
electromagnetic field strengths >1 V/m (±5 %).
15.3.2 Calibration The recommended calibration interval is 1 year.
Calibration should only be performed by qualified personnel.
15.3.3 Alternating current (AC)
Range
Resolution
Accuracy*
4.000 A
0.001 A
±(4.0 % +10) with a low-pass filter (VFC) ±(6.0 % + 20)
40.00 A
0.01 A
±(4.0 % +9) with a low-pass filter (VFC) ±(6.0 % + 20)
*Accuracy: 5 – 100 % of the measuring range Overload protection: 600 V 40 A Frequency range: 50-60 Hz Permitted display with unused measuring input: 5 counts (clamp closed)
True RMS Crest factor (CF) for non-sinusoidal signals: max. 3.0: Crest factor 1.0 2.0 : +4 % deviation Crest factor 2.0 2.5 : +5 % deviation Crest factor 2.5 3.0 : +7 % deviation
65
15.3.4 Direct current (DC)
Range 4.000 A
Resolution 0.001 A
Accuracy* ±(3.5 % +9)
40.00 A
0.01 A
±(3.5 % +5)
*Accuracy: 5 – 100 % of the measuring range; after successful zero adjustment Overload protection: 600 V, 40 A
15.3.5 AC voltage
Range
Resolution
Accuracy*
4.000 V
0.001 V
±(1.2 % +5)
40.00 V
0.01 V
±(1.5 % +5)
400.0 V
0.1 V
±(1.5 % +5)
with a low-pass filter
(VFC) ±(4.0 % + 10)
600 V
1 V
±(2.0 % +5)
with a low-pass filter
(VFC) ±(4.0 % + 10)
*Accuracy: 5 – 100 % of the measuring range Overload protection: 600 V Frequency range: 45-400 Hz Impedance: 10 M Permitted display with unused measuring input: 5 counts (clamp closed)
True RMS Crest factor (CF) for non-sinusoidal signals: max. 3.0: Crest factor 1.0 2.0 : +4 % deviation Crest factor 2.0 2.5 : +5 % deviation Crest factor 2.5 3.0 : +7 % deviation
66
15.3.6 DC voltage
Range 400.0 mV
Resolution 0.1 mV
Accuracy* ±(0.8 % +8)
4.000 V
0.001 V
±(1.2 % +5)
40.00 V
0.01 V
±(1.2 % +5)
400.0 V
0.1 V
±(1.2 % +5)
600 V
1 V
±(1.5 % +5)
*Accuracy: 5 – 100 % of the measuring range Overload protection: 600 V Impedance:10 M Permitted display with unused measuring input: 5 counts (clamp closed)
15.3.7 Resistance
Range
Resolution
Accuracy*
400.0
0.1
±(1.2 % +5)
4.000 k
0.001 k
±(1.0 % +5)
40.00 k
0.01 k
±(1.2 % +5)
400.0 k
0.1 k
±(1.2 % +5)
4.000 M
0.001 M
±(1.2 % +5)
40.00 M
0.01 M
±(2.0 % +5)
*Accuracy: 5 – 100 % of the measuring range, <400 after successful zero adjustment
Overload protection: 600 V
Measuring voltage: approx. 0.5 V
67
15.3.8 Capacitance
Range
Resolution
Accuracy*
4.000 nF
0.001 nF
±(4.0 % +10)
40.00 nF
0.01 nF
±(4.0 % +5)
400.0 nF
0.1 nF
±(3.0 % +5)
4.000 µF
0.001 µF
±(3.0 % +5)
40.00 µF
0.01 µF
±(3.0 % +5)
100.0 µF
0.1 µF
±(5.0 % +10)
*Accuracy: 10 – 100 % of the measuring range, 1 µF after successful zero adjustment Overload protection: 600 V Measuring voltage: approx. 0.5 V Permitted display with unused measuring input: 5 counts (clamp closed)
15.3.9 Diode test
Test voltage Approx. 3.2 V Overload protection: 600 V Test current: 1.7 mA
15.3.10 Acoustic continuity text
Resolution 0.001 V
Test voltage
Approx. 1 V Overload protection: 600 V Measuring range: max. 400 Test current:
<0.4 mA
Resolution 0.1
68
15.3.11 Non-contact (AC) voltage test
Test voltage 100 V/AC Frequency: 50-60 Hz
Distance max. 5 mm
References
- France.fr : Explore France and its wonders - Explore France
- RAD: Ethernet Access, Data Communications, Industrial IoT & Edge Computing
- Conrad Electronic » Alle Teile des Erfolgs
- Conrad Electronic » All parts of success
- Conrad Electronic – Alle Teile des Erfolgs
- Conrad Electronic » All parts of success
- Contact
- Download center
- Conrad Electronic Online Shop » Alle Teile des Erfolgs
- Conrad Electronic » All parts of success
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>