SUNSET HEALTHCARE SOLUTIONS VEN160100AM Closed Suction System for Adults Instruction Manual

June 10, 2024
SUNSET HEALTHCARE SOLUTIONS

VEN160100AM Closed Suction System for Adults
Instruction Manual

General Instructions for Use

Sunset Deluxe Closed Suction System is intended for endotracheal or tracheostomy suction of ventilator dependent adult patients. It is operated by health professionals in medical institutions for single patient use within 24 hours.

  • Maximum period of use is 24 hours and it is indicated 24 hrs on the pack.
  • For the series used with tracheostomy tubes, the length of catheter is 305mm and is indicated on the pack.

Contraindications

The device intends to suck the secretions from the patients with artificial airways. The pulmonary secretions from a patient’s artificial airways should be removed to prevent its obstruction, otherwise, they cause the patient’s death or to be lethal. Therefore, there is no absolute contraindication to the closed suction system.

Size

Product Code| Catheter Length| Remark| Outer Diameter| Packaging| ISO color
---|---|---|---|---|---
VEN160100A / VEN160100AM| 540mm| Endotracheal| 10FR /3.33 mm| 20ea/box|
VEN162100A / VEN162100AM
VEN161100A / VEN161100AM| 305mm| Tracheostomy
VEN163100A / VEN163100AM
VEN160120 / VEN160120M| 540mm| Endotracheal| 12FR /4.0     mm| 20ea/box|
VEN162120 / VEN162120M
VEN161120 / VEN161120M| 305mm| Tracheostomy
VEN163120 / VEN163120M
VEN160140 / VEN160140M| 540mm| Endotracheal| 14FR /4.67 mm| 20ea/box|
VEN162140 / VEN162140M
VEN161140 / VEN161140M| 305mm| Tracheostomy
VEN163140 / VEN163140M
VEN160160 / VEN160160M| 540mm| Endotracheal| 16FR / 5.3 mm| 20ea/box|
VEN162160 /VEN162160M
VEN161160 / VEN161160M| 305mm| Tracheostomy
VEN163160 / VEN163160M

M = with MDI port

EN / Caution EN / Sterilized using ethylene oxide
EN / Contains or presence of phthalate-DINP EN / Sale by or on the order of

a physician
EN / Does not contain natural rubber latex| EN / Manufacturer
EN / Do not use if package is damaged| EN / Date of manufacture
EN / Do not reuse| EN / Catalog number
EN / Do not desterilize| EN / Batch code
EN / Do not use in MRI| EN / Use by date
EN / Consult instructions for use| EN / Keep dry
EN / Temperature limitation
Upper limit: 49°C
Lower limit: 5°C| EN / Keep away from sunlight

Warning

  1. Do not use endotracheal length catheter in patient with tracheostomy tube. The extra length may cause mucosal tissue injured. Do not use tracheostomy length catheters for endotracheal tubes. The short length will result in insufficient suctioning.
  2. This is single use product. Do not reuse or desterilize this device. Reuse and desterilization may lead to the device not preforming as intended, malfunction or create a risk of cross infection.
  3. Do not CUT endotracheal tube while Closed Suction System is attached. Otherwise, the suction catheter may be cut and the cut portion of catheter may be aspirated into lower respiratory tract.

Caution

  1. Check the product package before use. Do not use if product packaging has been damaged.
  2. Withdraw suction catheter from the patients until the black marking can be seen on the distal side of patient-end adaptor.
  3. Do not pull suction catheter out of patient-end adaptor.
  4. The product should be handled by trained and qualified clinicians.
  5. Rx Only – Federal law restricts this device to sale by or on the order of a physician.
  6. For series used with tracheostomy tubes, the length of catheter is 305mm and is indicated on the pack.
  7. Do not leave catheter within the airway, it will cause increased airway resistance.
  8. Always place the vacuum control device in the locked position when not in use.
  9. Recommended vacuum settings 80 to 120mm/Hg.
  10. The entire suction procedure should last no longer than 10 to 15 seconds.
  11. When used in conjunction with other manufacture’s products always refer to their instruction for use.
  12. Excess fluid in heat and moisture exchangers (HMEs) or heat moisture exchanging filters (HMEFs) may increase resistance to flow. Follow manufacturer’s instructions and replace HME(F).

Pre-Use Check

  • There are no standard connections for suction tubing, hospitals should qualify the tubing fits the series of Sunset Deluxe Closed Suction Systems.
  • Inspect the package before opening. Ensure the product has not been compromised.
  • Select the appropriate catheter length and size following clinical guideline. It is recommended that the catheter size is no more than half the size of patient’s artificial airway.

Suggested Suction Procedure

SUNSET HEALTHCARE SOLUTIONS VEN160100AM Closed Suction System for Adults -
Fig

Suggested Suction Procedure

  1. (Fig. 1) Attach suction tubing to the connector on the vacuum control device (G ). Connect suction tubing to the hospital
    vacuum system.

  2. (Fig. 1) Attach a day label stick to the vacuum control device indicating the date of use.

  3. (Fig. 1) Attach the 15 mm male connector of the patient end adaptor (B) to the breathing system (machine-end); the 15 mm female connector (C) to the patient’s artificial airway (patient-end) .

  4. (Fig. 1) The included corrugated connector (L )may be attached between the 15 mm male connector (B )and the breathing system if needed.

  5. (Fig. 2) Unlock the cover of the vacuum control device(H ).

  6. (Fig. 2) Hold the patient end adaptor with one hand. Carefully advance the catheter down the airway with the other hand.

  7. Advance catheter(D )to desired depth. For endotracheal tube type, line up the marking numbers. There are no markings on tracheostomy tubes. Follow hospital procedures when advancing the catheter.

  8. (Fig. 3) Depress and hold the vacuum control device(H ).

  9. (Fig. 4) Then gently withdraw catheter(D )until the black ring(F )is visible on the distal side of the patient-end adaptor.

  10. Repeat steps 6-8 above as necessary.

Catheter Irrigation

  1. (Fig. 5) Open cap on irrigation port (J ). A prefilled saline vial or a 15ml syringe filled with saline is prepared to be used.
  2. (Fig. 5) Attached saline vial/syringe to the irrigation port.
  3. Depress the cover of vacuum control device (H )for saline to ender the irrigation chamber (A ). Continue to irrigate until the catheter is clear of secretion.
  4. Close the cap of irrigation port (J ).
  5. (Fig. 6) Lock the cover of vacuum control device (H )when not in use to prevent accidental depression.
  6. Turn off the vacuum after use. If the suction connection parts are disconnected, close cap (I )on the end of vacuum control device (G ).

Metered Dose Inhaler (MDI)

  1. Remove canister and close cap of MDI port (K ).
  2. Remove cap of MDI port (K ), and attach canister. Avoid discharge of canister when connecting.
  3. Hold canister in vertical position. Depress canister during inspiration cycle. Repeat as prescribed by physician or protocol.

Disconnection

  1. (Fig. 7) Disconnect the Sunset Deluxe closed suction system from the patient’s airway by the disconnect wedge(M ).
  2. Set the disconnect wedge (M )between the airway connector and 15mm female connector (C )of closed suction system.
  3. Lever up the wedge from the patient end adaptor to disconnect the closed suction system.
  4. After use, the product must be disposed in conformance with the local disposal protocols and regulation of hygiene and medical waste.

Manufactured for Sunset Healthcare Solutions
180 N Michigan Ave Ste 2000
Chicago, IL 60601
United States
www.SunsetHCS.com
Made in Taiwan

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

SUNSET HEALTHCARE SOLUTIONS User Manuals

Related Manuals