SUNSET HEALTHCARE SOLUTIONS HFN001AS Humidoflo High Flow Nasal Cannula Instruction Manual
- June 10, 2024
- SUNSET HEALTHCARE SOLUTIONS
Table of Contents
HFN001AS
Humidoflo High Flow Nasal Cannula
Instruction Manual
HFN001AS Humidoflo High Flow Nasal Cannula
HFN001AS • HFN001AM • HFN001AL
Intended use
Humidiflo High Flow Nasal Cannula (Non-sterile) is intended to provide high
flow oxygen therapy for adult patients with acute respiratory failure. The
Humidoflo Cannula is designed to deliver air,with or without blended oxygen,
up to a maximum flow of 60 l/min. The Humidoflo Cannula is for single- patient
use in the hospital/clinical environment.
Indication of this product, high flow nasal cannula: Since the Humidoflo
Cannula is noninvasive, PEEP (CPAP) is not measured. Despite these issues, a
growing body of evidence suggests that Humidoflo Cannula is an effective
modality for early treatment of patient with respiratory failure associated
with diverse underlying diseases.
Contraindications
High flow oxygen therapy may generate some positive airway pressure.
Contraindications for positive airway pressure devices apply: pneumothorax,
pathologically low blood pressure, cerebrospinal fluid leak, recent cranial
surgery or trauma, severe bullous lung disease, dehydration.
Fitting To change the tubing attachment side WARNINGS
- Do not reuse.
- Only use in approved configurations.
- Check for condensation regularly. Drain as required.
- Oxygen supports combustion. Do not use in the presence of an open flame.
- Do not use if components are damaged.
- Risk of minor nasal mucosa irritation and minor bleeding due to gas flow and mechanical contact.
- Recommended disposal after 7 days of use.
CAUTION
- Before connecting the nasal cannula check for adequate gas flow, oxygen and temperature.
- Ensure the nasal cannula is sized correctly and does not seal nares.
Technical specifications
Setup: For use with high flow oxygen therapy humidification systems which
deliver a warm, humid air flow ≤ 37°C (≤98°F) and ≤ 100% r.h. with a maximum
oxygen content of 100% O2 and a flow of up to 60 L/min.
Connection: 22/15 mm Conical Connector (complies with ISO 5356-1).
Flow range: ≤ 60 L/min
Environmental temperature range during use: 65-78°F / 18-26°C
Disposal
This nasal cannula does not contain any hazardous substances and may be
disposed of in accordance with local environmental regulations.
Manufactured for Sunset Healthcare Solutions
180 N Michigan Ave Ste 2000 • Chicago IL 60601
MADE IN TAIWAN
IFU-00068 REV 1.00
2022-04-07
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>