SUNSET HEALTHCARE SOLUTIONS TUB06HPR Heated CPAP Tube Instruction Manual
- June 10, 2024
- SUNSET HEALTHCARE SOLUTIONS
Table of Contents
SUNSET HEALTHCARE SOLUTIONS TUB06HPR Heated CPAP Tube
Introduction
The Sunset Heated CPAP Tube is designed to prevent condensation of humidified
air during CPAP treatment, also known as “rainout.” When combined with a
heated humidifier, the Sunset Heated CPAP Tube controls humidification by
ensuring that air temperature does not cool prior to inhalation. This supports
comfortable and consistent treatment.
The Sunset Heated CPAP Tube is manufactured exclusively for use with the
Philips Respironics DreamStation and Philips Respironics System One.
Understanding the Heated CPAP Tube
Components
- Humidifier connector cuff
- Connector lead
- Squeeze release tabs
- Mask-end cuff
- Flexible tubing
Indications for Use
The Sunset Heated CPAP Tube is a heated wire breathing tube intended to provide warmed and/or humidified breathing gases before they enter a patient’s airway. It is indicated for single-patient reuse in the home and in clinical settings, such as hospitals, institutions, sleep laboratories, and subacute care facilities. It may be used with non-invasive ventilation for adult patients. It is compatible with the Philips Respironics System One Heated Humidifier and Philips Respironics Dream Station Heated Humidifier.
Warnings and Cautions
Please retain these instructions for use for future reference.
- WARNING: The Sunset Heated CPAP Tube is validated for use with two specific humidifiers: Philips Respironics Dream Station and Philips Respironics System One. Do not use the Sunset Heated CPAP Tube in combination with other Humidifiers.
- WARNING: The Sunset Heated CPAP Tube used with incompatible parts can result in degraded performance.
- WARNING: Failure to use a mask or accessory that minimizes rebreathing of carbon dioxide or permits spontaneous breathing can cause asphyxiation.
- WARNING: The user is accountable for the compatibility of the Sunset Heated CPAP Tube and the humidifier and all of the parts and accessories used to connect to the patient before use.
- WARNING: Always connect-disconnect the tube from the humidifier by using the grips on the cuff. Do not pull directly on the tube.
- WARNING: To prevent disconnection of the tubing or tubing system during use, only use accessories in compliance with ISO 5367 or ISO 80601-2-74.
- WARNING: Do not add any attachments to the Sunset Heated CPAP Tube, otherwise, the humidifier and its accessories might not function correctly, affecting the quality of the therapy or cause injury to the patient.
- WARNING: When used routinely, check the tube daily for holes, deformation, or any other form of damage. When damaged, replace the tube immediately.
- WARNING: Operating the Sunset Heated CPAP Tube close to other heat sources including lamps, electric blankets, etc., can cause elevated airflow temperatures that can affect the quality of the therapy or cause injury to the patient
- WARNING: The tube shall be replaced after 90 days of use. Using the Sunset Heated CPAP Tube after the maximum 90 days duration of use can affect the quality of therapy or cause injury to the patient.
- WARNING: Do not sterilize the Sunset Heated CPAP Tube.
- WARNING: The Sunset Heated CPAP Tube is for single patient reuse only. Using the Sunset Heated CPAP Tube on multiple patients can lead to contamination and injure the patient.
- WARNING: Do not subject the tube to pressure, crushing or other mechanical stresses, as damage to the tube can affect the quality of the therapy or cause injury to the patient.
- WARNING: Replace damaged (having holes, kinks, tears, exposed heating wires, etc. or is not functional) tube immediately. Do not try to repair the tube yourself. Damaged CPAP tubes can affect quality of the therapy or injure the patient.
- WARNING: Using brushes or objects to clean the inside of the tube can damage the tubing and can affect the quality of the therapy or cause injury to the patient. See cleaning instructions.
- WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is required, this equipment and the other equipment should be observed to verify normal operation.
- WARNING: Use of accessories, transducers, and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and resulting in improper operation.
- WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Sunset Heated CPAP Tube, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
- WARNING: MR Unsafe and should remain outside the MRI scanner room.
- CAUTION: Dispose of the Heated CPAP Tube in accordance with your local government regulations. Upon permanent disposal, mark the tube so that it cannot be re-used by another patient.
- CAUTION: When using electrical products, always follow applicable safety precautions.
- CAUTION: The Sunset Heated CPAP Tube complies with applicable Radiated RF Emission levels to prevent adverse events to the patients and operator due to electromagnetic disturbances. In order to prevent adverse events to patient and operator due to electromagnetic disturbances do not modify or change The Sunset Heated CPAP Tube.
- CAUTION: Patients undergoing CPAP therapy may be susceptible to airway infections. Single patient reuse devices that are not replaced within the prescribed frequency may be a potential source of infection. Proper hand hygiene (e.g. hand washing or hand disinfection) is mandatory before and after replacement of the device.
- CAUTION: Do not use (microwave) ovens, fans, heating devices or any other appliances to dry the tube after washing. See cleaning instructions.
Use and Care
Directions for Use
The Sunset Heated CPAP Tube features a Mask-end Cuff and a Humidifier
Connector Cuff.
Before connecting, remove Sunset Heated CPAP Tube from packaging. Inspect tube for any damage. Discard and replace the tubing if damaged.
To Connect
Align the Sunset Heated CPAP Tube Humidifier Connector Cuff with the air
outlet port on the CPAP machine humidifier. Place the Connector Lead into the
mating connector located at the top of the humidifier air outlet port. Press
until the tabs on each side of the connector click into place.
Connect the Mask-end Cuff to your mask connector. For proper placement, please consult instructions included with your mask. Ensure there is airflow through the tubing.
Apply mask and use as directed by your physician or respiratory therapist, and in accordance with the instructions in your CPAP or mask ventilator system user manual.
For extended directions for use, please refer to your DreamStation or PR System-One CPAP humidifier manual.
To Disconnect
Lightly squeeze Release Tabs on side of Humidifier Connector Cuff and
carefully disconnect from air outlet port.
Caution: To avoid circuit leaks, do not pull or stretch tubing.
Home Cleaning & Disinfection Home Cleaning
Clean Sunset Heated CPAP Tube before first use and daily during use.
- Disconnect Heated CPAP Tube from CPAP machine and mask.
- Thoroughly rinse inside and outside of Heated CPAP Tube with cold tap water to remove excess soil.
- Gently wash Heated CPAP Tube and connectors by immersing ina solution of warm water and mild unscented liquid dish soap, gently agitating by hand and using a soft bristle brush to remove contaminants.
- Thoroughly rinse Heated CPAP Tube with clean water to remove all soap residue.
- With connector cuffs directed down, toward the floor, gently tap connector cuffs to remove excess water.
- Hang Heated CPAP Tube with connector cuffs directed downward, towards the floor, for drying.
- Allow Heated CPAP Tube to air dry thoroughly prior to next use.
- Visually inspect Heated CPAP Tube for cleanliness. Repeat cleaning process if not visually clean.
- Inspect Heated CPAP Tube for damage or wear. Discard and replace if necessary
Caution: Do not use harsh soaps, detergents, solvents, or alcohols to clean
the Sunset Heated CPAP Tube. Do not attempt to clean Heated CPAP Tube in a
dishwasher, ultrasonic cleaner or any other cleaning device.
Only reuse the Heated CPAP Tube when fully dry.
Home Disinfection
The disinfection procedure described is recommended for a maximum of 60
cycles.
Thermal Disinfection (Pasteurization)
Immerse Heated CPAP Tube in 75°C ± 2°C (167°F) water bath for 30 minutes.
Post Disinfection
Repeat steps 5-9 of the Home Cleaning procedure prior to use.
Institutional Cleaning & Disinfection Institutional Cleaning Prior to
Disinfection
Clean Sunset Heated CPAP Tube before first use and daily during use.
-
Disconnect Heated CPAP Tube from CPAP machine and mask.
-
Thoroughly rinse inside and outside of Heated CPAP Tube with cold tap water to remove excess soil.
-
Gently wash Heated CPAP Tube using an institutionally approved enzymatic cleaning solution and a soft bristle brush to remove contaminants.
Caution: Be sure all hard to reach surfaces, corners and crevices have been thoroughly cleaned. -
Thoroughly rinse Heated CPAP Tube by submersing in clean water and vigorously agitating. Repeat as necessary until all residue has been removed.
-
With connector cuffs directed downward, toward the floor, gently tap connector cuffs to remove excess water.
-
Hang Heated CPAP Tube with connector cuffs directed downward, toward the floor, for drying.
-
Allow Heated CPAP Tube to air dry.
-
Visually inspect Heated CPAP Tube for cleanliness. Repeat cleaning process if not visually clean.
-
Inspect Heated CPAP Tube for damage or wear. Discard and replace if necessary.
Institutional Disinfection
The disinfection procedures described below are recommended for a maximum of
60 cycles.
Thermal Disinfection (Pasteurization)
Immerse Heated CPAP Tube in 75°C ± 2°C (167°F) water bath for 30 minutes.
Chemical Disinfection
Immerse device completely in CIDEX® OPA, filling all lumens and eliminating
air pockets, in CIDEX® OPA Solution for a minimum of 12 minutes at 20°C (68°F)
or higher to destroy all pathogenic microorganisms. Remove device from the
solution and rinse thoroughly following the rinsing instructions on next page.
Rinsing Instructions
Following removal from CIDEX® OPA Solution, thoroughly rinse the medical device by immersing it completely in a large volume (e.g. 8 liters) of sterile water.
- Keep the device totally immersed for a minimum of 1 minute.
- Manually flush all lumens with large volumes (not less than 100ml) of rinse water.
- Remove the device and discard the rinse water. Always use fresh volumes of water for each rinse. Do not reuse the water for rinsing or any other purpose.
- Repeat the procedure TWO (2) additional times, for a total of THREE (3) RINSES, with large volumes of fresh water to remove CIDEX® OPA Solution residues.
- Refer to the disinfectant solution manufacturer’s instructions for use for additional information and cautions.
Caution: Follow disinfectant solution manufacturer’s instructions for use. Failure to do so may result in inadequate disinfection. Review all manufacturer cautions and warnings.
Errors and Troubleshooting
Consult your CPAP machine user manual.
Additional Requirements
Environmental Operating Conditions
- Ambient Temperature: 41°F – 86°F (5°C – 30°C)
- Relative humidity: 5% – 95%
- Altitude: less than 6,561 feet (2,000 meters)
Storage and Transport Conditions
Environmental conditions for transport and storage:
This device should be transported and stored in its original packaging and protected from direct sunlight, dust, or any other form of contamination. . Temperature during storage and transport should not exceed 140°F (60°C) or fall below -4°F (-20°C) with Relative humidity can range from 5 % to 95%.
Frequently Asked Questions
Question : Can I connect the heated tube directly to the CPAP machine?
Answer : Yes, the tubing will act like a standard CPAP tube. However,
rainout might occur.
Question : When do I replace my Sunset Heated CPAP Tube?
Answer : The heated tubing is intended to be replaced every 3 months, or
according to your provider’s recommended schedule. If damaged, leaking or
malfunctioning, replace your heated tube immediately
Question : With what humidifiers or flow generators can I use my Sunset
Heated CPAP Tube?
Answer : The Sunset Heated CPAP Tube is designed specifically for use
with the Respironics Dream Station and Philips Respironics System One.
Replacement Components
Please ask your physician, respiratory therapist, or home care provider for advice on replacement schemes. The Sunset Heated CPAP Tube should not be used for longer than 3 months.
Product Specifications
Description | Specification |
---|---|
Sunset Heated CPAP Tube Length | 6ft (1.83m) |
Internal Diameter | 15mm |
Flow resistance @60 l min-1 | ≤ 0.1 kPa |
Tubing Compliance Standards | ISO 5367:2000 |
ISO 8185:2007
ISO 80601-2-70:2015
ISO 80601-2-74:2017
Water resistance| IP22
Compatible
CPAP Machine Humidifiers
|
- Philips Respironics Dream Station
- Philips Respironics System One
Periodic visual safety inspections that should be performed by the OPERATOR|
Daily
HUMIDIFIER classification according to the ISO 80601-2-74| Humidifier classification: 2
Standards Compliance
This device is designed to conform to the following standards:
- IEC 60601-1 General Requirements for Basic Safety and Essential Performance of Medical Electrical Equipment
- EN 60601-1-2 Electromagnetic Compatibility
- ISO 80601-2-70 Sleep Apnea Breathing Therapy Equipment
- ISO 80601-2-74 Respiratory humidifying equipment
IEC 60601-1 Classification
- Degree of Protection Against Electric Shock: Type BF Applied Part
- Degree of Protection against Ingress of Water: Drip Proof, IPx2
- Mode of Operation: Continuous
Electrical
- Power Consumption 3.3 Vdc, max. 5A.
Warranty
The Sunset Heated CPAP Tube warranty is 30 days under normal operating conditions. This is based on daily usage of 8 hours.
EMC Table
Radiated RF Emission levels
Phenomenon
| Immunity test levels
---|---
Professional healthcare facility environment| Home healthcare environment
Emissions| Compliance
Radiated RF Emissions CISPR 11| Group 1
Radiated RF Emissions CISPR 11| Class B
Immunity levels
Phenom- enon
| Immunity test levels
---|---
Professional healthcare facility environment| Home healthcare environment
Immunity Test| Compliance1
Electrostatic Discharge IEC
61000-4-2
| ± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Radiated Immunity IEC 61000- 4-3| 3 V/m 80 MHz-2,7 GHz 80% AM at 1 kHz| 10 V/m
80 MHz-2,7 GHz
80% AM at 1 kHz
Power Frequency Magnetic Field IEC 61000- 4-8| 30 A/m ,50 Hz or 60 Hz
Proximity fields from RF wireless communication equipment, Immunity test levels:
Frequency Range and Level: RF wireless communication equipment
Frequency (MHz)| Modulation| Immunity Level (V/m)
Test| Compliance 2
385| Pulse Modulation: 18Hz| 27
450| FM + 5Hz deviation: 1kHz sine| 28
710
745
780| Pulse Modulation: 217Hz| 9
810
870
930| Pulse Modulation: 18Hz| 28
1720
1845
1970| Pulse Modulation: 217Hz| 28
2450| Pulse Modulation: 217Hz| 28
5240
5500
5785| Pulse Modulation: 217Hz| 9
2 Compliance with immunity test levels specified in IEC 60601-1-2, without deviations.
Symbol Glossary
Symbol | Meaning |
---|---|
Attention, consult accompanying documents |
| Read instruction manual carefully
| Catalog item number
| Batch code
| Name and address of manufacturer
| Date of manufacture
| Not made with natural latex rubber
IP22
| Protected from touch by fingers and objects greater than 12 millimeters.
Protected from water spray less than 15 degrees from vertical.
| Keep away from direct sunlight
| Keep dry
| Do not use if the package is damaged
| MR Unsafe and should remain outside the MRI scanner room.
Standard References
ISO 15223-1:2012, Medical Symbols to be used with medical device labels, labeling, and information to be supplied – General requirements
IEC 60601-1, Medical electrical equipment – General requirements for basic safety and essential performance
**** Sunset Healthcare Solutions
180 N Michigan Ave Ste 2000
Chicago, IL 60601
www.sunsethcs.com
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>