velleman CAT III 300 V Digital Multimeter User Manual
- June 4, 2024
- Velleman
Table of Contents
- velleman CAT III 300 V Digital Multimeter
- Introduction
- Used Symbols
- General Guidelines
- Maintenance
- During Use
- General Description
- Overvoltage/Installation Category
- Pollution Degree
- Specifications
- Operation
- Battery and Fuse Replacement
- Troubleshooting
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
velleman CAT III 300 V Digital Multimeter
Introduction
To all residents of the European Union Important environmental information
about this product
This symbol on the device or the package indicates that disposal of the device
after its lifecycle could harm the environment. Do not dispose of the unit (or
batteries) as unsorted municipal waste; it should be taken to a specialized
company for recycling. This device should be returned to your distributor or
to a local recycling service. Respect the local environmental rules.
If in doubt, contact your local waste disposal authorities.
Thank you for choosing Velleman! Please read the manual thoroughly before
bringing this device into service. If the device was damaged in transit, do
not install or use it and contact your dealer.
Used Symbols
General Guidelines
Refer to the Velleman® Service and Quality Warranty on the last pages of this manual.
-
This symbol indicates: Read instructions
Not reading the instructions and manual can lead to damage, injury or death. -
This symbol indicates: Danger
A hazardous condition or action that may result in injury or death -
This symbol indicates: Risk of danger/damage
Risk of a hazardous condition or action that may result in damage, injury or death -
This symbol indicates: Attention; important information Ignoring this information can lead to hazardous situations.
-
WARNING: To avoid electrical shock always disconnect the test leads prior to opening the housing. To prevent fire hazards, only use fuses with the same ratings as specified in this manual. Remark: refer to the warning on the battery compartment
-
Avoid cold, heat and large temperature fluctuations. When the unit is moved from a cold to a warm location, leave it switched off until it has reached room temperature. This to avoid condensation and measuring errors.
-
Protect this device from shocks and abuse. Avoid brute force when operating.
-
Pollution degree 2-device. For indoor use only. Keep this device away from rain, moisture, splashing and dripping liquids. Not for industrial use. Refer to §8 Pollution degree.
-
Keep the device away from children and unauthorized users.
-
Risk of electric shock during operation. Be very careful when measuring live circuits.
-
There are no user-serviceable parts inside the device. Refer to an authorized dealer for service and/or spare parts.
-
This is an installation category CAT III measuring instrument. Refer to §7 Overvoltage/installation category.
Read this addendum and the manual thoroughly. Familiarise yourself with the functions of the device before actually using it.
All modifications of the device are forbidden for safety reasons. Damage caused by user modifications to the device is not covered by the warranty.
Only use the device for its intended purpose. Using the device in an unauthorized way will void the warranty. Damage caused by disregard of certain guidelines in this manual is not covered by the warranty and the dealer will not accept responsibility for any ensuing defects or problems.
Maintenance
There are no user-serviceable parts inside the device. Refer to an authorized dealer for service and/or spare parts. Before performing any maintenance activities, disconnect the test leads from the jacks. For instructions on replacing battery or fuse, refer to §11 Battery and fuse replacement. Do not apply abrasives or solvents to the meter. Use a damp cloth and mild detergent for cleaning purposes.
During Use
Risk of electric shock during operation. Be very careful when measuring live circuits.
- Never exceed the limit value for protection. This limit value is listed separately in the specifications for each range of measurement.
- Do not touch unused terminals when the meter is linked to a circuit which is being tested.
- Never use the meter with CAT II installations when measuring voltages that might exceed the safety margin of 500 V above earth ground. Never use the meter with CAT III installations when measuring voltages that might exceed the safety margin of 300 V above earth ground.
- Set the range selector at its highest position if the intensity of the charge to be measured is unknown beforehand.
- Disconnect the test leads from the tested circuit before rotating the range selector in order to change functions.
- When carrying out measurements on a TV set or switching power circuits, always remember that the meter may be damaged by any high amplitude voltage pulses at test points.
- Always be careful when working with voltages above 60 VDC or 30 VAC rms. Keep your fingers behind the probe barriers at all times during measurement.
- Never perform resistance, diode or continuity measurements on live circuits. Make sure all capacitors in the circuit are depleted.
General Description
Refer to the illustration on page 2 of this manual:
- Display 3 ½ digits, LCD: 32 x 14 mm
- Selection button
- Fixed test leads
Overvoltage/Installation Category
DMMs are categorized depending on the risk and severity of transient
overvoltage that might occur at the point of test. Transients are short-lived
bursts of energy induced in a system, e.g. caused by a lightning strike on a
power line.
The existing categories according to EN 61010-1 are:
CAT I
| A CAT I-rated meter is suitable for measurements on protected electronic
circuits that are not directly connected to mains power, e.g. electronics
circuits, control signals…
---|---
CAT II
| A CAT II-rated meter is suitable for measurements in CAT I- environments and mono-phase appliances that are connected to the mains by means of a plug and circuits in a normal domestic environment, provided that the circuit is at least 10 m apart from a CAT III- or 20 m apart from a CAT IV-environment. E.g. household appliances, portable tools…
CAT III
| A CAT III-rated meter is suitable for measurements in CAT I- and CAT II- environments, as well as for measurements on (fixed) mono- or poly-phased appliances which are at least 10 m apart from of a CAT IV-environment, and for measurements in or on distribution level equipment (fuse boxes, lighting circuits, electric ovens).
CAT IV
| A CAT IV-rated meter is suitable for measuring in CAT I-, CAT II- and CAT III-environments as well as on the primary supply level.
Note that for all measurements on equipment for which the supply cables run outdoors (either overhead or underground) a CAT IV meter must be used.
Warning:
This device was designed in accordance with EN 61010-1 installation category
CAT II 500 V and CAT III 300 V. This implies that certain restrictions in use
applications that are related to voltages and voltage peaks that can occur
within the environment of use. Refer to the table above this device is only
suitable for measurements up to 500 V in CAT II and up to 300 V in CAT III.
Pollution Degree
IEC 61010-1 specifies different types of pollution environments, for which different protective measures are necessary to ensure safety. Harsher environments require more protection, and the protection against the pollution which is to be found in a certain environment depends mainly on the insulation and the enclosure properties. The pollution degree rating of the DVM indicates in which environment the device may be used.
Pollution degree 1| No pollution or only dry, nonconductive pollution occurs.
The pollution has no influence. (only to be found in hermetically sealed
enclosures)
---|---
Pollution degree 2| Only nonconductive pollution occurs. Occasionally,
temporary conductivity caused by condensation is to be expected.(home and
office environments fall under this category)
Pollution degree 3| Conductive pollution occurs, or dry nonconductive
pollution occurs that becomes conductive due to condensation that is to be
expected.
(industrial environments and environments exposed to outside air – but not in contact with precipitation)
Pollution degree 4| The pollution generates persistent conductivity caused by conductive dust or by rain or snow. (exposed outdoor environments and environments where high humidity levels or high concentrations of fine particles occur)
Warning: This device was designed in accordance with EN 61010-1 pollution degree 2. This implies that certain restrictions in use apply that are related to pollution which can occur within the environment of use. Refer to the table above. This device is only suitable for measurements in Pollution degree class 2 environments.
Specifications
This device is not calibrated when purchased! Regulations concerning the environment of use: Use this meter only for measurements in CAT I, CAT II, and CAT III environments (see §7). Use this meter only in a pollution degree 2 environment (see §8).
Ideal working conditions include:
-
temperature: 0 °C to 40 °C (32 °F to 104 °F)
-
relative humidity: max. 80 %
-
altitude: max. 2000 m (6560 ft)
-
voltage ………………………………………………………………… 500 V
fuse protection -
F0.5 A/500 V, 5 x 20 mm
-
power supply ………………………………………….. 1 x CR2032 (incl.)
-
display …………………………………………………… LCD, 1999 counts
-
display dimensions ……………………………………………61 x 32 mm
-
over-range ……………………………………………………………….. yes
-
continuity buzzer ………………………………………………………… yes
-
diode test …………………………………………………………………. yes
-
low-battery indication ………………………………………………….. yes
-
ranging mode ………………………………………………………. manual
-
data hold ………………………………………………………………….. no
-
backlight …………………………………………………………………… no
-
auto power-off ……………………………………………………………. no
-
dimensions ……………………………………………….97 x 51 x 27 mm
-
weight (with battery) …………………………………………………100 g
-
storage environment
-
temperature …………………………………………. -20 °C to 60 °C
-
humidity …………………………………………………… < 90 % RH
-
test lead probe ………. CAT II 500 V / CAT III 300 V, 10 A; L = 70 cm
DC VOLTAGE
****| Do not measure circuits that may contain voltages
500 VDC or > 500 VAC
---|---
range| resolution| accuracy
200 mV| 0.1 mV| ****
± (1.0 % rdg + 2 digits)
2 V| 1 mV
20 V| 10 mV
200 V| 100 mV
500 V| 1 V| ± (1.2 % rdg + 2 digits)
- Overload protection: 500 V DC or AC rms
- Impedance: 1 MΩ
AC VOLTAGE
****| Do not measure circuits that may contain voltages
500 VDC or > 500 VAC
---|---
range| resolution| accuracy
200 V| 100 mV| ± (1.5 % rdg + 3 digits)
500 V| 1 V
Average sensing, calibrated to rms of sine wave
- Frequency range: 40-500 Hz
- Overload protection: 500 V DC or AC rms
- Impedance: 450 kΩ
DC CURRENT
****| Do not measure circuits that may contain voltages
500 VDC or > 500 VAC
---|---
range| resolution| accuracy
2000 µA| 1 µA| ± (1.5 % rdg + 2 digits)
20 mA| 10 µA
200 mA| 100 µA| ± (2.0 % rdg + 2 digits)
Overload protection: F0.5 A/500 V fuse
RESISTANCE
**** | Do not conduct resistance measurements on live circuits |
---|---|
range | resolution |
200 Ω | 0.1 Ω |
2000 Ω | 1 Ω |
± (1.0 % rdg + 3 digits)
20 kΩ| 10 Ω
200 kΩ| 100 Ω
2000 kΩ| 1 kΩ| ± (1.5 % rdg + 3 digits)
Overload protection: F0.5 A/500 V fuse
DIODE AND CONTINUITY
**** | Do not conduct diode or continuity measurements on live circuits |
---|---|
range | description |
**** | display reads the approximate forward voltage of the diode |
forward DC current ± 10 µA reversed DC voltage ± 1.8 V
****| built-in buzzer sounds if resistance < 50 Ω| open-circuit voltage ± 1.8 V
Overload protection: F0.5 A/500 V fuse
BATTERY TEST
range | load current | resolution | accuracy |
---|---|---|---|
1.5 V | 50 mA | 10 mV | ± (5.0 % rdg + 5 digits) |
9 V | 5 mA | 10 mV |
Operation
VOLTAGE MEASUREMENT
Do not measure circuits that may contain voltages > 500 VDC or > 500 VAC Use extreme caution when measuring voltages higher than 60 VDC or 30 VAC rms. Always place your fingers behind the protective edges of the test probes while measuring!
- Set the rotary switch in the desired V= or V~ position.
- Connect the test leads to the source being measured.
- Read the voltage value and the polarity of the red test lead on the LCD display.
Notes
- If the range is not known beforehand, set the selector switch to a high range and lower gradually.
- An over-range is indicated by 1 or -1. Set to a higher range.
DC CURRENT MEASUREMENT
Do not measure circuits that may contain voltages > 500 VDC or > 500 VAC Use extreme caution when measuring voltages higher than 60 VDC or 30 VAC rms. Always place your fingers behind the protective edges of the test probes while measuring!
- Set the rotary switch in the desired mA= position.
- Open the circuit in which the current is to be measured and connect the test leads to the circuit IN SERIES.
- Read the current value and the polarity of the red lead connection on the LCD display.
Notes
- If the range is not known beforehand, set the selector switch to a high range and lower gradually.
- An over-range is indicated by 1 or -1. Set to a higher range.
- The maximum input current is 200 mA.
RESISTANCE MEASUREMENT
Do not conduct resistance measurements on live circuits. Make sure all capacitors in the circuit are depleted.
- Set the rotary switch in the desired Ω position.
- Connect the test leads to the resistor to be measured and read the LCD display.
Notes
- If the resistance being measured is connected to a circuit, turn off the power and discharge all capacitors before applying the test probes.
- An over-range is indicated by 1 or -1. Set to a higher range.
DIODE AND CONTINUITY TEST
Do not conduct diode or continuity measurements on live circuits. Make sure all capacitors in the circuit are depleted.
- Set the rotary switch in the (diode) or (continuity) position.
- For measuring the diode, connect the red test lead to the anode of the diode to be tested and the black test lead to the cathode of the diode. The approx. forward voltage drop of the diode will be displayed. For measuring continuity, connect the test leads to two points of the circuit to be tested. If continuity exists, the built-in buzzer will sound.
BATTERY TEST
- Set the rotary switch in the 1.5V or 9V position.
- Connect the test leads to the battery to be tested and read the LCD display
Battery and Fuse Replacement
- WARNING: To avoid electrical shock always disconnect the test leads prior to opening the housing. To prevent fire hazards, only use fuses with the same ratings as specified in this manual. Remark: refer to the warning on the battery compartment.
- There are no user-serviceable parts inside the device. Refer to an authorized dealer for service and/or spare parts.
- Disconnect the test leads from the test points and remove the test leads from the measuring terminals before replacing the batteries or fuses.
- When” ” is displayed, the battery should be replaced.
- Fuses rarely need replacement and blown fuses almost always result from human error.
To replace the battery:
- Switch of the meter.
- Remove the screw at the back of the case and gently open the housing.
- Remove the old battery and insert a new one.
- Close the housing and fasten the screw.
- Battery: 1x CR2032, make sure to respect the polarity
- Fuses: F0.5 A/500 V, 5 x 20 mm
Make sure the meter is closed tight and put the protective edge back in place before using the meter.
Troubleshooting
If the device acts abnormal while measuring, this means that the internal fuse
is defective.Keep in mind that a low battery level could lead to incorrect
measurements. Replace the battery on a regular basis. (tip: the reduced
luminosity of the backlight/LCD display indicates a low battery level) Use
this device with original accessories only. Velleman nv can not be held
responsible in the event of damage or injury resulting from (incorrect) use of
this device. For more info concerning this productand the latest version of
this manual, please visit our website
www.Velleman.EU. The information in this manual is
subject to change without prior notice.
© COPYRIGHT NOTICE
The copyright to this manual is owned by Velleman nv. All worldwide rights
reserved. No part of this manual may be copied, reproduced, translated or
reduced to any electronic medium or otherwise without the prior written
consent of the copyright holder.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>