PD1D Piezo Inertia Drive Monolithic Stage
User Guide
Chapter 1 Warning Symbol Definitions
Below is a list of warning symbols you may encounter in this manual or on your device.
Symbol | Description | Symbol | Description |
---|---|---|---|
Direct Current | Off (Supply) | ||
**** | Alternating Current | In Position of a Bi-Stable Push Control | |
Both Direct and Alternating Current | Out Position of a Bi-Stable Push |
Control
| Earth Ground Terminal| | Caution: Risk of Electric Shock
| Protective Conductor Terminal| | Caution: Hot Surface
| Frame or Chassis Terminal| | Caution: Risk of Danger
| Equipotentiality| | Warning: Laser Radiation
| On (Supply)| | Caution: Spinning Blades May Cause Harm
Chapter 2 Safety
2.1. Safety Information
For the continuing safety of the operators of this equipment, and the
protection of the equipment itself, the operator should take note of the
Warnings, Cautions and Notes throughout this manual and, where visible, on the
product itself.
The following safety symbols may be used throughout the manual and on the
equipment itself.
SHOCK WARNING
A shock warning is given when there is a risk of injury from electrical shock.
WARNING
A warning is given when there is danger of injury to users.
CAUTION
A caution is given when there is a possibility of damage to the product.
NOTE
This is a clarification of an instruction or additional information.
2.2. General Warnings
WARNING
If this equipment is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired. In particular, excessive
moisture and/or dust may impair operation. Spillage of fluid, such as sample
solutions, should be avoided. If spillage does occur, clean up immediately
using absorbent tissue. Do not allow spilled fluid to enter the internal
mechanism.
The equipment is for indoor use only.
The equipment is not designed for use in an explosive atmosphere.
When running custom move sequences, or under fault conditions, the stage may
move unexpectedly.
Operators should take care when working inside the moving envelope of the
stage.
2.3. General Cautions
CAUTION
Dust and debris can reduce the lifetime of the stage. If left running
unattended for extended periods of time, the application should be covered
where possible.
If the clamping force onto the moving platform is too high, this can cause
reduced stepping performance and possible stalling of the stage.
The stage’s screws are well adjusted and fixed. Loosening or removing any
screws will change the performance and invalidate the warranty, except where
instructed to do so in this manual.
2.4. General Notes
NOTE
The slip-stick function of the mechanism uses a very short pulse width.
Continuous stepping of the stage results in an audible noise at a typical
level of 60 to 70 dB.
The step size is defined as the distance moved in one step or pulse. This
distance can be increased by up to about 30% by changing the piezo drive
voltage. The actual step size achieved for a given drive voltage will be
dependent on the application. Due to the open loop design, piezo hysteresis,
component variance, and application conditions, the achieved step size of the
system may vary by over 20% and is not normally repeatable.
Chapter 3 Overview
3.1. Introduction
The PD1D(/M) stage features two piezo inertia motors that use friction and
inertia to translate the platform in discrete cycles. Each cycle consists of a
relatively slow piezo expansion step accompanied by a rapid piezo contraction
step. During the expansion step, the moving platform moves in the same
direction with the piezo element due to friction, while the rapid piezo
contraction causes the moving platform to slip and remain immobile due to
inertia. Continuous motion is achieved through repetition (“continuous
stepping”) of this cycle and reverse travel is achieved by performing the
cycle in reverse.
The PD1D(/M) stage is a monolithic stage and provides two axes of 20 mm travel
with open-loop operation and better perpendicularity than two combined,
discrete stages. When driven by the KIM101 or KIM001 Inertial Piezo
Controller, each axis of the PD1D(/M) stage offers continuous long term
stepping as well as periodic position and hold modes with 3 N clamping/holding
force. The stage can achieve a typical step size of 1 µm using different rise
and fall voltage rates that depend on the load.
3.2. Included Components
The PD1D(/M) stage is delivered with the following components:
- Quick Start Card
- 8-32 (M4) Cap Screws, 6 mm Long, 2 Pieces
- 2-56 (M2) Cap Screws, 4 mm Long, 2 Pieces
- Ø2 mm Dowel Pin, 3 mm Long, 2 Pieces
- 1.3 mm Hex Key
3.3. Unpacking
NOTE
Unpack the stage with care. Check the contents for signs of damage. If there
is any sign of damage or missing parts, contact Thorlabs immediately. The lock
plate and corresponding M1.4 screws are used to lock the stage during
transportation or storage.
Unlock the upper M1.4 screw indicated by the note card using the included 1.3
mm hex key and remove the note card before first use. Leave the M1.4 screw
loosened during use.
3.4. Accessories
The PD1B(/M) and PD1B2(/M) mounting adapters provide a flat surface for
mounting the PD1D(/M) stage and offer 8-32 (M4) mounting and 1/4″-20 (M6)
mounting features, respectively.
The PD1T(/M) adapter plate features a central 8-32 (M4) tapped hole, both
imperial and metric versions have four 4-40 tapped holes spaced for 16 mm cage
systems.
The PD1U(/M) adapter plate features four 6-32 and four 8-32 tapped holes (or
eight M4 tapped holes) in a cross pattern. The bottom surface features five #8
(M4) counterbores, providing a convenient means of attaching load to the
stage.
The PD1Z(/M) Z bracket allows a PD1(/M) or PDX1(/M) stage to be mounted on top
of the PD1D(/M) stage for an XYZ configuration.
The PAA101 1.5 m extension cable and T5026 male adapter can be used to extend
the drive cable length to 2.5 m.
CAUTION
Do not mount an imperial accessory to a metric stage or a metric accessory
to an imperial stage.
Chapter 4 Installation
4.1. Mounting
4.1.1. General
Mount the PD1D(/M) stage on an even surface. The recommended flatness of the
surface is ≤5 μm.
For applications with large temperature changes, only mount the PD1D(/M) stage
on a surface that has the same or similar thermal expansion properties as the
stage.
The mounting holes are accessible when the moving platform is at the ends of
the translation range. The stage can be mounted using two 2-56 (M2) screws
near the corners with a maximum of 0.35 N·m torque. Alternatively, two 8-32
(M4) screws can be used on either end of the stage with up to 0.55 N·m torque.
Take care to mount the stage before attaching components to the stage’s moving
platform.
Figure 3 Mounting Holes
Direction of Travel
- **** Through Holes for 2-56 (M2) Cap Screw (2 Places)
- **** Through Holes for 8-32 (M4) Cap Screw, 1″ (25.0 mm) Spacing (2 Places)
CAUTION
Mounting the PD1D(/M) stage on an uneven surface can warp the stage, degrading
the performance and possibly damaging the stage.
When mounting close to other equipment, ensure that the travel of the carriage
is not obstructed. If movement is obstructed the stage will stall, which could
cause excessive wear and premature failure.
When using the stage in the proximity of other objects or equipment, ensure
that movement of cables connected to the moving carriage is not impeded.
Avoid touching the ceramic bar under the carriage plate.
Thorlabs offers two mounting plates, PD1B(/M) and PD1B2(/M), to provide a flat
surface for mounting.
- First mount the plate to a breadboard or base plate or table using the screws included with the adapter plate. Washers are recommended to be used under the cap screws to enhance the mounting and reduce the cutting to the plate by the screws.
- Insert two 8-32 (M4), 6 mm long cap screws through the stage to the mounting plate. 8 mm long 8-32 (M4) screws could also be used here. Two 2-56 (M2), 4 mm long screws could also be used to mount the stage to the mounting plate.
4.1.2. Setting Up an XYZ Stage Configuration
CAUTION
In a multi-axis system, the Z-axis stage and any mounting adapters must also
be moved and should be considered as a part of the load.
Do not exceed the load capacity for any of the stages in a multi-axis system.
When a stage is mounted vertically, make sure that the installed load is lower
than the push/pull force of the stage.
Screws and locating dowel pins that are inserted too deep into the stage will
damage it. Do not thread screws into the top plate with excessive force as
they may break through the terminating lip at the base of each threaded hole.
Mounting the PD1D(/M) monolithic stage upright is not recommend as this will
lead to a 50% loss of drive capacity as well as a large decrease of holding
ability.
To set up an XYZ system, Thorlabs offers a Z bracket adapter plate, PD1Z(/M),
to mount a PD1(/M) stage vertically on a PD1D(/M) stage.
- Insert two Ø2 mm, 3 mm long dowel pins into the lower stage’s two slots in diagonal corners.
- Insert two 2-56 (M2), 4 mm long cap screws through the Z bracket to two adjacent corners on the lower stage.
- Insert another two Ø2 mm, 3 mm long dowel pins into the two diagonal corner slots in the bracket’s vertical face.
- Align the PD1(/M) stage to the vertical plane with its cable oriented vertically and away from the PD1D(/M) stage, then mount the PD1(/M) stage to the Z bracket by aligning and inserting the two dowel pins.
- Insert two 2-56 (M2), 4 mm long cap screws through the PD1(/M) stage to the vertical plane of the bracket, as shown above. Alternatively, two 8-32 (M4), 6 mm long cap screws can also be used to mount the PD1(/M) stage to the bracket.
4.1.3. Mounting Components to the Stage
CAUTION
Do not exceed the load capacity specified in Chapter 6.
The load’s center of mass should be over the platform.
Do not use screws and dowel pins that protrude past the maximum hole depth.
- **** 2-56 (M2) Threaded Hole, 3 mm Deep (4 Places)
- **** 8-32 (M4) Threaded Hole, 3 mm Deep, 1″ (25.0 mm) Spacing (4 Places)
- **** Ø2 mm Dowel Pin Hole, 1.5 mm Deep (4 Places)
Attach the load using screws and check that it is secure. We recommend
aligning the load with a set of dowel pins opposite each other relative to the
center of the stage, as well as ensuring the load’s center of mass is over the
platform before driving the stage.
Thorlabs offers a thin mounting adapter plate, PD1T(/M), to provide a
convenient means of attaching load to PD series stages by a central 8-32 (M4)
tapped hole and other 4-40 (M2 & M3) tapped holes. Both imperial and metric
version have four 4-40 tapped holes spaced for 16 mm cage systems. Following
is the example to show how to mount a PD1T(/M) to a PD1D(/M) stage.
- Insert two Ø2 mm, 3 mm long dowel pins into the moving platform’s two slots along the travel direction.
- Mount the adapter plate to the stage by the alignment of the two dowel pins.
- Insert two 2-56 (M2), 4 mm long cap screws through the adapter plate to the stage.
Thorlabs also offers the PD1U(/M) thick mounting adapter plate to provide a convenient means of attaching a load to PD series stages by five #8 (M4) counterbores. As shown in Figure 8:
- Use the five #8 (M4) counterbores to mount the load from the underside of the adapter. Notice that the load should not cover the two 2-56 (M2) cap screw counterbores on the top surface.
- Insert two Ø2 mm, 3 mm long dowel pins into the moving platform’s two slots along the travel direction.
- Mount the adapter plate to the stage by aligning the two dowel pins.
- Insert two 2-56 (M2), 8 mm long cap screws through the adapter plate to the stage.
4.2. Electrical Connections
CAUTION
The PD1D(/M) stage must be driven by a 2019 or newer version of Thorlabs
KIM101 controller with firmware revision 010004 or higher, or any version of
the Thorlabs KIM001 controller.
Earlier versions of the KIM101 controller or those with older firmware will
not function properly and may cause failure of the stage and/or the
controller. The previous-generation TIM101 controller is not compatible.
The stage is shipped with a 1 m (3.3 ft) cable that is terminated in an SMC
connector. Connect each axis of the PD1D(/M) stage to each of the MOT (motor)
terminals on the rear panel of the KIM101/KIM001 controller – see the KIM101
controller manual for pin out details.
Before driving the PD1D(/M) stage using the KIM101/KIM001 controllers, the
controller configuration must be set correctly, since the KIM101/KIM001
controllers are configured by default to drive our PIA series piezo inertia
actuators. Using the joystick and menu button, select the “Set stage” option
and change it from “PIA” to “PD(R)”, then restart the KIM101/KIM001
controller. The display will show “Stage is PD(R)” and the configuration will
be changed to drive the PD1D(/M) stage. Alternatively, using Kinesis software
with revision 1.14.19 or higher can also be used to switch the configuration
by chosing the stage type. See the KIM101 controller manual for additional
front panel configuration details.
CAUTION
The KIM101/KIM001 controller is configured by default to drive Thorlabs PIA
series actuators. The configuration must be changed to “PD(R)” before using
the controller to drive the PD1D(/M) stage, or else the stage may experience
degraded performance or damage may be caused to either the controller, the
stage, or both.
Confirm the configuration has been switched to PD(R) stage before using the
KIM101/KIM001 controller with the PD1D(/M) stage.
Thorlabs offers a 1.5 m extension cable (PAA101) and male adapter (T5026) to
extend the drive cable length to 2.5 m, see
www.thorlabs.com.
Due to the capacitance of the cables, do not use cables longer than 2.5 m in
total.
Chapter 5 Operation
5.1. General
CAUTION
Confirm the configuration of the KIM101/KIM001 controller has been switched to
“PD(R)” stage.
WARNING
The piezo actuators in this product use high voltages. Voltages up to 125 V
may be present at the SMC connector. This is hazardous and can cause serious
injury. Appropriate care should be taken when using this device. Persons using
the device must understand the hazards associated with using high voltages and
the steps necessary to avoid risk of electrical shock.
The piezo controller must be switched OFF before the stage is connected or
disconnected. Failure to switch the controller off may result in damage to
either the controller, the stage, or both.
For a complete tutorial on driving the stage, see the manual supplied with the
KIM101/KIM001controlle. Basic steps in controlling the stage are as follows:
-
Mount the PD1D(/M) stage as detailed in Section 4.1.
-
Make electrical connections as detailed in Section 4.2.
NOTE
During operation, the stage makes a high pitched noise and may generate some heat. This is normal behavior in the performance of the device and does not indicate a fault condition. -
Run the software and click the ‘zero’ button on the GUI panel. This establishes a datum at the current position, from which subsequent positional moves can be measured.
-
The stage can now be moved using the controls on the KIM101/KIM001, the GUI panel, or by sending commands to move each axis – see the manual supplied with the KIM101/KIM001 controller, and the help file supplied with the software for more information.
NOTE
The PD1D(/M) stage is an open loop device that has been designed to offer relative positioning which can be commanded via the number of steps.
CAUTION
If the stage is driven to the end of its travel range, do not drive it
further. Instead, stop or reverse the driving direction to avoid excessive
wear at the end stops.
Using a small working range permanently increases the wear in this area.
Select another part of the travel range for the working range if possible to
spread the wear evenly and extend stage life.
5.2. Maintenance
Periodically perform a maintenance run by driving the stage to travel back and
forth across the entire travel range. The frequency of this will depend on
usage. In applications involving continuous use or where a small travel range
is used at a high duty cycle, the maintenance run should be performed often to
ensure optimal performance of the stage.
Due to the stick-slip (friction-inertia) nature of the mechanism, wear of the
friction parts occurs during its lifetime. Periodic maintenance helps to
ensure the optimum stepping performance during its lifetime.
NOTE
The PD1D(/M) stage is assembled with rust preventative oil. Don’t wipe the
stage with organic solvent. When necessary, clean the surfaces of the stage
with a dry, dust-free cloth.
5.3. Transportation
CAUTION
When packing the unit for shipping, use the original packaging and lock the
moving platform of the stage using the lock plate and M1.4 screws. If the
original packaging is not available, use a strong box and surround the unit
with at least 20 mm of shock absorbent material.
Chapter 6 Specifications
Item # | PD1D(/M) |
---|---|
Travel Range(X & Y Axes) | 20 mm (0.78″) |
Typical Step Size1 | 1 pm |
Maximum Step Size2 | <3 pm |
Step Size Adjustability2 | Up to 30% |
Maximum Step Frequency3 | 2 kHz |
Speed (Typical Max, Continuous Stepping)4 | 3 mm/s |
Average Speed Variation Over Travel Range4 | ±10% |
Horizontal Load Capacity | 3 Ice 6.61 lbs |
Vertical Load Capacity5 | 100 g 3.5 oz |
Clamping / Holding Force | 3 N |
Pitch / Yaw Over Travel Range | 200 prad |
XY Stacked Orthogonality | 5_2 mrad |
Lifetime6 | >10 Billion Steps |
Connector Type | SMC Female |
Cable Length | 1 m (3.3 ft) Cable Included |
Piezo Capacitance | 170 nF |
Max Operating Voltage | 125 V |
Operating Temperature | 10 to 40 °C |
Dimensions | 32.5 mm x 32.5 mm x 20.0 mm (1.28″ x 1.28″ x 0.79″) |
Weight (Including Cable) | 137 g (4.83 oz) |
Note: all specifications are measured using the KIM101 controller and
with the stage mounted on a surface with flatness ≤5 μm.
*It’s not recommended to mount this monolithic stage vertically because one axis will get a lateral force to the rails and may influence its angular error and even reduce the lifespan of the rail. If it must be mounted vertically, long term running (>3 billion steps) may cause the creep of the rails in the vertical axis, leading to the decrease of travel range and an impact on the application. Recovery through horizontal mounting and several full range travel sequences is recommended after many steps (>1 billion steps) of vertical traveling.
1 This can vary by 20% due to component variance, change of direction, and
application conditions.
2 This is adjusted by changing the piezo drive voltage and frequency – see the
KIM101 controller manual for more details.
3 2 kHz is the maximum output from the KIM101 and KIM001 controllers.
4 Specified at 2 kHz Step Frequency
5 This is equivalent to the push/pull force along the axial/travel direction,
which is 1 N.
6 Driven by the KIM101 Controller Under 2000 Hz/85 V
Chapter 7 Frequently Asked Questions
How does a piezo inertia motor work?****
– A piezo inertia stage uses friction and inertia (“stick” and “slip”) to drive the moving platform. The piezo stack is driven by a sawtooth voltage waveform. As the voltage is ramped up, the piezo will extend and push the moving platform (stick). When the voltage is dropped to zero, the piezo returns to its original length. By using different rates of voltage rise and fall, the moving platform travels more in one direction than the other due to inertia and different frictional coefficients, resulting in residual linear travel.
What applications can they be used for?****
– Any ‘set-and-forget’ application, particularly where space is tight. The primary function of the PD1D(/M) stage is relative position and hold. The step size is dependent upon preload, and will differ between stages and applications. They are not suitable where repeatable step size is required.
What is the lifetime of a typical piezo inertial motor?****
– The piezo stack of the actuator is rated for a service life of over 10 billion steps. With proper maintenance (see Section 5.2), the wear performance of the stage’s friction parts should endure for this lifetime, however the speed variation could change by up to 50% or more over time.
What driver can I use?****
– The PD1D(/M) stage must be driven by a 2019 or newer version of Thorlabs
KIM101 controller (per the S/N label) with a firmware revision of 010007 or
higher, or any version of a Thorlabs KIM001 controller. Earlier versions of
the KIM101 controller or those with older firmware will not function properly
and may cause failure of the stage and/or the controller. The previous-
generation TIM101 controller is not compatible. It’s also recommended to use
latest version of Kinesis software to remotely control KIM101/KIM001.
Ensure the controller’s stage mode is set to “PD(R)” for KIM101/KIM001.
What is the maximum length of cable?****
– The stages are shipped with 1 m (3.3 ft) of cable. Thorlabs offers the PAA101 1.5 m extension cable and the T5026 male adapter, available at www.thorlabs.com, which can be used to extend the cable length to 2.5 m. Due to the capacitance of the cables, do not use cables longer than 2.5 m in total.
Why the speed differs between forward and backward directions?****
– Due to the open loop design, piezo hysteresis, component variance, and application conditions, there would be a speed difference between two directions. Improper drive signal (wrong stage type selection, low driving frequency), uneven mounting surface or unstable mounting, gravity of the carriage plate and load would also influence or increase the speed difference between two directions.
Chapter 8 Mechanical Drawings
|
---|---
Chapter 9 Regulatory
As required by the WEEE (Waste Electrical and Electronic Equipment Directive) of the European Community and the corresponding national laws, Thorlabs offers all end users in the EC the possibility to return “end of life” units without incurring disposal charges.
- This offer is valid for Thorlabs electrical and electronic equipment:
- Sold after August 13, 2005
- Marked correspondingly with the crossed out “wheelie bin” logo (see right)
- Sold to a company or institute within the EC
- Currently owned by a company or institute within the EC
- Still complete, not disassembled and not contaminated
As the WEEE directive applies to self-contained operational electrical and electronic products, this end of life take back service does not refer to other Thorlabs products, such as:
- Pure OEM products, that means assemblies to be built into a unit by the user (e. g. OEM laser driver cards)
- Components
- Mechanics and optics
- Left over parts of units disassembled by the user (PCB’s, housings etc.).
If you wish to return a Thorlabs unit for waste recovery, please contact
Thorlabs or your nearest dealer for further information.
Waste Treatment is Your Own Responsibility
If you do not return an “end of life” unit to Thorlabs, you must hand it to a
company specialized in waste recovery. Do not dispose of the unit in a litter
bin or at a public waste disposal site.
Ecological Background
It is well known that WEEE pollutes the environment by releasing toxic
products during decomposition. The aim of the European RoHS directive is to
reduce the content of toxic substances in electronic products in the future.
The intent of the WEEE directive is to enforce the recycling of WEEE. A
controlled recycling of end of life products will thereby avoid negative
impacts on the environment.
Chapter 10 Declaration of Conformity
Declaration of Conformity
We: Thorlabs Optical Electronic Technology (Shanghai) Co., Ltd
Of. Room A101, No.100, Lane 2891, South Qilianshan Rd, Shanghai
In accordance with the following Directive(s):
2014/35/EU: Low Voltage Directive (LVD)
2011/65/EU: Restriction of Use of Certain Hazardous Substances (RoHS)
hereby declare that:
Model: PD1D, PD1D/M
Equipment: Monolithic Piezo Inertia Drive Stage
Is in conformity with the applicable requirements of the following documents:
EN 61010-1: 2010 (Third Edition) + Al: 2019
and which, issued under the sole responsibility of Thorlabs, is in conformity
with Directive 2011/65/EU of the European Parliament and of the Council of
Eith June 2011 on the restriction of the use of certain hazardous substances
in electrical and electronic equipment, for the reason stated below: does not
contain substances in excess of the maximum concentration values tolerated by
weight in homogenous materials as listed in Annex II of the Directive.
I hereby declare that the equipment named has been designed to comply with the
relevant section of the above referenced specifications, and complies with all
applicable Essential Requirements of the Directives.
On: 21th Apr. 2021
Chapter 11 Thorlabs Worldwide Contacts
For technical support or sales inquiries, please visit us at www.thorlabs.com/contact for our most up- to-date contact information.
USA, Canada, and South America
Thorlabs, Inc.
sales@thorlabs.com
techsupport@thorlabs.com| UK and
Ireland
Thorlabs Ltd.
sales.uk@thorlabs.com
techsupport.uk@thorlabs.com
---|---
Europe
Thorlabs GmbH
europe@thorlabs.com| Scandinavia
Thorlabs Sweden AB
scandinavia@thorlabs.com
France
Thorlabs SAS
sales.fr@thorlabs.com| Brazil
Thorlabs Vendas de Fotônicos Ltda.
brasil@thorlabs.com
Japan
Thorlabs Japan, Inc.
sales@thorlabs.jp | China
Thorlabs China
chinasales@thorlabs.com
References
- Thorlabs, Inc. - Your Source for Fiber Optics, Laser Diodes, Optical Instrumentation and Polarization Measurement & Control
- Thorlabs, Inc. - Your Source for Fiber Optics, Laser Diodes, Optical Instrumentation and Polarization Measurement & Control
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>