Cardinal Health Kangaroo Joey Enteral Feeding Pump User Guide
- June 8, 2024
- Cardinal Health
Table of Contents
Cardinal Health Kangaroo Joey Enteral Feeding Pump
Cardinal-Health-Kangaroo-Joey-Enteral-Feeding-Pump-prodact-img
- A. Distal end of feeding set
- B. Feed only feeding set
- C. Feed and flush feeding set
- D. Pump
Do not use feeding sets for longer than 24 hours.
Proper loading and unloading of feeding pump
- Grasp thumb tab on valve and insert into left pocket — ensure valve is fully seated. Thumb tab should align with slot in blue door.
- Grasp black ring retainer, gently wrap tubing around rotor and insert retainer directly into right pocket. Do not overstretch tubing.
- To properly unload a pump set, power off the pump and wait 10 seconds before removing the feeding set.
Proper rinsing of feeding set
- Disconnect feeding set from 3. Empty contents of feed bag. patient before rinsing. The feeding set must remain loaded during the rinsing procedure.
- CAUTION — If set is removed from the pump while in operation, it may damage the set and prevent use, negatively affecting the system accuracy
- Empty contents of feed bag.
- Fill the feed bag with water to approximately half of the total capacity. Properly seal the bag and shake for 5 to 10 seconds.
- Empty the feed bag and repeat the last step until the bag contains clean water.
- Power on the pump. Select “Keep Settings” to continue with previous setting or “Clear Settings” to start programming new pump settings.
- Press the “Prime Pump” button to enter the “Prime Pump” screen.
- Press and hold the “Hold to Prime” button until only clean water flows through the tubing. Once the rotor starts this takes between 1 and 2 minutes.
- Empty bag of water. Prime to remove any remaining water from the feeding set.
- CAUTION — For proper accuracy, the fluid level in the feeding set bag must be 6 inches above the top of the feeding pump when the feeding cycle is initiated.
For additional operations and information, see the Operating Manual for the Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp, Programmable
© 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal
Health LOGO and Kangaroo are trademarks or registered trademarks of Cardinal
Health and may be registered in the US and/or in other countries. Patent
www.cardinalhealth.com/patents. Lit.
No. 2PATR18-764311 (10/2018) 15 Hampshire Street 800.962.9888
Mansfield, MA 02048 508.261.8000
References
Read User Manual Online (PDF format)
Loading......