BROOKFIELD VS-CRA-14S Laboratory Stand Instructions
- June 9, 2024
- BROOKFIELD
Table of Contents
VS-CRA-14S Laboratory Stand
Instruction Manual
VS-CRA-14S Laboratory Stand
Item | Part Number | Description | Qty. |
---|---|---|---|
1 | VS-CRA-14S |
Or
VS-CRA-18S| Rod (14″) and Clamp Assembly (Model qt Rod (18″) and Clamp
Assembly (Model G 18)t| 1
1
2| VSQA-100Y| Rod and Clamp Assembly (Model QB)t| 1
3| GV-1203| Leveling Screw, Model G and QB| 2
4| GV-1201| Base, Models G and QB (includes 2 GV-1203 leveling screws)| 1
5| N/A| 14″ or 18″ Upright Rod (Model A & G)| 1
6| N/A| Upright Rod (Model QB)| 1
7| 505311832501B| Screw, 5/16 -18 x 1″ hex head| 1
8| 502028071S33B| Flat washer 5/16 x 7/8 x .071″| 1
9| N/A| Universal Lab Stand Clamp Assembly| 1
10| VS-41 Y| Knob Assembly| 1
11| N/A| Tension Screw| 1
12| N/A| Knob Assembly (Model QB)| 1
Model G & Model G 18
Model G Lab Stand – Comes with 14” rod (VS-CRA-14S).
Used with all Viscometers/Rheometers.
Model G 18 Lab Stand – Comes with 18” rod (VS-CRA-18S).
Optional choice when additional height is required.
UNPACKING
Check to see that all the components are received with no concealed damage:
1 GV-1201 Base with two leveling screws
1 VS-CRA-14S (Model G) or
VS-CRA-18S (Model G 18) Rod and clamp assembly with one mounting screw and
washer.
ASSEMBLY
-
Remove the mounting screw and washer from the upright rod. Place the rod and clamp assembly into the hole in the top of the base.
Note: Be sure the clamp assembly has the ension screw (item 11) facing forward. -
Position the rod and clamp assembly until the slot on the bottom of the rod intersects the pin located in the base.
-
While holding the rod and base together, insert the screw and washer as shown and tighten securely.
-
Adjust the tension screw so that the clamp assembly is not loose on the upright rod.
VISCOMETER MOUNTING AND USE
Insert the Viscometer mounting rod into the hole (with the cut-away slot) in
the clamp assembly. Adjust the viscometer level until the bubble level is
centered and tighten the clamp knob (clockwise). Use the base leveling screws
to “fine” adjust the viscometer level. Note: The DVNext Rheometer has a
Digital Bubble Level.
CAUTION: Do not tighten the clamp knob unless the viscometer mounting rod
is inserted in the clamp assembly.
CAUTION: Do not use the DV2T or DVNext with any other laboratory stand
that does not utilize the GV-1201 base. This large base is necessary for
stability of the DV2T Viscometer and DVNext Rheometer during use. Earlier
versions of the Brookfield Laboratory Stand including Model A and Model S
should not be used with the DV2T or DVNext.
Model QB
Model QB Lab stand – Comes with 14” rod (VSQA-100Y).
Available as an optional choice for all Viscometers/Rheometers when easier
up/down movement of instrument head is desired.
CAUTION: The Model QB upright rod is a spring-loaded, non-serviceable
device. Do not disassemble the clamp and upright rod assembly.
UNPACKING
Check to see that all the components are received with no concealed damage:
1 GV-1201 Base with two leveling screws
1 VSQA-100Y Clamp and upright rod assembly with one mounting screw and washer
ASSEMBLY
- Remove the mounting screw and washer from the upright rod. Place the rod and clamp assembly into the hole in the top of the base with the release lever facing forward.
- Position the rod and clamp assembly until the slot on the bottom of the rod intersects the pin located in the base.
- While holding the rod and base together, insert the screw and washer as shown and tighten securely.
VISCOMETER MOUNTING AND USE
Insert the viscometer mounting rod into the cutaway slot of the clamp
assembly. Adjust the instrument until the viscometer/rheometer bubble level is
centered and tighten the clamp knob. Use the base leveling screws to “fine”
adjust the instrument level. Note: The DVNext Rheometer has a Digital Bubble
Level.
CAUTION: Do not tighten the clamp knob unless the viscometer mounting rod
is inserted in the clamp assembly.
Push the release lever on the clamp assembly. Move viscometer/rheometer head
to desired height and release lever.
Elevate your sample container, if necessary, to ensure that the proper spindle
immersion is achieved.
CAUTION: Do not use the DV2T or DVNext with any other laboratory stand
that does not utilize the GV-1201 base. This large base is necessary for
stability of the DV2T Viscometer and DVNext Rheometer during use. Earlier
versions of the Brookfield Laboratory Stand including Model A and Model S
should not be used with the DV2T or DVNext.
© 2019 AMETEK Brookfield Inc. All rights reserved.