PCE-DFG N Series Digital Force Gauge User Manual

June 4, 2024
PCE

PCE - logo DFG N Series Digital Force Gauge
User Manual

User manuals in various languages can be downloaded here: www.pce- instruments.com

PCE-DFG N Series Digital Force Gauge - qrhttp://www.pce-instruments.com

Safety notes

Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE  Instruments personnel.
Damage or injuries caused by non-observance of the manual are excluded from our liability and not covered by our warranty.

  • The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
  • The instrument may only be used if the environmental conditions (temperature, relative humidity, …) are within the ranges stated in the technical specifications. Do not  expose the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
  • The case should only be opened by qualified PCE Instruments personnel.
  • Never use the instrument when your hands are wet.
  • You must not make any technical changes to the device.
  • The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner, no abrasives or solvents.
  • The device must only be used with accessories from PCE Instruments or equivalent.
  • Before each use, inspect the case for visible damage. If any damage is visible, do not use the device.
  • Do not use the instrument in explosive atmospheres.
  • ATTENTION: For impact tests, the maximum measurable value of the force gauge should be twice as high as the applied impact load.
  • When doing impact tests, wear a mask and protective gloves to avoid injuries.
  • Do not use the test stand when it is bent or damaged. Dropping can cause injuries.
  • This device only measures tensile and compressive forces. The test head must not be bent or twisted.
  • Overloading, excessive impact loads or applied forces other than tensile and compressive forces can cause damage to the sensor.
  • Do not press the keys with pointed objects.
  • Keep the force gauge away from water, oil and other liquids.
  • Store the meter in a cool, dry place without any occurrence of vibration.
  • Wire the ports as described in this manual. Non-observance of the instructions can cause circuit failure or problems with your computer.
  • Make sure that the mains adaptor is securely connected to the power outlet as otherwise short circuits and thus electric shocks and fire can occur.
  • When the battery is fully charged, remove the mains adaptor immediately to avoid overheating, fire or accidents.
  • Non-observance of the safety notes can cause damage to the device and injuries to the user.

We do not assume liability for printing errors or any other mistakes in this manual.
We expressly point to our general guarantee terms which can be found in our general terms of business.
If you have any questions please contact PCE Instruments. The contact details can be found at the end of this manual.

Safety symbols
Safety-related instructions the non-observance of which can cause damage to the device or personal injury carry a safety symbol.

Symbol Designation/Description
General warning sign

Non-observance can cause damage to the device and injuries to the user.
| Warning: electrical voltage
Non-observance can cause electric shock.

Specifications

2.1 Technical specifications
Force gauge with internal dynamometric cell

S pecification V a l ue / version
Model PCE-DFG N

DFG N
5| 10| 20| 200| 500**
Range| 0 … 5 N| 0 … 10 N| 0 … 20 N| 0…200 N| 0…500 N
Resolution| 0.001 N| 0.002 N| 0.005 N| 0.01 N| 0.05 N
Accuracy| | | | |
Dimensions| 200 x 97 x 42 mm
Weight| approx. 540g

Force gauge with external dynamometric cell

Specification Value/version
Model PCE-DFG N 1K
Range 1.000 N / 100 kg
Resolution 0.5 N
Accuracy 0.1 % F.S.
Units N, kg. lb, KPa
Model PCE-DFG N 20K
Range 20.000 N / 2t
Resolution 0.002 kN / 2 N
Accuracy 0.2 % F.S.
Units kN, t. klb, MPa

Further versions on request
General specifications

Specification Value
Cell protection class IP 67
Device dimensions 162 x 82 x 41 mm
Device weight 325 q
Display 2.8″ TFT graphical display
Alarm modes within, beyond, fracture, off
Sampling rate 6 … 1600 Hz device

6…800 Hz software
Memory| 100 measurements
Power supply| Ni-Hi rechargeable battery
6 V, 1600 mAh
Battery life10 hours
Mains adaptor| 12 VDC 1 A;
Outputs| Communication via USB
Output port 12 V, 50 mA
Operating conditions| -10 … +50 °C; 5 … 95 % RH, non-condensing
Protection class| IP 54

2.2 Delivery contents

Version with internal cell: Version with external cell:
1 x force gauge PCE-DFG N 1 x force gauge PCE-DFG N
5 x test head adaptors (pointed head adaptor, 1 x dynamometric cell
notched head adaptor, chisel head adaptor, 2 x eyebolt (only up to 10 kN load

cell)
flat head adaptor, hook adaptor)| 1 x case for the meter
1 x extension rod incl. adaptor| 1 x case for the dynamometric cell
1 x USB cable| 1 x USB cable
1 x mains adaptor| 1 x mains adaptor
1 x software| 1 x software
1 x user manual| 1 x user manual

2.3 Optional accessories

PCE-C-S-1000N-C3 load cell  1,000 N / 100 kg / M10
PCE-C-S-2500N-C3 load cell  2,500 N / 250 kg / M12
PCE-C-S-5000N-C3 load cell  5,000 N / 500 kg / M12
PCE-C-S-10000N-C3 load cell 10,000 N / 1 t / M12
PCE-C-S-20KN-C3 load cell  20.000 N / 2 t / M20
PCE-C-S-50KN-C3 load cell  50.000 N / 5 t / M20
PCE-C-S-100KN-C3 load cell  100.000 N / 10 t / M30
PCE-C-A-1000N-EB eyebolt M10  1,000 N (set: 2 pieces)
PCE-C-A-2500N-EB eyebolt M12  2,500 N (set: 2 pieces)
PCE-C-A-5000N-EB eyebolt M12  5,000 N (set: 2 pieces)
PCE-C-A-10000N-EB eyebolt M12 10,000 N (set: 2 pieces)
PCE-C-A-60KN-EB eyebolt M20 / 60.000 N / 6 t (set: 2 pieces)
PCE-C-A-120KN-EB eyebolt M30 / 120.000 N / 12 t (set: 2 pieces)
PCE-C-A-1000N-RE eyebolt with joint head M10  1,000 N (set: 2 pieces)
PCE-C-A-10000N-RE eyebolt with joint head M12 10,000 N (set: 2 pieces)
PCE-C-A-1000N-HB clip bolt M10  1,000 N (set: 2 pieces)
PCE-C-A-2500N-HB clip bolt M12  2,500 N (set: 2 pieces)
PCE-C-A-5000N-HB clip bolt M12  5,000 N (set: 2 pieces)
PCE-C-A-10000N-HB clip bolt M12 10,000 N (set: 2 pieces)
CAL-PCE-DFG N calibration certificate

Further accessories on request

System description

3.1 Device
Version with internal dynamometric cell ![PCE-DFG N Series Digital Force Gauge

1 Measuring shaft 6 Pointed head adaptor
2 Display 7 Hook adaptor
3 Keypad 8 Notched head adaptor
4 Extension rod 9 Flat head adaptor
5 Chisel head adaptor 10 Adaptor for extension rod

3.2 Interfaces

  1. Input/output interface
  2. USB interface
  3. Power connection

Circuit diagram of output port

1   Switch of external input/output 3   Output upper limit
2   Output lower limit 4   GND

3.3 Display
In measurement mode

1 memory usage 7 battery level indicator
2 measurement curve 8-set sampling rate
3 force value 9 lower alarm value
4 upper alarm value 10 first peek
5 clear by key 11 minimum peak
6 maximum peak

3.4 Function keys

Ke y

|

Design Nation

|

Function

---|---|---
Si n gle measurement mode| Ca pture mode| Online measurement
mode
| Memory and query mode|

Menu mode

| On/ Off| Switch off| –| Switch off| –| –
| Back| –| Close capture mode| –| Exit| Exit/close parameter settings
| Zero| Zero setting| –| Zero setting| –| –
| Up| –| –| –| Up| Up
| Down| Activate memory and query mode| –| –| Switch to the upper window| Down
| OK| Open parameter settings| Stop capturing| –| Show report  and reading| Confirm parameter setting
| Left| Start curve capturing| –| Start curve capturin g| Move the flashing number left by one digit
| Right| Delete  peak value| –| Delete peak value| Move the flashing  number right by one digit

Getting started

4.1 Power supply
The PCE-DFG N is equipped with a rechargeable 1600 mAh 6 V Ni-Hi battery that should only be charged by means of the mains adaptor which is included in the standard delivery.
Charging can take 8 to 10 hours and should only be started when the battery is completely flat.
Excessively frequent or long-time charging shortens the battery life.
When the battery is fully charged, it will last up to 10 hours of continuous use. The device can also be used during charging. The battery can be charged approx. 500 times.

4.2 Settings
When you are in measurement mode, press the OK key to enter the settings screen which is divided into 2 pages:

Page 1 Page 2

PCE-DFG N Series Digital Force Gauge - satteng 2| PCE-DFG N Series Digital Force Gauge - satteng

In order to change settings, select the menu item with the arrow keys and confirm with the OK key.
The values can then be changed by means of the arrow keys. Then press “OK” to confirm the settings or the Back key to discard.

Function

|

Desc ription page 1

---|---
Display Unit| The display unit can be set to: „N“, „kg“, „lb“ or „KPa“

Force Area

| The force area can be set to a value between 999.99cm2 and 0.01cm2 and is included in the calculation if the display unit selected is „kPA“ (important for accuracy).
Zero Tracking| For zero tracking, you can set: „Off“, „0.1 N“, „0.2 N“, „0.3 N“, „0.4 N“, „0.5 N“
Values below the value set are automatically excluded before the zero point stabilises. After stabilisation of the reading, the sampling rate will be 1 x per second. Deviations from the measured value which are below the set value are automatically excluded in order to keep the displayed values.
Sampling Speed| You can set how many measurements are taken per second. A value between 6 and 1600 Hz can be set here.
Note: The higher the sampling rate, the lower the accuracy will be. Higher sampling rates are suitable for dynamic measurements whereas lower sampling rates are more suitable for static and slow measurements.
Calibrate Grav| Enter the gravity at the place of calibration.
Alarm Upper| The upper alarm can be set to +/- 9999.9.
Alarm Mode| You can choose “Within“ (within alarm limit), “Beyond“ (outside alarm limit), “Fracture“ (overload alarm) or “Off“.
| If you select “Within“ or “Beyond“, the display will show information on the alarm.
If you select “Fracture“, Alarm Upper LV and Alarm Lower LV will automatically be set to “Fracture Alarm” and „Fracture Stop of Peak“. Set these two parameters. When the force reaches the fracture alarm value or when the sample breaks, the display will show some information on the alarm.
---|---
Peak V. Hold| You can select „On“ or “Off“. If “Off” is selected, the peak value will not be indicated in the display.
Factory Set A| Only relevant for customer service.
Factory Set B| Only relevant for customer service.
Factory Set C| Only relevant for customer service.
Calibrate| Press OK to start the calibration. The calibration result will have a considerable influence on the accuracy of measurement. There are two possibilities to calibrate the meter:
1. Entering saved data:
The user enters the saved calibration data. The calibration is done without any other devices or weights.
2. Standard calibration:
The force gauge is calibrated by means of the calibration stand or calibration weight.
User Gravity| Here, you can set the gravity at the place of use. The value can be between 9.700 and 9.900 N/kg. This parameter is used for gravity correction. The following formula must be used for this:
Displayed value = reading + reading x (gravitation place of calibration –gravitation place of use)
Alarm Lower| The lower alarm can be set to +/- 9999.9.
External Input| You can select “On“ or “Off“. If “On” is selected, the external switch can be switched on and the force gauge enters curve capture mode.
Note: The capturing duration depends on the sampling rate. Capturing duration in seconds = number of recorded data/sampling rate
Peak Hold Time| You can select “Clr by Key“ or certain periods between 1 and 60 seconds. If “Clr by Key“ is selected, the peak value will not be changed until the “Arrow Right key or the “Zero Set“ button is applied.
If a period between 1 and 60 seconds is selected, the peak value will automatically be measured again after the set time has passed. The peak value can also be re-measured by applying the “Arrow Right key or the “Zero Set“ button.
Function| Desc ription page 2
---|---
Capture Length| You can set a value between 1 and 1280 seconds. This value represents the duration of curve capturing in capture mode which depends on the sampling rate:
Sampling rate 6 Hz: 1 ~ 1280 seconds Sampling rate 12 Hz: 1 ~ 640 seconds Sampling rate 25 Hz: 1 ~ 320  seconds Sampling rate 50 Hz: 1 ~ 160 seconds Sampling rate 100 Hz: 1 ~ 80 seconds Sampling rate 200 Hz: 1 ~  40 seconds Sampling rate 400 Hz: 1 ~ 20 seconds Sampling rate 800 Hz: 1 ~ 10 seconds Sampling rate 1600 Hz:  1 ~ 5 seconds
F/P Boundary| You can set a value between 1 and 99999. This setting is used during peak value measurement to determine the first peak value. When you press the Arrow Right key, a new peak value measurement will start. Meanwhile, the values peak-to-peak (Vmax), valley-to-peak (Vmin), and new peak (Vnew) are updated continuously. For example, if 10 is set as the criterion, Vmax or Vmin will be counted as the first peak value when the absolute value of (Vmax – Vnew) or (Vmin – Vnew) is above 10.
Serial Port| This port is used to control the real-time data transfer. The following parameters can be set:
Prohibit: The real-time data transfer of the serial interface is prohibited.
Key/Order: A single output will take place when you press the Up key or when an output command is received. When the force gauge is connected to a computer, the programs on the computer will automatically disable the output function.
Change: A single output will take place when the measuring data change.
Stabilize: A single output will take place when the reading stabilises. Consecutive: The measuring data are transferred without interruption.
Auto Power Off| This function reduces energy consumption. The force gauge will automatically power off when it has not been used for a certain period of time.
Max Charge V| This window shows the maximum voltage of the battery.
Clear Storage| Here, you can delete saved measurement reports and curves.
Im p ortant information:
When the memory is full, all data will automatically be deleted to enable new data to be saved.
Factory Test| Only relevant for customer service.
S/N| This window shows the serial number of the device which cannot be changed.
Capture Trigger| Here, you can set a value between -99999 and +99999. The range of values depends on the set unit. This parameter stipulates the
| a condition that triggers the capturing when the force gauge is in curve capture mode. When the maximum number of data has been recorded or capturing was discontinued early, a capture report is created and saved. The curve is deleted when you leave capture mode.
---|---
Baud Rate| The baud rate for the serial interface can be set to a value between 4800 and 230400 bps.
This setting will only be effective after restarting the force gauge. Note: To make sure that all data are retrieved  when the device is connected to a computer, the baud rate should be set as follows:
12 Hz: ≥9600 bps 25 Hz: ≥14400 bps  50 Hz: ≥19200 bps  100 Hz: ≥28800 bps
200 Hz: ≥38400 bps  400 Hz: ≥57600 bps  800 Hz: ≥115200 bps  1600 Hz: ≥230400 bps
Due to the limited speed of serial interfaces, some data get lost when transferred to a PC if the sampling rate is higher than 800 Hz. The readings will, however, not get lost in the device.
Display Angle| Here, you can set the display angle. You can select 0 or 180 °.
Auto Backlight| This function also reduces energy consumption. The brightness of the display backlight will automatically be turned off when the meter has not been used for a certain period of time.
Now Voltage| This window shows the current battery level.
Reset| You can reset the device to factory default settings, e. g. if you have made an incorrect setting or if other problems with the settings occurred.
Language| In this window, you can change the output language. English or

German can be selected.

Connection| In this window, you can see the circuit diagram of the input port (see chapter 3.2).

Operation

5.1 Measurement
Edition with internal sensor:
After making the settings, you can start a measurement. To do so, you first need to attach a suitable adaptor. If required, you can also use the extension rod. Do not apply too much force when attaching the adaptor and/or the extension rod to avoid damage to the sensor.

Edition with external sensor:
Connect the sensor to the force gauge.
Then attach the suitable adaptor to the sensor without applying too much force as this can damage the sensor.

Switch on the device by pressing the On / Off key. You are now in the measurement window. Start by checking the battery level in the upper right corner. If the battery level is low, charge the device using the mains adaptor that comes with the device. For charging, it does not matter if the device is switched on or off. It is possible to make measurements during charging. When the device is fully charged, a notification will appear on the display. You should then disconnect the device from the power supply immediately.
You can now set the parameters. Press “OK” when you are in measurement mode to enter the settings window. Set the unit, the force area, zero tracking, the sampling speed,  fracture alarm, the upper and lower limit alarm, the Peak Hold function, the capture trigger as well as the capture length (duration). Press the Back key to return to measurement mode.
To start your measurement, attach the device to a suitable test stand. Press the Zero key and the arrow Right key. You can now take a direct measurement or capture a curve. If you make a direct measurement, the force will be measured in real-time, as well as the peak values and further parameters. These will not be saved. The previously measured values will be lost when a new measurement is taken. If you want to capture a curve, enter capture mode by pressing the “Arrow Left” key when you are in measurement mode.  The measurement will start automatically as soon as the trigger condition is met. The measurement ends if you press “OK“ or when the set capture duration has been reached.  The measurement will return peak values, measurement curves, and further parameters.
These will be saved. Only one curve with the associated values can be saved. You can retrieve the saved data by pressing the “Query“ button. The curve will be deleted when the force gauge is restarted or when a new measurement is taken.
A maximum of 100 reports can be saved if no curve is saved.  These can also be retrieved by pressing the “Query“ button. Press the Back key to get back to the measurement window.
To switch off the device, press the On/Off key. Remove the sensor if you have a device with an external sensor and clean the force gauge.
It is recommended to store the device in its original carrying case.

Maintenance

6.1 Storage
Please charge the battery before storing the meter for a longer period of time and store the device as well as, if applicable, the external sensors and the accessories in the packaging/carrying case the PCE-DFG N comes with to protect the technical components.

Appendix: load cell (1 … 100 kN)

|

L [mm]

| H [mm]| W1[mm]| H1 [mm]|

M [mm]

---|---|---|---|---|---
1 kN| 50,8| 76,2| 20,8| 15,3| M10 x 1,5
2,5 … 5 kN| 20,8| 76,2| 20,8| 13,6| M12 x 1,75
10 kN| 50,8| 76,2| 27,2| 12,1| M12 x 1,75
20 … 50 kN| 76,2| 101,6| 27,2| 20| M20 x 1,5
100 kN| 112,78| 177,8| 42,93| –| M30 x 2

|

A [mm]

| B [mm]| H [mm]| L [mm]| D [mm]| Max|

Weight

---|---|---|---|---|---|---|---
1 kN| 24| 45| 45| 20| M10| 0,25 t| 110 g
2,5 kN| 29| 53| 53| 22| M12| 0,4 t| 175 g
5 kN| 35| 62| 63| 22| M12| 0,7 t| 260 g
10 kN| 39| 71| 73| 22| M12| 1 t| 395 g
20 kN … 50kN| 40| 72| 71| 30| M20| 6 t| 450 g
100 kN| 60| 108| 109| 45| M30| 12 t| 1700 g

Warranty

You can read our warranty terms in our General Business Terms which you can find here: https://www.pce-instruments.com/english/terms.

Disposal

For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament applies.
Due to the contained pollutants, batteries must not be disposed of as household waste. They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU, we take our devices back. We either re-use them or give them to a recycling company that disposes of the devices in line with the law.
For countries outside the EU, batteries, and devices should be disposed of in accordance with your local waste regulations.
If you have any questions, please contact PCE Instruments.

www.pce-instrument.com

PCE Instruments contact information

Germany
PCE Deutschland GmbH
Im Langel 4
D-59872 Meschede
Deutschland
Tel.: +49 (0) 2903 976 99 0
Fax: +49 (0) 2903 976 99 29
info@pce-instruments.com
www.pce-instruments.com/deutsch | France
PCE Instruments France EURL
23, rue de Strasbourg
67250 Soultz-Sous-Forets
F r ance
Téléphone: +33 (0) 972 3537 17
Numéro de fax: +33 (0) 972 3537 18
info@pce-france.fr
www.pce-instruments.com/french | Spain
PCE Ibérica S.L.
Calle Mayor, 53
02500 Tobarra (Albacete)
España
Tel. : +34 967 543 548
Fax: +34 967 543 542
info@pce-iberica.es
www.pce-instruments.com/espanol 
---|---|---
United Kingdom
PCE Instruments UK Ltd
Unit 11 Southpoint Business Park
Ensign Way, Southampton
Hampshire
United Kingdom, SO31 4RF
Tel: +44 (0) 2380 98703 0
Fax: +44 (0) 2380 98703 9
info@pce-instruments.co.uk
www.pce-instruments.com/english| Italy
PCE Italia s.r.l.
Via Pesciatina 878 / B-Interno 6
55010 Loc. Gragnano
Capannori (Lucca)
Italia
Telefono: +39 0583 975 114
Fax: +39 0583 974 824
info@pce-italia.it
www.pce-instruments.com/italiano| Turkey
PCE Teknik Cihazları Ltd.Şti.
Halkalı Merkez Mah.
Pehlivan Sok. No.6/C
34303 Küçükçekmece – İstanbul
Türkiye
Tel: 0212 471 11 47
Faks: 0212 705 53 93
info@pce-cihazlari.com.tr
www.pce-instruments.com/turkish
The Netherlands
PCE Brookhuis B.V.
Institutenweg 15
7521 PH Enschede
Nederland
Telefoon: +31 (0)53 737 01 92
info@pcebenelux.nl
www.pce-instruments.com/dutch| Hong Kong
PCE Instruments HK Ltd.
Unit J, 21/F., COS Centre
56 Tsun Yip Street
Kwun Tong
Kowloon, Hong Kong
Tel: +852-301-84912
jyi@pce-instruments.com
www.pce-instruments.cn | United States of America
PCE Americas Inc.
1201 Jupiter Park Drive, Suite 8
Jupiter / Palm Beach
33458 FL
USA
Tel: +1 561-320-9162
Fax: +1 561-320-9176
info@pce-americas.com 

User manuals in various languages
can be downloaded here: www.pce-instruments.com
Specifications are subject to change without notice.

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals