Ritter 204 Examination Table User Guide

June 3, 2024
ritter

Ritter 204 Examination Table

Ritter-204-Examination-Table-FIG- \(2\)

Product Information

(The information below is required when calling for service)

  • Dealer [name/phone]:
  • Date of Purchase:
  • Model / Serial Number:
  • Midmark Authorized Service Company [name/phone]:

Ritter-204-Examination-Table-FIG- \(3\)

Important Information

Safety Symbols

WARNING
Indicates a potentially hazardous situation which could result in serious injury

Caution
Indicates a potentially hazardous situation which may result in minor or moderate injury. It may also be used to alert against unsafe practices

Equipment Alert
Indicates a potentially hazardous situation which could result in equipment damage

Note
Amplifies a procedure, practice, or condition

  • Proper shipping orientation

  • Fragile

  • Type B, Applied Part (Upholstery)

  • Protective earth ground

  • Patient Weight Limit

  • Manufacturer

  • Serial Number

  • Keep dry

  • Maximum stacking height (palletted units)

  • Fuse rating specification

  • Dangerous voltage / shock hazard

  • Refer to manual

  • Catalogue Number

  • Medical Device

Warnings

  • To avoid injury, do not put head or extremities under the paper tear strap.
  • Connecting equipment to the multiple socket outlet effectively leads to creating a Medical Electrical System and the result can be a reduced level of safety.
  • Table must be positioned in normal use such that the power cords can be easily accessed to unplug.
  • To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
  • Equipment is not suitable for use in the presence of a flammable anesthetic mixture.
  • Clarification : Equipment is suitable for use in the presence of oxygen, air, or nitrous oxide.
  • No modification of this equipment is allowed.

Intended Use

This product is intended to be used as a table to provide positioning and support of patients during examination procedures conducted by qualified medical professionals in an office environment.

Electrical Requirements

Caution

To completely isolate the table from electrical mains supply, both power cords must be unplugged.

When using high frequency surgical devices or endocardial catheters:

  • Use non-conductive material to insulate patient from metal portions of the table.
  • Consult operating instructions for the device before using in conjunction with drawer or upholstery heater.

Failure to comply may result in electric shock or burns to patient.

Electromagnetic Interference

This product is designed and built to minimize electromagnetic interference with other devices. However, if interference is noticed between another device and this product:

  • Remove interfering device from room.
  • Plug table into an isolated circuit.
  • Increase separation between table and interfering device.
  • Contact Midmark if interference persists.

Disposal of Equipment

At the end of product life, the table, accessories, and other consumable goods may become contaminated from normal use. Consult local codes and ordinances for proper disposal of equipment, accessories and other consumable goods.

Transportation / Storage Conditions

  • Ambient Temperature Range :………………………………..-5°C to +38°C (+23°F to 100°F)
  • Relative Humidity ………………………………………………..10% to 90% (non-condensing)
  • Atmospheric Pressure ………………………………………….500hPa to 1060hPa (0.49atm to 1.05atm)

Authorized Representatives

WARNING
In the event of any serious incident in relation to the device, please contact Midmark and the appropriate competent authority.

Customers in the EU should direct all questions, incidents, and complaints to Midmark’s Authorized Representative listed below.

  • CEpartner4U
  • Esdoornlaan 13
  • 3951 DB Maarn, The Netherlands
  • Phone: +31 343 442 524
  • Fax : +31 343 442 162

Customers in the UK should direct all questions, incidents, and complaints to Midmark’s UK Responsible Person listed below.

  • UKCApartner4U Ltd.,
  • 7 CAMPION WAY, BINGHAM, NOTTINGHAM NG13 8TR, UK.
  • www.UKCApartner4U.co.uk
  • Phone : +44 7905 384429

Customers in Australia should direct all questions, incidents, and complaints to Midmark’s Sponsor listed below.

  • ICONA Pty Ltd.
  • 1A/2A Westall Road
  • Clayton, Victoria 3168
  • Phone: 1300 442 662

Customers in Qatar should direct all questions, incidents, and complaints to Midmark’s Authorized Representative listed below.

Tadmur Trading
Midmac Roundabout
Opposite Al Mana Petrol Station
Doha, Qatar
Tel : +974-44337000
Fax : +974-44337100

Operation

Back Section Positioning

To position the back section…

  • A) While holding the back section in place, squeeze the Back Release Handle (either side).
  • B) Move the back section to desired position.
  • C) Release handle to lock back section in position

Ritter-204-Examination-Table-FIG- \(6\)

Stirrups

Caution
Be sure the stirrups are locked in place before using. The stirrups will not support the patient’s entire weight

To position the stirrups

  • A) Pull the stirrup out, then unfold.
  • B) Lift the stirrup slightly, then move it left or right as desired.
  • C) Release stirrup to lock in position.

Ritter-204-Examination-Table-FIG- \(8\)

Footrest & Treatment Pan

Caution
Do not use footrest to support patient’s entire weightRitter-204
-Examination-Table-FIG- \(9\).

To access the treatment pan
Lift footrest pad slightly, then slide pad back into table.

To extend footrest
Pull footrest out to desired position

Footstep

Caution
Ensure patients place their feet fully on the flat surface of the footstepwhen getting aboard and exiting the table.

To access the footstep…
Lift footstep slightly, then pull outward.

Paper Roll & Tear Strap

Paper Roll Size (max.):

  • Seamless Upholstery ………………………………………………………….. 21 in. wide x 3,5 in. diameter (53 cm x 9 cm)
  • Soft Touch Upholstery …………………………………………………………. 21 in. wide x 4.75 in. diameter (53 cm x 12 cm)

Pelvic Tilt (optional)

WARNING
To avoid risk of injury do not put extremities under the seat

To elevate seat section

  • Lift seat until pelvic support drops into place.

To return seat section to flat position

  • A) Lift seat section slightly.
  • B) Pivot trigger handle upward.
  • C) Lower seat section.

Ritter-204-Examination-Table-FIG- \(12\)

Receptacle (optional)

The duplex receptacle on the side of the table, provides power for accessories used duringmedical procedures. There are two circuit breakers located under the seat section. If the receptacle’s maximum load is exceeded, the circuit breakers interrupt power to the receptacle

  • Maximum Load ……………………………………………………….. 5 amps

To reset circuit breaker(s)
Push the circuit breaker toggle switchRitter-204-Examination-Table-FIG-
\(13\)

Maintenance Cleaning Upholstery

Equipment Alert
The upholstery is resistant to most medicinal-type stains, but may be damaged by solvents and dyes. Immediately remove any fluids spilled on the upholstery

  • Wash your upholstery weekly with a mild liquid soap and water mixture, rinse with clear water
    and dry completely to remove disinfection cleaner build-up.

  • Disinfect your upholstery with a solution of standard bleach and water mixed 1 in 10 (10%) or chlorine based cleaners. Follow this with a clear water rinse and thorough drying of material. See current CDC Guideline for Disinfection & Sterilization in Healthcare Facilities.

  • To minimize damage caused by disinfectant cleaner residue build-up, do not allow disinfectants to pool on the upholstery surface. Once the approved contact time has been obtained, remove and dry excess liquid remaining on the surface.

  • Detailed care and maintenance instructions are included with your product. This information is also available on www.midmark.com in the Technical library under the User Information tab for your product.

Painted Metal / Plastic Surfaces

Clean the painted metal and plastic surfaces weekly using a clean soft cloth, and mild cleaner

Preventative Maintenance

Periodically inspect the following areas:

  • Power cord should be free of cuts or other visible damage.
  • All fasteners should be present and tightened securely.
  • All mechanical functions should operate properly.

Have a qualified service technician inspect your equipment every twelve months

Calling For Service
If service is required, contact your authorized Midmark dealer

Note
Model / Serial Number information is required when calling for service

To contact Midmark directly:

  • 1.800.MIDMARK (1.800.643.6275)
  • 8:00 am until 5:00 pm Monday – Friday (ET) [excluding standard US holidays]

Specifications

Weights & Dimensions

Maximum Patient Weight: 500 lbs (226 kg)
Paper Roll (maximum size): [See Paper Roll & Tear Strap page]
Range of Motion & Dimensions: (See illustration below)
Weight of Table: 220 lbs (99.8 kg)
Power Cord: 6 ft. (1.8 m) long

Ritter-204-Examination-Table-FIG- \(14\)

Upholstery DimensionsRitter-204-Examination-Table-FIG-
\(15\)

Voltage, Current, Frequency Ratings & Compliance Chart

__

__

Model

| __

__

Description

|  | Electrical Ratings:
---|---|---|---
__

ES 60601-1

| CAN / CSA C22.2, 60601-1| __

CE

| VAC

+/- 10%

| __

Amps

| Cycles (Hz)
204- 005

204- 013

| Ritter ® Exam Table – Export

(Pass-thru Drawers) w/ Pelvic Tilt

| | | | __

| __

| __

204- 006

204- 014

| Ritter ® Exam Table – Export

(Pass-thru Drawers)

w/ Receptacle & Pelvic Tilt

| | | | __

230

| __

5

| __

50/60

Midmark® + Ritter® Medical Equipment (Tables, Chairs, Physician Seating, Lights, Workstations and Sterilizers) Limited Warranty

Scope of Warranty

SCOPE OF WARRANTY

Midmark Corporation (“Midmark”) warrants to the original retail purchaser that it will, at Midmark’s option, repair or
replace components of the domestic and international medical products manufactured by Midmark (except for components not warranted under “Exclusions”) that are defective in material or workmanship under normal use and service. The sole remedy under this limited warran-ty is the repair or replacement, at Midmark’s option, of the applicable components. This limited warranty shall only apply to defects that are reported to Midmark within the applicable warranty period and which are determined to exist upon examination by Midmark,. This warranty extends only to the original retail purchaser of a product and is not transferable or assignable. Replacement components or products may be used and/or refurbished components or products, provided they are of like quality and specifi cations as new components or products.
Midmark warrants to the original retail purchaser that during the applicable warranty period it will repair or replace software contained within the products manufactured by Midmark (except for those not warranted under “Exclusions”) if: (1) the media on which the software is furnished exhibits defects in material or workmanship under normal use; or (2) the software does not substantially conform to its published specifi cations.

APPLICABLE WARRANTY PERIOD

The applicable warranty period, measured from the date of invoice to the original retail purchaser of the product and shall be one (1) year for all warranted products and components. QuickClean® Ultrasonic Cleaners are warranted for a period of three (3) years.

OBTAINING WARRANTY SERVICE

Warranty service must be obtained through either Midmark or an authorized dealer in the Midmark product line for which warranty service is requested. Midmark may be contacted for warranty service inquiries or issues via email at midmark.com; by phone at 1.800.MIDMARK or by mail to Midmark Corporation, 60 Vista Drive, Versailles, Ohio 45380. It is the retail pur-chaser’s obligation to arrange for delivery of a product to Midmark or one of its authorized dealers for warranty service, which delivery shall be at retail purchaser’s expense. It is also the retail purchaser’s obligation to comply with the warranty service instructions provided either by Midmark or its authorized dealer. The retail purchaser must provide Midmark with completed warranty registration information within thirty (30) days after purchase in order to obtain the benefi ts of this limited warranty.

EXCLUSIONS: This limited warranty does not cover and Midmark shall not be liable for the following:

  1. Defects, damage or other conditions caused, in whole or in part, by misuse, abuse, negligence, alteration, accident, freight damage, negligent storage, tampering or failure to seek and obtain repair or replacement in a timely manner;
  2. Products which are not installed, used, and properly cleaned and maintained as required or recommended in the Midmark Installation and/or Installation/Operation Manual for the applicable product, including the specifi ed structural and operational environmental conditions and electrical requirements;
  3. Products considered to be of a consumable or sterile nature;
  4. Accessories or parts not manufactured by Midmark;
  5. Charges by anyone for adjustments, repairs, replacement parts, installation or other work performed upon or in connection with such products which are not expressly authorized in writing in advance by Midmark;
  6. Costs and expenses of routine maintenance and cleaning; and
  7. Representations and warranties made by any person or entity other than Midmark.
  8. Matching of color, grain or texture except to commercially acceptable standards;
  9. Changes in color caused by natural or artifi cial light;
  10. Custom manufactured products;
  11.  Alterations or modifi cations to the product by any person or entity other than Midmark; and
  12. Products that would otherwise by covered under this limited warranty, but are acquired: (i) from a person or entity that is not Midmark or one of its authorized dealers; or (ii) from a Midmark dealer that is not authorized to sell the product at issue in the geographic territory where the purchaser is located, or is not authorized to sell the product at issue within the medical market.

SOFTWARE;

WITH RESPECT TO SOFTWARE THAT IS A PRODUCT OR COMPONENT THEREOF, MIDMARK DOES NOT WARRANT THAT THE SOFTWARE: (1) IS ERROR FREE; (2) CAN BE USED WITHOUT PROBLEMS OR INTERRUPTIONS; OR (3) IS FREE FROM VULNERABILITY TO INTRUSION OR ATTACK BY VIRUSES OR OTHER METHODS.

EXCLUSIVE REMEDY

CONSEQUENTIAL DAMAGES DISCLAIMER MIDMARK’S ONLY OBLIGATION UNDER THIS LIMITED WARRAN-TY IS THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS. MIDMARK SHALL NOT BE LIABLE FOR AND HEREBY DISCLAIMS ANY DIRECT, SPECIAL, INDIRECT, INCIDENTAL, EXEMPLARY OR CONSEQUENTIAL DAMAGES OR DELAYS, INCLUDING, BUT NOT LIMITED TO, DAMAGES FOR LOSS OF PROFITS OR INCOME, LOSS OF USE, LOSS OF DATA, DOWNTIME, COVER AND EMPLOYEE OR INDEPENDENT CONTRACTOR WAGES, PAYMENTS AND BENEFITS. THIS DISCLAIMER SHALL SURVIVE ANY FAILURE OR ASSERTED FAILURE OF THE ESSENTIAL PURPOSE OF THIS LIMITED WARRANTY OR ITS REMEDIES SPECIFIED HEREIN. WARRANTY DISCLAIMER THIS WARRANTY IS MIDMARK’S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRAN-TIES, EXPRESS OR IMPLIED. MIDMARK MAKES NO IMPLIED WARRANTIES OF ANY KIND INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACE-MENT OF DEFECTIVE PARTS.

STATUTE OF LIMITATIONS

No action may be brought against Midmark for breach of this limited warranty, an implied warranty, if any, or for any other claim arising out of or relating to the products, more than ninety (90) days following expiration of the limited warranty period.

NO AUTHORIZATION

No person or fi rm is authorized to create or approve for Midmark any other obligation or liability in connection with the products..

Midmark Corporation
60 Vista Drive Versailles, OH 45380 USA+1.800.643.6275
+1.937.526.3662
www.midmark.com