Roche cobas h 232 System User Guide

June 3, 2024
Roche

cobas h 232 System
User Guide

Quick Reference Guide

IQC Checking

IQC timeout at 8 AM daily

Roche cobas h 232 System fig 1

Power on the h 232| Press Control Test| Insert the IQC test strips (Low and High)
Roche cobas h 232 System fig 2
The measurement begins automatically| The result is displayed as Pass or Fail| Return the test strip quickly back into the container

■ Daily Cleaning (70% alcohol)

Roche cobas h 232 System fig 3

A) Clean the meter surface with moistened tissue B) Remove and wipe the green cover| Clean the inside area with cotton swab| * Be gentle & caution when cleaning the detection window on the side

Routine Testing

Roche cobas h 232 System fig 4

Use heparinized venous whole blood only| Gently swirl the tube for 10 times| Power on the h 232
Roche cobas h 232 System fig 5
Press the “Patient Test” Button| Enter or scan the patient ID| Insert the test strip
Roche cobas h 232 System fig 6
The test strip will warm up automatically| Draw exactly 150 μL sample using a Roche CARDIAC Pipette
in between the 2 blue arrows| Apply the sample within 5
minutes. Then press the check
button.*

Roche cobas h 232 System fig 7

  • Hold the pipette vertically over the sample application area. Slowly apply the sample to avoid spillage and losing of samples.
    The Roche CARDIAC Pipette tip should not touch the sample application area of the test strip.

https://vimeo.com/757443233
Technical Support:
+852 5729 6447 (Max Tsang)
+ 852 5729 3919 (Jeffrey Chan)
Roche Diagnostics (Hong Kong) Limited
Video Guide

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals