PeakTech 3296 Analog Voltmeter User Manual
- June 3, 2024
- PeakTech
Table of Contents
- PeakTech 3296 Analog Voltmeter
- Safety Precautions
- Precautions for Use
- Features
- Accessories
- Description
- Function Selection
- Reference Conditions
- Specifications
- General Characteristics
- Compliance with International Standards
- Electromagnetic Compatibility
- Warranty
- Maintenance
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
PeakTech 3296 Analog Voltmeter
Safety Precautions
This product complies with the requirements of the following directives of the
European Union for CE conformity: 2014/30/EU (electromagnetic compatibility),
2014/35/EU (low voltage), 2011/65/EU (RoHS).
Overvoltage Category II 1000V, Pollution degree 2.
To ensure safe operation of the equipment, the following safety precautions
must be observed.
Damages resulting from failure to observe these safety precautions are exempt
from any legal claims whatever.
- Do not place water filled containers on the equipment (danger of short- circuit in case of knock over of the container).
- Do not operate the equipment near strong magnetic fields (motors, transformers etc.).
- Do not operate the meter before the cabinet has been closed.
- Check instrument and accessories for any damages.
- Comply with the warning labels and other info on the equipment.
- Do not subject the equipment to direct sunlight or extreme temperatures, humidity or dampness.
- Do not subject the equipment to shocks or strong vibrations.
- Keep hot soldering irons or guns away from the equipment.
- Allow the equipment to stabilize at room temperature before taking up measurement (important for exact measurements).
- The meter is suitable for indoor use only
- Do not store the meter in a place of explosive, inflammable substances.
- Do not modify the equipment in any way.
- Opening the equipment and service – and repair work must only be performed by qualified service personnel
- Measuring instruments don’t belong to children hands.
Cleaning the cabinet
Clean only with a damp, soft cloth and a commercially available mild household cleanser. Ensure that no water gets inside the equipment to prevent possible shorts and damage to the equipment.
Meaning of the symbol
CAUTION! Please consult the operating instructions before using the
device. In these operating instructions, failure to follow or carry out
instructions preceded by this symbol may result in personal injury or damage
to the device and the installations.
Meaning of the symbol
This appliance is protected by double insulation or reinforced insulation. It
does not have to be connected to an earth protection terminal for electrical
safety.
To obtain the best service from your unit:
- Read these operating instructions carefully,
- Comply with the precautions for use.
Precautions for Use
- Never use on a voltage network over 1000Vpp with respect to the earth connection. This voltage surge category II voltmeter complies with stringent reliability and availability requirements, corresponding to fixed industrial and domestic installations (see IEC 664-1).
- Do not use on alternate and continuous voltages > 1000Vpp.
- Indoor use in environments with a maximum pollution level of 2 (EN 50419:2006), temperature of -10 °C to +50 °C and relative humidity below 90%.
- Respect the value and type of the fuses to avoid damaging the instrument and canceling the warranty.
- Fuse: 250V/0.5A quick-acting glass tube, 5x20mm.
- Use accessories corresponding to safety standards (EN 61010-1:2001) with 1000V min. voltage ratings and surge category II.
- Before any measurement, ensure correct positioning of the leads on the ammeter. When the value range of a measurement is not known, begin at the highest range, and then gradually reduce it until the appropriate range is achieved: the reading should preferably be in the upper 2/3 of the full scale deflection.
- The leads must be disconnected to open the lower half of the meter case.
- Never connect to the circuit to be measured if the casing is not properly closed.
Features
- Analog mirror scale
- Robust Plastic housing
- DC and AC voltage measurements
- No batteries required
Accessories
- Operation manual
Description
This voltmeter was developed for everyday use by electricity professionals and
persons acquainted with electricity. It offers the following functions:
Voltage measurement (VDC and VAC)
Safety Terminals, Ø4mm
- COM – Common terminals for VAC and VDC, respectively
- 0.1V, 1V, 10V, 100V, 1kV – Terminals for VDC Ranges, 3V, 10V, 100V, 1kV – for VAC Ranges
Analogue Scale
One linear black scale, with anti-parallax mirror, for VDC & VAC. One scale for the 3V VAC range.
Function Selection
Reference Conditions
- Temperature: 23°C ± 2°C
- Humidity: 45% RH ± 5%
- Position: horizontal ± 2°
Ensure that the pointer is at zero before starting any measurements.
Mechanical zero adjustment is carried out by turning the screw on the front of
the voltmeter.
When an estimated voltage value is unknown, start measurement at the highest
range, and then gradually reduce it until the appropriate range is achieved:
the reading should preferably be in the upper 2/3 of the full scale
deflection.
Specifications
DC Voltage
V-DC
Range
| Reading coefficient| Internal Resistance| Accuracy| Admissive Overload
---|---|---|---|---
0.1V| ×1| 2k| ****
2.0% of F.S.
| 250V
1V| ×0.01| 20k| 250V
10V| ×0.1| 200k| 1kV
100V| ×1| 2M| 1kV
1000V| ×10| 20M| 1kV
AC Voltage (Sine Wave)
A-AC
Range
| Reading coefficient| Internal Resistance| Accuracy| Admissive Overload
---|---|---|---|---
3V| ×0.01| 27k| ****
3.0% of F.S.
| 250V
10V| ×0.1| 90k| 1kV
100V| ×1| 900k| 1kV
1000V| ×10| 9M| 1kV
The presence of a DC component falsifies the measurement. Bandwidth: 40Hz- 1kHz.
General Characteristics
Dimensions and Weight
- Dimensions: 200×120×5 mm
- Weight: 450 g
Maximum Climatic Conditions
- Temperature (use): -0°C to +40°C
- Temperature (storage): -20°C to +50°C
- Relative humidity (use): ≤80% RH
- Altitude (use): <2000m
Compliance with International Standards
Electrical safety (EN 61010-1 -2001)
CEI 1010-1 EN61010 NF-C 42020 VDE 0411
- Double insulation:
- Pollution level: 2
- Installation category: II according to CEI 664
- Allocated voltage: 1000V
Electromagnetic Compatibility
- EN 61326-1:2013
- EN 61326-2-2:2013
Warranty
Our warranty lasts three years for the voltmeter after the delivery date.
Maintenance
Only use the specified spare parts for maintenance. The manufacturer shall not
be held liable for any incident occurring following repairs carried out by a
party other than its after-sales service or approved repairers.
Fuse Change
Open the device. Use the same type of fuse to ensure the safety of users and
of the device.
All rights, also for translation, reprinting and copy of this manual or parts
are reserved. Reproductions of all kinds (photocopy, microfilm or other) only
by written permission of the publisher.
This manual is according the latest technical knowing. Technical alterations
reserved.
We herewith confirm that the units are calibrated by the factory according to
the specifications as per the technical specifications. We recommend to
calibrate the unit again, after 1 year.
© PeakTech® 07/2021 Mi./Ehr.
PeakTech Prüf- und Messtechnik GmbH – Gerstenstieg 4 – DE-22926 Ahrensburg /
Germany
Tel: +49-(0) 4102-97398 80
+49-(0) 4102-97398 99
E-mail: info@peaktech.de
Website: www.peaktech.de
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>