Time Electronics 1067 Precision Resistance Decade Box User Manual
- June 3, 2024
- Time Electronics
Table of Contents
User Manual 1067
Precision Resistance Decade Box
Version 1.2 11-22
User Manual
1067 Precision Resistance Decade Box
© 2022 Time Electronics Ltd.
All rights reserved.
Nothing from this manual may be multiplied, or made public in any form or
manner, either electronically or hard copy, without prior written consent from
Time Electronics Ltd. This also applies to any schematics, drawings and
diagrams contained herein.
This manual provides operating and safety instructions for the Time
Electronics product. To ensure correct operation and safety, please follow the
instructions in this manual.
Time Electronics reserves the right to change the contents, specifications and
other information contained in this manual without notice.
Time Electronics
1067 Resistance Decade Box
Introduction
A precision resistance decade box suitable for a wide range of simulation
work. High accuracy, long term stability, and low temperature coefficient make
the 1067 ideal for
simulating and calibrating precision Pt100 sensors and temperature
indicators/meters that use resistive sensors.
Special care has been taken in the construction of this precision decade box
to ensure that the residual end resistance is as low and as stable as
possible. Multiple self-cleaning silver alloy contacts are used for each
position to ensure outstanding performance and long life. Housed in a robust
metal case the 1067 is fully screened and low thermal EMF terminals are used.
The slimline design means it takes up minimum bench space and is easily
transportable. Resistance is selected by dialling the value required using the
rotary switches. This enables precise setting with a clear indication. Each
decade is scaled from 0 to 11 and therefore allows convenient overlap of the
set values. The maximum value settable is 12,222.21 ohms.
Features
- 10 mΩ to 12 kΩ
- 0.01 % accuracy
- Precision Pt100 simulation
- 6-digit resolution
- Better than 20 ppm/year stability
- Residual resistance < 10 mΩ
- Fully screened
Specifications
Range / Resolution …………………. 0 to 12 kΩ / 10 mΩ steps.
Number of decades …………………. 6. Each settable from 1 to 11.
Accuracy ………………………………… ± 0.01 % of setting ± 2 mΩ, after deduction of residual
end
(At calibration temperature of 22°C) …………….. resistance ± 1 mΩ for residual
variation.
Current rating …………………………. 10 mΩ range: 3 A. 100 mΩ range: 2 A. 1 Ω range:
600 mA. 10 Ω range: 200 mA. 100 Ω range: 60 mA. 1 kΩ range: 20 mA.
Residual resistance ………………… Less than 10 mΩ.
Residual resistance stability ……. Less than 1 mΩ.
Voltage rating …………………………. 200 V DC at maximum resistance setting.
Temperature coefficient ………….. Less than 10 ppm per °C (> 1 Ω). .. Less than
20 ppm per °C (< 1 Ω).
Stability ………………………………….. Better than 20 ppm per year (> 1 Ω). Better than 100
ppm per year (< 1 Ω).
Insulation ……………………………….. Case to resistance terminals 2 kV DC maximum.
Connections …………………………… 2 x 4 mm active safety terminals. A third safety
terminal is to enable the case to be earthed or connected to either output.
Contacts …………………………………. Make before break – silver alloy.
Operating torque …………………….. Less than 0.1 Nm.
Dimensions …………………………….. W 355 x H 63 x D 82 mm.
Weight ……………………………………. 1.1 kg.
Options ………………………………….. Calibration certificates: Traceable (factory) and
accredited (ISO 17025).
Country of origin …………………….. United Kingdom.
Operation
3.1 Safety Precautions
** Observe proper safety operational guidelines when working with high
voltages.
To minimize shock hazard connect the case terminal to an electrical ground.
Always take precautions to avoid and prevent contact with live components.
Handle the unit with care and use as per the instructions in this manual.
3.2 Front Panel Controls and Connections**
- Resistance Terminals: Resistance is connected via the safety terminal binding posts that are suitable for twisted stripped wire compression connection, spade terminals, or by 4 mm shrouded or normal plug insertion.
- Case Terminal: The case terminal is isolated from the two active resistance terminals. When connected to ground/earth, it may be used as a guard or shield connection, this can help to reduce unwanted electrical noise pickup, and help maintain the case at a safe voltage in certain modes of use.
- Resistance Setting Dials: Used for selecting the required resistance by setting the dial to the value of the decade range. Each dial can be set from 0 to 11. The value is clearly displayed in the white setting window below the decade range.
3.3 Operating Instructions
3.3.1 Connections![Time Electronics 1067 Precision Resistance Decade Box
- fig](https://manuals.plus/wp-content/uploads/2022/11/Time-Electronics-1067 -Precision-Resistance-Decade-Box-fig.png)
Connection to the decade box is via 4 mm safety terminal posts, using 4 mm
shrouded or standard plugs. Alternatively, crocodile clips or stripped wire
connections can be used. Whatever method is used, the connection must be tight
to the terminal posts to avoid introducing unwanted additional resistance.
The red and black active terminals connect to the resistance elements, and the
green terminal is connected to the case for screening purposes.
** For certain applications, the user may want to connect the case terminal
to either of the active terminals. This can be done, but the case would then
be at the same potential as the active terminals. The user should be aware
that this could be hazardous and safety precautions must be taken to prevent
electric shock.
3.3.2 Setting Resistance**
Use the front panel dials to set the required values according to the decade
ranges. The selected value shown in the white setting window under the range
arrow. This enables
precise setting with a clear unambiguous indication.
For example, to set 138.51 Ω equivalent to 100 °C Pt100 (ITS-90):
- Set the 1 kΩ dial to 0.
- Set the 100 Ω dial to 1.
- Set the 10 Ω dial to 3.
- Set the 1 Ω dial to 8.
- Set the 0.1 Ω dial to 5.
- Set the 0.01 Ω dial to 1.
For a quick reference to the setting values of the 1067, please see the Resistance Setting Table on the following page.
Note: All resistance boxes have a residual resistance, meaning that when
the dials are set to zero a small resistance remains. If you are making
precision measurements or recalibrating the instrument, this residual value
must be subtracted from all measurements.
Typical values of residual resistance are shown in the specifications.
For precise values relating to your specific unit, refer to calibration
certificate (if ordered).
3.4 Resistance Setting Table
The below table shows the settings of each resistance decade using the
dials.
1 K Ω Decade| 100 Ω Decade| 10 Ω Decade| 1 Ω Decade|
0.1 Ω Decade| 0.01 Ω Decade
---|---|---|---|---|---
Dial| Ω| Dial| 0 Ω| Dial| Ω| Dial| Ω| Dial| Ω| Dial| Ω
o| 1k Ω| o| O Ω| 0| 0| Ω| 0 Ω| 0| 0Ω| 0| 0 Ω
1| 1 k Ω| 1| 100 Ω| 1| 100| 1| 10 Ω| 1| 0.1Ω| 1| 0.01 Ω
2| 2k Ω| 2| 200 Ω| 2| 200| 2| 20 Ω| 2| 0.2Ω| 2| 0.02 Ω
3| 3 k Ω| 3| 300 Ω| 3| 300| 3| 30 Ω| 3| 0.3Ω| 3| 0.03 Ω
4| 4k Ω| 4| 400 Ω| 4| 400| 4| 40 Ω| 4| 0.4Ω| 4| 0.04 Ω
5| 5 k Ω| 5| 500 Ω| 5| 500| 5| 50 Ω| 5| 0.5Ω| 5| 0.05Ω
6| 6k Ω| 6| 600 Ω| 6| 600| 6| 60 Ω| 6| 0.6Ω| 6| 0.06Ω
7| 7 K Ω| 7| 700 Ω| 7| 700| 7| 70 Ω| 7| 0.7Ω| 7| 0.07Ω
8| 8k Ω| 8| 800 Ω| 8| 800| 8| 80 Ω| 8| 0.8Ω| 8| 0.08Ω
9| 9k Ω| 9| 900 Ω| 9| 900| 9| 90 Ω| 9| 0.9Ω| 9| 0.09Ω
10| 10k Ω| 10| 1 k Ω| 10| 1000| 10| 100 Ω| 10| 1.0Ω| 10| 0.10Ω
11| 11 k Ω| 11| 11 k Ω| 11| 1100| 11| 110 Ω| 11| 1.1Ω| 11| 0.11Ω
Applications
Pt100 Resistance Simulation
Accurate simulation of low ohm values, such as PRT, can be performed by using
the 1067.
Resistance vs Temperature Relationship for Platinum Resistance Thermometer
Detector Element (ITS-90)
°C | Ω | °C | Ω | °C | Ω | °C | Ω |
---|---|---|---|---|---|---|---|
-200 | 18.52 | 60 | 123.24 | 320 | 219.15 | 580 | 307.25 |
-180 | 27.1 | 80 | 130.9 | 340 | 226.21 | 600 | 313.71 |
-160 | 35.54 | 100 | 138.51 | 360 | 233.21 | 620 | 320.12 |
-140 | 43.88 | 120 | 146.07 | 380 | 240.18 | 640 | 326.48 |
-120 | 52.11 | 140 | 153.58 | 400 | 247.09 | 660 | 332.79 |
-100 | 60.26 | 160 | 161.05 | 420 | 253.96 | 680 | 339.06 |
-80 | 68.33 | 180 | 168.48 | 440 | 260.78 | 700 | 345.28 |
-60 | 76.33 | 200 | 175.86 | 460 | 267.56 | 720 | 351.46 |
-40 | 84.27 | 220 | 183.19 | 480 | 274.29 | 740 | 357.59 |
-20 | 92.16 | 240 | 190.47 | 500 | 280.98 | 760 | 363.67 |
0 | 100 | 260 | 197.71 | 520 | 287.62 | 780 | 369.71 |
20 | 107.79 | 280 | 204.9 | 540 | 294.21 | 800 | 375.7 |
40 | 115.54 | 300 | 212.05 | 560 | 300.75 | 820 | 381.65 |
Typical connections for 2, 3 and 4 wire resistance thermometers
Connection Diagram | Notes |
---|---|
2-wire Pt100 Connection |
The connecting lead/wire resistance must be factored.
Residual resistance is factored to calculate best setting
accuracy, if required.
| 3-wire Pt100 Connection
Leads/wires must be of the same length, gauge, and resistance, meaning the
connection is compensated. Residual resistance is factored to calculate best
setting accuracy, if required.
| 4-wire Pt100 Connection
Provides the most accurate measurement, not being affected by any differences
in the wires/leads used. Residual resistance is factored to calculate best
setting accuracy, if required.
Warranty and Servicing
Warranty
Time Electronics products carry a one-year manufacturer’s warranty as
standard. Time Electronics products are designed and manufactured to the
highest standards and
specifications to assure the quality and performance required by all sectors
of industry. Time Electronics products are fully guaranteed against faulty
materials and workmanship. Should this product be found to be defective,
please contact us using the below details. Inform us of the product type,
serial number, and details of any fault and/or the service required. Please
retain the supplier invoice as proof of purchase.
This warranty does not apply to defects resulting from action of the user such
as misuse, operation outside of specification, improper maintenance or repair,
or unauthorized
modification. Time Electronics’ total liability is limited to repair or
replacement of the pr oduct. Note that if Time Electronics determine that the
fault on a returned product has been caused by the user, we will contact the
customer before proceeding with any repair.
Product Registration
You can register your product at: www.timeelectronics.com/contact/product-
registration.
Registering your product will enable us to maintain a record of purchase for
your warranty. You can also use the web form to provide feedback about our
products and services.
Calibration and Repair Services
Time Electronics offers repair and calibration services for all the products
we make and sell. Routine maintenance by the manufacturer ensures optimal
performance and condition of the product. Periodic traceable or accredited
calibration is available.
Contacting Time Electronics
Online:
Please visit www.timeelectronics.com and
select Technical Support from the Contact links.
From this page you will be able to send information to the Time Electronics
service team who will help and support you.
By phone: +44 (0) 1732 355993
By email: mail@timeelectronics.co.uk
Returning Instruments
Prior to returning your product please contact Time Electronics. We will issue
a return merchandise authorization (RMA) number that is to accompany the goods
returning.
Further instructions will also be issued prior to shipment. When returning
instruments, please ensure that they have been adequately packed, preferably
in the original packing supplied. Time Electronics Ltd will not accept
responsibility for units returned damaged. Please ensure that all units have
details of the service required and all relevant paperwork. Send the
instrument, shipping charges paid to: Time Electronics Ltd
Unit 5, TON Business Park, 2-8 Morley Road, Tonbridge, Kent, TN9 1RA.
United Kingdom.
Tel: +44(0)1732 355993
Fax: +44(0)1732 350198
Email: mail@timeelectronics.co.uk
Web Site: www.timeelectronics.com
Disposal of your old equipment
- When this crossed-out wheeled bin symbol is attached to a product it means the product is covered by the European Directive 2002/96/EC.
- All electrical and electronic products should be disposed of separately from the municipal waste stream via designated collection facilities appointed by the government or the local authorities.
- The correct disposal of your old appliance will help prevent potential negative consequences for the environment and human health.
- For more detailed information about disposal of your old appliance, please contact your city office, waste disposal service or return to Time Electronics.
Time Electronics Ltd
Unit 5, TON Business Park, 2-8 Morley Road,
Tonbridge, Kent, TN9 1RA, United Kingdom.
T: +44 (0) 1732 355993
F: +44 (0) 1732 350198
mail@timeelectronics.co.uk
www.timeelectronics.com
References
- Time Electronics – Precision Test Instruments, Calibration Benches
- Product Registration Form | Time Electronics
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>