Time Electronics 7115 Contamination Trap User Manual
- June 3, 2024
- Time Electronics
Table of Contents
User Manual
7115 Contamination Trap
Version 1.2
9-21
7115 Contamination Trap
All rights reserved.
Nothing from this manual may be multiplied, or made public in any form or
manner, either electronically or hard copy, without prior written consent from
Time Electronics Ltd.
This also applies to any schematics, drawings and diagrams contained herein.
This manual provides operating and safety instructions for the Time
Electronics product.
To ensure correct operation and safety, please follow the instructions in this
manual.
Time Electronics reserves the right to change the contents, specifications and
other information contained in this manual without notice.
Warnings
The 7115 contamination trap is design for use with pressure controllers,
calibrators and regulated pneumatic pressure systems. These systems control a
supplied pressure source and output them to a unit under test. The 7115 is
connected between the controller and the UUT. Please observe all warnings and
safety instructions when dealing with high pressure supplies and outputs.
Please ensure that all users of the instrumentation have fully read and
understood operating instructions and safety procedures for handling high
pressure systems.
Observe all warnings before operating pressure calibration systems (ie 8030,
RMTEG etc).
The 7115 shall only be used with clean dry air or nitrogen.
Never use hazardous media as pressure medium.
Do not exceed 210bar input to the 7115.
Before pressurisation of the 7115, check all components and connections are in
good condition, fully functional and that all screw fittings are firmly. All
connections should be suitable for the applied maximum pressure.
In the case of an error in connections or use, a high pressure or vacuum may
be on the input and output connections. Ensure supplies are vented before
working on pressure connections.
During operation, do not disconnect the 7115 when pressure is being applied.
When working on and with the device wear safety glasses.
Connections and UUTs damaged from overpressure can cause high velocity
shrapnel.
Note that when a device (such as an analogue pressure gauge) is to be tested
or calibrated, it is referred to as the UUT (Unit Under Test) in this manual.
The UUT should be connected to the 7115 using suitable fittings and hose
ensuring they are rated for at least the maximum pressure available from the
pressure controller or calibrator supplying pressure.
Time Electronics Ltd do not assume liability for damages that arise from
incorrect use of the instrument device or from disregard of the information
contained in this manual.
Introduction
The 7115 is a contamination trap designed to protect pressure equipment, such
as controllers, calibrators and pressure generators, from impurities, liquids
(water, oil etc.) and
residues released by manifolds, pressure gauges, pressure transmitters and
other process instruments under test.
It has been designed primarily for use with the Time Electronics 8030 pressure
controller.
The 7115 provides a level of protection for the internal system components
that require mechanical operation to be smooth and unobstructed, to ensure
correct operation of the controller. Contamination can damage the 8030 control
mechanisms.
During transitions from a high pressure setpoint to a smaller one, liquid and
residues present in the pneumatic system may flow into the 8030.
For example, consider an 8030 controlling to 20 bar on a UUT, and the setpoint
is then changed to 2 bar. In order to decrease the pressure, the 8030 must
remove air from the system. Such air removal can cause liquids and residues
present in the UUT to flow into the 8030, damaging it and compromising the
operation of the control valves. This may also occur when selecting negative
setpoints (vacuum).
The contamination trap must be placed between the 8030 and the UUT in order to
avoid that liquid (water, oil etc.) and residues from these process
instruments damage the internal system of the controller. The 7115 is a trap
that is supplied as standard with the 8030 controllers.
2.1 Specifications
Pressure Range: | -1 to 210bar |
---|---|
Pressure Media: | Pneumatic only – Gas or air |
Material: | Stainless steel, polycarbonate and nitrile rubber seals. |
Connections: | 2 x Adaptors – minimess, hoses supplied. |
Weight: | 1 kg |
Operating Temperature: | 10 to 40°C (50 to 104°F) |
Storage Temperature: | 0 to 70°C (32 to 158°F) |
Air Humidity: | 0 to 95% (relative humidity without moisture condensation) |
2.2 7115 Photos
Using the 7115
The Contamination Trap must be used always in the upright position to prevent
liquid from flowing into the pressure controller or calibrator.
This guide will show the connections and usage with the 8030/8030B Pressure
Controllers.
8030B is the module version of the 8030.
The CLEAN port must be connected to the 8030 pressure controller. The DIRTY
(UUT) port must be connected to the manifold, manometer, pressure switch,
transmitter or other process instruments that will be calibrated or adjusted.
Example Connections:
The 8030 should be used with a liquid trap to ensure that no contamination
reaches the instrument. UUT’s can be dirty, with oil and water capable of
getting into hoses and affecting instrument performance.
Minimess hoses are supplied for connection from the 8030 to the 7115. An additional hose is supplied for connection from the 7115 to the UUT. If an manifold is supplied this will be a minimess to minimess type. A minimess to ¼ BSP or ¼ NPT hoses will be supplied if required.
The 7198 pressure adaptor kit can also be used to make connections from the 7115 to various UUTs with different thread types.
3.1 Example Set Ups
3.1.1 8030 > 7115 > Manifold > UUT
3.1.2 8030 > 7115 > UUT
3.1.3 8030B > 7115 > 8100 Manifold > UUT
Cleaning and Maintenance
When the liquid accumulated inside the contamination trap reaches the level of the boundary line, the liquid can be removed as follows:
-
Disconnect both hoses from the contamination trap.
-
Using two 19 mm (3/4”) spanners loosen the contamination trap via the hexagonal nuts so that it may be dismantled by hand.
-
Hold the water trap over a sink or receptacle and allow the water/contaminant to flow out.
-
Dismantle the water trap completely.
|
---|--- -
Using clean water, rinse the components.
-
Using clean tissue dry thoroughly, ensuring all contaminates are removed.
Note: Do not use solvents.
Note: Do not use grease on the sealing “O”-rings. -
Reassemble hand tight and then using the two 19mm (3/4”) spanners tighten the assembly by ¼ of a turn clockwise. Do not over-tighten the hexagonal nuts.
Components List
Warranty and Servicing
Warranty
The Time Electronics products carry a one-year manufacturer’s warranty as
standard.
Time Electronics products are designed and manufactured to the highest
standards and specifications to assure the quality and performance required by
all sectors of industry.
Time Electronics products are fully guaranteed against faulty materials and
workmanship.
Should this product be found to be defective, please contact us using the
below details.
Inform us of the product type, serial number, and details of any fault and/or
the service required. Please retain the supplier invoice as proof of purchase.
This warranty does not apply to defects resulting from action of the user such
as misuse, operation outside of specification, improper maintenance or repair,
or unauthorized modification. Time Electronics’ total liability is limited to
repair or replacement of the product.
Note that if Time Electronics determine that the fault on a returned product
has been caused by the user, we will contact the customer before proceeding
with any repair.
Calibration and Repair Services
Time Electronics offers repair and calibration services for all the products
we make and sell.
Routine maintenance by the manufacturer ensures optimal performance and
condition of the product. Periodic traceable or accredited calibration is
available.
Contacting Time Electronics
Online:
Please visit www.timeelectronics.com and
select Support Request from the Contact links.
From this page you will be able to send information to the Time Electronics
service team who will help and support you.
By phone: +44 (0) 1732 355993
By email: mail@timeelectronics.co.uk
Returning Instruments
Prior to returning your product please contact Time Electronics. We will issue
a return merchandise authorization (RMA) number that is to accompany the goods
returning.
Further instructions will also be issued prior to shipment. When returning
instruments, please ensure that they have been adequately packed, preferably
in the original packing supplied.
Time Electronics Ltd will not accept responsibility for units returned
damaged.
Please ensure that all units have details of the service required and all
relevant paperwork.
Send the instrument, shipping charges paid to:
Time Electronics Ltd
Unit 5, TON Business Park, 2-8 Morley Road,
Tonbridge, Kent, TN9 1RA.
United Kingdom.
Tel: +44(0)1732 355993
Fax: +44(0)1732 350198
Email: mail@timeelectronics.co.uk
Web Site: www.timeelectronics.com
Disposal of your old equipment
- When this crossed-out wheeled bin symbol is attached to a product it means the product is covered by the European Directive 2002/96/EC.
- All electrical and electronic products should be disposed of separately from the municipal waste stream via designated collection facilities appointed by the government or the local authorities.
- The correct disposal of your old appliance will help prevent potential negative consequences for the environment and human health.
- For more detailed information about disposal of your old appliance, please contact your city office, waste disposal service or return to Time Electronics.
Time Electronics Ltd
Unit 5, TON Business Park, 2-8 Morley Road,
Tonbridge, Kent, TN9 1RA, United Kingdom.
T: +44 (0) 1732 355993 | F: +44 (0) 1732 350198
mail@timeelectronics.co.uk |
www.timeelectronics.com
© 2021 Time Electronics Ltd.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>