Introduction

Product Overview

Description

The 7191 is a hand operated pressure pump designed to generate either vacuum to -0.95 bar (-14 psi) or pressures to 4 bar (60 psi). A high-quality screw press is designed for fine pressure adjustment, with adjusting resolution up to 0.0001 mbar (0.01 Pa). To reach the pressure stability as high as possible during the pressure calibration, an isothermal bellows chamber is designed for reducing the possible heat effects such as environmental temperature change.

The 7191 is designed without a non-returning valve meaning minimal maintenance is required. Two finger-tight connectors installed on the pump allow the user using fingers to connect and disconnect the reference gauge and devices under test quickly and easily without the need for PTFE tape or wrenches.

The 7191 is ideal for applications requiring accurate low pressure testing and calibration. It can be used in calibrating low pressure gauges, transmitters, or other pressure measuring instruments. To accompany the pump a wide range of digital pressure gauges are available. The TEG-CV4b covers the full range of the pump (-1 to 4 bar) with accuracies of either 0.2 %, 0.1 %, 0.05 %, or 0.02 % of full scale.

Specifications
  • Media: Air

  • Generated Pressure Range: 0.95 bar vacuum to 4 bar pressure (14 psi vacuum to 60 psi pressure)

  • Pressure Resolution: 0.0001 mbar / 0.01 Pa

  • Material: Ram/adapters: SST
    Body: SST/aluminum
    Seals: Buna-N

  • Connection: Finger-tight connectors on both output ports

  • Test Gauge connection: 1/4” BSP (1/4” NPT or other optional)

  • Reference Gauge Connection: 1/4” BSP (1/4” NPT or other optional)

  • Height: 145 mm

  • Base: 275 mm x 260 mm

  • Weight: 2.8 kg (6.2 lbs)

  • Optional Extras: Digital pressure gauges
    7191-CC Carry Case (with pump & 2 gauge storage)
    Pressure adaptor kits (7198 and 7198-NPT)
    Accessories

Controls

Product Controls

  1. 1/4” BSP port (1/4” NPT, M20 x 1.5 if specified)
  2. Release valve (clockwise = close / anti-clockwise = open)
  3. Fine adjust handle (turn clockwise to increase pressure)
  4. 1/4” BSP port (1/4” NPT, M20 x 1.5 if specified)
  5. Pressure/vacuum selector (pull out for pressure, push in for vacuum)
  6. Security bolt pin
  7. Isolation valve
  8. Pressure lever
  9. Coarse adjustment handle (turn clockwise to increase pressure)
Pressure Routine

Pressure Routine

Operation

  • A Connection
    Operating Instructions

  • B Open to atmosphere

  • C Pre-pressure

  • D Pressurizing process
    Operating Instructions

  • E Repeat pressurizing process
    Operating Instructions

  • F Decrease pressure process
    Operating Instructions

  • G Prepare for vacuum
    Operating Instructions

  • H Vacuum process
    Operating Instructions

  • I Repeat vacuum process
    Operating Instructions

  • J Decrease vacuum process
    Operating Instructions

Operating Precautions & Troubleshooting

Operating Precautions
  • Do not exceed the safety pressure limit
  • Compressed air may cause problems in explosive or hazardous environments
  • Close the valves and tighten plugs when 7191 is being transported
  • Do not shift the selector to vacuum mode whilst under pressure
  • Do not over tighten connectors
  • Store the pump in a dry and clean environment
Troubleshooting
Problem Causes Solution
Cannot pressurise device under test Hard to pump pressure lever

Check isolating valve is open Check airline for blockages
Pressure lever pumps but
pressure is not increasing| Check release valve is closedCheck pressure/vacuum selector is pulled out (for pressure). Check for leaks on device under testCheck for leaks on output connection
Difficult to use coarse adjust| The isolating valve is open| Close isolating valve
Sealing ring in broken or loose| Replace or tighten sealing ring
Difficult to use fine adjust| Output pressure is unstable when using fine adjust| Check the isolating valve is closed
Slow leak, output pressure dropping.| Check reference gauge and device under test is tightly screwed on and connection seal washer.Check release valve is closedCheck pressure/vacuum selector is engaged correctly.

Warranty and Servicing

Warranty

Time Electronics products carry a one-year manufacturer’s warranty as standard.

Time Electronics products are designed and manufactured to the highest standards and specifications to assure the quality and performance required by all sectors of industry. Time Electronics products are fully guaranteed against faulty materials and workmanship.

Should this product be found to be defective, please contact us using the below details. Inform us of the product type, serial number, and details of any fault and/or the service required. Please retain the supplier invoice as proof of purchase.

This warranty does not apply to defects resulting from action of the user such as misuse, operation outside of specification, improper maintenance or repair, or unauthorized modification. Time Electronics’ total liability is limited to repair or replacement of the product. Note that if Time Electronics determine that the fault on a returned product has been caused by the user, we will contact the customer before proceeding with any repair.

Calibration and Repair Services

Time Electronics offers repair and calibration services for all the products we make and sell. Routine maintenance by the manufacturer ensures optimal performance and condition of the product. Periodic traceable or accredited calibration is available.

Contacting Time Electronics

Online:
Please visit www.timeelectronics.com and select Technical Support from the Contact links. From this page you will be able to send information to the Time Electronics service team who will help and support you.

By phone: +44 (0) 1732 355993
By email: mail@timeelectronics.co.uk

Support

© 2021 Time Electronics Ltd.

All rights reserved.
Nothing from this manual may be multiplied, or made public in any form or manner, either electronically or hard copy, without prior written consent from Time Electronics Ltd. This also applies to any schematics, drawings and diagrams contained herein.

This manual provides operating and safety instructions for the Time Electronics product. To ensure correct operation and safety, please follow the instructions in this manual.
Time Electronics reserves the right to change the contents, specifications and other information contained in this manual without notice.

Time Electronics Ltd
Unit 5, TON Business Park, 2-8 Morley Road,
Tonbridge, Kent, TN9 1RA, United Kingdom.