ResMed 10 ST-A Air Curve User Guide
- September 26, 2024
- ResMed
Table of Contents
10 ST-A Air Curve
“`html
Product Specifications
-
Device Name: AirCurve 10 ST-A
-
Type: Bilevel positive airway pressure device
-
Indications for Use: Non-invasive ventilation for patients
weighing more than 30 lb (13 kg) with respiratory insufficiency or
obstructive sleep apnea (OSA) -
iVAPS Mode Indication: Patients weighing more than 66 lb (30
kg) -
Intended Use: Home and hospital use
-
Humidifier: Intended for single patient use in the home
environment and re-use in a hospital/institutional environment
Product Usage Instructions
Getting Started
1. Read the entire user guide before using the device.
2. Ensure the device is set up on a stable surface near a power
source.
Controls and Indicators
Air outlet: Where the air is released from the
device.
Air filter cover: Protects the air filter.
Retention clip: Secures tubing in place.
Power inlet: Connects the device to a power
source.
Screen: Displays various icons and
settings.
Start/Stop button: Press to start/stop
therapy.
Dial: Navigates menu options and adjusts
settings.
Home button: Returns to the Home screen.
Using the Device
1. Press the Start/Stop button to initiate therapy.
2. Adjust settings using the Dial as needed.
Cleaning and Maintenance
1. Regularly clean the tubing assembly, water tub, and mask for
optimal therapy and to prevent the growth of germs.
2. Follow the cleaning instructions provided in the user
guide.
Frequently Asked Questions (FAQ)
Q: Can the AirCurve 10 ST-A be used for pediatric
patients?
A: The AirCurve 10 ST-A is indicated for patients weighing more
than 30 lb (13 kg) with respiratory insufficiency or obstructive
sleep apnea (OSA). The iVAPS mode is indicated for patients
weighing more than 66 lb (30 kg).
Q: How often should I clean the device?
A: It is recommended to regularly clean the tubing assembly,
water tub, and mask for optimal therapy and to prevent the growth
of germs that can adversely affect the patient’s health.
“`
ResMed Confidential Proprietary Information, not to be reproduced or made
available to third parties without prior consent from ResMed and not to be
used in any unauthorised way. ResMed Ltd 2018.
SPECIFICATION PRINTED MATERIALS
DOC NO PAGE REV
288156 1 of 134 1
Rev Change Note
1 K001608-00
Date document
drafted
24 July 2018
Document prepared by (Name)
Melanie Theaker
Document Checked by (Name)
Lisa Oliver or delegate refer to change note.
AirCurve 10 ST-A User Guide AMER Multi
1. ORACLE DESCRIPTION AirCurve10 ST-A UG AMER MULTI
2. TRANSLATION DESCRIPTION Refer to the Change History on the English source listed for summary of changes in each version.
Rev
English source
This translation replaces…
1
288009r4 plus changes shown 288131
below
3. PRINT DETAILS Print according to master spec R370-7240
4. RESMED QC INSPECTION For general sampling and inspection requirements,
refer to AWI203-002 Appendix A, Printed or Unprinted Materials and Labels,
Criteria 1, 2, 3, 4, 5, 6, 8 and 9 Additional requirement a) None
5. FOR ALL OTHER REQUIREMENTS, REFER TO RESMED SUPPLIER MANUAL AQP119.
6. CHANGE HISTORY
Section Update
Tech Specs
Update of Wireless approval certificate id to add information for 4G FCC ID: 2ACHL-AIR104G IC ID: 9103A-AIR104G
Tech Specs
Update of Classification to be IEC 60601-1:2005+A1:2012
Tech Specs
Update of Wireless information to match Machine data for Tech comms rev 65
Cleaning Add warning `Regularly clean the tubing assembly, water tub and mask for optimal therapy and to prevent the growth of germs that can adversely the patient’s health.
User guide
User guide
288156 covers.indd 2
User guide
English | Français | Español | Português
24/07/2018 12:09:57 PM
288156 covers.indd 3
24/07/2018 12:09:57 PM
ENGLISH
Welcome
The AirCurveTM 10 ST-A is a bilevel positive airway pressure device.
WARNING
Read this entire guide before using the device.
CAUTION
In the US, Federal law restricts this device to sale by or on the order of a
physician.
Indications for use
AirCurve 10 ST-A The AirCurve 10 ST-A is indicated to provide non-invasive
ventilation for patients weighing more than 30 lb (13 kg) with respiratory
insufficiency or obstructive sleep apnea (OSA). The iVAPS mode is indicated
for patients weighing more than 66 lb (30 kg). The AirCurve 10 ST-A is
intended for home and hospital use. The humidifier is intended for single
patient use in the home environment and re-use in a hospital/institutional
environment.
Contraindications Positive airway pressure therapy may be contraindicated in
some patients with the following preexisting conditions: · severe bullous lung
disease · pneumothorax or pneumomediastinum · pathologically low blood
pressure, particularly if associated with intravascular volume depletion ·
dehydration · cerebrospinal fluid leak, recent cranial surgery, or trauma.
Adverse effects You should report unusual chest pain, severe headache, or
increased breathlessness to your prescribing physician. An acute upper
respiratory tract infection may require temporary discontinuation of
treatment. The following side effects may arise during the course of therapy
with the device: · drying of the nose, mouth, or throat · nosebleed · bloating
· ear or sinus discomfort · eye irritation · skin rashes.
English
1
At a glance
The AirCurve 10 includes the following: · Device with HumidAirTM integrated
humidifier · Water tub · Air tubing · Power supply unit · Travel bag · SD card
(already inserted). Contact your care provider for a range of accessories
available for use with the device including: · Air tubing (heated and non-
heated): ClimateLineAirTM, ClimateLineAir Oxy, SlimLineTM, Standard · Water
tub: Standard water tub, cleanable water tub (can be disinfected) · Side cover
for use without the humidifier · Filter: Hypoallergenic filter, standard
filter · Air10TM DC/DC converter (12V/24V) · SD card reader · Air10 oximeter
adapter · Air10 USB adapter · Power Station II · Air10 tubing elbow
2
About your device
1 Air outlet 2 Air filter cover 3 Retention clip 4 Power inlet 5 Serial number
and device number
About the control panel
Start/Stop button
Dial
6 HumidAir humidifier (water tub) 7 Screen 8 Adapter cover 9 SD card cover 10
LED alarm indicator
Press to start/stop therapy. Press and hold for three seconds to enter power
save mode.
Turn to navigate the menu and press to select an option. Turn to adjust a
selected option and press to save your change.
Home button
Press to return to the Home screen.
Different icons may be displayed on the screen at different times including:
Ramp Time
Wireless signal strength (green)
Humidity
Wireless transfer not enabled (gray)
Humidifier warming
No wireless connection
Humidifier cooling
Airplane Mode
Alarm muted
English
3
Setup
CAUTION
Do not overfill the water tub as water may enter the device and air tubing. 1.
With the device on a stable level surface, grip the retention clip on the back
of the device and
pull up to open. Note: The retention clip is shown in the open position. 2.
(a) Plug the power connector into the device power inlet then (b) push down
the retention clip to
secure in place. Connect one end of the power cord into the power supply unit
and the other end into the power outlet. 3. Connect the air tubing firmly to
the air outlet located on the rear of the device. 4. Open the water tub and
fill it with distilled water up to the maximum water level mark. Do not fill
the water tub with hot water. 5. Close the water tub and insert it into the
side of the device. 6. Connect the free end of the air tubing firmly onto the
assembled mask. See the mask user guide for detailed information. Recommended
masks are available on www.resmed.com. Note: Ensure the device is placed so
that the LED alarm indicator is clearly visible.
4
Starting therapy
1. Fit your mask. 2. Press Start/Stop or breathe normally if SmartStartTM is
enabled.
You will know that therapy is on when the Monitoring screen is displayed.
The pressure bar shows the inspiratory and expiratory pressures in green. The
green bar will expand and contract as you breathe in and out.
The screen will go black automatically after a short period of time. You can
press Home or the dial to turn it back on. If power is interrupted during
therapy, the device will automatically restart therapy when power is restored.
The AirCurve 10 device has a light sensor that adjusts the screen brightness
based on the light in the room.
Stopping therapy
1. Remove your mask. 2. Press Start/Stop or if SmartStart is enabled, therapy
will stop automatically after a few seconds. Note: If Confirm Stop is enabled,
a message is displayed asking if you want to stop therapy. Turn the dial to
select Yes and then press the dial to stop therapy. Once therapy has stopped,
the Sleep Report gives you a summary of your therapy session.
Usage hoursIndicates the number of hours of therapy you received last
session. Mask SealIndicates how well your mask sealed:
Good mask seal.
Needs adjusting, see Mask Fit.
HumidifierIndicates if your humidifier is working properly:
Humidifier working.
Humidifier might be faulty, contact your care provider. If set by your care
provider, you will also see: Events per hourIndicates the number of apneas
and hypopneas experienced per hour. More InfoTurn the dial to scroll down to
view more detailed usage data.
English
5
Power save mode
Your AirCurve 10 device records your therapy data. In order to allow it to
transmit the data to your care provider, you should not unplug the device.
However, you can put it into power save mode to save electricity. To enter
power save mode: · Press and hold Start/Stop for three seconds.
The screen goes black. To exit power save mode: · Press Start/Stop once.
The Home screen is displayed.
My Options
Your AirCurve 10 device has been set up for your needs by your care provider,
but you may find you want to make small adjustments to make your therapy more
comfortable.
Highlight My Options and press the dial to see your current settings. From
here, you can personalize your options.
Ramp Time
Designed to make the beginning of therapy more comfortable, Ramp Time is the
period during which the pressure increases from a low start pressure to the
prescribed treatment pressure. You can set your Ramp Time to Off or between 5
to 45 minutes.
To adjust Ramp Time: 1. In My Options, turn the dial to highlight Ramp
Time and then press the dial. 2. Turn the dial to adjust the ramp time to your
preferred setting and press the dial to save the change.
6
Ramp Down
Ramp Down is intended to make stopping therapy more comfortable by reducing
your pressure over a fixed 15 minute period. This option will only be
available to you via your care provider.
To enable Ramp Down: 1. In My Options, turn the dial to highlight Ramp Down
and then press the
dial. 2. Turn the dial to select On and then press the dial to save the
change.
To start Ramp Down: 1. Press the Start/Stop button.
Note: If Confirm Stop is enabled, a message is displayed asking if you want to start Ramp Down. Turn the dial to select Yes and then press the dial to start Ramp Down.
The Ramp Down icon
and time remaining will be displayed at the bottom left of the screen.
Once Ramp Down is complete, the device will continue to run at low pressure. To stop therapy at any time, press Start/Stop.
Humidity Level
The humidifier moistens the air and is designed to make therapy more
comfortable. If you are getting a dry nose or mouth, turn up the humidity. If
you are getting any moisture in your mask, turn down the humidity.
You can set the Humidity Level to Off or between 1 and 8, where 1 is the
lowest humidity setting and 8 is the highest humidity setting.
To adjust the Humidity Level:
1. In My Options, turn the dial to highlight Humidity Level and then press
the dial.
2. Turn the dial to adjust the humidity level and press the dial to save the
change.
If you continue to get a dry nose or mouth, or moisture in your mask, consider using ClimateLineAir heated air tubing. ClimateLineAir together with Climate Control delivers more comfortable therapy.
English
7
Mask Fit
Mask Fit is designed to help you assess and identify possible air leaks around
your mask.
To check Mask Fit: 1. Fit the mask as described in the mask user guide. 2. In
My Options, turn the dial to highlight Run Mask Fit and then press the
dial. The device starts blowing air. 3. Adjust the mask, mask cushion and
headgear until you get a Good result.
To stop Mask Fit, press the dial or Start/Stop. If you are unable to get a good mask seal, talk to your care provider.
More options
There are some more options on your device which you can personalize.
Mask
This option shows your mask type setting. If you use more than one type of mask, adjust this setting when switching between masks.
Run Warm Up
This option allows you to pre-heat the water before starting therapy, so that the air is not cold or dry at the beginning of therapy.
Ramp Down*
This option is intended to make stopping therapy more comfortable by reducing your pressure over a fixed 15 minute period.
SmartStart*
When SmartStart is enabled, therapy starts automatically when you breathe into your mask. When you remove your mask, it stops automatically after few seconds.
*When enabled by your care provider.
Working with alarms
The device is fitted with an alarm feature that monitors your therapy and
alerts you to changes that may affect your treatment.
When power is connected to the device, the yellow LED alarm indicator will
flash and the alarm will sound to confirm that the alarm is working.
When an alarm is activated, the yellow LED alarm indicator will flash, the
alarm will sound and a message will appear on the screen.
8
Muting activated alarms
To mute the alarm: 1. Press the dial. A list of activated alarms will appear
and the flashing alarm
mute icon will be displayed in the top right corner of the screen. The alarm
will be muted for 2 minutes. 2. To return to the previous screen, highlight OK
and press the dial.
Once the condition that activated the alarm is corrected, the alarm sound and
flashing icon will stop. If the condition that activated the alarm remains
after 2 minutes, the alarm will re-occur. See the Alarms troubleshooting
section for help with managing common alarm conditions.
Multiple alarms
If multiple alarms are activated at the same time, the most recent alarm
message will be displayed on the screen and any other activated alarms will be
shown in the Alarms list.
Viewing the alarms
To view the alarm list: 1. From the Monitoring screen, turn the dial clockwise
until the last
Monitoring screen is displayed. 2. To view the alarm details, highlight the
alarm and press the dial.
Caring for your device
It is important that you regularly clean your AirCurve 10 device to make sure
you receive optimal therapy. The following sections will help you with
disassembling, cleaning, checking and reassembling your device.
WARNING
Regularly clean the tubing assembly, water tub and mak for optimal therapy and
to prevent the growth of germs that can adversely affect the patient’s health.
English
9
Disassembling
1. Hold the water tub at the top and bottom, press it gently and pull it away
from the device. 2. Open the water tub and discard any remaining water. 3.
Hold the cuff of the air tubing and gently pull it away from the device.
Grip the retention clip and pull up to release the power cord. 4. Hold both
the cuff of the air tubing and the swivel of the mask, then gently pull apart.
Cleaning
You should clean the device weekly as described. Refer to the mask user guide
for detailed instructions on cleaning your mask. 1. Wash the water tub and air
tubing in warm water using mild detergent.
Do not wash in a dishwasher or washing machine. 2. Rinse the water tub and air
tubing thoroughly and allow to dry out of direct sunlight and/or heat. 3. Wipe
the exterior of the device with a dry cloth.
Checking
You should regularly check the water tub, air tubing and the air filter for
any damage. 1. Check the water tub:
· Replace it if it is leaking or has become cracked, cloudy or pitted. ·
Replace it if the seal is cracked or torn. · Remove any white powder deposits
using a solution of one part household vinegar to
10 parts water. 2. Check the air tubing and replace it if there are any holes,
tears or cracks. 3. Check the air filter and replace it at least every six
months. Replace it more often if there are any
holes or blockages by dirt or dust.
10
To replace the air filter:
1. Open the air filter cover and remove the old air filter. The air filter is
not washable or reusable.
2. Place a new air filter onto the air filter cover and then close it. Make
sure the air filter is fitted at all times to prevent water and dust from
entering the device.
Reassembling
When the water tub and air tubing are dry, you can reassemble the parts. 1.
Connect the air tubing firmly to the air outlet located on the rear of the
device. 2. Open the water tub and fill it with distilled room temperature
water up to the maximum water
level mark. 3. Close the water tub and insert it into the side of the device.
4. Connect the free end of the air tubing firmly onto the assembled mask.
Therapy data
Your AirCurve 10 device records your therapy data for you and your care
provider so they can view and make changes to your therapy if required. The
data is recorded and then transferred to your care provider wirelessly, if
wireless network is available, or via an SD card.
Data transmission
Your AirCurve 10 device has the capability of cellular communication so that
your therapy data can be transmitted to your care provider to improve the
quality of your treatment. It also allows your care provider to update your
therapy settings in a more timely manner or upgrade your device software to
ensure you receive the best therapy possible. The data is usually transmitted
after therapy has stopped. In order to make sure that your data is
transferred: · Leave your device connected to the power outlet at all times. ·
Make sure that it is not in Airplane Mode. · Check the Wireless signal
strength icon for adequate wireless coverage. Notes: · Therapy data might not
be transmitted if you use it outside of the country or region of purchase. ·
Devices with cellular communication might not be available in all regions.
Please be aware that within the wireless network, the availability and quality
of the network may be affected by terrain, buildings, and the weather.
Wireless communication depends on network availability. Coverage is not
available everywhere and varies by service.
English 11
SD card
An alternative way for your therapy data to be transferred to your care
provider is via the SD card. Your care provider may ask you to send the SD
card by mail or to bring it in. When instructed by your care provider, remove
the SD card. Do not remove the SD card from the device when the SD light is
flashing because data is being written to the card. To remove the SD card:
1. Open the SD card cover. 2. Push in the SD card to release it. Remove the
SD card from the device.
Place the SD card in the protective folder and send it back to your care
provider. For more information on the SD card refer to the SD card protective
folder provided with your device. Note: The SD card should not be used for any
other purpose.
Traveling
You can take your AirCurve 10 device with you wherever you go. Just keep the
following in mind: · Use the travel bag provided to prevent damage to the
device. · Empty the water tub and pack it separately in the travel bag. · Make
sure you have the appropriate power cord for the region you are traveling to.
For
information on purchasing, contact your care provider. · If you are using an
external battery, you should turn off the humidifier in order to maximize the
life of your battery. Do this by turning the Humidity Level to Off.
Traveling by plane
Your AirCurve 10 device may be taken on board as carry-on luggage. Medical
devices do not count toward your carry-on luggage limit. You can use your
AirCurve 10 device on a plane as it meets the Federal Aviation Administration
(FAA) requirements. Air travel compliance letters can be downloaded and
printed from www.resmed.com. When using the device on a plane: · Make sure the
water tub is completely empty and inserted into your device. The device will
not
work without the water tub inserted. · Turn on Airplane Mode.
12
To turn on Airplane Mode:
1. In My Options, turn the dial to highlight Airplane Mode and then press the
dial.
2. Turn the dial to select On and then press the dial to save the change. The
Airplane Mode icon is displayed at the top right of the screen.
CAUTION
Do not use the device with water in the water tub on a plane due to the risk
of inhalation of water during turbulence.
Troubleshooting
If you have any problems, have a look at the following troubleshooting topics.
If you are not able to fix the problem, contact your care provider or ResMed.
Do not try to open the device.
General troubleshooting
Problem/possible cause
Solution
Air is leaking from around my mask
Mask may be fitted incorrectly.
Make sure your mask is fitted correctly. See your mask user guide for fitting instructions or use the Mask Fit function to check your mask fit and seal.
I am getting a dry or blocked nose Humidity level may be set too low.
Adjust the Humidity Level.
If you have ClimateLineAir heated air tubing, see the ClimateLineAir user
guide.
I am getting droplets of water on my nose, in the mask and air tubing
Humidity level may be set too high.
Adjust the Humidity Level.
If you have ClimateLineAir heated air tubing, see the ClimateLineAir user guide.
My mouth is very dry and uncomfortable Air may be escaping through your mouth.
Increase the Humidity Level.
You may need a chin strap to keep your mouth closed or a full face mask.
Air pressure in my mask seems too high (it feels like I am getting too much air)
Ramp may be turned off.
Use the Ramp Time option.
English 13
Problem/possible cause
Solution
Air pressure in my mask seems too low (it feels like I am not getting enough air)
Ramp may be in progress
.
Wait for air pressure to build up or turn Ramp Time off.
Ramp Down may be in progress
.
Non-vented mask is used. Mask vents might be blocked.
Expiratory pressure (EPAP) may be set too low.
Press Start/Stop to stop therapy then press Start/Stop to restart and continue
therapy.
Only use a vented mask.
Check if you have sufficient venting. Unblock mask vents if necessary.
Talk to your care provider about your settings.
My screen is black
Backlight on the screen may have turned off. It turns off automatically after
a short period of time.
Power may not be connected.
Press Home or the dial to turn it back on.
Connect the power supply and make sure the plug is fully inserted. Note: The
retention clip should be in the open position when inserting the plug. For
instructions refer to the Setup section.
I have stopped therapy, but the device is still blowing air
Device is cooling down.
Device blows a small amount of air in order to avoid condensation in the air tubing. It will stop automatically after 30 minutes.
My water tub is leaking Water tub may not be assembled correctly. Water tub may be damaged or cracked.
Check for damage and reassemble the water tub correctly. Contact your care provider for a replacement.
My therapy data has not been sent to my care provider
Power may not be connected.
Connect the power supply and make sure the plug is fully inserted.
Note: The retention clip should be in the open position when inserting the plug. For instructions refer to the Setup section.
Wireless coverage may be poor.
Make sure that the device is placed where there is coverage (ie, on your
bedside table, not in a drawer or on the floor) and away from other
communication devices.
The Wireless signal strength icon indicates good coverage when all bars are
displayed, and poor coverage when fewer bars are displayed.
The No wireless connection icon is displayed on the top right of the screen. no wireless network available.
Make sure that the device is placed where there is coverage (ie, on your bedside table, not in a drawer or on the floor) and away from other communication devices. If instructed to do so, send the SD card to your care provider. The SD card also contains your therapy data.
14
Problem/possible cause Device may be in Airplane Mode. Data transfer is not enabled for your device.
Solution Turn off Airplane Mode, see Traveling by plane. Talk to your care provider about your settings.
My screen and buttons are flashing but there is no alarm sound or message
Software upgrade is in progress.
Software upgrade takes approximately 10 minutes to complete.
Displays message: Read only card, please remove, unlock and re-insert SD card
SD card switch may be in the lock (read-only) position.
Move the switch on the SD Card from the lock position to the unlock position and then re-insert it.
Alarms troubleshooting
Problem/possible cause
Display disappears and an alarm is activated Power failure.
Power cord is disconnected or mains power has been turned off during therapy.
Solution
Remove your mask until power is restored. Ensure the power cord is connected
and the mains power switch (if available) is on.
Displays message: High leak detected, check your water tub, tub seal or side cover
Water tub may not be inserted properly.
Make sure the water tub is correctly inserted.
Water tub seal may not be inserted properly.
Open the water tub and make sure that the seal is correctly inserted.
Displays message: High leak detected, connect your tubing
Air tubing may not be connected properly.
Make sure the air tubing is firmly connected at both ends.
Mask may be fitted incorrectly.
Make sure your mask is fitted correctly. See your mask user guide for fitting instructions or use the Mask Fit function to check your mask fit and seal.
Displays message: Tubing blocked, check your tubing
Air tubing may be blocked.
Check the air tubing and remove any blockages. Press the dial to clear the message and then press Start/Stop to restart the device.
Displays message: Leak detected, check your system setup and all connections
Mask may be fitted incorrectly.
Make sure your mask is fitted correctly. See your mask user guide for fitting instructions or use the Mask Fit function to check your mask fit and seal.
Displays message: Low MV detected
Minute ventilation level has dropped below the pre-set alarm level.
Contact your care provider.
English 15
Problem/possible cause
Solution
Displays message: Apnea detected
The device detects an apnea that has exceeded the pre-set Breath normally to disable the alarm.
alarm.
If the problem persists contact your care provider.
Displays message: Low Sp02 detected SpO2 has dropped below the pre-set alarm level.
Check the attachment of the sensor. If the problem persists contact your care provider.
Displays message: No SpO2 data, check your oxi sensor attachment to module/finger
Oximeter sensor is not attached properly.
Ensure that the oximeter sensor is attached properly to the module and to your finger.
Oximeter sensor may be faulty.
If the message appears repeatedly, the oximeter sensor might be faulty. Replace the oximeter.
Displays message: Non-vented mask detected, use vented mask or unblock your mask vents
Non-vented mask is used.
Only use a vented mask.
Mask vents might be blocked.
Check if you have sufficient venting. Unblock mask vents if necessary.
Expiratory pressure (EPAP) may be set too low.
Talk to your care provider about your settings.
Displays message: System fault, refer to user guide, Error 004
Device may have been left in a hot environment.
Allow to cool before re-use. Disconnect the power supply and then reconnect it to restart the device.
Air filter may be blocked.
Check the air filter and replace it if there are any blockages. Disconnect the power supply and then reconnect it to restart the device.
Air tubing may be blocked.
Check the air tubing and remove any blockages. Press the dial to clear the message and then press Start/Stop to restart the device.
There may be water in the air tubing.
Empty the water from the air tubing. Disconnect the power supply and then reconnect it to restart the device.
Displays message: System fault, refer to user guide, Error 022
Power cord may not be correctly inserted into the device.
Remove the power cord from the device and then re-insert it. Ensure that the power cord is fuly inserted into the device.
Note: the retention clip should be in the open position when inserting the plug. For instructions refer to the Setup section.
If the problem continues, contact your local ResMed dealer or ResMed office. Do not open the device.
All other error messages, for example, System fault, refer to user guide, Error 0XX
An unrecoverable error has occurred on the device.
Contact your care provider. Do not open the device.
16
Reassembling parts
Some parts of your device are designed to easily come off in order to avoid
damage to the parts or the device. You can easily reassemble them as described
below. To insert the water tub seal:
1. Place the seal into the lid. 2. Press down along all edges of the seal
until it is firmly in place. To reassemble the water tub lid:
1. Insert one side of the lid into the pivot hole of the base. 2. Slide the
other side down the ridge until it clicks into place.
English 17
General warnings and cautions
WARNING
· Make sure that you arrange the air tubing so that it will not twist around
the head or neck. · Make sure the power cord and plug are in good condition
and the equipment is not
damaged. · Keep the power cord away from hot surfaces. · If you notice any
unexplained changes in the performance of the device, if it is making
unusual sounds, if the device or the power supply are dropped or mishandled,
or if the enclosure is broken, discontinue use and contact your care provider
or your ResMed Service Center. · Do not open or modify the device. There are
no user serviceable parts inside. Repairs and servicing should only be
performed by an authorized ResMed service agent. · Beware of electrocution. Do
not immerse the device, power supply or power cord in water. If liquids are
spilled into or onto the device, unplug the device and let the parts dry.
Always unplug the device before cleaning and make sure that all parts are dry
before plugging it back in. · Supplemental oxygen must not be used while
smoking or in the presence of an open flame. · Always make sure that the
device is turned on and airflow generated before the oxygen supply is turned
on. Always turn the oxygen supply off before the device is turned off, so that
unused oxygen does not accumulate within the device enclosure and create a
risk of fire. · Do not perform any maintenance tasks while the device is in
operation. · The device should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the device should be
observed to verify normal operation in the configuration in which it will be
used. · The use of accessories other than those specified for the device is
not recommended. They may result in increased emissions or decreased immunity
of the device. · Regularly check the antibacterial filter for signs of
moisture or other contaminants, particularly during nebulization or
humidification. Failure to do so could result in increased breathing system
resistance. · The device has not been tested or certified for use in the
vicinity of X-ray, CT or MRI equipment. Do not bring the device within 13 ft
(4 m) of X-ray or CT equipment. Never bring the device into an MR environment.
· Therapy settings should not be changed remotely for patients in a hospital
setting. Remote changes in a hospital setting may not be appropriate for
certain patients, as these setting changes may not be communicated to all
hospital personnel treating the patient. Hospital staff should liaise with the
patient’s regular care provider such that the desired therapy outcome is
achieved.
CAUTION
· Use only ResMed parts and accessories with the device. Non-ResMed parts may
reduce the effectiveness of the treatment and/or damage the device.
· Use only vented masks recommended by ResMed or by the prescribing doctor
with this device. Fitting the mask without the device blowing air can result
in rebreathing of exhaled air. Make sure that the mask vent holes are kept
clear and unblocked to maintain the flow of the fresh air into the mask.
· Be careful not to place the device where it can be bumped or where someone
is likely to trip over the power cord.
18
· Blocking the air tubing and/or air inlet of the device while in operation
could lead to overheating of the device.
· Keep the area around the device dry, clean and clear of anything (eg,
clothes or bedding) that could block the air inlet or cover the power supply
unit.
· Do not place the device on its side as water might get into the device. ·
Incorrect system setup may result in incorrect mask pressure reading. Ensure
the system is
correctly set up. · Do not use bleach, chlorine, alcohol, or aromatic-based
solutions, moisturizing or
antibacterial soaps or scented oils to clean the device, the humidifier or air
tubing. These solutions may cause damage or affect the humidifier performance
and reduce the life of the products. · If you use the humidifier, always place
the device on a level surface lower than your head to prevent the mask and air
tubing from filling with water. · Leave the water tub to cool for ten minutes
before handling to allow the water to cool and to make sure that the water tub
is not too hot to touch. · Make sure that the water tub is empty before
transporting the device.
Note: The device is not intended to be operated by persons (including
children) with reduced physical, sensory or mental capabilities without
adequate supervision by a person responsible for the patient’s safety.
Technical specifications
90W power supply unit AC input range:
DC output: Typical power consumption: Peak power consumption:
100240V, 5060Hz 1.01.5A, Class II 115V, 400Hz 1.5A, Class II (nominal for aircraft use)
24V
3.75A
53W (57VA)
104W (108VA)
Environmental conditions Operating temperature:
Operating humidity: Operating altitude: Storage and transport temperature:
Storage and transport humidity:
+41°F to +95°F (+5°C to +35°C)
Note: The air flow for breathing produced by this therapy device can be higher
than the temperature of the room. Under extreme ambient temperature conditions
(104°F/40°C) the device remains safe.
10 to 95% relative humidity, non-condensing
Sea level to 8,500′ (2,591 m); air pressure range 1013 hPa to 738 hPa
-4°F to +140°F (-20°C to +60°C)
5 to 95% relative humidity, non-condensing
Electromagnetic compatibility
The AirCurve 10 complies with all applicable electromagnetic compatibility
requirements (EMC) according to IEC 60601-12:2014, for residential, commercial
and light industry environments. It is recommended that mobile communication
devices are kept at least 1 m away from the device.
For further details see the Guidance and manufacturer’s declaration
electromagnetic emissions and immunity section.
Classification: IEC 60601-1:2005+A1:2012 Class II (double insulation), Type BF, Ingress protection IP22.
English 19
Sensors Pressure sensors:
Flow sensor:
Internally located at device outlet, analog gauge pressure type, -5 to +45 cm
H2O (-5 to +45 hPa)
Internally located at device inlet, digital mass flow type, -70 to +180 L/min
Maximum single fault steady pressure Device will shut down in the presence of a single fault if the steady state pressure exceeds: 30 cm H2O (30 hPa) for more than 6 sec or 40 cm H2O (40 hPa) for more than 1 sec.
Sound
Pressure level measured according to ISO 80601-2-70:2015 (CPAP mode):
SlimLine:
25 dBA with uncertainty of 2 dBA
Standard:
25 dBA with uncertainty of 2 dBA
SlimLine or Standard and humidification:
27 dBA with uncertainty of 2 dBA
Power level measured according to ISO 80601-2-70:2015 (CPAP mode):
SlimLine:
33 dBA with uncertainty of 2 dBA
Standard:
33 dBA with uncertainty of 2 dBA
SlimLine or Standard and humidification:
35 dBA with uncertainty of 2 dBA
Declared dual-number noise emission values in accordance with ISO 4871:1996.
Alarm volume settings Low (nominal 54 dBA), Medium (nominal 60 dBA), High (nominal 73 dBA)
Physical – device and water tub Dimensions (W x D x H):
Air outlet (complies with ISO 5356-1:2004): Weight (device and standard water
tub): Weight (device and cleanable water tub): Housing construction: Water
capacity: Standard water tub – material: Cleanable water tub – material:
10.04″ x 5.91″ x 4.57″ (255 mm x 150 mm x 116 mm) 22 mm 47.1 oz (1336 g) 47.1 oz (1336 g) Flame retardant engineering thermoplastic To maximum fill line 380 mL Injection molded plastic, stainless steel and silicone seal Injection molded plastic, stainless steel and silicone seal
Temperature Maximum heater plate: Cut-out: Maximum gas temperature:
154°F (68°C) 165°F (74°C) 106°F ( 41°C)
Air filter Standard:
Hypoallergenic:
Material: Polyester non woven fiber Average arrestance: >75% for ~7 micron
dust
Material: Acrylic and polypropylene fibers in a polypropylene carrier
Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron dust
Aircraft use
ResMed confirms that device meets the Federal Aviation Administration (FAA)
requirements (RTCA/DO-160, section 21, category M) for all phases of air
travel.
20
Wireless module Technology used:
Frequencies:
Max RF power output:
CDMA (USA only), 4G (USA and Canada) 3G, 2G
CDMA (800/1900 MHz) 4G (700/850/1900 Mhz) 3G (850/900/1700/1900/2100 MHz) 2G
(850/900/1800/1900 MHz)
CDMA 24.5 dBm 4G 23.0 dBm 3G 24.0 dBm 2G 33.0 dBm (850/900MHz) 30,0 dBm
(1800/1900 MHz)
FCC ID: 2ACHL-A10STACD, 2ACHL-A10STA3G, 2ACHL-AIR104G IC: 9103A-A10STA3G, 9103A-AIR104G The AirCurve 10 device complies with FCC Rules and Industry Canada Rules. The AirCurve 10 device should be used at a minimum distance of 0.8″ (2 cm) from the body during operation. Additional information regarding the FCC Rules and IC compliance for this device can be found on www.resmed.com/downloads/devices
Operating pressure range S, ST, T, PAC, iVAPS: CPAP
3 to 30 cm H2O 4 to 20 cm H2O
Supplemental oxygen Maximum flow:
15 L/min (S, ST, T, PAC, CPAP), 4 L/min (iVAPS)
Pneumatic flow path
1. Flow sensor 2. Blower 3. Pressure sensor 4. Mask 5. Air tubing 6. Water tub 7. Device 8. Inlet filter
Design life Device, power supply unit: Cleanable water tub: Standard water tub, air tubing:
5 years 2.5 years 6 months
General The patient is an intended operator.
Operator position
The device is designed to be operated within arm’s length. An operator should
position their line of sight within an angle of 30 degrees from a plane
perpendicular to the screen.
English 21
Humidifier performance
Mask Pressure cm H2O (hPa)
3 4 10 20 25 30
Nominal RH output %
Setting 4
Setting 8
85
100
85
100
85
100
85
90
85
90
85
90
1 AH – Absolute Humidity in mg/L 2 BTPS – Body Temperature Pressure Saturated
Nominal system output AH1, BTPS2
Setting 4
Setting 8
6
10
6
10
6
10
6
10
6
10
6
10
Air tubing
Air tubing
Material
ClimateLineAir
Flexible plastic and electrical components
ClimateLineAir Oxy Flexible plastic and electrical components
SlimLine
Flexible plastic
Standard
Flexible plastic
3 m
Flexible plastic
Heated air tubing temperature cut-out: 106°F ( 41°C)
Length 6’6″ (2 m) 6’4″ (1.9 m) 6′ (1.8 m) 6’6″ (2 m) 9’10” (3 m)
Inner diameter 0.6″ (15 mm) 0.75″ (19 mm) 0.6″ (15 mm) 0.75″ (19 mm) 0.75″ (19 mm)
Notes: · The manufacturer reserves the right to change these specifications
without notice. · The electrical connector end of the heated air tubing is
only compatible with the air outlet at the device end and should
not be fitted to the mask. · Do not use electrically conductive or antistatic
air tubing. · The temperature and relative humidity settings displayed are not
measured values.
Displayed values
Value Pressure sensor at air outlet: Mask pressure Flow derived values: Leak Tidal volume Respiratory rate Minute ventilation Ti I:E ratio
Range
330 cm H2O
0120 L/min 04000 mL 050 BPM 030 L/min 0.14.0 sec 1:1002:1
Display resolution
0.1 cm H2O
1 L/min 1 mL 1 BPM 0.1 L/min 0.1 sec 0.1
22
Value
Accuracy1
Pressure measurement1:
Mask pressure2 Flow and flow derived values1:
±[0.5 cm H2O (0.5 hPa) + 4% of measured value]
Flow
±6 L/min or 10% of reading, whichever is greater, at 0 to 150 L/min positive flow
Leak2
±12 L/min or 20% of reading, whichever is greater, 0 to 60 L/min
Tidal volume2,3
±20%
Respiratory rate2,3
±1.0 BPM
Minute ventilation2,3
±20%
1 Results are expressed as STPD (Standard Temperature and Pressure, Dry). 101.3kPa at an operating temperature of 68°F (20°C), dry.
2 Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL or minute ventilation <3 L/min.
3 Measurement accuracy verified as per ISO 10651-6:2004 for Home Care Ventilatory Support Devices (Figure 101 and Table 101) using
nominal ResMed mask vent flows.
Measurement system uncertainties
In accordance with ISO 80601-2-70:2015 the measurement uncertainty of the manufacturer’s test equipment is:
For measures of flow
± 1.5 L/min or ± 2.7% of reading (whichever is greater)
For measures of volume (< 100 mL)
± 5 mL or 6% of reading (whichever is greater)
For measures of volume ( 100 mL) For measures of pressure For measures of time
± 20 mL or 3% of reading (whichever is greater) ± 0.15 cm H2O (0.15 hPa) ± 10 ms
Pressure accuracy – CPAP
Maximum static pressure variation at 10 cm H2O (10 hPa) according to ISO 80601-2-70:2015
Without humidification With humidification
Standard air tubing ± 0.5 cm H2O (± 0.5 hPa) ± 0.5 cm H2O (± 0.5 hPa)
SlimLine air tubing ± 0.5 cm H2O (± 0.5 hPa) ± 0.5 cm H2O (± 0.5 hPa)
Maximum dynamic pressure variation according to ISO 80601-2-70:2015
Device without humidification and Standard air tubing / Device with humidification and Standard air tubing
Pressure [cm H2O (hPa)] 10 BPM
15 BPM
20 BPM
4
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
8
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
12
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
16
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
20
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
Device without humidification and SlimLine air tubing / Device with humidification and SlimLine air tubing
Pressure [cm H2O (hPa)] 10 BPM
15 BPM
20 BPM
4
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
8
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
12
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
16
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
20
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
English 23
Pressure accuracy – bilevel
Maximum dynamic pressure variation according to ISO 80601-2-70:2015. Device without humidification and Standard air tubing / Device with humidification and Standard air tubing Breath rate Inspiratory pressure (cm H2O [hPa]) (Means, Standard Deviations)
10 BPM 15 BPM 20 BPM Breath rate
6
10
16
21
25
-0.09, 0.01 / – -0.01, 0.07 / – 0.07, 0.05 / – -0.03, 0.09 / – 0.12, 0.01 / –
0.22, 0.01
0.22, 0.01
0.24, 0.01
0.29, 0.03
0.26, 0.02
0.02, 0.08 / – 0.12, 0.01 / – 0.15, 0.01 / – 0.15, 0.01 / – 0.16, 0.12 / –
0.22, 0.01
0.22, 0.01
0.26, 0.01
0.31, 0.02
0.30, 0.02
0.17, 0.01 / – 0.21, 0.01 / – 0.25, 0.01 / – 0.21, 0.17 / – 0.32, 0.02 / –
0.23, 0.01
0.28, 0.01
0.34, 0.01
0.38, 0.02
0.40, 0.03
Expiratory pressure (cm H2O [hPa]) (Means, Standard Deviations)
30
0.12, 0.01 / 0.14, 0.02
0.20, 0.05 / 0.22, 0.02
0.34, 0.02 / 0.34, 0.03
2
6
12
17
21
25
10 BPM
-0.14, 0.01 / – -0.16, 0.01 / – -0.11, 0.10 / – -0.16, 0.05 / – -0.17, 0.05 / – 0.04, 0.17 / –
0.27, 0.01
0.29, 0.02
0.34, 0.02
0.33, 0.01
0.33, 0.02
0.21, 0.01
15 BPM
-0.16, 0.01 / – -0.20, 0.01 / – -0.20, 0.05 / – -0.21, 0.05 / – -0.23, 0.08 / – 0.04, 0.21 / –
0.25, 0.01
0.33, 0.02
0.35, 0.01
0.38, 0.02
0.38, 0.02
0.25, 0.01
20 BPM
-0.27, 0.01 / – -0.26, 0.02 / – -0.25, 0.01 / – -0.29, 0.01 / – -0.31, 0.01 / – -0.13, 0.23 / –
0.37, 0.01
0.34, 0.01
0.38, 0.01
0.43, 0.02
0.45, 0.03
0.31, 0.01
Device without humidification and SlimLine air tubing / Device with humidification and SlimLine air tubing
Breath rate Inspiratory pressure (cm H2O [hPa]) (Means, Standard Deviations)
10 BPM 15 BPM 20 BPM Breath rate
6
10
16
21
25
-0.26, 0.01 / – -0.25, 0.02 / – -0.24, 0.02 / – -0.25, 0.02 / – -0.20, 0.02 / –
0.52, 0.01
0.53, 0.02
0.53, 0.01
0.54, 0.02
0.51, 0.02
-0.26, 0.01 / – -0.25, 0.01 / – -0.26, 0.01 / – -0.31, 0.03 / – -0.30, 0.05 / –
0.51, 0.01
0.54, 0.01
0.56, 0.01
0.58, 0.02
0.60, 0.03
-0.25, 0.02 / – -0.29, 0.02 / – -0.34, 0.02 / – -0.36, 0.02 / – -0.36, 0.03 / –
0.52, 0.01
0.58, 0.01
0.62, 0.01
0.67, 0.02
0.69, 0.02
Expiratory pressure (cm H2O [hPa]) (Means, Standard Deviations)
30
-0.07, 0.09 / 0.18, 0.02
0.18, 0.08 / 0.25, 0.02
0.36, 0.02 / 0.40, 0.02
10 BPM 15 BPM 20 BPM
2
-0.28, 0.01 / 0.43, 0.01
-0.24, 0.02 / 0.37, 0.01
0.05, 0.21 / 0.38, 0.01
6
-0.30, 0.03 / 0.50, 0.01
-0.29, 0.02 / 0.47, 0.01
-0.31, 0.02 / 0.50, 0.02
12
-0.30, 0.01 / 0.54, 0.01
-0.35, 0.01 / 0.55, 0.01
-0.37, 0.02 / 0.57, 0.02
17
-0.33, 0.01 / 0.58, 0.01
-0.38, 0.01 / 0.62, 0.02
-0.43, 0.02 / 0.65, 0.02
21
-0.34, 0.01 / 0.60, 0.02
-0.42, 0.02 / 0.66, 0.01
-0.48, 0.02 / 0.68, 0.02
25
-0.27, 0.01 / 0.30, 0.01
-0.33, 0.01 / 0.36, 0.01
-0.43, 0.02 / 0.45, 0.01
Note: The table above is based on data that covers between 60.1 and 88.8% of the inspiratory phase and 66.1 and 93.4% of the expiratory phase durations. These data time slots start immediately after the initial transient overshoot/undershoot periods and end at the point that flow diminishes to an equivalent absolute value of its starting point, towards the end of the breath phases (this corresponds to the % ranges of values given immediately above).
24
Flow (maximum) at set pressures
The following are measured accordingly to ISO 80601-2-70:2015 at the end of the specified air tubing:
Pressure cm H2O (hPa)
AirCurve 10 and Standard L/min
AirCurve 10, humidification and Standard L/min
AirCurve 10 and SlimLine L/min
AirCurve 10, humidification and ClimateLineAir L/min
4
180
143
162
151
8
168
135
151
142
12
157
136
140
135
16
144
134
128
121
20
131
123
117
109
25
120
115
96
84
Guidance and manufacturer’s declaration electromagnetic emissions and immunity
Medical electrical equipment needs special precautions regarding EMC and needs
to be installed and put into service according to EMC information provided in
this document.
The AirCurve 10 device has been designed to meet EMC standards. However,
should you suspect that the device performance (eg, pressure or flow) is
affected by other equipment, move the device away from the possible cause of
interference.
Guidance and manufacturer’s declaration–electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that the device is used in such an environment.
Emissions test
Compliance Electromagnetic environment — guidance
RF emissions CISPR 11
Group 1
The device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The device is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage network that supplies buildings used for domestic purposes.
Harmonic Emissions IEC 61000-3-2
Class A
Voltage Fluctuations/Flicker Emissions IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that the device is used in such an environment.
Immunity test
Electrostatic discharge (ESD) IEC 61000-4-2 Electrical fast transient/burst
IEC 61000-4-4
IEC60601-1-2 test level
±6 kV contact ±8 kV air
Compliance level
±8 kV contact ±15 kV air
±2 kV for power
±2 kV
supply lines
±1 kV for input/output ±1 kV for input/output
lines
lines
Electromagnetic environment — guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30%. Mains power
quality should be that of a typical commercial or hospital environment.
English 25
Immunity test
Surge IEC 61000-4-5
Voltage dips, short interruptions and voltage variations on power supply input
lines IEC 61000-4-11
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
IEC60601-1-2 test level
±1 kV differential mode ±2 kV common mode
<5% Ut (>95% dip in Ut) for 0.5 cycle
40% Ut (60% dip in Ut) for 5 cycles
70% Ut (30% dip in Ut) for 25 cycles
<5% Ut (>95% dip in Ut) for 5 sec
3 A/m
Compliance level ±1 kV differential mode ±2 kV common mode 100V 240V
30 A/m
Electromagnetic environment — guidance
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the device requires continued operation during
power mains interruptions, it is recommended that the device be powered from
an uninterruptible power source.
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
Conducted RF 3 Vrms
3 Vrms
IEC 61000-4-6 150 kHz to 80 MHz 150 kHz to 80 MHz
Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Radiated RF 3 V/m IEC 61000-4-3 80 MHz to 2.5 GHz
10 V/m 80 MHz to 2.5 GHz
Recommended separation distance d = 0.35 P
d = 0.35 P 80 MHz to 800 MHz
d = 0.70 P 800 MHz to 2.5 GHz
Where (P) is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. b Interference may occur in the vicinity of equipment marked with the following symbol:
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the device.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. Notes:
· Ut is the AC mains voltage prior to application of the test level.
· At 80 MHz and 800 MHz, the higher frequency range applies.
· These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
26
Recommended separation distances between portable and mobile RF communications
equipment and the device
The device is intended for use in an environment in which radiated RF
disturbances are controlled. The customer or the user of the device can help
prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the device
as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output Separation distance according to frequency of transmitter (m)
power of transmitter (W)
150 kHz to 80 MHz d = 0.35 P
80 MHz to 800 MHz d = 0.35 P
800 MHz to 2.5 GHz d = 0.7 P
0.01
0.035
0.1
0.11
1
0.35
10
1.1
100
3.5
0.035 0.11 0.35 1.1 3.5
0.070 0.22 0.70 2.2 7.0
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be determined using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Notes:
· At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
· These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
Symbols
The following symbols may appear on the product or packaging.
Read instructions before use. Indicates a warning or caution. Follow instructions before
use. Manufacturer.
European Authorized Representative.
Batch code.
Catalog number.
Serial number.
Device number. On / Off. Device weight.
Protected against finger sized objects and against dripping water when tilted up to 15 degrees
from specified orientation.
Direct current. Type BF applied part. Class II
equipment. Humidity limitation. Temperature limitation. Non-ionising radiation.
China pollution control logo 1.
China pollution control logo 2.
Prescription only
(In the US, Federal law restricts these devices to sale by or on the order of a physician).
Maximum water level.
Use distilled water only. Operating altitude.
Atmospheric pressure limitation. Complies with RTCA DO-160 section 21, category M.
MR unsafe (do not use in the vicinity of an MRI device).
English 27
Environmental information
This device should be disposed of separately, not as unsorted municipal waste.
To dispose of your device, you should use appropriate collection, reuse and
recycling systems available in your region. The use of these collection, reuse
and recycling systems is designed to reduce pressure on natural resources and
prevent hazardous substances from damaging the environment.
If you need information on these disposal systems, please contact your local
waste administration. The crossed-bin symbol invites you to use these disposal
systems. If you require information on collection and disposal of your ResMed
device please contact your ResMed office, local distributor or go to
www.resmed.com/environment.
Servicing
The AirCurve 10 device is intended to provide safe and reliable operation when
operated in accordance with the instructions provided by ResMed. ResMed
recommends that the AirCurve 10 device be inspected and serviced by an
authorized ResMed Service Centre if there is any sign of wear or concern with
device function. Otherwise, service and inspection of the products generally
should not be required during their design life.
Limited warranty
ResMed Ltd (hereafter ‘ResMed’) warrants that your ResMed product shall be
free from defects in material and workmanship from the date of purchase for
the period specified below.
Product
Warranty period
· Mask systems (including mask frame, cushion, headgear and tubing)–excluding 90 days single-use devices
· Accessories–excluding single-use devices
· Flex-type finger pulse sensors
· Humidifier water tubs
· Batteries for use in ResMed internal and external battery systems
6 months
· Clip-type finger pulse sensors
1 year
· CPAP and bilevel device data modules
· Oximeters and CPAP and bilevel device oximeter adapters
· Humidifier cleanable water tubs
· Titration control devices
· CPAP, bilevel and ventilation devices (including external power supply units)
2 years
· Humidifiers
· Battery accessories
· Portable diagnostic/screening devices
This warranty is only available to the initial consumer. It is not transferable.
If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components.
This Limited Warranty does not cover: a) any damage caused as a result of improper use, abuse, modification or alteration of the product; b) repairs carried out by any service organization that has not been expressly authorized by ResMed to perform such repairs; and c) any damage or contamination due to cigarette, pipe, cigar or other smoke.
Warranty is void on product sold, or resold, outside the region of original
purchase.
28
Warranty claims on defective product must be made by the initial consumer at
the point of purchase. This warranty replaces all other expressed or implied
warranties, including any implied warranty of merchantability or fitness for a
particular purpose. Some regions or states do not allow limitations on how
long an implied warranty lasts, so the above limitation may not apply to you.
ResMed shall not be responsible for any incidental or consequential damages
claimed to have resulted from the sale, installation or use of any ResMed
product. Some regions or states do not allow the exclusion or limitation of
incidental or consequential damages, so the above limitation may not apply to
you. This warranty gives you specific legal rights, and you may also have
other rights which vary from region to region. For further information on your
warranty rights, contact your local ResMed dealer or ResMed office.
Further information
If you have any questions or require additional information on how to use the
device, contact your care provider.
English 29
FRANÇAIS
Bienvenue
L’appareil AirCurveTM 10 ST-A est un appareil de ventilation à deux niveaux de
pression positive.
AVERTISSEMENT
Veuillez lire le présent guide entièrement avant d’utiliser cet appareil.
ATTENTION
Selon la loi fédérale américaine, cet appareil ne peut être vendu aux USA que
par un médecin ou sur prescription médicale.
Indications d’utilisation
AirCurve 10 ST-A L’appareil AirCurve 10 ST-A est indiqué pour la ventilation
non invasive des patients d’un poids de plus de 13 kg (30 lb) atteints
d’insuffisance respiratoire ou du syndrome d’apnées obstructives du sommeil
(SAOS). Le mode iVAPS est indiqué pour les patients d’un poids de plus de 30
kg (66 lb). L’appareil AirCurve 10 ST-A est destiné à un usage à domicile ou
en milieu hospitalier.
L’humidificateur est conçu pour un usage unique à domicile et pour un usage
multiple dans un hôpital ou dans un autre établissement.
Contre-indications Le traitement par pression positive est contre-indiqué chez
certains patients présentant l’une des pathologies préexistantes suivantes : ·
affection pulmonaire bulleuse grave; · pneumothorax ou pneumomédiastin ·
tension artérielle pathologiquement basse, en particulier si associée à une
déplétion du volume
intravasculaire · déshydratation · fuite de liquide céphalo-rachidien, récent
traumatisme ou intervention chirurgicale crânienne.
Effets secondaires Les patients doivent avertir leur médecin traitant en cas
de douleurs thoraciques inhabituelles, de maux de tête sévères ou d’une
dyspnée accrue. Une infection aiguë des voies respiratoires supérieures peut
nécessiter l’arrêt temporaire du traitement.
Les effets secondaires suivants peuvent apparaître au cours du traitement sous
VPAP : · sécheresse nasale, buccale ou de la gorge; · saignements de nez; ·
ballonnements; · gêne au niveau de l’oreille ou des sinus; · irritation des
yeux; · érythèmes cutanés.
Français
1
Description rapide
L’appareil AirCurve 10 comprend les accessoires suivants : · Appareil avec
humidificateur HumidAirTM intégré · Réservoir d’eau · Circuit respiratoire ·
Module d’alimentation · Sac de transport · Carte SD (déjà insérée) Communiquer
avec le fournisseur de soins pour connaître la gamme d’accessoires pouvant
être utilisés avec l’appareil, notamment : · Circuit respiratoire (chauffant
et non chauffant) : ClimateLineAirTM, ClimateLineAir Oxy, SlimLineTM,
standard · Réservoir d’eau : Réservoir d’eau standard, réservoir d’eau
nettoyable (peut être désinfecté) · Couvercle latéral pour une utilisation
sans humidificateur · Filtre : Filtre hypoallergénique, filtre standard ·
Convertisseur CC/CC Air10TM (12 V/24 V) · Lecteur de carte SD · Adaptateur
d’oxymètre Air10 · Adaptateur USB Air10 · Power Station II · Circuit coudé
Air10
2
Présentation de l’appareil
1 Sortie d’air 2 Couvercle du logement du filtre à air 3 Attache de retenue 4 Entrée d’alimentation électrique 5 Numéro de série et numéro de l’appareil
6 Humidificateur HumidAir (réservoir d’eau) 7 Écran 8 Couvercle de l’adaptateur 9 Panneau de la carte SD 10 Voyant d’alarme DEL
Présentation du panneau de contrôle
Touche On/Off
Appuyer sur cette touche pour démarrer/arrêter le
traitement. Appuyer sur cette touche et la maintenir enfoncée
pendant trois secondes pour activer le mode d’économie d’énergie.
Molette de réglage
Tourner la molette pour naviguer dans le menu et appuyer dessus pour sélectionner une option. Tourner la molette pour régler l’option sélectionnée et appuyer dessus pour enregistrer la modification.
Touche accueil
Appuyer sur cette touche pour revenir à l’écran d’accueil.
Différentes icônes peuvent s’afficher à l’écran à différents moments, notamment :
Durée de rampe
Puissance du signal sans fil (vert)
Humidité
Préchauffage de l’humidificateur Refroidissement de l’humidificateur
Alarme désactivée
Transfert sans fil non activé (gris) Aucune connexion sans fil Mode avion
Français
3
Installation
ATTENTION
Ne pas trop remplir le réservoir d’eau sans quoi de l’eau risque de s’écouler
dans l’appareil et dans le circuit respiratoire. 1. Une fois l’appareil posé
sur une surface plane et stable, saisir l’attache de retenue située à
l’arrière
de l’appareil et la tirer vers le haut pour ouvrir. Remarque : L’attache de
retenue est présentée en position ouverte. 2. (a) Brancher le connecteur
d’alimentation dans l’entrée d’alimentation électrique de l’appareil et (b)
pousser l’attache de retenue vers le bas pour bien la fixer en place. Brancher
une extrémité du cordon d’alimentation dans le module d’alimentation et
l’autre extrémité dans une prise de courant. 3. Raccorder fermement le circuit
respiratoire à la sortie d’air située à l’arrière de l’appareil. 4. Ouvrir le
réservoir d’eau et le remplir d’eau distillée jusqu’au repère de niveau d’eau
maximal. Ne pas remplir le réservoir d’eau chaude. 5. Fermer le réservoir
d’eau et l’insérer sur le côté de l’appareil. 6. Raccorder fermement
l’extrémité libre du circuit respiratoire au système de masque assemblé.
Consulter le guide de l’utilisateur du masque pour obtenir des renseignements
détaillés.
4
Une liste des masques recommandés se trouve à l’adresse www.resmed.com.
Remarque : S’assurer que l’appareil est placé de manière à ce que l’indicateur
d’alarme DEL soit bien visible.
Démarrage du traitement
1. Mettre le masque. 2. Appuyer sur la touche On/Off ou respirer normalement
si la fonction SmartStart est activée.
L’affichage de l’écran Monitorage indique que le traitement est en marche.
La barre de pression affiche les pressions inspiratoires et expiratoires en
vert. La barre verte s’agrandira et se rétrécira lorsque vous inspirerez et
expirerez.
L’écran s’éteindra automatiquement après un bref délai. Pour le rallumer,
appuyer sur Accueil ou sur la molette de réglage. En cas de panne de courant
au cours du traitement, l’appareil redémarrera automatiquement lorsque le
courant reviendra.
L’appareil AirCurve 10 comporte un capteur de luminosité qui ajuste la
luminosité de l’écran en fonction de la lumière ambiante.
Arrêt du traitement
1. Retirer le masque. 2. Appuyer sur la touche On/Off ou, si la fonction
SmartStart est activée, le traitement s’arrêtera
automatiquement après quelques secondes.
Remarque : Si la fonction de confirmation d’arrêt est activée, un message
s’affiche pour demander si l’arrêt du traitement est souhaité. Tourner la
molette de réglage pour sélectionner Oui, puis appuyer sur la molette pour
arrêter le traitement.
Après l’arrêt du traitement, le Rap. somm. présente alors un résumé de la
séance de traitement.
Heures d’utilisationIndique le nombre d’heures de traitement reçues au cours
de la dernière séance. Étanchéité masqueIndique le degré d’étanchéité du
masque :
Bonne étanchéité masque.
Un ajustement est nécessaire, consulter la section Ajustement masque.
HumidificateurIndique si l’humidificateur fonctionne correctement :
L’humidificateur fonctionne.
L’humidificateur est peut-être défectueux, communiquer avec le fournisseur de soins.
Français
5
S’ils ont été réglés par le fournisseur de soins, les paramètres suivants
seront également affichés : Événements par heureIndique le nombre d’apnées et
d’hypopnées qui sont survenues par heure. Infos détailléesTourner la molette
de réglage et faire défiler pour voir des données détaillées sur
l’utilisation.
Mode d’économie d’énergie
L’appareil AirCurve 10 enregistre les données de traitement. Afin de lui
permettre de transmettre ces données au fournisseur de soins, il ne faut pas
débrancher l’appareil. Cependant, on peut le mettre en mode d’économie
d’énergie pour économiser de l’électricité. Pour activer le mode d’économie
d’énergie : · Appuyer sur la touche On/Off et la maintenir enfoncée pendant
trois secondes.
L’écran s’éteint. Pour quitter le mode d’économie d’énergie : · Appuyer une
fois sur la touche On/Off.
L’écran Accueil s’affiche.
Mes options
L’appareil AirCurve 10 a été réglé par le fournisseur de soins pour convenir
aux besoins de l’utilisateur; il se pourrait cependant que l’utilisateur
veuille apporter de légers ajustements afin de rendre le traitement plus
confortable.
Mettre en surbrillance Mes options et appuyer sur la molette de réglage pour
voir les réglages actuels. Dans cet écran, il est possible de personnaliser
les options.
Durée de rampe
Conçue pour rendre le début du traitement plus confortable, la durée de rampe
est la période durant laquelle la pression passe d’une faible pression de
départ à la pression de traitement prescrite. Vous pouvez régler votre durée
de rampe (Dur. rampe) à Off ou entre 5 et 45 minutes.
Pour régler la durée de rampe : 1. Dans Mes options, tourner la molette de
réglage
pour mettre en surbrillance Dur. rampe, puis appuyer sur la molette. 2.
Tourner la molette de réglage pour ajuster la durée de rampe au réglage
souhaité et appuyer sur la molette pour enregistrer la modification.
6
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>