BRASCH TGK-NCM Generation 2 Sensor Modules Instructions
- September 20, 2024
- BRASCH
Table of Contents
BRASCH TGK-NCM Generation 2 Sensor Modules Instructions
Overview
Brasch Generation 2 sensor boards are calibrated at the factory and have an
expected calibration duration of approximately two years. Environmental
conditions may shorten or prolong this period of time, but once the sensor
parameters drift outside of acceptable tolerances, the sensor board must be
either be recalibrated or replaced.
This document outlines the procedure for performing field calibration of all
Brasch Generation 2 sensor boards including:
Carbon Monoxide Models
- GSE2-CM-24
- GSE2-CM-120
- GSE2-CM-Remote
- TRNS2-CM-Analog
- TRNS2-CM-Digital
- GDCP-CM-Remote
Nitrogen Dioxide Models
- GSE2-ND-24
- GSE2-ND-120
- GSE2-ND-Remote
- TRNS2-ND-Analog
- TRNS2-ND-Digital
- GDCP-ND-Remote
Combination Models
- GSE2-NCM-24
- GSE2-NCM-120
- GSE2-NCM-Remote
- TRNS2-NCM-Analog
- TRNS2-NCM-Digital
- GDCP-NCM-Remote
Required
- Voltmeter
- Potentiometer Adjustment Tool
- 100 PPM Carbon Monoxide Test Gas (for CO sensors)
- 5.0 PPM Nitrogen Dioxide Test Gas (for NO2 sensors)
- Gas Flow Regulator
- Tubing
- Calibration Hood
Note: All of the above required parts (except a voltmeter) are included with Brasch CO and NO2 Test Gas Kits.
Procedure
-
Turn power OFF to the sensor board
-
Move the calibration jumpers (JP4) so that both pins are covered by the jumpers
-
Turn power ON to the sensor board
-
Connect the negative voltmeter lead to COM (TP3)
-
Connect the positive voltmeter lead to AN OUT (TP7)
-
Set the voltmeter to measure DC voltage (VDC)
a) If possible, set the meter to display the average voltage to smooth out readings -
Use the potentiometer adjustment tool on R2 until the meter reads approximately 5.0 V
a) This number will fluctuate but should remain stable at or near 5.0 V -
Remove the top calibration jumper on JP4
-
Apply test gas directly to the sensor
a) The voltage at TP7 will begin to drop -
Once the voltage stabilizes, remove the bottom calibration jumper on JP4
a) Stabilization requires approximately 5 minutes from the application of gas -
Remove test gas from the sensor
-
Verify that the voltage at TP7 returns to approximately the same value as was set in step 7
a) Stabilization may take approximately 5 minutes from the removal of gas -
Cycle power to the sensor board to verify that calibration data has been saved
-
Repeat this procedure for any remaining sensor boards
Troubleshooting
- If any of the following errors are present, calibration data was corrupted and this procedure needs to be repeated.
-
- Standalone Detectors: 9997
- Control Panel Remotes: Sensor # Calibration Invalid
- If achieving stable voltages is not possible, it may be necessary to replace the sensor and/or the sensor board.
-
- Over time, control electronics can degrade or suffer damage. If this is the case, replacing the sensor board should yield more stable voltages.
General Contact Information
Phone:
314-291-0440">314-291-0440
Fax:
314-291-0646">314-291-0646
Email:
customerservice@braschenvtech.com
Website: www.braschenvtech.com
Technical Support Contact Information
Phone:
314-291-0440">314-291-0440
ext. 1006
Fax:
314-291-0646">314-291-0646
Email:
customerservice@braschenvtech.com
Website: www.braschenvtech.com/support-for-existing-
systems
Revision: 1.2
Issue Date: January 10, 2022
Document Number: FG01
© Brasch Environmental Technologies, LLC All Rights Reserved
FG01
Rev 1.2 – January 10, 2022
Read More About This Manual & Download PDF
References
- Home - Brasch Environmental Technologies
- Support for Existing Systems - Brasch Environmental Technologies