F2 ZR Touch S Cordless Endodontic Motor Instruction Manual

September 18, 2024
f2

ZR Touch S Cordless Endodontic Motor

“`html

Specifications

  • Manufacturer: F2 Medical Supplies
  • Model: ZR Touch
  • Weight: Varies based on components
  • Speed range: 150-1000rpm
  • Torque range: 0.6-5.0Ncm
  • Operating conditions:
    • Temperature: +10 to +40 degrees Celsius
    • Humidity: 30% – 75%
    • Pressure: 70kPa – 106kPa
  • Storage conditions:
    • Temperature: -10 to +50 degrees Celsius
    • Humidity: 10% – 85%
    • Pressure: 70kPa – 106kPa

Product Usage Instructions

1. Equipment Introduction

The ZR Touch is designed for dentists to drive root canal
instruments for shaping and cleaning during root canal treatment.
It must be used by qualified dentists in hospitals or dental
clinics.

2. Contraindications

a. Not recommended for use by individuals with pacemakers.

b. Patients with pacemakers or other electrical devices should
not use the equipment.

c. Prohibited for patients with hemophilia.

d. Use caution with patients having heart conditions, pregnant
women, and young children.

3. Warning

Read the manual carefully before use. Ensure the device is
turned off before installing or removing components.

4. Product Introduction

The product includes a motor handpiece, a 1:1 contra angle, and
an adapter.

5. Device Installation

5.1 Installation & Removal of Contra Angle

Installation: Connect the contra-angle at
adjustable head positions by aligning the positioning pins with the
slots and inserting until it clicks.

Removal: Pull out the contra-angle straight
when removing.

5.2 Inserting and Removing the File

File Insertion: Press the top cover of the
contra-angle, insert the file until it locks into place, then
gently pull out to confirm locking.

Frequently Asked Questions (FAQ)

Q: Is the ZR Touch suitable for home use?

A: No, the ZR Touch is designed for professional use by
qualified dentists in hospitals or dental clinics.

Q: How should I clean and sterilize the ZR Touch?

A: Refer to the “Cleaning, disinfection, and sterilization”
section in the manual for detailed instructions on proper cleaning
and sterilization procedures.

“`

ZR TOUCH
F2 Medical Supplies www.f2medicalsupplies.com Find Endodontic Files and more
Endo Equipment Online

ZR TOUCH
ENDO MOTOR
INSTRUCTION MANUAL
Please read the instruction manual before operating.
F2 Medical Supplies www.f2medicalsupplies.com
For model ZR Touch

Content

Product instroduction

4

Product Introduction

6

Device installation

8

Device function and operation

10

Maintenance

18

Troubleshooting

19

Cleaning, disinfection and sterilization 20

Disposal of products

22

Warranty

23

Symbol description

24

EMC Information

25

CONGRATULATIONS!
Thank you for your purchase. For optimum safety and performance, read this manual carefully and pay close attention to warnings and precautions before using the instrument. Please keep this manual in a convenient place for quick and easy reference. Hot Line: +86 0755-28540953 E-mail: sales@dental- perfect.com
ShenzhenPerfect Medical Instruments Co., Ltd. Room 103, Building 3, No. 2, Weiqun Road, 4th Community, Henggang Street, Longgang District, Shenzhen

CONTENTS
1 EQUIPMENT INTRODUCTION ………………………………………………….. 02 2 CONTRAINDICATIONS …………………………………………………………….. 02 3 WARNING ………………………………………………………………………………… 02 4 PRODUCT INTRODUCTION …………………………………………………….. 03 5 DEVICE INSTALLATION ……………………………………………………………. 05 6 EQUIPMENT FUNCTION AND USE …………………………………………… 07 7 MAINTENANCE ……………………………………………………………………….. 11 8 TROUBLESHOOTING ………………………………………………………………. 12 9 CLEANING, DISINFECTION AND STERILIZATION ……………………… 12 10 DISPOSAL OF THE PRODUCT …………………………………………………. 13 11 STATEMENT ……………………………………………………………………………. 13 12 SYMBOL DESCRIPTION …………………………………………………………… 14 13 EMC DECLARATION …………………………………………………………………. 15
01

1.Equipment Introduction
The device is specially designed for dentists and used to drive root canal instruments to rotate and reciprocate.
It is clinically used in root canal shaping and cleaning in root canal treatment.
1.1 Scope of application This equipment must be used in hospitals or dental clinic by qualified dentists.
2.Contraindications
a. Disabled by doctors with pacemakers. b. Patients with pacemakers (or other electrical equipment) and are warned not to
use small appliances (such as electric shavers, hair dryers, etc.) are disabled. c. Prohibited for patients with hemophilia. d. Use with caution in patients with heart disease, pregnant women and young
children.
3.Warning
Please read the instruction manual carefully before using this device for the first time. This equipment must be used in hospitals or dental clinic by qualified dentists. The patient shall not be an intended operator. This device can only be used with its accompanying accessoriesdo not use any other power adapters, contra angle, lithium batteries, etc. There is an unacceptable risk of improper replacement lithium batteries.please use the lithium batteries provided by us and replace it according to the correct steps in the instructions. Do not disassemble or repair the equipment by yourself. Any disassembly or repair by yourself may violate safety regulations and cause injury to the patient. Any self-dismantling and repair will not be guaranteed by any promise. Do not remove the contra angle and press the push button before the motor stops rotating,otherwise the contra angle may be damaged. Do not remove the file before the motor stops rotating, otherwise the operator may be injured. Before starting the motor,make sure that the file is in place and locked. Please set the motor output speed and torque according to the parameters recommended by the file manufacturer.
02

Using the reciprocating mode for a long time may cause the motor handpiece to overheat. At this time, it should be left to cool down to reuse. If the motor handpiece frequently overheats, please contact your local dealer. Do not put the equipment near flammable materials. Do not operate the device in the presence of a flammable anesthetic mixture of air, oxygen or nitrogen oxides. Portable and mobile RF communication equipment may affect the performance of this equipment. Avoid strong electromagnetic interference when in use, such as close to mobile phones, microwave ovens, etc. Do not place this equipment near a heat source. The device must be operated and stored in a reliable environment. The equipment is usually operated at +10°C to +40°C, relative air humidity 30%-75%, and atmospheric pressure (70~106) kPa. This equipment should be sterilized and disinfected for the first time and after the patient is used to avoid cross-infection. The motor handpiece cannot be sterilized by high temperature and high pressure, only neutral disinfectant or alcohol can be used to wipe the surface.
4.Product introduction
1.Motor handpiece

2.1:1 Contra angle

3.Adapter

Manufacturers Model Form factor Weight Contra angle Power supply The adapter output The adapter input Electrical safety classification Applied part
Apply part of the material Speed range Torque range Operating conditions
Storage conditions

ShenzhenPerfect Medical Instruments Co.,Ltd. ZR Touch 184mm104mm75mm 112g±1g(with contra angle) transmission ratio 11 3.7V/800mAh lithium battery 5V DC, 2A AC100-240~50/60Hz Class Il devices with internal power
Type B: Contra angle Application part contact time:1 to 10 minutes
Contra angle: copper, stainless steel
150-1000rpm 0.6-5.0Ncm Temperature+10 to +40 Humidity30% – 75% Pressure70kPa – 106kPa Temperature-10 to +50 Humidity10% – 85% Pressure70kPa – 106kPa

4.Base

5.Oil injectors

03

04

5.Device installation
5.1 Installation & removal of contra angle 5.1.1 Installation The contra – angle can be connected at adjustable head positions , Align the positioning pins of the contra – angle with the positioning slots of the contra – angle and insert the head until it clicks.

5.2.2 File removal Press the top cover of the contra-angle , and at the same time pull out the file at an angle perpendicular to the drive shaft of the hand piece.

5.1.2 Removal When removing the contra – angle , pull it straight out .
Warning When install or remove the contra angle , make sure to turn off the device .
5.2 Inserting and Removing the File 5.2.1 File insertion Press the top cover of the contra – angle and insert the file until the file tail is loaded into the inner Iock slot of the contra – angle .Release the contra – angle top cover and gently pull the file out to confirm that the file is locked.
05

Warning
When installing and removing files, the motor needs to be turned off to prevent damage to the operator.
When installing and removing the file, press the top cover of the hand piece until the installation or removal action is completed. Otherwise, the locking needle part inside the contra angle will be damaged.
Please use file that meets ISO standards.( ISO standards; 2.3342.350mm)
06

6.Equipment function and use

6.1 Button definition and settings main button

Select button LCD screen

Adjust button”<” Adjust button”>”

6.2 Power on and off Short press the main key” “to turn on the device. Press the key” “then press the main key” “to turn off the device. The device will automatically shut down after 5 minutes without any operation.
6.3 Motor start and stop When the device is on, select the appropriate mode, press the main key” “to start the motor, press the main button” “again to stop the motor.

6.4 Terms and definitions
When the device is on, select the appropriate mode, press the main key” “to start the motor, press the main button” “again to stop the motor.

RPM Ncm

Rotation speed of file Torque setting value

Forward clockwise rotation

Reserve Anticlockwise rotation

Reciprocating rotation

07

6.5 Mode introduction

Serial number

Mode

Display content

M1

Continuous rotation

M2

Continuous rotation

M3

Continuous rotation

M4

Continuous rotation

M5

Continuous rotation

M6

Reciprocating rotation Rec

M7

Reciprocating rotation Rec

M8

Reciprocating rotation Rec

Mode function
Rotate in one directioncan select clockwise and counterclockwise directions.Speed range:150-1000rpm step value 50rpm. Torque range0.6-5.0N·cmstep value: 0.1N·cm
Rotate in one directioncan select clockwise and counterclockwise directions. Speed range: 150-1000rpm step value 50rpm. Torque range0.6-5.0N·cmstep value: 0.1N·cm
Rotate in one directioncan select clockwise and counterclockwise directions. Speed range: 150-1000rpm step value 50rpm. Torque range0.6-5.0N·cmstep value: 0.1N·cm
Rotate in one directioncan select clockwise and counterclockwise directions.Speed range:150-1000rpm step value 50rpm. Torque range0.6-5.0N·cmstep value: 0.1N·cm
Rotate in one directioncan select clockwise and counterclockwise directions. Speed range: 150-1000rpm, step value 50rpm. Torque range: 0.6-5.0N·cm, step value: 0.1N·cm
Reciprocating rotation Rec Speed range300rpm Angle range150° forward30°reversal
170°forward50°reversal 50°forward170°reversa
Reciprocating rotation Rec Speed range300rpm Angle range150° forward30°reversal
170°forward50°reversal 50°forward170°reversal
Reciprocating rotation Rec Speed range300rpm Angle range150° forward30°reversal
170°forward50°reversal

08

6.6 Parameter settings
Mode selection, press adjust key”<*”>”select different mode, the device has 8 mode, M1-M8.
Speed setting, press the selection key ” “the speed parameter flashes,and then press the adjustment keys “<” “>” to select the desired speed, and press the” “key to confirm the selection.
Torque setting, press the selection key” “, the torque parameter flashes,and then press the adjustment keys “<” “>” to select the desired torque, and press the” “key to confirm the selection.
Direction setting, press the selection key” “the direction parameter flashes,and then press the adjustment keys “<” “>” to select the desired direction, and press the” “key to confirm the selection.

6.7 Preset the file program selection

For convenience, the device is equipped with different file procedures

of 5 manufacturers and 37 models, the speed and torque are set to the

1

manufacturer’s recommended range values. press the adjustment

key”<“”>”to select the desired direction to confirm the file program.

Select file manufacturer: Long press the selection key ” ” to enter the

file selection interface, and press the adjustment keys “<“”>” to select

2

a different file manufacturer. This machine includes Perfect, Dentsply,

VDW, MicroMega, FKG.

Select the directory: Press the selection key ” ” to select the

corresponding file manufacturer, and press the adjustment keys “<”

3

“>” to select different file type programs.

Select the file type: press the selection key ” ” to select the

corresponding file type, and press the adjustment keys “<“”>” to select

4

a different file type program.Press the key” “to confirm the selection.

6.8 Motor handpiece function settings
In order to satisfy the left-handed and right-handed operators, the viewing angle of the display screen can be adjusted by 180°. Press the “>” key without releasing it, and then press the main button” “to enter the adjustment interface. Then press the adjustment keys “<“”>” to select.the key” “to confirm the selection.
09

6.9 Protective function of automatic reverse
In the M1~M5 mode, the motor rotates in one direction. When the motor load torque reaches the preset value, the motor will automatically reverse; when the motor load returns to half of the preset torque value, the motor returns to the original direction of rotation.

Warning
1.The torque trigger automatic reverse protection function only works in M1~M5 (one-way rotation) modes 2.This function does not take effect in M6,M7,M8 modes. 3.When the power of the device is less than 20%, it is not enough to support the maximum torque of the motor 4Ncm, please charge it in time.
6.10 Device charging
This product has a built-in rechargeable lithium battery. Insert the power adapter into the USB port at the bottom of the handle, and the device will enter the charging state.

After fully charged, the battery icon shows full and in a static state, please unplug the power adapter in time when fully charged.

Warning
Keep away from heat sources when charging.
6.11 Device calibration 6.11.1 Calibration method Press the”<“button , then the main button to enter the calibration mode.

Warning
Please keep the battery above 50% to calibration.
10

7. Maintenance
7.1 Battery replacement To replace the battery, use the lithium battery provided by the original manufacturer. Please contact your local dealer or manufacturer for replacement.
Replacement procedure: 1Ensure that the device is turned off 2Remove the silicone plug of the battery cover with tweezers, and then remove the fixing screws with a screwdriver 3Remove the battery cover and remove the old battery 4Put the new battery into the battery compartment and confirm that the battery is normal when you turn it on. 5Install the battery cover and lock it with screws, and then plug the silicone plug into the screw hole.
Warning
Do not disassemble parts that are not related to the replacement of the battery. Do not replace the battery with wet hands, otherwise the battery may short-circuit and damage the device. Do not use lithium batteries other than the original manufacturer, otherwise the equipment may be damaged. If the device is not used for a long time, the lithium battery should be removed.
7.1 Battery replacement To extend the use of the contra angle, it should be lubricated each time it is used or before sterilization.
1)Rotate the oiler into the nozzle of the oiler bottle.

2)Insert the oil injector into the tail of the contra angle and then fill for 2 to 3 seconds until the oil is flowing out of the contra angle.
3)Place the tail of the contra-angle up right for more than 30 minutes and use gravity to drain excess lubricant.

The motor handpiece can Warning not be filled with oil.
11

Note

When filling oil, hold the Contra-angle tightly to prevent the Contra-angle from leaving the oil injector due to the oil injection pressure.

8.Troubleshooting
8.Troubleshooting

Fault status

Possible causes

Solution

Motor calibration failed Low battery
Motor handle gets hot
The use time becomes shorter when the battery is fully charged

Low battery The resistance of the contra angle is too large.
Excessive use time under load Long reciprocating use time
Battery capacity becomes smaller

Charge the device to full power. Lubricate the contra angle.
Stop using, wait until the handle temperature returns to normal
Contact your local dealer or manufacturer to replace the battery

OVER LOAD

Over load

Stop the motor, exit the root tube and re-start

If you have not found the necessary information,you may consult the manufacturer via hotline+ 86-755-28540953E-mailendoservice@dental-perfect.com or address to the service department.

9.Cleaning, disinfection and sterilization

For the purpose of hygiene and safety, the components must be cleaned, disinfected and sterilized before the first use and after each use to prevent cross-infection.

Wipe disinfection
The device should be cleaned and wiped and disinfected regularly. After use, use 75% medical alcohol to thoroughly wipe the contact surface of it to remove the residual substances or debris on the surface of the device.

Sterilization
The contra angle must be cleaned, disinfected and sterilized for the first use and after each use. Please seal the contra angle in a sterilized packaging that meets the EN-868-5 standard. At 134°C, the sterilization time is at least 4 minutes. At 121°C, the sterilization time is at least 20 minutes, and the drying time is at least 15 minutes.

Note

The contra angle can be sterilized repeatedly for at least 250 cycles. The useful life is 1 year, and its lifespan is also different according to the frequency and time of clinical use, and the difficulty of the patient’s oral root canal during treatment.

Do not disinfect the device by steaming. Warning Do not heat and sterilize the device.
12

10.Disposal of the product
Do not throw the device into the household garbage system. Dispose of waste in accordance with local laws and regulation
11. Statement
ZR Touch S warranty for 12 months from the date of purchase by the user. The company does not provide technical data(such as circuit principles, component lists,etc.) to other organizations. If vou have any questions, please contact your local dealer or manufacturer. The manufacturer is not responsible for: 1.Use ZR Touch S for purposes and purposes that violate the specific provisions in this instruction manual. 2.Use methods that are contrary to those stated in this manual for cleaning, disinfection and sterilization operations. 3.Disassemble and repair by unauthorized personnel.
13

12. Symbol Description
Serial number
Manufacturer Date of manufacture
Class II equipment Type B applied part
Attention; Consult operation instruction Warning, if the operation is improper, it may endanger the user or patient Refer to the instructions for use Do not throw the device into the household garbage system Direct current (power connection)
Can only be used indoors
Alternating current Follow the instruction manual
14

13. EMC Declaration

Electromagnetic Compatibility
Attention
1)ZR Touch endo motor meets the relevant requirements of IEC60601-1 standard electromagnetic compatibility; 2)The user should install and use it according to the electromagnetic compatibility information provided by the accompanying document 3)Portable and mobile RF communication equipment may affect the performance of ZR Touch endo motor. Avoid strong electromagnetic interference when using it, such as close to mobile phones, microwave ovens, etc.; 4)The guidelines and manufacturer’s declaration are detailed in the attachment
Warning
1)ZR Touch endo motor should not be used close to or stacked with other equipment. If it must be used close or stacked, it should be observed and verified that it can operate normally under the configuration used 2)Except for the cables sold by the manufacturer of the ZR Touch as spare parts for internal components, the use of accessories and cables other than those specified may result in an increase in emission or a decrease in immunity of the ZR Touch S endo motor.

Part name

Cable Lengthm

Whether to block

ADAPTER CABLE

1.5

yes

Guidelines and manufacturer’s declaration-electromagnetic emissions

The model ZR Touch is intended for use in the electromagnetic environment specified below. The user should assure that it is used in such an environment.

Emissions test Compliance Electromagnetic environment guidance

RF Emissions GB 4824

Group 1

The model ZR Touch uses RF energy only for its internal functions. Therefore, its radio frequency emission is very low and are not likely to cause any interference in nearby electronic equipment.

RF Emissions GB 4824

Class B

Harmonic emission GB 17625.1

Class A

The model ZR Touch uses RF energy only for its internal functions. Therefore, its radio frequency emission is very low and are not likely to cause any interference in nearby electronic equipment.

Voltage fluctuation/flicker emission GB 17625.2

Class A

15

16

Guidance & Declaration — electromagnetic immunity

The model ZR Touch is intended for use in the electromagnetic environment specified below. The user should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment – guidance

Electrostatic discharge GB/T 17626.2

±6 kV contact ±8 kV air

±6 kV contact ±8 kV air

Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.

Electrical fast transient/burst GB/T 17626.4

±2kV for power ±2kV for

supply lines

power supply

lines

Surge GB/T 17626.5

±1 kV line to line ±2 kV line to earth

±1 kV line to line

Mains power quality should be that of a typical commercial or hospital environment.

Voltage dips, short interruptions and voltage variations on power supply input lines GB/T 17626.11

<5 % UT,for

<5 % UT,for

0.5cycle

0.5cycle

UT>95%

UT>95%

40 % UT,for 5 40 % UT,for 5

cycleUT 60% cycleUT 60%

70 % UTfor 25 70 % UTfor 25

cycleUT30% cycleUT30%

<5 % UT,for

<5 % UT,for

5sec

5sec

UT>95%

UT>95%

Mains power quality should be that of a typical commercial or hospital environment. If the user requires continued operation during power mains interruptions, it is recommended that the device be powered from an uninterruptible power supply or a battery.

Power frequency Magnetic field 50/60Hz GB/T 17626.8

3A/m

3A/m,50/60Hz

The power frequency magnetic field should have the level characteristics of the power frequency magnetic field in a typical place in a typical commercial or hospital environment.

NOTEUT is the a.c. mains voltage prior to application of the test level.

17

Guidance & Declaration – Electromagnetic immunity

The model ZR Touch is intended for use in the electromagnetic environment specified below. The user should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance Electromagnetic

level

environment – guidance

RF conduction GB/T 17626.6
RF radiation GB/T 17626.3

3 V 150 kHz80 MHz
3 V/m 80 MHz2.5 GHz

3VEffective value
3V/m

ZR Portable and mobile RF communications equipment should be used no closer to any part of the models ZR Touch S, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance d =1.2 p d =1.2 p 80 MHz800 MHz d =2.3 p 800 MHz2.5 GHz where Pis the maximum output power rating of the transmitter In watts(W)according to the transmitter manufacturer r and dls the recommended separation distance in meters(m). The field strength of the fixed radio frequency transmitter is determined by surveying the electromagnetic field c, and in each frequency range d should be lower than the compliance level. Interference may occur in the vicinity of equipment marked with the following symbols.

NOTE1At 80MHz & 800MHzthe higher frequency range applies. NOTE2The guidelines may not apply in all situations.Electromagnetic propagation is affected by absorption and reflection from structures,objects and people.
1.Stationary transmitters, such as base stations for wireless (cellular/cordless) telephones and terrestrial mobile radios, amateur radios, AM and FM radio broadcasts, and television broadcasts, whose field strength cannot be accurately predicted in theory. In order to assess the electromagnetic environment of fixed radio frequency transmitters, electromagnetic field surveys should be considered. If the measured field strength of the place where ZR Touch S is located is higher than the above applicable RF compliance level, ZR Touch S should be observed to verify its normal operation. If abnormal performance is observed, supplementary measures may be necessary, such as re-adjusting the direction or position of ZR Touch S. 2.In the entire frequency range of 150KHz80MHz, the field strength should be lower than 3 V/m.

18

Recommended isolation distance between portable and mobile RF communication quipment and ZR Touch S

ZR Touch is expected to be used in an electromagnetic environment where radio frequency radiation disturbance is controlled. According to the maximum rated output power of the communication equipment, the purchaser or user can prevent electromagnetic interference by maintaining the minimum distance between the portable and mobile radio frequency communication equipment (transmitter) and ZR Touch as recommended below.

maximum output power of transmitter
W

Separation distance according to frequency of transmitter/m

150 kHz 80 MHz 80 MHz 800 MHz 8800 MHz 2.5 GHz

d =1.2 p

d =1.2 p

d = 2.3 p

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For the rated maximum output power of the transmitter not listed in the above table, the recommended isolation distance d, in meters (m), can be determined by the formula in the corresponding transmitter frequency column, where P is the emission provided by the transmitter manufacturer The maximum output rated power of the machine, in watts (W).

NOTE 1 At 80MHz and 800MHz frequency points, the higher frequency band formula is used.
NOTE 2: The engineering medical frequency band between 150kHz and 80MHz refers to 6.765MHz6.795MHz13.553MHz13.567MHz 26.957MHz27.283MHz40.66MHz40.70MHz.
NOTE 3An additional factor of 10/3 is used to calculate the recommended isolation distance for transmitters in the 150kHz80MHz engineering and medical frequency band and 80MHz2.5GHz frequency range to reduce portable/mobile radio frequency communication equipment accidentally brought into the patient area The possibility of interference caused by time.
NOTE 4These guidelines may not be suitable for all situations. Electromagnetic propagation is affected by absorption and reflection from buildings, objects and human bodies.

19

F2 Medical Supplies www.f2medicalsupplies.com Find Endodontic Files and more
Endo Equipment Online
Shenzhen Perfect Medical Instruments Co., Ltd. AddressRoom 103, Building 3, No. 2, Weiqun Road, 4th Community, Henggang Street, Longgang District, Shenzhen.
Post Code518115 Tel+8675528540953 Fax+8675528540953 Websitewww.dental- perfect.com E-mailsales@dental-perfect.com