airmet 5, 17159437-1 Ventis Pro User Guide
- September 14, 2024
- airmet
Table of Contents
airmet 5, 17159437-1 Ventis Pro
Specifications
- Product Name: Standby Clip Guide
- Manufacturer: Distributed by Air-Met Scientific Pty Ltd
- Part Number : 17159437-1
- Revision Date: April 17, 2020
- Firmware Requirement : Version 4.1 or later
Product Usage Instructions
Instrument Requirements Ensure the following settings on the instrument for proper Standby Clip functionality:
- NFCa (Maintenance): On
- Edit User/Site (Operation): On
- Standby Clip (Alarm): On
Attaching the Standby Clip
- Hook the lip of the Standby Clip over the top, back edge of the instrument.
- Grip the bottom tab of the clip.
- Press downward on the clip’s tab, snapping the clip in place.
Detaching the Standby Clip
- While holding the instrument with one hand…
- Grip the bottom tab of the Standby Clip with your other hand.
- Pull upward on the clip’s tab, separating it from the instrument.
Note: The panic alarm remains active with the Standby Clip attached. To deactivate the panic alarm, detach the clip and press and hold the Enter button.
FAQ
- Q: What should I do if my instrument’s firmware version is not 4.1 or later?
- A: Update your instrument’s firmware to version 4.1 or later to use the Standby Clip feature.
- Q : Can I remove or replace the iAssign Tag from the Standby Clip?
- A: No, the iAssign Tag on the Standby Clip is factory-programmed and should not be removed or replaced.
- Q: How do I know if the Standby Clip is properly attached?
- A: Ensure the clip’s lip is securely hooked over the back edge of the instrument and that it snaps in place when pressing down on the tab.
Standby Clip Guide
WARNING – When attached to a Ventis® Pro, the Standby Clip™ will place man down or selected instrument alarm events on standby. The instrument will NOT alert the operator to hazards associated with these events when the Standby Clip is in use.
- The Standby Clip (part number 17159316) is an iAssign® accessory that snaps over the front of a Ventis Pro instrument.
- The clip uses NFC (Near Field Communication) to place on standby man down or selected instrument alarm events. When the clip is detached from the instrument, any alarms that had been on standby are generally re-enabled within 5 seconds.
- Safety specialists should review the instrument settings below and select which alarm events to place on standby. When set up as described below and the Standby Clip is attached, the audible, vibration, and visual signals for the selected alarm events will be
- placed on standby. Additionally, corresponding standby symbols for the selected alarm events will be displayed on the Ventis Pro’s screen to indicate which events are on standby when the clip is attached.
Instrument requirements
- Setting (menu) Value
- NFCa (Maintenance) On
- Edit User/Site (Operation) On
- Standby Clip (Alarm) Select from:
- Man down
- Man down and gas
- Man down and peer
- Man down, gas, and peer
Firmware version : The instrument must be running firmware version 4.1 or later to use Standby Clip
Note: If the instrument battery charge falls to a critical low level, NFC
is automatically disabled. If NFC is disabled
the Standby Clip will not place alarms on standby.
The instrument panic alarm remains active when the Standby Clip is attached.
Panic alarm signals will activate and display on the instrument when the panic
button is pressed. To deactivate the panic alarm, first detach the clip, then
press and hold the instrument Enter button
The Standby Clip contains a permanently locked, factory-programmed iAssign Tag. Do not attempt to replace or remove this tag from the clip. A lanyard is attached to the Standby Clip. When the clip is not in use, you can attach it to any (ferromagnetic) metal object using the magnets on its front or hang it by the lanyard.
Figure 1 Standby Clip features
Attaching the Standby Clip
- Hook the lip of the Standby Clip over the top, back edge of the instrument.
- Next, grip the bottom tab of the clip.
- Press downward on the clip’s tab, snapping the clip in place.
Detaching the Standby Clip
- While holding the instrument with one hand…
- Grip the bottom tab of the Standby Clip with your other hand.
- Pull upward on the clip’s tab, separating it from the instrument.
Figure 2 Attaching and Detaching Standby Clip
OUR MISSION
Preserving human life on, above, and below the earth Delivering the highest
quality, best customer service … every transaction, every time
CONTACT
Distributed by:
- Air-Met Scientific Pty Ltd
- Air-Met Sales/Service
- P: 1800 000 744
- F: 1800 000 774
- E: sales@airmet.com.au
Air-Met Rental
- P: 1300 137 067
- E : hire@airmet.com.au
- W : www.airmet.com.au
Industrial Scientific Corporation 1 Life Way
- Pittsburgh, PA 15205-7500, USA Phone: 412-788-4353
- 1-800-DETECTS (338-3287)
- Fax: 412-788-8353
- www.indsci.com
Industrial Scientific Corporation
- 1 Life Way, Pittsburgh, PA 15205-7500 USA
- E-mail: info@indsci.com
Industrial Scientific France S.A.S.
- Téléphone : +33 (0)1 57 32 92 61
- E-mail: info@eu.indsci.com
- +86 21 5899 3279
- E-mail: iscapinfogroup@indsci.com
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>