airmet 17154524-1 Slide On Pump Owner’s Manual

September 14, 2024
airmet

airmet 17154524-1 Slide On Pump

Specifications

  • Model: Ventis Slide-on Pump
  • Certifications: ATEX, CSA, GOST, IECEx, INMETRO, UKExd, ULe China Ex
  • Temperature Range: T4
  • Battery Type: AAA alkaline (Energizer EN92 and Duracell MN2400)

Product Usage Instructions

Safety Warnings

  • Read and understand the manual before use. Failure to follow instructions may result in unsafe conditions.

Connecting Sample Tubing

  • When connecting sample tubing to the pump, ensure the pump is powered off to prevent drawing sample air with an open pump door, which can be unsafe.

Powering On and Off

  • Ensure no instrument is installed when powering on the pump to avoid drawing sample air unnecessarily. Power off the pump before removing the instrument.

Battery Maintenance

  • Service and replace battery packs only in nonhazardous areas. Do not touch battery contacts or stack battery packs. Use only approved AAA alkaline batteries and replace all batteries at the same time.

Battery Replacement

  • When replacing battery packs, use only the approved part numbers: 17148313-1, 17134453-X1, 17138041, or 17150608. Part numbers 17150608 and 17138041 are approved for use with specific AAA battery types.

FAQ

  • Q : Can I use rechargeable batteries with the Ventis Slide-on Pump?
  • A: No, only use AAA alkaline batteries of approved types (Energizer EN92 and Duracell MN2400) with the Ventis Slide-on Pump.
  • Q: How should I store the Ventis Slide-on Pump when not in use?
  • A: Store the pump in a safe, dry place away from hazardous atmospheres. Remove batteries if stored for an extended period.

General Information

Each Ventis™ Slide-on Pump is certified by one or more certifying bodies (CBs). Certifications for the Ventis Slide-on Pump, at the time of this document’s publication, are listed below in Table 1. To determine the hazardous-area classifications for which a pump is certified, refer to its label or the product order.

Table 1 Hazardous-area Certifications

Certifying Body (CB)| Area Classifications| Standards| Approved

Temperature Range

ATEXa| Ex ia I Ma| EN IEC 60079-0: 2018| -20 °C to +50 °C
| Ex ia IIC T4 Ga| EN 60079-11: 2012| (-4 °F to +122 °F)
| Equipment Group and Category: I M1 and II| EN 50303: 2000|
| 1G;| |
CSAb| Class I, Division 1, Group A B C D, T4| CAN/CSA-C22.2 No. 157-92| -20 °C to +50 °C
| Ex ia IIC T4| CSA-C22.2 No. 30-M1986| (-4 °F to +122 °F)
| | CSA-C22.2 No. 152-M1984|
| | CSA-C22.2 No. 60079-0:19

CAN/CSA-C22.2 No. 60079-1: 11

|
| | CAN/CSA-C22.2 No. 60079-11: 14|
| | CAN/CSA-C22.2 No. 61010-1-12|
| | UPD1: 2015, UPD2: 2016, AMD1: 2018|
GOST- EAC| 0 Ex ia IIC T4 Х

PO Ex ia I X

| GOST R 51330.0,

GOST R 51330.10,

GOST R 51330.20,

GOST 24032,

GOST 14254

| -20 °C to +50 °C

(-4 °F to +122 °F)

IECExc| Ex ia IIC T4 Ga| IEC 60079-0: 2017| -20 °C to +50 °C
| | IEC 60079-11: 2011| (-4 °F to +122 °F)
INMETRO| Ex ia IIC T4 Ga| ABNT NBR IEC 60079-0: 2020| -20 °C to +50 °C
| UL-BR 13.0870| ABNT NBR IEC 60079-11: 2013| (-4 °F to +122 °F)
UKExd| Ex ia I Ma

Ex ia IIC T4 Ga

Equipment Group and Category: I M1 and II 1G

| EN IEC 60079-0: 2018

EN 60079-11: 2012

EN 50303: 2000

| -20 °C to +50 °C

(-4 °F to +122 °F)

ULe| Class I, Division 1, Groups A, B, C, and D; T4

Class I, Zone 0, AEx ia IIC T4 Ga Class II, Group F, and G; T4

| UL 913 8th Ed.

UL 60079-0 6th Ed.

UL 60079-11 6th Ed. CSA C22.2 No. 157

| -20 °C to +50 °C

(-4 °F to +122 °F)

China Ex| Ex ia IIC T4 Ga| GB/T 3836.1-2021,

GB/T 3836.4-2021

| -20 °C to +50 °C

(-4 °F to +122 °F)

The EC type-examination certificate is DEMKO 12 ATEX 1204290 with marking code Ex ia I Ma and Ex ia IIC T4 Ga for the equipment group and category II 1G andI M1. The Ventis Slide-on Pump (VSP) complies with relevant provisions of European ATEX directive 2014/34/EU and EMC directive 2014/30/EU. The VSP is constructed with reference to published standards of directive 2014/35/EU, to eliminate electrical risks and fulfill 1.2.7 of ANNEX II of directive 2014/34/EU.  bThe VSP is CSA certified according to the applicable CSA standards for use in Class I, Division 1, and Class I, Zone 0 hazardous locations within an ambient-temperature range of Tamb: -20 °C to +50 °C.

  • The IECEx examination certificate is IECEx UL 12.0021 with marking code Ex ia IIC T4 Ga for hazardous locations with an ambient temperature range of Tamb: -20 °C ≤ Ta ≤ +50 °C. The charging contact parameters are: Um = 6.2V; Ii = 1.3A.
  • The VSP is UKEx certified according to the UL22UKEX2724 certification standard.
  • The VSP is UL certified according to the applicable standards of UL File E218330, Volume 1, Section 10.

Table 2 Warnings and cautionary statements

| Read and understand the manual before use. Failure to perform certain procedures or note certain conditions may impair the performance of this product.
| When sample tubing is connected to the pump and the pump is powered on, it continues to draw sample air even if the pump door is open. This can cause an unsafe condition. Remove the sample tubing from the pump inlet or power off the pump before opening the pump door.
| When the pump is powered on and no instrument is installed, the pump continues to draw sample air. This can cause an unsafe condition. Power off the pump before removing the instrument.
| Service the pump and replace or charge the battery packs only in an area known to be nonhazardous. Not for use in oxygen-enriched atmospheres.
| Battery contacts are exposed on battery packs when they are removed from the instrument. Do not touch the battery contacts and do not stack battery packs on top of each other.
| WARNING –Do not use the Ventis Slide-on Pump when sampling for these gases: Chlorine (CL2), Chlorine Dioxide (CLO2), Hydrogen Chloride (HCL), and volatile organic compounds (VOC), or when a sensor for any of these gases is installed and the target gas is unknown. The use of the Slide-On Pump with these gases will result in inaccurate gas readings due to their susceptibility to absorption.
| Insert the alkaline batteries with the correct positive (+) and negative (-) orientation. The Ventis Slide-on Pump is approved for use with the AAA alkaline battery types Energizer EN92 and Duracell MN2400 only. Do not mix battery types; when replacing alkaline batteries, replace each and every battery.
IECEx
| Replace battery packs with these part numbers only.

IECEx: 17148313-1, 17134453-X1, 17138041, or 17150608.

| Alkaline battery pack part numbers 17150608 and 17138041 are only approved for use with three AAA battery types Duracell MN2400 and Energizer EN92. Replace all batteries at the same time.

Motorized pump and sampling line guidelines
When sampling with a motorized pump and sampling line, Industrial Scientific recommends the following:

  • When the target gases are unknown, use Teflon-lined tubing. Otherwise, use either urethane or Teflon-lined tubing.
    Note: WARNING–Do not use the Ventis Slide-on Pump when sampling for chlorine (Cl2), chlorine dioxide (ClO2), hydrogen chloride (HCl), or volatile organic compounds (VOCs) or when the target gas is unknown and sensors for these gases are installed. The use of the Slide-On Pump with these gases will result in inaccurate gas readings due to their susceptibility to absorption.

  • Know the length of the sample line as it is a factor in determining sampling time. A sample line may consist of tubing, a probe, or a probe and tubing. It should also have a dust filter-water stop installed at the line’s end that will extend into the sample area. Sample-line length is defined as the distance from the dust filter-water stop opening to the point where the line connects to the pump’s inlet. Ensure sample-line length does not exceed the pump’s maximum draw.

  • Before and after each air sample, perform a test of the full sampling line.

    • Use your thumb to block the end of the sampling line at the water-stop opening. This should cause a pump fault alarm.

    • Unblock the water-stop opening. After the alarm cycle completes, the pump should resume normal operation.
      Note: If a pump fault does not occur, check and correct for cracks or other damage, debris, and proper installation in these areas: the sampling line and its connections, the pump’s inlet cap and inlet barrel, and the dust filter- water stop items at the end of the sampling line and inside the pump inlet barrel.

    • Based on sample-line length, calculate the minimum time recommended for the air sample to reach the instrument’s sensors. As shown in Table 3, use a base time of 2 minutes, and add 2 seconds for each 30 cm (1 ‘) of line length. Watch the display screen for gas readings and, if present, allow them to stabilize to determine the reading.

Table 3 Minimum sample time for common sample-line lengths

Sample-line length| Base time (minutes)| +| Sample-line-length factor| =| Minimum sample time (mm: ss)
3.05 m (10 ‘)| 2 min| +| (10 ‘ x 2 s)| =| 02:20
6.10 m (20 ‘)| 2 min|

| (20 ‘ x 2 s)| =| 02:40
9.14 m (30 ‘)| 2 min|

| (30 ‘ x 2 s)| =| 03:00
12.10 m (40 ‘)| 2 min|

| (40 ‘ x 2 s)| =| 03:20
15.24 m (50 ‘)| 2 min|

| (50 ‘ x 2 s)| =| 03:40
18.29 m (60 ‘)| 2 min|

| (60 ‘ x 2 s)| =| 04:00
21.34 m (70 ‘)| 2 min|

| (70 ‘ x 2 s)| =| 04:20
24.38 m (80 ‘)| 2 min|

| (80 ‘ x 2 s)| =| 04:40
27.43 m (90 ‘)| 2 min|

| (90 ‘ x 2 s)| =| 05:00
30.48 m (100 ‘)| 2 min|

| (100 ‘ x 2 s)| =| 05:20

Product Overview

  • The Ventis Slide-on Pump is an in-field attachable and removable pump. This enables the use of a single compatible instrument for both personal monitoring (diffusion) and confined space applications (aspirated).

Installation

  • The pump’s spring-assisted, hinged door is opened and securely closed by the body-mounted latch; therefore, no tools are needed for instrument installation or removal. The pump also allows for the installation of a clip-equipped instrument.

Power source

The pump is powered by its own battery and it does not draw power from the instrument battery. When using Li-ion battery packs, the pump and instrument batteries should be charged separately.
Sample draw
50 feet. Flow rate (nominal): 0.27 liters per minute (LPM).

Table 4 Compatibilities

Category| Part number| Item| Notes
Instruments| Varies| Ventis Pro Series, Ventis MX4| Diffusion unit only

Use of the VSP may result in sensor readings that are within ± 5% of their published accuracy specifications; sensor T90 response times may increase by 10 seconds.

Pump Door| 17157329-X| Door| Ventis Pro; Ventis MX4
| 17154396a| Door| Ventis MX4; MX4 iQuad
Tubing| 18109206-XX| Air-sample tubing kit (Teflon-lined)| When the target gases are unknown , use Teflon-lined tubing. Otherwise, use either urethane or Teflon-lined tubing.

Note: Do NOT use the Ventis Slide-on Pump when sampling for chlorine (Cl2), chlorine dioxide (ClO2), hydrogen chloride (HCl), or volatile organic compounds (VOCs) or when the target gas is unknown and sensors for these gases are installed. The use of the Slide-On Pump with these gases will result in inaccurate gas readings due to their susceptibility to absorption.

| 18109207-XX| Air-sample tubing kit (Urethane)
Batteries| 17134453-XY| Ventis Standard Battery| Run time is 18 hours at 20º C; recharge time is 3 to 5 hoursc.
| 17148313-Y| Ventis Extended Run-time Battery| Run time is 36 hours at 20º C; recharge time is 3 to 7.5 hoursc.
| 17151184-XY| Cover| For use with Ventis Extended Run-time Battery.
| 17154577-XY| Alkaline battery kit| Includes pack, batteries, and cover. Run time: 10 hours at 20º C (typical)c.
“X” denotes color where 0=black and 1=orange. “Y” denotes approvals where 1 = ATEX, CSA, IECEX, and UL; and 5 = INMETRO.
Chargers| 18108191| Single-unit charger| —
| 18108209| Single-unit charger with datalink| —
| 18108650-A| 6-unit charger| —
| 18108651| 12VDC Single-unit automotive charger| With plug
18108652| 12VDC Single-unit truck-mount charger| With plug
---|---|---
18108653| 12VDC Single-unit truck-mount charger| Hard wired

  • aMX4 iQuad can only be used with pump door 17154396.
  • bShips with pump as ordered.
  • cBattery run-times are typical for operation at room temperature

Hardware overview

The main hardware components of the Ventis Slide-on Pump are identified below in Table 5.
Table 5 Slide-on pump diagram and parts list

NOTE: If any ordered item is missing or appears to have been damaged, contact a local distributor of ISC products or ISC (See” Contact Information”). — indicates part is not field replaceable. See “Contact Information” to locate an ISC Service Center.

Setup, Operation, and Service

Figure 1, below, details key setup, operation, and service tasks for the Ventis Slide-on Pump including pump setup, instrument installation, pump operation, and service activities.

Figure 1 Setup, operation, and service tasks

Pump setup

WARNING: This task must be completed in an area known to be nonhazardous.

  • Attach the battery to the upper portion of the pump case back; the battery contacts align with the top of the pump.
  • Tighten the battery’s four captive screws to a torque setting of 0.39 Newton-meters (55-ounce-inch).
  • To attach the wrist strap to the pump:
  • Thread the shorter strap loop through the pass-through on the back of the pump. Thread the longer strap loop through the shorter loop, and pull it to tighten.

Tip. The battery charge indicators are:

  • Solid green indicates charging is complete.
  • Blinking green indicates charging is in progress.
  • Blinking amber indicates a charge fault.

Charge the pump in a compatible charger (single-unit charger shown). The pump’s orientation is contact-side down so its battery contacts touch the charger’s contact pins.

Instrument installation and removal

  • To open the door, move the latch up to its open position.

  • Check for and clear the pump inlet, door openings, exhaust filter, and audible alarm filter (see inset photo), of any dirt or debris.

  • To install the instrument, place it in the pump cradle at a slight angle; the instrument top meets the cradle top. Press down on the instrument to secure it in the cradle.
    NOTE: Remove the instrument in a similar fashion by pulling it up from and out of the pump.
    Tip. To prevent damage to or loss of the door, transport or store the pump with its door closed and latched.

  • To close the pump door, lower it and hold it closed. Move the latch to its closed position to secure the door.

  • After installing the instrument and before taking an air sample, perform a pump test.

Pump test and operation

  1. Power on the pump: press and hold the power button for three seconds. The pump will emit a series of audio and visual indicators as it goes through the start-up sequence. When the pump’s power status LED turns green, it is ready for use.
  2. Check the pump’s battery status indicator.
  3. Solid green indicates the battery’s charge level is between 50% and 100%.
  4. Solid amber indicates the battery’s charge level is at 50%or less.
  5. A blinking amber light with audible beep indicates a critically low battery; the battery’s remaining charge will operate* the pump for less than one hour.
  6. Attach one end of the sample tubing to the pump inlet nipple. Use a compatible water stop at the other end of the sampling line.
  7. Use your thumb to block the end of the sampling line at the water-stop opening. This should cause a pump fault alarm. Unblock the water-stop opening. After the alarm cycle completes, the pump should resume normal operation.

Note : If a pump fault does not occur, first check the pump’s inlet cap and ensure that it is tight. Then check and correct for cracks or other damage in the sampling line and its connections, the pump’s inlet cap and inlet barrel, and the dust filter-water stop items at the end of the sampling line and inside the pump inlet barrel.
Operation at room temperature.

Shutting Down

To power off the pump, press and hold the pump power button The audible indicator will sound five times before the complete shutdown.

Service Tasks

Door Installation and removal

  1. Power off the pump.
  2. Open the door.
  3. To remove the door, lightly press it past the fully open position.
  4. A wire extends from the spring coil inside the covered hinge. To attach the door, ensure the wire rests on the inside of the door. Press the door’s flexible pegs into the pump case grooves located above and below the covered hinge.

Battery removal and installation

WARNING: This task must be completed in an area known to be nonhazardous.

  1. Power off the pump.
  2. To remove the battery from the pump, loosen its four captive screws.
  3. Attach the battery to the pump case back; align the battery contacts with the top of the pump.
  4. Tighten the battery’s four captive screws to a torque setting of 0.39-newton meters (55 ounce-inch).

Water barrier

  1. Power off the pump.
  2. Turn the pump inlet cap counterclockwise to remove.
  3. Remove the water barrier from the inlet barrel.
  4. Place the new water barrier inside the inlet barrel; the side with the larger filter surface should face the instrument operator.
  5. Reattach the cap and turn it clockwise to tighten. Ensure that the inlet cap is tight.

NOTE: The pump will not operate without a water barrier.

Exhaust filter or audible alarm filte r

  1. Using tweezers, peel off the audible alarm filter (shown) or the exhaust filter and discard.
  2. Lift to remove a new filter from the sheet.
  3. Place the filter on the alarm (or exhaust), adhesive side down. Press and hold for five seconds to ensure the adhesive is activated.

Figure 1 Setup, operation, and service tasks

Alarms

The following table describes the types of alarms that may occur when using the pump. Indicators, possible causes, and recommended actions are described for each alarm type. For additional assistance in resolving any alarm condition, see an on-site supervisor or contact ISC Technical Support (see “Contact Information”)

Table 6 Alarms and indicators

Alarm type and indicators| Possible causes| Recommended actions
Battery failure| |
·         Audible alarm beeps on and off| The battery pack can no longer operate| Charge or replace the battery pack.
continually.| the pump unit.|
·         Battery status LED blinks amber| |
continually.| |
System alarm| |
·         Audible alarm turns on.

·         Each LED blinks twice.

· The audible alarm turns off.

| There is a system-level fault.| Check the inlet barrier for blockage. See an on-site supervisor or contact ISC Technical Support (see “Contact

Information”).

·         The above sequence repeats continuously.| |
Pump flow fault| |
·         Audible alarm turns on.

·         Each LED blinks once.

· The audible alarm turns off.

·         The above sequence repeats continuously.

| The pump nipple, inlet barrel, water barrier, exhaust (inside pump cradle), or sample tubing is blocked.| Check for debris at the pump nipple, inside the inlet barrel, at the water barrier, inside the pump cradle, and in the tubing. Attempt to clear any blockage.

If the condition persists, the filter or the tubing may need to be replaced.

Charge fault| |
·         Battery status LED blinks amber continually.| The pump unit is not properly seated in the charger or the battery is alkaline and cannot be charged.| Remove the pump from the charger. Redock a Li-ion-powered pump.
| The battery temperature is outside the allowable temperature range for charging.| Allow the battery to reach a temperature that is within the allowable range for charging.

Warranty

Industrial Scientific Corporation’s Ventis™ Slide-on Pumps are warranted to be free from defects in material and workmanship for a period of two years after purchase. This warranty includes the pump and the battery pack as shipped with the Ventis™ Slide-on Pump.
Filters are warranted to be free from defects in material and workmanship for 18 months from date of shipment, or one year from date of first use, whichever occurs first, except where otherwise stated in writing in Industrial Scientific literature.

Limitation of Liability

  • THE WARRANTY SET FORTH ABOVE IS STRICTLY LIMITED TO ITS TERMS AND IS IN LIEU OF ALL OTHER WARRANTIES, GUARANTEES, EXPRESS OR IMPLIED, ARISING BY OPERATION OF LAW, COURSE OF DEALING, USAGE OF TRADE OR OTHERWISE. INDUSTRIAL SCIENTIFIC MAKES NO OTHER WARRANTIES, EITHER EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE WARRANTIES OF MERCHANTABILITY OR FITNESS FOR PARTICULAR PURPOSE.
  • SHOULD THE PRODUCT FAIL TO CONFORM TO THE ABOVE WARRANTY, BUYER’S ONLY REMEDY AND INDUSTRIAL SCIENTIFIC’S ONLY OBLIGATION SHALL BE, AT INDUSTRIAL SCIENTIFIC’S SOLE OPTION, REPLACEMENT OR REPAIR
  • OF SUCH NON-CONFORMING GOODS OR REFUND OF THE ORIGINAL PURCHASE PRICE OF THE NONCONFORMING GOODS. IN NO EVENT WILL INDUSTRIAL SCIENTIFIC BE LIABLE FOR ANY OTHER SPECIAL, INCIDENTAL, CONSEQUENTIAL, PUNITIVE, OR OTHER SIMILAR DAMAGES, INCLUDING LOSS OF PROFIT OR LOSS OF USE, ARISING OUT OF THE SALE, MANUFACTURE OR USE OF ANY PRODUCTS SOLD HEREUNDER WHETHER SUCH CLAIM IS PLEADED IN CONTRACT OR IN TORT, INCLUDING STRICT LIABILITY IN TORT AND WHETHER INDUSTRIAL SCIENTIFIC HAS BEEN ADVISED OF THE POTENTIAL FOR SUCH DAMAGES.

Industrial Scientific’s total liability hereunder from any cause whatsoever (except liability from personal injury caused by Industrial Scientific’s negligence), whether arising under contract, warranty, tort (including negligence), strict liability, products liability or any other theory of liability, will be limited to the lesser of Buyer’s actual damages or the price paid to Industrial Scientific for the Products that are the subject of Buyer’s claim. All claims against Industrial

  • Scientific must be brought within one year after the cause of action arises, and Buyer expressly waives any longer statute of limitations.
  • It shall be an express condition to Industrial Scientific’s warranty that all products be carefully inspected for damage by Buyer upon receipt, be properly calibrated for Buyer’s particular use, and be used, repaired, and maintained in strict accordance with the
  • instructions set forth in Industrial Scientific’s product literature. Repair or maintenance by non-qualified personnel will invalidate the warranty, as will the use of non-approved consumables or spare parts.
  • As with any other sophisticated product, it is
  • essential and a condition of Industrial Scientific’s warranty that all personnel using the products be fully acquainted with their use, capabilities, and limitations as set forth in the applicable product literature.
  • Buyer acknowledges that it alone has determined the intended purpose and suitability of the goods purchased. It is expressly agreed by the parties that any technical or other advice given by Industrial Scientific with respect to the use of the goods or services is
  • given without charge and at Buyer’s risk; therefore, Industrial Scientific assumes no obligations or liability for the advice given or results obtained.

Contact Information

Industrial Scientific Corporation

Industrial Scientific France S.A.S.

  • 5 Rue Frédéric Degeorge, CS 80097 62002 Arras Cedex, France
  • Web: www.indsci.com
  • Téléphone : +33 (0)1 57 32 92 61
  • E-mail: info@eu.indsci.com
  • Fax: +33 (0)1 57 32 92 67

To locate a nearby distributor of our products or an Industrial Scientific service center or business office, visit us at www.indsci.com.

www.airmet.com.au

www.indsci.com/ventis/vsp

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals