Fieldpiece ST4 Dual Temp Digital Thermometer Instruction Manual
- August 29, 2024
- Fieldpiece
Table of Contents
Fieldpiece ST4 Dual Temp Digital Thermometer Instruction Manual
Description
The ST4 Dual Temp Digital Thermometer is ruggedized for field use and simple
to use. The rubber boot protects against drop impacts and is comfortable to
hold.
Plug in 2 Type K thermocouples and see the calculated difference in real time.
Each thermocouple jack has its own offset pot for accurate calibration. Strap
a thermocouple to a pipe using the included velcro straps. Use the ST4 in the
included clear case for added protection and easy viewing with magnet.
WARNING
To avoid electrical shock, do not use this instrument when voltages at the
measurement surface exceed 24VAC or 60VDC.
To avoid electrical shock, disconnect thermocouples from the ST4 before
opening case or battery cover.
To avoid damage or burns, do not take temperature measurements in microwave
ovens.
When testing hot temperatures, the thermocouple and velcro may become hot. Do
not handle the thermocouple or the velcro when hot.
**Velcro Pipe Stabilizing Straps
ST4 Hanging in Case
**
Controls
- Select T1 to display.
- Select T2 to display.
- Select real time T1-T2 to display.
- Display minimum or maximum captured measurement. Press and hold to clear.
- Hold measurement.
- Select resolution: 1° or 0.1°
- Select temperature scale: °C or °F
- Power ON or power OFF.
- Adjust calibration offset of T1 Type K jack.
- Adjust calibration offset of T2 Type K jack.
Indicators
- LED shines Red when a required
- thermocouple for selected display mode is not connected.
- Displaying held measurement.
- Displaying maximum captured measurement.
- Displaying minimum captured measurement.
- Displaying T1 thermocouple.
- Displaying T2 thermocouple.
- Displaying T1-T2 in real time.
- Temperature set to Fahrenheit.
- Temperature set to Fahrenheit.
- Battery needs to be replaced.
Maintenance
CLEANING: Clean the exterior with a damp cloth. Do not use detergents or
solvents.
BATTERIES: The 9V battery must be replaced when the low battery icon
is displayed. Pull off the rubber boot, unscrew the battery cover, and replace
the old battery.
FIELD CALIBRATION: Calibration should be performed regularly for the
highest accuracy. When you calibrate the ST4, you’re calibrating that jack (T1
or T2) to the specific thermocouple. If you change thermcouples, calibrate it
to that jack.
- Stabilize distilled ice water.
- Immerse the bead of the T1 thermocouple.
- Select T1.
- Use the T1 calibration pot to dial in 32.0°F.
- Repeat for T2.
FIXING A BROKEN WIRE: Due to frequent bending, a thermocouple wire may
break or come loose near the plug. To repair, cut and strip the thermocouple
wire near the plug.
The red wire is the (-) wire and is connected to the wider prong. The yellow
wire is the (+) wire and is connected to the thinner prong.
Loosen the screws on the plug and wrap the conductors around the appropriate
screws and tighten.
Finally, position the plugs into the tab and screw the tab back together.
Specifications
Specifications good in ambient conditions of 73°F ±9°F (23°C ±5°C), <70% RH
Sensor type: Type K thermocouples (nickel chromium/nickel aluminum)
Display: 3.5 digit LCD, 2000 count
Over range: “OL” or “-OL” is displayed
Range: Meter: -58 to 2000°F (-50 to 1300°C);
Thermocouple: -58 to 400°F (-50 to 204°C)
Resolution: 0.1°
Accuracy:
±4°F @ -58 to 32°F ±(0.3%rdg +2°F) @ 32 to 1100°F ±(0.5%rdg +2°F) @ 1100 to
2000°F; ±2°C @ -50 to 0°C ±(0.3%rdg +1°C) @ 0 to 600°C ±(0.5%rdg +1°C) @ 600
to 1300°C
Accuracy after field calibration: ±1°F @ 30 to 120°F, ±0.6°C @ 0 to 50°C
Temperature coefficient: 0.1 x (specified accuracy) per °C (0 to 18°C, 28
to 50°C)
Measurement rate: 2.5 readings/second
Battery type: 9V, NEDA 1604, JIS 006P, IEC 6F22
Battery life: 200 hours typical alkaline.
is displayed when the battery needs to be replaced.
Operating environment: 32°F to 122°F (0°C to 50°C) at <70% RH
Storage temperature: -4°F to 140°F (-20°C to 60°C), 0 to 80% RH (with
battery removed)
Limited Warranty
This meter is warranted against defects in material or workmanship for one
year from date of purchase from an authorized Fieldpiece dealer. Fieldpiece
will replace or repair the defective unit, at its option, subject to
verification of the defect.
This warranty does not apply to defects resulting from abuse, neglect,
accident, unauthorized repair, alteration, or unreasonable use of the
instrument.
Any implied warranties arising from the sale of a Fieldpiece product,
including but not limited to implied warranties of merchantability and fitness
for a particular purpose, are limited to the above. Fieldpiece shall not be
liable for loss of use of the instrument or other incidental or consequential
damages, expenses, or economic loss, or for any claim of such damage,
expenses, or economic loss.
State laws vary. The above limitations or exclusions may not apply to you.
Obtaining Service
Visit www.fieldpiece.com/rma for the
latest information on how to obtain service.
More Fieldpiece
There’s always something new coming from Fieldpiece Instruments.
Find the latest news and updates at www.
fieldpiece.com
Quick Start
- Install the 9V battery.
- Press ON OFF for 1 second to power ON the ST4.
- Plug thermocouples into the top.
- Press T1, T2, or T1-T2 to select thermocouple display mode.
- Read temperature on display.
Certifications
EN61326-1, CE-EMC
RCM
WEEE
RoHS Compliant