ANALOG DEVICES CN-0565 ARDZ EIT Measurement System Owner’s Manual
- August 20, 2024
- Analog Devices
Table of Contents
- ANALOG DEVICES CN-0565 ARDZ EIT Measurement System
- Product Information
- Product Usage Instructions
- Electrical Impedance Tomography Measurement System
- CIRCUIT FUNCTIONS AND BENEFITS
- CIRCUIT DESCRIPTION
- Crosspoint Switch
- INPUT MEASUREMENT
- POWER ARCHITECTURE
- COMMON VARIATIONS
- CIRCUIT EVALUATION AND TEST
- LEARN MORE
- FAQ’S
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
ANALOG DEVICES CN-0565 ARDZ EIT Measurement System
Product Information
Specifications
- Devices Connected/Referenced : AD5940, ADG2128, ADuM5020, ADM7150, ADM8829, ADuM3151, ADuM1250
- Measurement Frequency Range : 0.015 Hz to 200 kHz
- Isolation : Complete power and signal isolation from the host controller
- Form Factor : Arduino-compatible
Product Usage Instructions
Circuit Functions and Benefits:
The EIT measurement system constructs images by placing an array of electrodes
across a sample’s surface. Impedance profiles are measured by exciting a pair
of electrodes with a current and measuring the resulting voltage between
another pair of electrodes. The system allows for 2D or 3D image production
based on internal impedance profiles.
Impedance Measurement:
Multiple, independent impedance measurements are necessary to create
tomographic images. The system supports both DC resistance and AC impedance
measurements. Impedance measurements can be performed using either two or four
electrodes.
2-Wire Impedance Measurement:
For the simplest form of impedance measurement, connect two wires to the
unknown sample as shown in Figure 4.
Bioimpedance Considerations:
Bioimpedance measurements have additional considerations such as conforming to
the IEC 60601 standard for medical devices. Electrode impedance variations
must also be considered.
Circuits from the Lab® reference designs are engi- neered and tested for quick and easy system integration to help solve today’s analog, mixed-signal, and RF design challenges. For more information and/or support, visit www.analog.com/CN0565. Devices Connected/Referenced
AD5940 | High Precision, Impedance & Electrochemical Front End |
---|---|
ADG2128 | I2 C CMOS 8×12 Analog Switch Array with Dual/Single Supplies |
ADuM5020 | Low Emission |
500 mW Isolated DC-to-DC Converter
ADM7150| 800 mA, Ultralow Noise/High PSRR LDO
ADM8829| Switched Capacitor Voltage Inverter
ADuM3151| 3.75 kV, 7-Channel, SPIsolator Digital Isolators for SPI (with 2/1
Aux Channel Directionality)
ADuM1250| Hot Swappable Dual I2 C Isolator
Electrical Impedance Tomography Measurement System
EVALUATION AND DESIGN SUPPORT
- Circuit Evaluation Boards
- CN0565 Circuit Evaluation Board (EVAL-CN0565-ARDZ)
- Design and Integration Files
- Schematics, Layout Files, Bill of Materials, Software
CIRCUIT FUNCTIONS AND BENEFITS
Electrical impedance tomography (EIT) is a type of imaging in which an image is constructed from multiple impedance measurements taken between various points across the surface of a specimen. Compared with other imaging techniques such as X-ray, ultrasound, or computerized axial tomography (CAT) scan, EIT is considerably lower cost, lower power, radiation-free, and easily adapted to portable, wearable, and continuous monitoring applications. While EIT is typically lower resolution than X-ray, ultrasound, or CAT scan, it finds broad application in medical imaging, often augmenting other methods. EIT can be as effective as other imaging methods in certain pulmonary function measurements. EIT images are constructed by placing an array of electrodes across the surface of a sample.
A pair of electrodes is then excited with a current, and the resulting voltage is measured between a second pair of electrodes. A series of measurements is taken across the sample, with the exact sequence dependent on the imaging technique. From these measurements, a calculation determines the internal impedance profile, from which a two- or three-dimensional (2D or 3D) image is produced. The circuit shown in Figure 1 is a 24-electrode EIT measurement system for measuring the impedance profiles of various materials. The electrode configuration is arbitrary; excitation can be applied to any two electrodes, and measurements taken across any two electrodes.
Excitation frequency is adjustable between 0.015 Hz and 200 kHz, allowing the measurement of frequency-dependent effects. The design features complete power and signal isolation from the host controller, essential for medical, bioimpedance analysis (BIA), and certain industrial applications. The module uses an Arduino compatible form factor, and the software example is based on the industry-standard industrial input/output (IIO) framework for easy integration into end-user systems.
Circuits from the Lab circuits from Analog Devices have been designed and built by Analog Devices engineers. Standard engineering practices have been employed in the design and construction of each circuit, and their function and performance have been tested and verified in a lab environment at room temperature. However, you are solely responsible for testing the circuit and determining its suitability and applicability for your use and application. Accordingly, in no event shall Analog Devices be liable for direct, indirect, special, incidental, consequential or punitive damages due to any cause whatsoever connected to the use of any Circuits from the Lab circuits. (Continued on last page)
CIRCUIT DESCRIPTION
ELECTRICAL IMPEDANCE TOMOGRAPHY
In its simplest form, EIT is implemented by placing an array of electrodes
across the surface of a sample and exciting pairs of electrodes with a current
signal while measuring the voltages induced on other electrodes. From these
measurements, a calculation determines the sample’s internal impedance profile
and uses this input to create a 2D or 3D image. Figure 2 shows a typical EIT
system. An array of 8 electrodes is placed around the test object, and an AC
current at a certain frequency is injected into a pair of forcing electrodes
(F+ and F-). The resultant voltage between a pair of sense electrodes (S+ and
S-) is measured and stored. The sequence continues to the next pair of
electrodes until all required inputs to the impedance calculation are
collected. The impedance readings are then mapped to an EIT image showing the
sample’s internal impedance profile.
IMPEDANCE MEASUREMENT
Multiple, independent impedance measurements must be taken in order to create
a tomographic image. The electrical techniques for measuring DC resistance and
AC impedance are well established; conceptually, a circuit element is excited
with a current, and the resulting voltage is measured. In general, impedance
measurements can be performed with either two electrodes or four electrodes,
shown in Figure 3.
Bioimpedance measurements have additional considerations; medical devices must conform to the IEC 60601 standard, which limits the amount of DC and AC voltage that can be applied to the human body. Furthermore, electrodes have their own impedance that can vary considerably from unit to unit, contact to a sample, and over time.
2-Wire Impedance Measurement
The simplest form of impedance measurement uses two connections to the
unknown sample, shown in Figure 4.
A voltage excitation signal is applied across the unknown impedance (ZUNKNOWN) in series with a current limiting resistor (RLIMIT). The voltage across the electrodes (VZUNKNOWN) is measured, which does not include the drop across the limiting resistor. At the same time, current flow through the unknown impedance (IZUNKNOWN) is converted to a voltage by a transimpedance amplifier (TIA). In practice, the 2-wire method is not practical for impedance measurements in medical applications. While the impedances in these applications are typically high enough that the lead resistance is insignificant, the electrode impedance appears in series with the measured impedance. Also in tomography applications, multiple sense measurements at other locations on the sample may be required for a given pair of force electrodes.
4-Wire Impedance Measurement
Figure 5 shows how the 4-wire impedance measurement technique overcomes the
electrode impedance problem by using separate force and sensing electrodes.
Any voltage drop across the force electrodes drops out of the calculation,
similar to the drop across RLIMIT. The IN+ and IN- sense inputs are high-
impedance, so drop across the sense electrodes and their corresponding RLEAD
is negligible.
ANALOG FRONT END
The core of the CN0565 is the AD5940 high precision, impedance, and
electrochemical front end. Table 1 shows the two operation modes of AD5940’s
digital-to-analog converter (DAC). Table 1. AD5940 DAC Operation
| Low Bandwidth Operation| High Bandwidth Operation
---|---|---
DAC| low power, dual-output DAC| high-speed DAC
Operation| generates the bias voltage for the low power transimpedance
amplifier and the bias voltage for the potentiostat amplifier| generates a
high frequency, AC excitation signal for highspeed transimpedance amplifier
designed to convert high bandwidth current signals
Excitation Frequency Range| < 200 Hz| > 200 Hz
Application Examples| electrodermal activity measurements| body impedance
measurements, battery impedance measurements
The AD5940 includes a flexible switch matrix that can select the electrodes, high speed DAC excitation amplifier, and the high speed TIA inverting input, as well as an external calibration resistor. Figure 6 shows the AD5940 configured for a 4-wire impedance measurements, including 0.47 µF DC-blocking capacitors.
Excitation Signal
The CN0565 includes a high speed DAC and waveform generator to provide a
precision AC voltage source. A digital waveform generator produces a sine
wave, square wave, or trapezoid pattern which is routed to the 12-bit, high
speed DAC and excitation amplifier, which then drives the electrodes. The
impedance under test is connected between the counter electrode pin (CE0) and
reference electrode input pin (RE0). An internal differential sense
configuration inside the AD5940 guarantees the accuracy of the voltage source
by connecting CE0 and RE0 back into the excitation buffer. The excitation
signal amplitude and frequency are programmable up to ±607 mV and 200 kHz,
respectively.
Current Limit
The CN0565 includes a current limit resistor, which limits the amount of AC
current that enters the sample. To conform to IEC 60601 standards for medical
applications, the maximum allowable AC current is 500 µA at 50 kHz and 600 µA
at 60 kHz. When calculating the RLIMIT resistor value, the maximum output
voltage from the AD5940 is 1.2 Vp-p (0.4243 VRMS). Set the maximum allowable
AC current to 80% of maximum, or 400 µARMS. Calculate the current limit using
Equation 1: As such, a 1 kΩ RLIMIT is selected and is connected to the CE0 pin
on the AD5940. The 0.47 µF CISOx has an impedance of 6.8 Ω at 50 kHz and is
ignored in this calculation.
Isolation Capacitors
The IEC 60601 standards set 10 µA as the maximum DC current allowed to enter
the human body. The DC current of the CN0565 is guaranteed to be zero due to
the addition of isolation capacitors. A value of 0.47 µF is selected for the
isolation capacitors because 0.47 µF is a sufficiently large capacitance that
is also available in small packages suitable for wearable electronics. See the
EDA section of AN-1557 application note for details on capacitor size
calculations. While the AD5940 excitation frequency can be programmed as low
as 0.015 Hz, at low frequencies the impedance of the isolation capacitor is
high enough to attenuate the excitation current. This reduces the signal-to-
noise ratio (SNR) of the impedance measurement, resulting in noisy readings.
Figure 7 shows the impedance of two 0.47 µF in series vs. excitation frequency and can be used to estimate the lowest practical frequency that can be used for a given sample impedance. For example, using a guideline that the drop across the isolation capacitor hould be less than or equal to the drop across the sample, an excitation frequency of 150 Hz will be suitable for impedances as low as 4515 Ω.
High Speed TIA Gain Resistor and ADC Input Range
To achieve an accurate, low-noise impedance measurement, the voltage and
current measurements should exercise a large fraction of the analog-to-digital
converters (ADC) input range. The sense lead voltages are measured directly,
and the ADC input range of 900 mV peak accommodates the 600 mV peak DAC output
voltage.
Note that an infinite impedance (open circuit) would result in the full DAC voltage at the ADC input. The current measurement is scaled by the transimpedance amplifier’s RTIA, which can be set to one of ten values between 50 Ω and 160.1 kΩ. With the excitation set to maximum (600 mV), the value of the minimum impedance can be calculated using Equation 2: ZMIN = RTIA × 2/3 (2) where: ZMIN is the minimum impedance in the force path.
RTIA is the gain resistor value of the transimpedance amplifier. Note that the impedance includes the lead resistance, isolation capacitors, and RLIMIT, as all of these reduce the magnitude of the excitation current. The LTspice simulation shown in Figure 8 is a conceptual model of the AD5940, isolation capacitors, switch resistance, and can be used to estimate an optimal excitation amplitude and RTIA value for any application. Figure 9 shows the TIA output voltage for a given resistor value and excitation frequency.
Crosspoint Switch
The CN0565 includes two ADG2128 8×12 analog crosspoint switches, which allow force and sense electrodes to be assigned arbitrarily. The switch resistances are less than 35 Ω, and has a typical capacitance of 18.5 pF. The switches are powered from a ±5 V bipolar supply, allowing a ±5 V common-mode range of the electrodes with respect to isolated circuit ground.
As shown in Figure 10, the four Y input/output of the two ADG2128 switches are routed to the four analog pins of the AD5940. The first 12 electrode connections (E0 to E11) are connected to the input/output nodes of the first ADG2128, and the second set of 12 electrodes are connected to input/output nodes of the second ADG2128.
INPUT MEASUREMENT
The AD5940 uses a high speed, high precision TIA to convert the current from the electrode into a voltage measured by the ADC at a speed of 800 kSPS. The TIA is internally biased to 1.1 V, centering the output in the ADC’s full scale range. A discrete Fourier transform (DFT) is performed on the ADC data for the current and voltage values, and is configurable up to 16,384 points. The magnitude and phase of the voltage and current are then calculated from the real and imaginary outputs of the DFT in Equation 3 and Equation 4, respectively. The AD5940 then calculates the real and imaginary parts and the unknown impedance is derived using Equation 5:
where: ZUNKNOWN is the unknown impedance of the specimen under test. VMAG is the magnitude of the voltage across the ZUNKNOWN. IMAG is the magnitude of the current through the ZUNKNOWN. rv and iv are the real and imaginary components from the voltage DFT measurements. ri and ii are the real and imaginary components from the current DFT measurements. RTIA is the value of the high speed TIA gain resistor in Ω.
POWER ARCHITECTURE
Figure 11 shows the complete power tree of the CN0565. The host platform board provides 5 V to the ADuM5020 isolated DC-to-DC converter, producing an isolated 5 V, 500 mW output voltage. The isolated 5 V output is used as input to the whole section of the circuit responsible for the actual impedance measurement and processing. The isolated 5 V is fed into the ADM7150 low dropout linear regulator, producing a regulated 3.3 V with 120 mA output. The 3.3 V is used to power the AD5940 as well as provide the digital power supply to the ADG2128 crosspoint switches. The isolated 5 V rail is also used as input to the AD8829 voltage inverter, which produces a dual supply, ± 5 V for the two ADG2128 analog crosspoint switches.
ISOLATION
The entire measurement circuit is galvanically isolated via the ADuM5020 DC- to-DC converter, ADuM1250 I 2 C isolator, and AD-uM3151 7-channel digital isolator for serial peripheral interface (SPI) with aux channels. No explicit stitching capacitance is included, for a total isolated capacitance of 18 pF between grounds. The minimum creepage on the CN0565 is 20 mils due to layout constraints, providing a basic isolation of 150 V. When designing a board to meet a specific regulatory standard, refer to the standard for creepage requirements, and refer to the Safety and Regulation for Digital Isolation for agency certifications.
SYSTEM PERFORMANCE
The fidelity of a tomography image is dependent on several factors including the accuracy of electrode placement, the imaging mode, and fundamentally, the accuracy of the impedance measurements themselves. The 20% typical RTIA tolerance directly affects the measurement and must be calibrated. An external calibration resistor between RCAL0 and RCAL1 pins provides an accurate reference. During calibration, a software routine sets the multiplexer to connect RCAL between the excitation buffer and TIA input, and an excitation signal is applied. Calibration values are measured, calculated, and stored, removing the RTIA error from subsequent impedance measurements. Figure 12 and Figure 13 show the post-calibration measurement error for various resistances and capacitances at a typical excitation frequency of 20 kHz.
IMAGE RECONSTRUCTION ALGORITHMS
From the collected data, the image of the electric impedance distribution may be constructed by use of certain reconstruction algorithms. In EIT, the conductivity distribution of the unknown object is estimated from boundary voltages induced by different current patterns with the aid of various image reconstruction algorithms. The CN0565 example software can perform three image reconstruction algorithms: back-projection (BP), Jacobian calculations (JAC), and Graz consensus reconstruction EIT (GREIT). With back-projection, reconstructed images can be simply understood as a superposition of boundary measurements. JAC algorithm preserves sharp edges, while GREIT uses training data to generate images with more uniform amplitude response. Figure 14 shows a sample EIT image using different reconstruction techniques.
Sensing Area
The CN0565 supports up to 24 electrodes and various impedance measurement
methods are supported. Figure 15 shows the neighboring method, whereby the
current is applied through neighboring electrodes and the voltage is measured
successively from all other adjacent electrode pairs. Lines of equipotential
are shown, and the measured impedance is a weighted average of the red shaded
area.
COMMON VARIATIONS
As the number of electrodes vary depending on the application, variations in array sizes and number of switches are possible using different crosspoint switches. For applications requiring only 8 electrodes, the ADG2188 crosspoint switch is an alternative to the ADG2128. This switch offers eight columns and eight rows in the switch array, providing a total of 64 switch channels. For applications using 10 electrodes, the ADG2108 crosspoint switch can be used. This part has an array size of 8 × 10, providing a total of 80 switch channels.
CIRCUIT EVALUATION AND TEST
This section describes the setup and procedure for evaluating the EVAL- CN0565-ARDZ. For complete information on the circuit evaluation setup, refer to the CN0565 User Guide.
EQUIPMENT NEEDED
- EVAL-CN0565-ARDZ circuit evaluation board
- EVAL-ADICUP3029 development platform
- Host PC (Windows)
- Micro-USB cable
SETUP AND TEST
- Connect the EVAL-CN0565-ARDZ to the EVAL-ADICUP3029 via the Arduino headers.
- Connect the impedance test board to the main EVAL-CN0565ARDZ board via the 30-pin, 100-mil headers.
- Power the boards by connecting the micro-USB cable into P10 on the EVAL-ADICUP3029 and then plug the other end of the cable to the host PC, as shown in Figure 16. a. The DS2 in the EVAL-CN0565-ARDZ must turn green to indicate power.
- Install Python and the pyadi-iio library which includes CN0565 hardware support and example scripts. a. Ensure the version of Python is supported in pyadi-iio.
- Load the CN0565 firmware to the EVAL-ADICUP3029 by copying the pre-built .hex file directly into the DAPLink drive. a. Ensure that the DAPLink is visible in the file explorer to indicate a proper connection between boards. b. Drag and drop the CN0565.hex file to the DAPLink. The DAPLink will automatically disconnect and connect again, indicating a successful upload.
- Open a command terminal and navigate to the working directory where the example script is saved by entering: cd
- Execute the script by typing the following command into the terminal window.
For detailed instructions on how to run the impedance measurement for a single pair, electrode tomography and for the graphical user interface (GUI) demo by using a phantom or a petri dish, see the CN0565 User Guide.
LEARN MORE
CN0565 Design Support Package EVAL-ADICUP3029 User Guide AN-1557. “Implementing the AD5940 and AD8233 in a Full Bioelectric System.” Analog Devices. CN-0510. “Electrochemical Impedance Spectroscopy (EIS) for Batteries.” Analog Devices. Bera, Tushar Kanti. “Bioelectrical Impedance Methods for Noninvasive Health Monitoring: A Review.” 2014. Broeders, Jan- Hein. “Jack of All Trades in Impedance Measurement.” Analog Devices. GitHub. “OpenEIT: Python-based Dashboard for Real-time Electrical Impedance Tomography Including Image Reconstruction Using Back Projection, Graz Consensus and Gauss-Newton methods.” Holder, David and Adler, Andy. “Electrical Impedance Tomography Methods, History and Applications.” 2022. IEEE.
“ Electrical Impedance Tomography for Biomedical Applications
Circuits and Systems Review.” 2021. Liu, Benyuan, et al. “pyEIT: A Python-
based Framework for Electrical Impedance Tomography.” 17th International
Conference on Electrical Impedance Tomography, 2016. Madden, Maria, Andrews,
Penny, and Habashi, Nader. “Comparison of Chest Radiograph to Electrical
Impedance Tomography.” Respiratory Care, 2018. Malmivuo, Jaako.
“Bioelectromagnetism: Principles and Applications of Bioelectric and
Biomagnetism Fields.” Oxford University Press, 1995. Safety and Regulatory
Certification for Digital Isolation. Analog Devices.
DATA SHEETS AND EVALUATION BOARDS
AD5940 Data Sheet AD5940 Evaluation Board ADG2128 Data Sheet ADG2128
Evaluation Board ADuM5020 Data Sheet ADuM5020 Evaluation Board ADM7150 Data
Sheet ADM7150 Evaluation Board ADM8829 Data Sheet ADuM3151 Data Sheet ADuM3151
Evaluation Board
REVISION HISTORY 08/2023—Revision 0: Initial Version
ESD Caution
ESD (electrostatic discharge) sensitive device.
Charged devices and circuit boards can discharge without detection. Although
this product features patented or proprietary protection circuitry, damage may
occur on devices subjected to high energy ESD. Therefore, proper ESD
precautions should be taken to avoid performance degradation or loss of
functionality.
(Continued from first page)
Circuits from the Lab circuits are intended only for use with Analog Devices
products and are the intellectual property of Analog Devices or its licensors.
While you may use the Circuits from the Lab circuits in the design of your
product, no other license is granted by implication or otherwise under any
patents or other intellectual property by application or use of the Circuits
from the Lab circuits. Information furnished by Analog Devices is believed to
be accurate and reliable. However, Circuits from the Lab circuits are supplied
“as is” and without warranties of any kind, express, implied, or statutory
including, but not limited to, any implied warranty of merchantability, non-
infringement or fitness for a particular purpose and no responsibility is
assumed by Analog Devices for their use, nor for any infringements of patents
or other rights of third parties that may result from their use. Analog
Devices reserves the right to change any Circuits from the Lab circuits at any
time without notice but is under no obligation to do so.
©2023 Analog Devices, Inc. All rights reserved. Trademarks and registered trademarks are the property of their respective owners. One Analog Way, Wilmington, MA 01887-2356, U.S.
FAQ’S
Q: What is the frequency range for impedance measurement?
A: The frequency range for impedance measurement is adjustable between
0.015 Hz and 200 kHz.
Q: What standards do bioimpedance measurements need to conform to?
A: Bioimpedance measurements for medical devices must conform to the IEC
60601 standard.