PHONAK Charger RIC I BTE Rechargeable Hearing Aid User Guide
- August 13, 2024
- PHONAK
Table of Contents
Charger RIC I BTE Rechargeable Hearing Aid
“`html
Specifications
- Device: Phonak Charger I
- Compatible hearing aid models:
- Phonak ChargerGo RIC I
- Phonak ChargerGo RIC SPH I
- Phonak Charger RIC I (Non-Rechargeable)
Product Information
1. Parts of the Charger
-
Charger case lid
-
Hearing aid charging slots with left/right indicators
-
Cavity for earpieces
-
Indicator light for internal battery status (only applicable
for rechargeable models) -
USB-C connection to external power supply
-
External power supply
-
USB cable
2. Preparing the Charger
-
Plug the larger end of the charging cable into the power
adapter. -
Plug the smaller end into the USB port on the charger.
-
Plug the power adapter into a power outlet.
-
The indicator light around the USB port is green when the
charger is connected to the power supply.
3. Charging
3.1 Using and Charging the Charger
If the built-in battery is fully charged, the hearing aids can
be charged up to 3 times*. Always make sure that the hearing aids
and charger are dry and clean before charging.
Before using your charger, pull the insulating tab out to
activate the internal battery. After pulling out the insulating
tab, the indicator light will be red for 10 seconds**.
Battery Status
-
Solid green – Fully charged: No action needed.
-
Blinking green – Charging >1 charging cycle: The internal
battery of the charger has enough power to charge your hearing aids
at least once while not connected to a power supply. -
Blinking red – Charging in progress.
FAQ
Q: How many times can I charge my hearing aids with a fully
charged built-in battery?
A: If the built-in battery is fully charged, the hearing aids
can be charged up to 3 times.
Q: What should I do if the indicator light is blinking
green?
A: The blinking green light indicates that the internal battery
of the charger has enough power to charge your hearing aids at
least once while not connected to a power supply.
“`
Phonak Charger I
User Guide
Phonak Charger RIC I Phonak ChargerGo RIC I Phonak ChargerGo RIC SPH I
This user guide is valid for: Charging accessories Rechargeable: Phonak
ChargerGo RIC I Phonak ChargerGo RIC SPH I Non Rechargeable: Phonak Charger
RIC I
2
Your charger details
If no box is checked and you do not know the model of your charging accessory,
please ask your hearing care professional. Please also read safety information
regarding handling the charger (chapter 10). The charger contains a built-in
rechargeable lithium-ion battery*.
Device:
Compatible hearing aid models:
c Phonak Charger RIC I Phonak Audéo I-R c Phonak ChargerGo RIC I Phonak CROS I-R
c Phonak ChargerGo RIC SPH I
Phonak Audéo I-Sphere
- only applicable for rechargeable models
3
Your charger has been developed by Phonak a world leader in hearing
solutions based in Zurich, Switzerland.
This charger is a device intended to charge rechargeable hearing aids.
Please read the user guide carefully to make sure that you understand and get
the best out of your charger.
For more information regarding features, benefits, setup, use and maintenance
or repairs contact your hearing care professional or manufacturer
representative. Training is not required to handle the device. Additional
information can be found in the datasheet of your product.
Phonak – life is on www.phonak.com
Contents
Your charger
1. Parts of the charger
6
Using the charger
2. Preparing the charger
8
3. Charging
9
Further information
4. Environmental conditions
19
5. Care and maintenance
21
6. Service and warranty
23
7. Compliance information
25
8. Information and description of symbols
30
9. Troubleshooting
35
10. Important safety information
36
4
5
1. Parts of the charger
Charger case lid
Hearing aid charging slots with left/right indicators
R
Cavity for earpieces
L
Indicator light for internal
battery status*
*only applicable for rechargeable models
6
USB-C connection to external power supply
External power supply
USB cable
Power supply with wall plug adapter
Power adapter and charger specification
Power adapter output voltage:
5 VDC +/-5%, current 500 mA – 1 A
Power adapter input voltage:
100-240 VAC, 50/60Hz, current 0,2 A
Charger input voltage:
5 VDC +/-5%, current 350 mA
USB cable specifications:
5 V min. 1 A, USB-A to USB-C, maximum length 3 m
7
2. Preparing the charger
Connecting to the power supply
c
d
a
b
a) Plug the larger end of the charging cable into the power adapter.
b) Plug the smaller end into the USB port on the charger. c) Plug the power
adapter into a power outlet. d) The indicator light around the USB port is
green when
the charger is connected to the power supply.
8
3. Charging
3.1 Using and charging the charger If the built-in battery is fully charged,
the hearing aids can be charged up to 3 times*. Always make sure that the
hearing aids and charger are dry and clean before charging.
Before using your charger, pull the insulating tab out to activate the
internal battery. After pulling out the insulating tab, the indicator light
will be red for 10 seconds**.
R /
- For specific information please check the hearing aids User Guide or consult your hearing care professional. **Only applicable for rechargeable models.
9
Rechargeable models have an internal battery, which means they can’t be
switched off. Even when unplugged from a power outlet, it will continue
charging the hearing aids until its internal battery runs out. Once it runs
out, the charger effectively switches off as it can no longer provide power to
charge the hearing aids. Even when the charger’s internal battery is
completely discharged, the charger can still charge the hearing aids once
connected to the power supply. The indicator light on the charger shows the
status of the charger’s internal battery. To switch off non rechargeable
models, unplug the charger from the power outlet.
10
Rechargeable charger connected to a power supply
Indicator light of the charger
Battery status
Description and actions
Solid green Fully charged
The internal battery of the charger is fully charged.
No action needed.
Blinking green
Charging >1 charging cycle
The internal battery of the charger has enough power to charge your hearing aids at least once while not connected to a power supply.
Blinking red
Charging <1 charging cycle
The internal battery of the charger does not have enough power to charge your
hearing aids at least once while not connected to a power supply.
Keep the charger connected to a power supply until it is completely charged.
Solid red
Battery error / charging error
An error was detected while trying to charge the internal battery of the
charger.
Check if your hearing aids still can be charged. Contact your hearing care
professional.
11
Rechargeable charger not connected to a power supply Indicator light turns off after a short time to save battery
Indicator light of the charger
Battery status
Description and actions
Solid green
1 charging cycle left
The internal battery of the charger has enough power to charge your hearing
aids at least once.
No action needed.
Blinking green <1 charging cycle left
No indicator Discharged
light
battery
The internal battery of the charger is running low.
Connect the charger to a power supply.
The internal battery is completely discharged.
Connect the charger to a power supply.
3.2 Charging the hearing aids
Low power: You will hear two beeps when the hearing aids battery is low. You
will have approximately 60 minutes before you have to charge the hearing aids
(this can vary, depending on the hearing aid settings). Before using your
hearing aids for the first time, it is recommended to charge them for 3 hours.
The hearing aids must be dry before charging. Please ensure to charge and use
the hearing aids within operating temperature range: +5° to +40° Celsius (41°
to 104° Fahrenheit).
12
13
Please refer to chapter 2 for how to set up the charger.
1. Insert the hearing aids into the charging slots. The hearing aids will
automatically switch off when inserted into the charger.
2. The indicator light on the hearing aids will show the charging state of their battery until the hearing aids are fully charged. When fully charged, the green indicator light on the hearing aids will remain on.
R L
Make sure the left and right hearing aid markings match the left (L) and right
(R) indicators next to the charging slots.
14
R L
15
The charging process will automatically stop when the batteries are fully
charged, so the hearing aids can be safely left in the charger. It can take up
to 3 hours to charge the hearing aids. The charger lid shall be closed while
charging the hearing aids.
Charging times (representative values)*
Indicator light of the hearing aid
Charging state
Blinking red
0 10%
Blinking orange 11 70%
Blinking green
71 99%
Solid green
100%
Approx. charging times
30 min (20%) 60 min (40%) 90 min (70%)
3 h (charging time may be affected by temperature conditions)
*For specific charging times please check your hearing aids User Guide or consult your hearing care professional.
16
3. Remove the hearing aids from the charging slots by lifting them out of the
charger.
Do not hold the tubes to remove the hearing aids from the charger as this can
damage the tubes.
The hearing aids switch on automatically when removed from the charger. The
indicator light starts blinking. The solid green light indicates the hearing
aids are ready. By default, the hearing aids are programmed so that they turn
on automatically when they are removed from a charger that is connected to the
power supply. Your hearing care professional can turn this setting off. In
this case, you will need to switch the hearing aid on manually.
17
The hearing aids stored in the charger will go on Sleep mode after 15 minutes
if fully charged. Always close the lid when charging your hearing aids in a
humid environment.
18
4. Environmental conditions
Please ensure to use, charge, transport and store the charging accessories according to the following conditions:
Charging Operating
Transport
Storage
Temperature + 5 to + 40 °C
– 20 to + 60 °C – 20 to + 60 °C
(+ 41 to + 104° F) (- 4 to + 140 °F) (- 4 to + 140 °F)
Humidity 0 % to 90 %
(non-condensing)
0 % to 93 % 0 % to 93 %
Atmopheric 500 hPa to
pressure
1060 hPa
500 hPa to 1060 hPa
500 hPa to 1060 hPa
Prolonged storage at temperature lower than +10 °C and higher than +30 °C may negatively affect the battery performance of the product.
19
For information on the operating condition of your hearing aids, please refer
to the user guide of your hearing aids.
If the device has been subject to storage and transportation conditions
outside of the recommended use conditions, wait 15 minutes before starting the
device.
Non rechargeable models are classified as IP21. This means that the chargers
are protected from vertically falling drops of water e.g. condensation. IP21
is satisfied when the charger lid is closed.
Rechargeable models are classified as IP22. This means that the charger is
protected from vertically falling drops of water when the charger is tilted up
to 15 degrees from the horizontal. IP22 is satisfied when the charger lid is
closed.
5. Care and maintenance
Expected service life The charger has the expected service life of five years.
The device is expected to remain safe for use for this duration.
Commercial service period Routine and diligent care of your charger will
contribute to the outstanding performance for the duration of the expected
service life of your devices. Sonova AG will provide a minimum of a five years
repair service period after the respective charger and essential components
have been phased out of the product portfolio.
Please use the following specifications as a guideline. Further information
regarding product safety, see chapter 10.2.
20
21
General information Please make sure that the charger is used and maintained
as described in this user guide.
Daily Make sure that charging slots are clean. Otherwise, clean the surfaces
with a soft, lint-free cloth. Never use cleaning agents such as household
detergents, soap, etc. for cleaning your charger. Do not rinse your charger
with water.
Weekly Remove dust or dirt from the charging slots with a soft, lint-free
cloth.
Always make sure that the hearing aids and charger are dry and clean before
charging.
Always make sure that the charger is disconnected from the power mains before
cleaning.
22
6. Service and warranty
Local warranty Please ask the hearing care professional where you purchased
your charger about the terms of the local warranty. International warranty
Sonova AG offers a one year limited international warranty, valid from the
date of purchase. This limited warranty covers manufacturing and material
defects in the charger. The warranty only comes into force if a proof of
purchase is shown. The international warranty does not affect any legal rights
that you might have under applicable national legislation governing sale of
consumer goods.
23
Warranty limitation This warranty does not cover damage from improper handling
or care, exposure to chemicals or undue stress. Damage caused by third parties
or non-authorized service centers renders the warranty null and void. This
warranty does not include any services performed by a hearing care
professional in their office.
Charger:
Serial number:
Date of purchase:
Authorized hearing care professional (stamp/signature):
7. Compliance information
Declaration of Conformity Hereby Sonova AG declares that this product meets
the requirements of the Medical Devices Regulation (EU) 2017/745 and the UK
Medical Device Regulation 2002.
This device complies with part 15 of the FCC Rules and ICES-003 of Industry
Canada. Operation is subject to the following two conditions:
- This device may not cause harmful interference, and 2) this device must
accept any interference received, including interference that may cause
undesired operation.
Notice: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules and ICES-003 of Industry Canada. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
25
This equipment generates, uses and can radiate radio frequency energy and, if
not installed and used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no guarantee that
interference will not occur in a particular installation.
If this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference by one or more of the
following measures:
· Reorient or relocate the receiving antenna. · Increase the separation
between the equipment and
receiver. · Connect the equipment into an outlet on a circuit
different from that to which the receiver is connected. · Consult the dealer
or an experienced radio/TV
technician for help.
Please keep at least 20 cm separation distance between the charger and the
human body during operation.
26
Electromagnetic emissions
Manufacturer’s Declaration: The device is intended for use in the
electromagnetic environments listed below. The user should assure that they
are used in such environments.
Emissions Test Compliance Electromagnetic environment guidance
RF emissions CISPR 11
Group 1
The device generates or uses RF energy only for their internal functioning. Therefore, the RF emissions are very low and are not likely to cause any interference to nearby electronic equipment.
RF emissions CISPR 11
Class B
The device is suitable for use in domestic establishment and in establishment directly connected to a low voltage power supply network which supplies buildings used for domestic purposes.
27
Electromagnetic immunity
Immunity Test
IEC60601-1-2 Test Level
Electrostatic Discharge (ESD) IEC 61000-4-2
+/- 8 kV contact
+/- 2 kV , +/- 4 kV, +/- 8 kV, +/- 15 kV air
Compliance Level
+/- 8 kV contact +/- 2 kV , +/- 4 kV, +/- 8 kV, +/- 15 kV air
Radiated RF Electromagnetic Field Immunity
Test IEC 61000-4-3
10 V/m 80 MHz 2.7 GHz 80% AM at 1 kHz
10 V/m 80 MHz 2.7 GHz 80% AM at 1 kHz
9 V/m; 710, 745, 780 MHz; 9 V/m; 710, 745, 780 MHz; Pulse modulation 18 Hz Pulse modulation 18 Hz
9 V/m; 5240, 5500, 5785 MHz; Pulse modulation 217 Hz
9 V/m; 5240, 5500, 5785 MHz; Pulse modulation 217 Hz
Proximity Magnetic Field IEC 61000-4-3
27 V/m; 385 MHz; Pulse modulation 18 Hz
28 V/m; 450 MHz FM +/- 5 kHz deviation; 1 kHz sine
27 V/m; 385 MHz; Pulse modulation 18 Hz
28 V/m; 450 MHz FM +/- 5 kHz deviation; 1 kHz sine
28 V/m; 810, 870, 930 MHz; 28 V/m; 810, 870, 930 MHz; Pulse modulation 18 Hz Pulse modulation 18 Hz
28 V/m; 1720, 1845, 1970, 2450 MHz;
Pulse modulation 217 Hz
28 V/m; 1720, 1845, 1970, 2450 MHz;
Pulse modulation 217 Hz
Immunity Test
Rated power frequency magnetic fields IEC 61000-4-8
IEC60601-1-2 Test Level
30 A/m 50 Hz or 60 Hz
Compliance Level
30 A/m 50 Hz or 60 Hz
Test frequency 30kHz,
modulation: CW,
Enclosure
Immunity test level 8
Port
Test frequency
Immunity to
134, 2 kHz,
Proximity Pulse modulation: 2.1 kHz,
Magnetic Immunity test level 65
Fields IEC
61000-4-39 Test frequency 13.56 kHz,
Pulse modulation: 50kHz,
Immunity test level 7.5
Test frequency 30kHz, modulation: CW,
Immunity test level 8
Test frequency 134, 2 kHz,
Pulse modulation: 2.1 kHz, Immunity test level 65
Test frequency 13.56 kHz, Pulse modulation: 50kHz, Immunity test level 7.5
29
8. Information and description of symbols
With the CE symbol, Sonova AG confirms that this product including
accessories meets the requirements of the Medical Devices Regulation MDR
2017/745.
Indicates the medical device manufacturer, as defined in Medical Device
Regulation (EU) 2017/745.
Indicates the date when the medical device was manufactured.
Indicates the Authorized representative in the European Community. The EC REP
is also the importer to the European Union.
This symbol indicates that it is important for the user to read and take into
account the relevant information in these user guides.
30
This symbol indicates that it is important for the user to pay attention to
the relevant warning notices in these user guides.
Important information for handling and product safety.
This symbol confers that the electromagnetic interference from the device is
under limits approved by the US Federal Communications Commission.
Regulatory compliance mark (RCM). Indicates a device’s compliance with
electrical safety legislation applicable to electrical and electronic
equipment in Australia and New Zealand.
With the UK Conformity Assessed mark, Sonova AG confirms that the product is
conform to applicable Great Britain legislations.
31
Indicates the manufacturer’s serial number so that a specific medical device
can be identified.
Indicates the manufacturer’s catalogue number so that the medical device can
be identified.
Indicates that the device is a medical device.
An indication that electronic instructions for use are available.
IIP Rating = Ingress Protection Rating.IP21 rating indicates that the device
is protected against solid foreign objects of 12,5 mm diameter and greater and
is additionally protected against vertically falling water drops..
32
IP Rating = Ingress Protection Rating. IP22 rating indicates that the device
is protected against solid foreign objects of 12,5 mm diameter and greater and
is additionally protected against vertically falling water drops when
enclosure tilted up to 15°.
Indicates the temperature limits to which the medical device can be safely
exposed.
Indicates the range of humidity to which the medical device can be safely
exposed.
Indicates the range of atmospheric pressure to which the medical device can be
safely exposed.
During transportation keep dry.
33
This symbol indicates a device that is not safe to use in magnetic resonance
(MR) environment (e.g. during an MRI scan).
The symbol with the crossed-out garbage bin is to make you aware that this
charger may not be thrown away as normal household waste. Please dispose of
old or unused charger, at waste disposal sites intended for electronic waste,
or give your charger to your hearing care professional for disposal. Proper
disposal protects the environment and health.
Symbols applicable to the European power supply only
Power supply with double isolation.
Device designed for indoor use only.
Safety isolating transformer, short circuit proof.
34
9. Troubleshooting
Cause
Possible remedy
Issue: Charging of the hearing aid does not start.
Internal battery of the charger discharged*.
Connect the charger to the power supply (chapter 2).
Hearing aids not inserted Insert hearing aids correctly into the correctly into charger. charger (chapter 3).
*Only applicable for rechargeable models.
Please check https://www.phonak.com/com/en/support.html for further
information.
If the problem persists, contact your hearing care professional for
assistance.
35
10. Important safety information
Please read the relevant safety information and limitationsof-use information
on the following pages before using your device.
Intended use The chargers are intended to charge the battery of the
rechargeable hearing aid.
Intended user Intended for people with hearing loss using the compatible
rechargeable hearing aid, their caregivers and hearing care professional.
Intended patient population The intended patient population for the chargers
is the intended patient population of the compatible hearing aids.
Intended medical indication Indications are not derived from charger but from
the compatible hearing aids.
Intended medical contra-indications Contraindications are not derived from the
charging accessories, but from the compatible hearing aids.
Clinical benefit The charger itself does not provide a direct clinical
benefit. An indirect clinical benefit is provided by enabling the parent
device (rechargeable hearing aids) to achieve their intended purpose.
Side effects Side effects are not derived from the charger, but from the
compatible hearing aids.
Any serious incident that has occurred in relation to the charging
accessories, should be reported to the manufacturer representative and the
competent authority of the state of residence. The serious incident is
described as any incident that directly or indirectly led, might have led or
might lead to any of the following:
36
37
a) the death of a patient, user or other person b) the temporary or permanent
serious deterioration of a patient’s, user’s or other person’s state of health
c) a serious public health threat
To report an unexpected operation or event, please contact the manufacturer or
a representative.
10.1 Warnings
Charger and power supply must be protected from shocks. If the charger or
power supply is damaged after a shock, the devices must not be used anymore.
Protect your charging accessories from heat and sunlight (never leave near a
window or in the car). Never use a microwave or other heating devices to dry
your hearing aid(due to risk of fire or explosion). Ask your hearing care
professional about suitable drying methods.
Only authorized personnel are allowed to open the charging accessory due to
possible electrical hazard.
Cables and cords wrapped around the neck may cause strangulation. Do not leave
the device and its components unattended with children, individuals with
cognitive impairment or pets.
38
39
Do not use the device in explosive areas (mines or industrial areas with
danger of explosions, oxygenrich environments or areas where flammable
anesthetics are handled).The device is not ATEX certified.
Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to verify
that they are operating normally.
Do not place these rechargeable devices in your checked-in luggage when
traveling by plane as they contain lithium-ion batteries. The devices should
be placed in your carry-on/hand luggage.
Ship the rechargeable devices according to local regulations as the devices
contain lithium-ion batteries. For shipping, they are classified as dangerous
goods. If in doubt ask the courier responsible for shipping how to ship the
devices properly.
*Only applicable for rechargeable models
40
Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation
To supply power to the charging accessories, use only IEC 60950-1*, IEC
62368-1 or IEC60601-1 certified charging equipment (e.g. power supply) within
the specification in Chapter 1 Parts of the charger on page 7.
Do not place the charging accessories close to an induction cooking surface.
Conductive structures inside the charging accessories could absorb inductive
energy resulting in thermal destruction.
Always make sure that the hearing aids and charger are dry and clean. If the
charger or power supply is damaged, the device must not be used anymore as it
may cause electric shock.
- Note: this standard may not be valid anymore in your country, please refer to IEC 62368-1
41
The charger contains Lithium-Ion rechargeable battery that might swell and
deform the housing of the charger. If you notice any swelling or deformation
in the charger, stop using it immediately. Swollen batteries can be a serious
safety hazard, posing risks such as leakage, overheating, or even explosion.
Do not attempt to remove, charge, puncture or modify the swollen battery. Do
not ignore signs of battery damage. Using the charger with a damaged battery
can affect the charger performance and permanently damage the charger. Avoid
handling the charger with bare hands if it shows signs of damage or leakage.
Avoid direct contact with the leaked electrolyte. If contact occurs, rinse
with water and seek medical attention if needed. If you have any questions or
concerns, seek professional guidance at your hearing care professional before
using the charger. Dispose of the charger at waste disposal sites intended for
electronic waste, or give your charger to your hearing care professional for
disposal.
Only applicable for rechargeable models
42
The following is only applicable for persons with active implantable medical
devices (i.e. pacemakers, defibrillators, etc.): · Keep the charging device at
least 15 cm (6 inches) away from the active implant. If you experience any
interference do not use the charging device and contact the manufacturer of
the active implant. Please, note that the interference can also be caused by
powerlines, electrostatic discharge, airport metal detectors etc.
Do not touch the contacts of the charger when powered.
Do not expose your device to temperatures outside the operation conditions
indicated in the User Guide (see chapter 4). They contain lithium-ion
batteries, which might swell, leak or explode upon exposure to temperatures
outside those conditions.
Only applicable for rechargeable models
43
Do not use your hearing aid or charger in areas where electronic equipment is
prohibited.
Only use approved chargers described in this user guide for charging your
hearing aids, otherwise the devices may be damaged.
Changes or modification to the device that were not explicitely approved by
Sonova AG are not permitted. Such changes may damage the device.
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches)
to any part of the hearing aid or accessory, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment
could result.
10.2 Cautions
Only use the device according to chapter 4 Environmental conditions page 18.
Using the device otherwise may heat the hearing aid up to temperatures, which
in the worst-case can lead to skin burns.
10.3 Notes
Do not drop your charging accessories! Dropping onto a hard surface can damage
your charging accessories.
Always keep the USB socket of the device and the USB plug of the cable clean
and dry. Foreign particles or moisture in the USB connection can lead to an
electrical fault, which can heat up the USB connection and, in the worst case,
cause it to melt. Use the provided cable to supply power to the device. If you
notice the indicator light around the USB port of the device went off during
charging, check the USB connection for proper conditions.
44
45
The built-in battery of the rechargeable charger must be recharged every 3 months. This will prevent a deep discharge of the built-in battery. Deep discharged batteries may be damaged if recharged and could catch fire.
Notes
46
47
029-1382-02/V1.00/2024-04/TM © 2024 Sonova AG All rights reserved
Your hearing care professional:
Sonova Deutschland GmbH Max-Eyth-Strasse 20 70736 Fellbach-Oeffingen Germany
Manufacturer: Sonova AG Laubisrütistrasse 28 CH-8712 Stäfa Switzerland
www.phonak.com
Australian Sponsor: Sonova Australia Pty Ltd 12 Inglewood Place, Norwest NSW
2153 Australia
UK Responsible person and UK Importer: Sonova UK Limited Sonova House Lakeside
Drive, Centre Park WA1 1RX Warrington United Kingdom
613389 759396
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>