Attogene AU2048 Staphylococcus Enterotoxin B SEB Lateral Flow Detection
Kit Instructions
Introduction ****
Attogene’s SEB detection kit is a ready to use, convenient, and simple way to test food or water samples for Staphylococcal Enterotoxin B. Being one of the most frequent causes of food poisoning, it can be important to test food or water samples to prevent the spread of contaminants. SEB can also be used as a biological weapon in the form of an aerosol highlighting the importance of a simple ready to use kit. Using our kit, which employs the usage of an Anti-SEB Toxin Antibody covalently bound to colloidal gold, you will be able to accurately be able to detect extremely minute quantities of SEB antigen down to 20ppb.
Features & Benefits
- Can be used for development of a lateral flow assay for detection of SEB
- No expensive equipment
Cost-effective way to screen for further downstream lateral flow assay development
Kit Contents
Component Name | Volume | Storage |
---|---|---|
lateral flow dipsticks | 50 | RT |
Assay Running Buffer | 10 mL | RT |
96 well plate | 1 each | RT |
Manual | 1 each | RT |
Storage and Stability ****
- The kit should be stored at 2℃ – 30℃ until ready to
- The test must remain in the sealed pouch until
Required Materials Not Supplied
- Samples (contrived sample, human sera, or plasma).
- Centrifuge – For preparation of clear
- Pipettor and pipette tips – to transfer samples and
- Tubes or microtiter plates to run the
Precautions
- Prior to use, ensure that the product has not expired by verifying that the date of use is prior to the expiration date on the
- The test strips are packaged in a vial that contains
- Avoid cross-contamination of samples by using a new tube and disposable pipette tip for each
- Use only Lateral Flow Kit reagents from one kit lot, as they have been adjusted in
- It is good laboratory practice to use positive and negative controls to ensure proper test
- Due to the hook effects, if no signal is detected in the test line, a serial dilution may be necessary to bring the antigen into the appropriate concentration ratio/stoichiometry.
Procedure
Prepare serially diluted antigen in phosphate buffered saline (i.e., 1mg/ml to 1ug/mL). Perform the following:
- Add 150μL of Lateral Flow Running Buffer into a well of a 96 well
- Always run a positive and negative control well with sample:
- (SAMPLE) mix a designated amount (a volume 1μL-10μL are good starting points) of sample which potentially contains SEB toxin into the
- Mix each well completely by pipetting up and down several
- Add your dip stick to the
- Incubate for 10-15
- Visually analyze the strip by eye, photography, or read in a lateral flow
NOTE: If a higher analytical sensitivity is required, it could be helpful to increase the volume and/or concentration of the sample added into the well. Volume and concentration of analyte-specific solu- tions, and incubation time are always part of the individual test development.
Interpretation of Results
This test is a lateral flow assay containing test lines that are dependent on the concentration of antigen in the sample. Adjustment of sample may be required to understand the dynamic range of the test.
What to expect at the test lines:
The higher the concentration of SEB antigen will facilitate greater intensity of the test line compared to the strip run in a mixture lacking sample (negative control strip).
What to expect at the control line:
The intensity of the control will not decrease as the test lines increase.
Determination made using strips which have dried for more or less than the
required time may be inaccurate, as line intensities may vary with drying time