Sonel CMP-402 Multimeter Clamp Meter Universal with LCD Display User Manual

July 24, 2024
Sonel

Sonel CMP-402 Multimeter Clamp Meter Universal with LCD Display

Sonel-CMP-402-Multimeter-Clamp-Meter-Universal-with-LCD-Display-
image

Product Specifications

  • Model: CMP-402 / CMP-403
  • Version: 1.04
  • Date: 13.05.2024
  • Measurement Functions: Voltage (AC/DC), Current (AC/DC), Resistance, Capacitance, Frequency, Duty Cycle, Temperature, Diode Test, Continuity

Product Usage Instructions

1. Preparation of the Meter

Before use, ensure the meter is turned off. Insert the required probes into the corresponding measurement sockets.

2. Functional Description

  • Measurement Sockets: Connect probes to appropriate sockets for different measurements.
  • Display: View measurement results and additional information on the display.
  • Probes: Use appropriate probes for accurate measurements.

3. Special Functions

  • REL Button: Allows relative measurements.
  • RANGE Button: Adjusts measurement range.
  • MODE/VFD Button: Changes measurement mode and activates VFD function.
  • PEAK/INRUSH Button: Enables peak/max/min and inrush current functions.
  • H Button: Activates hold and flashlight functions.
  • Auto Power Off: Automatically turns off the device after a period of inactivity.

4. Battery Replacement

When the battery is low, open the battery compartment and replace with a new battery following the polarity markings.

5. Maintenance and Care

Keep the meter clean and dry. Avoid exposure to extreme conditions or liquids. Regularly calibrate the meter for accurate readings.

6. Storage

Store the meter in a cool, dry place away from direct sunlight and moisture. Protect it from dust and physical damage.

Frequently Asked Questions

Q: What is the maximum input value for different functions?

A:

  • A DC/AC: 400 A DC/AC
  • V DC/AC, Frequency, Duty Cycle: 1000 V DC/AC RMS
  • Resistance, Continuity, Diode Test, Capacitance, Temperature:
    300 V DC/AC RMS

Q: How should I handle safety precautions while using the product?

A: Ensure that only qualified personnel handle the product.
Follow specified safety precautions for different measurement conditions and input signal limits.

The icon with the meter name is placed next to sections of the text that refer to specific features of the device. All other parts of the text relate to all types of the instrument.

Introduction

Thank you for purchasing Sonel multimeter. CMP-402 / 403 meter is a modern, easy and safe measuring device. Please acquaint yourself with this manual in order to avoid measuring errors and prevent possible problems in operation of the meter.

This manual contains three types of warnings. They are presented as a framed text describing the possible risks for the user

and the device. Texts WARNING describe situations, which may endanger user’s life or health, when instructions are not fol-

lowed. Texts

CAUTION! begin a description of a situation,

which may result in device damage, when instructions are not fol-

lowed. Indication of possible problems is preceded by symbol .

WARNING · CMP-402 / 403 meter is designed to measure the AC/DC
current and voltage, frequency, resistance, capacitance, as well as to test the circuit continuity and diodes. Any application that differs from those specified in the present manual may result in a damage to the device and constitute a source of danger for the user. · CMP-402 / 403 meter must be operated only by appropriately qualified personnel with relevant certificates authorising the personnel to perform works on electric systems. Unauthorized use of the meter may result in its damage and may be a source of serious hazard to the user. · Before operating the device, read thoroughly this manual and observe the safety regulations and guidelines provided by the producer. Failure to follow instructions specified in this manual may result in a damage to the device and be a source of serious hazard to the user.

38

CMP-402 CMP-403 ­ USER MANUAL

Safety

2.1 General rules
In order to provide conditions for correct operation and the correctness of the obtained results, the following recommendations must be observed: · before using the meter read carefully this manual, · the meter should be operated only by qualified persons that have
passed health and safety training, · be very careful when measuring voltages exceeding (as per IEC
61010-1:2010/AMD1:2016):

Normal locations 60 V DC 30 V AC RMS 42.4 V AC of peak value

Wet locations 35 V DC 16 V RMS 22.6 V AC of peak value

as they generate a potential risk of electric shock, · do not exceed the maximum limits of the input signal, · during the voltage measurements do not switch the device in the
current or resistance measuring mode and vice versa, · when changing ranges, always disconnect the test leads from the
tested circuit, · hold the measuring probes by the spot provided, restricted by a spe-
cial barrier to avoid accidental contact with exposed metal parts, · If during the measurement symbol OL appears on the screen, it indi-
cates that the measured value exceeds the measurement range, · It is unacceptable to operate:
a damaged meter which is completely or partially out of order, a device with damaged insulation of test leads, a meter stored for an excessive period of time in disadvan-
tageous conditions (e.g. excessive humidity). · repairs may be carried out only by an authorised service point.

CMP-402 CMP-403 ­ USER MANUAL

39

WARNING
· Never start the measurements if you have wet or damp hands.
· Do not perform measurements in explosive atmosphere (e.g. in the presence of flammable gases, vapours, dusts, etc.). Using the meter in such conditions may result in sparking and cause an explosion.

The limit values of the input signal

Function

The maximum input value

A DC (

), A AC

400 A DC/AC

V DC, V AC, voltage frequency, duty cycle

1000 V DC/AC RMS

Resistance, continuity, diode test, capacitance, temperature

300 V DC/AC RMS

2.2 Safety symbols This symbol located near another symbol or terminal, indicates that the user should read the further information contained in the manual.
This symbol located near the terminal, indicates that in normal use there is a possibility of dangerous voltages.
Protection class II ­ double insulation
Terminals with this marking cannot be connected to a circuit where the voltage to ground exceeds the maximum safe voltage of the device.

40

CMP-402 CMP-403 ­ USER MANUAL

Preparing the meter for operation

After purchasing the meter, check whether the content of the package is complete.
Before performing the measurement: · make sure that the battery level is sufficient for measurements, · check whether the meter casing and insulation of the test leads
are not damaged, · to ensure consistent measurement results it is recommended to con-
nect black lead to COM terminal and red lead to other terminals, · when the meter is not in use, set the function switch in OFF position.
The device has the AUTO-OFF function triggered after 15 minutes of user inactivity. To turn the meter on again, set the function switch to OFF position and then set it at the desired function.
WARNING
· Connecting wrong or damaged leads may cause electric shock.
· The meter must not be connected to the voltage source when it is set to current or resistance measurement or to diode test. Failure to observe this precaution may damage the meter!
When using the meter, be sure to: · discharge capacitors in the tested power sources, · disconnect the power supply when measuring the resistance and
diode tests, · turn off the meter and disconnect test leads before removing the
back cover to replace the batteries.

CMP-402 CMP-403 ­ USER MANUAL

41

WARNING
Do not use the meter if the cover of battery compartment is removed.
It is possible that in certain low ranges of AC or DC voltage, when the meter is not connected to the leads, the screen will show random and variable readings. This is a normal phenomenon, which results from the input sensitivity with high input resistance. When connected to a circuit, the read- out will stabilize and the meter will provide the correct value.

42

CMP-402 CMP-403 ­ USER MANUAL

Functional description

4.1 Measuring terminals and functions

CMP-402 CMP-403 ­ USER MANUAL

43

Non-contact voltage detector

Current clamp

Flashlight

Indicator light of the non-contact voltage detector

Button H /
· HOLD mode ­ freezing the measurement results on the display (press briefly)
· flashlight mode (press and hold)
Clamp-opening trigger

Rotary switch

Function selection:

·

400A ~ ­ measurement of alternating current up to 400 A

·

400A ­ measurement of direct and alternating current

up to 400 A

·

40A ~ ­ measurement of alternating current up to 40 A

·

40A ­ measurement of direct and alternating current

up to 40 A

· Temp ºC ºF ­ temperature measurement

· CAP ­ measurement of resistance, capacitance

·

­ measurement of continuity, diode test

· ­ measurement of direct voltage

· Hz% VFD ­ measurement of alternating voltage, measurement of frequency and duty cycle, measurement of current and voltage behind the inverter, frequency converter, in the VFD system
· OFF ­ the meter is switched off

44

CMP-402 CMP-403 ­ USER MANUAL

Button REL
o REL mode ­ press briefly:
Reset display (DC current measurement) Displaying the measurement result related to the reference
value (other measuring functions)
o Activating the backlight of the display (press and hold)
LCD display
Function buttons
· RANGE button Ustawianie zakresu pomiarowego:
o automatic (press and hold) o manual (press briefly)
· MODE / VFD button The choice of sub-functions and modes assigned to the selected measurement function
o Changing the measurement mode in functions: A / temperature
measurement / resistance / capacitance / continuity / diode test / V / frequency / duty cycle (press briefly)
o Measurement of current and voltage behind the inverter, fre-
quency converter, in the VFD system (press and hold)
· PEAK / INRUSH button
o Displays peak value of the measured signal (press briefly) o Displays the starting current (press briefly)
COM measuring terminal
Measuring input, common for all measuring functions excluding current.

Measurement terminal VCAP

Hz%Temp

Measuring input for measurements other than current measurement.

CMP-402 CMP-403 ­ USER MANUAL

45

4.2 Display

AUTO H
VFD APO INRUSH
P MAX / MIN
°C / °F

Automatic range setting HOLD function activated Measurement behind the inverter, frequency converter, in the VFD system Auto-off mode Inrush current Peak value Maximum / Minimum value Diode test Continuity test Temperature measurement in Celsius / Fahrenheit degrees Relative measurement
Alternating signal Constant signal

Low battery

n / µ / m / k / M The prefix of multiple measurement unit

V

Voltage measurement

A

Current measurement

F

Measurement of capacitance

Measurement of resistance

Hz

Measurement of frequency

%

Duty cycle measurement

­

Negative read-out value

OL

Exceeded measurement range

46

CMP-402 CMP-403 ­ USER MANUAL

4.3 Leads The manufacturer guarantees the correctness of read-outs only
when original test leads are used.
WARNING Connecting wrong leads may cause electric shock or measurement errors.
· The probes are equipped with additional removable tip guards.
· The probes must be stored in a designated area.

CMP-402 CMP-403 ­ USER MANUAL

47

Measurements

The content of this chapter should be thoroughly read and understood since it describes methods of measurements and basic principles of interpreting measurement results.
5.1 Current measurement
WARNING Disconnect the test leads before measuring current using clamp.

To perform the current measurement: · set the rotary switch at:

40A ~ / 400A ~,

40A / 400A ,

· press MODE/VFD button to display the following symbol: , if you are measuring alternating current, , if you are measuring direct current,

· use the clamp-opening trigger and attach the clamps on the

tested conduit. Only one conduit must be within the testing range of the clamps, · read the measurement result on the display.

If DC current is measured and the meter is not attached to the tested circuit, but it still indicates a non-zero value, then you must reset it by pressing REL button.

48

CMP-402 CMP-403 ­ USER MANUAL

5.2 Non-contact voltage detector
WARNING
· The detector is designed to detect the presence of a voltage, not for determining its absence.
· Electric shock hazard. Before using the tester, check if its operational by testing it on a known AC voltage (i.e. next applicable socket with live voltages).
To activate the detector: · set the rotary switch at any position, · touch the tip of the detector to the tested object. If the AC voltage is present, the indicator light will glow red.
· The wires in the extension cords are often twisted. For best results, move the tip of the detector along the wire to locate the live line.
· The indicator has a high sensitivity. It can be randomly actuated by static electricity or other energy sources. This is normal.
· The type and thickness of the insulation, distance from the power source, shielded cables and other factors may affect the operation of the tester. If you are unsure about the test result, check the presence of voltage in a different way.

CMP-402 CMP-403 ­ USER MANUAL

49

5.3 Voltage measurement

WARNING · Electric shock hazard. The ends of measuring
probes, due to their length, may not reach the live parts inside some network connections of lowvoltage electrical equipment, because the contacts are arranged inside the sockets. In such a case, the read-out will be 0 V with the simultaneous presence of voltage in the socket. · Before acknowledging the absence of voltage in the socket make sure that the ends of the of the probe touch the metal contacts inside the socket.

CAUTION! Do not measure the voltage when an electric motor located within the circuit is being switched on or off. Resulting voltage spikes may damage the meter.

To perform AC voltage measurement:

· set the rotary switch at ing voltage),

(direct voltage) or Hz% VFD (alternat-

· connect black test lead to COM terminal, and red test lead to

VCAP

Hz%Temp terminal,

· contact the tips of test probes to the points of measurement,

· read the measurement result on the display.

5.4 Frequency measurement

To perform frequency measurement:

· set the rotary switch at Hz% VFD,

· press MODE/VFD button to display Hz on the screen,

· connect black test lead to COM terminal, and red test lead to

VCAP

Hz%Temp terminal,

· contact the tips of test probes to the points of measurement,

· read the measurement result on the display.

50

CMP-402 CMP-403 ­ USER MANUAL

5.5 Measurement % of duty cycle (pulse filling indicator)

To perform the measurement:

· set the rotary switch at Hz% VFD,

· press MODE/VFD button, until symbol % is shown on the display,

· connect black test lead to COM terminal, and red test lead to

VCAP

Hz%Temp terminal,

· contact the tips of test probes to the points of measurement,

· read the measurement result on the display.

5.6 Measurement of resistance
WARNING
Do not perform measurements on the circuit under the voltage. Before the measurement disconnect the power and discharge capacitors.

To perform measurement of resistance:

· set the rotary switch at CAP,

· connect black test lead to COM terminal, and red test lead to

VCAP

Hz%Temp terminal,

· contact the tips of test probes to the points of measurement; the

best solution is to disconnect one side of the tested element, to

prevent the remaining part of the circuit interfere with the read-out

of the resistance value,

· read the measurement result on the display.

CMP-402 CMP-403 ­ USER MANUAL

51

5.7 Circuit continuity test
WARNING
Do not perform measurements on the circuit under the voltage. Before the measurement disconnect the power and discharge capacitors.

To perform the continuity test:

· set the rotary switch at

,

· connect black test lead to COM terminal, and red test lead to

VCAP

Hz%Temp terminal,

· contact the tips of test probes to the points of measurement,

· read the measurement result on the display; the beep will be acti-

vated when resistance values are below approx. 50 .

5.8 Diode test
WARNING
Do not perform measurements on the circuit under the voltage. Before the measurement disconnect the power and discharge capacitors. Do not test the diode under voltage.

To perform the diode test:

· set the rotary switch at

,

· press MODE/VFD button, to display V on the screen,

· connect black test lead to COM terminal, and red test lead to

VCAP

Hz%Temp terminal,

· contact the tips of test probes to the diode. The red test probe

should contact the anode and the black should contact cathode,

· read the test result on the display ­ the forward voltage is displayed.

For a typical silicon rectifier diode, it is approx. 0.7 V, and for

a germanium diode it is approx. 0.3 V

For LEDs with a low power, typical voltage value is in the

range of 1.2…5.0 V depending on the colour.

52

CMP-402 CMP-403 ­ USER MANUAL

If the diode is polarized in the reverse direction, or there is a break in the circuit, the display will show OL.
When the diode is shorted, the meter will show a value near 0 V, · after completing the measurements, remove test leads from the
terminals of the meter.

5.9 Measurement of capacitance
WARNING
Risk of electric shock. Disconnect the power supply from the tested capacitor and discharge all capacitors before any starting capacity measurements.

To perform the measurement:

· set the rotary switch at CAP, · press MODE/VFD button to display nF on the screen,

· connect black test lead to COM terminal, and red test lead to

VCAP

Hz%Temp terminal,

· contact the probe tips to the tested capacitor,

· read the measurement result on the display.

CMP-402 CMP-403 ­ USER MANUAL

53

5.10 Temperature measurement

To perform the measurement:

· set the rotary switch at Temp ºC ºF, · to change the unit, press MODE/VFD,

· place the adapter of the temperature probe in COM terminal

(black leg) and VCAP

Hz%Temp (red leg):

· place the temperature probe in the adapter, as shown in the

figure:

thin pin of the probe (marked as +) fits to terminal +;

thick pin of the probe (marked as K) fits to terminal ­;

reversed connection of the probe is mechanically

impossible,

· contact the head of the temperature probe to the device under

test. Maintain the contact of the probe head with the part of the

device under test, until the reading stabilizes.

· read the measurement result on the display,

· after completing the measurements, disconnect the probe from

the meter.

CAUTION!
Risk of burns. The temperature probe heats up, adapting to the temperature of tested object.

54

CMP-402 CMP-403 ­ USER MANUAL

Special features

6.1 Button REL
6.1.1 REL function This mode enables a measurement relative to a reference value. · To enable the mode, press briefly REL . Then, the displayed
readout value is taken as the reference value, and the readout will be reset. · From this moment, the readings will be presented as the ratio of the measured value to the reference value. · To disable the mode, press REL .
The displayed main result is the difference between the reference value (read- out at the moment of activating REL mode) and the current read-out. Example: if the reference value is 20 A, and the current reading is 12.5 A, then the main result on the display will be -7.5 A. If the new reading is identical to the reference value, then the result will be zero.
· When the function is activated, the automatic adjustment of the measuring range is not available.
· If the reading is outside the measurement range, symbol OL is displayed In this situation, turn off the function and manually switch to a higher range.
· This function is not available for diode test, continuity test and duty cycle.
6.1.2 Display backlight
Pressing and holding REL button for 2 seconds will turn ON/OFF the display backlight function.

CMP-402 CMP-403 ­ USER MANUAL

55

6.2 RANGE button The button is used to set the measuring range.
To enable the automatic function, press and hold RANGE button longer than 1 second.
To manually toggle through the measuring ranges, press RANGE button.
6.3 Button MODE/VFD
6.3.1 Changing the measurement mode Press briefly MODE/VFD button to switch between the available
measurement modes.
6.3.2 VFD function To measure the AC voltage behind the inverter, frequency con-
verter or in the VFD system: · set the rotary switch to the voltage or current measurement position, · press and hold the MODE/VFD button until the “VFD” symbol appears.
To disable the mode, press and hold the MODE/VFD button.

56

CMP-402 CMP-403 ­ USER MANUAL

6.4 PEAK/INRUSH button
6.4.1 PEAK MAX/PEAK MIN function PEAK function allows user to record very short alternating volt-
age surges. The meter will update the display each time a lower negative, or
higher positive peak occurs. Auto Power Off feature will be disabled automatically in this mode.
To activate the mode, press briefly the PEAK/INRUSH button.
To disable the mode, press and hold the PEAK/INRUSH button.
· This function is available only when measuring AC voltage. · While PEAK is active, autoranging is disabled, therfore it is advised to
start the function after connecting test leads to the measurement point. Running PEAK before that may cause overrange symbols to appear.
6.4.2 INRUSH function The INRUSH function captures the starting current precisely in
the beginning of 100-millisecond period when the device is just started. To perform the measurement: · activate the AC measurement, · press briefly PEAK/INRUSH button, · fasten the clamp on the cord supplying power to the tested object, · turn ON the tested object, · read the results.
To disable the mode, press and hold the PEAK/INRUSH button.
· This function is available only when measuring AC current. · While INRUSH is active, autoranging is disabled, therfore it is advised
to start the function after connecting test leads to the measurement point. Running INRUSH before that may cause overrange symbols to appear.

CMP-402 CMP-403 ­ USER MANUAL

57

6.5 Button H

6.5.1 HOLD function

This function is used to ‘freeze’ the measurement result on the

display. To do this, briefly press H button. When the function is

enabled, the display shows symbol H.

To return to the normal operation mode of the device, press

H

button again.

6.5.2 Flashlight function Briefly press H , to turn on or off the flashlight mode.

6.6 Auto-Off
The meter automatically shuts off after 15 minutes of user inactivity. Symbol APO in the display indicates activated function.
Auto-off function may be temporarily disabled. For this purpose: · set the rotary switch at OFF position, · press and hold MODE/VFD button, · set the rotary switch at the desired measuring function, · wait until the meter reaches the measurement readiness, · release MODE/VFD button. When the automatic shutdown is de-
activated, the display does not show APO.
Each pass of the rotary switch through “OFF” position with non-pressed MODE/VFD button, will activate again the Auto-Off function.

58

CMP-402 CMP-403 ­ USER MANUAL

Replacing the batteries

WARNING To avoid electric shock, do not use the meter if the battery compartment cover is not in place or is not properly fastened.
CMP-402 / 403 is powered by by three LR03 AAA 1.5 V batteries. It is recommended to use alkaline batteries.
To replace the batteries: · set the rotational function selector at OFF, · remove test leads from the terminals of the meter. · turn the fixing screw of the compartment cover to the position:
· remove the cover, · remove the batteries and insert a new ones, observing the polarity, · put on the cover and turn the fixing screw to the position:

· While performing the measurements with the low battery symbol displayed, the user must be aware of additional measurement uncertainties or unstable operation of the device.
· If the meter does not work properly, check the batteries in order to ensure that they are in proper condition and properly installed in the device.

CMP-402 CMP-403 ­ USER MANUAL

59

Maintenance and care

The digital multimeter has been designed for many years of reliable use, provided that the following recommendations are observed for its maintenance and care:
1. THE METER MUST BE DRY. Wipe the dampened meter.
2. THE METER MUST BE USED AND STORED IN NORMAL TEMPERATURES. Extreme temperatures may shorten the life of electronic components and distort or melt plastic parts.
3. THE METER MUST BE HANDLED CAREFULLY AND GENTLY. Dropping the meter may damage its electronic elements or the housing.
4. THE METER MUST BE KEPT CLEAN. From time to time wipe the housing with a damp cloth. DO NOT use chemicals, solvents or detergents.
5. USE ONLY NEW BATTERIES OF RECOMMENDED SIZE AND TYPE. Remove the old or discharged batteries from the meter to avoid leakage and damage.
6. IF THE METER IS TO BE STORED FOR LONGER THAN 60 DAYS, remove the batteries and keep them separately.
The electronic system of the meter does not require maintenance.

60

CMP-402 CMP-403 ­ USER MANUAL

9 Storage
During the storage of the device, the following recommendations must be observed: · disconnect the test leads from the meter, · make sure that the meter and accessories are dry, · when the device is to be stored for longer time, remove the batteries.
10 Dismantling and disposal
Worn-out electric and electronic equipment should be gathered selectively, i.e. it must not be placed with waste of another kind.
Worn-out electronic equipment should be sent to a collection point in accordance with the law of waste electrical and electronic equipment.
Before the equipment is sent to a collection point, do not dismantle any elements.
Observe local regulations concerning disposal of packages, waste batteries and accumulators.

CMP-402 CMP-403 ­ USER MANUAL

61

Technical data

11.1 Basic data “m.v.” means a standard measured value.
True RMS measurement for AC current

Range

Resolution

Accuracy

40.00 A

0.01 A

(2.0% m.v. + 8 digits)

400.0 A

0.1 A

(2.5% m.v. + 8 digits)

· All AC current ranges are specified from 5% to 100% of range

· Frequency range: 50 Hz…60 Hz · Overload protection: 400 A

DC current measurement

Range

Resolution

40.00 A

0.01 A

400.0 A

0.1 A

· Overload protection: 400 A

Accuracy (2.0% m.v. + 8 digits) (2.5% m.v. + 8 digits)

True RMS voltage and VFD measurement
Accuracy Range Resolution for f = 50 Hz…60 Hz
(all waveforms)

Accuracy
for f = 50 Hz…1 kHz (sine waveforms)

4.000 V 0.001 V

40.00 V

0.01 V

(1.2% m.v. + 5 digits) (1.2% m.v. + 5 digits)

400.0 V

0.1 V

1000 V

1 V

(1.5% m.v. + 5 digits) (1.5% m.v. + 5 digits)

· All AC voltage ranges are specified from 5% to 100% of range

· Input impedance:

9,5 M,

9 M

· Frequency range: 50 Hz…1000 Hz

· Overload protection: 1000 V DC/AC RMS

· AC voltage range for VFD function: 100 V…600 V

62

CMP-402 CMP-403 ­ USER MANUAL

DC voltage measurement

Range

Resolution

Accuracy

4.000 V 40.00 V

0.001 V 0.01 V

(1.0% m.v. + 3 digits)

400.0 V

0.1 V

1000 V

1 V

(1.2% m.v. + 5 digits)

· Input impedance: 10 M · Overload protection: 1000 V DC/AC RMS

Resistance measurement

Range

Resolution

Accuracy

400.0

0.1

(1.0% m.v. + 4 digits)

4.000 k

0.001 k

40.00 k

0.01 k

(1.5% m.v. + 2 digits)

400.0 k

0.1 k

4.000 M

0.001 M

(2.0% m.v. + 5 digits)

40.00 M

0.01 M

(3.0% m.v. + 8 digits)

· Overload protection: 300 V DC/AC RMS

Capacitance measurement

Range

Resolution

Accuracy

9.999 nF

0.001 nF

unspecified

99.99 nF

0.01 nF

(4.5% m.v. + 20 digits)

999.9 nF

0.1 nF

9.999 µF

0.001 µF

99.99 µF

0.01 µF

(3.0% m.v. + 5 digits)

999.9 µF

0.1 µF

9.999 mF

0.001 mF

99.99 mF

0.01 mF

(5.0% m.v. + 5 digits)

· Overload protection: 300 V DC/AC RMS

CMP-402 CMP-403 ­ USER MANUAL

63

Frequency measurement ­ current

Range

Resolution

99.99 Hz

0.01 Hz

999.9 Hz

0.1 Hz

· Sensitivity: >20A, >45 Hz

Accuracy (1.0% m.v. + 5 digits)

Frequency measurement ­ voltage

Range

Resolution

Accuracy

99.99 Hz

0.01 Hz

999.9 Hz 9.999 kHz

0.1 Hz 0.001 kHz

(1.0% m.v. + 5 digits)

99.99 kHz

0.01 kHz

· Sensitivity: >2 V RMS · Frequency measured starting from 1 Hz

· Overload protection: 1000 V DC/AC RMS

Duty cycle measurement

Range

Resolution

20.0… 80.0%

0.1%

· Pulse amplitude: 5 V · Pulse width: 0.1 ms…100 ms · Frequency: 45 Hz…10 kHz

Accuracy (1.2% m.v. + 10 digits)

Temperature measurement

Range

Resolution

Accuracy

-20.0…+1000C 0.1 or 1C

± (3% m.v. + 3C)

-4.0…+1832F 0.1 or 1F

± (3% m.v. + 5F)

· The accuracy of the temperature probe is not taken into account · Overload protection: 300 V DC/AC RMS

64

CMP-402 CMP-403 ­ USER MANUAL

11.2 Operating data
a) measurement category according to IEC 61010-1………………………………………….CAT III 600 V (II 1000 V) b) type of insulation ……………………………………………………………………………………………………..double, Class II c) housing type ………………………………………………………………………………………………………… double-composite d) degree of housing protection acc. to EN 60529 …………………………………………………………………………..IP30 e) pollution degree……………………………………………………………………………………………………………………………. 2 f) opening of measurement clamp ……………………………………………………………………………………. 30 mm (1.2″) g) power supply of the meter ………………………………………………………………………………..3 x AAA 1.5 V battery h) diode test …………………………………………………………………………………………………I = 1.0 mA, U0 < 3.0 V DC i) continuity test …………………………………………………………………………………………acoustic signal for R < 50
…………………………………………………………………………………………………………….. measuring current <1.5 mA j) indication for range exceeding ………………………………………………………………………………………….OL symbol
k) low battery indication ………………………………………………………………………………………………….. symbol l) measurement rate ……………………………………………………………………………….. 3 measurements per second m) INRUSH function
sampling time…………………………………………………………………………………… 48 Hz (RMS), 400 kHz (clock) integration time …………………………………………………………………………………………………………………. 100 ms sensitivity………………………………………………………………………………………………………………………….>2 A AC n) VFD function operating voltage ……………………………………………………………………………………………………100…600 V AC o) range of the non-contact voltage detector ………………………………………………. 100…1000 V AC (50/60 Hz) p) response time for PEAK function …………………………………………………………………………………………. <10 ms q) temperature sensor ………………………………………………………………………………………….. type K thermocouple r) input impedance CMP-402 ­ V AC …………………………………………………………………………………………………………….. >9.5 M CMP-402 ­ V DC………………………………………………………………………………………………………………10 M CMP-403 ­ V AC ……………………………………………………………………………………………………………….. >9 M CMP-403 ­ V DC………………………………………………………………………………………………………………10 M s) compatibility with HVDC adapters ………………………………………………………………………………………………. yes t) AC read-out ………………………………………………………………………………………….. True RMS (A AC and V AC) u) AC bandwidth sine waveforms………………………………………………………………………………………………………….50…2000 Hz all waveforms………………………………………………………………………………………………………………..50…60 Hz v) display…………………………………………………….4 digit LCD with backlit, 4000 counts with function indicators w) dimensions……………………………………………………………………………………………………………220 x 80 x 39 mm x) meter weight CMP-402 …………………………………………………………………………………………………………………………….. 266 g CMP-402 (without batteries) …………………………………………………………………………………………………. 230 g CMP-403 …………………………………………………………………………………………………………………………….. 270 g CMP-403 (without batteries) …………………………………………………………………………………………………. 234 g y) operating temperature……………………………………………………………………………………………………… +5..+40C
z) operating humidity…………………………..< 80% for temp. 31C decreasing linearly to 50% at temp. 40C aa) storage temperature……………………………………………………………………………………………………….. -20..+60C
bb) storage humidity…………………………………………………………………………………………………………………….< 80% cc) maximum operating altitude………………………………………………………………………………………………….. 2000 m dd) Auto-Off function …………………………………………………………………………………………………………………. 15 min ee) compliance with the requirements of the following standards …………………………EN 61326-1, EN 61326-2
……………………………………………………………………………IEC 61010-1, EN 61010-02-032, EN 61010-02-033 ………………………………………………….RoHS 2011/65/EU, (EU) 2015/863, EN 62479:2010, EN 50663:2017 ff) quality standard ………………………………………………………………………………………………………………..ISO 9001

CMP-402 CMP-403 ­ USER MANUAL

65

Manufacturer

The provider of guarantee and post-guarantee services is:
SONEL S.A. Wokulskiego 11 58-100 widnica
Poland tel. +48 74 884 10 53 (Customer Service)
e-mail: customerservice@sonel.com web page: www.sonel.com
CAUTION!
Service repairs must be performed only by the manufacturer.

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>