Posey 2791, 2791Q Twice As Tough Cuffs Instructions

July 13, 2024
Posey

Posey 2791, 2791Q Twice As Tough Cuffs

Specifications

  • Product Name: Posey Twice-As-Tough Limb Holder
  • Variants: 2791 D-Ring Cuff (single strap with D-Rings), Large/Red; 2791Q Quick-Release Buckle Cuff (single strap with quick-release buckle), Large/Red

Product Usage Instructions
Applying Limb Restraints

  1. Bring the ends of the strap down through the inside of the side rails.
  2. Attach the strap to the frame that moves with the patient.
  3. Thread the end of the strap over the top, around the frame, and through the D-rings or close the quick-release buckle.
  4. Ensure a secure connection by pulling firmly on the straps.

Applying the Cuffs

  1. Wrap the neoprene piece around the wrist or ankle and secure with hook-and-loop closures.
  2. Check for proper fit by sliding one finger between the cuff and the patient’s wrist/ankle.
  3. Attach the limb strap to the cuff by threading it through the D-rings or closing the quick-release buckle.
  4. Adjust strap length as needed to allow movement without compromising safety.

Adjusting Strap Length
Adjust connecting strap length from the frame to ensure desired freedom of movement without compromising safety. Secure excess strap with an overhand knot below the quick-release buckle and D-ring.

FAQ
Q: Can these limb restraints be used for patients outside of a bed or stretcher?
A: No, these limb restraints are designed for bed and stretcher use only.

Posey® Twice-As-Tough Cuffs 2791, 2791Q Application Instructions for Wrist and Ankle

DESCRIPTION OF PRODUCT
Padded limb restraints with double-security closure. For bed and stretcher use only.

INTENDED USE

  • To protect patients assessed to be at risk of disrupting life-saving treatments (e.g., chronic tube pulling) or in danger of injury to themselves or to others by limiting limb movement in a bed or stretcher.
  • Follow your hospital’s restraint policies and procedures that are compliant with USA CMS guidelines and state laws, or other governing agencies outside the USA.

Posey-2791,-2791Q-Twice-As-ough-Cuffs- \(2\)

CONTRAINDICATIONS
None

APPLICATION INSTRUCTIONS
The order of either application or removal of the straps to the bed OR the cuff to the patient is to be determined by the clinician. Follow these steps to apply device (repeat steps 1-2 on each side):

  1.  Bring the ends of the strap down through the inside of the side rails so they do not interfere when the side rails are raised. Make sure that the straps are attached to the frame that moves with the patient (Fig. 1) (do not attach to side rail or head/footboard), out of the patient’s reach.
  2.  2791: Thread the end of the strap over the top, around the frame, at least once, and through the two D-rings. Bring the strap back over the first ring and through the second D-ring (Fig. 2), or attach with a quick-release tie. To limit unwanted adjustment, tie an overhand knot with the excess strap directly below the D-ring.
    2791Q: Thread the end of the strap over the top, around the frame, at least once, and close the quick-release buckle. Listen for a “snapping” sound. Pull firmly on straps to ensure a good connection (Fig. 3). If buckle does not fit an opening, use the alternate method in Fig. 4 by looping the webbing through the frame opening, inserting the buckle through the loop, and securing the male buckle to the female buckle to secure. Close the quick-release buckle by sliding the male end into the female end (Fig. 4). To limit unwanted adjustment, tie an overhand knot with the excess strap directly below the quick-release buckle.

APPLYING THE CUFFS: (Repeat steps 1-3 for each side)

  1. Wrap the neoprene piece around wrist/ankle so the buckle and connecting strap is on the ulnar side of the wrist or lateral malleolus of the ankle. Attach the black hook-and-loop pieces together and pull back the black pull tab on the red fuzzy piece, followed by the red hook-and-loop pieces. The fuzzy piece should be sandwiched between the two pieces of hook (Fig. 5). Be sure to overlap at least one inch (3 cm).
  2. Press the hook-and-loop closure together firmly and make sure it adheres securely. Slide ONE finger (flat) between the cuff and the inside of the patient’s wrist/ankle to ensure proper fit (Fig. 6).
  3. 2791: Pass the end of the limb strap over the top of the cuff and through the two D-rings on the cuff. Bring the strap back over the first ring and through the second D-ring on the cuff (Fig. 2). Slide one finger flat between the cuff and limb strap.
    2791Q: Close the quick-release buckle on the cuff. Insert ONE finger (flat) under the buckle and pull the strap snug, but not so tight as to restrict circulation (Fig. 7). Release the quick-release buckle, twist buckle 180°, and reconnect (Fig. 8). Listen for a “snapping” sound.

POSEY TWICE-AS-TOUGH LIMB HOLDER

Posey-2791,-2791Q-Twice-As-ough-Cuffs- \(5\)

ADJUSTING STRAP LENGTH:
Adjust the connecting strap(s) length from the frame to allow desired freedom of movement without compromising patient or caregiver safety. To limit unwanted adjustment, tie an overhand knot with the excess strap directly below the quick-release buckle and D-ring. Ensure the remaining strap end(s) are secured and out of the reach of the patient.

  1.  Attach the cuff that is secured to the bottom right side of the frame, that moves with the patient, out of the patient’s reach to the left ankle.
  2. Criss-cross the straps and attach the cuff secured to the bottom left side of the frame, that moves with the patient, out of the patient’s reach to the right ankle.
  3. Adjust connecting straps as necessary

PRECAUTIONS

  1. A clinical assessment and decision are required when used with monitoring lines or if patient has a wound or dislocated/fractured limb.
  2. Check the patient regularly to ensure that circulation is not impaired. Serious injury may occur if the cuffs restricts circulation when the limb holder is applied.

TO REMOVE:
Release the cuff from the patient. Release the anchor strap from the bed frame.
TO LIMIT LOWER LIMB RANGE OF MOTION FOR LEGS (FIG. 9): The clinician applying the strap’s criss-cross application method can determine which is applied first (cuff or strap) and which end of the restraint is criss-crossed.

WARNING: Additional or different body or limb restraints may be needed (visit tidiproducts.com):

  • Some patients may require additional interventions in conjunction with a restraint in order to prevent injury to self or others.
    • If the patient pulls violently against the bed straps.
    • To reduce the risk of the patient getting access to the line/wound/tube site.
    • To prevent the patient from flailing or bucking up and down and causing self-injury.
  • It is a clinical decision to decide when a restraint is no longer deemed safe.

BED SAFETY:
Refer to the Food and Drug Administration (FDA), or other governing agencies outside the USA, for the most recent Hospital Bed Safety Guidelines and the bed manufacturer’s instructions for use.
Report any serious incident related to device use to TIDI Products and the Member State Competent Authority. | Translations of this document in French, Spanish, Dutch, German, Italian, and Portuguese are located at www.tidiproducts.com/IFU

ADDITIONAL SAFETY AND LAUNDERING INSTRUCTIONS ON OTHER SIDE.

  • TIDI Products, LLC • 570 Enterprise Drive, Neenah, WI 54956 USA
    Phone: 1.800.521.1314 • International: +1.920.751.4036 • www.tidiproducts.com

  • © 2023 TIDI Products, LLC. All rights reserved. Posey is a registered trademark of TIDI Products, LLC.

  • MDSS GmbH Schiffgraben 41
    30175 Hannover, Germany

  • UK Responsible Person
    Emergo Consulting (UK) Limited c/o Cr360 – UL International Compass House, Vision Park Histon Cambridge CB249BZ
    United Kingdom

Safety Information for the use of Posey® Restraining Products

  • For RX PRODUCTS ONLY. Federal law (USA) restricts RX devices to sale by  or on order of a physician. For use in a licensed healthcare facility only.
  • STAFF TRAINING: TIDI offers in-service training aids at no charge. Contact TIDI online at www.tidiproducts.com or call toll-free at 1.800.521.1314.
  • BEFORE APPLYING ANY RESTRAINT: Follow your hospital’s restraint policies and procedures that are compliant with USA CMS guidelines and state laws or other governing agencies outside the USA.

ADDITIONAL WARNINGS

  1. ALWAYS monitor patient per facility policy. Improper application or use of any restraint may result in serious injury or death. Be aware that constant monitoring may be required for:
    • Aggressive or agitated patients; and
    • Patients in a supine position; ensure that being in this position would not cause any complications for those who are at risk of vomiting while being restrained.
  2.  NEVER alter or repair this product. ALWAYS Inspect before each use: Check for broken stitches or parts; torn, cut or frayed material; or locks, buckles, or hook-and-loop fasteners that do not hold securely. DO NOT use soiled or damaged products. Doing so may result in serious injury or death. Dispose of damaged products per facility policy for BIOHAZARDOUS material.
  3. ALWAYS secure straps to a part of the frame that moves with the patient (not foot/head board or side rails), out of the patient’s reach, using quick-release ties (see drawing below) or buckles. These allow easy release in the event of an accident or fire.

Test to make sure straps cannot tighten, loosen, or slip and create excess slack. If this occurs, the patient may slide off the chair or bed, increasing the risk of serious injury or suffocation. Restraint release is an important part of facility fire and disaster drills. Straps can be cut with scissors in an emergency.

  1. NEVER secure restraint strap to side rail or head/footboard.
  2. NEVER use Posey products on toilets, or on any chair or furniture that does not allow proper application as directed in the Application Instructions. DO NOT use at home.
  3.  NEVER expose this product to open flame, fire, smoking materials, or high heat sources. Some products may melt or ignite and burn. The facility smoking/no smoking policy should be strictly enforced.
  4. NEVER use a Posey product as a seat belt in a moving vehicle. Posey products are not designed to withstand the force of a crash or sudden stop.
  5. Test Zippers or hook-and-loop fasteners before each use. DISCARD device if it does not fasten securely.

LAUNDERING INSTRUCTIONS (if applicable)

  • Fasten all buckles and locks to reduce risk of damage during wash and dry cycles. DO NOT put buckles or locks through extractors. For maximum life, launder in a laundry bag.

  • Stainless Steel Locks: Posey recommends that locks be treated with a dryfilm lubricant (silicone) after each cleaning. For best results, spray a small amount into the lock. Work the action of the lock with a key several times to ensure that all the surfaces are well lubricated.

  • Before laundering, zip up and turn the product inside out to protect zipper.

  • Hook-and-loop fasteners may collect lint after repeated use or laundering, reducing grip strength. Fasten the “hook” to the “loop” before laundering to help prevent lint buildup. As needed, use a stiff-bristle brush to remove lint from the “hook” side.

  • These products, other than foam products, can be machine washed under CDC guidelines for material soiled with blood or bodily fluid

  • For non-contaminated material, use lower temperature wash and dry cycles to extend product life.

  • For foam products:

  • Make sure products are completely dry before use.

STORAGE AND HANDLING:

  • This device is designed for use in normal indoor environments.
  • This device may be stored in ambient warehouse temperatures at normal humidity levels. Avoid excess moisture or high humidity that may damage product materials.

HOW TO TIE THE POSEY QUICK-RELEASE TIE

Posey-2791,-2791Q-Twice-As-ough-Cuffs- \(1\)

  1. Wrap strap once around the frame that moves with the patient (do not attach to side rail or head/footboard), leaving at least an 8” (20 cm) tail. Fold the loose end in half to create a loop and cross it over the other end.
  2. Insert the folded strap where the straps cross over each other, as if tying a shoelace. Pull on the loop to tighten.
  3.  Fold the loose end in half to create a second loop.
  4. Insert the second loop into the first loop.
  5.  Pull on the loop to tighten. Test to make sure strap is secure and will not slide in any direction.
  6. Repeat on other side. Practice quick-release ties to ensure the knot releases with one pull on the loose end of the strap.

References